INTERNAL CALIBRATION PROCEDURE

Approved by:D.R.Vickers Date:2003


BMTS-CP-018, Issue 1, June 2003

INTERNAL CALIBRATION PROCEDURE AND RESULTS


Steel Rule

ITEM DETAILS
Bodycote Serial Number Description Range
Reference

1 Procedure
1.1 Equipment/Consumables required

REFERENCE EQUIPMENT REQUIRED/USED


Description Bodycote Reference Date of Last Calibration Stated
Uncertainty
Reference Steel ruler
Glass plate -- -- --
Square

b. Method

2.1 Visually inspect the working rule. Take particular note of the condition of the following:

corners between the measuring scale and the datum edge,


damage to the datum edge,
damage to the scale edge(s)
Distortion or evidence of bending
Corrosion damage or wear that makes scale difficult to read.

Record findings in the box below and report any serious deficiency to the Quality Manager.

Condition of Criteria for Normal Use Acceptable (Yes/No) and comments


and limitations if applicable
Corners Rounding <0.5mm

Datum Edge Free from dents and burrs,


square to scale edge(s)
Scale Edges Free from dents and burrs

Distortion or bend Rule lies flat on glass plate

Scales Easily readable, no corrosion,


No obliteration / wear

2.2 Allow the reference rule and working rule to stabilize in the same temperature environment for
at least 3 hours before calibration.

2.3 The working rule and the reference rule shall be laid flat on a glass plate and aligned using
engineers square. Only the metric face shall be checked.

Page 1 of 3 BMTS-CP-18 Issue 1


INTERNAL CALIBRATION PROCEDURE
Approved by:D.R.Vickers Date:2003

2.4 The accuracy of adjacent graduation lines on both are compared at 10, 25, 50mm and at
increments of 50mm thereafter. Use a low power magnifying glass to assist in the comparison
of the graduation marks. Estimate any deviations to the nearest 0.2mm

3 Results of Calibration

Reference rule Top scale Bottom scale Permissible


(mm) (mm) (mm) deviation
(mm)
10 0
25 0
50 0
100 0
150 0
200 0
250 0
300 0
350 0
400 0
450 0
500 0
550 0
600 0
650 0
700 0
750 0
800 0
850 0
900 0
1000 0

Date of Calibration Ambient Temperature (C)

4 Assessment of Results

Suitability for Use/Restrictions on Use

Name of Operator Signature of Operator Checked by, Name Checked by, Signature

Page 2 of 3 BMTS-CP-18 Issue 1


INTERNAL CALIBRATION PROCEDURE
Approved by:D.R.Vickers Date:2003

5 Uncertainty of Measurement

5.1 The Uncertainty of measurement has been assessed as 0.19mm for the calibration process.
.
5.2 For subsequent measurements using the rule, measuring to the nearest 0.5mm, the Uncertainty
of measurement is estimated as 0.7mm.

Details of these assessments are held by the Quality Manager.

Page 3 of 3 BMTS-CP-18 Issue 1