Beruflich Dokumente
Kultur Dokumente
DOI 10.1007/s00330-011-2062-1
COMPUTED TOMOGRAPHY
Received: 29 September 2010 / Revised: 19 November 2010 / Accepted: 1 December 2010 / Published online: 20 January 2011
# European Society of Radiology 2011
Abstract
Objective The aim of the study was to assess the
performance and diagnostic value of a dual energy CT
approach to reduce metal artefacts in subjects with metallic
implants.
Methods 31 patients were examined in the area of their
metallic implants using a dual energy CT protocol
(filtered 140 kVp and 100 kVp spectrum, tube current
relation: 3:1). Specific post-processing was applied to
generate energies of standard 120 and 140 kVp spectra
as well as a filtered 140 kVp spectrum with mean photon
energies of 64, 69 and 88 keV, respectively, and an
optimized hard spectrum of 95150 keV. Image quality
and diagnostic value were subjectively and objectively
determined.
Results Image quality was rated superior to the standard
image in 29/31 high energy reconstructions; the diagnostic
value was rated superior in 27 patients. Image quality and
diagnostic value scores improved significantly from 3.5 to
2.1 and from 3.6 to 1.9, respectively. In several exams
decisive diagnostic features were only discernible in the
high energy reconstructions. The density of the artefacts
decreased from 882 to 341 HU.
F. Bamberg : A. Dierks : K. Nikolaou : M. F. Reiser :
C. R. Becker : T. R. C. Johnson
Department of Clinical Radiology, Ludwig-Maximilians
University,
Klinikum Grosshadern,
Munich, Germany
T. R. C. Johnson (*)
University of Munich,
Grosshadern Campus, Marchioninistrasse 15,
81377 Munich, Germany
e-mail: thorsten.johnson@med.uni-muenchen.de
Introduction
Since the beginning of computed tomography (CT), metal
artefacts have been representing a significant limitation in
diagnostic image evaluation [8]. In the presence of metallic
joint prostheses or osteosynthetic material, it would be
desirable to evaluate the metal implant itself, the interface
between implant and bone and the surrounding tissue.
Important diagnostic criteria include exclusion of fractures
or loosening and verification of sufficient coverage of the
implant [13] and ruling out of hematoma or inflammation in
the adjacent soft tissue. However, due to the occurrence of
metal artefacts, the evaluation of these features remains
challenging with many cases rendered uninterpretable, even
with hard convolution kernels and widened CT density
ranges [8].
The metal artefacts actually comprise two main different
components. The one component is photon starvation due
to full absorption of the x-ray quanta, causing zerotransmission projections. The other component is beam
hardening due to absorption of low energy quanta [1].
While there is no remedy for these artefacts, in clinical
practice, higher energy quanta reduce the extent of artefacts
to some degree and would constitute a potential approach to
improve diagnostic CT image quality in metallic implants.
Specific algorithms such as linear interpolation of reprojected metal traces and multi-dimensional adaptive filtering
of the raw data can also improve image quality but may
also yield other artefacts [12].
Dual Energy CT as a relatively new approach has
brought about several advances in clinical CT interpretation, largely by improving the specificity of diagnostic
information. This is accomplished mainly by displaying the
presence, amount or distribution of specific substances
based on their specific absorption spectrum [6]. For
instance, these methods make it possible to differentiate
kidney stones, to characterize renal lesions, to map lung
perfusion, to depict arteries without superimposing bones or
to differentiate plaque components [2, 3, 911]. Moreover,
images can be optimized in certain aspects based on the
spectral information, i.e. the difference between both
datasets. For instance, it is possible to accentuate the
iodine-related contrast and decrease the noise in the image
[5]. Also, it feasible to extrapolate the beam hardening to
generate images as though they had been acquired with
monoenergetic high energy quanta. This latter approach
may substantially reduce metal artefacts in the acquired
data sets.
Thus, the purpose of this study was to initially evaluate
the efficacy of this technique in removing metal artefacts
based on quantitative CT density measurements and
subjective grading of artefacts and adequacy for diagnostic
evaluation.
1425
1426
Fig. 1 Two sets of images showing screws in the spine and in the
tibia reconstructed at 64, 69, 88, 105 and an optimal keV setting (left
to right). Note that the spinal canal the thin layer of bone covering the
left screw are only discernible in the two reconstructions at the highest
energy. Similarly, the screw in the tibia is optimally depicted in the
rightmost image
Statistical analysis
Descriptive statistics are provided as mean standard
deviation for continuous and n (%) for categorical variables.
To determine differences of the artefact and diagnostic image
Fig. 2 Improvement of subjectively defined diagnostic image
quality (a) and objectively
measured HU in the most
significant artefact zone
adjacent to the metal implant
(b) at different keV settings
from 64 keV to 105 keV as
well as the individually
selected optimal keV
setting (optimal)
Results
Overall, 31 consecutive subjects were included in the
analysis. They were predominantly elderly women (mean
age: 64.819, 54% female). No adverse event occurred
during the CT examination. Regarding the implants, there
were 22 in the trunk of the body (6 spine, 12 hip, 4 femur)
and 9 in the extremities (2 humerus, 2 radius/ulna, 5 ankle),
examined with the respective protocol. Average CT dose
index (CTDI) was 11.0 mGy in the extremities and
1427
Table 1 Observed diagnostic image quality and extent of artifacts across the different keV settings. Delta indicates the difference between 64 keV
and optimal setting, p the probability of equivalence
Variable
64 keV
69 keV
88 keV
105 keV
Optimal
P-value
Diagnostic Value
Extent of Artifacts
Artifact Density (HU)
Reference Soft Tissue Density (HU)
3.60.7
3.70.7
900.1222
48.044
3.50.7
3.60.7
875.8244
54.946.5
2.80.8
2.90.9
495.4251
56.822
2.20.9
2.01.1
374.8287
52.924.1
2.10.9
1.91.1
341309
45.820.2
1.60.8
1.80.9
558269
2.234
<0.001
<0.001
<0.001
0.67
Fig. 4 68-year old male patient with a locked intramedullary nail who
underwent dual energy CT for assessment of pseudarthrosis. Images
show a coronal reconstruction (a) and a volume rendered image (b) at
an extrapolated photon energy of 105 keV. There are minimal residual
artefacts at the largest diameter of the metal in axial plane due to
screws
1428
Discussion
1429
Conclusion
High energy reconstructions of Dual Energy CT datasets can
significantly reduce metal artefacts and improve image quality
and diagnostic value. The evaluation of metallic implants
and adjacent bone or tissue is considerably enhanced.
References
1. Barrett JF, Keat N (2004) Artifacts in CT: recognition and
avoidance. Radiographics 24:16791691
2. Graser A, Becker CR, Staehler M, et al. Single-phase dual-energy
CT allows for characterization of renal masses as benign or
malignant. Invest Radiol 45:399405
3. Graser A, Johnson TR, Bader M et al (2008) Dual energy CT
characterization of urinary calculi: initial in vitro and clinical
experience. Invest Radiol 43:112119
4. Haramati N, Staron RB, Mazel-Sperling K et al (1994) CT scans
through metal scanning technique versus hardware composition.
Comput Med Imaging Graph 18:429434
5. Holmes DR 3rd, Fletcher JG, Apel A et al (2008) Evaluation of
non-linear blending in dual-energy computed tomography. Eur J
Radiol 68:409413
6. Johnson TR, Krauss B, Sedlmair M et al (2007) Material
differentiation by dual energy CT: initial experience. Eur Radiol
17:15101517
7. Krauss B, Schmidt B, Sedlmair M, Flohr T (2010) Evaluation of an
image-based method to calculate monoenergetic images from dual
energy images(ed)^(eds) Abstract Book, European Congress of
Radiology 2010. European Society of Radiology, Vienna, p S201
8. Lee MJ, Kim S, Lee SA et al (2007) Overcoming artifacts from
metallic orthopedic implants at high-field-strength MR imaging
and multi-detector CT. Radiographics 27:791803
9. Thieme SF, Hoegl S, Nikolaou K, et al. Pulmonary ventilation and
perfusion imaging with dual-energy CT. Eur Radiol 20:28822889
10. Uotani K, Watanabe Y, Higashi M et al (2009) Dual-energy CT
head bone and hard plaque removal for quantification of calcified
carotid stenosis: utility and comparison with digital subtraction
angiography. Eur Radiol 19:20602065
11. Watanabe Y, Uotani K, Nakazawa T et al (2009) Dual-energy
direct bone removal CT angiography for evaluation of intracranial
aneurysm or stenosis: comparison with conventional digital
subtraction angiography. Eur Radiol 19:10191024
12. Watzke O, Kalender WA (2004) A pragmatic approach to metal
artifact reduction in CT: merging of metal artifact reduced images.
Eur Radiol 14:849856
13. White LM, Buckwalter KA (2002) Technical considerations: CT
and MR imaging in the postoperative orthopedic patient. Semin
Musculoskelet Radiol 6:517
14. Yu L, Li H, Mueller J et al (2009) Metal artifact reduction from
reformatted projections for hip prostheses in multislice helical
computed tomography: techniques and initial clinical results.
Invest Radiol 44:691696