Sie sind auf Seite 1von 9

G Model

CLAE 925 No. of Pages 9

Contact Lens and Anterior Eye xxx (2016) xxxxxx

Contents lists available at ScienceDirect

Contact Lens and Anterior Eye


journal homepage: www.elsevier.com/locate/clae

The impact of lens care solutions on corneal epithelial changes during


daily silicone hydrogel contact lens wear as measured by in vivo
confocal microscopy
Xiaolin Zhanga , Christine Marchettia , Jessica Leea , Yan Suna , Sara Debanneb , Ying Jiangb ,
Jami Kernc , Mark Harroda , Beth Ann Benetza , Eric Pearlmana,d , Loretta Szczotka-Flynna,*
a
Department of Ophthalmology & Visual Sciences and University Hospitals Eye Institute, University Hospitals Case Medical Center, Cleveland, OH, United
States
b
Department of Epidemiology & Biostatistics, Case Western Reserve University, Cleveland, OH, United States
c
Alcon Research, Fort Worth, TX, United States
d
Department of Ophthalmology, University of California, Irvine, United States

A R T I C L E I N F O A B S T R A C T

Article history: Purpose: To assess corneal epithelial microstructure via confocal microscopy and determine if cellular
Received 3 March 2016 changes are associated with lens care solutions during daily wear of silicone hydrogel contact lenses.
Received in revised form 9 November 2016 Methods: Corneal in vivo confocal microscopy with the Nidek ConfoScan4 was performed at baseline and
Accepted 11 November 2016
after 5 months of lotralcon A daily contact lens wear. Enrolled participants were randomized to use
either a polyhexamethylene biguanide (PHMB) preserved multipurpose care solution (MPS) or a peroxide
Keywords: based solution system. Lens and storage case bioburden were assessed with aerobic culture methods.
Confocal microscopy
Univariate and multivariable analyses were done to evaluate the association between solution use, or
Silicone hydrogel contact lens
Hyper-reective corneal epithelial cells
solution-related clinical covariates, and morphologic differences (hyper-reectivity) in the supercial
epithelial cells and epithelial basal cell density.
Results: Data on 139 participants were available for analysis of supercial epithelial cells while data on 92
participants were available for epithelial basal cell density. Five months after randomization to the
solution groups, 33% of participants had visible hyper-reective cells. More participants using MPS had
1 hyper-reective cells compared to peroxide users at 5 months (44% vs. 22%; p = 0.006). Similarly at 5
months, more participants with solution-induced corneal staining (SICS) had 1 hyper-reective cells
compared to non-SICS participants (57% vs. 29%; p = 0.010). The adjusted odds ratios (ORs) for risk of
presenting with hyper-reective cells in MPS users or SICS participants was 2.7 (95% CI; 1.275.65) and
3.4 (95% CI; 1.298.97), respectively. Basal cell density decreased by over 350 cells/mm2 over time (about
6%) in participants who had substantial bioburden on their lenses or in their storage case.
Conclusion: The confocal microscope can detect epithelial cellular changes in vivo during contact lens
wear. Hyper-reective supercial epithelial cells are associated with a PHMB preserved solution and
decreases in basal epithelial cell density may be associated with bacterial bioburden.
2016 British Contact Lens Association. Published by Elsevier Ltd. All rights reserved.

1. Introduction cornea have not been systematically addressed. We examined


these variables using in vivo confocal microscopy as a non-invasive
There is a wealth of literature evaluating the effects of lens care way to examine the corneal epithelium at the cellular level.
solutions and microbial bioburden on corneal structure and Confocal microscopy is clinically used to diagnose atypical and
function [17]. However, the effects of multiple clinical covariates rare corneal infections [810] and corneal endothelial density and
during contact lens wear including solution type, lens or solution- morphology [11,12]. In the research setting, it is additionally used
induced corneal staining, and bioburden on cellular changes in the to evaluate corneal epithelial structure and reectivity [13], sub-
basal nerve plexus [14], stromal reectivity [15], and pre-clinical
stromal corneal inltrates [16,17]. In vivo confocal microscopy can
* Corresponding author at: University Hospitals Eye Institute, 11100 Euclid Ave
also examine corneal layers during contact lens wear, after
Lakeside 4126C, Cleveland, Ohio 44106, United States. refractive surgery, and determine the effect of immune-altered
E-mail address: loretta.szczotka@uhhospitals.org (L. Szczotka-Flynn). disease states [1820].

http://dx.doi.org/10.1016/j.clae.2016.11.006
1367-0484/ 2016 British Contact Lens Association. Published by Elsevier Ltd. All rights reserved.

Please cite this article in press as: X. Zhang, et al., The impact of lens care solutions on corneal epithelial changes during daily silicone hydrogel
contact lens wear as measured by in vivo confocal microscopy, Contact Lens & Anterior Eye (2016), http://dx.doi.org/10.1016/j.clae.2016.11.006
G Model
CLAE 925 No. of Pages 9

2 X. Zhang et al. / Contact Lens & Anterior Eye xxx (2016) xxxxxx

Some of the more recent applications of confocal microscopy analysis, bacterial bioburden from lenses collected after 1 and 5
have been in the analysis of the corneal epithelium. For example, months of wear and storage cases collected after 2 and 5 months of
mean diameter of supercial epithelial cells have been shown to use were assessed as exploratory covariates potentially impacting
increase with age using confocal microscopy [21]. In clinical disease, epithelial cellular changes. Bacterial bioburden was dened as
confocal microscopy has been used to assess corneal epithelial substantial if there was presence of high levels of commensal
changes after refractive surgery and collagen crosslinking [22], and ocular biota or organisms of low pathogenicity, and/or moderate
Gabbriellini et al. recently assessed epithelial integrity, density, and levels (10 colony forming units (CFU)) of pathogenic organisms.
sub-basal plexus morphology in Sjgren's syndrome patients using The levels of coagulase negative staphylococci (CNS) considered to
confocal microscopy [23]. Of relevance to this study, confocal be substantial varied by the device cultured. If the CFU count of
microscopy was used by Situ et al. to study hyper-reective corneal CNS cultured from the lenses was 10, it was considered
epithelial cells during and after contact lens use [24], and Harvey substantial. If the CFU count of CNS cultured from the case or
et al. found hyper-reective supercial corneal epithelial cells in residual case uid was 60, it was considered substantial.
participants exposed to a specic contact lens/solution combination,
and suggested a correlation between solution-induced corneal 2.1. Confocal microscopy
staining (SICS) and hyper-reective cells [25].
Due to the physical proximity of contact lenses on the ocular The ConfoScan4 (Nidek Technologies, Srl, Padova, Italy) and 40X
surface, the corneal epithelium, as opposed to the deeper stromal probe with a gel immersion exam was used in this study. All
or endothelial layers, is most likely to demonstrate morphological participants had a baseline scan at the initial visit and again at the 5
changes with lens wear. Two indices have been proposed capable month visit, to allow a sufcient time period of use in the lens and
of detecting cell stress at the epithelial corneal surface; supercial care products to detect cellular changes. Scans were done by a
epithelial cell hyper-reectivity and basal cell density [23,25]. We single ophthalmic photographer in the University Hospitals Eye
utilize them here as a mechanism to detect corneal cellular Institute. Scans were done at the end of each study visit with a daily
changes secondary to variables related to daily contact lens wear disposable 9.0 mm +0.50 D One-Day Acuvue (Vistakon, Jackson-
including use of care solutions and microbial contamination of ville, FL) lens worn during the probe immersion exam to lessen
lenses. Specic goals of the current study were to use confocal epithelial trauma to the subject and provide better epithelial
microscopy to determine whether contact lenses in combination images [28]. Following placement of a drop of the topical
with care solutions over ve months of wear impacted the corneal anesthetic (proparacaine) and placement of a bandage contact
epithelium, controlling for confounding effects of other stress lens, a drop of Genteal1 gel (Alcon, Fort Worth, TX) was placed on
factors, such as microbial bioburden, when possible. the central cornea. Topical anesthetic was used as we found that
participants remained more comfortable and were more coopera-
2. Materials and methods tive when imaging both eyes, which ultimately resulted in better
images, and topical proparacaine has not been shown to induce any
This was an ancillary study to the Daily Wear Corneal Inltrative cellular changes on confocal microscopy [13]. The objective piece
Event (DWCIE) Study, which was a prospective, investigator masked, of the confocal microscope was placed in contact with the gel.
cohort study with an embedded randomized clinical trial of solution Three complete scans of the entire central cornea from the
use. In the DWCIE Study, 218 participants were t to the Air Optix 1 endothelium to the epithelium in 5 mm steps were obtained from
Night & Day1 Aqua Lens (lotralcon A, Alcon, Duluth, GA) for up to each eye. The images were electronically stored and sent to the
30 days of daily wear with monthly disposal, and followed for 1 year Cornea Image Analysis Reading Center (CIARC) for masked
at the University Hospitals Eye Institute. The DWCIE and all ancillary evaluation. The CIARC identied the best image(s) for each
studies were approved by the University Hospitals Case Medical respective corneal layer, saved and labeled these images for
Center Institutional Review Board and followed all the Tenets of the masked analysis.
Declaration of Helsinki. All participants provided written informed Epithelial images were analyzed by masked, experienced
consent prior to participation. readers. The number of epithelial hyper-reective cells were
At the baseline DWCIE study visit, enrolled participants were independently read and counted by authors XZ and CM, who were
block randomized to one of two solutions: a preserved multipur- masked to solution group and all other covariates. In the
pose (MPS) solution (Renu Fresh, Bausch & Lomb, Rochester, NY) or adjudication process, a third masked reader, author JL, reviewed
a 1-step hydrogen peroxide care system (Clear Care1; Alcon, Fort the ndings of the two masked readers. If the number of epithelial
Worth, TX) using rub and rinse regimens with companion hyper-reective cells as read by the rst two readers differed by <5
rewetting drops. No wash out period from previous lens wear cells per eye-visit, the nal number of cells was taken as the
was required. At baseline, 50% of participants habitually used a average of the 2 readings. When differences between the counts
PHMB-preserved MPS care product, 21% used a polyquaternium- were >5, the adjudicator re-interpreted the ndings and entered
preserved MPS product, and 10% reported using a peroxide based the nal count. If the images were unanalyzable they were
care product. Enrolled participants returned for follow-up visits excluded from analysis. Fig. 1 displays a representative image of
after 2 weeks, and after 1, 2, 5, 8 and 12 months of daily wear; supercial epithelial hyper-reective cells.
however, in the ancillary study, confocal microscopy images were The CIARC imported the confocal images into Konan Analysis
taken only at the baseline and 5 month visit, therefore, data up to software (Konan KSS300 S v2.62, Konan Medical, Inc.). The Konan
the 5 month visit are used in this analysis. A modied version of the Center method was used to analyze epithelial basal cell density.
Institute for Eye Research (IER) grading scale was used for corneal Two masked readers (adjudicated by third in cases of >5%
staining to quantify staining density in each of 5 corneal zones, as discrepancy) selected the largest area of contiguous cells in the
was done in our previous studies [26]. Solution-induced corneal image where the centers could be accurately distinguished, and
staining (SICS) was dened as present when micropunctate stain marked the center of all contiguous, analyzable cells. One to three
encompassing more than 15% surface area was present in at least 4 unique images were analyzed per time point to maximize the total
zones. area and number of cells analyzed per time point. The software
Aerobic cultures were routinely performed on the ocular then calculated the mean epithelial basal cell density and standard
surface (lid margins and conjunctival surface), worn contact deviation (cells/mm2). Fig. 2 displays a representative image of
lenses and storage cases as previously published [27]. In this epithelial basal cells.

Please cite this article in press as: X. Zhang, et al., The impact of lens care solutions on corneal epithelial changes during daily silicone hydrogel
contact lens wear as measured by in vivo confocal microscopy, Contact Lens & Anterior Eye (2016), http://dx.doi.org/10.1016/j.clae.2016.11.006
G Model
CLAE 925 No. of Pages 9

X. Zhang et al. / Contact Lens & Anterior Eye xxx (2016) xxxxxx 3

For basal epithelial cell density, the change between the 5 month
and baseline values were calculated. For analysis of supercial
epithelial cell hyper-reectivity, only the 5 month measurements
were compared because the outcome counts were very low with
probable variation in the corneal point analyzed between the two
visits, and thus a change value would not be clinically meaningful.
We assessed if the number of hyper-reective cells or change in
basal epithelial cell density signicantly differed when stratied by
the covariates in Table 1. Univariate comparisons were done using
the t-test for normally distributed data; otherwise, the non-
parametric Wilcoxon test was used. Hyper-reective cell counts
were also stratied into a binary covariate (no hyper-reective
cells or 1 hyper-reective cells) and used in univariate or
multivariate logistic regression. The chi-square test and Fishers
exact test were also used for comparisons of categorical variables.
Signicant covariates (p < 0.05) and biologically plausible cova-
riates of age and sex were then entered into multivariable linear or
logistic regression models; solution use and SICS are known to be
Fig 1. Representative image of epithelial hyper-reective cells.
highly correlated therefore both were not used in the same model
due to potential multicollinearity. P-values were adjusted for
2.2. Denition of clinical covariates and statistical analysis multiple uses of the data. All analyses were done using SAS 9.3.

Multiple clinical covariates (Table 1) were analyzed to 3. Results


determine if there was any effect on the supercial or basal
epithelium as viewed by confocal microscopy. These covariates Baseline demographics of the overall cohort and results of the
were drawn from any visit up to and including the 5 month visit main analyses have been published [27]. In brief, the participants
and considered positive if present on at least one of these visits. represented standard contact lens wearers found in routine clinical
Outcome measurements were drawn from one eye of each patient. practices. Participants eyes were clinically normal with no signs of
anterior segment disease or abnormal slit lamp signs. The mean
age of participants was 34.7 years (range 16.5 66.3) and 50
participants (22.9%) were 25 years of age. The majority of
participants were female (70%), Caucasian (59%), and existing or
recent lens wearers (88%). Almost half (44%) of participants had
experienced previous adverse events with contact lens wear,
which were equally distributed between the lens care groups by
design via block randomization. Of the 218 participants in the
parent DWCIE Study, 215 had baseline and 165 had 5-month in vivo
confocal microscopy examinations. Fig. 3 outlines the ow of the
participants and their available images throughout this ancillary
study by lens care group. Data on 139 participants were available
for analysis of supercial epithelial cells, while data on 92
participants were available for epithelial basal cell density
analyses.

3.1. Supercial epithelial cell hyper-reectivity

For the analysis of supercial epithelial cell hyper-reectivity,


there were 135 and 139 participants with readable images
available at the baseline and 5 month visits, respectively. Fig. 3
lists numbers of available images by solution group. The mean
number of hyper-reective cells within the best scan (representing
a 460  345 mm inspected eld in the central cornea) for all
participants was 0.45 (range 08) and 0.56 (range 06) at the
baseline and 5 month visits, respectively. The majority of
participants had zero hyper-reective cells visible at baseline
(69%) or 5 months (67%). Univariate analyses showed statistically
signicant differences in number of hyper-reective cells when
stratied by solution type, presence of SICS staining, and lens
bioburden. Table 2 provides the baseline and 5 month data for
mean number of supercial hyper-reective cells counted in each
central scan for staining, bioburden and solution covariates. At the
5 month visit, MPS users and SICS stainers had a greater mean
number of hyper-reective cells compared to the peroxide users
and non-SICS stainers, respectively. There were no differences
noted between the randomized solution groups at baseline, prior
Fig. 2. Representative image of epithelial basal cells. to randomization, however, SICS stainers at baseline had more

Please cite this article in press as: X. Zhang, et al., The impact of lens care solutions on corneal epithelial changes during daily silicone hydrogel
contact lens wear as measured by in vivo confocal microscopy, Contact Lens & Anterior Eye (2016), http://dx.doi.org/10.1016/j.clae.2016.11.006
G Model
CLAE 925 No. of Pages 9

4 X. Zhang et al. / Contact Lens & Anterior Eye xxx (2016) xxxxxx

Table 1
Clinical Covariates.

Corneal Staining
Solution Induced Corneal type Staining (SICS) vs. not
Overall Grade 2 Corneal Staining vs. not
Two or more visits (over 5 months) with increased corneal staining compared to baseline
Two or more visits (over 5 months) where less corneal staining was noted compared to baseline

Conjunctival Staining
Grade 1 Circumlimbal lissamine green staining vs. not
Grade 2 Circumlimbal lissamine green staining vs. not
Grade 1 General lissamine green staining vs. not
Grade 2 General lissamine green staining vs. not

Microbiology
Substantial bioburden on lids, conjunctivae, lenses or cases vs. not
Substantial CNS or gram positive bioburden on lids or conjunctivae vs. not
Normal bioburden on lids or conjunctivae at 5 months and substantial bioburden at baseline
Substantial bioburden on lids or conjunctivae at 5 months and normal bioburden at baseline

Solution Use
MPS vs. Peroxide

DWCIE Baseline/Randomization visit (n = 218)


Confocal microscopy examinations (n=215)

Preserved multipurpose (MPS) solution Hydrogen peroxide care system


(n=107) (n=108)

BASELINE: BASELINE
Readable superficial epithelial images (n=67) Readable superficial epithelial images (n=68)
Readable basal epithelial cell images (n =62 ) Readable basal epithelial cell images (n =54 )

5 month visit (n=82) 5 month visit (n=83)


Readable superficial epithelial images (n=71) Readable superficial epithelial images (n=68)
Readable basal epithelial cell images (n =63) Readable basal epithelial cell images (n = 64)

Number of paired basal cell image sets Number of paired basal cell image sets
available (n =47) available (n =45)

Fig. 3. Flow of participants and their available images by lens care group.

hyper-reective cells. Substantial lens bioburden was associated covariates, multivariate logistic regression was used. After
with a lower mean hyper-reective cell count. Chi-square analyses controlling for age and sex, the MPS and presence of SICS remained
were then performed to determine if there were differences in the signicantly associated with the presence of hyper-reective cells
number of participants with 1 hyper-reective cells within these (Table 3). Participants who used the MPS were 2.7-fold more likely
same covariates. Only solution type and SICS staining remained to have hyper-reective cells while presence of SICS was associated
signicantly associated with presence of hyper-reective cells. with a 3.4-fold increased odds of having hyper-reective cells.
More participants using MPS had 1 hyper-reective cells in either Recognizing that SICS is most prevalent in the corneal
eye compared to peroxide users at 5 months (44% vs. 22%; periphery, and the confocal probe is placed centrally, we assessed
p = 0.006). Similarly at 5 months, more SICS participants had 1 SICS in the central corneal zone at the 5 month visit for association
hyper-reective cells compared to non-SICS participants (57% vs. with hyper-reectivity at this visit using Fisher exact tests. Central
29%; p = 0.010). SICS stainers were dened as those that had >45% surface area of
To ascertain whether solution use and presence of SICS micropunctate staining in the central corneal zone. Of the 6
remained associated with increased hyper-reectivity after participants that had evidence of central SICS and usable
controlling for potentially confounding biologically plausible supercial confocal images at the 5 month visit, 5 of them (83%)

Please cite this article in press as: X. Zhang, et al., The impact of lens care solutions on corneal epithelial changes during daily silicone hydrogel
contact lens wear as measured by in vivo confocal microscopy, Contact Lens & Anterior Eye (2016), http://dx.doi.org/10.1016/j.clae.2016.11.006
G Model
CLAE 925 No. of Pages 9

X. Zhang et al. / Contact Lens & Anterior Eye xxx (2016) xxxxxx 5

Table 2
Comparisons of Supercial Hyper-reective Epithelial Cells at Baseline and 5 Months for Solution Group and Solution Induced Corneal Staining.

Covariate of Interesta Time Point Mean (standard deviation) with sample sizeb available for each comparison and timepoint P value

Randomized Solution Group MPS Peroxide


n = 67 at baseline n = 68 at baseline
n = 71 at 5 months n = 68 at 5 months
Baseline 0.40 (0.76) 0.51 (1.25) 0.780
5 months 0.70 (1.22) 0.32 (0.82) 0.007
Solution induced corneal staining YES NO
n = 25 at baseline n = 110 at baseline
n = 21 at 5 months n = 118 at 5 months
Baseline 0.72 (1.02) 0.40 (1.03) 0.010
5 months 1.10 (1.65) 0.41 (0.88) 0.009
Overall Grade 2 Staining YES NO
n = 43 at baseline n = 92 at baseline
n = 45 at 5 months n = 94 at 5 months
Baseline 0.59 (1.44) 0.40 (0.78) 0.782
5 months 0.60 (0.99) 0.47 (1.09) 0.128
Improved Corneal Staining Over Time YES NO
n = 8 at baseline n = 127 at baseline
n = 10 at 5 months n = 129 at 5 months
Baseline 0.25 (0.38) 0.47 (1.06) 0.968
5 months 0.35 (0.94) 0.53 (1.06) 0.377
Worsened Corneal Staining Over Time YES NO
n = 27 at baseline n = 108 at baseline
n = 33 at 5 months n = 106 at 5 months
Baseline 0.31 (0.71) 0.50 (1.10) 0.324
5 months 0.74 (1.40) 0.44 (0.92) 0.202
Signicant lens bioburden YES NO
n = 19 at baseline n = 116 at baseline
n = 21 at 5 months n = 118 at 5 months
Baseline 0.32 (0.73) 0.48 (1.08) 0.368
5 months 0.10 (0.26) 0.59 (1.12) 0.029
Signicant case bioburden YES NO
n = 57 at baseline n = 67 at baseline
n = 61 at 5 months n = 68 at 5 months
Baseline 0.47 (1.22) 0.45 (0.90) 0.640
5 months 0.52 (1.07) 0.50 (1.01) 0.827
a
covariates refer to variables dened in Table 1 drawn from data across the 5 months of the study period.
b
one eye per participant was analyzed thus sample sizes refers to participants or eyes.

Table 3
Multivariate Logistic Regression Models with Hyper-reective Cells (1 cell) at 5 months as Outcome.

A: Model Exploring Solution Use as Covariate, signicant covariate in bold


Covariate Odds Ratio 95% Wald Condence Limits P value
Randomized Solution Group 2.68 1.27 5.65 0.010
(multipurpose vs. peroxide)
Age 0.81 0.34 1.96 0.645
(25 vs. >25)
Sex 0.63 0.27 1.44 0.271
(male vs. female)

B: Model Exploring Solution Induced Corneal Staining (SICS) as Covariate; signicant covariate in bold
Covariate Odds Ratio 95% Wald Condence Limits P value
SICS 3.40 1.29 8.97 0.013
(yes vs. no)
Age 0.74 0.31 1.8 0.511
(25 vs. >25)
Sex 0.60 0.26 1.4 0.229
(male vs. female)

had 1 hyper-reective cells compared to 37/141 (26%) of non- of interest in Table 1 and basal epithelial cell density are shown in
central SICS participants, and this difference was signicant Table 4. Improved corneal staining was associated with a change
(p = 0.008). (increase) in epithelial basal cell density from baseline to 5 months.
This table can be interpreted as follows; there was an increase in
3.2. Basal epithelial cell density epithelial basal cell density from baseline to 5 months (by a mean
of 585 cells/mm2) in the group that showed improved corneal
As shown in Fig. 3, there were 92 participants with paired staining over the rst 5 months of the study compared to those that
images from both the baseline and 5 month time-points; only did not. Participants with substantial storage case bioburden
these 92 participants that had usable paired images were used in through the 5 month visit had a signicant decrease in mean basal
the basal cell analyses. Univariate associations between covariates cell density over time. The multivariate linear regression model

Please cite this article in press as: X. Zhang, et al., The impact of lens care solutions on corneal epithelial changes during daily silicone hydrogel
contact lens wear as measured by in vivo confocal microscopy, Contact Lens & Anterior Eye (2016), http://dx.doi.org/10.1016/j.clae.2016.11.006
G Model
CLAE 925 No. of Pages 9

6 X. Zhang et al. / Contact Lens & Anterior Eye xxx (2016) xxxxxx

Table 4
Change in Basal Epithelial Cell Density from Baseline to the 5 Month Visit by Various Covariates of Interest.

Covariate of Interest Mean (standard deviation)* P value

Randomized Solution Group MPS Peroxide


n = 47 n = 45
58.45 (623.26) 43.41 (740.10) 0.507
Solution induced corneal staining YES NO
n = 14 n = 78
43.03 (615.88) 2.46 (695.60) 0.892
Overall Grade 2 Staining YES NO
n = 32 n = 60
58.77 (638.39) 18.11 (706.52) 0.443
Improved Corneal Staining Over Time YES NO
n=4 n = 88
584.88 (89.58) 17.56 (684.77) 0.030
Worsened Corneal Staining Over Time YES NO
n = 22 n = 70
15.5 (543.82) 6.47 (722.28) 0.982
Signicant lens bioburden YES NO
n = 15 n = 77
17.48 (1067.12) 13.72 (587.74) 0.305
Signicant case bioburden YES NO
n = 36 n = 49
261.16 (628.76) 173.21 (684.74) 0.005

Change value derived by subtracting baseline value from the 5 month value, positive values indicate increased cell density over time.

(Table 5) including age and sex found that only abnormal compared to the peroxide group whereas presence of SICS was
microbiology within storage cases remained independently associated with a 3.4 fold increased risk of having hyper-reective
associated with basal cell density changes from baseline to 5 cells. These covariates could not be assessed independently and we
months. Specically, the basal cell density decreased by 366 cells/ believe the mechanism of preservative induced corneal staining
mm2 (about 6%) in those participants with storage case bioburden. and epithelial cell hyper-reectivity are probably linked.
Analysis of solution and other staining variables did not show Hyper-reective corneal epithelial cells have not been exten-
signicant effects or trends on change in basal epithelial cell sively reported or studied in the literature and to our knowledge
density (p > 0.100) when controlling for storage case bioburden. this is the rst published study on the link between hyper-
Although lens bioburden was not signicant in univariate analysis, reective cells and solution use although others have previously
multivariate analysis showed similar trends to storage case found the same association [13,29]. Scanning electron microscopy
bioburden. That is, basal cell density decreased by 416 cells/ and confocal microscopy have identied different shades of
mm2 in those participants with substantial lens bioburden supercial cells. Dark cells possess fewer surface features, and
(p=0.017) and in this model, improved corneal staining was are assumed to have lost microvilli [3032]. Lighter cells may be
associated with increased epithelial basal cell density (p = 0.025) the most recently recruited to the supercial epithelial cell layer,
because they have surface features consistent with microvilli as it
4. Discussion has been proposed that microdesmosomes could explain brighter
cell membranes of epithelial cells [33,34]. Therefore, darker cells
We used confocal microscopy to assess hyper-reectivity of the are presumed to represent mature cells that are in process of being
supercial corneal epithelial cells, which has been reported in desquamated [30,32]. In that regard, Wilson et al. suggested that
several studies to be associated with various contact lens and differences in cytoplasm reectivity could represent stages of
solution combinations [13,29]. We found that the presence of progression towards cell death, with the less reective or darker
hyper-reective supercial epithelial cells was associated with a cells about to desquamate [35].
PHMB-preserved MPS as well as SICS. We calculated that MPS Hyper-reective cells are different than the lightly shaded
users were 2.7 fold more likely to have hyper-reective cells epithelial cells previously described using confocal microscopy. In

Table 5
Linear Regression Models with Basal Epithelial Cell Density Change from Baseline to 5 months as Outcome, signicant covariates in bold.

Covariate Estimate 95% CI P value


Improved corneal staining vs. not 564.50 188.3 to 1317.2 0.142
Substantial bacterial bioburden on cases vs. not 366.70 649.5 to 83.9 0.011
Age 156.00 481.5 to 169.5 0.348
(25 vs. >25)
Sex 162.60 439.2 to 113.9 0.249
(male vs. female)

Covariate Estimate 95% CI P value


Improved corneal staining vs. not 708.77 87.09 to 1330.45 0.025
Substantial bacterial bioburden on lenses vs. not 416.17 -758.21 to 74.13 0.017
Age 163.51 461.66 to 134.64 0.282
(25 vs. >25)
Sex 135.12 392.51 to 122.27 0.304
(male vs. female)

Please cite this article in press as: X. Zhang, et al., The impact of lens care solutions on corneal epithelial changes during daily silicone hydrogel
contact lens wear as measured by in vivo confocal microscopy, Contact Lens & Anterior Eye (2016), http://dx.doi.org/10.1016/j.clae.2016.11.006
G Model
CLAE 925 No. of Pages 9

X. Zhang et al. / Contact Lens & Anterior Eye xxx (2016) xxxxxx 7

known circumstances of ocular surface disease, hyper-reective [53]. However, Sindt et al., using laser-scanning confocal micros-
cells (in various locations within corneal tissue) have been linked copy (which uses a wavelength of 670 nm), found that uorescein
to pro-inammatory states such as viral keratitis [36,37], pterygi- staining did not interfere with confocal imaging of corneal
um [38], vernal keratoconjunctivitis [39], Salzmanns nodular epithelium, sub-basal nerves, or dendritic immune cells [54].
degeneration [40], and ocular changes in systemic inammatory Schneider, using white-light slit scanning confocal microscopy as
states such as Crohns disease[41] and Stevens Johnsons Syndrome we did, reported no difference in cellular appearance between eyes
[42]. Hamrah et al. proposed that an increase in hyper-reective that had sodium uorescein from those without, implying that
epithelial cells in patients with herpes simplex viral keratitis is due sodium uorescein and its residue are not associated with hyper-
to supercial epithelial cell desquamation [37]. Outside of active reective cells. In fact, even after provocatively inducing SICS, the
ocular surface disease, hyper-reectivity has also been shown to be use of sodium uorescein prior to confocal microscopy did not
associated with dry eye and normal cellular turnover [43,44]. affect appearance of hyper-reective cells [13]. While topical
Additionally, hyper-reective supercial epithelial cells have been anesthetics have been associated with inhibition of corneal
associated with use of certain preserved anti-glaucoma eye drops epithelial migration and disruption of supercial corneal epithelial
and benzalkonium chloride [4547]. Therefore, although there is microvilli [55,56], Schneider showed that topical anesthetic was
little data on mechanisms causing hyper-reectivity of supercial not linked to the appearance of hyper-reective cells because even
epithelial cells, in general, hyper-reectivity has been postulated to when confocal microscopy was done without anesthetic, hyper-
correlate with inammation or toxicity. When present on the reective cells were observed [13].
supercial cornea in greater numbers in patients that use Basal epithelial cell density was also evaluated with confocal
preserved anti-glaucoma eye drops or benzalkonium chloride, microscopy. These deep epithelial cells undergo mitosis with cell
its presumed the changes observed are a form of preservative differentiation as they move anteriorly. We found a mean epithelial
toxicity [4547]. Additionally, hyper-reective stromal keratocytes basal cell density across participants of 6147  171 cells/mm2. This
as imaged on confocal microscopy are assumed to correspond to a nding is consistent with previous reports; Schneider had
state of metabolic activation induced by pro-inammatory reviewed others work and reported basal epithelial cell density
cytokines [48,49]. Thus, although the signicance of hyper- using confocal microscopy ranged from 3601  408 cells/mm2 to
reective cells is not known, in other layers of the cornea, they 6188  427 cells/mm2 [13]. Participants with substantial bacterial
are presumed to correspond to a pro-inammatory state as well. burden on their lenses or within their storage cases were
In contact lens wearers, Harvey et al. found hyper-reective independently associated with decreased basal epithelial cell
supercial corneal epithelial cells in participants exposed to a density. Only one of our staining variables was associated with
specic contact lens/solution combination, and suggested a changes in basal cell density. Participants with decreased corneal
correlation between SICS and hyper-reective cells [25]. Schneider, staining over time were associated with an increase in their basal
expanded upon these ndings and found that normal non-contact epithelial cell density. In contrast, worsening of corneal staining
lens wearers had a few random supercial hyper-reective cells over time, SICS or overall staining were not associated with
which did not change over 9 months [13]. She found increased decreases in basal cell density. Therefore, further research is
numbers of hyper-reective cells associated with higher levels of required in this regard to understand the relationship between
SICS in a PHMB-based solution compared to a polyquaternium- corneal staining and impact on basal epithelial cell density.
based and hydrogen peroxide-based solution during short term Corneal epithelial basal cell density has been identied as an
use (2 weeks-1 month) [13]. SICS has been observed with certain indicator of cell stress in other studies. Many of these studies
combinations of contact lens solutions and silicone hydrogel lenses collectively suggest that inammatory conditions (atopic and
[4,50], and has been hypothesized by some to be a result of a toxic vernal keratoconjunctivitis, dry eyes, and overnight orthokeratol-
reaction [4,51]. SICS is typically represented by ne micro- ogy) can result in decreased epithelial basal cell density [39,57
punctate uorescein staining of the supercial epithelium usually 59]. In eyes with atopic keratoconjunctivitis and dryness, corneal
in the corneal periphery, is typically asymptomatic, and is most staining was found to be correlated with decreased basal epithelial
evident during the rst 24 h of lens wear. It is well established cell density [58]. Additionally, Zhang et al. discovered a signicant
that certain lens types in combination with the same polyhexa- decrease in basal epithelial cell density in mild, moderate, and
methylene biguanide-based solution we used in our study, is severe dry eye participants when compared with controls [58]. The
associated with signicantly more SICS than other non-PHMB authors suggested that increased tear osmolarity elicits pro-
preserved solutions due to uptake and release of preservatives into inammatory cytokines, which decrease basal cell density even
the lens matrix [4,52], and hydrogen peroxide-based solutions are before disrupting epithelial barrier function. Hamrah et al. noted
associated with the least amount of SICS [32,51]. that basal epithelial cell density was lower in herpes simplex viral
In our study, based on the fact that the percentage of keratitis patients when compared with controls [37], and others
participants that have hyper-reective cells is greater in the described a signicant decrease in basal epithelial cell density in
PHMB-based MPS and SICS participants, and much greater in clear corneas adjacent to pterygium when compared with controls
participants with central SICS, our data is consistent with the [60]. Therefore, ocular surface inammation appears to be
hypothesis that the hyper-reective cells are likely representative associated with decreased epithelial basal cell density, and our
of the stained cells induced by the solution. Our ndings reect lens bioburden ndings support this theory.
that hyper-reective cells co-exist in participants with SICS. We One limitation of the study includes no washout period prior to
conclude that hyper-reective cells occurred more commonly in study entry. However, this should not inuence the hyper-
our MPS users, especially in MPS users who exhibited SICS. The reective cell analyses because only the 5 month visit data were
hyper-reective cells may reect the same cells that will stain with used, thus not inuenced by habitual lens or solution type. The
sodium uorescein. baseline basal cell density may have been affected by habitual lens
Since sodium uorescein and topical anesthetic were used prior use, and changes seen by 5 months could be inuenced by retting
to confocal microscopy in this study, it has been suggested that all participants to a new lens type or solution system. However, we
these topical agents could be responsible for the hyper-reectivity. controlled for factors associated with lens type or solution use at
In that regard, Mocan et al. suggested that sodium uorescein baseline (staining and bacterial bioburden) in the linear regression
damaged epithelial tight junctions which resulted in increased models (Table 5), therefore, outcomes of this analyses were not
hyper-reective cells from epithelial sodium uorescein diffusion confounded.

Please cite this article in press as: X. Zhang, et al., The impact of lens care solutions on corneal epithelial changes during daily silicone hydrogel
contact lens wear as measured by in vivo confocal microscopy, Contact Lens & Anterior Eye (2016), http://dx.doi.org/10.1016/j.clae.2016.11.006
G Model
CLAE 925 No. of Pages 9

8 X. Zhang et al. / Contact Lens & Anterior Eye xxx (2016) xxxxxx

5. Conclusions [18] D.V. Patel, C.N. McGhee, In vivo confocal microscopy of human corneal nerves
in health, in ocular and systemic disease, and following corneal surgery: a
review, Br. J. Ophthalmol. 93 (7) (2009) 853860.
Confocal microscopy can identify multiple cellular changes in [19] J.W. McLaren, et al., Automated assessment of keratocyte density in clinical
the cornea during daily contact lens wear. We have identied confocal microscopy of the corneal stroma, J. Microsc. 229 (Pt 1) (2008) 2131.
potential cellular stresses on the ocular surface by recognition of [20] T. Tervo, J. Moilanen, In vivo confocal microscopy for evaluation of wound
healing following corneal refractive surgery, Prog. Retin Eye Res. 22 (3) (2003)
hyper-reective supercial epithelial cells and decreases in basal 339358.
epithelial cell density. We propose that SICS staining secondary to a [21] C. Gambato, et al., Aging and corneal layers: an in vivo corneal confocal
PHMB-preserved MPS results in hyper-reective supercial microscopy study, Graefes Arch. Clin. Exp. Ophthalmol. 253 (2) (2014)
267275.
epithelial cells noted during confocal microscopy. Furthermore, [22] G. Alessio, et al., Confocal microscopy analysis of corneal changes after
lens microbial bioburden is associated with decreases in basal photorefractive keratectomy plus cross-linking for keratoconus: 4-year
epithelial cell density; one explanation may be that this represents follow-up, Am J Ophthalmol 158 (3) (2014) 476484 (e1).
[23] G. Gabbriellini, et al., In vivo confocal scanning laser microscopy in patients
decreased proliferation or increased apoptosis of corneal basal
with primary Sjogren's syndrome: a monocentric experience, ModRheumatol
epithelial cells. Although these ndings are exploratory in nature (2014) 15.
and deserve further study, the confocal microscope is a powerful [24] P. Situ, et al., Confocal imaging of hyper-reective corneal epithelial cells
instrument that can detect cellular changes in vivo during contact during and after contact lens wear, Invest. Ophthalmol. Vis. Sci. 52 (2012) (e-
abstract 4698).
lens wear. [25] K.M.N. Harvey, The Effects of Contact Lens Care Solutions on the Corneal
Epithelium: a Comparative Investigation Using Confocal Microscopy, in
Source of funding Department of Vision Science, University of Waterloo, Waterloo, Ontario,
Canada, 2006.
[26] L. Szczotka-Flynn, et al., Risk factors for corneal inltrative events during
This work was supported by Alcon Research Ltd. with indirect continuous wear of silicone hydrogel contact lenses, Invest. Ophthalmol. Vis.
support for laboratories and coordination from the Ohio Lions Eye Sci. 51 (11) (2010) 54215430.
[27] L. Szczotka-Flynn, et al., Corneal inammatory events with daily silicone
Research Foundation and Research to Prevent Blindness. The hydrogel lens wear, Optom. Vis. Sci. 91 (1) (2014) 312.
clinicalTrials.gov identier is NCT00937105. [28] K.Y. Chan, S.W. Cheung, P. Cho, Nidek ConfoScan 4 (z-ring) measurements over
soft contact lenses, Eye Contact Lens 38 (2) (2012) 8085.
[29] S. Schneider, et al., Hyper-reective cells observed by confocal microscopy as
Conicts of interest
an indicator of lens and lens care interactions, Optom. Vis. Sci. (2008) e
abstract 80028.
Dr. LSF has received research support from Alcon Laboratories [30] W.D. Mathers, et al., Morphologic effects of contact lens wear on the corneal
surface, CLAO J. 18 (1) (1992) 4952.
and Johnson & Johnson Vision Care, Inc. in the last 12 months. J.
[31] F. Hoffmann, The surface of epithelial cells of the cornea under the scanning
Kern is employed by Alcon Research. electron microscope, Ophthal. Res. 3 (1972) 207214.
[32] N. Efron, Contact lens-induced changes in the anterior eye as observed in vivo
References with the confocal microscope, Prog. Retin Eye Res. 26 (4) (2007) 398436.
[33] R. Bochert, et al., Contribution to comprehension of image formation in
confocal microscopy of cornea with Rostock cornea module, Br. J. Ophthalmol.
[1] S. Duench, et al., Impact of silicone hydrogel lenses and solutions on corneal 89 (10) (2005) 13511355.
epithelial permeability, Optom. Vis. Sci. 90 (6) (2013) 546556. [34] A. Eckard, J. Stave, R.F. Guthoff, In vivo investigations of the corneal epithelium
[2] M. Gorbet, et al., Human corneal epithelial cell shedding and uorescein with the confocal Rostock Laser Scanning Microscope (RLSM), Cornea 25 (2)
staining in response to silicone hydrogel lenses and contact lens disinfecting (2006) 127131.
solutions, Curr. Eye Res. 39 (3) (2013) 245256. [35] G. Wilson, P.M. Ladage, H.D. Cavanagh, The epithelium in extended wear, in
[3] M. Gorbet, C. Postnikoff, The impact of silicone hydrogel-solution silicone hydrogels: continuous wear contact lens, in: D.F. Sweeney (Ed.),
combinations on corneal epithelial cells, Eye Contact Lens 39 (1) (2013) Butterworth-Heinemann, 2004, pp. 2854.
4247. [36] G. Martone, et al., In vivo confocal microscopy in bilateral herpetic keratitis: a
[4] L. Jones, N. MacDougall, L.G. Sorbara, Asymptomatic corneal staining case report, Eur. J. Ophthalmol. 18 (6) (2008) 994997.
associated with the use of balalcon silicone-hydrogel contact lenses [37] P. Hamrah, et al., Cellular changes of the corneal epithelium and stroma in
disinfected with a polyaminopropyl biguanide-preserved care regimen, herpes simplex keratitis: an in vivo confocal microscopy study,
Optom. Vis. Sci. 79 (12) (2002) 753761. Ophthalmology 119 (9) (2012) 17911797.
[5] J.R. Paugh, et al., A preliminary study of silicone hydrogel lens material and [38] A. Labbe, et al., An in vivo confocal microscopy and impression cytology
care solution bioincompatibilities, Cornea 30 (7) (2011) 772779. evaluation of pterygium activity, Cornea 29 (4) (2010) 392399.
[6] L.C. Posch, M. Zhu, D.M. Robertson, Multipurpose care solution-induced corneal [39] A. Leonardi, et al., Corneal confocal microscopy in patients with vernal
surface disruption and Pseudomonas aeruginosa internalization in the rabbit keratoconjunctivitis, Ophthalmology 119 (3) (2012) 509515.
corneal epithelium, Invest. Ophthalmol. Vis. Sci. 55 (7) (2014) 42294237. [40] A.M. Roszkowska, P. Aragona, R. Spinella, Morphological and confocal
[7] N.C. Tanti, L. Jones, M.B. Gorbet, Impact of multipurpose solutions released investigation on Salzmann nodular degeneration of the cornea, Invest.
from contact lenses on corneal cells, Optom. Vis. Sci 88 (4) (2010) 483492. Ophthalmol. Vis. Sci. 52 (2011) 59105919.
[8] Y.C. Hsiao, et al., Diagnosis of microsporidial keratitis with in vivo confocal [41] G. Ceresara, et al., Corneal involvement in Crohn's disease: an in vivo confocal
microscopy, J. Xray Sci. Technol. 21 (1) (2013) 103110. microscopy study, Cornea 30 (2) (2011) 136142.
[9] A. Labbe, et al., Contribution of in vivo confocal microscopy to the diagnosis [42] L.S. Vera, et al., In vivo confocal microscopic evaluation of corneal changes in
and management of infectious keratitis, Ocul. Surf. 7 (1) (2009) 4152. chronic Stevens-Johnson syndrome and toxic epidermal necrolysis, Cornea 28
[10] P.K. Vaddavalli, et al., Role of confocal microscopy in the diagnosis of fungal (4) (2009) 401407.
and acanthamoeba keratitis, Ophthalmology 118 (1) (2010) 2935. [43] J.M. Benitez del Castillo, et al., An in vivo confocal masked study on corneal
[11] M.O. Price, et al., Effect of incision width on graft survival and endothelial cell epithelium and subbasal nerves in patients with dry eye, Invest. Ophthalmol.
loss after Descemet stripping automated endothelial keratoplasty, Cornea 29 Vis. Sci. 45 (9) (2004) 30303035.
(5) (2010) 523527. [44] H.E. Kaufman, B.A. Barron, M.B. McDonald, The Cornea, 2nd ed., Butterworth-
[12] M.O. Price, et al., Descemet's stripping automated endothelial keratoplasty Heinemann, Newton, 1997.
outcomes compared with penetrating keratoplasty from the Cornea Donor [45] H. Ichijima, et al., Confocal microscopic studies of living rabbit cornea treated
Study, Ophthalmology 117 (3) (2010) 438444. with benzalkonium chloride, Cornea 11 (3) (1992) 221225.
[13] S. Schneider, The Appearance of Hyper-reective Supercial Epithelial Cells [46] A. Labbe, et al., Comparison of toxicological proles of benzalkonium chloride
Observed Using in Vivo Confocal Microscopy, in Department of Vision Science, and polyquaternium-1: an experimental study, J. Ocul. Pharmacol. Ther. 22 (4)
University of Waterloo, Waterloo, Ontario, Canada, 2010. (2006) 267278.
[14] C.W. Sindt, et al., Dendritic immune cell densities in the central cornea [47] G. Martone, et al., An in vivo confocal microscopy analysis of effects of topical
associated with soft contact lens types and lens care solution types: a pilot antiglaucoma therapy with preservative on corneal innervation and
study, Clin. Ophthalmol. 6 (2012) 511519. morphology, Am. J. Ophthalmol. 147 (4) (2009) 725735 (e1).
[15] R.A. Schachar, et al., Subclinical increased anterior stromal reectivity with [48] E. Villani, et al., Corneal involvement in rheumatoid arthritis: an in vivo
topical taprenepag isopropyl, Cornea 32 (3) (2013) 306312. confocal study, Invest. Ophthalmol. Vis. Sci. 49 (2) (2008) 560564.
[16] Y. Sun, et al., Inhibition of corneal inammation by liposomal delivery of short- [49] E. Villani, et al., The cornea in Sjogren's syndrome: an in vivo confocal study,
chain: c-6 ceramide, J. Leukoc Biol. 83 (6) (2008) 15121521. Invest. Ophthalmol. Vis. Sci. 48 (5) (2007) 20172022.
[17] Y. Sun, E. Pearlman, Inhibition of corneal inammation by the TLR4 antagonist [50] G. Andrasko, K. Ryen, Corneal staining and comfort observed with traditional
Eritoran tetrasodium (E5564), Invest. Ophthalmol. Vis. Sci. 50 (3) (2009) and silicone hydrogel lenses and multipurpose solution combinations,
12471254. Optometry 79 (8) (2008) 444454.

Please cite this article in press as: X. Zhang, et al., The impact of lens care solutions on corneal epithelial changes during daily silicone hydrogel
contact lens wear as measured by in vivo confocal microscopy, Contact Lens & Anterior Eye (2016), http://dx.doi.org/10.1016/j.clae.2016.11.006
G Model
CLAE 925 No. of Pages 9

X. Zhang et al. / Contact Lens & Anterior Eye xxx (2016) xxxxxx 9

[51] R.J. Garofalo, et al., Corneal staining and subjective symptoms with [56] D.G. Ezra, B.D. Allan, Topical anaesthesia alone versus topical anaesthesia with
multipurpose solutions as a function of time, Eye Contact Lens 31 (4) intracameral lidocaine for phacoemulsication, Cochrane Database Syst. Rev. 3
(2005) 166174. (2007) CD005276.
[52] N. Dassanayake, R. Schlitzer, C. Carey, Pre-clinical Techniques for Identifying [57] Y. Hu, et al., Corneal in vivo confocal scanning laser microscopy in patients
Incompatibilities Between Contact Lens Materials and Disinfection Systems, with atopic keratoconjunctivitis, Ophthalmology 115 (11) (2008) 20042012.
ARVO, 2004, pp. 1572 (B383). [58] X. Zhang, et al., Tear dynamics and corneal confocal microscopy of subjects
[53] M.C. Mocan, M. Irkec, Fluorescein enhanced confocal microscopy in vivo for with mild self-reported ofce dry eye, Ophthalmology 118 (5) (2011) 902907.
the evaluation of corneal epithelium, Clin. Exp. Ophthalmol. 35 (1) (2007) [59] A. Nieto-Bona, et al., Long-term changes in corneal morphology induced by
3843. overnight orthokeratology, Curr. Eye Res. 36 (10) (2011) 895904.
[54] C.W. Sindt, et al., Effects of uorescein staining on laser in vivo confocal [60] Y. Wang, et al., In vivo confocal microscopic evaluation of morphologic changes
microscopy images of the cornea, J. Ophthalmol. 2012 (2012) 541974. and dendritic cell distribution in pterygium, Am. J. Ophthalmol. 150 (5) (2010)
[55] R.L. Grant, D. Acosta, Comparative toxicity of tetracaine, proparacaine and 650655 (e1).
cocaine evaluated with primary cultures of rabbit corneal epithelial cells, Exp.
Eye Res. 58 (4) (1994) 469478.

Please cite this article in press as: X. Zhang, et al., The impact of lens care solutions on corneal epithelial changes during daily silicone hydrogel
contact lens wear as measured by in vivo confocal microscopy, Contact Lens & Anterior Eye (2016), http://dx.doi.org/10.1016/j.clae.2016.11.006

Das könnte Ihnen auch gefallen