Beruflich Dokumente
Kultur Dokumente
p
y Stimulated Luminescence
(OSL) dosimetry in radiotherapy
Joanna E.Cygler
yg 1 and Eduardo Yukihara2
The Ottawa Hospital Regional Cancer Centre, Ottawa, Canada
2Department of Physics, Oklahoma State University, Stillwater, Oklahoma
1
The Ottawa
LHopital
Hospital
d
dOttawa
Ottawa
Regional Cancer Centre
Disclosure
The authors have received research
support from Landauer Inc.
Inc
Outline
Environmental corrections
Linearity of dose response
Dose-rate dependence
Energy dependence
Directional dependence
Fading
g
Introduction to luminescence
dosimetry
EXPOSURE
Radiation
Radiation sensor
(insulating crystal)
Introduction to luminescence
d i t
dosimetry
STORAGE
Introduction to luminescence
d i t
dosimetry
READOUT
Light emission
(e.g., blue, UV)
Thermal
stimulation
ti l ti
(heating)
Introduction to luminescence
d i t
dosimetry
READOUT
Light
stimulation
(e.g., green)
Light emission
(e.g., blue, UV)
Introduction to luminescence
d i t
dosimetry
READOUT
Optically Stimulated
Luminescence detectors
(OSLD )
(OSLDs):
Al2O3:C (TLD500)
Thermoluminescence detectors
(TLDs):
LiF:Mg,Ti, CaF2
Light
stimulation
(e.g., green)
Thermal
stimulation
ti l ti
(heating)
Light emission
(e.g., blue, UV)
PMT
Detection filters
Stimulation
filters
OSL
Dosimeter
(OSLD)
POSL (pulsed
( ls d OSL)
CWCW
-OSL readout method
CW
W-OSL (a
arbitratry u
units)
400
300
constant
200
100
0
0
100
200
300
Time (s)
400
500
600
Stimulatio
on intensity
50
100
150
200
250
300
350
400
450
500
550
50
100
150
200
250
300
350
400
450
500
550
Gate S
State
on
off
Time (s)
One manufacturer
MicroStarTM
www.Landauer.com, www.osldosimetry.com
OSL dosimeters
Dot
nanoDot
OSLD reproducibility
40
35
% number
30
25
20
15
10
5
0
0.94
0.96
0.98
1.00
1.02
1.04
Relative sensitivity
y
Courtesy of C. Yahnke
1.06
1.08
Environmental corrections
Temperature dependence
During irradiation
During readout
Rdg
g (arbitrary
y units)
1000
800
600
400
200
0
0
100
200
300
400
Dose / cGy
500
DoseDose
-rate dependence
6 MV
gamma rays
Mean Energy
(keV)
Q
F
Al 2O3 Co
Q
F
LiF Co
Ratio
Al 2 O3 /LiF
1250
1.000
1.000
1.00
50 kV X-rays
29
3.219 0.3%
1.463
2.20
100 kV X-rays
60
2.861 0.3%
1.376
2.08
150 kV X-rays
105
1.607 0.3%
1.245
1.29
250 kV X-rays
170
1.449 0.3%
1.192
1.19
6 MV X
X-rays
2020
0 990 0.3%
0.990
0 3%
0 987
0.987
1.00
1 00
10 MV X-rays
3050
0.983 0.3%
0.976
1.00
15 MV X-rays
4180
0.980 0.3%
0.976
1.00
25 MV X-rays
6600
0.973 0.3%
0.976
1.00
Directional dependence
Q(t))/Q(1min)
1.0
0 8
0.8
0.6
0.4
0.2
0
Time / min
10
12
OSL dosimetry
Advantages
Ad
t
vs. disadvantages
di d
t
Advantages
Disadvantages
High sensitivity
Sensitivity to light
High precision
Non-tissue equivalent energy
Size
d
dependence
d
Convenience
Only 1 material currently
Readout flexibility
available (only 1 provider)
Fast non-destructive
Fast,
non destructive readout
Narrow stimulating beams may
could allow dose mapping
No significant fading - dose
storage
No need for annealing
Although it can be bleached
and re-used if needed*
60Co
In vivo dosimetry
y
M Aznar,
M.
Aznar Phys.
Phys Med.
Med Biol.
Biol 49,
49 1655
165569
69, 2004
OSL c e t dt
OSL c e t dt
courtesy
t
off C
C. Andersen,
A d
Ri
Ris
Summary
OSLD have linear dose response and good
reproducibility (screened) for standard clinical doses
Minimal energy dependence in megavoltage photon
beams
Suitable
bl for
f accurate d
dosimetric measurements
individual calibration factors
Can
C b
be used
d in
i variety
i t of
f clinical
li i l applications
li ti
surface dose detectors
entrance
t
and
d exit
it dose
d
measurements
t
brachytherapy
dose mapping
Are suitable for remote dosimetry
Acknowledgements
Claus Andersen, Ris National Laboratory
Cliff Yahnke, Landauer Inc.
Thank you
Energy dependence
Al2O3:C
water
Zeff=10.2
Zeff=7.2
LM-OSL arbitrary un
nits
LMLM
-OSL readout method
Time (s)