Beruflich Dokumente
Kultur Dokumente
gehealthcare.com
Technical
Publication
5116384
Revision 8
Operating Documentation
GE HEALTHCARE
REV 8
5116384
(ZH-CN)
VSTRAHA
(CS)
ADVARSEL
(DA)
WAARSCHUWING
(NL)
WARNING
(EN)
HOIATUS
(ET)
GE HEALTHCARE
VAROITUS
(FI)
ATTENTION
(FR)
WARNUNG
(DE)
(EL)
FIGYELMEZTETS
(HU)
GE HEALTHCARE
AVRUN
(IS)
AVVERTENZA
(IT)
(JA)
(KO)
BRDINJUMS
(LV)
.
,
.
.
,
, .
apkalpes rokasgrmata ir pieejama tikai angu valod.
Ja klienta apkalpes sniedzjam nepiecieama informcija cit valod, nevis
angu, klienta pienkums ir nodroint tulkoanu.
Neveiciet aprkojuma apkalpi bez apkalpes rokasgrmatas izlasanas un
sapraanas.
brdinjuma neievroana var radt elektrisks strvas trieciena, mehnisku
vai citu risku izraistu traumu apkalpes sniedzjam, operatoram vai pacientam.
GE HEALTHCARE
SPJIMAS
(LT)
ADVARSEL
(NO)
OSTRZEENIE
(PL)
ATENO
(PT-BR)
ATENO
(PT-PT)
GE HEALTHCARE
ATENIE
(RO)
!
(RU)
UPOZORENJE
(SR)
UPOZORNENIE
(SK)
ATENCION
(ES)
GE HEALTHCARE
VARNING
(SV)
DIKKAT
(TR)
(BG)
GE HEALTHCARE
REV 8
5116384
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write Damage In Shipment on ALL
copies of the freight or express bill BEFORE delivery is accepted or signed for by a GE representative or hospital receiving
agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event,
within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company
will not pay a claim for damage if an inspection is not requested within this 14 day period.
To file a report:
Fill out a report on http://egems.med.ge.com/edq/home.jsp
Contact the local service coordinator.
GE HEALTHCARE
REV 8
5116384
CAUTION
Danger of explosion if battery is incorrectly replaced. Replace only with the same or equivalent
type recommended by the manufacturer. Discard used batteries according to the manufacturers
instructions.
GE HEALTHCARE
REV 8
5116384
3 (Table/Gantry)
General Description
1-1 to 1-3
Sub-Assembly Description
2-1 to 2-16
1-1
Detector
2-1 to 2-6
3-1 to 3-12
4-1 to 4-12
General Description - I
1-1 to 1-2
General Description - II
2-1 to 2-5
3-1
A-1 to A-4
Blank/Rear cover
Page
REV
Title page
Title page rear
8
blank
a toh
(Revision Information)
A to B
1 (System)
General Description
1-1 to 1-8
Power Distribution
2-1 to 2-10
Scan Operation
3-1 to 3-9
General Description
1-1 to 1-7
Typical Signals
Host Processor
2-1 to 2-5
Connector Boards
3-1 to 3-6
4-1
2 (Operator Console)
Other OC Components
4 (DAS/Detector)
General Description
5 (X-ray Generator)
Appendix
Symbols and Classification
GE HEALTHCARE
REV 8
5116384
REVISION HISTORY
REV
Date
12/29/08
System
OC
Table/Gantry
DAS/Detector
Updated Figure for DAS Signal Connection (Flex Cables & Connectors) according to
PQR13206154. (Sec. 3 & Sec. 4)
X-ray Generator
10/24/07
OC
04/16/07
OC
X-ray Generator
OC
Updated VSPL Board as an option in 5114678-3 & 5114679-2 in Illustration 1-1 Operator Console
Block Diagram (Sec. 1)
09/11/06
05/14/06
System
03/04/06
System
DAS/Detector
07/25/05
OC
10/15/04
Initial Release.
GE HEALTHCARE
REV 1
5116384
SYSTEM
TABLE OF CONTENTS
SECTION
PAGE
1-1
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SYSTEM OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SYSTEM SPECIFICATIONS AND DATA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SYSTEM OPTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-1
1-1
1-4
1-6
2-1
2-1
2-1
2-7
2-9
3-1
2-2
3-1
3-2
3-3
WARM-UP SCANS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
AXIAL SCANS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-2-1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-2-2
Scan Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-2-3
Asymmetric Scan. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-2-4
Twin Cine Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-2-5
Auto mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SCOUT SCANS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-1
3-4
3-4
3-4
3-6
3-8
3-8
3-9
SYSTEM
GE HEALTHCARE
SYSTEM
GE HEALTHCARE
REV 4
5116384
INTRODUCTION
The features of this CT (Computed Tomography) system, one of the CT scanners of this series, include the
following:
Workstation type information processing system,
Solid-state x-ray intensity detector,
Continuous rotation type gantry with slip rings and high frequency coupling.
This CT system is comprised of the following main components (called subsystems):
DAS/Detector
The system may include some of the following customer option equipments:
1-2
Image Camera
SYSTEM OVERVIEW
Illustration 1-1 and Illustration 1-2 shows the system block diagram.
The operator console (OC) controls the entire system, according to the operators operations. The OC sends
instructions to the processor of the TGP board, which then controls the gantry and table subsystems according to
the instructions.
The TGP board processor also passes the OC instructions to the processor of the OGP board which is equipped on
the gantry rotative frame. The OGP board controls the DAS subsystem, the collimator aperture, or positioning lights
according to the passed instructions.
The OGP board processor also passes the instructions from the TGP board to the processor of the x-ray generator
subsystem. The XG processor controls the x-ray generator according to the instructions (originally from the OC).
Reversely, the OC receives status information from the TGP board or other processors (via the TGP board).
1-1
SYSTEM
GE HEALTHCARE
REV 4
5116384
Illustration 1-1
Tube
PDU
High Voltage
JEDI
Auxiliary Unit
SR
AC115
XG Power
AC308-480
Power Unit
Gantry
Rotational
Temperature
Control
Twin Detector
TG/OC
AC115
Twin DAS
Step Motor
HSDCD
Receiver
Slip Ring
HSDCD
Transmitter
DCP2
Axial Drive
AC200
Cover
Boards
Tilt Pot.
Servo Motor
Servo
Amp.
Azimuth Encode
Twin Collimator
Gantry
Stationary
OGP
Board
Step Motor
Step Motor
Driver
Foot SW
Table
Board
Table
Connector
Board
Pump/Value for
Table Elevation
DASIFN
Cradle
Encoder
Speaker
Pump/Value
for Gantry tilt
Aperture Encoder
Step Motor
Power
Supply
Cover Switch
SUB
Board
TGP
Board
Touch Sensors
DBPCI 1
Host Processor
Key B'd
Latch Switch
Pot. For
Horizontal
DBPCI 0
Monitor
SCSI1
SCSI0-00 SCSI0-01
Rear CN HDD
DASM
DVD
OC
Table
1-2
SYSTEM
GE HEALTHCARE
REV 4
5116384
Illustration 1-2
Tube
PDU
High Voltage
JEDI
Auxiliary Unit
SR
AC115
XG Power
AC308-480
Power Unit
Gantry
Rotational
Temperature
Control
Twin Detector
TG/OC
AC115
HSDCD
Transmitter
DCP2
Step Motor
HSDCD
Receiver
Slip Ring
Twin DAS
Axial Drive
AC200
Cover
Boards
Tilt Pot.
Servo Motor
Servo
Amp.
Azimuth Encode
Twin Collimator
Gantry
Stationary
OGP
Board
Step Motor
Step Motor
Driver
Foot SW
Table
Board
Table
Connector
Board
Pump/Value for
Table Elevation
DASIFN
Cradle
Encoder
Speaker
Pump/Value
for Gantry tilt
Aperture Encoder
Step Motor
Power
Supply
Cover Switch
SUB
Board
TGP
Board
Touch Sensors
DBPCI 1
Host Processor
Key B'd
Latch Switch
Pot. For
Horizontal
DBPCI 0
Monitor
SCSI1
SCSI0-00 SCSI0-01
Rear CN HDD
DASM
DVD
OC
Table
1-3
SYSTEM
GE HEALTHCARE
REV 4
1-3
5116384
Table 1-1
Scan Time
Scan/Recon/Cal FOV
Scan/Recon FOV
Cal FOV
P-Head (18)
Small (25)
Head (25)
Small (25)
Body (43)
Large (43)
Table 1-3
Half Scan
No. of Detector
Active Channels
Recon Matrix
972
635
685 X 2
512 X 512
256 X 256
1-4
SYSTEM
GE HEALTHCARE
REV 4
1-3
5116384
KV-MA Stations
KV
MA
120
140
Slice Thickness
1.2*, 2, 4, 6, 10
0.6*, 1, 2, 3, 5, 7, 10
Other Specifications
1-5
SYSTEM
GE HEALTHCARE
REV 4
1-4
5116384
SYSTEM OPTIONS
Table 1-7
Category
Name
Description
Helical scan
90 sec Helical
120 sec Helical
Collimator
Magnetic optical
disk drive (MOD)
5 inch MOD
5 MOD
Monitor
LCD monitor
22 inch CRT monitor
Filming
Laser camera
Network
Ethernet
Advantage Windows
Software
3D Imaging
Denta Scan
SmartRecon
SmartPrep
Navigator
Others
Remote Tilt
Scan/Recon
Storage Device
1-6
SYSTEM
GE HEALTHCARE
REV 1
5116384
2-1
The power distribution unit (PDU) provides powers to all the subsystems.
Two types of PDUs are available: 400VPDU and 200VPDU. Either one of two types is used according to the mains
voltage (main power from the power distribution box of sites). (200VPDUs are used in Japan only.)
See Illustration 2-1 (400VPDU) or 2-2 (200VPDU). These illustrations describe the following:
All the powers except XG Power are output from the system transformer on the PDU.
(200VPDUs generate XG Power by a step-up transformer.)
XG Power powers the x-ray generator (XG) equipped on the gantry rotational block. Powers for x-ray
tube high voltage, x-ray tube anode rotation, XG control circuit boards, etc. are derived from this power.
However, the fan and pump for the x-ray tube are powered by SR115/120.
SR115/120 powers all the components equipped on the gantry rotational block except the x-ray
generator.
TG115 powers all the components equipped on the gantry stationary block except the axial motor. The
gantry tilt pump also is operated by this power. TG115 also powers the table subsystem.
TG200 powers the axial motor which rotates the gantry rotational block.
2-1
SYSTEM
GE HEALTHCARE
REV 1
2-1
5116384
Slip Rings
Mains
AC 380 V
480V
X-ray
Generator
Rotational Block
(Stationary
Block)
SW1
TG200 (AC 200 V)
OC115 (AC 115 V)
400VPDU
BRK2
Axial
Motor
Gantry
Table
CB1
To Internal
and External
components
Noise
Filter
To DC Power
Suppliers
2-2
SYSTEM
GE HEALTHCARE
REV 1
2-1
5116384
Main s
Step-up
Trans
fo rmer
AC 200 V
Sys
tem
Trans
fo rmer
Slip Rings
X-ray
Generat or
Rotational Bloc k
SW1
(Stationary
Block)
200VPDU
BRK2
Axial
Motor
Gantry
Table
CB1
Noise
Filter
To internal
com ponent s
Monitor
CB2
External
com ponent s
2-3
SYSTEM
GE HEALTHCARE
REV 1
2-1
5116384
Illustration 2-3 (400VPDU) or 2-4 (200VPDU) shows the simplified circuit diagram of the PDU.
As shown;
(400VPDU):
The voltage of mains power should be within a range of 380 V 480 V. The system transformer provides
terminals for 380 V, 400 V, ... , 480 V to connect power to the nearest voltage terminal.
(200VPDU):
The system transformer provides a terminal for 200 V to connect the mains power.
The SR115/120 output is not controlled by any relay contacts; the power is always present in the CT
system unless circuit breakers CB2 or CB5 are turned off.
The relay K1 is turned on some delay time later after K11 is turned on.
The resisters connected to the relay K11 suppress rush currents.
Circuit breakers CB1, CB2, ... , CB6 can turn OFF powers, as written below:
-
CB1: XG Power
CB3: TG200
CB4: TG115
CB5: SR115
CB6: OC115
CB7: PS
2-4
SYSTEM
GE HEALTHCARE
REV 1
2-1
5116384
Mains (3 Phase,
380 ~ 480 V)
K1
CB1
CB2
480
XG Power
K11
F1
System
Transformer
CB3
K2
TG200
200
460
440
120
415
115
CB5
SR115/120
400
380
CB4
K2
TG115
CB6
K3
OC115
P-ON
FAN AL
From
OC
E-OFF-O
RST-EM
RMT CNT
Board
K19
K1
K7
K3
K6
K2
K5
K11
SAFE-O
SAFE-G
From
Gantry
RST-EMG
E-OFF-G
2-5
SYSTEM
GE HEALTHCARE
REV 1
2-1
5116384
Mains (3 Phase,
200 V)
K1
XG Power
Step-up
Transformer
CB1
CB2
200
K11
F1
System
Transformer
CB3
K2
TG200
200
120
SR115/120
CB5
115
CB4
K2
TG115
CB6
K3
OC115
P-ON
FAN AL
From
OC
E-OFF-O
RST-EM
RMT CNT
Board
K19
K1
K7
K3
K6
K2
K5
K11
SAFE-O
SAFE-G
From
Gantry
RST-EMG
E-OFF-G
2-6
SYSTEM
GE HEALTHCARE
REV 1
2-1
2-1-1
5116384
The power on/off controls of XG Power, TG200, TG115, etc. shown in Illustration 2-1 or 2-3 (or, 2-2 or 2-4, for
200VPDU) are shown in Illustration 2-5.
In the illustration;
E-OFF turns close or open when either E-OFF-O or E-OFF-G turns close or open.
That Safety Loop turns close or open means the system safety loop is closed or opened, including both SAFEO and SAFE-G turns close or open. (See the Safety Loop (2-2) description)
Reset Emergency turns close or open when either RST-EM or RST-EMG turns close or open.
2-7
SYSTEM
GE HEALTHCARE
REV 1
2-1
5116384
INPUT Control
PON SW
E-OFF
Safety-Loop
RESET Emer
FAN ALA
OUTPUT Control
OC115(K3)
XG400(K1)
Resistor(k11)
Timer1(Smart Gantry)
TG115(K2)
TG200(K2)
Timing2(general)
TG115(K2)
TG200(K2)
Pins 2 and 3 of JP1 are shorted on the RMT board
Pins 2 and 3 of JP2 are shorted on the RMT board
Note
Safety Loop is always closed in normal conditions after the system is switched ON or reset; this
means that XG Power is always supplied to the gantry in normal conditions.
2-8
SYSTEM
GE HEALTHCARE
REV 1
2-2
5116384
SAFETY LOOP
Safety Loop
Switch on the
Gantry Rear Base
TGP
Board
SAFE-G
Gantry
RMT CNT
Board
PDU
Operator Console
Host
Processor
REAR CN1
Board
PCI Bus
DBPCI
Board
SAFE-O
K5
Relays
K1
XG Power
2-9
SYSTEM
GE HEALTHCARE
REV 1
2-2
5116384
When the TGP board receives from the OGP board a safety loop open demand due to an
overtime of x-ray exposure.
When the TGP board detects abnormal communication with the OGP board or the operator
console (host processor) during x-ray exposure.
When the host processor receives a safety loop open demand from the TGP board.
When the host processor detects an extra scan (other than scans which the host processor
instructed the TGP board to perform) performed.
2-10
SYSTEM
GE HEALTHCARE
REV 1
5116384
WARM-UP SCANS
The system requires the Warmup scans to warmup the xray tube prior to starting the first scan after poweron,
or when more than two hours have elapsed since the last scan.
The warmup scans also have to be performed before performing phantom calibration which updates the calibration
files (CAL files), if more than three hours have elapsed since the last scan.
The operator selects either the Tube Warmup or Daily Calibration warmup scans, as listed in Table 3-1.
Table 3-1
Type
Tube Warm-up
Warm-up Scans
Purpose
No. of Scans
13 (Max.)
- Including 8 scans (calibration) + 5 scans (Xray alignment)
- Calibration scans contain 10mm, 7mm, 5mm, 3mm, 2mm,
1mm, 0.6mm, and Asymmetric scans. (0.6mm and Asymmetric
are optional.)
- Using 7 mm scan data, warmup error is analyzed.
6
for system diagnostics
(IQ Performance Measurement Scan)
The Tube Warmup type scans or the first series of scans of Daily Calibration warmup the xray tube.
This means that Tube Warmup only warms the xray tube.
During the scans for calibration (eight scans) the system collects data to analyze it, and update CAL files:
The system stores the data and checks whether the data falls within the specified range. If the data is out of range,
the system reports a WARMUP ERROR. If the data is within range, the system compares the data with the previous warmup data and calculates correction factors for CAL files, and updates them.
During the scans for xray alignment (three scans), the system collects data to analyze the xray alignment.
IIllustration 3-1 shows the warm-up scan sequence.
3-1
SYSTEM
GE HEALTHCARE
REV 1
3-1
5116384
Warm-up Sequence
Start
Tube warm-up
scans
Error Message
Error &
Error code
Yes
Any error?
No
Scans &
Data collection
Data check
Raw data
files
Error Message
Warm-up
Error
No
Within
range?
Yes
Calculates
correction
factors
CAL
files
System Disk
CAL files
correction
End
3-2
SYSTEM
GE HEALTHCARE
REV 1
3-1
5116384
Analysis Items
Descriptions
Specifications
Ref Count
3000 ~ 60000
SD/MEAN
0 ~ 0.03
Ratio
0.95 ~ 1.05
HPF (Ratio)
HPF of Ratio
0.01 ~ 0.01
Aperture Check
0.5 ~ 1.5
When you analyze a warmup calibration file, convert the related vector file into the Mean file, then use the Generic
System Analyzer of the Service tool.
3-3
SYSTEM
GE HEALTHCARE
REV 1
3-2
3-2-1
5116384
AXIAL SCANS
General
Prior to the actual x-ray exposure, the system collects 256 views of offset data generated from the DAS. The offset
data is used to correct actual x-ray data.
The system performs a full 360 deg. scan (clockwise direction only) during 1.0, 1.5, 2.0, 3.0 or 5.0 sec. These are
called scan speeds or scan times, and not all of the them are available to all the systems; the available scan speeds
or times vary according to system models or options installed on the systems.
During a scan, the system collects 972 views of data. A view period differs according to the scan speed. The
following table shows this relation:
Scan Speed (sec)
1.0
1029
1.5
1543
2.0
2058
3.0
3086
5.0
5144
Scan Sequence
3-4
SYSTEM
GE HEALTHCARE
REV 1
3-2
5116384
Parameters
change?
(Yes)
Scan parameters setting
Cradle positioning
Rotor start
Gantry acceleration
High Voltage ON
Das enable
Data collection
(No)
3-5
SYSTEM
GE HEALTHCARE
REV 1
3-2
3-2-3
5116384
Asymmetric Mode
In scanning, you can choose either symmetric or asymmetric slices from the twin-detector.
There are three choices in the Asymmetric Mode.
[OFF] mode: (Symmetric is OFF) acquires slices with the same thickness (symmetric).
3-6
SYSTEM
GE HEALTHCARE
REV 1
3-2
5116384
Asymmetric Slice
Radiation area
Center between
detectors A and B
1mm slice
9mm
Detector B
Detector A
10mm slice
Radiation area
Center between
detectors A and B
2mm slice
8mm
Detector B
Detector A
10mm slice
Radiation area
Center between
detectors A and B
1mm slice
5mm
Detector B
Detector A
6mm slice
3-7
SYSTEM
GE HEALTHCARE
REV 1
3-2
3-2-4
5116384
In Cine Scan, continuous multiple 360-degree data acquisition is achieved without table travel or at the same location.
In Twin Cine Scan, images are displayed in the 1i mode during prospective recon; on the other hand, in retrospective
recon, images from Detector A and images from Detector B can be displayed alternately (2i mode).
The 1i mode display is also possible in retrospective recon.
3-2-5
Auto mA
The Auto mA feature enables the system to reduce mA and patient dose with negligible effect on image quality. The
system varies mA according to patient shape and a predetermined algorithm for scan location.
The Auto mA feature requires one Scout scan, either AP or LAT in the illustration below, to obtain necessary information on X-ray attenuation. (The last scout scan data is used for the Auto mA scan.)
The system automatically increases mA at the scan locations where X-ray attenuation is relatively high and decreases mA at the scan locations where X-ray attenuation is relatively low. In the example below, the mA is set low at the
start scan location (lung area), then the mA will increase at the location of liver area, and then it will decline at the
location of abdomen area.
Illustration 3-4
Auto mA
Patient
Start
End
mA
Scan location
End
Start
3-8
SYSTEM
GE HEALTHCARE
REV 1
3-3
5116384
SCOUT SCANS
Prior to the actual x-ray exposure, the system collects 64 views of offset data generated from the DAS. The offset
data is used to correct actual x-ray data.
The system advances the cradle and collects data from 685 active channels. The cradle speed is 75 mm/sec, and
data collection timing is synchronized with the cradle encoder pulse.
Illustration 3-5 shows the scout scan sequence.
Illustration 3-5
Cradle positioning
Rotor start
Cradle acceleration
High Voltage ON
Das enable
Data collection
Start
position
Cradle
move start
Cradle stop
High
Voltage
ON
DAS
enable
Data collection
Cradle
acceleration
3-9
SYSTEM
GE HEALTHCARE
3-10
SYSTEM
GE HEALTHCARE
REV 7
5116384
OPERATOR CONSOLE
TABLE OF CONTENTS
SECTION
PAGE
1-1
OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
MAIN COMPONENT DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
OPERATIONAL DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-1
1-5
1-7
2-1
2-1
2-2
OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
BUILT-IN I/O (INPUT/OUTPUT) DESCRIPTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-1
2-4
3-1
3-1
3-2
CONNECTOR BOARDS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
REAR CN1 BOARD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-2-1
Filter Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-2-2
Audio Mixer Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-1
3-1
3-1
3-3
4-1
4-1
4-2
4-3
KEYBOARD, MOUSE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DISPLAY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
NAA1 (NP AUDIO AMPLIFIER) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-1
4-1
4-1
OPERATOR CONSOLE
GE HEALTHCARE
ii
OPERATOR CONSOLE
GE HEALTHCARE
REV 7
5116384
OVERVIEW
The information processing system (operator console) is based on the Intel 32 bit architecture.
The host processor of the operator IA console is a Linux computer, which is on the market.
(You connect it to a monitor, keyboard, and mouse to configure a working system.)
Illustration 1-1 shows a block diagram of the operator console of this CT system.
IA Host System Software
IA Host System Software is the standard Linux operating system software for the IA Host that come on the system
disk and on the DVD that you use in the event of a system crash.
PCI Bus
PCI stands for Peripheral Component Interconnect - a bus specification. The PCI bus is a high-performance local
bus used to connect peripherals to memory and a microprocessor. A wide range of vendors make devices that plug
into the PCI bus.
The PCI bus can be operated at its intended high-performance by using DMA transfers.
To perform DMA transfers with the PCI bus, a DMA controller should be equipped on the connected peripheral.
Address Map
For a PCI bus based computer system, memory address mapping and I/O address mapping of PCI devices are
determined during Configuration Cycle which the host processor generates when the system is powered on. Each
of the PCI devices requests its required address space size from the host processor during the cycle.
The PCI devices also (such are DBPCI, others) within this operator console are relocatable in the PCI address
space. This means that address mappings of the PCI devices are not fixed. The physical addresses assigned by
the host processor can be known by reading a set value on a base address register within configuration registers of
each PCI device. (With a VME bus, base addresses are determined by dip switches.)
1-1
OPERATOR CONSOLE
GE HEALTHCARE
REV 7
5116384
Illustration 1-1
HIS/RIS
10/100/1000 Mbps
Analog
RGB
Video Card
SCSI1
DBPCI 0#
Monitor
DBPCI 1#
PS/2
Mouse
PS/2
Host Processor
5" MOD
(Option)
PDU
SCSI0-00
System
Disk
Scan Panel
SCSI0-01
DVD
Laser Camera
Interface (DASM)
SCSI Tower
Raw Data
Disk
103 Keyboard
Keyboard
Service Key
Modem
Scan Room
TGP
NAA1
Gantry Mic
Table Speaker
OC Speaker
DAS
DASIFN 2
Heart Gate
Trigger
Optical
Fiber
1-2
OPERATOR CONSOLE
GE HEALTHCARE
REV 7
5116384
Illustration 1-2
HIS/RIS
10/100/1000 Mbps
Analog
RGB
SCSI1
Video Card
DBPCI 0#
Monitor
PS/2
Host Processor
DBPCI 1#
Mouse
5" MOD
(Option)
Serial Port
PDU
SCSI0-00
System
Disk
Scan Panel
SCSI0-01
DVD
Laser Camera
Interface (DASM)
SCSI Tower
Raw Data
Disk
PS/2
103 Keyboard
Keyboard
Service Key
Modem
Scan Room
TGP
NAA1
Gantry Mic
Table Speaker
OC Speaker
DAS
DASIFN 2
Heart Gate
Trigger
Optical
Fiber
1-3
OPERATOR CONSOLE
GE HEALTHCARE
REV 7
5116384
Illustration 1-3
HIS/RIS
10/100/1000 Mbps
Analog
RGB
SCSI1
Video Card
DBPCI2 0#
Monitor
PS/2
Host Processor
DBPCI2 1#
Mouse
5" MOD
(Option)
Serial Port
PDU
SATA-00
System
Disk
Scan Panel
SATA-01
DVD
Laser Camera
Interface (DASM)
SCSI Tower
Raw Data
Disk
PS/2
103 Keyboard
Keyboard
Service Key
Modem
Scan Room
TGP
NAA1
Gantry Mic
Table Speaker
OC Speaker
DAS
DASIFN 2
Heart Gate
Trigger
Optical
Fiber
1-4
OPERATOR CONSOLE
GE HEALTHCARE
REV 7
1-2
5116384
Host Processor
The host computer performs all of the following operations:
User interface
Image processings/display
Communication
Database control
Raw data/image data flow control
Image storage
Scan/recon control
Recon operation
System Disk
For HP8000/HP8200
SCSI, within the Host Processor
The Host Processor includes two ports. SCSI0-00 is connected to a hard disk drive which is used as the system
disk.
For HP8400
SATA, within the Host Processor
The Host Processor includes two ports. SATA-00 is connected to a hard disk drive which is used as the system disk.
HP8000/HP8200 Capacity: 73 GB
Image Storage: >100000 images
DVD Drive (IDE; Outside the Host Processor by using SCSI Tower)
DVD drive is connected to the SCSI card through an SCSI to IDE Adapter.
Raw Data Disk (SCSI; within the Host Processor)
SCSI0-01 is connected to a hard disk drive which is a high-speed one for storing raw data.
Capacity: 146 GB
Raw Data Storage: >5000 Groups
1-5
OPERATOR CONSOLE
GE HEALTHCARE
REV 7
1-2
5116384
DBPCI (DAS Buffer for PCI Bus) (two pieces for Dual slice, one piece for one slice)
This is a circuit board, serving as a ring buffer whose capacity is 4 MB. This buffer temporarily stores DAS data.
This board also includes control registers for the OC keyboard, safety loop, others.
DASIFN2 (DAS Interface for NP)
This is a circuit board which converts raw data sent from the gantry in optical serial signals into electric parallel
signals. During conversion, this board performs transfer error corrections.
This board also includes one channel for receiving Heart Gate Trigger inputs.
REAR CN1
This is a circuit board which mainly has two functions: Signal filtering and Audio control.
This board is connected to OC external components such are keyboard, TGP board on the gantry, or Power
Distribution Unit (PDU), and is also connected to OC internal components such are DBPCI board, NAA1 board, or
others.
Archive Equipment (MOD)
5 MOD is an option, it is used for archiving images.
Capacity: >2.3 GB
Images: DICOM format
1-6
OPERATOR CONSOLE
GE HEALTHCARE
REV 7
1-3
5116384
OPERATIONAL DESCRIPTION
Monitor
Video Signal
Video Card
System/Image
Disk
Host
Processor
Image Data
PCI Bus
SCSI Bus
Raw Data
Raw Data
Disk
DBPCI # 0/1
Raw Data from DAS
1-7
OPERATOR CONSOLE
GE HEALTHCARE
1-8
OPERATOR CONSOLE
GE HEALTHCARE
REV 7
5116384
OVERVIEW
Controls data transfer between data storage units and memory devices, etc.
Controls the entire scan sequence; it sends necessary instructions to relevant devices, and receives
status information from those devices.
Recon operation.
The host processor includes a DC power supply (>450 watts); it requires AC 115 V power input.
2-1
OPERATOR CONSOLE
GE HEALTHCARE
REV 7
2-1
5116384
OVERVIEW (CONTINUED)
Table 2-1 and 2-2 shows main specifications of the host processor.
Table 2-1
Central Processor
Specification
Model
Operating Frequency
Cache Memory
Other
Main Memory
Capacity (standard)
2 GB (512 MB X 4)
Type
Capacity
DVD
Type
Graphics Card
Resolution
Built-in Graphics
Built-in I/O
(Input/Output)
Built-in Audio
Operating System
2-2
OPERATOR CONSOLE
GE HEALTHCARE
REV 7
5116384
Table 2-2
Central Processor
Specification
Model
Operating Frequency
Cache Memory
Other
Main Memory
Capacity (standard)
Type
Capacity
Capacity
146GB(10K rpm)SCSI
Graphics Card
Resolution
Built-in I/O
(Input/Output)
Operating System
Table 2-3
Central Processor
Specification
Model
Operating Frequency
Main Memory
Cache Memory
Other
Capacity (standard)
Type
2 GB (1GB X 2)
DDR-2 667MHz, with ECC
Capacity
Capacity
Graphics Card
Resolution
Built-in I/O
(Input/Output)
Operating System
2-3
OPERATOR CONSOLE
GE HEALTHCARE
REV 7
2-2
5116384
2-4
OPERATOR CONSOLE
GE HEALTHCARE
REV 7
5116384
Built-in Audio
Audio in and Audio out is connected to RCN1 board.
PCI Bus Slot (HPxw8000 only)
The host processor is equipped with 5 PCI slot.
PCI
(32bit/33M)
DBPCI #1
PCI
(32bit/33M)
DBPCI #0
PCI-x
(64bit/133M)
empty
PCI-x
(64bit/100M)
PCI-x
(64bit/100M)
empty
PCI
DBPCI#1
Slot2
PCI-Express X16
Graphic card
Slot3
PCI
DBPCI# 0
Slot4
PCI-Express X4
empty
Slot5
PCI-X133
empty
Slot6
PCI-X100
Slot7
PCI-X100
empty
PCI
Slot2
PCI-Express X16
Graphic card
Slot3
PCI-Express X18
empty
Slot4
PCI-Express X16
empty
Slot5
PCI-X133
SCSI Card
Slot6
PCI-X100
DBPCI 2
Slot7
PCI-X100
DBPCI 2
2-5
OPERATOR CONSOLE
GE HEALTHCARE
2-6
OPERATOR CONSOLE
GE HEALTHCARE
REV 1
5116384
CONNECTOR BOARDS
All the signal cables connected between the operator console (OC) and subsystems/components outside the OC
run through these connector boards.
The connector boards mainly include filters for each signal line to improve the EMC (Electro-Magnetic Compatibility)
performance of the OC. The REAR CN1 board also includes an audio mixer, relay for the safety loop, a video
amplifier for slave monitors, and others.
3-2
Illustration 3-1 (or Illustration 3-2) shows a block diagram of the REAR CN1 board.
The REAR CN1 board is connected to a slot of the connector box of the OC, and as shown in Illustration 3-1 (or
Illustration 3-2) , the board is connected to OC external components such are the keyboard, TGP board on the
gantry, or the Power Distribution Unit (PDU), and is connected to OC internal components such are DBPCI board,
NAA1 board, or others.
The board functionally consists of the following two blocks:
3-2-1
Filter Block
Filter Block
As shown in Illustration 1-1, this block includes the following interface circuits:
Keyboard Interface:
Mainly interfaces between the keyboard (including the scan buttons) and the host processor.
TGP Interface:
Mainly interfaces RS232/RS422 communications and Direct Lines between the TGP board located on
the gantry and the host processor or keyboard.
PDU Interface:
Mainly interfaces Power On/Off, Emergency, Safety Loop, or other signals between the Power
Distribution Unit (PDU) and the keyboard or the FRONT PNL board of the OC.
3-1
OPERATOR CONSOLE
GE HEALTHCARE
REV 1
3-2
5116384
REAR CN1
RS232
RS232
Scan
Panel
Keyboard
RS232
KB R eset
Emergenc y Stop/Rest
Direc t Line
PS/2
Gantry
from/to Patient,
Auto Voice, Foot
SW, X-ray On
TGP Bo ard
Direc t Line
TGP Reset
X-ray On
PCI Bus
Filter Block
Talk On
Fan Alarm
Safety Loop
Power On
Ther mal
Guard
Power On
Shut down
Front
Panel
PCI Bus
DBPCI
Board
DBPCI
Board
X-ray On
Al ert On
CSA/WRA
ACKA
Addres s (4:0)
Data (7:0)
from Patient
to Patient
PDU
Fan Alarm
KB R eset
Safety Loop
Shut dow n
TGP Reset
X-ray LED
NAA2
Emergency Stop/Rest
Host
Processor
PS/2
RS422
Speak er on T abl e
Audi o Mix
er Bloc k
Line OUT
Line IN
External
Speak er
(opt ion)
CD Line OUT
from Patient
DVD Drive
from Patient
3-2
NAA1
OPERATOR CONSOLE
GE HEALTHCARE
REV 1
3-2
3-2-2
5116384
The audio mixer block of the REAR CN1 board Includes registers for audio function, and also includes an audio
amplifier for headphones or external speakers (which are not the OC speaker which is driven by the NAA1 board).
See Illustration 3-2.
As shown, audio signals from several sources are mixed in this block.
Speaker on the OC
Audio input on the host processor (for recording Auto Voice messages)
3-3
OPERATOR CONSOLE
GE HEALTHCARE
REV 1
3-2
5116384
An intercom function is provided between the OC room and the gantry room. The Talk On button on
the keyboard switches communication direction, as described below:
-
While Talk On is pressed, communication of OC room gantry room is enabled; and both of
Auto Voice (both in the OC room and the gantry room) and sound from the CD-ROM drive (both
in the OC room and the gantry room) are disabled.
The X-ray On sound and alert sounds in the OC room are always enabled, even when Talk On is
pressed.
Auto Voice messages can be recorded by using the microphone equipped on the keyboard while
pressing the Talk On button.
While Auto Voice is being played, communication of gantry room OC room can be disabled. This
prevents Auto Voice sound degradation on the OC side, because: Auto Voice is played both in the OC
room and the gantry room. If communication of gantry room OC room is enabled, the Auto Voice
messages are doubled on the OC side, which degrades the sounds.
Volume Control
The following two kinds of volume control are provided.
The following can be adjusted with the manual volumes on the keyboard:
-
Sound volume from the CD-ROM drive both in the gantry room and the OC room
3-4
OPERATOR CONSOLE
GE HEALTHCARE
REV 1
3-2
5116384
from/to DBPCI
Analog Switch
OP. Amp.
X-ray On/Off
Alert On
X-ray
Sound
Address (4:0)
Programmable
Timer
Modu
lation
Data (7:0)
Frequency &
Alert pulse
width controller
Alert Sound
(one shot)
Electronic
Volume
Sound Volume
Control Register
to Electronic
volume control
from CD Drive
from Host Processor
Line OUT
from Gantry Mic
from OC Mic
to Patient Volume
Lowpass
Filter
Scan Keyboard
Volumes
to Host Processor
Line IN
3-5
to Gantry
Speaker
to OC Speaker
OPERATOR CONSOLE
GE HEALTHCARE
REV 1
3-2
5116384
D7
$b4
$b0
CD GA Volume Register
$ac
CD OC Volume Register
$a8
AV OC Volume Register
$a4
$a0
$90
PIT0 Register
$80
$60
IO Register Base
AUX0
Auto Voice Playback Register
$54
$50
3-6
Note:
CD: CD-ROM Drive
GA: Gantry
OC: Operator Console
AV: Auto Voice
PIT: Sound Pulse Width Control
OPERATOR CONSOLE
GE HEALTHCARE
REV 1
5116384
KEYBOARD, MOUSE
The keyboard used for this operator console integrates the following components:
The keyboard and the TGP board are connected by Direct Lines for the remote cradle and gantry tilt control.
4-2
4-3
DISPLAY
CRT (22 color monitor): The resolution used for the CRT Display is 1280 x 1024 matrix. The refresh
rate used is 75 Hz.
LCD (18): The resolution used for LCD Displayis 1280x1024 matrix. The refresh rate used is 60-75 Hz.
This is an amplifier board for the OC speaker. See the following block diagram.
Illustration 4-1
+
+
OUT
OUT
CN2
-
CN1
To OC Speaker
Muting
Voltage Regulator
4-1
OPERATOR CONSOLE
GE HEALTHCARE
blank
4-2
OPERATOR CONSOLE
GE HEALTHCARE
REV 1
5116384
TABLE/GANTRY
TABLE OF CONTENTS
SECTION
PAGE
1-1
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
COMPONENTS INTERCONNECTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-1
1-2
1-3
2-1
2-1
2-2
2-3
2-4
2-5
OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
STATIONARY GANTRY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-2-1
TGP board Service Switches. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-2-2
TGP Processors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ROTATIONARY GANTRY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SLIP RING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-4-1
MECHANISM. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TABLE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-1
2-2
2-9
2-10
2-12
2-14
2-14
2-15
TABLE/GANTRY
GE HEALTHCARE
ii
TABLE/GANTRY
GE HEALTHCARE
REV 1
5116384
INTRODUCTION
The TGP (Table Gantry Processor) board is the main controller of the Table/Gantry subsystem. The TGP board
receives command from the Operator Console and sends control commands to Table, Stationary Gantry and
Rotating Gantry including OGP (On Gantry Processor), DAS and X-ray Generator. Although the DAS/Detector and
X-ray Generator are in the Gantry, they are not explained in this subsystem.
The Table/Gantry consists of the following components:
TGP board
OGP board
SUB board
Table board
Gantry Display
Gantry Switch
Collimator
Positioning Light
Touch Sensors
1-1
TABLE/GANTRY
GE HEALTHCARE
REV 1
1-2
5116384
COMPONENTS INTERCONNECTION
Table/Gantry
GANTRY
SWITCH
PANEL
GANTRY DISPLAY
TILT POTENTIOMETER
TABLE HEIGHT
POTENTIOMETER
CRADLE POSITION
POTENTIOMETER
CRADLE ENCODER
AZIMUTH ENCODER
AXIAL MOTOR
TGP BOARD
OGP BOARD
SUB BOARD
SERVO AMP/MOTOR
STEPPING MOTOR
DRIVER
TABLE BOARD
CRASDLE DRIVER
TILT HYDRAULIC
SYSTEM
APERTURE
MOTOR
CRADLE
MOTOR
TABLE VALVE
SOLENOID
TABLE HY
DRAULIC SYSTEM
GANTRY PULSE
LOOK GUIDE
APERTURE SENSORS
EMERGENCY SWITCHES
FOOT SWITCHES
1-2
TABLE/GANTRY
GE HEALTHCARE
REV 1
1-3
5116384
SPECIFICATIONS
Tilt Angle
Tilt Speed
Table Height
400 ~ 900 mm
UP/Down Speed
Cradle Speed
Slow 20 mm/sec
Fast 100 mm/sec
Scout 75 mm/sec
Home 100 mm/sec
1-3
TABLE/GANTRY
GE HEALTHCARE
1-4
TABLE/GANTRY
GE HEALTHCARE
REV 1
5116384
OVERVIEW
The Table/Gantry consists of table and Gantry. The Gantry is divided into Stationary Gantry and Rotationary Gantry.
The communication between stationary and rotationary Gantry is made by slip-ring, a rotation mechanism that
permits electrical power and signals exchange.
The illustration 2-1 shows the Table/Gantry block diagram.
Illustration 2-1
GANTRY
PDU
STATIONARY GANTRY
ROTATIONARY GANTRY
POWER
GANTRY
DISPLAY
OC
TGP BOARD
GANTRY
SWITCH
COLLIMATOR
OGP BOARD
POSITIONING
LIGHT
GANTRY
TILT
VALVE
SUB BOARD
SLIP
RING
DAS
SERVO
MOTOR
SERVO AMP
TABLE
XG
CRADLE
IN/OUT
TABLE
BOARD
MAT
SWITCH
TOUCH
SENSORS
TABLE ELEVATION /
GANTRY TILT PUMP
AND TABLE VALVE
2-1
TABLE/GANTRY
GE HEALTHCARE
REV 1
2-2
5116384
STATIONARY GANTRY
The stationary Gantry is mainly controlled by the TGP (Table Gantry Processor) board. The Illustration 2-2 shows
the main components of the stationary Gantry.
The TGP board communicates with Operator Console and Table via cables and with the OGP board located in
rotationary Gantry via cable and slipring.
Illustration 2-2
OGP
(SLIP RING)
MICROPHONE
GANTRY SWITCH
115VAC
PS1
DISPLAY BOARD
FCV
BOARD
PS2
TOUCH SENSOR
BREATH NAVI
EMERGENCY SWITCH
PS 3
OC
MICROPHONE
TGP BOARD
GANTRY SWITCH
TABLE
XDISP BOARD
TOUCH SENSOR
RCV
BOARD
BREATH NAVI
EMERGENCY SWITCH
GANTRY TILT
PUMP & VALVE
TABLE ELEVATION
PUMP & VALVE
SIDE COVER
SWITCHES
SUB BOARD
AXIAL
MOTOR
M
FRONT AND
REAR
COVER
SWITCHES
2-2
TABLE/GANTRY
GE HEALTHCARE
REV 1
2-2
5116384
+24V,+5V,GND,SHIELD,XLED,PLLED,
A0,A1,B0,B1,B2,B3,SL0,SL1,SL2
TGP BD
ASSY
(CN8)
+24V,+5V,SHIELD,FLED,SLED
CUNTL0,CUNTL1,
CUNTL2,CUNTL3,
CUNTH0,CUNTH1,
CUNTH2,CUNTH3
7seg DECORD
& DRIVE
CIRCUIT
GNTDISP
BD ASSY
CN5
LG
La,Lb,Lc,Ld,Le,Lf,Lg,
Ha,Hb,Hc,Hd,He,Hf,Hg
SHIELD,FWDLED,BWDLED,EXLMLED,IMSLED
+5V
SWENBL*
CN2
DISP
CN3
LSW
LSW BD
ASSY
+5VSW
K1
FLG BD
ASSY
CN4
RSW
EMERGENCY SW
INR,OUTR,FAST,IMSR,INLMR,EXLMR,RANGE,
UPR,DNR,FWDR,BWDR,POSL,PRACTICE
EMERGENCY
TCH*
NAA2
ASSY
(CN2)
CN6
TSW
TSNS,CONECTOR
TOUCH SW
DETECT
CIRCUIT
CN8
MICOUT
TOUCH
SW ASSY
CN7
MIC
MIC+,MIC-,MICGND
2-3
MIC
ASSY
TABLE/GANTRY
GE HEALTHCARE
REV 1
2-2
5116384
+24V,+5V,GND,SHIELD,XLED,PLLED,
A0,A1,B0,B1,B2,B3,SL0,SL1,SL2
CN2
DISP
CN8
XDISP
TGP BD
ASSY
(CN9)
+24V,XLED,PLLED
XDISP
BD ASSY
+24V,+5V,SHIELD,FLED,SLED
CUNTL0,CUNTL1,
CUNTL2,CUNTL3,
CUNTH0,CUNTH1,
CUNTH2,CUNTH3
7seg DECORD
& DRIVE
CIRCUIT
CN5
LG
La,Lb,Lc,Ld,Le,Lf,Lg,
Ha,Hb,Hc,Hd,He,Hf,Hg
SHIELD,FWDLED,BWDLED,EXLMLED,IMSLED
RLG BD
ASSY
CN3
RSW
+5VSW
TEMP+5V
+5V
SWENBL*
K2
K1
CN4
LSW
RENBL
POSL,INLMR,RANGE,PRACTICE (+5VSW)
INR,OUTR,FAST,EXLMR,UPR,DNR,FWDR,BWDR (TEMP+5V)
EMERGENCY
TCH*
NAA2
ASSY
(CN3)
TOUCH SW
DETECT
CIRCUIT
CN9
MICOUT
CN10
E-OFF
SW
CN6
TSW
TSNS,CONECTOR
TOUCH
SW ASSY
CN7
MIC
MIC+,MIC-,MICGND
2-4
MIC
ASSY
TABLE/GANTRY
GE HEALTHCARE
REV 1
2-2
5116384
2-5
TABLE/GANTRY
GE HEALTHCARE
REV 1
TABLE DOWN
VALVE
GANTRY TILT(BWD)
VALVE
Illustration 2-5
2-2
5116384
2-6
TABLE/GANTRY
GE HEALTHCARE
REV 1
2-2
5116384
TGP BOARD
There are three micro processors on the TGP Board. One controls the GANTRY Rotation, one controls the Table
and the other controls the arbitration of commands from the gantry processor, table processor, scan processor
(located on the OGP board) and the OC. See Illustration 2-6.
The Management processor communicates with the OC through serial links. The other processors (Gantry
Processor and Table Processor) communicate with the Management Processor, and can receive commands from
the OC via the Management Processor. The Gantry Processor and Table Processor also contain a serial
communication port which is not used for communication between OC or other Processors, but rather is used for
Data correction from the sensor switches.
The TGP Board issues commands to and receives status from the components it controls. The TGP oversees the
following functions:
Error Detection
2-7
TABLE/GANTRY
GE HEALTHCARE
REV 1
2-2
5116384
TGP Bd
Emergency
PDU
PDU
Safety
Loop
Rot Safe
Gantry
Processor
(uPD78310)
G-Pulse
Rot Count
SRAM
Test 0,1
Reset SW
Dual
Port RAM
Reset
Patarn
OC
OC Comm
unication
OC
OC Direct
Line
G Test
SW
Management
Processor
(uPD78310)
SRAM
Flash
Memory
Flash
Memory
Dual
Port RAM
Rotate
Control
Flash
Memory
Status
LED
OC
12bit
DAC
Azimuth
Servo
AMP
M
Enc
Slice Count
DAS
Trig
Slip
Ring
9600bps
SIO
Slip
Ring
CT Lock
&Alarm
Control
Breath
Navi
Control
Relay
Breath
Navi
Service
SW
M Test
SW
Flash
Add Cont
Status
LED
Table
Processor
(uPD78310)
HIT Pot
TILT Pot
SRAM
Flash
Memory
T Test
SW
Status
LED
12bit
ADC
Cradle Pot
Cradle
Motor
Driver
Cradle
Control
M
Enc
Table
Control
Up/Dwn
Relay
Tilt
Control
Tilt
Relay
P,V
P,V
Display
Control
Gantry
Display
T/G SW
Port
Gantry
SW
Table
SW
Flash
Add Cont
Mat SW
Audio
Amp
Fluoro
SW
Mic
Spk
2-8
TABLE/GANTRY
GE HEALTHCARE
REV 1
2-2-1
5116384
TGP board Service Switches
There are five service switches at right upper side of the TGP board which are used to control the Gantry Rotation
manually.
TGP Board Service Switches
MNL
SPD0
SLOW
HOME
SVE
FIX
CNT
FAST
SPD2
MID
ABT
OFF
SYS
SPD1
Illustration 2-7
90DEG
TGP BOARD
SYS : System Position. When the switch is in this position, the Gantry is controlled by the
Operator Console.
OFF
SYS
MNL
Disable the Gantry rotation. Servo Amplifier is disabled and Static brake activated.
MNL
Manual Position. When the switch is in this position, the Gantry rotation is controlled
manually, by the other switches described below.
ABT : Arbitrary Mode. This switch is a spring loaded switch that enables Gantry rotation
toward the direction selected by the rotation Direction switch (CNT/HOME/90DEG),
by the fixed speed of XXX seconds per revolution.
SVE
ABT
FIX
OFF
SVE : Servo Enable. Servo Amplifier is enabled and voltage command is 0 volts, that
means there is no rotation at this position.
FIX
2-9
TABLE/GANTRY
GE HEALTHCARE
REV 1
5116384
2-2-1
SPD2
and
SPD0
SLOW
CNT
SPD2/SPD1/SPD0
SPD2
SPD2
SPD2
SPD1
SPD1
SPD1
SPD0
SPD0
SPD0
FAST/MID/SLOW
FAST
MID
SLOW
FAST
MID
SLOW
FAST
MID
SLOW
SPEED (sec/rot)
1.0
1.0
1.0
1.5
10
15
HOME
CNT
: Clockwise rotation
90DEG
2-2-2
MID
SPD1
FAST
TGP Processors
Tilt position, table height and cradle position adjustment. There is a ROM connected to management
processor that contains adjustment data. The management processor reads this data, compares with
the current positions and make adjustment as needed.
2-10
TABLE/GANTRY
GE HEALTHCARE
REV 1
2-2-2
5116384
Gantry rotation
Cradle In/Out
Gantry Tilt
Table Up/Down
Gantry Display
Gantry Switch
Interlock
2-11
TABLE/GANTRY
GE HEALTHCARE
REV 1
2-3
5116384
ROTATIONARY GANTRY
The Illustration 2-8 shows the main components of the rotationary Gantry.
The functions of the parts of the stationary Gantry are as follows:
OGP Board
The OGP (On Gantry Processor) performs the following functions:
Collimator control
Scan control
The automatic shut-off of the Tube Fan/Pump Power Supply after 30 minutes when the OC has been
powered down.
Illustration 2-8
PHOTO
SENSOR
DAS
MOTOR
DRIVER
MOTOR
MOTOR
DRIVER
MOTOR
GPLS
TO TGP
(SLIP
RING)
COLLIMATOR
ENOCODER
OGP BOARD
POSITIONING
LIGHT
PHOTOSENSOR
RELAY
SWITCHING
POWER
SUPPLY
TEMP
CONT
ASSY
115VAC
TO DAS BUFFER
RF XMT
DCP2
DAS/DETECTOR
2-12
TABLE/GANTRY
GE HEALTHCARE
REV 1
2-3
5116384
2-13
TABLE/GANTRY
GE HEALTHCARE
REV 1
2-4
5116384
SLIP RING
2-4-1
DAS Data
DAS Trigger
Safety Loop
MECHANISM
Every slip ring has a corresponding terminator located on the opposite side of the transmitter terminal. See
illustration 2-9.
TERMINATE
RESISTOR
SIGNAL
SLIP
RING
Illustration 2-9
2-14
TABLE/GANTRY
GE HEALTHCARE
REV 1
2-5
5116384
TABLE
Table
HEIGHT
POTENTIOMETER
PUMP
FROM
SUB
BOARD
VALVE
TABLE UP/DOWN &
GANTRY TILT
HYDRAULIC
SYSTEM
MAT SWITCHES
FROM
TGP
BOARD
TABLE
CONN
BOARD
TABLE
BOARD
LATCH SWITCHES
TOUCH SENSORS
SPEAKER
E
FROM
GANTRY
NF2
STEP
MOTOR
(CRADLE)
SW
P.S.
STEP MOTOR
DRIVER
2-15
ENCODER
M
ENCODER
POTENTIOMETER
TABLE/GANTRY
GE HEALTHCARE
REV 1
2-5
5116384
TABLE (CONTINUED)
Hydraulic System
Consists of pump, hose and cylinder. it is used for Table Up/Down and Gantry Tilt.
Cradle
Portion of the Table that executes longitudinal movement. It is controlled by step motor and motor driver.
Latch Switches
Positioned at both right and left sides of the Table, these switches release or latch the cradle. When both switches
are pressed, the cradle is latched. When one of the switches is depressed, the cradle is released. They are for safety
purpose.
Potentiometers
There are three potentiometers: Height potentiometer and Encoder (tilt) potentiometer. They are used for feedback
purpose.
Touch Sensors
Located at under surface of the Table. They are for safety purpose.
Mat Switches
Located at both right and left sides of the table base. The Gantry panel switch can be available with the mat switch
ON.
2-16
TABLE/GANTRY
GE HEALTHCARE
REV 3
5116384
DAS/DETECTOR
TABLE OF CONTENTS
SECTION
PAGE
1-1
OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-1
SECTION 2 - DETECTOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-1
2-1
2-2
2-3
2-4
2-5
2-6
2-7
2-8
OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HILIGHT DETECTOR MODULE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CHANNEL DISTRIBUTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DAS/DETECTOR CONNECTOR. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
QCAL CHANNELS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
AFTER GLOW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HEATER AND TEMPERATURE CONTROL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-1
2-1
2-2
2-3
2-5
2-6
2-6
2-7
3-1
3-1
3-2
3-3
OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DATA FLOW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DCP2 BOARD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-3-1
DCP2 Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-3-2
DCP2 Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CONVERTER BOARD. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-1
3-5
3-6
3-6
3-7
3-11
4-1
3-4
4-1
4-2
4-3
4-4
OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DATA FLOW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DCP CARD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-3-1
DCP Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-3-2
DCP Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CONVERTER BOARD. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-1
4-5
4-6
4-6
4-7
4-10
DAS/DETECTOR
GE HEALTHCARE
ii
DAS/DETECTOR
GE HEALTHCARE
REV 1
5116384
OVERVIEW
DAS/Detector subsystem consists of Solid State Detector and Data Acquisition System (DAS) that are located on
the rotationary Gantry. X-ray data acquired by the detector is converted to light, then to electrical signal in the
detector and then sent to DAS. The DAS digitizes, serializes and performs offset corrections on the signal and sends
to the Operator Console for image reconstruction.
The X-ray exposure through the scanning object is detected by these modules, converted to light, then converted
to electrical signal and sent to the DAS. The DAS digitizes and sends this data by slipring and fiber optics to the
Operator Console, where it is processed for image reconstruction.
Illustration 1-1
OBJECT
X-RAY
X-RAY
TUBE
DATA
ACQUISITION
SYSTEM
DETECTOR
Illustration 1-2
DAS Data
DATA
PROCESSING
UNIT
X-RAY TUBE
X-RAY BEAM
DETECTOR
DAS
1-1
DAS/DETECTOR
GE HEALTHCARE
1-2
DAS/DETECTOR
GE HEALTHCARE
REV 3
5116384
SECTION 2 - DETECTOR
2-1
OVERVIEW
The detector consists of housing, collimator and detector modules. The detector module is made of scintillator
(lumex crystal), photo-diode and connector. When the X-ray exposure is performed, the X-ray enters through the
collimator and hits the scintillator. The scintillator is made of a material that emits light when hit by X-ray. The
scintillator is in contact with photo-diode so that the light generated by X-ray is converted into electrical current. The
generated current is sent to the DAS (Data Acquisition System) to be processed.
2-2
SPECIFICATIONS
Channel Distribution:
MDAS
GDAS
Slice
A Slice
B Slice
A Slice
B Slice
Active Channel
685
685
685
685
Reference Channel
9+9
9+9
3+3
3+3
QCAL Channel
2-1
DAS/DETECTOR
GE HEALTHCARE
REV 3
2-3
5116384
The module consists of Scintillator cells, Collimator, Photo Diode, Reflex Films and Connector (illustration 2-1). The
X-ray enters through the collimator (tungsten plates) and is absorbed by the Scintillator. The Scintillator is made of
Lumex Crystal, the material that emits light when absorbs X-ray radiation. The Scintillator is optically coupled with
Photo Diode which converts light to electrical current. The Photo Diode is electrically connected to DAS through
the connector.
Illustration 2-1
DETECTOR MODULE
HILIGHT MODULE
LUMEX CRYSTAL
LIGHT
PHOTO DIODE
Slice B
Slice A
2-2
DAS/DETECTOR
GE HEALTHCARE
REV 3
2-4
5116384
CHANNEL DISTRIBUTION
MDAS
J01
J05
36~48
CH
49~51 52~61
CH
CH
62~64
CH
J44
J45
J46
J47
65~640
CH
GND
13~19
GND
3
3:1
13+
4CH 3CH
2:1
3:1
J51
2:1
13+
2:1
1:1
10
576CH
1:1
16
12+
4 2+
14
2:1
+10
3:1
3:1
3:1
1 MOD
J08
1 MOD
3:1
J07
1 MOD
DAS
1:1
3 3
1 MOD
4~12
REF
1~3CH
QCAL
3CH
1 MOD
1 MOD
Detector
1 MOD
J06
1 MOD
3:1
3:1
A Slice
1:1
3:1
3:1
2:1
3:1
2:1
2:1
1:1
1:1
2:1
3:1
3:1
3:1
3:1
3:1
B Slice
CH:
576
14
3:1
Three detector channels are connected in parallel into one DAS channel.
DAS channel current is the sum of the three detector channel current.
2:1
Two detector channels are connected in parallel into one DAS channel.
DAS channel current is the sum of the two detector channel current.
2-3
DAS/DETECTOR
GE HEALTHCARE
REV 3
2-4
5116384
GDAS
J01
J05
J07
49~51 52~61
CH
CH
62~64
CH
J45
J46
J47
65~640
CH
NC
1~3CH 4~12CH
QCAL REF
3CH
DAS
CH:
1:1
3 3
3:1
1:1
J51
NC NC
4CH 3CH
3:1
13+
2:1
2:1
13+
2:1
1:1
10
576CH
1:1
16
12+
4 2+
2:1
14
3:1
10
3:1
NC
1 MOD
3:1
1 MOD
36~48
CH
J44
1 MOD
J08
1 MOD
1 MOD
1 MOD
1 MOD
Detector
1 MOD
J06
3:1
1:1
3:1
$6OLFH
1:1
3:1
1:1
3:1
2:1
2:1
2:1
1:1
1:1
2:1
576
14
3:1
3:1
3:1
NC
3:1
1:1
3:1
Three detector channels are connected in parallel into one DAS channel.
DAS channel current is the sum of the three detector channel current.
2:1
Two detector channels are connected in parallel into one DAS channel.
DAS channel current is the sum of the two detector channel current.
2-4
3:1
%6OLFH
DAS/DETECTOR
GE HEALTHCARE
REV 3
2-5
5116384
DAS/DETECTOR CONNECTOR
The detector module consists of 32 channels(A slice and B slice). The illustration 2-3 shows the flexible PWB that
connects detector modules to the DAS Converter Board and its pin configuration.
Illustration 2-3
DAS/DETECTOR CONNECTOR
48 PIN
CONNECTOR
D C B A
FLEXIBLE
PWB
1
2
3
4
5
6
7
8
9
10
11
12
2SG
2SG
2SG
2SG
2SG
2SG
2SG
2SG
10
11
FG
FG
FG
FG
12
FG
FG
FG
FG
Note
SG: Signal Ground
FG: Frame Ground
2-5
DAS/DETECTOR
GE HEALTHCARE
REV 3
2-6
5116384
QCAL CHANNELS
The X-ray exposure can move along the Z-axis. It can be due to mechanical adjustment or to tube temperature (the
angle of the X-ray output can vary according to the temperature). There is three channels in the left most positioned
detector module that is used for Z-axis data correction. The outputs of the three channels are tied together so that
only one data is taken for three channels. The corresponding intensity according to the position of the X-ray
exposure along the Z-axis is detected (as can be seen at right side of the illustration 2-4), so that the system
recognizes the current X-ray position (along the Z-axis). There is a remaining data for Z-axis data correction in the
system. Therefore, the correction is made according to the data acquired by the Q-CAL channels.
Illustration 2-4
Q-CAL CHANNELS
QCAL CHANNELS
XRAY INTENSITY
DETECTED ACCORDING TO THE
BEAM POSITION
ALONG THE ZAXIS
A Slice
B Slice
A Slice
XRAY
EXPOSURE
XRAY
EXPOSURE
B Slice
QCAL CHANNELS
X
2-7
AFTER GLOW
The Lumex crystal scintillator has a property represented in the illustration 2-5. The scintillator continues emitting
light even after the X-ray radiation is stopped. This phenomenon is called AFTER GLOW. This remaining scintillation
provides incorrect data, but the system performs the error correction to eliminate this portion.
Illustration 2-5
AFTER GLOW
XRAY ON
scintillation
XRAY OFF
2-6
DAS/DETECTOR
GE HEALTHCARE
REV 3
2-8
5116384
The detector comprises a heater unit. The heater keeps the detector temperature constant, what is needed for
detector accuracy. The heater is controlled by the Temperature Control Box which supplies power to the heater. The
detector also comprises a thermistor that feeds back the temperature information to the Temperature Control box.
Approximately three hours is needed after turning ON the power to the Temperature Control box to stabilize the
detector temperature. The temperature is controlled to 35 0.5 C. The illustration 2-6 shows its block diagram.
Illustration 2-6
Thermistor
DETECTOR
Heater
DC24V 60W
INPUT POW
ER 24V DC
Driver for
Heater
PS
(+12V -> +5V)
PS
(+24V -> +12V)
(OP Amp
Coparato)
CPU
Switch
ROM
LED indication
2-7
DAS/DETECTOR
GE HEALTHCARE
REV 8
5116384
OVERVIEW
The DAS consists of the following boxes: RBB (Right Back Board), CBB (Center Back Board), and LBB (Left Back
Board). Each box contains the set of boards as follows:
RBB
- 2 Converter Boards
CBB
- 6 Converter Boards
LBB
- 2 Converter Boards
RBB ASSY
CBB ASSY
rter
nve
Co r d # 1
a
o
B
CONNECTER
(TO DETECTOR)
rter
nve
Co r d # 7
a
Bo
rter
nve
Co rd #2
a
o
B
Converter
Board #8
Converter
Board #3
DCP2 ASSY
Con
Boa verter
rd #
9
Con
Boa verter
rd #
4
Con
Boa verter
rd #
10
Con
Boa verter
rd #
5
LBB ASSY
rter
nve
Co rd #6
a
o
B
Illustration 3-1
Slice
A Slice
B Slice
Active Channel
685
685
Reference Channel
9+9
9+9
QCAL Channel
GND Channel
14
14
3-1
DAS/DETECTOR
GE HEALTHCARE
REV 8
3-1
5116384
OVERVIEW (CONTINUED)
DAS/Detector Structure
#45
#35
#5
#10
#27
#4
#26
#9
#19
#3
#18
#8
#11
#2
#10
#7
#1
#1
#6
DC+/-12V Analog
DC+/-5V Analog
Converter
Boards
DC+5V
Digital
Data Flow
Data Flow
POWER
SUPPLY
#34
DCP2
Data Flow
BOARD
Control Signal
3-2
DAS/DETECTOR
110V AC
N
3-3
1A
1A
2A
AC
AC
AC
AC
AC
AC
+5V
DIGITAL
+/- 5 V
ANALOG
+/- 12 V
ANALOG
+
COM
-
+
COM
6
7
8
3
4
5
1
2
4
5
6
1
2
3
7
8
CN81
5
6
1
2
3
7
8
......
......
DCP2
DIGITAL GND
+5V DIGITAL
-5V ANALOG
ANALOG GND
+5V ANALOG
-12V ANALOG
ANALOG GND
+12V ANALOG
6
7
8
3
4
5
1
2
JP3
8-8
JP4
JP1
8-8
LBB
CBB
Illustration 3-3
CN80
3-1
RBB
GE HEALTHCARE
HiSpeed Dual, ProSpeed FII
THEORY OF OPERATION
REV 8
5116384
OVERVIEW (CONTINUED)
DAS Power Connection
JP5
DAS/DETECTOR
GE HEALTHCARE
REV 8
OVERVIEW (CONTINUED)
DAS Signal Connection
LBB ASSY
FC45
J51
Converter
Board #5
FC44
J47
Converter
Board #10
FC35
J38
CN45
CN44
CN35
JS4
CBB ASSY
Converter
JS3
Board #4
J37
4onverter
Board #9
FC34
CN34
Converter
Board #3
Converter
Board #8
Converter
Board #2
FC11
Converter J14
Board #7 JS2
RBB ASSY
CN11
DETECTOR
OGP ASSY
Illustration 3-4
DCP2 ASSY
3-1
5116384
JS1
FC10
J13
CN10
Converter
Board #1
Converter
Board #6
FC2
J5
FC1
J1
3-4
CN2
CN1
DAS/DETECTOR
GE HEALTHCARE
REV 8
3-2
5116384
DATA FLOW
The illustration 3-5 shows the data flow. The Detector generates an electrical current proportional to X-ray intensity.
The electrical current is converted to voltage signal in the Converter board. The voltage signal is amplified to an
appropriate level, converted to digital data (serial). The data is then sent to to DCP2 board where it is converted
back to serial, conditioned and then sent to Operator Console through the RF Slipring.
Illustration 3-5
DAS
FPGA
DETECTOR
Electrical
Current
Analog
Data
A/D
Converter
Data
Control
DCP Board
3-5
DAS/DETECTOR
GE HEALTHCARE
REV 8
3-3
3-3-1
5116384
DCP2 BOARD
DCP2 Functions
Interfaces with the OGP for Rx reception and scan completion via RS422 bus
Sets up gain and offset trim and controls the converter cards, via the converter CAN interface
Performs serial-to-parallel conversion on data streams from the converter cards, does parity checking
on the data, and runs it through a translation table for view data ordering
Adds Forward Error Correction (FEC) to the channel data and sends it across the slipring to the Scan
Reconstruction Unit (SRU) via the high-speed serial data interface
Monitors the power supply voltages to make sure they are within the software programmable limits
Z-axis beam tracking: controls the X-ray intensity between the Converters A and B Slices
3-6
DAS/DETECTOR
GE HEALTHCARE
REV 8
3-3-2
5116384
DCP2 Interface
The DCP2 is the main control board for Data Acquisition System (DAS). It packages all the converter card data into
a single high-speed serial data stream, compatible with the receive function on the DASIFN2 on OC. Illustration 36 shows the interface of DCP2.
Illustration 3-6
DCP2 Interface
DCP2
Fiber Optical
Cable
FW LMT
BW LMT
BW-
BW+
FW-
FW+
3-7
DAS/DETECTOR
GE HEALTHCARE
REV 8
3-3-2
5116384
Inputs
From Converter Boards:
Converter CAN Bus communication for status, faults, chassis temperature readings, and serial number
information
From OGP:
From JEDI:
From Collimator
3-8
DAS/DETECTOR
GE HEALTHCARE
REV 8
3-3-2
5116384
Outputs
To Converter Boards:
Shift clock
Trigger signal
Reset signal
Analog test voltage (for MDAS compatibility) or external reference current (for GDAS compatibility)
To OGP:
RS422 Bus communication for DAS status, error, or scan complete information
High-speed serial data stream containing the view data with embedded FEC CRC.
To Collimator
3-9
DAS/DETECTOR
OGP
Interface
Converter
Interface
3-10
Collimator
Control
Collimator
Dual Port
Control
Control
Trig
Dual Port
RAM
A/D Data
KV mA
Pre AMP
A/D
FEC
Control
Data
Receiver
AT 55800A
AT 55800A
Bus Interface Core Controller
FEC
TAXI
Serial
Data
Output
Illustration 3-7
LUT
3-3
FPGA
GE HEALTHCARE
HiSpeed Dual, ProSpeed FII
THEORY OF OPERATION
REV 8
5116384
DAS/DETECTOR
GE HEALTHCARE
REV 8
3-4
5116384
CONVERTER BOARD
The signal from the detector with the level proportional to the X-ray intensity is fed to the Converter board. The
signal go through the electrical current to voltage signal conversion and conditioning circuits then is fed to the
Analog/Digital(A/D) converter. The A/D converts analog signal to 16 bits digital serial data.
Illustration 3-8
Preamplifier Architecture
150 pF
C_EXI
C_EXO
V_Ref
A_GND
72 pF
36 pF
18 pF
9 pF
R_N_red
A_IN
+
TEST_V_IN
R_h
C_h
18 pF
From 7
Other
Channel
Outputs
A_GND
OP_COM
S_GND
A_OUT
A_GND
3-11
DAS/DETECTOR
GE HEALTHCARE
REV 8
3-4
5116384
TEST_V +/-
PREAMPLIFIERS
Chan_1
Odd # Channels
Chan_127
Chan_2
O
U
T
P
U
T
Anti-Alias
Filter
Resetable
Integrator
With S/H
2
2
2
2
M
U
X
Even # Channels
Chan_128
VOLTAGE
REFERENCES (2)
(16 X 8 CH.)
I
N
P
U
T
M
U
X
E
S
ADC's
FPA's
FLOATING
POINT
AMPLIFIERS
ANALOG
TO
DIGITAL
CONVERTERS
4
(4)
(4)
S_DAT
PECL
Differential
Receivers
SH_CK +/-
PECL
Differential
SEQUENCER
Drivers
(FPGA or ASIC)
X_PRG*
CV_FLT*
CV_RST
CVB_ADDR
6
CON_HI
CONVERTER
CON_LO
I/F
I2C
TEMPERATURE
SENSOR
2
MICROCONTROLLER
SCC
I2C I/F
I2C
SERIAL
EEPROM
WR_PRTCT
3-12
DAS/DETECTOR
GE HEALTHCARE
REV 8
5116384
OVERVIEW
The DAS consists of the following boxes: RBB (Right Back Board), CBB (Center Back Board), and LBB (Left Back
Board). Each box contains the set of boards as follows:
RBB - 2 Converter Boards
CBB - 6 Converter Boards
LBB - 2 Converter Boards
The Illustration 4-1 shows the DAS boxes.
DAS (RBB, CBB, LBB)
LBB ASSY
rter
nve
Co rd #7
a
Bo
rter
nve
Co rd #2
a
o
B
Converter
Board #8
Converter
Board #3
Con
Boa verter
rd #
9
Con
Boa verter
rd #
4
SS
Con
Boa verter
rd #
10
PA
rter
nve
Co r d # 1
a
o
B
CONNECTER
(TO DETECTOR)
DC
Con
Boa verter
rd #
5
RBB ASSY
CBB ASSY
rter
nve
Co rd #6
a
o
B
Illustration 4-1
GDAS
Slice
A Slice
B Slice
Active Channel
685
685
Reference Channel
3+3
3+3
QCAL Channel
GND Channel
26
26
4-1
DAS/DETECTOR
GE HEALTHCARE
REV 8
4-1
5116384
OVERVIEW (CONTINUED)
DAS/Detector Structure
#45
#35
#5
#10
#27
#4
#26
#9
#19
#3
#18
#8
#11
#2
#10
#7
#1
#1
#6
DC+/-5V Analog
Converter
Boards
DC+5V
Digital
Data Flow
DCP
Card
Data Flow
POWER
SUPPLY
#34
Data Flow
Control Signal
4-2
DAS/DETECTOR
110V AC
N
15A
4-3
AC
AC
AC
AC
+5V
DIGITAL
+/- 5 V
ANALOG
+
COM
-
5
6
1
2
3
7
8
4
6
7
8
3
4
5
1
2
......
......
DIGITAL GND
+5V DIGITAL
-5V ANALOG
ANALOG GND
+5V ANALOG
6
7
8
3
4
5
1
2
JP3
8-8
JP4
JP1
8-8
LBB
CBB
Illustration 4-3
CN80
4-1
RBB
GE HEALTHCARE
HiSpeed Dual, ProSpeed FII
THEORY OF OPERATION
REV 8
5116384
OVERVIEW (CONTINUED)
DAS Power Connection
JP5
DAS/DETECTOR
GE HEALTHCARE
REV 8
OVERVIEW (CONTINUED)
DAS Signal Connection
LBB ASSY
DCP
OGP
J51
J47
FC45
FC44
CN45
CN44
Converter
Board #5
Converter
Board #10
J38
FC35
CN35
JS4
CBB ASSY
Converter
JS3
Board #4
J37
4onverter
Board #9
FC34
CN34
Converter
Board #3
Converter
Board #8
Converter
Board #2
Converter J14
Board #7 JS2
RBB ASSY
FC11
CN11
DETECTOR
Collimator
Illustration 4-4
Collimator Driver
4-1
5116384
JS1
J13
FC10
CN10
Converter
Board #1
Converter
Board #6
J5
J1
4-4
FC2
FC1
CN2
CN1
DAS/DETECTOR
GE HEALTHCARE
REV 8
4-2
5116384
DATA FLOW
The illustration 4-5 shows the data flow. The Detector generates an electrical current proportional to X-ray intensity.
The electrical current is converted to voltage signal in the Converter board. The voltage signal is amplified to an
appropriate level, converted to digital data (serial). The data is then sent to DCP Card where it is converted back to
serial, conditioned and then sent to Operator Console through the RF Slipring.
Illustration 4-5
DAS
Offset
Trim
DAC
FPGA
DETECTOR
Electrical
Current
Fold
Control
4-5
DCP Board
DAS/DETECTOR
GE HEALTHCARE
REV 8
4-3
4-3-1
5116384
DCP CARD
DCP Functions
Interfaces with the OGP for Rx reception and scan completion via RS422 bus.
Sets up gain and offset trim and controls the converter cards, via the converter CAN interface.
Performs serial-to-parallel conversion on data streams from the converter cards, does parity checking
on the data, and runs it through a translation table for view data ordering.
Adds Forward Error Correction (FEC) to the channel data and sends it across the slip ring to the OC via
the high speed serial data interface.
It provides the A/D converter chip, it acquires the kV and mA values for each scan.
Communication with Detector Heater Control Board to monitor the Detector Temperature.
4-6
DAS/DETECTOR
GE HEALTHCARE
REV 8
4-3-2
5116384
DCP Interface
The DCP is the main control board for Data Acquisition System (DAS). It is applicable for GDAS converter cards.
It packages all the converter cards data into a single high speed serial data stream to OC, compatible with the receive function on the DASIFN at 95/110 Mbps slip ring communication. Illustration shows the interface of DCP.
Illustration 4-6
DCP Interface
SDCOM/+-
CAN_LO
DSCOM/+-
CAN_HI
DATA 2_ODD/+-
OGP
DASTRIG/+-
DATA 2_EVEN/+-
GPLS2/+-
DATA 1_ODD/+-
XRON/+-
DATA 1_EVEN/+-
AC/OFF/+-
STRIG/+-
DCP
RESET/+-
DAS
SH_CK/+-
DASENABLE/+-
CN_FLT
CVTR
CN_RST
X-PROG
POWER
SUPPLY
RAW+5VA
W-PRTCT
RAW-5VA
IREF.EXT
RAW+5VL
AGND
AGND
LGND
Fibre Optical
Cable
From JEDI
0~10V
Slip Ring
KV Signal
mA Signal
FW LMT
BW LMT
BW-
BW+
FW-
FW+
4-7
DAS/DETECTOR
GE HEALTHCARE
REV 8
4-3-2
5116384
Inputs
From Converter Boards:
Converter CAN Bus communication for status, faults, chassis temperature readings, and serial number
information (CAN)
From OGP:
From Collimator:
From Backplane:
Power Supply
Outputs
To Converter Boards:
Converter CAN Bus communication for control information (CAN Differential Pairs)
To OGP:
4-8
DAS/DETECTOR
GE HEALTHCARE
REV 8
4-3-2
5116384
To Collimator:
To Slip-ring (OC):
High-speed serial data stream containing the view data with embedded FEC (High speed optical fiber
link).
Illustration 4-7
CLK
GDAS DCP
CARD ASSY
Fiber
32MHz
Reset
ARM
JTAG
RAM
RAM
ARM
AT91
TG com
Flash-ROM
PLL CLK
30MHz
Pulse/
status
FEC
FPGA
TAXI
CY9689A
Cyclone
CFG
CAN
ADC
LVDS
LVDS
Iref
AD977
Mx9129
Ds032
Ref200
KV/mA
DHCBCollimator
OGP
JEDI
CVTR
JTAG
+5V_Logic
+5V_Analog
-5V_Analog
Power Supply
4-9
DAS/DETECTOR
GE HEALTHCARE
REV 8
4-4
5116384
CONVERTER BOARD
The signal from the detector with the level proportional to the X-ray intensity is fed to the Converter board. The
signal go through the electrical current to voltage signal conversion and conditioning circuits then is fed to the
Analog/Digital(A/D) converter. The A/D converts analog signal to 16 bits digital serial data.
Illustration 4-8
Preamplifier Architecture
4-10
DAS/DETECTOR
GE HEALTHCARE
REV 8
4-4
5116384
VOLTAGE
REFERENCE
IREF_INT
IREF_EXT
..
.
Chan_127
Anti_Alias
Filter
..
.
Chan_128
(2 x 64 CH.)
Chan_1
Odd # Chan
CHARGE TO
DIGITAL
CONVERTERS
Chan_2
10
Even # Chan
D0:D9
ASIC/FPGA INTERFACE
10
LVDS
Differential
Receivers
3.3
V
O
L
T
(FPGA or ASIC)
1.5V Core
2.5, 3.3 VIO
7
CVB_ADDR
CAN_HI
CAN_LO
B
A
N
K
CAN
I/F
3.3
V
O
L
T
B
A
N
K
CV_FLT*
I C I/F
2
I C
TEMPERATURE
SENSOR
SHARC DSP
SDRAM
Flash Boot
Prom
CPLD
INTERFACE
CV_RST
CPLD
14 Pin JTAG
HEADER
4-11
DAS/DETECTOR
GE HEALTHCARE
4-12
DAS/DETECTOR
GE HEALTHCARE
REV 1
5116384
X-RAY GENERATOR
TABLE OF CONTENTS
SECTION
PAGE
1-1
GLOSSARY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
JEDI HIGH LEVEL BLOCK DIAGRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-1
1-2
2-1
2-1
2-2
2-3
2-4
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
STANDARD FEATURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
APPLICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ARCHITECTURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-4-1
Kernel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-4-2
Options Depending on the Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-4-3
A Packaging Architecture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-1
2-1
2-2
2-3
2-3
2-3
2-4
3-1
3-1
3-1
X-RAY GENERATOR
GE HEALTHCARE
ii
X-RAY GENERATOR
GE HEALTHCARE
REV 1
5116384
GLOSSARY
Term
Definition
ABC
AEC
CAN
CPU
EPLD
EMC
Electro Magnetic Compatibility. The EMC function prevents the generator from polluting the power source.
FPGA
HV Ripple
State Machine
IGBT
Ilp
Ilr
MOS
OGP
PDU
RMS
1-1
X-RAY GENERATOR
1-2
Inverter
Tank
heater supply
bus
HV Cables
DC Bus
Heater
High Voltage
Control Bus
kV Control
High Voltage
1 phase
3 phase power
tube cooling
input
input
Filter
EMC
System
Interface
X-Ray Tube 1
Low Voltage
Power
Supply
rotation phases
Rotation
Illustration 1-1
AC/DC
System
1-2
GE HEALTHCARE
HiSpeed Dual, ProSpeed FII
THEORY OF OPERATION
REV 1
5116384
X-RAY GENERATOR
GE HEALTHCARE
REV 6
5116384
INTRODUCTION
Jedi is the engineering name for a family of compact high frequency X-Ray generators. This generator
family covers a wide range of applications from mobile equipment up to vascular systems:
JEDI 12-25 kW:
Mobile applications
JEDI 24-48 kW:
CT applications
JEDI 32-50 kW:
RAD applications
JEDI 50-65-80 kW: RF applications
JEDI 100 kW:
VASCULAR applications
2-2
STANDARD FEATURES
Jedi is a family of 150 kV generators operating from 12 kW up to 100 kW for all the major radiological,
fluoroscopic and CT applications. The family handles 1 ms to continuous exposures with tube currents
ranging from 0mA up to 1000 mA.
The generators feature the very latest technology available:
Constant potential independent of line voltage variations
Power generation by a high-frequency converter (High voltage ripple: 40 kHz-140 kHz)
Distributed micro-processor controlled functions (CAN bus)
Other features include:
Single phase, three phase or battery power source
Very low kV and mA ripple, excellent accuracies and dose reproducibility
Compatible with a wide range of tubes, high speed or low speed, can supply up to 3 different tubes.
Thermal load interactive integrator ensuring optimum use of the heat protection curve of the x-ray
tube
Available in various packaging configurations: gantry, under-table, cabinet
Serviceability: high reliability, fast installation (no generator calibration), application error codes
ensure fast troubleshooting
Meets CE marking (and in particular EMC), IEC, UL, CSA, MHW regulations (if required)
Optional pulsed fluoroscopy
2-1
X-RAY GENERATOR
GE HEALTHCARE
REV 6
2-3
5116384
APPLICATIONS
RAD
Surgery
RF
Vascular/
Cardiac
3 Points Mode
2 Points Mode
1 Point Mode
x
x
0 Point Mode
AEC
Tomography
AET
Continuous/Pulsed Fluoroscopy
Rad Exposures
Cinema 30 fr/s
Cinema 90 fr/s
CT
ABC
Variable mA Scans
Low mA Fluoroscopy
Legend:
-
2-2
X-RAY GENERATOR
GE HEALTHCARE
REV 6
2-4
5116384
ARCHITECTURE
Kernel
These functions are the Jedi core. They are present in all versions of the generator.
A function can be unique for all products, or can have several different releases based on product specification.
Examples:
The anode rotation function is available in 2 releases:
low speed rotation for applications where the tube has a max rotation of 3000 rpm
high speed/low speed rotation for applications where at least one of the tubes can use
8000-10000 rpm
The control bus is unique.
2-4-2
CT interface
RAD interface (console interface, room interface, AEC management present or not)
- ATLAS interface
EMC function
Grid function (RF, vascular)
Bias function (RF, vascular)
Tube management (2 tubes or 3 tubes option)
2-3
X-RAY GENERATOR
GE HEALTHCARE
REV 6
2-4-3
5116384
A Packaging Architecture
The packaging architecture consists in a set of boxes which can be put together in several ways to make
Jedi fit either in a cabinet, or a console foot, or a table foot. The boxes can also be split in 2 units distant
of several meters (example: CT gantry).
The boxes normally consist of the following:
Auxiliaries Box
Rotation function
Heater function
Low voltage power supply (which can also be in the AC/DC box)
HV tank
HV power inverter
kV control
System interface (for the less complex system interface)
MC filter (optional)
AC/DC function
Low voltage power supply (which can also be in the auxiliaries box)
2-4
X-RAY GENERATOR
GE HEALTHCARE
REV 6
5116384
Illustration 2-1
Inverter Assembly
Dual Snub
AC/DC
Board
Tube
Gate Command
Board
HV
+ KV Measure
CT
Inter
face
HV Tank
Rectifier
Block
KV Control Board
EMC Filter
AC/DC BOX
POWER BOX
LVPS
Board
Heater Board
Rotation
Capacitor
Rotation Board
Rotation
Capacitor
AUXILIARIES BOX
2-5
X-RAY GENERATOR
GE HEALTHCARE
2-6
X-RAY GENERATOR
GE HEALTHCARE
REV 1
5116384
The following table describes the main signals of the generator. For each of them, typical values are presented for the main applications:
Signal/Application
Anode Rotation:
Phase P current during high speed run
2 A peak / 144 Hz
2 A peak / 144 Hz
Filament Drive:
Inverter current in standby 2.5 A
160V
kV Control:
DC bus measure at 400 VAC
560 V
80 A peak / 25 kHz
3-1
X-RAY GENERATOR
GE HEALTHCARE
3-2
X-RAY GENERATOR
GE HEALTHCARE
REV 1
5116384
Publication
Description
417-5032
Alternating Current
335-1
335-1
3N
Direct Current
417-5019
348
417-5008
417-5007
Emergency Stop
A-1
GE HEALTHCARE
REV 1
5116384
Symbol
Publication
Description
Type B
417-5339
417-5009
Standby
Start
Table Set
Abort
Intercom
Warning sign :
Radiation of Laser Apparatus
A-2
GE HEALTHCARE
REV 1
5116384
Symbol
Description
Microphone (Mic)
Contrast
Brightness
50C
-10C
Humidity 5-95%
Excluding Condensation
Air Pressure
700-1060hPa
A-3
GE HEALTHCARE
REV 1
5116384
CLASS 1 EQUIPMENT
Any permanently installed equipment containing operator or patient accessible surfaces must provide backup
protection against electric shock,in case the BASIC INSULATION fails. In addition to BASIC INSULATION,Class1
equipment contains a direct connection to a PROTECTIVE(EARTH) CONDUCTOR which prevents shocks when a
person touches a broken piece of equipment or touches two different equipment surfaces simultaneously.
TYPE B EQUIPMENT
CLASS I, II, or III EQUIPMENT or EQUIPMENT with INTERNAL ELECTRICAL POWER SOURCES provide an
adequate degree of protection against electric shock arising from (allowable) LEAKAGE CURRENTS or a
breakdown in the reliability of the protective earth connection.
ORDINARY EQUIPMENT
Enclosed EQUIPMENT without protection against the ingress of water.
OPERATION 0f EQUIPMENT
CONTINUOUS OPERATION WITH INTERMITTENT LOADING.
Operation in which EQUIPMENT is connected continuously to the SUPPLY MAINS.The stated permissible loading
time is so short that the long term on-load operating temperature is not attained.The ensuing interval in loading is,
however, not sufficiently long for cooling down to the long term no-load operating temperature.
EQUIPMENT not suitable for use in the presence of a FLAMMABLE ANESTHETIC MIXTURE WITH AIR or WITH
OXYGEN or NITROUS OXIDE
CLEANING
The system is NOT WATERPROOF. It is NOT designed to protect internal components against the ingress of
liquids.Clean external system surfaces(Gantry,table,consoles and accessories)with a soft cloth dipped in hot water
and wrung DAMP/DRY. (NOT dripping!) IF NECESSARY, use only mild (dish washing liquid) soap to remove dirt.
NOTICE
Avoid damage to equipment! Some spray and wipecleaners etch and permanently cloud
clear plastic surfaces!! Use only warm water and mild liquid soap to clean surfaces.
A-4
www.gehealthcare.com
Imagination at work