Beruflich Dokumente
Kultur Dokumente
2.
Steam Sterilization
Of all the methods available for sterilization, moist heat in the form of saturated
steam under pressure is the most widely used and the most dependable. Steam
sterilization is nontoxic, inexpensive rapidly microbicidal, sporicidal, and rapidly heats
and penetrates fabrics. Like all sterilization processes, steam sterilization has some
deleterious effects on some materials, including corrosion and combustion of lubricants
associated with dental hand pieces; reduction in ability to transmit light associated with
laryngoscopes; and increased hardening time (5.6 fold) with plaster-cast.
The two basic types of steam sterilizers (autoclaves) are the gravity displacement
autoclave and the high-speed prevacuum sterilizer. In the former, steam is admitted at the
top or the sides of the sterilizing chamber and, because the steam is lighter than air, forces
air out the bottom of the chamber through the drain vent. The gravity displacement
autoclaves are primarily used to process laboratory media, water, pharmaceutical
products, regulated medical waste, and nonporous articles whose surfaces have direct
steam contact. For gravity displacement sterilizers the penetration time into porous items
is prolonged because of incomplete air elimination. This point is illustrated with the
decontamination of 10 lbs of microbiological waste, which requires at least 45 minutes at
121oC because the entrapped air remaining in a load of waste greatly retards steam
permeation and heating efficiency . The high-speed prevacuum sterilizers are similar to
the gravity displacement sterilizers except they are fitted with a vacuum pump (or
ejector) to ensure air removal from the sterilizing chamber and load before the steam is
admitted. The advantage of using a vacuum pump is that there is nearly instantaneous
steam penetration even into porous loads. The Bowie-Dick test is used to detect air leaks
and inadequate air removal and consists of folded 100% cotton surgical towels that are
clean and preconditioned. A commercially available Bowie-Dick-type test sheet should
be placed in the center of the pack. The test pack should be placed horizontally in the
front, bottom section of the sterilizer rack, near the door and over the drain, in an
otherwise empty chamber and run at 134oC for 3.5 minutes 813, 819. The test is used each
day the vacuum-type steam sterilizer is used, before the first processed load. Air that is
not removed from the chamber will interfere with steam contact. Smaller disposable test
packs (or process challenge devices) have been devised to replace the stack of folded
surgical towels for testing the efficacy of the vacuum system in a prevacuum
sterilizer.833 These devices are "designed to simulate product to be sterilized and to
constitute a defined challenge to the sterilization process" 819, 834. They should be
representative of the load and simulate the greatest challenge to the load 835. Sterilizer
vacuum performance is acceptable if the sheet inside the test pack shows a uniform color
change. Entrapped air will cause a spot to appear on the test sheet, due to the inability of
the steam to reach the chemical indicator. If the sterilizer fails the Bowie-Dick test, do not
use the sterilizer until it is inspected by the sterilizer maintenance personnel and passes
the Bowie-Dick test.
Another design in steam sterilization is a steam flush-pressure pulsing process,
which removes air rapidly by repeatedly alternating a steam flush and a pressure pulse
above atmospheric pressure. Air is rapidly removed from the load as with the prevacuum
sterilizer, but air leaks do not affect this process because the steam in the sterilizing
chamber is always above atmospheric pressure. Typical sterilization temperatures and
times are 132oC to 135oC with 3 to 4 minutes exposure time for porous loads and
instruments.
Like other sterilization systems, mechanical, chemical, and biological monitors
monitor the steam cycle. Steam sterilizers usually are monitored using a printout (or
graphically) by measuring temperature, the time at the temperature, and pressure.
Typically, chemical indicators are affixed to the outside and incorporated into the pack to
monitor the temperature or time and temperature. The effectiveness of steam sterilization
is monitored with a biological indicator containing spores of Geobacillus
stearothermophilus (formerly Bacillus stearothermophilus). Positive spore test results are
a relatively rare event and can be attributed to operator error, inadequate steam
delivery or equipment malfunctions.
Portable (table-top) steam sterilizers are used in outpatient, dental, and rural
clinics. These sterilizers are designed for small instruments, such as hypodermic syringes
and needles and dental instruments. The ability of the sterilizer to reach physical
parameters necessary to achieve sterilization should be monitored by mechanical,
chemical, and biological indicators.
The oldest and most recognized agent for inactivation of microorganisms is heat.
D-values (time to reduce the surviving population by 90% or 1 log10) allow a direct
comparison of the heat resistance of microorganisms. Because a D-value can be
determined at various temperatures, a subscript is used to designate the exposure
temperature (i.e., D121C). D121C-values for Geobacillus stearothermophilus used to monitor
the steam sterilization process range from 1 to 2 minutes. Heat-resistant non spore-
forming bacteria, yeasts, and fungi have such low D121C values that they cannot be
experimentally measured
Steam sterilization should be used whenever possible on all critical and semi
critical items that are heat and moisture resistant (e.g., steam sterilizable respiratory
therapy and anesthesia equipment), even when not essential to prevent pathogen
transmission. Steam sterilizers also are used in healthcare facilities to decontaminate
microbiological waste and sharps containers but additional exposure time is required in
the gravity displacement sterilizer for these items.
3.
Sterility testing may also be carried out for quality assurance purposes as a means
of continuously monitoring the process, rather than as a pre-release test. The development
of a statistically valid sampling plan is equally important, since it must be able to detect
any deviations from the acceptable contamination rate. Compendial sterility test methods
do not usually contain guidance for the development of this type of sampling plan, but the
statistical principles involved are well documented. The key is to decide exactly what the
sampling plan should be able to detect and then design a plan that will achieve the desired
detection level while taking into account any other relevant characteristics of the
manufacturing and sterilisation processes.
Compendial methods for sterility testing require that a sample be cultured using
two separate media. These are usually fluid thioglycollate medium (FTM), to culture both
anaerobic and some aerobic bacteria, and soybean casein digest medium (SCDM) to
culture fungi and aerobic bacteria. The cultures are incubated for 14 days at 32.5oC and
22.5oC and then examined. Any turbidity in the culture may indicate growth and must be
investigated.
There are two recommended methods for carrying out the test. The first is by
direct inoculation, whereby a small volume of sample is removed aseptically from the
sample unit and inoculated directly into a suitable volume of growth medium prior to
incubation. This method has some significant disadvantages. Firstly, only small volumes
of product can be inoculated into the culture medium, limiting the sensitivity of the test.
Secondly, if the sample appears milky or turbid, it can be very difficult to detect turbidity
caused by microbial growth at the end of the incubation period.
To overcome these drawbacks the recommended method wherever possible is
membrane filtration. Here the sample is passed through a 0.45 m membrane filter and
the filter is then transferred to the culture medium for incubation. Membrane filtration
allows the whole sample, or a composite sample, to be passed through a single filter and
is therefore potentially much more sensitive than direct inoculation. Filtration also
provides an opportunity to rinse away components in the sample that may cause turbidity
and any growth inhibitors, such as antibiotics or preservatives, which may be present.
The membrane filtration method may be carried out using a traditional open
filtration system, or by using one of the commercially available closed systems, where
the sample is never exposed to the test environment, thus minimising the opportunities
for contamination and false positive results. A widely used example of a closed
membrane filtration system is the Millipore Steritest filtration unit, which includes
connection devices and tubing so that samples can be withdrawn aseptically from
ampoules, collapsible bags and other containers without being exposed to the external
environment. The sample is then pumped through two individual filter canisters each
containing a 47 mm, 0.45 m membrane filter. Once filtration is complete each canister is
filled with 100 ml of medium (FTM and SCDM) and incubated. A range of filters is
available to suit products with different characteristics and the canisters are colour coded.
The system can also be used to sterility test medical devices.
Whichever method is used, it must be properly validated for the product being
tested to ensure that it does not increase the probability of recording a false negative
result. A number of different test microorganisms, including Staphylococcus aureus,
Bacillus subtilis, Pseudomonas aeruginosa, Clostridium sporogenes and Candia albicans,
should be used to demonstrate that the method is able to detect the required range of
organisms.
Cytometry does not rely on microbial growth to detect contamination, but instead
uses cell labelling techniques to detect viable microorganisms. This approach has the
potential to detect a wide range of organisms, including yeasts and moulds, within
minutes. Commercial systems utilise combined fluorescent cell labelling and flow
cytometry or solid phase cytometry to detect viable microbial cells. Typically, the cells
are labelled using a fluorescent dye or a non-fluorescent substrate, which is converted to
a fluorochrome in viable cells. Detection of the labelled cells occurs by laser scanning in
either a flow cell (flow cytometry), or on a solid phase platform such as a membrane
filter (solid phase cytometry). AES Chemunex has developed solid phase cytometry
detection systems. The companys ScanRDI (also known as ChemScan RDI) system is
capable of detecting 1 CFU per sample and has been evaluated as a possible RMM for
sterility testing. The technology has been developed for the Stereal-T sterility testing
system.
4.
Clean rooms and clean air devices should be routinely monitored in operation and
the monitoring locations based on a formal risk analysis study and the results obtained
during the classification of rooms and/or clean air devices. For Grade A zones, particle
monitoring should be undertaken for the full duration of critical processing, including
equipment assembly, except where justified by contaminants in the process that would
damage the particle counter or present a hazard, e.g. live organisms and radiological
hazards. In such cases monitoring during routine equipment set up operations should be
undertaken prior to exposure to the risk. Monitoring during simulated operations should
also be performed. The Grade A zone should be monitored at such a frequency and with
suitable sample size that all interventions, transient events and any system deterioration
would be captured and alarms triggered if alert limits are exceeded. It is accepted that it
may not always be possible to demonstrate low levels of 5.0 m particles at the point of
fill when filling is in progress, due to the generation of particles or droplets from the
product itself. (c) It is recommended that a similar system be used for Grade B zones
although the sample frequency may be decreased. The importance of the particle
monitoring system should be determined by the effectiveness of the segregation between
the adjacent Grade A and B zones. The Grade B zone should be monitored at such a
frequency and with suitable sample size that changes in levels of contamination and any
system deterioration would be captured and alarms triggered if alert limits are exceeded.
(d) Airborne particle monitoring systems may consist of independent particle counters; a
network of sequentially accessed sampling points connected by manifold to a single
particle Working document QAS/09.295 Rev.1 page 8 counter; or multiple small particle
counters located near to monitoring points and networked to a data acquisition system.
Combinations of systems can also be used. The system selected should be appropriate for
the particle size considered. Where remote sampling systems are used, the length of
tubing and the radii of any bends in the tubing should be considered in the context of
particle losses in the tubing. The selection of the monitoring system should take account
of any risk presented by the materials used in the manufacturing operation, for example
those involving live organisms or radiopharmaceuticals. The sample sizes taken for
monitoring purposes using automated systems will usually be a function of the sampling
rate of the system used. It is not necessary for the sample volume to be the same as that
used for formal classification of clean rooms and clean air devices. The airborne particle
conditions given in Table 2 for the at rest state should be achieved in the absence of the
operating personnel after a short clean-up or recovery period of about 1520 minutes
(guidance value), after completion of the operations. The particulate conditions given in
Table 2 for grade A in operation should be maintained in the zone immediately
surrounding the product whenever the product or open container is exposed to the
environment. In order to demonstrate control of the cleanliness of the various clean areas
during operation, they should be monitored for airborne particles and microbiological
contamination. In addition to at rest and in operation classification, airborne particles
should be monitored periodically in operation at critical locations.
The sampling plan need not be the same as that used for classification. Locations
and sample sizes should be determined based on assessment of the process and
contamination risk. The monitoring of Grade C and D areas in operation should be
performed in accordance with the principles of quality risk management. The
requirements and alert/action limits will depend on the nature of the operations carried
out, but the recommended clean up period should be attained. Other characteristics
such as temperature and relative humidity depend on the product and nature of the
operations carried out. These parameters should not interfere with the defined cleanliness
standard.