Beruflich Dokumente
Kultur Dokumente
ABSTRACT
851
ator between the compressor and the cold end of the tube. These compo-
nents make up a system called the basic pulse tube, which was studied
extensively by Gifford and coworkers. 2 In 1984 Mikulin inserted an ori-
fice at the closed end of the tube which allowed gas to flow into and
out of a large reservoir volume. 3 Figure 1 shows the NBS version of
this new pulse tube and is referred to as the orifice pulse tube. The
refrigeration cycle begins as the piston moves forward and the gas is
cooled to the cold end temperature as it flows through the regenerator
and into the tube. The gas in the tube is compressed adiabatically and
is heated as it travels toward the closed end. During this high pres-
sure period, heat is rejected from the system in the hot end heat ex-
changer. In the basic pulse tube, gas in the tube is cooled by transfer-
ring heat with the tube wall, while in the orifice pulse tube the gas is
cooled by adiabatic expansion due to flow through the orifice. The pis-
ton then moves back and gas flows out of the tube and back through the
regenerator. Gas in the tube is then cooled due to adiabatic expansion
and gas flowing through the cold end heat exchanger absorbs heat from
the refrigeration load. The cycle ends with a low pressure period dur-
ing which gas in the tube is heated in the basic pulse tube by heat
transfer from the tube wall. In the orifice pulse tube, gas in the tube
is heated by adiabatic compression due to flow through the orifice. The
cycle results in an average enthalpy flow from the cold end to the hot
end which establishes a constant temperature gradient in the tube and
provides a continuous refrigeration effect.
Previous experimental work indicates that the orifice creates a
much greater enthalpy flow in the tube than heat transfer to the tube
wall. 1 Therefore, the subject of this study is the orifice pulse tube
and the purpose is to develop an analytical model of the refrigeration
power. Such a model will identify the various parameters which
influence performance and will be useful in optimization and design.
servoir
Volume
Piston
Pulse Tube
852
Pulse Tube
<!h. (2)
The energy balances reveal two interesting points about the pulse tube
refrigerator. First, unlike other refrigerators both the net
refrigeration, Qc, and the gross refrigeration, Qh, can be measured.
Second, since there is no addition of heat at any point along the tube,
the enthalpy flow must be constant in this region.
The average enthalpy flow over one cycle assuming an ideal gas is
given by
PHASOR ANALYSIS
853
The total mass flow rate is written in terms of the phasors for
flow at the cold end (x=O) d~e to the compressor <~c) and flow at the
hot end due to the orifice (~0 ),
-M = {1-X(x)}M
-c + X(x)M,
-o
(4)
where X(x) is the ratio of mass in the tube up to point x to the total
mass. The term X(x) satisfies the conditions X(O) = 0 and X(1) = 1 and
is a function of the temperature profile in the pulse tube. The phasor
diagram in figure 3a demonstrates how ~0 lags ~c by the phase angle e.
(7)
-sin -sin
-cos cos
sin sin
(a) (b)
Figure 3. Phasor diagram for (a) mass flow rate and (b) dynamic pressure.
854
-sin
where R is the gas constant. For small pressure ratios, 6P/P and 6T/T
are small and T and P are equal to the average values Ta and Pa
Equation (7) is then easily integrated over time to arrive at the phasor
representation of the dynamic temperature in terms of Pd and M, the mass
moved past point x, - -
(9)
An expression for cos6 is written in terms of the mass flow ratio using
figure 3b where Tcp is the compressor temperature and Tc is the cold end
temperature,
cos6 = P /P ( 10)
0 c
Then using the following equation for the mass flow rate at the cold end
where Vcp is the fixed swept (total) volume of the compressor and vis
the frequency,
.
m (P V /RT hv, ( 11)
c a cp cp
the enthalpy flow is put in the dimensionless form
855
The following equation is applied for a reversible, adiabatic pro-
cess with a small pressure ratio and zero dead volume between the com-
pressor and the tube,
where Vt is the pulse tube volume and Y is the ratio of heat capacities.
Two dimensionless forms for a fixed Vt in terms of the pressure ratio
and the volume ratio are
<H>IPa vt v ~(1/Y)(Pd/P )2
a
(mo/mc ), ( 15)
<H>IPa vt v (~/4)Y(V
cp
/Vt)2 (mo /mc ), (16)
The average enthalpy flow given in equations (13), (15), and (16) repre-
sent the analytical model for the gross refrigeration power produced by
the orifice pulse tube.
DISCUSSION OF RESULTS
40 40
Dependence of Refrigeration Power
Dependence of Refrigeration on Pressure Ratio
Power on Average Pressure
3: 3:
ri 30 Tc= 200 K ri 30 1c = 200 K
w w
v =6Hz 3: v =6Hz
3:
0 ~I~= 0.29 0 P8 = 1.0 MPa
Q. Q.
ril0 /rflc = 0.28 z rilofrilc = 0.26
z
0 20 0 20
1- i=
<( <(
a: a:
w w
(!) (!)
a: a:
u. 10 u. 10
w w
a: a:
(a) (b)
856
60
1.0 10 12
rilofrhc FREQUENCY, Hz
(a) (b)
CONCLUSIONS
The analytical model presented here for the orifice pulse tube
refrigerator results in a simple expression for the gross refrigeration
power which agrees with experiments regarding the dependence on all the
parameters. The magnitude predicted by the model is 3-5 times higher
than experiment and is presumably due to the simplifying assumptions
used in the model.
857
4001 ~--~----~--~----~--~
Temperature Profiles
in Pulse Tube
::.::
300
w
a:
::l
1-
<(
a:
w
Q. Figure 7. Average gas
::2 temperature as a
w
1- function of
position in the
orifice pulse
tube for differ-
ent cold end
temperatures.
0.2 0.4 0.6 0.8 1.0
RELATIVE POSITION
Figure 8. Comparison of
experimental
temperature wave
with theoretical
sine wave at the
cold and hot ends.
TIME,ms
858
REFERENCES
859