Beruflich Dokumente
Kultur Dokumente
90000094 R1 Page 1 of 21
IMPORTANT NOTICES
Read and understand this operating manual before installing or using the system.
Only use cables from H2scan with this system.
If this equipment is used in a manner not specified by H2scan, the protection provided by this equipment
may be impaired.
Hydrogen is flammable at 4% in air. Take indications seriously and be prepared to take action. In the event
of detection of 4% or higher of a hydrogen gas concentration there is a high probability of a hazard to
safety. Inform local emergency response personnel immediately.
LIMITATION OF LIABILITY
IN THE EVENT OF A DEFECT IN THE SYSTEM, H2SCAN SHALL NOT BE RESPONSIBLE FOR ANY DIRECT,
INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING THEREFORE, INCLUDING, BUT NOT
LIMITED TO, LOSS OF REVENUE AND/OR PROFIT.
LIMITED WARRANTY
H2scan Limited Warranty: Each hydrogen instrument (System) will conform, as to all substantial operational features, to the System
specifications set forth in this Manual and will be free of defects which substantially affect such Systems performance for twelve (12)
months from the ship date for such System.
Must Provide Notice of Defect: If you believe the System is defective, you must notify H2scan in writing, within ten (10) days of receipt of
such System, of your claim regarding any such defect.
Return of System to H2scan for Repair, Replacement or Credit: If the System is found defective by H2scan, H2scans sole obligation under
this warranty is to either (i) repair the System, (ii) replace the System, or (iii) issue a credit for the purchase price for such System, the
particular remedy to be determined [by H2scan] on a case-by-case basis.
Voided Warranty:. H2scans 12 Month Limited Warranty is void for any of the following:
The system is opened and the manufacturing seal is broken
Unauthorized repair work performed at the customers location or carried out by anyone other than H2scans factory trained
technicians
Equipment or parts that have been tampered with, misused, neglected, mishandled, improperly adjusted, or modified in any way
without the written consent of H2scan.
Equipment or parts that have been damaged due to shipping, misuse, accidents, mishandling, neglect, or problems with electrical
power sources.
Repair work performed during the warranty period does not extend the warranty beyond the originally stated duration.
System operation in incorrect or inappropriate environments.
Usage that is not in accordance with system guidelines or an operators failure to follow manual instructions.
Limitation of Warranty: THE ABOVE IS A LIMITED WARRANTY AS IT IS THE ONLY WARRANTY STATED BY H2SCAN. H2SCAN
MAKES NO OTHER WARRANTY EXPRESS OR IMPLIED AND EXPRESSLY EXCLUDES ALL WARRANTIES OF MERCHANTABILITY
AND FITNESS FOR A PARTICULAR PURPOSE. YOUR SOLE REMEDY HEREUNDER IS REPAIR OR REPLACEMENT OF THE
SYSTEM OR A CREDIT FOR THE PURCHASE PRICE FOR SUCH SYSTEM, THE PARTICULAR REMEDY TO BE DETERMINED BY
H2SCAN ON A CASE-BY-CASE BASIS. H2SCAN SHALL HAVE NO LIABILITY WITH RESPECT TO ITS OBLIGATIONS UNDER THIS
AGREEMENT FOR CONSEQUENTIAL, EXEMPLARY, OR INCIDENTAL DAMAGES EVEN IF IT HAS BEEN ADVISED OF THE
POSSIBILITY OF SUCH DAMAGES. THE STATED EXPRESS WARRANTY IS IN LIEU OF ALL LIABILITIES OR OBLIGATIONS OF
H2SCAN FOR DAMAGES ARISING OUT OF OR IN CONNECTION WITH THE DELIVERY, USE OR PERFORMANCE OF THE
SYSTEMS.
Limitation of Liability: SELLER SHALL UNDER NO CIRCUMSTANCES BE LIABLE FOR ANY INCIDENTAL, CONSEQUENTIAL, SPECIAL,
PUNITIVE, OR OTHER DAMAGES, INCLUDING, BUT NOT LIMITED TO, LOSS OF BUSINESS OR PROFIT, PROMOTIONAL OR
MANUFACTURING EXPENSES, INJURY TO REPUTATION, OR LOSS OF CUSTOMER, BASED ON ANY ALLEGED NEGLIGENCE,
BREACH OF WARRANTY, STRICT LIABILITY, BREACH OF CONTRACT, OR ANY OTHER LEGAL THEORY ARISING OUT OF THE
USE, MISUSE, PURCHASE, SALE OR POSSESSION OF ITS GOODS OR ITS PERFORMANCE OF THIS CONTRACT TO THE EXTENT
THAT SUCH LIABILITY EXTENDS SELLERS OBLIGATIONS BEYOND THE PRICE PAID BY BUYER TO SELLER FOR THE ITEM ON
WHICH SUCH CLAIM IS BASED. SELLER ADVISES BUYER TO PERFORM ACCEPTABLE TESTS ON ALL HARDWARE PRIOR TO
DEPLOYMENT AND TO PERFORM MAINTENANCE AS DESCRIBED IN THE SELLERS INSTRUCTION GUIDE. UNDER NO
CIRCUMSTANCES SHALL THE EQUIPMENT PROVIDED HEREUNDER BE USED IN A MANNER WHERE IT IS THE SOLE
PROTECTIVE SYSTEM FOR FACILITIES, EQUIPMENT AND PERSONNEL SAFETY; THE EQUIPMENT IS INTENDED FOR USE IN
CONUNCTION WITH OTHER APPROPRIATE PROTECTIVE SYSTEMS.
90000094 R1 Page 2 of 21
CONTENTS
1 DESCRIPTION ........................................................................................................................................ 4
2 SPECIFICATIONS ................................................................................................................................... 5
3 PARTS ..................................................................................................................................................... 6
4 INSTALLATION INSTRUCTIONS ........................................................................................................... 7
5 CONNECTIONS .................................................................................................................................... 10
6 STARTUP .............................................................................................................................................. 10
7 OPTIMUM PERFORMANCE ................................................................................................................. 10
8 PRESSURE EFFECTS.......................................................................................................................... 10
9 OPERATION MODE .............................................................................................................................. 10
9.1 NORMAL ..................................................................................................................................... 10
9.2 Inert Gas purge (option): ............................................................................................................. 10
9.3 Alarm ........................................................................................................................................... 11
9.4 Visual Status Indicator................................................................................................................. 11
10 TROUBLESHOOTING ......................................................................................................................... 12
11 SERIAL COMMUNICATION ................................................................................................................ 12
11.1 Software .................................................................................................................................. 12
11.2 Connection .............................................................................................................................. 12
11.3 Command Levels .................................................................................................................... 12
11.4 Command Summary ............................................................................................................... 13
12 SET ALARM PARAMETERS ............................................................................................................... 13
13 DISPLAY PRODUCT INFORMATION ................................................................................................. 14
14 DISPLAY STREAMING DATA ............................................................................................................. 14
15 BUMP TEST......................................................................................................................................... 15
16 CALIBRATION ..................................................................................................................................... 15
16.1 Factory Calibration .................................................................................................................. 15
16.2 Field Calibration ...................................................................................................................... 15
17 FIELD CALIBRATION .......................................................................................................................... 15
17.1 Gases ...................................................................................................................................... 15
17.2 Gas Connection ...................................................................................................................... 16
17.3 Procedure ............................................................................................................................... 16
17.4 Calibration Using the Serial Interface ..................................................................................... 16
18 CLEAR FIELD CAL .............................................................................................................................. 18
18.1 Clear Field Cal Using the Serial Interface............................................................................... 18
APPENDIX 1 - FOXTERM SETUP ............................................................................................................... 19
Installation .............................................................................................................................................. 19
Setup...................................................................................................................................................... 19
APPENDIX 2 FAULT DIAGNOSTIC CHART ............................................................................................. 20
90000094 R1 Page 3 of 21
H2scan 720AS-GC Hydrogen Sensor System
OPERATING MANUAL
1 DESCRIPTION
The Model 720AS-GC Hydrogen Sensor System is designed to detect and respond to hydrogen leaks in
Gas Chromatography systems. The hydrogen specific solid-state sensing element is designed for long term
reliable operation. The Model 720AS-GC detects hydrogen alarm indications at a user selectable level.
The factory pre-set level is 1.0% H2, or 25% LEL (Lower Explosive Limit). The alarm activates a tower with
flashing red lights and a buzzer and closes a valve to shut off the supply of hydrogen to the Gas
Chromatography system.
90000094 R1 Page 4 of 21
H2scan 720AS-GC Hydrogen Sensor System
OPERATING MANUAL
2 SPECIFICATIONS
Operating: 0C to 40C
Ambient Temperature
Storage: 0C to 50C
Hydrogen Sensitivity
0.5% to 5% hydrogen by volume at 1 ATM
Range
90000094 R1 Page 5 of 21
H2scan 720AS-GC Hydrogen Sensor System
OPERATING MANUAL
3 PARTS
Inspect the system and make sure the following parts are included.
720AS-GC Hydrogen Sensor System
Power Cord
Serial Interface Cable
Mounting screws, lock washers and nuts (5ea)
(Optional) Sensing Tube (See Figure 7)
(Optional) Sensing Tube Pass-Thru (See Figure 6)
(Optional) Sensing Tube Pass-Thru Nuts (2)
720AS-GC System Manual
Internal View of the 720AS-GC sensor system components is illustrated in figure 2 below:
GC CONNECTOR
PORT
BUZZER
ALARM
PUMP
POWER
SUPPLIES
SENSOR
SOLENOID
VALVES
CALIBRATION
PORT EXHAUST
90000094 R1 Page 6 of 21
H2scan 720AS-GC Hydrogen Sensor System
OPERATING MANUAL
4 INSTALLATION INSTRUCTIONS
Select the mounting location:
GC SIDE MOUNTED INSTRUCTIONS: Provide access to the GC oven. If knock-outs are available on the
GC side panel remove the desired knock-out. Align sensing tube hole with selected knock-out hole on GC
side panel. If no knock-out is available drill a 7/16 (12 mm) hole through left or right side panel and oven
wall.
NOTE: If drilling a hole the recommended position is 1 below the top of the GC oven and in a location that
will not interfere with column access. See Figure 3.
Remove the system cover mounting screws and set cover aside.
Insert threaded tube through hole in system back plate and into GC oven to align system. Level the system
and the mark location of 4 mounting holes on the GC side panel. Remove the GC side panel. Drill 4 holes
(only through the side panel) and mount the system to the GC using the screws provided. See figure 4.
90000094 R1 Page 7 of 21
H2scan 720AS-GC Hydrogen Sensor System
OPERATING MANUAL
90000094 R1 Page 8 of 21
H2scan 720AS-GC Hydrogen Sensor System
OPERATING MANUAL
Inside the Model 720AS-GC enclosure, disconnect the end of the 1/8 flex tube attached to the GC
Connect fitting. Connect the end of 1/8 flex tube to the sensing tube with 1/8 plastic fittings. For
reference, the flex tube will also be connected to a plastic T fitting.
Inspect the installation and ensure no debris or loose parts are present, and reinstall the cover. Figure 8
shows the complete setup of Model 720AS-GC installed on a gas chromatograph.
WALL MOUNTED INSTRUCTIONS: Use the mounting holes inside the Model 720AS-GC enclosure to
mount the system to a wall. Typical installations may hang the Model 720AS-GC on uni-strut mounted to a
wall behind the GC system.
Run a 1/8 tube from the external GC Connect port to the inside of a GC thermal chamber.
Inspect the installation and ensure no debris or loose parts are present, and reinstall the cover.
INLET GAS LINES SETUP: Two solenoid valves have been provided to accommodate control of the
hydrogen carrier gas and also the option of enabling an inert gas purge.
The hydrogen carrier gas shall be connected to the bottom solenoid valve marked H2 on the Model
720AS-GC. Ensure that the supplied hydrogen line is connected to the valve Inlet (facing forward) using
the provided 1/8 compression tube fitting. Connect the solenoid outlet (facing backward) to the Gas
Chromatograph carrier gas line in using the provided 1/8 compression tube fitting.
Optional Connect the supplied inert gas line to the topmost solenoid valve (the port facing to the right)
marked N2 using the provided 1/8 compression tube fitting.
NOTE: The solenoid valves will not work as intended if the inert gas line pressure exceeds the hydrogen
gas line pressure. The pressure of the nitrogen gas supply line must be less than the pressure of the
hydrogen gas supply line.
DISASSEMBLY: To remove the 720AS-GC sensor system from a GC or uni-strut, proceed through the
following instructions:
Unplug to power off and then disconnect the cable and plumbing connected to the solenoid valve.
Remove the front panel
Disconnect the flexible tubing from the sensing tube
Loosen the sensing tube pass-thru nuts inside the GC oven and remove the pass-thru along with the
sensing tube.
Remove the GC side-wall panel.
Remove the 720AS-GC sensor system from the GC side-wall panel.
90000094 R1 Page 9 of 21
H2scan 720AS-GC Hydrogen Sensor System
OPERATING MANUAL
5 CONNECTIONS
A standard 115 VAC instrument power cord is supplied.
A standard serial cable is provided so the unit can be monitored, controlled, and calibrated using standard
PC.
Hydrogen gas connection and inert gas connection (optional) to the solenoid valve are made using
standard 1/8 compression fittings.
6 STARTUP
Plug the standard instrument power cable into 720AS-GC sensor system. Once power is applied, the unit
executes a warm-up sequence lasting about 10 minutes. The status LED will be amber in color during
warm-up and will change to green when the unit is ready. The following operations will be performed:
Sensor stabilizes to its operating temperature
Perform the Power-On-Self-Test (POST)
The supply of hydrogen to the GC system is shut off.
At the completion of warm-up the hydrogen flow to GC is turned on.
Turn the pump switch ON. A green valve LED will activate indicating hydrogen is now allowed to flow.
NOTE: THE SYSTEMS OPERATIONAL AND OUTPUT SETTINGS HAVE BEEN CONFIGURED AT THE
MANUFACTURER. SETTINGS MAY BE CHANGED THROUGH THE SERIAL PORT.
7 OPTIMUM PERFORMANCE
The analyzer is designed to be exposed to an air background with no hydrogen most of the time. For
optimal performance keep any hydrogen exposure to the sensor to less than one hour. If the sensor is
exposed to high levels of hydrogen for a long period of time, it may need to be reconditioned by leaving it
powered on in air for a few days.
For maximizing the performance of the sensor:
Perform a Field calibration after the initializing sequence described in section 6 Start UP. See Section 17
for field calibration instructions.
8 PRESSURE EFFECTS
The analyser is hydrogen specific and sensitive to only the hydrogen partial pressure in the gas stream.
Since changes in total gas pressure will affect the hydrogen partial pressure, they will also affect the sensor
readings. For instance, at one atmosphere pressure, a 1% H2/Air mixture will be reported as 1% from the
unit. At 0.9 atm, the reading will decrease to 0.9%.
At the factory, the units are calibrated at one atmosphere pressure at sea level.
Performing field calibration will adjust H2 measurement to installation elevation and atmospheric pressure.
9 OPERATION MODE
9.1 NORMAL
The status LED will be green indicating 720AS-GC Hydrogen Sensor System is operating normally.
90000094 R1 Page 10 of 21
H2scan 720AS-GC Hydrogen Sensor System
OPERATING MANUAL
allow inert gas to flow. If an inert gas purge is not desired, the system will only cut off hydrogen gas
flow.
NOTE: The solenoid valves will not work as intended if the inert gas line pressure exceeds the
hydrogen gas line pressure. The pressure of the nitrogen gas supply line must be less than the
pressure of the hydrogen gas supply line.
9.3 ALARM
If the hydrogen level exceeds the set point, or there is an instrument fault, the towers red light will
light, buzzer will sound, and the supply of hydrogen to the GC system is shut off and inert gas
solenoid valve (if connected) is turned on and GC system will be using it as the carrier gas otherwise
there will be no carrier gas flow.
Turn off the hydrogen supply at the source.
Unplug the power cable to silence the alarm. The hydrogen supply will remain shut off.
Investigate and repair the cause of the alarm by referring to troubleshooting and diagnostic chart.
ALARM
ACTIVATED
INERT GAS
SOLENOID OPENS
H2 SUPPLY
SOLENOID SHUTS
WARNING: IF FAULT CONDITION (HIGH HYDROGEN EVENT, PUMP, SENSOR FAILURE) OCCURS THE
HYDROGEN (CARRIER GAS) SUPPLY TO GC WILL BE SHUT OFF.
90000094 R1 Page 11 of 21
H2scan 720AS-GC Hydrogen Sensor System
OPERATING MANUAL
10 TROUBLESHOOTING
The primary function of the 720AS-GC sensor system is to detect an unexpected buildup of hydrogen in the
GC oven and shut off the supply line to prevent an explosion.
Hydrogen supply (solenoid) to GC is shut off under the following scenarios:
720AS-GC sensor system is not functioning (no power or internal fault detected).
When a hydrogen event is detected the hydrogen supply is shut off and the alarm-light and
sound are turned on.
The only way to silence the alarm is to turn off the 720AS-GC sensor system by unplugging the AC power
line. The system should not be plugged in (turned on) until after whatever caused the hydrogen event has
been determined and corrective actions have been taken. It is highly recommended to connect the Model
720AS-GC Hydrogen Sensor System to a UPS (Backup Battery) to prevent power spikes and inconvenient
resets.
11 SERIAL COMMUNICATION
The sensor can be monitored and controlled via the serial interface.
11.1 SOFTWARE
Software that communicates over the serial port is typically called a terminal emulator. Any two-way
terminal emulators (HyperTerminal, Telnet, etc.) or purpose-built software (using LabView, Visual
Basic, C++, etc.) can be used to establish serial communication with the unit.
XP and previous versions of Windows came with a terminal emulator called Hyper Terminal. Newer
versions of Windows do not. These versions require installing a third-party program. There are many
inexpensive or free programs available. FoxTerm, a free program, is used extensively by H2scan.
Refer to appendix 1 for setup instructions.
11.2 CONNECTION
Connect the sensor to the serial port, using direct COM port or USB adapter.
If a USB adapter is used, install the driver according to its instructions.
In Windows, look at the ports using the Device Manger.
A built-in port will be COM1.
The port assigned to the USB adapter could have any number, but it will be called something like
USB Serial Port (COM7) or I-756X (COM3). Below is an example.
.
The RETURN or ENTER key is the last character of the command string. If either key is pressed
without a command string the result is an invalid command and will resume continuous display if in
Level 0 or return to prompt if in Level 1.
Level 0 Commands
Keystroke Description
Stops continuous display to enter a password or command. If in level zero, the
ESC
continuous display will resume after executing one command.
Pressing the Space key while the serial output is active will display a label line
sp(spacebar)
showing the heading for each column of data.
Enter the password to change security level. A null or invalid password returns to
the default security level.
=<password>
Level 0 password = 0
Level 1 password = h2scan
Level 1 Commands
Keystroke Description
A Set the alarm parameters
Display Product Information. Enter page number 0 or 1
D <page> 0 Product information
1 User configuration
F Field Calibration
G Start or resume the sensor operation
X Clear field calibration data (returns to last factory calibration data).
WARNING: IF SETTINGS ARE CHANGED FROM THOSE SET BY THE MANUFACTURER THEN IT IS THE
USERS RESPONSIBILITY TO UNDERSTAND THE IMPLICATIONS OF THESE CHANGES.
90000094 R1 Page 13 of 21
H2scan 720AS-GC Hydrogen Sensor System
OPERATING MANUAL
H2scan: d 1
User configuration is:
Hydrogen reporting range 0.4-5.0000% H2
Isolated Output is enabled: 4.000 to 20.000 mAmps
Error output is 3.000 mAmps
Not-Ready output is 2.000 mAmps
Voltage Output is disabled
Alert threshold is 0.8% Hydrogen; Fault Indication is enabled
Alarm threshold is 0.8% Hydrogen; Fault Indication is enabled
Alert/Alarm outputs are latched
90000094 R1 Page 14 of 21
H2scan 720AS-GC Hydrogen Sensor System
OPERATING MANUAL
To have the Overall Hydrogen Reading, Peak Hydrogen, Calculations Errors and Heater State columns
continuously reported, type the following G serial command at the H2scan: command prompt:
H2scan: G 03 06
; %H2 %H2_Pk Messages
5.29087 5.29087
5.29610 5.29610
5.28849 5.29610
To only have the overall Hydrogen column continuously reported, type the following G serial command at
the H2scan: command prompt:
H2scan: G 02
; %H2
1.24837
1.25051
1.24814
15 BUMP TEST
A bump test can be performed at any time by connection through the calibration port. Turn off the system
pump (Figure 8, this will close the solenoid valve) and supply an H2/air concentration sufficient to trigger
the alarm. For example using the default 1% alarm level, 2%H2/air (recommended) works well. The alarm
will trigger when the sensor measures H2 over the set point. If the unit does not, perform a field calibration.
Make sure to turn the pump on when testing or when calibration is complete.
16 CALIBRATION
Calibrations do not cause any wear on the sensor and can be accomplished as often as desired by
connecting a certified concentration of H2/air through the calibration port. Analog outputs can be monitored.
H2scan suggests that the calibration of the system be carried out every 90 days.
17 FIELD CALIBRATION
17.1 GASES
Calibration requires the availability of one or two certified known gases at values specific to the
application. One gas will correct any offset error while two gasses will also correct offset and gain
error. It is recommended to use 0.5%H2/air and 2% H2/air to obtain superior calibration accuracy.
DO NOT use 0% H2 as the lower calibration gas. The recommended flow rate required is 0.5 0.2
slpm.
Recommended time is 10 min.
Calibrate the sensor in an air background.
Custom Field Calibration Kits are available from H2scan.
90000094 R1 Page 15 of 21
H2scan 720AS-GC Hydrogen Sensor System
OPERATING MANUAL
The accuracy of the gas bottle concentration will directly affect the measured accuracy by the units.
During factory calibration, the units are calibrated with high accuracy gases (as high as +/- 0.02%). It
is strongly recommended that similar high accuracy gases are used to maintain the highest
accuracy.
17.3 PROCEDURE
Before performing calibration please turn pump off (see figure 10). During the Field Calibration
process any previously completed Field Calibrations are cancelled. As a result, during the routine the
unit may display a hydrogen concentration that is different from the applied gas concentration. This is
normal. Once the procedure is completed, the readings will be corrected to display accurately for all
subsequent exposures.
PUMP SWITCH
CALIBRATION
PORT
90000094 R1 Page 16 of 21
H2scan 720AS-GC Hydrogen Sensor System
OPERATING MANUAL
Streaming data
Streaming data
Follow the instructions in Section 12 of this manual, set both of the Alarm Thresholds back
to the desired alarm activation threshold (usually 1.0%).
Turn on the pump.
90000094 R1 Page 17 of 21
H2scan 720AS-GC Hydrogen Sensor System
OPERATING MANUAL
90000094 R1 Page 18 of 21
H2scan 720AS-GC Hydrogen Sensor System
OPERATING MANUAL
SETUP
Start FoxTerm
Close the default session window (if needed).
Open a new session window.
Select the correct port as determined above
(COM3 in this example).
Setup the session as shown below
Select a log file name and location. The file
name should start with the current date
(YYYMMDD) followed by any particular
information required. This way, the files will be
easy to sort. For example
20110920Yukon6.log would be the file name
for the Yukon6 sensor that had logging started
September 20th, 2011. The .log extension is
the default, but any extension could be used.
Newline Behavior must be set to CRLF.
Click OK.
Save the session as H2scan.xml in the
FoxTerm program location.
The setup should look similar to that shown.
90000094 R1 Page 19 of 21
H2scan 720AS-GC Hydrogen Sensor System
OPERATING MANUAL
90000094 R1 Page 20 of 21
H2scan 720AS-GC Hydrogen Sensor System
OPERATING MANUAL
QUESTIONS?
PLEASE CONTACT US AT
90000094 R1 Page 21 of 21