PRODUCT DESCRIPTION
This autoclave series is designed to cover a large field of
applications for hospitals, CSSD and medical centers.
Narrow T-Max Mid Range Sterilizer
The autoclave series has a chamber volume range from 310
liters to 706 liters.
APPLICATIONS
Central Sterilizing Supply Department (CSSD)
Operating Theater
Out Patient Clinic
Rev. 19-07-2009 Tuttnauer - Your Sterilization & Infection Control Partners Page 1 of 9
TECHNICAL SPECIFICATION
T-Max Mid Range Sterilizer Series
STEAM GENERATOR
T-Max 10 660 660 1620 700 10 MODEL
POWER (kW)
T-Max 4 24
All automatic doors have a robust hydraulic mechanism for Customized electrical specifications are available (additional
operating doors. cost).
LANGUAGE
SERVICE ACCESS The operator display is available in English or Spanish.
From loading side and left or right side.
STANDARDS AND CODES
INSTALLATION SELECTIONS Tuttnauer products meet the following provisions and
standards:
Installed in a wall [Single Door]
Cabinet enclosure [Single Door] Europe:
Cabinet enclosure through one wall [Double Door] EN 285:2006 for Large Autoclaves
Installed through two walls [Double Door] USA:
ST8 Hospital Sterilizers
Standard Configuration Materials: Built-in Steam Generator Safety: A water level control
Chamber + Door: 316L stainless steel system maintains a constant water level that ensures safe
operation of the heaters. Furthermore, a sensor ensures that
Jacket: 304L stainless steel the heater does not operate without a minimum water intake.
Optional Materials (Stainless Steel): Emergency Shut-Off: Easily accessible emergency switches
Chamber + Door: 316Ti for immediate cycle shut-off.
Jacket: 316L or 316Ti
Double Independent Monitoring: The combined digital and
mechanical monitoring systems provide a cross-reference and
Chamber Design
guarantee accurate monitoring of temperature and pressure.
The chamber is fully jacketed. The chamber floor is slightly
sloped towards an internal drain to facilitate drainage. A
stainless steel mesh strainer protects the drain port from
blockage by debris. The chamber is mounted on a steel WATER RING VACUUM PUMP
framework which is height adjustable. The vacuum pump effectively removes more than 99% air from
the chamber. The pump is mounted on a damping mechanism
Door Gasket to minimize vibration.
Manual hinged door gasket: a silicone gasket is permanently
fixed in the door. During vacuum action the flow path of the fluids is through the
condenser (heat exchanger) and the vacuum pump/ejector to
Automatic sliding door gasket: a silicone gasket is fixed in a drain. This protects the vacuum system from excessive
groove in the door frame and the gasket is sealed against the temperatures.
door by means of steam pressure.
The insulation reduces the energy consumption by keeping the ** Primary piping system that connects to the chamber
heat inside the jacket and chamber. up to and including the first valve.
AIR FILTER
CITY TAP WATER CONSUMPTION
A disposable microbiological air filter is provided to filter all air
Average Consumption
entering the sterilizer chamber. The air is used to equalize the Model
(liters/cycle)
chamber pressure to atmospheric pressure at the end of the
sterilization cycle. T-Max 4 160
T-Max 6 210
The filter is rated for a particle size of 0.2 m. T-Max 8 210
T-Max 10 260
Note: Tuttnauers optional water recycling system (mentioned
on page 9) significantly reduces water usage.
STEAM SUPPLY
The autoclave is supplied with either a steam generator or an
option to connect to building steam supply. A combined option CONTROL SYSTEM
of steam generator and building steam supply is also available. The main board controls and monitors the physical parameters
of the sterilization process and performs the operation
Additional specialized steam supply configurations are listed in sequence of the machine, according to the selected program,
the Special Options and Accessories section below. and features:
Steam Generator: supplies steam for the sterilization process. Supports up to 6 PT100 sensors and 7 analog
The stainless steel steam generator is an electrically heated inputs (4 20 mA)
type generator equipped with immersion heaters, divided in Digital inputs and outputs for sterilizer control.
three-phases. The electrical system of the steam generator is Analog inputs for control and reading temperature and
located in a separate electric box. pressure.
Measures chamber pressure, jacket pressure, steam
Water to the steam generator is supplied by a single-phase generator pressure, and door gasket pressure.
pump, from a water reservoir, which is connected to a source Two serial ports for connection of up to two command
of mineral-free water. panels, one 9-pin serial connection for PC communication,
and one RS232 (optional RS485) serial port for modem
communication. The 9-pin interface connected to a PC
STEAM GENERATOR WATER QUALITY computer in conjunction with the ADMC software (optional)
Various water systems can be used to supply mineral-free enables a service technician to remotely control the
water to the steam generator such as Reverse Osmosis (RO), autoclave.
distilled water, etc. Water supplied to the steam generator FLASH memory and real-time clock backup store cycle
should be in compliance with the EN 285 standard which data for at least 5 years even if there is a power failure.
includes the following hardness and conductivity requirements:
Hardness < 0.1 mmol/l The control system controls all system functions, monitors
Conductivity < 50 S/cm system operations, both visually and audibly alerts the operator
of cycle malfunctions and, on demand, provides visual
Note: Soft water should not be used since its use may result in indication of the chamber temperature and pressure.
corrosion of the steam generator and chamber.
TEMPERATURE AND PRESSURE SENSORS
STEAM GENERATOR WATER CONSUMPTION The temperature and pressure measuring circuits are both
linear and designed with components having a high precision.
Consumption The PT100 sensors conform to Class A of the IEC751
Model standard.
(liters/cycle)
T-Max 4 24 When the exhaust temperature increases above 40C
T-Max 6 36 (measured by a PT100 sensor at the exhaust drain) then tap
T-Max 8 54 water is mixed with the exhaust to drain.
T-Max 10 60 The control system allows for the calibration of temperature
and pressure to be performed digitally.
The system uses FLASH memory in which the offset and gain
data of the sensors are stored. This data may be entered into
the system, through programming or through the autoclave
CAT 2007 control panel.
CONTROL PANEL ON LOADING SIDE The printing format is 24 characters per line. The following
The operators control panel is found at the loading side of the preliminary information and set parameters are printed when
sterilizer. the sterilization cycle begins:
The data is printed from the bottom up, beginning with the date
and ending with O.K. for a complete cycle or FAIL for an
aborted cycle.
Note: Upon request Tuttnauer can provide special configuration so that the autoclave will fit through narrow
passage-ways / elevators during transportation in and out of buildings.
Notes:
Par 3, 4 Residual air is displaced by 4 vacuum pulses (down to 25 kPa) and 3 steam pulses up
Atmospheric
to 160 kPa. Drying is performed during a "low-pressure" phase. When the pressure
Pressure
reaches atmospheric pressure the vacuum pump is activated until the pressure drops
to 25 kPa. Only the number and type of vacuum pulses can be modified.
Program 12 Air Leakage/Vacuum Test Air leakage test to chamber through door and other seals
Parameters Units Range Factory Setting
Atmospheric 1. Stabilizing Time min - 5
Pressure 2. Test Time min - 10
Par 1, 2
Notes:
Chamber vacuum is brought to 10 kPa. Thereafter all valves and motors are closed for
5 minutes enabling pressure stabilization. The acceptable pressure change over the
5 min 10 min following 10 minutes is 1.3 kPa.