Beruflich Dokumente
Kultur Dokumente
Water Research
journal homepage: www.elsevier.com/locate/watres
a r t i c l e i n f o a b s t r a c t
Article history: For the elimination of potentially harmful micropollutants, powdered activated carbon (PAC) adsorption
Received 25 January 2016 is applied in many wastewater treatment plants (WWTP). This holds the risk of PAC leakage into the
Received in revised form WWTP efuent and desorption of contaminants into natural water bodies. In order to assess a potential
27 May 2016
PAC leakage, PAC concentrations below several mg/L have to be detected in the WWTP efuent. None of
Accepted 3 June 2016
Available online 6 June 2016
the methods that are used for water analysis today are able to differentiate between activated carbon and
solid background matrix. Thus, a selective, quantitative and easily applicable method is still needed for
the detection of PAC residues in wastewater.
Keywords:
Micropollutants
In the present study, a method was developed to quantitatively measure the PAC content in waste-
Wastewater water by using ltration and thermogravimetric analysis (TGA), which is a well-established technique for
Powdered activated carbon the distinction between different solid materials. For the sample ltration, quartz lters with a tem-
Thermogravimetric analysis perature stability up to 950 C were used. This allowed for sensitive and well reproducible measure-
ments, as the TGA was not affected by the presence of the lter. The samples mass fractions were
calculated by integrating the mass decrease rate obtained by TGA in specic, clearly identiable peak
areas. A two-step TGA heating method consisting of N2 and O2 atmospheres led to a good differentiation
between PAC and biological background matrix, thanks to the reduction of peak overlapping. A linear
correlation was found between a samples PAC content and the corresponding peak areas under N2 and
O2, the sample volume and the solid mass separated by ltration. Based on these ndings, various
wastewater samples from different WWTPs were then analyzed by TGA with regard to their PAC content.
It was found that, compared to alternative techniques such as measurement of turbidity or total sus-
pended solids, the newly developed TGA method allows for a quantitative and selective detection of PAC
concentrations down to 0.1 mg/L. The method showed a linearity coefcient of 0.98 and relative standard
deviations of 10%, using small water sample volumes between 0.3 and 0.6 L.
2016 Elsevier Ltd. All rights reserved.
http://dx.doi.org/10.1016/j.watres.2016.06.008
0043-1354/ 2016 Elsevier Ltd. All rights reserved.
ver et al. / Water Research 101 (2016) 510e518
T. Krahnsto 511
Table 1
Characteristics of the WWTPs for water sampling and PAC quantication.
WWTP Location Population Origin municipal: PAC PAC separation Plant type DOC concentration in the inuent to the PAC
equivalents industrial dosage treatment
Ergolz 1 Sissach 40,000 100:0 12 mg/L Flocculation deep bed Technical 6.7 mg/L
ltration scale
Flotation surface ltration Pilot scale 6.7 mg/L
Bachwis Herisau 35,000 50:50 15 mg/L Sedimentation deep bed Full scale 13.2 mg/L
ltration
Basel Basel 250,000 90:10 15 mg/L Membrane ltration Pilot scale 11.6 mg/L
matrix prior to oxidation (Table 2). The resulting two-stage method is the known ratio of oxidative and pyrolytic decomposition of
was compared to the conventional approach during method background matrix and with
development, and was then applied for sample analysis during the
method application stage.
mO2
FTGA (5)
The sample weight msample and temperature T were digitally mO2 mN2
logged during the whole TGA measurement and were then
analyzed with the TGA software (PerkinElmer, Pyris Software, being a quantitative variable that can be obtained from the TGA
version 2009). For method development purposes, the sample curve analysis, the samples PAC content can be expressed by
degradation was assessed using both the curve msample vs. tem- inserting (2)e(5) into (1), which yields the linear equation:
perature and its time-derivative m_ sample vs. temperature, in order to
XTGA 1 FMatrix $FTGA FMatrix : (6)
compare both options. During method application, the samples
were analyzed using the derivative curve. On the other hand, with regard to the sample preparation, the
samples theoretical PAC content XSample can be calculated as:
2.2. Calculation of the PAC content of a sample.
CPAC $V
Xsample (7)
In order to quantitatively assess the PAC content of a sample, the mtot
curve m_ sample vs. time during combined N2 and O2 treatment was
with CPAC being the concentration of powdered activated carbon in
used. The mass loss was calculated by integration of the clearly
the water sample, V the ltrated water volume and mtot the mass of
identiable mass decrease peaks in the specic time interval. A
solids separated by ltration. Based on a mass balance, it is
dimensionless variable XTGA was then dened as an indicator for
postulated that the TGA parameter XTGA correlates with the PAC
the samples PAC fraction:
content Xsample in a water sample:
mPAC
XTGA (1) Xsample XTGA $FDegr ; (8)
mPAC mMatrix; O2 mMatrix; N2
where FDegr is a measure for the degree of sample degradation
with mPAC being the mass decrease caused by PAC oxidation, while
during TGA. Hence, by inserting (6) and (7) into (8), a water sam-
mMatrix; O2 and mMatrix; N2 are the background mass decreases under
ples PAC concentration can be calculated from a TGA curve as:
O2 and N2, respectively. While the pyrolytic mass decrease is
entirely caused by background matrix: FDegr $mtot
CPAC $1 FMatrix $FTGA FMatrix (9)
mMatrix; N2 mN2 ; (2) V
The values for FDegr and FMatrix were obtained by measuring
the oxidative mass decrease is composed of both activated carbon several samples of known composition, as described in the results
oxidation and the oxidation of background matrix (Fig. 4): and discussion section.
mMatrix; O2 In the second stage of this study, wastewater samples from three
FMatrix (4) Swiss WWTPs were analyzed in triplicates using TGA and
mMatrix; N2
comparative methods (Fig. 1). All of these plants comprise a me-
chanical pre-treatment, a biological sludge treatment and a
Table 2
Temperature programs of the TGA heating methods.
1 Heat from 30 C to 900 C with 5 C/min 20 mL/min oxygen 1 Heat from 30 C to 350 C with 5 C/min 20 mL/min nitrogen
2 Cool down to 30 C 20 mL/min oxygen 2 Constant at 350 C for 300 min 20 mL/min nitrogen
3 Cool down to 30 C 20 mL/min nitrogen
4 Constant at 30 C for 300 min 20 mL/min oxygen
5 Heat from 30 C to 900 C with 5 C/min 20 mL/min oxygen
6 Cool down to 30 C 20 mL/min oxygen
ver et al. / Water Research 101 (2016) 510e518
T. Krahnsto 513
PAC
Primary and Deep bed
FeCl(SO4)
secondary treatment filtration
Tertiary
A Wastewater
effluent
PAC stage
PAC
Polymer Belt
filter
Tertiary
effluent
PAC flotation
PAC
Primary and
FeCl(SO4)
secondary treatment Sedimentation
Tertiary
B Wastewater
effluent
PAC stage Deep bed
filtration
PAC recirculation
Primary and
PAC Membrane
secondary treatment Ozonation
filtration
Tertiary
C Wastewater
effluent
PAC stage
O3
Sampling point
Fig. 1. Flow schemes of the PAC processes on WWTPs Ergolz 1 (A), Bachwis (B) and Basel (C) with sampling points (grey).
Temperature [C]
0 100 200 300 400 500 600 700 800 900
6 0.0
5
-0.1
sample [mg/min]
4
msample [mg]
-0.2
3
pure PAC
-0.3
2
background + 2 mg/L PAC
1 -0.4
background without PAC
0 -0.5
0 20 40 60 80 100 120 140 160 180
Time [min]
Fig. 2. Typical TGA curves of samples with varying composition (centrifuged, O2 method).
0
msample [mg]
5
4 -0.05
3 -0.1
2
-0.15
1 A B
0 -0.2
0 50 100 150 0 50 100 150
Time [min] Time [min]
Fig. 3. TGA curves of triplicates prepared by centrifugation (A) and ltration (B) (2 mg/L PAC, O2 method).
514 ver et al. / Water Research 101 (2016) 510e518
T. Krahnsto
7
0
6.5
sample [mg/min]
msample [mg] 6
5.5 -0.05
5
4.5
A B
4 -0.1
0 50 100 0 50 100 150
Time [min] Time [min]
Fig. 4. TGA curves of samples analyzed with the combined method under N2 (A) and O2 (B) (ltrated).
secondary clarication followed by a PAC adsorption step. The PAC specic temperature of this mass decrease is found to vary
processes of these plants mainly differ in their volume ow ca- depending on the sample composition, but could be clearly iden-
pacity, the origin of the treated wastewater and the method used tied in all cases. Finally, at temperatures above 700 C, the sample
for PAC separation (Table 1). The PAC dosage was at 12e15 mg/L in mass reaches a nearly constant value which corresponds to the ash
all cases. content of the sample. In the case of a constant temperature
Several alternative techniques were applied in order to assess gradient, the temperature axis can be directly correlated to a time
the suspended solids content of wastewater samples and to vali- axis. In order to take into account mass transport and reaction ki-
date and compare the TGA method developed in this study. netics and allow for the evaluation of the mass loss rate, the time
Turbidity measurements were performed in glass vials with a axis is of particular interest for analyzing a TGA curve.
portable turbidimeter (WTW, Turb 355 IR/T) and the concentration
of particles (>1 mm) was measured with an online particle counter 3.2. Methods for sample preparation
(PAMAS, OLS50P). The amount of total suspended solids was
determined according to DIN 38 409-2 using cellulose nitrate lters In the studies performed by Vu et al. (2012) and Vogel (2013),
(Macherey-Nagel, PORA NC) and a precision scale (Mettler Toledo, the solids were concentrated from sewage sludge samples by
XS 205). Additionally, the lter color after ltration and drying was evaporation. However, those samples contained around 10 g/L solid
visually compared to reference lters, as proposed by Metzger et al. matter, while in the here presented study, the solids content of the
(2011). water samples was always below 0.1 g/L. In preliminary experi-
ments, it was found that vaporization does not allow for the com-
3. Results and discussion plete recovery and mass balance of solid matter in this
concentration range. Thus, two alternative methods were tested for
3.1. Qualitative evaluation of TGA curves sample preparation, namely centrifugation and ltration, and were
then compared with regard to their practical feasibility and
The TGA with a heating step in oxygen atmosphere showed a reproducibility. The loss of solids during the decanting, centrifu-
characteristic curve that is in accordance with the results of pre- gation and drying procedure was found to be considerable: On
vious studies (Vu et al., 2012; Otero et al., 2008). The behavior of a average, the centrifugation of 2.2 L water yielded the same solid
sample exposed to a TGA heating method is characterized by de- sample mass as the ltration of 1.0 L. This incomplete solid sepa-
creases in sample mass msample, which each correspond to a peak of ration increases the workload and calls for an additional, reliable
the mass loss rate m _ sample (Fig. 2). In many cases, the sample mass determination of the samples solids content. In contrast, in the
decreases slightly during the rst minutes of measurement at case of ltration, a complete removal of solids was achieved and the
temperatures lower than 150 C. This effect is related to residual separated solid mass correlated well with the TSS, making an
moisture in the lter and can be clearly differentiated from any additional determination redundant. Furthermore, it was observed
other mass decreases at higher temperatures. A signicant decrease that the solid separation by centrifugation is not always consistent
is observed in the temperature range 200 Ce400 C. This can be for biological matrix and activated carbon. By qualitatively
explained by the volatilization and subsequent oxidation of comparing the TGA curves obtained with these methods, it was
wastewater background matrix. Wastewater background matrix found that a better reproducibility could be achieved if the water
consists of carbohydrates, proteins and fatty acids which are char- samples were ltrated rather than centrifuged (Fig. 3). If the water
acterized by a high fraction of volatiles. Thermal treatment results sample is ltrated, the solid sample partly consists of lter material,
in a decomposition of macromolecules into smaller volatile mole- which reduces the degree of sample decomposition and causes a
cules (Basu, 2010). Evaporation of these molecules is detected as a slight mass decrease above 700 C. However, at this temperature,
mass loss in the TGA curve. Another sharp mass decrease at 400 C the mass decrease attributed to activated carbon and biological
e 700 C is caused by the oxidation of powdered activated carbon. background matrix is already completed. This means that the
Activated carbon consists of a network of carbon rings and contains oxidation of lter material can be clearly differentiated from the
negligible amounts of volatiles, thus oxidation reactions are gov- sample oxidation and that the presence of the lter during analysis
erned by solid-gas reactions (Marsh and Rodrguez-Reinoso, 2006). does not affect the analyzability of the TGA curve.
At temperatures below 500 C, the reaction rate is kinetically
controlled by a high activation energy (Luo and Jean-Charles, 2006). 3.3. TGA heating methods
It is assumed that, at 600e700 C, reaction kinetics increase rapidly
and are then controlled by in-pore oxygen and carbon monoxide For the analysis of solid samples, two different TGA methods
diffusion which, however, increases at progressing burn-off. The were applied. The O2 method, in which the sample is exposed to
ver et al. / Water Research 101 (2016) 510e518
T. Krahnsto 515
Xsample [mg/L ]
fraction. For this reason, the N2/O2 method was developed, which
leads to partial pyrolysis in a rst step under N2. Due to its chemical
composition, PAC is not expected to decompose at the applied 0.3
conditions (Marsh and Rodrguez-Reinoso, 2006), which was
veried by the present experiments (Fig. 4 A). The mass decrease 0.2
caused by pyrolytic reactions can therefore be selectively attributed
to the decomposition of water background matrix. In contrast, the 0.1
simultaneous oxidation of biomass and PAC results in an over-
lapping mass decrease under O2 atmosphere (Fig. 4 B). Still, the 0.0
mass decrease in this second heating step can be used as a measure 0.0 0.2 0.4 0.6 0.8 1.0
for the amount of oxidable solids in the TGA, including both
FTGA [ - ]
background matrix and PAC. This allows to differentiate between
the solids extracted from the original water sample and the lter Fig. 5. FTGA and Xsample obtained from samples of known composition, tted to a linear
material, which does not degrade under these conditions. If both correlation.
heating steps are taken into account, the combined N2/O2 heating
method allows for the identication of lter material, background
matrix and PAC, and for the quantication of the samples 2.0
composition by simply determining the pyrolytic and oxidative
mass decreases during TGA.
1.5
FDegr $mtot
CPAC $1 FMatrix $FTGA FMatrix (10)
V 0.5
with FTGA being obtained from the TGA curve analysis, mtot and V
being parameters of the sample preparation step and FDegr and
FMatrix being dimensionless characteristics of PAC and background
0.0
matrix, respectively. The values obtained from several samples of 0 0.5 1 1.5 2
known composition were found to correlate with a linear trendline CPAC dosed[ mg/L ]
(n 19). If this trendline is assumed to satisfy equation (10) with
Fig. 6. PAC detection by TGA (CPAC measured) in samples that contained a known
FDegr $1 FMatrix being the slope and FDegr $FMatrix being the Y-
amount of PAC (CPAC dosed).
intercept, it is possible to calculate the values (Fig. 5):
FMatrix 0:55 TGA curves due to sample inhomogeneities. The calculations based
on the tted values for FMatrix and FDegr correlated well with the
and
known sample composition (R2 0.959) and showed a Pearson
FDegr 0:60: correlation coefcient of 0.98, indicating a good method linearity.
However, it should be noted that the tted parameters could vary
In order to validate the developed method, CPAC was calculated for efuents from different WWTPs, e.g. depending on the nitrogen
for samples with known PAC concentrations between 0 and 2 mg/L content of the sewage sludge (Go mez-Rico et al., 2005). It is thus
and compared to the actual PAC dosage (Fig. 6). The method recommended for future measurements on different WWTPs to
repeatability was assessed on the basis of the standard deviations rst determine FMatrix and FDegr before applying the calculation
of triplicate measurements at each concentration. The average method to TGA curves.
standard deviation was 0.04 mg/L, which corresponds to 10% of
the actual concentration, and it was observed that the relative
deviation decreased with increasing PAC concentration. Taking the 3.5. Method application for PAC retention assessment
background noise into account, the methods detection limit was
estimated to be around 0.1 mg/L. The average recovery of PAC was The TGA method was applied to assess and compare the PAC
83.8%, which indicates a satisfactory method accuracy. However, retention of four different separation technologies (Fig. 1). The re-
the accuracy was notably inuenced by the sample preparation sults give an outlook on the targeted application of the developed
step, where the complete and reproducible separation of solids is method and on the range of PAC concentrations in WWTP efuents.
the determining factor. It is expected that the increase of the The PAC removal of the nal separation stage was above 99% in all
sample volume and the optimal selection and conditioning of the cases, resulting in efuent concentrations close to the detection
lter could increase the PAC recovery in the sample preparation limit (Fig. 7).
step in particular. The calculation of FTGA from TGA peak areas was On WWTP Ergolz 1, two process options for PAC removal were
found to be relatively robust, even in the case of slight variations in compared. It was found that by applying otation, the PAC
516 ver et al. / Water Research 101 (2016) 510e518
T. Krahnsto
1.2
1.0
A B C 17.5
0.8
PAC [mg/L]
0.6
0.4
0.2
0.0
Secondary Filtration Flotation Flotation+ Secondary Sediment. Filtration Ozonation PAC Filtration
effluent effluent effluent filtration effluent effluent effluent effluent reactor effluent
effluent
Fig. 7. Characterisation of wastewater samples taken from the WWTPs Ergolz 1 (A), Bachwis (B) and Basel (C) by N2/O2 TGA (n 3).
concentration can only be reduced to 0.4 mg/L. Thus, in this case, an methods, the wastewater samples from several WWTPs with PAC
additional nal ltration step is necessary to achieve the same treatment were analyzed not only by TGA, but also by particle
efuent quality as by deep bed ltration. The PAC removal was counting, turbidity and TSS measurements, as well as visually after
above 99% for both process options (Fig. 7 A). Metzger et al. (2011) (Fig. 8).
On WWTP Bachwis, the PAC separated in the sedimentation step The visual analysis of laboratory lters showed no blackness at
is fed back into the biological treatment and discharged together all, which corresponds to the absence of PAC according to Metzger
with the waste activated sludge (Fig. 1). However, it was found that et al. (2011). This was the case for all sampled PAC stage efuents
some PAC residues pass through the secondary clarier, resulting in (Fig. 8), indicating that the visual method is not sensitive enough to
a PAC concentration of 0.5 mg/L in the secondary efuent. As on detect PAC traces in wastewater. The results are considered semi-
WWTP Ergolz 1, the efuent still contains PAC in concentrations quantitative and of limited accuracy, so that this method is not
close to the detection limit, corresponding to a PAC removal of recommended for PAC detection in WWTP efuents.
above 99% (Fig. 7 B). The TSS removal of the studied processes was between 76% and
The secondary efuent on WWTP Basel is rst treated in a pilot 91% and always lower than the corresponding PAC removals (Figs. 7
scale ozonation unit before entering the PAC stage. In the adsorp- and 8). On WWTP Ergolz I, the TSS and turbidity measurements
tion reactor, a PAC concentration of 17.5 mg/L was measured, which qualitatively conrmed that the otation and subsequent surface
is nearly consistent with the targeted PAC dosage of 15 mg/L. The ltration yields a similar efuent quality as the deep bed ltration
ultraltration unit ensured a full PAC removal to values below the with direct PAC dosage. However, only 2% of the solids leaking
limit of detection (Fig. 7 C). through the lter were attributed to PAC (Fig. 8A). In the deep bed
ltration of WWTP Bachwis, comparable reductions of TSS (72%)
and of turbidity (67%) were observed with regard to the sedimen-
3.6. Method evaluation and comparison tation efuent, while the TGA method revealed a PAC retention by
almost 95% (Figs. 7 and 8B). As the ltration step on WWTP Basel
In order to compare the developed TGA method to reference
Fig. 8. Characterisation of wastewater samples taken from the WWTPs Ergolz 1 (A), Bachwis (B) and Basel (C) by different methods (n 3).
ver et al. / Water Research 101 (2016) 510e518
T. Krahnsto 517
was realized by ultraltration the TSS concentration of 3.8 mg/L in Ofce for the Environment for funding this study under contract
the ltration efuent was surprisingly high (Fig. 8C). However, no number UTF 450.06.13/IDM 2004.2423.391. Special thanks go to
PAC was detected on the permeate side of the membrane unit. In all Felix Thiele, Antje Langbein and Jan Stemann for their helpful
separation processes, the percentage removal value of TSS was collaboration and valuable advice.
found to be lower than that of PAC. This shows clearly that the
determination of standard parameters such as TSS or turbidity is References
not sufcient to assess PAC leakage into WWTP efuent. The main
disadvantage of the reference methods is that they represent the Abegglen, C., Siegrist, H., 2012. Mikroverunreinigungen aus kommunalem
Abwasser. Verfahren zur weitergehenden Elimination auf Kla ranlagen. Bunde-
total concentration of solids in the water without differentiating samt fr Umwelt, Bern.
between PAC and background matrix. However, combining TGA Altmann, J., Ruhl, A.S., Sauter, D., Pohl, J., Jekel, M., 2015a. How to dose powdered
and TSS results reveals that the leaking solids on WWTP Ergolz 1 activated carbon in deep bed ltration for efcient micropollutant removal.
Water Res. 78, 9e17.
and on WWTP Bachwis are partially composed of PAC, while the
Altmann, J., Sperlich, A., Jekel, M., 2015b. Integrating organic micropollutant
efuent of the pilot plant on WWTP Basel contains no PAC at all. removal into tertiary ltration: combining PAC adsorption with advanced
This example shows very well that conventional parameters are no phosphorus removal. Water Res. 84, 58e65.
suitable indicators for PAC in wastewater, and that the PAC speci- Basu, P., 2010. Biomass Gasication and Pyrolysis - Practical Design and Theoriy.
Elsevier, ISBN 978-0-12-374988-8.
city of the developed TGA method is an important contribution to Bohler, M., Zwickenpug, B., Siegrist, H., Hollender, J., Joss, A., Ternes, T., 2012.
assessing the PAC retention of ltration processes. Removal of micropollutants in municipal wastewater treatment plants by
TGA curves obtained from samples of different WWTPs showed powder-activated carbon. Water Sci. Technol. 66.10, 2115e2121.
Bolong, N., Ismail, A.F., Salim, M.R., Matsuura, T., 2009. A review of the effects of
the same characteristic behavior and intervals of mass losses, emerging contaminants in wastewater and options for their removal. Desali-
indicating that the TGA measurement was not affected by the origin nation 239.1e3, 229e246.
or the characteristics of the sampled wastewater or the applied PAC Brown, M., 2001. Introduction to Thermal Analysis. Kluwer Academic Publishers,
ISBN 978-1-4020-0211-3.
product. However, variations of FDegr and FMatrix may explain the Coats, A., Redfern, J., 1963. Thermogravimetric analysis. A review. Analyst 88,
slightly higher standard deviations in Bachwis and Basel compared 906e924.
to WWTP Ergolz I, the water of which was used for method Eidgeno ssisches Departement fr Umwelt, Verkehr, Energie und Kommunikation,
Revision der Gewa sserschutzverordnung fr bessere Wasserqualit at, [Online].
development. Available: https://www.news.admin.ch/message/index.html?langde&msg-
From a practical point of view, the developed method was found id55786. [accessed 25.09.15.].
to be well workable under laboratory conditions. It takes around Folgueras, B., Daz, R., Xiberta, J., Prieto, I., 2003. Thermograimetric analysis of the
co-combustion of coal and sewage sludge. Fuel 82, 2051e2055.
18 h to carry out one measurement, which is mainly due to lter mez-Rico, M.F., Font, R., Fullana, A., Martn-Gullo n, I., 2005. Thermogravimetric
Go
drying (minimum 12 h, e.g. over night) and TGA analysis (5.5 h). study of different sewage sludges and their relationship with the nitrogen
However, laboratory staff is only required for sample ltration, lter content. J. Anal. Appl. Pyrolysis 74, 421e428.
weighing and TGA curve analysis, which add up to an effective Jean-Charles, X.R., 2006. Research of oxidation properties of graphite used in HTR-
10. Inst. Nucl. New Energy Technol. 20e45.
workload of 15e20 min per sample. In the development stage, the Jekel, M., Ruhl, A.S., Meinel, F., Zietzschmann, F., Pugmacher Lima, S., Baur, N.,
proposed method required visual peak interpretation which, Wenzel, M., Gnir, R., Sperlich, A., Dnnbier, U., Bo ckelmann, U., Hummelt, D.,
however, most likely can be automized once more data sets are von Baar, P., Wode, F., Petersohn, D., Grummt, T., Eckhardt, A., Schulz, W.,
Heermann, A., Reemtsma, T., Seiwert, B., Schlittenbauer, L., Lesjean, B.,
available. Miehe, U., Remy, C., Stapf, M., Mutz, D., 2013. Anthropogenic organic micro-
pollutants and pathogens in the urban water cycle: assessment, barriers and
4. Conclusion risk communication (ASKURIS). Environ. Sci. Eur. 25, 1e8.
Leusch, F., Khan, S., Laingam, S., Prochazka, E., Froscio, S., Trinh, T., Chapman, H.,
Humpage, A., 2014. Assessment of the application of bioanalytical tools as
In the present study, a method was developed to quantitatively surrogate measure of chemical contaminants in recycled water. Water Res. 49,
measure the PAC content in wastewater by using thermogravi- 300e315.
wenberg, J., Zenker, A., Baggenstos, M., Koch, G., Kazner, C., Wintgens, T., 2014.
Lo
metric analysis (TGA). Prior to TGA, the samples were ltered using
Comparison of two PAC/UF processes for the removal of micropollutants from
a quartz lter with a temperature stability of up to 950 C, so that wastewater treatment plant efuent: process performance and removal ef-
the separated solids could be analyzed by TGA without the mea- ciency. Water Res. 56, 26e36.
wenberg, J., Zenker, A., Krahnsto
Lo ver, T., Bo hler, M., Baggenstos, M., Koch, G.,
surement being affected by the presence of the lter. Based on the
Wintgens, T., 2016. Upgrade of deep bed ltration with activated carbon dosage
conducted laboratory experiments, a mathematical correlation was for compact micropollutant removal from wastewater in technical scale. Water
found for the calculation of a samples PAC content from the cor- Res. 94, 246e256.
responding TGA curve. In comparison to alternative methods, such Magdeburg, A., Stalter, D., Schlsener, M., Ternes, T., Oehlmann, J., 2014. Evaluating
the efciency of advanced wastewater treatment: target analysis of organic
as measuring turbidity, suspended solids or particle concentration, contaminants and (geno-)toxicity assessment tell a different story. Water Res.
the developed method combining vacuum ltration and TGA is able 50, 35e47.
to selectively differentiate between PAC and background particle Margot, J., Kienle, C., Magnet, A., Weil, M., Rossi, L., de Alencastro, L.F., Abegglen, C.,
Thonney, D., Che vre, N., Scha rer, M., Barry, D., 2013. Treatment of micro-
matrix, which allows for assessing the retention of processes with pollutants in municipal wastewater: ozone or powdered activated carbon? Sci.
regard to PAC in particular. Different separation processes were Total Environ. 461-462, 480e498.
shown to achieve PAC removals between 95% and 100%. The TGA Marsh, H., Rodrguez-Reinoso, F., 2006. Activated Carbon. Elsevier, ISBN 0-08-
044463-6.
method yields quantitative results with a detection limit around Metzger, S., Ro ssler, A., Kapp, H., 2011. Erweiterung kommunaler Kla ranlagen durch
0.1 mg PAC/L. One measurement requires small sample volumes eine adsorptive Stufe zur Elimination organischer Spurenstoffe. Optimierung
around 0.5 L and an effective workload of less than 20 min. Given der Pulveraktivkohleabtrennung durch Filtration als Grundlage zur
Anlagendimensionierung.
the incomplex sample preparation and relatively low TGA system Otero, M., Go mez, X., Garca, A., Moran, A., 2007. Effects of sewage sludge blending
prices, it seems manageable to include the analysis of PAC con- on the coal combustion: a thermogravimetric assessment. Chemosphere 69,
centration by the proposed method in the regular monitoring 1740e1750.
Otero, M., Calvo, L., Gil, M., Garca, A., Mor an, A., 2008. Co-combustion of different
routine of WWTPs equipped with PAC stages for advanced
sewage sludge and coal: a non-isothermal thermogravimetric kinetic analysis.
treatment. Bioresour. Technol. 99, 6311e6319.
Pasquini, L., Munoz, J.-F., Pons, M.-N., Yvon, J., Dauchy, X., France, X., Le, N.D., France-
Lanord, C., Go rner, T., 2014. Occurrence of eight household micropollutants in
Acknowledgement
urban wastewater and their fate in a wastewater treatment plant. Statistical
evaluation. Sci. Total Environ. 481, 459e468.
The authors express their gratitude towards the Swiss Federal Snyder, S., Westerhoff, P., Yoon, Y., Sedlak, D., 2003. Pharmaceuticals, personal care
518 ver et al. / Water Research 101 (2016) 510e518
T. Krahnsto
products, and endocrine disruptors in water: implications for the water in- von Klaranlagen. NETZSCH-Gera tebau GmbH, pp. 6e8.
dustry. Environ. Eng. Sci. 20, 449e469. Vu, T.T., Vogel, A., Fglein, E., Platz, S., Menzel, U., 2012. Nachweis von Pulver-
Snyder, S., Adham, S., Redding, A., Cannon, F., DeCarolis, J., Oppenheimer, J., Wert, E., aktivkohle in Abwa ssern mithilfe der Thermogravimetrie. Korresp. Abwasser,
Yoon, Y., 2007. Role of membranes and activated carbon in the removal of Abfall 59, 208e218.
endocrine disruptors and pharmaceuticals. Desalination 202, 156e181. Zietzschmann, F., Altmann, J., Ruhl, A.S., Dnnbier, U., Dommisch, I., Sperlich, A.,
~ as, B.J., Snoeyink, V.L., Ng, W.J., 2008. Effect of strongly competing
To, P.C., Marin Meinel, F., Jekel, M., 2014. Estimating organic micro-pollutant removal potential
background compounds on the kinetics of trace organic contaminant desorp- of activated carbons using UV absorption and carbon characteristics. Water Res.
tion from activated carbon. Environ. Sci. Technol. 42.7, 2606e2611. 56, 48e55.
Vogel, A., 2013. Thermogravimetrischer Nachweis von Pulveraktivkohle im Ablauf