Sie sind auf Seite 1von 18

4

RLDE

Right Leg Drive (4)

RLDout
Cable_gnd

Lead_selector (3)

buffer_v1_C

Preamplifier (4)

V1

toRLD

out

V1E

buffer_v2

In

out

V2E

LeadOut+

Lead+
Sat_Test

LeadOut-

LeadtoAmplifier

Amplifier (4)

In
V3E

V2
V3
V4
V5

buffer_v3
In

Out

buffer_v4

All the schematics are contained in the book:


"Medical Instrument
Design and Development",
Becchetti, Neri,
John Wiley & Sons Ltd.

RLDIn

Cable_GND

Low Pass Filter (5)

AmplifierIn
toFilters

FiltersIn
VRefLow

out

V4E
In

Cal

buffer_v5_C
V5E

buffer_v6

In

Control1.C
Control1.B
Control1.A

out
V6E

In

Calibration

Voltage reference

VRefLow

Test
This work is licensed under the
Creative Commons
Attribution-NonCommercialShareAlike 3.0 Unported License

Voltage
reference (3)

RA_ON
LA_ON
LL_ON
CAL_ON

out

Control2.A
Control2.B
Control2.C

V6
RA
LA
LL

toDCelev

RAE
In
out
out

LAE

Controller
ADC
(6)

In
B

Control1.C
Control1.B
Control2.A
Control1.A
Control2.B
Control2.C

buffer_LL
out
In

Circuit_Protection (1)
Buffer (2)

J2

V2E
V3E
V4E
V5E
V6E
Cable_GND
RAE
LAE
LLE
V1E
RLDE

232-RX

232-TX UsbGND
UsbD+
232-RXD UsbDVUsb+

232-TX

RB7
RB6
VPP

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15

PC->USB, Isolation (8),


Power (9)
PC->USB
USB
FB2

PC USB
Connection
120R
600R
600R
120R

FB3
FB4
FB5

UsbShield

4
3
2
1

J1

Shield
case

CAL_ON
RA_ON
LL_ON
LA_ON

Patient Cable DSUB 15 female

ADC Input

In

LLE

to_HP
Sat_Test

PIC (7)

buffer_LA

Test

buffer_RA

FB6

600R

USB-FEMALE-B
4
3
2
1

Title

CG ECG: Main scheme


Size
A
Date:

Document Number

Rev

Copyright Gamma Cardio Soft all rights reserved 30.2


Tuesday, January 21, 2014

Sheet

of

19

J3

Programming Plug
1

Pay Attention, critical path,


minimize length

All the schematics are contained in the book:


"Medical Instrument
Design and Development",
Becchetti, Neri,
John Wiley & Sons Ltd.

VCC

10

AmplifierIn

V+

TLC2274

0.1uF CB1

U1C OUT

11

toFilters

V-

0.1uF CB2

0
VDD
R1
R2
464

22.6K
All resistors are precision type

This work is licensed under the Creative Commons


Attribution-NonCommercial-ShareAlike 3.0 Unported
License.
To view a copy of this license, visit
http://creativecommons.org/licenses/by-nc-sa/3.0/

Title
Amplification Module
Size
A
Date:

Document Number
Copyright Gamma Cardio Soft all rights reserved
Tuesday, January 21, 2014
5

Sheet

Rev
30.0
2

of
4

19
3

Warning, very critical


path for all the buffers
Minimize length
and keep similar length
for all the paths of the
buffer modules

For all the buffers use power


resistance suitable for
defibrillator's discharge

VCC

10

In

V+

U3C

R12

100K
9

Out

All the schematics are contained in the book:


"Medical Instrument
Design and Development",
Becchetti, Neri,
John Wiley & Sons Ltd.

TLC2274

11

Q17
2n3904

V-

OUT
Q18
2n3904

VDD

This work is licensed under the Creative Commons


Attribution-NonCommercial-ShareAlike 3.0 Unported
License.
To view a copy of this license, visit
http://creativecommons.org/licenses/by-nc-sa/3.0/

Warning: all these buffer


ground points must be
connected together and then
they can be connected to the
ground plane in one point only

Title
Buffer

Size
A
Date:

Document Number
Copyright Gamma Cardio Soft all rights reserved
Tuesday, January 21, 2014

Sheet

Rev
30.0
3

of

19

Warning, very critical


path for all the buffers
Minimize length
and keep similar length
for all the paths of the
buffer modules

For all the buffers use power


resistance suitable for
defibrillator's discharge

VCC

In

V+

U3A

R13

100K
2

Out

All the schematics are contained in the book:


"Medical Instrument
Design and Development",
Becchetti, Neri,
John Wiley & Sons Ltd.

TLC2274

11

Q19
2n3904

V-

OUT
Q20
2n3904

VDD

This work is licensed under the Creative Commons


Attribution-NonCommercial-ShareAlike 3.0 Unported
License.
To view a copy of this license, visit
http://creativecommons.org/licenses/by-nc-sa/3.0/

Warning: all these buffer


ground points must be
connected together and then
they can be connected to the
ground plane in one point only

Title
Buffer

Size
A
Date:

Document Number
Copyright Gamma Cardio Soft all rights reserved
Tuesday, January 21, 2014

Sheet

Rev
30.0
4

of

19

Warning, very critical


path for all the buffers
Minimize length
and keep similar length
for all the paths of the
buffer modules

For all the buffers use power


resistance suitable for
defibrillator's discharge

VCC

In

V+

U1B

R11

100K
6

Out

All the schematics are contained in the book:


"Medical Instrument
Design and Development",
Becchetti, Neri,
John Wiley & Sons Ltd.

TLC2274

11

Q15
2n3904

V-

OUT
Q16
2n3904

VDD

This work is licensed under the Creative Commons


Attribution-NonCommercial-ShareAlike 3.0 Unported
License.
To view a copy of this license, visit
http://creativecommons.org/licenses/by-nc-sa/3.0/

Warning: all these buffer


ground points must be
connected together and then
they can be connected to the
ground plane in one point only

Title
Buffer

Size
A
Date:

Document Number
Copyright Gamma Cardio Soft all rights reserved
Tuesday, January 21, 2014

Sheet

Rev
30.0
5

of

19

Warning, very critical


path for all the buffers
Minimize length
and keep similar length
for all the paths of the
buffer modules

For all the buffers use power


resistance suitable for
defibrillator's discharge

VCC

In

V+

U1A

R38

100K
2

Out

All the schematics are contained in the book:


"Medical Instrument
Design and Development",
Becchetti, Neri,
John Wiley & Sons Ltd.

TLC2274

11

Q21
2n3904

V-

OUT
Q22
2n3904

VDD

This work is licensed under the Creative Commons


Attribution-NonCommercial-ShareAlike 3.0 Unported
License.
To view a copy of this license, visit
http://creativecommons.org/licenses/by-nc-sa/3.0/

Warning: all these buffer


ground points must be
connected together and then
they can be connected to the
ground plane in one point only

Title
Buffer

Size
A
Date:

Document Number
Copyright Gamma Cardio Soft all rights reserved
Tuesday, January 21, 2014

Sheet

Rev
30.0
7

of

19

Warning, very critical


path for all the buffers
Minimize length
and keep similar length
for all the paths of the
buffer modules

For all the buffers use power


resistance suitable for
defibrillator's discharge

VCC

12

In

V+

U1D

R4

100K
13

14

Out

All the schematics are contained in the book:


"Medical Instrument
Design and Development",
Becchetti, Neri,
John Wiley & Sons Ltd.

TLC2274

11

Q3
2n3904

V-

OUT
Q4
2n3904

VDD

This work is licensed under the Creative Commons


Attribution-NonCommercial-ShareAlike 3.0 Unported
License.
To view a copy of this license, visit
http://creativecommons.org/licenses/by-nc-sa/3.0/

Warning: all these buffer


ground points must be
connected together and then
they can be connected to the
ground plane in one point only

Title
Buffer

Size
A
Date:

Document Number
Copyright Gamma Cardio Soft all rights reserved
Tuesday, January 21, 2014

Sheet

Rev
30.0
8

of

19

Warning, very critical


path for all the buffers
Minimize length
and keep similar length
for all the paths of the
buffer modules

For all the buffers use power


resistance suitable for
defibrillator's discharge

VCC

In

V+

U3B

R5

100K
6

Out

All the schematics are contained in the book:


"Medical Instrument
Design and Development",
Becchetti, Neri,
John Wiley & Sons Ltd.

TLC2274

11

Q5
2n3904

V-

OUT
Q6
2n3904

VDD

This work is licensed under the Creative Commons


Attribution-NonCommercial-ShareAlike 3.0 Unported
License.
To view a copy of this license, visit
http://creativecommons.org/licenses/by-nc-sa/3.0/

Warning: all these buffer


ground points must be
connected together and then
they can be connected to the
ground plane in one point only

Title
Buffer

Size
A
Date:

Document Number
Copyright Gamma Cardio Soft all rights reserved
Tuesday, January 21, 2014

Sheet

Rev
30.0
9

of

19

Warning, very critical


path for all the buffers
Minimize length
and keep similar length
for all the paths of the
buffer modules

For all the buffers use power


resistance suitable for
defibrillator's discharge

VCC

In

V+

U2B

R10

100K
6

Out

All the schematics are contained in the book:


"Medical Instrument
Design and Development",
Becchetti, Neri,
John Wiley & Sons Ltd.

TLC2274

11

Q13
2n3904

V-

OUT
Q14
2n3904

VDD

This work is licensed under the Creative Commons


Attribution-NonCommercial-ShareAlike 3.0 Unported
License.
To view a copy of this license, visit
http://creativecommons.org/licenses/by-nc-sa/3.0/

Warning: all these buffer


ground points must be
connected together and then
they can be connected to the
ground plane in one point only

Title
Buffer

Size
A
Date:

Document Number
Copyright Gamma Cardio Soft all rights reserved
Tuesday, January 21, 2014

Sheet

Rev
30.0
10

of

19

All the schematics are contained in the book:


"Medical Instrument
Design and Development",
Becchetti, Neri,
John Wiley & Sons Ltd.

This work is licensed under the Creative Commons


Attribution-NonCommercial-ShareAlike 3.0 Unported
License.
To view a copy of this license, visit
http://creativecommons.org/licenses/by-nc-sa/3.0/

VCC
C

Warning: critical trace for all buffers


minimize lenght and assure equal
lenght for all buffers

0.1uF

10

In

100K
9

Out

TLC2274

11

Q1
2n3904

V-

OUT
Q2
2n3904

V+

U2C

R3

C2

C3 0.1uF

VDD

Warning: all ground points to be connected


together and they have to be connected to
ground plane in one point only

Title
Buffer

Size
A
Date:

Document Number
Copyright Gamma Cardio Soft all rights reserved
Tuesday, January 21, 2014

Sheet

Rev
30.0
of

11

19

All the schematics are contained in the book:


"Medical Instrument
Design and Development",
Becchetti, Neri,
John Wiley & Sons Ltd.

This work is licensed under the Creative Commons


Attribution-NonCommercial-ShareAlike 3.0 Unported
License.
To view a copy of this license, visit
http://creativecommons.org/licenses/by-nc-sa/3.0/

VCC
C

Warning: critical trace for all buffers


minimize lenght and assure equal
lenght for all buffers

0.1uF

12

In

100K
13

14

Out

TLC2274

11

Q7
2n3904

V-

OUT
Q8
2n3904

V+

U3D

R6

C42

C43 0.1uF

VDD

Warning: all ground points to be connected


together and they have to be connected to
ground plane in one point only

Title
Buffer

Size
A
Date:

Document Number
Copyright Gamma Cardio Soft all rights reserved
Tuesday, January 21, 2014

Sheet

Rev
30.0
of

12

19

All the schematics are contained in the book:


"Medical Instrument
Design and Development",
Becchetti, Neri,
John Wiley & Sons Ltd.

VCC
R49
VRefLow

51K
R50

D5

124K

Calibration

LT1004CD-1-2
R51

100

This work is licensed under the Creative Commons


Attribution-NonCommercial-ShareAlike 3.0 Unported
License.
To view a copy of this license, visit
http://creativecommons.org/licenses/by-nc-sa/3.0/

0
All resistences are precision type

Title
Calibration
Size
A
Date:
5

Document Number
Copyright Gamma Cardio Soft all rights reserved
Tuesday, January 21, 2014
2

Sheet

Rev
30.0
13

of
1

19

HP filter reset
to_HP
CAL_ON
Test

232-TX
D

12

10

Sat_Test

RB7
RB6

2N7002

14
15

0
C

28
27
RB7
RB6
VPP

VDD

AVDD

MCLR/VPP

M1

A
D4

D3

In circuit
programming
port

Saturation
signal ADC

RA4/TOCKI
OSC1/CLKIN

17
18

20

Lead Selector
control lines

RA3/AN3/VREF
RBO/INT
RB1
RC6/TX
RB2
RC7/RX/DT
U7
RB3
RB4
RC3
RB5
RC4
RC0/T1OSO/T1CKI
RC5
RA0/AN0
PIC16C773

5
21
22
23
24
25
26
11
16
2

Control2.A
Control2.B
Control2.C
Control1.A
Control1.B
Control1.C
RA_ON
LL_ON
LA_ON

Rs 232 port

GND

RA1/AN1
RA2/AN2

OSC2/CLKOUT

3
4

RC2/CCP1

1K

1K
D2
1N4148

In

232-RX

RC1/T1OSI/CCP2

13

Test signals

19

R23

AVSS

ECG ADC

R22

X2
B

R27
27

C15

0.1uF
22PF
C16

R28

4.7K

22PF
C17
VCC

0
Title
Controller
Size
A

Document Number
Copyright Gamma Cardio Soft all rights reserved

Date:

Tuesday, January 21, 2014

Sheet

Rev
30.0
14

of

19

Analog Switch
R29 820
LA

R30

820

LL

R31

820

NC1

14

NC2

11

NC3

NC4

U7

MAX364

COM1
COM2
COM3
COM4

2
15
10
7

V+
VL
GND
V-

13
12
5
4

critical path,
Minimize length
D

1
16
9
8

0
RA

C20
0.1uF

VDD
VDD
VCC

GND

4
2
5
1
12
15
14
13

0.1uF
VCC

LeadOut+
3

MUX OUT
R32

C
B
A
EN

IN7
IN6 U8
IN5
OUT
IN4 MC14051
IN3
IN2
IN1
IN0

820

9
10
11
6

Cal

0.1uF

CAL_ON
RA_ON
LL_ON
LA_ON

16
7
8

C19

VDD
VEE

C18

LeadOut-

C21
0.1uF

All the schematics are contained in the book:


"Medical Instrument
Design and Development",
Becchetti, Neri,
John Wiley & Sons Ltd.

IN1
IN2
IN3
IN4

Critical path,
Minimize length

0
Control1.A
Control1.B
Control1.C

0.1uF

Test

VCC

VDD

C23

EN
C
B
A

V6
V5
V4
V3
V2
V1

IN7
IN6 U9 OUT
IN5
IN4 MC14051
IN3
IN2
IN1
IN0

This work is licensed under the Creative Commons


Attribution-NonCommercial-ShareAlike 3.0 Unported
License.
To view a copy of this license, visit
http://creativecommons.org/licenses/by-nc-sa/3.0/

0
3

MUX OUT
Use precision resistence for all
resistances into this scheme

Title

6
9
10
11

4
2
5
1
12
15
14
13

40.2

C24
0.1uF

VDD
VEE

R34

GND

100K

16
7

R33

Lead selector
Control2.A
Control2.B
Control2.C

Size
A
Date:

Document Number
Copyright Gamma Cardio Soft all rights reserved
Tuesday, January 21, 2014
2

Sheet

Rev
30.0
of

15
1

19

C28

C29

0.082UF

0.15UF

C31
6

OUT

10
5.9K

R42
12.1K

C30

VCC

U11C

0.022uF

VDD

TLC2274

11

0.068uF

R41
7

V+

U11B

V-

FiltersIn

12.1K

5.36K

OUT

TLC2274

11

R40
5

VCC

V+

R39

V-

7.32K

VCC

C32
U11D
OUT 14
0.068uF
13 TLC2274

11

C35
0.15UF

VDD

R46

C33
1.8NF

C34

TLC2274
2 -

R47

A
95.3K

VDD

R45
95.3K

V-

4.02K

243K

U11A

8.2nf

OUT

toDCelev

C36

V+

C37
3

0.1uF

11

12

V-

R44

V+

R43

VDD

0.1uF
C38
0.1uF

All the schematics are contained in the book:


"Medical Instrument
Design and Development",
Becchetti, Neri,
John Wiley & Sons Ltd.

VCC

VRefLow

Title

This work is licensed under the Creative Commons


Attribution-NonCommercial-ShareAlike 3.0 Unported
License.
To view a copy of this license, visit
http://creativecommons.org/licenses/by-nc-sa/3.0/
5

Bessel 8th low pass filter


Size
A
Date:
3

Document Number
Copyright Gamma Cardio Soft all rights reserved
Tuesday, January 21, 2014
2

Sheet

Rev
30.0
16

of

19
1

VUsb+

Title

FB1

PCUsb

Sheet

17

Verrify that CX1 and X2 are 0805 and


non 0603

of

19

C8
10UF

VUSB

R14

-Vout

C13
0.1uF

1.5K

USBDP

DP

7
R19

27

USBDM

USBDP

ISO1
6n138

0.033UF
C41

27

CTS#

22

DTR#

21

DSR#

20

DCD#

19

RI#

18

TXDEN

16

Test

PWREN#

15

RESET#

PWRCTL

14

TXLED#

12

RXLED#

11

SLEEP#

10

3V3OUT

XTIN

X1
close to chip
CX2

31
VUSB
4

VUSB

EESK

EEDATA

VCC

8
R20
270

ISO2
6n138

FT8U232BM

232-TX

R21

3
390

232-RXD

51K
RS31
VUSB

keep 4mm distance between


isolated non-isolated
pins
For 6n138 place both
THT and SMD option

GND
GND2

EECS

9
17

32

AGND

0.01UF

29

C39

3
1.5K

VUSB

R16
820

XTOUT

22PF
UsbShield

25

23

22PF

28

TXD

RTS#

CX1
47pf

RS30
51K

24

U5

All the schematics are contained in the book:


"Medical Instrument
Design and Development",
Becchetti, Neri,
John Wiley & Sons Ltd.

CB4
0.1uF VDD

R18

RXD

USBDM

C14

This work is licensed under the Creative Commons


Attribution-NonCommercial-ShareAlike 3.0 Unported
License.
To view a copy of this license, visit
http://creativecommons.org/licenses/by-nc-sa/3.0/

C12
0.1uF

VUSB

close to the chip

C40
47pf

1uF

3
13
26

R15
27

DM

1uF
C9

CB3
0.1uF

VCC

2
R17

100UF
2 -Vin

C5
0.1uF

VUSB

470

COM

VCC

C4

AVCC

use quartz like Murata Electronics North


America CSTLS6M00G53Z-B0
RESONATOR 6.00MHZ CERAMIC

UsbD-

+Vout

keep 4mm distance between


isolated non-isolated
pins

Ferrite bead type to be chose on required current 100ma.


use component like Panasonic - ECG
EXC-ELSA35 LEAD CORE SINGLE 3.5X5MM AXIAL

C11
0.1uF

use same length


traces

UsbD+

RY-0505D
6

+Vin

C6 U4
C10
0.1uF

VCC
VCCIO
VCC2

Tuesday, January 21, 2014

C7
0.01UF

30

Date:

USB
Power
UsbGND

Rev
5

AVCC

Document Number
Copyright Gamma Cardio Soft all rights reserved

VUSB

RSTOUT

Size
A

All the schematics are contained in the book:


"Medical Instrument
Design and Development",
Becchetti, Neri,
John Wiley & Sons Ltd.

VCC

Warning: critical trace


for all buffers
minimize lenght and
assure equal lenght for
all buffers

VCC
RS1 2K
C25
A
Feed

12

OUT

11

CA

1N4148
DS2

V+
7

All resistors in
this block are
precision type

1N4148
QS1
2n3904

REF

Lead+

GS1
GS2 +

DS1

U10

10

2
15

V-

Lead-

Sat_Test
C

0.1uF
13

RS2
2K

INA114/BB

R36
2.8K
B

R37
2.8K

0C27

VDD

toRLD

C26
0.1uF

toAmplifier
1UF

Warning:critical trace

VDD

Capacitor C26 is critical. A very low


dielectric absorption and very low
leakage ( Mylar /polystyrene dielectrics
types)

This work is licensed under the Creative Commons


Attribution-NonCommercial-ShareAlike 3.0 Unported
License.
To view a copy of this license, visit
http://creativecommons.org/licenses/by-nc-sa/3.0/
5

R35
3.3M

Title
Preamplifier
Size
A
Date:
3

Document Number
Copyright Gamma Cardio Soft all rights reserved
Tuesday, January 21, 2014
2

Sheet

Rev
30.0
18

of
1

19

All the schematics are contained in the book:


"Medical Instrument
Design and Development",
Becchetti, Neri,
John Wiley & Sons Ltd.

C1
R9

100K

1K

0.0022uF

Q10
2n3904

R7
U2A

OUT
V+

Q9
2n3904

V-

VDD

2
100K

TLC2274

11

VCC

14

Q11
2n3904

13
TLC2274

12

RLDIn

VCC

Warning: critical trace


minimize lenght and assure
equal lenght for all buffers

Q12
2n3904

VDD

OUT
V+

Cable_gnd

U2D

V-

R8

11

RLDout

This work is licensed under the Creative Commons


Attribution-NonCommercial-ShareAlike 3.0 Unported
License.
To view a copy of this license, visit
http://creativecommons.org/licenses/by-nc-sa/3.0/

Title
RLD

This resistor (R8 100k)


must resist
at defibrillator's
discharge
5

Size
A
Date:
4

Document Number
Copyright Gamma Cardio Soft all rights reserved
Tuesday, January 21, 2014
2

Sheet

Rev
30.0
19

of
1