Beruflich Dokumente
Kultur Dokumente
This standard has been approved for use by agencies of the Department of Defense.
1 3
This test method is under the jurisdiction of ASTM Committee D15 on Engine Reagent Chemicals, American Chemical Society Specifications, American
Coolants and is the direct responsibility of Subcommittee D15.04 on Chemical Chemical Society, Washington, DC. For suggestions on the testing of reagents not
Properties. listed by the American Chemical Society, see Analar Standards for Laboratory
Current edition approved June 15, 1991. Published October 1991. Originally Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia
published as D 1287 53 T. Last previous edition D 1287 85. and National Formulary, U.S. Pharmaceutical Convention, Inc. (USPC), Rockville,
2
Annual Book of ASTM Standards, Vol 15.05. MD.
Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1
D 1287 91 (2002)
NOTE 4Commercially available buffer tablets or prepared buffer 9.1.1 Turn on the instrument, allow it to warm up, and
solutions may be used. adjust according to the manufacturers instructions. Immerse
6.4 Potassium Hydrogen Phthalate Buffer Solution (0.05 the tips of the electrodes in the standard buffer solution chosen
M, pH = 4.01 at 25C)Dissolve 10.21 g of potassium and allow sufficient time for the temperature of the buffer
hydrogen phthalate (KHC8H4O4) in water, and dilute to 1 L. solution and the electrodes to equalize. Set the temperature
6.5 Neutral Phosphate Buffer Solution (0.025 M with re- knob at the temperature of the buffer solution. Adjust the
spect to each phosphate salt, pH = 6.86 at 25C)Dissolve standardization or asymmetry potential control until the meter
3.40 g of potassium dihydrogen phosphate (KH2PO4) and 3.55 registers a scale reading, in pH units, equal to the known pH of
g of anhydrous disodium hydrogen phosphate (Na2HPO4) in the standardizing buffer solution.
water and dilute to 1 L. 9.1.2 Rinse the electrodes with distilled water and touch the
6.6 Borax Buffer Solution (0.01 M, pH = 9.18 at 25C) tips with a dry cloth or tissue to remove excess water. Immerse
Dissolve 3.81 g of disodium tetraborate decahydrate (Na2B4O the electrodes in a second standard buffer solution. The reading
710H2O) in water, and dilute to 1 L. Protect this solution from of the meter shall agree with the known pH of the second
atmospheric carbon dioxide; stopper the bottle at all times standard buffer solution within 60.05 unit without changing
except when actually in use. the setting of the standardization of asymmetry potential
6.7 Alkaline Phosphate Buffer Solution (0.01 M trisodium control. If it does not agree or if the meter gives a slow
phosphate, pH = 11.72 at 25C)Dissolve 1.42 g of anhy- response time and drifts, the electrode should be cleaned in
drous disodium hydrogen phosphate (Na2HPO4) in 100 mL of accordance with the manufacturers instructions.
a 0.1 M carbonate-free solution of sodium hydroxide and dilute 9.1.3 The presence of a faulty electrode is indicated by
to 1 L with water. failure to obtain a correct value for the pH of the second
6.8 Potassium Chloride ElectrolytePrepare a saturated standard buffer solution after the meter has been standardized
solution of potassium chloride (KCl) in water. with the first. A cracked electrode will often yield pH values
that are essentially the same for both standard buffer solutions.
7. Sampling
10. Procedure
7.1 Sample material in accordance with Practice D 1176.
10.1 Make the pH determination on (1) the unused concen-
8. Preparation of Electrode System trated engine coolant or antirust (as received), or (2) any
desired dilution, with distilled water, of the concentrated
8.1 Maintenance of ElectrodesClean the glass electrode at
product, or (3) used engine coolant or antirust solution. For
frequent intervals (not less than once every week during
dilute solutions, suggested engine coolant concentrations are
continual use) in accordance with the manufacturers instruc-
50 and 33 volume %; antirust solution concentration should be
tions. Drain the calomel electrode at least once each week and
in the range of proposed usage. Prepare solutions as follows:
fill with KCl electrolyte. Keep the electrolyte level in the
Unused engine coolant or antirust solution samples to be tested
calomel electrode above that of the solution being tested at all
should be obtained following the applicable portions of the
times. When not in use, immerse the lower halves of the
procedure outlined in Practice D 1176.
electrodes in distilled water, or the appropriate buffer solution
10.1.1 When a small, insoluble liquid layer is present,
recommended by the manufacturer. Do not immerse the
remove it prior to testing the unused sample or preparing
electrodes in the solution being tested for any appreciable
diluted samples for testing. Using a 100-mL mixing cylinder
period of time between determinations. Although the elec-
and a pipet or other suitable measuring device, make up a
trodes are not extremely fragile, they should be handled
solution of the concentration desired for testing by measuring
carefully at all times.
the necessary amount of antifreeze or antirust solution into the
8.2 Preparation of ElectrodesBefore and after using,
cylinder and adding sufficient distilled water to bring the
wipe the electrode thoroughly with a clean cloth, or a soft
solution volume to 100 mL.
absorbent tissue, and rinse with distilled water. Follow the
10.1.2 Thoroughly mix the contents of the cylinder by
manufacturers recommendations if further preparation is re-
shaking, and transfer into a 250-mL tall-form spoutless or other
quired. Prior to each pH determination, soak the prepared
suitable beaker.
electrode in distilled water for at least 2 min. Immediately
before use, touch the tips of the electrodes with a dry cloth or NOTE 5Some pH apparatus is equipped with a sample cup. It is
tissue to remove excess water. satisfactory to use these cups and the smaller sample required with them,
as an alternative to the 100-mL sample and beaker specified.
9. Standardization and Testing of Electrode System in 10.2 Prepare the calomel and glass electrodes as directed in
pH Units Section 8. Immerse the tip of the electrode in the solution and
9.1 The assembly shall always be standardized with one of agitate. Where the instrument has a mechanical stirrer, this may
the standard buffer solutions. A second standard buffer solution be used. Allow the system to come to equilibrium, and record
shall be used to check the linearity of the response of the the pH value. Remove and wash the electrodes with distilled
electrodes at different pH values and to detect a faulty glass water and immerse them in distilled water until the next
electrode or incorrect temperature compensation. The two determination is made.
buffer solutions selected should bracket the anticipated pH of NOTE 6In the event that the glass electrodes become coated with an
the solutions being tested. oil film, as the result of testing two-phase products, it is necessary to
2
D 1287 91 (2002)
remove such film with a suitable solvent, such as 1 + 1 mixture of toluene 12. Precision and Bias
and ethyl acetate.
12.1 RepeatabilityDuplicate results by the same operator
NOTE 7With silicate-containing engine coolant, it is important to
specify that the pH reading be taken either immediately after preparing the
should not be considered suspect unless they differ by more
dilution or after a specified period of time, such as 30 min. In contrast to than 60.1 pH units.
an immediate pH, the 30-min value has been shown to be approximately 12.2 ReproducibilityThe results submitted by two or more
0.2 units lower and a measurement made after 24 h can be from 0.3 to 0.5 laboratories should not be considered suspect unless the results
pH units lower. differ by more than 60.2 pH units.
11. Report 13. Keywords
11.1 Report the pH. 13.1 antirusts; engine coolants; pH of engine coolants
ANNEX
(Mandatory Information)
A1. APPARATUS
3
D 1287 91 (2002)
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).