Sie sind auf Seite 1von 14

Support Note 393, Revision A

NanoScope Version 6 Addendum

393.1 Introduction
Your microscope ships with several pieces of documentation: a microscope manual (MultiMode,
BioScope, Dimension 3100 or Dimension 5000 Manual, depending on your system), a software
manual (NanoScope 6.12 Users Guide), and any supplemental documents specific to your
microscope and/or options you have purchased. NanoScope Software Version 6.12 introduces new
features to Version 6 software that belong in the microscope manual. This support note summarizes
changes to your microscope manual due to the release of this software.

See the following sections:

Version 5 Enabled Functions: Section 393.2

Version 6.12 New Functions: Section 393.3

Image Browse Window: Section 393.3.1

Capture Now: Section 393.3.2

Fast Z Scan Calibration: Section 393.3.3

Remove File on Close: Section 393.3.4

Multiple Channel Print: Section 393.3.5

Export View as JPEG: Section 393.3.6

Point and Shoot: Section 393.3.7

Document Revision History: NanoScope Version 6 Addendum

Revision Date Section(s) Affected Reference Approval

A 6-23-04 New Release C. Kowalski

Veeco Instruments Inc., 2004 Part Number: 013-393-000


112 Robin Hill Road
Santa Barbara, CA 93117
805.967.1400
Version 5 Enabled Functions

Improved Cantilever Tune Dialog Box: Section 393.3.8

Electric Tune: Section 393.3.9

XY Drift Correction: Section 393.3.10

ASCII Export: Section 393.3.11

393.2 Version 5 Enabled Functions


NanoScope Version 6.12r1 introduced new features to Version 6 software that are already described
in your microscope manual. See your microscope manual for more information about these
features.

Table 393.2a NanoScope 6.12r1 Additions

NanoScope 6.12
Opening/ Enabling Reference for Version 6.12
Addition

Frequency Feedback Use with LiftMode; MFM


Force Modulation In the Feedback panel of the Scan View, select TM Deflection
for the SPM Feedback parameter.
Scanner Calibration To open the Scanner Calibration dialog box, select Tools >
Calibrate > X-Y.
Autocalibration To open the Autocalibration dialog box, select Tools > Cali-
brate > Auto Calibration.
Capture Calibration To open the Capture Calibration dialog box, select Realtime
> Capture Calibration.
STM To enable STM, select Tools > Select Microscope > Edit or
New > Microscope > STM. Refer to your microscope manual
(e.g., MultiMode and Dimension 3100 Manuals have STM
chapters) in addition to the CRM
Surface Potential Use with LiftMode, Tapping Mode

2 NanoScope Version 6 Addendum Rev. A


Version 6.12 New Functions

393.3 Version 6.12 New Functions


Some features are new to NanoScope software. These features do not yet appear in the your
microscope manual, and are described below.

393.3.1 Image Browse Window

Directory icons now appear in the browse window. Double-click a folder icon to browse
that directory (see Figure 3a).

Figure 393.3a Directory Icons

Capture Directory Icon

Directory Folder Icon

Added the Capture Directory icon to the Image Browse Area (see Figure
393.3a). Click this icon to automatically open the Capture Directory.

You can now configure the Capture Directory. You can change the current capture
directory in the Capture File dialog box (Realtime > Capture Filename) (see Figure
393.3b).

Rev. A NanoScope Version 6 Addendum 3


Version 6.12 New Functions

Figure 393.3b Capture File Dialog Box

393.3.2 Capture Now

Added Realtime > Capture Now menu item. Capture Now creates a partial capture of the current
scan and saves the image in the Capture Directory.

393.3.3 Fast Z Scan Calibration

1. Set the COM Port to Fast Scan

a. Select Tools > Select Microscope.

b. Select the Edit button in the Microscope Select dialog box.

c. Select the Serial button in the Equipment dialog box.

d. Select the Edit button in Serial Port Configuration dialog box.

e. In the Edit Port Setup dialog box, set the COM 1 Equipment Type parameter to Fast
Scan.

2. On the Other panel of the Scan View, set the Fast Scan parameter to Enabled.

3. Select Calibrate > Fast Scan Z. The Force Calibration dialog box will open.

4 NanoScope Version 6 Addendum Rev. A


Version 6.12 New Functions

393.3.4 Remove File on Close

Added Tools > Options > Remove File on Close menu item (see Figure 3c). You can also right-
click the workspace and select Remove File on Close. When selected, this option removes the
workspace item of the offline file when all views on that file are closed.

Figure 393.3c Remove File on Close Menu Item

393.3.5 Multiple Channel Print

NanoScope will now print up to three channels simultaneously to an external printer.

Rev. A NanoScope Version 6 Addendum 5


Version 6.12 New Functions

393.3.6 Export View as JPEG

You can now export a view as a JPEG. To print multiple channels simultaneously, all must be open
in the same window. Export a view one of two ways:

Right-click the view and select Export > View.

-OR-

Select File > Export > View.

393.3.7 Point and Shoot

Point and Shoot is a new view to the NanoScope Version 6.12 software. The Point and Shoot
View allows you to select specific points on an image (see Figure 3d). Use Point and Shoot to
capture an image and/or collect a force curve for every point you designate.

When you click a point on an image, a crosshair (+) marks the location. You can designate
individual points, or use the tools in the Point and Shoot dialog box to assign multiple points
simultaneously. See Section 2.2.3: Point and Shoot View of the NanoScope Software 6.12 Users
Guide for a description of the Point and Shoot parameters.

Offline Point and Shoot is also available in Multi-Curve Analysis (MCA).

6 NanoScope Version 6 Addendum Rev. A


Version 6.12 New Functions

Figure 393.3d Point and Shoot View

Rev. A NanoScope Version 6 Addendum 7


Version 6.12 New Functions

393.3.8 Improved Cantilever Tune Dialog Box

The Cantilever Tune dialog box has been rearranged to provide more consistent accessibility (see
Figure 393.3e). The Cantilever Tune parameters are now in collapsible sections in the box on the
right side of the dialog box. For additional parameters, click the More button at the bottom of the
dialog box.

Figure 393.3e Cantilever Tune Dialog Box

Cantilever Tune
Parameters

More button

8 NanoScope Version 6 Addendum Rev. A


Version 6.12 New Functions

393.3.9 Electric Tune

Theory

When evaluating Surface Potential, you can now tune the cantilever electrically. Traditional
cantilever tune oscillates the cantilever via tapping piezo material (see Figure 393.3f). Electric Tune
sweeps the frequency of the electric field surrounding the cantilever. The force of the electric field
moves the cantilever. The cantilever amplitude is then plotted in the Generic Sweep dialog box.

Figure 393.3f Cantilever Tune Method Comparison

Traditional Cantilever Tune Electric Tune

Electric Field

Piezo Material Oscillates the Cantilever Electric Field

Change the Electric Field Frequency


to Oscillate e the Cantilever

Rev. A NanoScope Version 6 Addendum 9


Version 6.12 New Functions

Procedure

1. While evaluating Surface Potential in Realtime mode, select RealTime > Generic Sweep.
The Generic Sweep dialog box will display.

Figure 393.3g Typical Generic Sweep

Interleave Mode

2. In the Generic Sweep dialog box, set the Sweep parameters Input igain and Input pgain to
0. This disables the Surface Potential feedback loop, which works to keep the cantilever's
amplitude at zero.

3. Select Interleave as the Mode (see Figure 393.3g).

4. You can adjust several parameters to acquire interesting data:

Try changing the Data Scale, Drive Phase and Drive Amplitude (see Figure 393.3h).

Try changing the channel. For example, the Data Type can be changed to Potential
Input.

Try sweeping another channel. For example, set the Graph parameter Sweep Output to
Bias (see Figure 393.3h).

10 NanoScope Version 6 Addendum Rev. A


Version 6.12 New Functions

Figure 393.3h Electric Tune Generic Sweep while adjusting the Drive Amplitude and Drive Phase

Note: Remember to re-enable the Input gains before collecting further Surface
Potential data.

Rev. A NanoScope Version 6 Addendum 11


Version 6.12 New Functions

393.3.10 XY Drift Correction

Due to temperature differences, XY drift can occur between two successive images while scanning.
You can capture two images in Realtime mode for later XY Drift analysis in Offline mode. In
Offline XY Drift analysis, the software can automatically calculate the lateral shift between two
images (see Figure 393.3i). You can also manually enter the drift. See the NanoScope Software 6.12
Users Guide for a description of the XY Drift parameters.

Figure 393.3i XY Drift Dialog Box

393.3.11 ASCII Export

When transporting the results of using NanoScope software to another computing platform, the
most generic format is an ASCII text file. The File > Export > ASCII command builds such a file.

1. Select a directory, then an image file within it, from the file browsing window at the right of
the NanoScope Version 6 main window. Right-click in the thumbnail image to open the
menu shown in Figure 393.3j.

12 NanoScope Version 6 Addendum Rev. A


Version 6.12 New Functions

Figure 393.3j The ASCII Export Command

2. Click Export > ASCII to open the Export dialog box (see Figure 393.3k).

Figure 393.3k The File Export Dialog Box

Rev. A NanoScope Version 6 Addendum 13


Version 6.12 New Functions

3. Select the units in which to record the data in the new file by checking the appropriate boxes.
You can also export image header, ramp, or time information by selecting those check boxes.

4. Click Save As..., designate a directory path and filename, and click Save.

Figure 393.3l shows the start of an exported ASCII image file saved without a header.

Figure 393.3l Sample ASCII Image File

14 NanoScope Version 6 Addendum Rev. A

Das könnte Ihnen auch gefallen