TECHNICAL SPECIFICATION

Narrow T-Max Mid Range Sterilizer Series

FEATURE SUMMARY DEVICE PICTURE


Chamber volume from 310 to 700 liters
StU capacity 4 to 10
Working pressure meets PED and ASME requirements
Temperature range 105 C (221 F) to 137 C (279 F)
Electrical steam generator or external steam source
User friendly control system with touch screen display
12 preset cycle programs
Built-in printer
PC connection port for direct PC access
Independent pressure gauges on front panel
316L stainless steel chamber and door
Chamber has a mirror-like finish
Stainless steel piping
Conforms to Medical Device Directive 93/42 EEC
Conforms to standards: EN 285, EN13445, ASME,
AAMI/ANSI-ST8, UL, EN 61010-1 & EN 61010-2-040,
Company approved for ISO 9001:2000, ISO 13485:2003
(Medical Devices) and 21 CFR 820
Optional configurations available

PRODUCT DESCRIPTION
This autoclave series is designed to cover a large field of
applications for hospitals, CSSD and medical centers.
Narrow T-Max Mid Range Sterilizer
The autoclave series has a chamber volume range from 310
liters to 706 liters.

The autoclave operates with saturated steam as the sterilizing


agent with a temperature range from 105 C (221F) to 137 C
(279 F) and a working pressure that meets PED and ASME
requirements. The autoclave is equipped with a 24 kW / 32 kW
/ 36 kW / 63 kW electrical steam generator. These steam
generators supply the necessary steam for the sterilization
process. The sterilizer may also be steam fed from an external
steam source.

APPLICATIONS
Central Sterilizing Supply Department (CSSD)
Operating Theater
Out Patient Clinic

Information furnished by Tuttnauer is believed to be accurate and reliable.


However, no responsibility is assumed by Tuttnauer for its use.

This specification is subject to change without notice.

Rev. 19-07-2009 Tuttnauer - Your Sterilization & Infection Control Partners Page 1 of 9
TECHNICAL SPECIFICATION
T-Max Mid Range Sterilizer Series

PRODUCT SPECIFICATION VOLTAGE SUPPLY


Three phase or one phase electricity supply is available.
CHAMBER VOLUME & SIZE
415V 3 Phase with Neutral
CHAMBER DIMENSIONS VOLUME StU 380V / 400V 3 Phase with Neutral
MODEL
(mm) (liter) Capacity 230V 3 Phase with Neutral
Width Height Depth 208V 3 Phase without Neutral
220V / 230V 1 Phase *
T-Max 4 660 660 705 310 4
110V / 115V 1 Phase *
T-Max 6 660 660 990 430 6 * Must have building steam supply and vacuum by ejector

T-Max 8 660 660 1295 565 8 Frequency: 50 Hz or 60 Hz

STEAM GENERATOR
T-Max 10 660 660 1620 700 10 MODEL
POWER (kW)
T-Max 4 24

DOOR SELECTION T-Max 6 32


Single door T-Max 8 36
Double door (pass through)
T-Max 10 63
Automatic vertical sliding (control on top)

All automatic doors have a robust hydraulic mechanism for Customized electrical specifications are available (additional
operating doors. cost).

LANGUAGE
SERVICE ACCESS The operator display is available in English or Spanish.
From loading side and left or right side.
STANDARDS AND CODES
INSTALLATION SELECTIONS Tuttnauer products meet the following provisions and
standards:
Installed in a wall [Single Door]
Cabinet enclosure [Single Door] Europe:
Cabinet enclosure through one wall [Double Door] EN 285:2006 for Large Autoclaves
Installed through two walls [Double Door] USA:
ST8 Hospital Sterilizers

ELECTRICAL COMPONENTS Directives & Guidelines:


The following parts are located in the electric boxes: Medical Device Directive MDD 93/42 EEC
Pressure Equipment Directives PED 97/23 EEC
The Solid State Relays for the command of the vacuum EN 13445
pump, and the water pump. Low Voltage Directive 73/23 EEC
The Circuit Breakers for protection of all power circuits. EMC Directive 89/336
The power supplies for powering of electronic circuitry and WEE & RoHS Directives 2002/95, 96
solenoid valves. FDA Clearance
An overload switch is used to protect the device from high
overload currents. Safety and EMC Standards:
The connection elements (sockets, plugs, etc) and IEC/UL/EN 61010-1:2001, IEC/UL/EN 61010-2-041
components have a degree of protection conforming to the IEC/EN 61010-2-040:2005 Autoclave Safety
IP44 protection level and conform to the IEC60529 EN 61326 EMC
standard. International Protection Level IPx4
EN 50082-1:97 (EMC) Immunity Compatibility

Pressure Vessel and Steam Generator Construction


Standards:
ASME Code, Section VIII, Division 1, unfired pressure vessels
ASME Code, Section I, for boilers

Good Practice Standards:


ISO 17665-1 Validation and Routine Control
ST79 Sterility Assurance in Healthcare Facilities

Quality System Standards:


ISO 9001:2000 (Quality Systems)
ISO 13485:2003 (Quality Systems for Medical Devices)
21 CFR 820

Rev. 19-07-2009 Page 2 of 9


TECHNICAL SPECIFICATION
T-Max Mid Range Sterilizer Series
CHAMBER CONSTRUCTION
Materials Safety Valves: Both chamber and jacket are equipped with
The sterilizer chamber is constructed from solid, high quality pressure relief safety valves. If the pressure exceeds the
materials. Chamber and door thickness is no less than 6 mm. allowed limit then the safety valves will discharge.

Standard Configuration Materials: Built-in Steam Generator Safety: A water level control
Chamber + Door: 316L stainless steel system maintains a constant water level that ensures safe
operation of the heaters. Furthermore, a sensor ensures that
Jacket: 304L stainless steel the heater does not operate without a minimum water intake.

Optional Materials (Stainless Steel): Emergency Shut-Off: Easily accessible emergency switches
Chamber + Door: 316Ti for immediate cycle shut-off.
Jacket: 316L or 316Ti
Double Independent Monitoring: The combined digital and
mechanical monitoring systems provide a cross-reference and
Chamber Design
guarantee accurate monitoring of temperature and pressure.
The chamber is fully jacketed. The chamber floor is slightly
sloped towards an internal drain to facilitate drainage. A
stainless steel mesh strainer protects the drain port from
blockage by debris. The chamber is mounted on a steel WATER RING VACUUM PUMP
framework which is height adjustable. The vacuum pump effectively removes more than 99% air from
the chamber. The pump is mounted on a damping mechanism
Door Gasket to minimize vibration.
Manual hinged door gasket: a silicone gasket is permanently
fixed in the door. During vacuum action the flow path of the fluids is through the
condenser (heat exchanger) and the vacuum pump/ejector to
Automatic sliding door gasket: a silicone gasket is fixed in a drain. This protects the vacuum system from excessive
groove in the door frame and the gasket is sealed against the temperatures.
door by means of steam pressure.

Surface Treatment PIPES & COMPONENTS


A passivation layer is applied to the internal surface through an The piping system of the autoclave consists of air-operated ball
electro-chemical treatment which results in a smooth and shiny valves, which control the condensate and steam flow in and out
stainless steel. The resultant surface is polished to less than of the chamber, operates the vacuum, and the air inlet valve.
0.8 m which is highly protected against corrosion. The
internal corners are rounded to aid cleaning. Standard configuration:
Piping and fittings: 304 threaded stainless steel
Chamber Heating Components*: Brass
The fully jacketed chamber is pre-heated by steam in the
jacket. Optional configuration (stainless steel):
Insulation Primary piping & fittings **: 316 + triclamp fittings
The sterilizer jacket and door are completely insulated with a 1" Primary components: 316 triclamps or 316 threaded
chloride free glass wool thereby keeping the autoclave cool on
the outside. Customized thicker insulation is optional. The * Components Manual valve, non-return valve,
insulation is completely enclosed within a stainless steel cover. pressure regulator, pneumatic valves, steam trap, etc.

The insulation reduces the energy consumption by keeping the ** Primary piping system that connects to the chamber
heat inside the jacket and chamber. up to and including the first valve.

SAFETY FEATURES PRESSURE GAUGES


Door Safety Systems: The single door configuration has four pressure gauges
that indicate the pressure in the chamber, jacket, door
A pneumatic safety component (pressure switch) is
gasket, and the steam generator line.
installed in the autoclave which prevents opening of the
doors until pressure in the chamber reaches room
These pressure gauges are located on the autoclave's front
pressure.
panel (loading side).
Chamber door cannot be opened when chamber is
pressurized. The double door configuration has two additional pressure
Steam is not allowed into the chamber when the door is gauges on the unloading side that indicate the pressure in
open. the chamber and the pressure in the gasket of the second
A cycle cannot start if the door is open or not properly door.
locked.
The sliding door movement will stop immediately if an
object is detected in the way of the closing door.
Double door safety is implemented through interlocks which
prevent both doors from being opened simultaneously.

Rev. 19-07-2009 Page 3 of 9


TECHNICAL SPECIFICATION
T-Max Mid Range Sterilizer Series
VALIDATION PORTS VACUUM AND DRAIN COOLING WATER QUALITY
The chamber is provided with two 1" threaded connections for City tap water supply:
optional vacuum/pressure gauges and test sensors. The Hardness between 0.7 and 2 mmol/l
Validation Ports are each found on the right and left side of the The tap water pressure should be in the range of
sterilizer from the control side. 3 5 bar (40 - 72 psi)
Recommended temperature: 15C

AIR FILTER
CITY TAP WATER CONSUMPTION
A disposable microbiological air filter is provided to filter all air
Average Consumption
entering the sterilizer chamber. The air is used to equalize the Model
(liters/cycle)
chamber pressure to atmospheric pressure at the end of the
sterilization cycle. T-Max 4 160
T-Max 6 210
The filter is rated for a particle size of 0.2 m. T-Max 8 210
T-Max 10 260
Note: Tuttnauers optional water recycling system (mentioned
on page 9) significantly reduces water usage.
STEAM SUPPLY
The autoclave is supplied with either a steam generator or an
option to connect to building steam supply. A combined option CONTROL SYSTEM
of steam generator and building steam supply is also available. The main board controls and monitors the physical parameters
of the sterilization process and performs the operation
Additional specialized steam supply configurations are listed in sequence of the machine, according to the selected program,
the Special Options and Accessories section below. and features:
Steam Generator: supplies steam for the sterilization process. Supports up to 6 PT100 sensors and 7 analog
The stainless steel steam generator is an electrically heated inputs (4 20 mA)
type generator equipped with immersion heaters, divided in Digital inputs and outputs for sterilizer control.
three-phases. The electrical system of the steam generator is Analog inputs for control and reading temperature and
located in a separate electric box. pressure.
Measures chamber pressure, jacket pressure, steam
Water to the steam generator is supplied by a single-phase generator pressure, and door gasket pressure.
pump, from a water reservoir, which is connected to a source Two serial ports for connection of up to two command
of mineral-free water. panels, one 9-pin serial connection for PC communication,
and one RS232 (optional RS485) serial port for modem
communication. The 9-pin interface connected to a PC
STEAM GENERATOR WATER QUALITY computer in conjunction with the ADMC software (optional)
Various water systems can be used to supply mineral-free enables a service technician to remotely control the
water to the steam generator such as Reverse Osmosis (RO), autoclave.
distilled water, etc. Water supplied to the steam generator FLASH memory and real-time clock backup store cycle
should be in compliance with the EN 285 standard which data for at least 5 years even if there is a power failure.
includes the following hardness and conductivity requirements:
Hardness < 0.1 mmol/l The control system controls all system functions, monitors
Conductivity < 50 S/cm system operations, both visually and audibly alerts the operator
of cycle malfunctions and, on demand, provides visual
Note: Soft water should not be used since its use may result in indication of the chamber temperature and pressure.
corrosion of the steam generator and chamber.
TEMPERATURE AND PRESSURE SENSORS
STEAM GENERATOR WATER CONSUMPTION The temperature and pressure measuring circuits are both
linear and designed with components having a high precision.
Consumption The PT100 sensors conform to Class A of the IEC751
Model standard.
(liters/cycle)
T-Max 4 24 When the exhaust temperature increases above 40C
T-Max 6 36 (measured by a PT100 sensor at the exhaust drain) then tap
T-Max 8 54 water is mixed with the exhaust to drain.
T-Max 10 60 The control system allows for the calibration of temperature
and pressure to be performed digitally.

Each sensor circuit is calibrated with individual constants to


correct the deviation in manufacturing and aging.

The system uses FLASH memory in which the offset and gain
data of the sensors are stored. This data may be entered into
the system, through programming or through the autoclave
CAT 2007 control panel.

Rev. 19-07-2009 Page 4 of 9


TECHNICAL SPECIFICATION
T-Max Mid Range Sterilizer Series
CONTROL PANEL ALARMS
Control and Monitoring The autoclave uses both audible and visual alarm indicators.
Automatic process checking and failure detection are provided
The control system is operated via the CAT 2007 panel which by the control system. In the event of a failure during the
uses an easy-to-use LCD screen found on the operator control sterilization cycle, the system enters an alarm phase which will
side (loading side). safely end the process automatically. The range of alarms
includes:
Five password access levels (0-4) are provided to control
access/operation of the machine preventing unauthorized Temperature & pressure sensor failure
access. These access levels are customizable. Access control Phase time-outs
can be applied to functions, such as running test cycles, setting Door(s) not properly closed
parameters, calibration, service and maintenance, cycle Power failure
selection, cycle start and door control. No water in the water reservoir
Optional utility alarms: no water / no steam / no air
With the standard factory configuration, calibration of the
temperature circuits and calibration of the pressure circuits
require an access code. CYCLE DOCUMENTATION - PRINTER
The autoclave is equipped with a printer which prints a detailed
Double door autoclaves are provided with status indicators on history of each cycle performed by the instrument. Various
the unloading side. types of printers are listed in the Special Options and
Accessories section towards the end of this document.

CONTROL PANEL ON LOADING SIDE The printing format is 24 characters per line. The following
The operators control panel is found at the loading side of the preliminary information and set parameters are printed when
sterilizer. the sterilization cycle begins:

Software version (printed only when power is on)


Real time
Selected program
Sterilization temperature
Sterilization time
Dry time

Thereafter, the autoclave starts performing the sequence of


operations of the cycle. The measured values of temperature
and pressure are printed at fixed time intervals, according to
various phases of the sterilization process: 4 minute time
interval for vacuum, 1 minute time interval for sterilization, and
the start and end time of the drying phase. All interval times
can be user defined. Furthermore, the customer may request
customized time intervals prior to order delivery.

The data is printed from the bottom up, beginning with the date
and ending with O.K. for a complete cycle or FAIL for an
aborted cycle.

OPTIONAL LOADING EQUIPMENT


Two Pull Out Trays
Stainless steel trays equipped with tracks for easy loading and
unloading. The tracks are designed to prevent the trays from
rolling over.

Loading Cart and Transfer Carriage


CAT 2007 Operators Control Panel The loading cart rolls off the transfer carriage onto the chamber
rails for easy handling of heavy loads.
The control panel consists of the following:
The transfer carriage uses heavy duty revolving wheels
4 row display, 20 characters per row
(castors) to maximize mobility in limited space, and wheel
18 key pads
brakes to prevent rolling. There is a lock to prevent the loading
2 LED's (Fail, Run)
cart from sliding. The legs of the transfer carriage are
adjustable to match the height of the loading cart with the
The following information is displayed:
tracks in the sterilizer chamber.
Temperature, chamber pressure and jacket pressure
Door status Both the loading cart and transfer carriage are made of durable
Sterilization time count down AISI 304 or AISI 316 stainless steel.
Autoclave status: standby, ready, pre-vacuum, heating,
sterilization, exhaust, dry time, air inlet, cycle ended See Special Options and Accessories section for more options.

Rev. 19-07-2009 Page 5 of 9


TECHNICAL SPECIFICATION
T-Max Mid Range Sterilizer Series

STERILIZER DOCUMENTATION Program 8, 9, 10 with Slow 121:


A number of copies of the manuals are provided. Operator and Materials which require a slow pressure exhaust in open
service manuals are in selected language. Other manuals are or closed containers (not sealed).
in English. Manuals include electrical and piping diagram. For loads which its manufacturer declares to be in compliance
the following sterilization conditions:
Furthermore, a CD is provided containing the following:
Sterilization temperature 121 C
Operators manual
Sterilization time 15 minutes (programs 8 and 9)
Technical manual
Sterilization time 30 minutes (program 10)
Serial number for specific autoclave
Slow exhaust
XPCS software for calibration and software download
Slow discharge of pressure
Factory test report prior to shipping
Optional - technical specifications for component, etc.
TEST CYCLE PROGRAMS
Program 11: Bowie & Dick Test
STERILIZATION PROGRAMS This program is intended to test the efficiency of air removal
The autoclave is supplied with 10 sterilization programs and 2 from within the chamber.
test programs. An authorized user can change the parameters
of each program cycle. The number and type of vacuum Test parameters:
pulses, the atmospheric pressure value, and the pressure Sterilization temperature 134 C
profile are also programmable. There are 6 selectable exhaust Sterilization time 3.5 minutes
modes. The cycle programs are summarized as follows: Fast exhaust
Drying time of 2 minutes

Program 1 with Fast 121:


Unwrapped rigid (instruments), etc. Program 12: "Air Leakage Test (Vacuum Test)
For loads which its manufacturer declares to be in compliance This program is intended to test air leakage from the chamber
the following sterilization conditions: through the door seal or any other seals. This test is
performed during the vacuum phase.
Sterilization temperature 121 C
Sterilization time 15 minutes
Fast exhaust
Drying time 1 min
MAINTENANCE / SERVICE PLAN
A global network of skilled service specialists can provide
periodic inspections and adjustments to help assure low-cost
Program 2 and 3 with Fast 134:
peak performance. A detailed service and maintenance plan is
Unwrapped rigid (instruments), etc.
included in the operator manual.
For loads which its manufacturer declares to be in compliance
the following sterilization conditions:

Sterilization temperature 134 C


Sterilization time 3 minutes PACKAGING FOR SHIPMENT
Fast exhaust The sterilizer is packed in a wooden crate for
Drying time 1 min shipping/transportation.

Program 4 and 5 with Dry 134:


Wrapped materials (instruments), porous load, etc. WARRANTY
For loads which its manufacturer declares to be in compliance Tuttnauer warrants that each device is carefully tested,
the following sterilization conditions: inspected and that it leaves the factory in proper working
condition.
Sterilization temperature 134 C
Sterilization time 7 minutes Tuttnauer certifies that the device is guaranteed to be free from
Fast exhaust defects in material and workmanship, for one year from
Drying time 15 min installation date but not more than 18 months from shipping
date, against faulty components and assembly. Extended
warranty periods are optional.
Program 6 with Dry 121:
Wrapped materials (instruments), porous load, etc. The warranty does not include and does not replace routine
For loads which its manufacturer declares to be in compliance treatment and preventive maintenance to be performed
the following sterilization conditions: according to "Preventive and Periodical Maintenance"
instructions mentioned in the device's accompanying manual.
Sterilization temperature 121 C
Sterilization time 15 minutes The user must ensure that all utilities used, including the
Fast exhaust water, meet all the specifications mentioned in the
Drying time 15 min operator manual.

The user is subject to the full warranty statement found in the


documentation delivered with the equipment.

Rev. 19-07-2009 Page 6 of 9


TECHNICAL SPECIFICATION
T-Max Mid Range Sterilizer Series

UTILITY DATA / REQUIREMENTS


External Steam Source Mineral-Free Feed Water for Steam Generator
1. 97 100 % dry saturated steam 1. Water intended for the steam generator must have a water
2. Pressure 4 to 6 Bar quality in accordance with EN285/ISO 17665
3. Final steam pipe to the autoclave at least pipe 1" 2. Install 1/2" pipe with a shut-off valve at its end
4. Install a shut-off valve at end of steam pipe 3. Regulations may require a Back-Flow protection device
5a. T-Max 8 steam rate (water supply) 30 Kg/hour 4a. T-Max 4 mineral-free water supply 30 liters/hour
5b. T-Max 8 steam rate (water supply) 46 Kg/hour 4b. T-Max 6 mineral-free water supply 46 liters/hour
5c. T-Max 8 steam rate (water supply) 46 Kg/hour 4c. T-Max 8 mineral-free water supply 46 liters/hour
5d. T-Max 10 steam rate (water supply) 80 Kg/hour 4d. T-Max 10 mineral-free water supply 80 liters/hour

Electrical Tap Water


With generator (all models) 1. Intended for vacuum pump and cooling
3-Phase, 208 / 220 / 230 / 380 / 400 / 415 Volt, with or 2. Install 3/4" pipe with shut-off valve at its end
without neutral, 24 kW / 32 kW / 36 kW / 63 kW 3. 1 4 bar pressure supply at approximately 15C
4. Local regulations may require a Back-Flow protection
Without generator (all models) device
Note: must have building steam supply 5. Hardness (free of alkaline earth ions) should be between
3-Phase, 208 / 220 / 230 / 380 / 400 / 415 Volt, 0.7 mmol/l and 2.0 mmol/l
with or without neutral, not less than 10 A 6. Supply: approximately 17 l/min
1-Phase, 110 / 115 / 220 / 230 Volt, with neutral, not less
than 10 A, vacuum by ejector

Additional customizable electrical specifications are available. Compressed Air


1. At least 3/8" hose to supply the compressed air
2. The compressed air supply line coming from the building
Drainage (Sewage) must have a shut-off valve
1. At least 2" sewage pipe. 3. Pressure: 5 to 8 bar, free from liquid water, filtered to
2. The sewage shall be able to withstand continuous 25 m, free from oil droplets greater than 2 m
temperature of 80 C and for short time periods at 100 C.
3. Each autoclave will have a separate sewage line connected
to a vertical vent pipe. Vertical pitch is 1000 mm. Ambient Temperature
Plant room temperature should be in the range from 5 to 40 C
Note: Local national regulations may require that the drain be and 85% RH (relative humidity).
tapped and vented, and not connected to other drains which
may cause back pressure or obstruct flow. An air break may Floor
also be necessary. The unit mass shall be considered no less than 10000 N/m2
according to the DIN 58949-7:2004 standard.
Ventilation
10 air replacements per hour

OVERALL EXTERNAL DIMENSIONS


DOOR CONFIGURATION EXTERNAL DIMENSIONS (mm)
MODEL
Door Type No. of Doors Width Height Depth
1 1030
T-Max 4 Vertical Sliding 999 1980
2 1050
1 1400
T-Max 6 Vertical Sliding 999 1980
2 1310
1 1690
T-Max 8 Vertical Sliding 999 1980
2 1620
1 2100
T-Max 10 Vertical Sliding 999 1980
2 1950

Note: Upon request Tuttnauer can provide special configuration so that the autoclave will fit through narrow
passage-ways / elevators during transportation in and out of buildings.

Rev. 19-07-2009 Page 7 of 9


TECHNICAL SPECIFICATION
T-Max Mid Range Sterilizer Series

PROGRAM CYCLE DATA


Program 1 with Fast 121 Unwrapped rigid (instruments), etc.
Par 1, 2 Parameters Units Range Factory Setting
1. Sterilization Temperature C 105 - 137 121
2. Sterilization Time min 1 - 99 15
3. Exhaust Mode 1-6 1 (fast)
4. Drying Time (Post-Vacuum) min 0 - 99 1
Par 3, 4 Notes:
Atmospheric
Pressure Residual air is displaced by 4 vacuum pulses (down to 25 kPa) and 3 steam pulses up
to 160 kPa. When the pressure reaches atmospheric pressure, the vacuum pump is
activated until the pressure drops to 25kPa.
Program 2 and 3 with Fast 134 Unwrapped rigid (instruments), etc.
Parameters Units Range Factory Setting
Par 1, 2 1. Sterilization Temperature C 105 - 137 134
2. Sterilization Time min 1 - 99 3
3. Exhaust Mode 1-6 1 (fast)
4. Drying Time (Post-Vacuum) min 0 - 99 1
Notes:
Atmospheric Par 3, 4
Residual air is displaced by 4 vacuum pulses (down to 25 kPa) and 3 steam pulses up
Pressure to 160 kPa. When the pressure reaches atmospheric pressure the vacuum pump is
activated until the pressure drops to 25 kPa.
Program 4, 5 and 7 with Dry 134 Wrapped materials (instruments), porous load, etc.
Parameters Units Range Factory Setting
Par 1, 2
1. Sterilization Temperature C 105 - 137 134
2. Sterilization Time min 1 - 99 7
3. Exhaust Mode 1-6 1 (fast)
4. Drying Time (Post-Vacuum) min 0 - 99 15
5. Dryvac Time min 0 - 99 5
Atmospheric Par 3, 4, 5 Notes:
Pressure Residual air is displaced by 4 vacuum pulses (down to 25 kPa) and 3 steam pulses up
to 160 kPa. When the pressure reaches atmospheric pressure the vacuum pump is
activated until the pressure drops to 25 kPa.
Program 6 with Dry 121 Wrapped materials (instruments), porous load, etc.
Parameters Units Range Factory Setting
Par 1, 2
1. Sterilization Temperature C 105 - 137 121
2. Sterilization Time min 1 - 99 15
3. Exhaust Mode 1-6 1 (fast)
4. Drying Time (Post-Vacuum) min 0 - 99 15
5. Dryvac Time min 0 - 99 5
Atmospheric Par 3, 4, 5 Notes:
Pressure Residual air is displaced by 4 vacuum pulses (down to 25 kPa) and 3 steam pulses up
to 160 kPa. When the pressure reaches atmospheric pressure the vacuum pump is
activated until the pressure drops to 25 kPa.
Program 8, 9, 10 with Slow 121 Liquids: slow pressure exhaust for open or closed containers
Parameters Units Range Factory Setting
Par 1, 2 1. Sterilization Temperature C 105 - 137 121
2.a Sterilization Time (program 8 and 9) min 1 - 99 15
2.b Sterilization Time (program 10) min 1 - 99 30
Par 3 3. Exhaust Mode 1-6 4 (slow)
Notes:
Atmospheric Residual air is displaced by 1 vacuum pulse (down to 25 kPa). A slow exhaust is used
Pressure to discharge pressure from the chamber very slowly. Exhaust time is dependent on
load volume and configuration of containers. This program is not designed in
accordance with the EN285 and must not be used for routine sterilization of medical
devices.
Program 11 Bowie & Dick Test Efficiency test of air removal from within the chamber
Parameters Units Range Factory Setting
Par 1, 2
1. Sterilization Temperature C - 134
2. Sterilization Time min - 3.5
3. Exhaust Mode - 1 (fast)
4. Drying Time (Post-Vacuum) min - 2

Notes:
Par 3, 4 Residual air is displaced by 4 vacuum pulses (down to 25 kPa) and 3 steam pulses up
Atmospheric
to 160 kPa. Drying is performed during a "low-pressure" phase. When the pressure
Pressure
reaches atmospheric pressure the vacuum pump is activated until the pressure drops
to 25 kPa. Only the number and type of vacuum pulses can be modified.
Program 12 Air Leakage/Vacuum Test Air leakage test to chamber through door and other seals
Parameters Units Range Factory Setting
Atmospheric 1. Stabilizing Time min - 5
Pressure 2. Test Time min - 10
Par 1, 2

Notes:
Chamber vacuum is brought to 10 kPa. Thereafter all valves and motors are closed for
5 minutes enabling pressure stabilization. The acceptable pressure change over the
5 min 10 min following 10 minutes is 1.3 kPa.

Rev. 19-07-2009 Page 8 of 9


TECHNICAL SPECIFICATION
T-Max Mid Range Sterilizer Series

SPECIAL OPTIONS AND ACCESSORIES ADVANCED OPERATION FEATURES


Multipurpose Valve
ADVANCED CONTROL FEATURES Compressed air operated control valve drains the water/steam
10" Touch Screen (industrial PC with Windows XP system) and releases pressure in order to open the chamber if a cycle
The screen is placed on the front. In addition to cycle name, failure occurs due to electric power failure.
number, stage and parameters, the screen displays cycle
information in a graphical or textual format showing progress Dual Steam Source Valve
of each stage. Previously stored cycle data can be recalled. To connect central steam supply and backup steam generator.
The piping system can also be monitored visually.
Foot Pedal to Open Door
The price includes ADMC software (see ADMC below). User The foot pedal feature allows one to open the door by foot.
interface languages are: English, Spanish, and Russian.
Automatic Restart and Shutdown
ADMC Software Feature enables a preprogrammed restart and/or complete
Enables complete control of the autoclave, including receiving shutdown of utilities at cycle end or set time, closing all utility
data, creating printable graphs and parameter tables. The valves, returning chamber and jacket to ambient temperature.
software can be used for calibration and indicating valve and
switch status. The software creates an electronic file (.mbd)
accessible with suitable database software, such as Microsoft ADVANCED STEAM RELATED OPTIONS
Access, to control the autoclave or show real time data. Automatic Blow Down
Automatic drainage can be accomplished through the use of a
TAS - Tuttnauer Archive System (Datalogger) programmable timer.
This system provides a complete archive solution storing up to
100,000 hours of logs (log rate from 1 to 999 seconds). It Separate Connection to Jacket and Chamber
features automatic printout. Its user-friendly software allows Enables reaching sterilization at a faster rate and improved
use of the data file for statistics, reports, etc. The system temperature control when working with liquids, especially with
operates independently of the autoclave with its own power 1050C and less. This feature is vital for clean steam systems.
supply, 4MB flash memory, two USB ports, two RS232 ports,
one parallel port, battery backup real time clock, LED with
three color indications, and Windows software. ACCESSORIES
Air Compressor
Independent Cycle Documentation (Chart Recorder) For operating the pneumatic valves and automatic door/s
A high-speed chart recorder, with independent microprocessor (must be ordered if there is no local source of compressed
control and power supply, is suitable for independent cycle air).
documentation. This multi-range input recorder can record 12
points at once from RTDs and DC voltage input signals, Hotwell
producing analog trend records and print-outs. Simple The hotwell heats water to about 80-90C to remove non-
operation with an easy-to-view display allows one to key-in condensable gases prior to being pumped to the steam
various items of set data. The unit operates independently of generator or steam-to-steam generator.
the autoclave (obligatory for meeting EN285).
Bio Shield Frame
Barcode Reader and Driver The sterilizer is surrounded by a frame which serves as a
The barcode reader allows identification and tracing of the preparation for a cross-contamination seal made of Neoprene
material flow in the CSSD by reading the barcode on the sheet that is attached between the frame and the wall at site.
container/basket. The data is transferred to the ADMC and
linked to the sterilization cycle parameters and batch. One Water Recycling System
driver can support up to 4 barcode reader devices. This system recycles water necessary for the heat exchanger
to cool down the steam exhaust from the chamber. The water
Two Color Printer saving system recycles the water and reuses it. Tuttnauer's
The Star SP2000 uses a 9-pin dot matrix head and prints up water saving system saves more than 50% of the tap water
to 3.1 lines per second. It uses a simple paper loading system. used.

Graphic Printer - DPU 414


The DPU-414 is a silent high-quality printer that prints
characters and high-density graphics. The DPU-414 prints a CONTACT INFORMATION
detailed history of each cycle performed by the instrument. International Sales and Marketing
The price includes the TAS (Datalogger).
E-mail: info@tuttnauer-hq.com
Website: www.tuttnauer.com
Full Control System for Second Door (Two Door Models)
Double door sterilizers may be installed with an optional Tuttnauer Europe b.v.
control interface on the unloading side. If requested, a double Paardeweide 36, 4824 EH,
door sterilizer may have a fully operated console on both P.O.B. 7191, 4800 GD Breda
sides. The Netherlands
Tel: +31 (0) 765 423 510 Fax: +31 (0) 765 423 540
E-mail: info@tuttnauer.nl
COMMUNICATIONS
Modem
Tuttnauer - Your Sterilization & Infection Control Partners
Internal modem for direct factory on-line diagnostics (ADMC
software is a pre-requisite and should be requested).

Rev. 19-07-2009 Page 9 of 9