Sie sind auf Seite 1von 16

Med. J. Cairo Univ., Vol. 81, No. 2, September: 81-94, 2013 www.medicaliournalofcairouniversitv.

net

Knowledge and Performance of Critical Care Nurses Toward


Nebulizer Therapy in the Intensive Care Unit at Assiut University
Hospital
FARES A.A. KHODISH, B.Sc.N*; MONA A. MOHAMMED, D.N.Sc.*;
MERVAT ANWAR ABD EL-AZIZ, D.N.Sc.* and HANY AHMED IBRAHEM, M.D.**
The Departments of Critical Nursing* and Anesthesiology**, Faculties of Nursing* and Medicine**, Assiut University

Abstract Correspondence to: Dr. Mona A. Mohammed, The


Background: A nebulizer is an important device used to Department of Critical Nursing, Faculty of Nursing, Assiut
inhalant the drugs to treat some pulmonary diseases. The University delivered pi. A nebulizer is an electrically
examination of nurse's knowledge and performance powered machine that turns liquid medication into a mist
regarding the nebulizer therapy has not been conducted so that it can be breathed directly into the lungs through a
before at Assiut University Hospital. face mask or mouthpiece [21. It is generally believed that
nebulizers are superior to MDIs during mechanical
Aim of Study: Asses critical care nurse's knowledge and ventilation. The delivery of bronchodilators with MDI in
performance related to nebulizer therapy. mechanically ventilated patients has received considerable
Methods: Descriptive design was adapted for this interest in recent years. This is because the use of metered
study. The sample of this study consisted of 80 nurses dose inhalation (MDI) has several advantages over the
working in three intensive care units (38 nurses in nebulizer, such as reduced cost, ease of administration,
traumatic ICU, 20 nurses in postoperative ICU and 22 in less personnel time, reliability of dosing and a lower risk
general ICU). Two tools were used to collect the required of contamination. The use of nebulizers under certain
data; nurses' knowledge assessment questionnaire and circumstances may lead to patient/ventilator
skills observation checklist related to nebulizer therapy. dyssynchrony. Finally, nebulizers may damage the
expiratory transducer of some ventilators, rendering the
Results: The findings revealed that (56.2%) of nurses
have knowledge about nebulizer therapy at unsatisfactory expiratory volume measurement unreliable [31.
level (MeanSD=27.4412.09) also the findings revealed Bronchodilators are frequently used in ICU patients
that (88.8%) of nurses had performance about nebulizer receiving invasive mechanical ventilation (IMV) and
therapy at unsatisfactory level (MeanSD=29.2515.97). commonly delivered through the inhalation route [4].
There was a positive strong relationship between Medication that is inhaled in the form of small droplets is
knowledge and performance regarding nebulizer therapy absorbed immediately into the mucosa and blood stream
(r=0.277, p=0.013). and is available to the body within minutes. This method
of medication delivery is one of the fastest, noninvasive
Conclusions: Nurse's knowledge and performance methods of medications designed to ease respiratory
about nebulizer therapy were unsatisfactory so that the distress symptoms such as those seen with asthma [51.
nurses should be trained enough to give effective and safe Some studies reflect that, there are advantages of
care to such patients.
nebulizertreatment including: Patient coordination
Key Words: Nursing knowledge Nursing performance required, effective with tidal breathing, high dose possible,
Nebulizer therapy. dose medication possible, can be used with supplemental
oxygen, systemic side effects are less frequent and severe
Introduction with inhalation compared to systemic delivery (injection,
AEROSOLIZED medications can greatly benefit oral); e.g., less muscle tremor, tachycardia, Can deliver
critically ill patients with respiratory disorders. However, combination therapies if compatible and Ultrasonic
for an aerosol to be clinically useful, it must be delivered nebulizer is quiet, and has faster delivery are smaller and
more portable
effectively to the airways as well as produce a desired
therapeutic effect once
[6] . Aerosolized medication if not performed with [7] . Aerosolized medication must be performed
appropriate technique, will lead to several according to right standards and codes in order
consequences and perils such as mucosa to reduce its side effects [8]. Diggory & Vallone
irritation, bron- chospasm, dyspnea, airway [9],
burns (when heating elements are used), demonstrated in a study that performing
headaches, coughing, tachycardia, palpitations, aerosolized
nausea and precipitation of bronchoconstriction medication by well-educated nurses and after
most common in asthmatic and COPD patients checking patient's need has better effect and
and may result in hypoxemia fewer side effects than performing it routinely
[10]. Also revealed in a study that performing

1
2 Knowledge & Performance of Critical Care Nurses

aerosolisolized medication by themselves should only be carried out following careful


according to the codified protocol will minimize its assessment of the need for suctioning and should
side effects. Despite the existence of several only be undertaken by practitioners who are
studies, manual on the basis of evidences about competent. Patient indications for suctioning may
aerosolized medication by nebulizer is not include: Visible or audible secretions that patient
available in most units. There for performing are unable to clear, patient distress, clammy skin or
aerosolized medication by nebulizer with the sweating, increased heart rate/respiratory rate
observance of such experimental evidences by reduced oxygen saturation [1 6] . The role of the
nurses is on question 1111. Day et al., [12] in two nurse toward the patient with nebulizer therapy for
separate studies in intensive and acute care units, promoting respiratory care based. The nurse
realized nurses lacking knowledge which was instructs the patient to breathe through the mouth,
reflected in their performance Nurses are one of taking slow, deep breaths, and then to hold the
the major components of manpower in the field of breath for a few seconds at the end of inspiration
healthcare. The skills and knowledge of nursing for increasing the intrapleural pressure and reopen
personnel may be directed towards health promo- collapsed alveoli, therapy increasing function
tion, crisis intervention, maintenance, residual capacity, the nurse encourage the patient
rehabilitation/ restoration or palliation in care of to cough and to monitor the effectiveness of the
critically ill patient [13] . Nursing rolebefore using therapy [17]. So, the significance of the study from
Nebulizer; assess the patients respiratory status. an organizational perspective, the large and
Note if the patient is using accessory muscles for increasing number of noninvasive procedures at
respiration or if there is flaring of the nares. intensive care unit may result in capacity problems
Auscultate the clients chest for wheezes and such as respiratory infection. A possible solution
crackles. Respiratory distress is the primary reason for these problems might be a reduction in length
to administer nebulized treatment, the nurse must hospital stay. The number of nebulizer therapy
Place the patient in an upright position (40 to 90 performed at Assiut University ICU units in 2010-
degrees), which allows deep ventilation and 2011 more than 450 intervention procedures
maximal diaphragmatic movement Put the Observation of nebulizer therapy at ICU showed
mouthpiece in the mouth, between teeth, and close that most patient who performed nebulizer therapy.
the lips (if using a mask, cover mouth and nose Therefore the aim of this study is to investigate
with the mask). Assess the breath sound, pulse rate, Knowledge and practices levels of nurses in ICU at
respiratory status, oxygen saturation; assess the Assiut University Hospital. This study could be
heart rate during the treatment. In pregnant patient, beneficial in many ways; first, it will provide data-
the fetal heart rate should also be assessed, instruct base that can be utilized by health team members
the patients to take slow deep breaths for 10 to raise nurses awareness, and initiate more active
seconds through the mouth and hold at end nurses roles at intensive care unit. Second, Health
inspiration and wash face if using a mask to professionals can utilize such information in the
prevent rash [1 4] . The nurse must follow the care plan for such group of nurses in the future.
following steps when preparing a medication by a Third, proper nursing assessment and monitoring
nebulizer machine; if you have the multiple dose can improve patients outcome.
vial be; Squeeze the dropper to fill it to the line Subjects and Methods
that matches your prescribed dose, squeeze the Research design: Descriptive design was adapt-
medicine into the medicine cup, most medicines ed for this study.
that comes in multiple dose vials needs to be
diluted with normal saline [5] . The nurse instructs Studyaim: Investigate critical care nurses
the patient to do breathing exercise for two or three knowledge and performance related to nebulizer
seconds to give the medication time to fully enter therapy.
the airways of The lungs. After this first breath Research questions: What are nurses
continue this cycle of exhalation and inhalation knowledge and performance of toward nebulizer
until the dose of medication is finished. Typically, therapy in the intensive care unit at Assiut
treatment will take roughly five to 10 minutes [15]. University Hospital.
During the nebulizing; therapy it may be necessary
for the nurse to suction the patients tracheostomy Material:
to enable removal of secretions. This should only Setting:
be carried out if necessary and if the patient is This study was conducted in three intensive
unable to cough the secretions out of the tube care units (general intensive care units, postoper-
themselves. It should be remembered that suction- ative intensive care units & traumatic intensive
ing is an invasive procedure, which may be care units) at Assiut University Hospital.
uncomfortable and frightening for the patient. It
Fares A.A. Khodish, et al. 3

Subjects: nebulizer therapy. Second section was concerned


A sample of convenience included all the avail- with the preparation of equipment and patient.
able nursing staff working at the general intensive Third section was for the application of the
care units (22 nurses), postoperative intensive care nebulizer therapy including practices such as,
units (20 nurses) & traumatic intensive care units breathexercises, vital- singes, suction, mechanical
I.C.0 (38 nurses) at Assiut University Hospital ventilation, pulse oximetry, oxygen therapy and
(Total 80 nurses). arterial blood gas. Section four involves post care
practices such as; washing hands and regular care
Tools: while the nebulizer therapy was maintained.
Two tools were used to collect the required Finally, section five was; documentation of the
data; nurses knowledge assessment questionnaire type of nebulizer, medications, time, indications
and skills observation checklist related to nebulizer and unexpected outcomes for nebulizer therapy.
therapy. Satisfactory levelof performance equal or more
Tool (1): Nurses knowledge assessment Question- than 60% of the maximum score. Unsatisfactory
naire: level of performance less than 60% of the
Arabic interview questionnaire sheet was con- maximum score.
structed after reviewing the relevant literature to Methods:
assess nurses knowledge about a nebulizer ma-
chine. It was developed and translated into Arabic
language. Thequestionnaire was included three
parts. Part (1): Is for nurses characteristics such
asage, marital status, years of experience, depart-
ment, qualifications and their attendance of related
course training about a nebulizer machine. Part (2):
Thispartassesses nurses knowledge related to
nebulizer machine in general such as type, indica-
tions for applications, and complications of nebu-
lizer machine. Part (3): Assess nurses knowledge
concerning nebulizer therapy related practices
including knowledge regarding assessment, prep-
aration, application, post nursing care, and docu-
mentation of nebulizer therapy.
The total score for this questioner tool was 24
points, the correct response answer was scored as
"1" and incorrect response or do not know as "o".
The Satisfactory level of knowledge equal or more
than 60% of the maximum score. Unsatisfactory
level of knowledge less than 60% of the maximum
Tool (2): Skills observation checklist tool:
This tool was adapted from Potter et al., it was
used to assess nursing performance while applying
and providing maintenance care of nebulizer ther-
apy. The check list items included 71 items. It
consists of three parts. Part (1): Includes type of
the ICU and the shift (morning, evening or night)
when the observation checklist was taken. Part (2):
Involves patients characteristics, such as; age,
medical diagnosis and past history. Part (3): Was
used to observe nurses practices while applying
and maintaining the nebulizer therapy. It contains
five main section covering the main steps of neb-
ulizer use and care; assessment, preparation, post-
care, maintenance and documentation. First section
was for assessment which involves items to be
assessed before the application of nebulizer
therapy, such as indication of nebulizertherapy,
physician s order and medications which used by
4 Knowledge & Performance of Critical Care Nurses

Permission was obtained from the hospital observation technique. At initial interview the researcher
administrative authority to collect the necessary explain the nature & purpose of the study, prior to answering
data. Validation of the study tools was assessed by the questions to gain their oral consent & cooperation, and fill
presenting them to five experts from the critical out the questionnaire sheet by nurses were on duty during
care nursing field. A pilot study was carried out in morning,afternoon and night shifts. Nurses were observed two
march-July 2012 to test the tool. A pilot study was hours for each nurse, while they were performance nursing
conducted for purpose of testing clarity, complete- intervention for their patient during this period. The whole
ness, and validity, practicability of the study tools period for implementation of study was morning shift.
of (5) nurses to determine the time involvement. Statistical analysis:
Also it was done to show the possibility and effec- Data entry was done using compatible personal computer
tiveness of using the observation checklist. Neces- by the investigator. The statistical analysis was done using
sary modification was done. Testing the questions SPSS-16 statistical software package. The content of each tool
ware done to see if they are relevant and if they was analyzed, categorized and then coded by the investigator.
elicit the type of information, which are thought
Data were presented using descriptive statistics in the form of
according to the answers and comments made by
frequencies and percentages for qualitative variables,
nurse. Some details not required were omitted,
Quantitative continuous data were compared using student t-
either to advance or reduce the tool lengthy test.
Also to show the possibility and the effectiveness of test in case of comparisons between three groups. Pearson
using the observation checklist regarding the policy of hospital correlation analysis was used for assessment of the inter-
performance relationships among quantitative variables. Statistical signifi-
cance was considered at p-value<0.05.
The observation checklist was carried out while the nurses
were working during their shifts in these units. Each Results
observation lasted for 20 minutes by using the direct

Table (1): Presentation of study group at three units in intensive care units.
Postoperative I.C.0 (n=20) General I.C.0 (n=22)
DF=2
Characteristics I.C.0 trauma (n=38)
No. % No. % No. % X2 p-value
Age in years:

<25 years 10 26.3 6 30.0 9 40.9 3.187


25-30 years 21 55.2 9 45.0 7 31.8
>30 years 7 18.4 5 25.0 6 27.2 0.527

Qualification:
Nursing Bachelor 10 26.3 4 20.0 2 9.0
Nursing Technical Institute 7 18.42 5 25.0 6 27.2 2.825 0.588
Nursing Diploma 21 55.2 11 55.0 14 63.6

Years of experience:
<5 years 4 10.5 7 35.0 4 18.1
5-10 years 25 65.7 5 25.5 7 31.8 13.219 0.010*
10 years and more 9 23.6 8 40.0 11 50.0

Training course:
Attended 0 00.0 0 00.0 0 00.0
Not attended. 38 10.0 20 10.0 22 10.0
(*) Statistical significant difference (p<0.05).

Table (1) showed that the total number of nurses are (80) (65.7%) were between 5-10 years in
nurses from different of intensive care units, (38) nurses from trauma ICU.
trauma intensive care unit, (20) from post operative & (22) L,
from general intensive care unit. The majority of nurses Unsatisfactory
(55.2%) were between 25-30 years old in traumatic ICU, The 56.2%
majority of the nurses were nursing diploma (63.6%) in
general ICU, and all nurses had no previous training IN Fig. (1):
courseabout nebulizer (100%). Illustrate nurses' knowledge
about nebulizer.
There was statistical significant difference related to years
of experience (p-value=0.010) and the majority of nurses
Fares A.A. Khodish, et al. 5

This figure shows that (56.2%) of nurses at unsatisfactory


level of knowledge.
Table (2): Description of nurses' knowledge related to nebulizer therapy at three units.

Units
Postoperati p-
I.C.0 General value
Nurses' knowledge v
Unsatisf. Satisfact. Unsatisf. Satisfact. Unsatisf. Satisfact.
No. % No % No. % No. % No. % N %
. o.
Define the nebulizer 20 52.6 18 47.4 10 50.0 10 50.0 17 77.3 5 22.7 0.115
Indications of nebulizer 3 7.9 35 92.1 1 5.0 19 95.0 2 9.1 20 90.9 0.874
Assessment:
Sings of hypoxemia 12 31.6 26 68.4 8 40.0 12 60.0 16 72.7 6 27.3 0.007*
The lung sound. 16 42.1 22 57.9 9 45.0 11 55.0 16 72.7 6 27.3 0.059
Chest secretions 34 89.4 4 10.5 18 90.0 2 10.0 21 95.4 1 4.5 0.542
The arterial blood gases 12 31.5 28 73.6 5 25.0 15 75.0 10 45.4 12 54.5 0.214
Pulse oximetry 7 18.4 31 81.6 5 25.0 15 75.0 12 54.5 10 45.5 0.011*
Pulmonary function test 7 18.4 31 81.6 3 15.0 17 85.0 10 45.5 12 54.5 0.033*
Prepare:
The equipment 12 31.5 26 68.4 9 45.0 11 55.0 17 77.2 5 22.7 0.003*
The medication 4 10.5 34 89.5 1 5.0 19 95.0 4 18.2 18 81.8 0.394
Checkthe connections 14 36.8 24 63.1 14 70.0 6 30.0 13 59.0 9 40.9 0.039*
Explain breathing exercises 22 57.9 16 42.1 10 50.0 10 50.0 17 77.3 5 22.7 0.163
Complications of nebulizer 28 73.6 10 26.3 16 80.0 4 20.0 19 86.3 3 13.6 0.048*
(*) Statistical significant difference (p<0.05).

Table (2) shows that there were significant difference in 1. 003) and the majority of nurses in general ICU were at
nurses' knowledge about Sings of hypoxemia patient (p- Unsatisfactory level (77.2%).
value=0.007), and the majority of nurses in General I.C.0
There was significant difference in nurses' knowledge
were at unsatisfactory level (72.7%). Also, there was
about Check all connections of nebulizer (p-value=0.039) and
significant difference in nurses' knowledge about pulse
the majority of nurses in postoperative I.C.0 were at
oxime- tery (p-value=0.011) and the majority of nurses in
Unsatisfactory level (70%).
I.C.0 trauma were at satisfactory level (81.6%).

Shows there were significant difference in nurses' There was significant difference in nurses' knowledge
knowledge aboutpulmonary function test (p-value=0.033) and about complications of nebulizer (p- value=0.048) and the
the majority of nurses in postoperative I.C.0 were at majority of nurses in general
satisfactory level (85.0%). There was significant difference in 2. C.0 were at unsatisfactory level (86.3%).
nurses' knowledge about prepare theequipment (p-value=

Table (3):Difference between nurses' knowledge related to nebulizer therapy according to qualification.

Qualification
Nursing Bachelor (n=l6) Technical Institute (n=18) Nursing Diploma (n=46)
Nurses' knowledge p-value
Unsatis Satis Unsatis- Satisfactory
factory factory Unsatis- Satisfactory factory factory
No. % No. % No. % No. % No. % No. %
Define of nebulizer 13 81.3 36 78.3 10 21.7 4 8.7 42 0.000*
Indications nebulizer 3 18.8 2 12.5 14 87.5 12 66.7 6 33.3 4 22.2 14 77.8 91.3 0.194
Assessment:
Signs of hypoxemia 7 43.8 9 56.3 4 22.2 14 77.8 11. 54.3
25 10 45.7
21 0.067
The lung sound. 13 81.2 3 18.7 0. 16 88.8 2 1 46 0.0 0 0.0 0.026*
0 16 88.9 2 11.1 44 95.7 2 0 0.317
Thechest secretions. 16 100.0 0 30.432 30 4.3
Arterial blood gases 5 31.3 11 68.8 950.1 22 9 50.0 14 0.317
Pulse oximetry 4 25.0 12 75.0 4 .2 14 77.8 34.8
16 28. 33 69. 0.546
Pulmonary function test 4 25.0 12 75.0 3 16. 15 83.3 13 3 6 0.629
40 65.2
1 7 17 94.4 613.1 77. 5 0.685
Prepare medication 2 12.5 14 87.5 71.7
Cleans the equipments 12 31.5 26 9 5.6 11 55.0 17 0.003*
68.4 2 87.0
Explain: 45.0 42 22.7
Breathing exercises. 3 81.8 13 2 14 12 4 8.7 17
18.2 611.1 33 29
Procedures of neubilizer 5 31.2 11 68.7 88.9 63.0 0.548
Complications of nebulizer 12 75.0 3 .3 91.3
25.0 66.6 37.0 0.024*.
14 77.7 4 22.2 42 91.3 4 0.872
Fares A.A. Khodish, et al. 7

Qualification
Nursing Technical
Nursing Bachelor Institute (n=18) Nursing Diploma
Nurses' knowledge (n=16) (n=46) p-value
No. No. No.

Hand washing:
Satisfactory 4 25.0 2 11.1 6 13.0
Unsatisfactory 12 75.0 16 89.0 40 87.0 0.019*

Flow rate of 02:


Satisfactory 13 81.3 15 83.3 40 87.0
Unsatisfactory 3 3 16.7 13.0 0.838
18.8 6
Arterial blood gases:
Satisfactory 13 81.3 13 72.2 22 47.8
Unsatisfactory 3 18.8 5 27.7 24 52.2 0.007*

Reassess lung sounds, pulse:


Satisfactory 3 18.8 7 38.9 5 10.9
Unsatisfactory 13 81.3 41 89.1 0.036*
11 61.1
Clean of equipment:
Satisfactory 7 43.7 12 66.6 8 17.3
Unsatisfactory 9 56.2 6 33.3 38 82.6 0.001**

Table (3) shows there were significant difference in nurses' knowledge aboutdefine of nebulizer (p- value=0.000) and the
majority of nurses in Nursing Bachelorwere at satisfactory level (81.3%), there were significant difference in nurses' knowledge
aboutdefine of lung sound (p-value=0.026) and the majority of nurses in Nursing Bachelorwere at unsatisfactory level (81.2%),
there were significant

Table (4): Differences nurses' knowledge and nurses' characteristics according to department, education, and experience.
Level of Knowledge Nurses' l
characteristics Unsatisfactory Satisfactory P-value
No. % No. %
Department:
I.C.0 trauma 20 52.6 18 47.4 47.4
I.C.0 postoperative 10 50.0 10 50.0 50.0
General I.C.0 15 68.2 7 31.8 31.8
Education:
Nursing Bachelor 4 25.0 12 75.0 75.0
Nursing Diploma 32 69.6 14 30.4 30.4
Nursing Technical 9 50.0 9 50.0 50.0
Institute
Experience:
12 80.0 3 20.0 20.0
<5 years 5-10 years 22 59.5 15 40.5 40.5
10 years 11 39.3 17 60.7 60.7
(*) Statistical significant difference (p<0.05).

Table (4) shows that there was significant differences regarding to


education (p-value=0.07) and experiences (p-value=0.32).
difference in nurses' knowledge aboutdefine of Cleans the equipment's (p-
value=0.003) and the majority of nurses in Nursing Bachelorwere at satisfactory level (68.4%), there were significant difference in
(*) Statistical significant difference (p<0.05).

Table (5): Relationship between nurses' performance and nurses' qualifications related to nebulizer therapy.
nurses' knowledge aboutexplained procedure of nebulizer (p-value=0.024) and the majority of nurses in Nursing Bachelorwere at
satisfactory level (68.7%).

100 90 80 70 60 50 40 30 20 10 0
ICU Trauma Post operation General ICU
ICU
Unsatisfactory Satisfactory

Fig. (2): Nurses' performance according to department.


Fares A.A. Khodish, et al. 9

Table (5) shows significant difference in nurses' performance about washing hands (p-value =0.019) and the majority of nurses
innursing Technical Institute were at unsatisfactory level (89.0%).

There was significant difference in nurses' performance about Arterial blood gases (p-value=
1. 007) and the majority of nurses in nursing bachelor were at satisfactory level (81.3%).
There was significant difference in nurses' performance aboutreassessing breath sounds, pulse, oxygen saturation, respiratory
rate, (p-value=0.036) and the majority of nurses in nursing diploma were at unsatisfactory level (89.1%).

There was significant difference in nurses' performance aboutwashing all parts of the nebulizer (p-value=0.001) and the
majority of nurses in nursing diploma were at unsatisfactory level (82.6%).

Table (6): Nurses performance related to nebulizer of MV (mechanical ventilation) according to qualifications.

Qualification
N
Nurses' knowledge
&
ursmg B^for Nurnng T^hnical Nursmg Dmloma p-value
(n=16) Institute (n=18) (n=46) 1

No. No. No.


Prepare suction equipment:
Satisfactory 14 87.5 14 77.77 33 71.73
2 12.5 4 22.22 13 28.26 0.436
Unsatisfactory
Reassess vital signs: 11 68.7 11 61.11 22 47.82
Satisfactory 5 31.25 0.035*
7 38.88 24 52.17
Unsatisfactory
Cleans the equipment: 4
12 25.0 11 23.39
Satisfactory 75.0 633.33 35 76.08 0.950
Unsatisfactory 1266.66
Attach MV to endotracheal tube/tracheostomy tube: 14
2 87.5 17 00.0
Satisfactory Unsatisfactory 12.5 15
83.33 29 100.0
0.010*
Secure ventilator tabling: 12 3 16.66
Satisfactory 4 75.1 25. 00.0
0 100,0
Unsatisfactory Setup nebulizer 5 46
11 61.11 0.010*
12
which attached with MV: 4 7 38.88 69.56
Satisfactory 75.0 32 30.43
25.0 14
Unsatisfactory Document all finding 12 13 72.22 0.913
4 73.91
5 27.77
before & after therapy: 75.0 34 26.08
Satisfactory 25.0 12
Unsatisfactory 15 83.33
16.66 0.817
3

(*) Statistical significant difference (p<0.05).

(*) Statistical significant difference (p<0.05).


Table (6) shows there were significant difference in tracheostomy tube (p-value =0.10) and the majority of nurses
nurses' performance about Reassess vital signs (p- in Nursing Bachelor were at satisfactory level (87.5%).
value=0.035) and the majority of nurses in Nursing Technical
There was significant difference in nurses' performance
Institute were at satisfactory level (77.7%).
aboutsecure ventilator tabling (p- value=0.010) and the
There was significant difference in nurses' performance majority of nurses in Nursing Bachelor were at satisfactory
about Attach mechanical ventilator to endotracheal tube or level (75%).
Table (7): Nurses' performance related to nebulizer of mechanical ventilation according to units.
Units

Nurses' performance I.C.0 trauma General I.C.0 p-value


(n=38) Postoperative I.C.0 (n=20) (n=22)
No. % No. % No. %
Prepared suction equipment: Satisfactory 78.9
Unsatisfactory 30 21. 15 68.2
8 1 14 70.0 6 30.0 7 31.8 0.598
7.9
3 92. 1 4.5
Lung sound: Satisfactory Unsatisfactory 35 1 2 10.0 18 90.0 21 95.5 0.792
21. 4.5
Reassess vital signs: Satisfactory Unsatisfactory 8 1 3 15.0 17 85.0 1 95.5
30 78.9 21 0.226
Cleans the equipment at least once daily: 28.9
Satisfactory Unsatisfactory 11 71. 4 18.1
27 1 6 30.0 14 70.0 18 81.8 0.012*

Attach MV to endotracheal tube or tracheostomy


tube: Satisfactory Unsatisfactory 33 86.8 0 0.0
5 13.2 16 30.0 4 70.0 22 100.0 0.000*
Secure ventilator tabling: Satisfactory
Unsatisfactory 28 73.7 0 0.0
10 26.3 17 85.0 3 15.0 22 100.0 0.000*
Documentation:
Satisfactory 33 86.8 18 81.8
Unsatisfactory 5 13.2 17 85.0 3 15.0 4 18.2 0.871

Table (7) shows that there was significant difference in


nurses' performance about Cleans the equipment at least once =0.000) and the majority of the nurses in general I.C.0 were at
daily (p-value=0.012) and the nurses of the majority in general unsatisfactory level (0.100%).
I.C.0 were at unsatisfactory level (81.8%).
There was significant difference in nurses' performance about
There was significant difference in nurses' performance Secure ventilator tabling to reduce pull on tracheostomy or
about Attach mechanical ventilator to endotracheal tube or endotracheal tube (p-value = 0.000) and the majority of nurses in
tracheostomy tube (p-value general I.C.0 were at unsatisfactory level (100%).

Table (8): Nurses performance related to nebulizer of mechanical ventilation (MV) according to experiences.
Year of experiences
Nurses' performance <5 p-value
(n years 5-10 years = 15) (n=37) >10 years (n=28)
No. % No. % No . %

Prepared suction equipment readily available: 12 18 64.28


Satisfactory Unsatisfactory 3 80 28 75.67 20 9 24.32 10 35.71 0.314
Fares A.A. Khodish, et al. 11

Reassess vital signs, breathing and lung sound:


Satisfactory Unsatisfactory 4 26.6 10 27.0 73.3 27 73.0 39.28
11 11 60.71
17 0.048*
Cleans the equipment: Satisfactory Unsatisfactory
5 33.33 10 27.0 66.66 27 73.0 57.14
10 16 42.85
12 0.042*
Attach MV to endotracheal /tracheostomy tube:
Satisfactory Unsatisfactory 13 86.66 33 89.18 13.33 4 10.81 27 96.42
2 1 3.57
0.465

Secure MV tabling: Satisfactory Unsatisfactory


7 46.66 29 78.37 53.33 8 21.62 71.42
8 20 28.57
8 0.076
Setup nebulizer which attached with MV:
Satisfactory Unsatisfactory 9 60.1 30 81.08 22 78.57
6 40.1 7 18.91 6 13.04 0.253
Documentation:
Satisfactory 13 86.6 29 78.37 13.33 8 21.62 19 67.85
Unsatisfactory 0.353
2 9 3.57

Table (8) shows there were significant difference in Institute


nurses' performance about Reassess vital signs (p- Experience:
value=0.048) and the majority of the nurses at the age less <5 years 13 86.7 2 13.3
than 5 years old were at satisfactory level (73.3%). 5-10 years 32 86.5 5 13.5
10 years 26 92.9 2 7.1 0.695
There was significant difference in nurses' performance Nurses' performance
about Cleans the equipments at least once daily (p- Nurses' Unsati Satis p-value
value=0.042) and the majority of the nurses at the age characteristics s factor
between 5 to 10 years old (30.0%). factory y

Table (9) shows that significant differences in nursing


performance regarding to department (p- v alue=0 .027) and
education (p-value=0.001).
Discussion
Nebulisers are commonly used in the management of
acute exacerbation airways disease. A study was therefore
conducted to ascertain what nurses knew about this form of
therapy. Several areas of deficient knowledge were revealed,
including how to clean the equipment properly, what the flow
rate of gas should be and what type of gas to use. And nursing
care for nebuliser therapy is included [17] .
5.00 10.00 15.00 20.00 25.00 30.00 35.00
(*) Statistical significant difference (p<0.05).
(*) Statistical significant difference
Fig. (3): Correlation between nurses' (p<0.05).
knowledge and performance.
There was a significant positive correlation between The results of the present study showed that the majority
score of knowledge and score of performances (p- of nurses (55.2%) were between 25 -30 years old in traumatic
value=0.013). ICU. the majority of nurses were having nursing diploma
(63.6%) in general ICU, and all nurses had no previous
Table (9): Relationship between nurses' performance and training about nebulizer (100%).
nurses' characteristicsaccording to department,
education, and experience This study indicated that there was statistical a significant
No. % No. % difference related to years of experience and the majority of
nurses were between 510 years in trauma ICU. These findings
Department: are congruent with those of who [18] reported that; years of
I.C.0 trauma 30 78.9 8 21.1
Postoperative I.C.0 19 95.0 5.0 experience in the hospital significantly correlated to increased
1
General I.C.0 22 100.0 0 0.0 0.027* knowledge, attitudes and practices among the various
categories of staff but this did not translate into good clinical
Education:
62.5 practice in the ward and stated that the education and training
Nursing Bachelor 10 6 37.5
Nursing Diploma 45 97.8 1 2.2 are two components of staff development that occur after
Nursing Technical 16 88.9 2 11.1 0.001* employee indoctrination (which refers to planned, guided
adjustment of employees' organization and work
(*) Statistical significant difference (p<0.05).
environment). The staff knowledge level and capabilities are a
major factor in determining the number of staff required to It was found in the present study that there was
carry out unit goals. The better trained and more competent significant difference in nurses' knowledge about
the staff, the fewer staff required, which in turn saves the cleaning the equipments at least once daily These
organization money and rise reproductively. results are in agreement with the study of Canadian
Committee on Antibiotic Resistance, [23], they
The findings revealed that total knowledge was at an
found that reusable medical equipment must be
unsatisfactory level. The reasons might be due to education
thoroughly cleaned before disinfection or steriliza-
and training of the nurses, all nurses (100%) had not attended
tion, factors that affect the ability to effectively
training courses about nebulizer. The majority of the nurses in
clean medical equipment must be considered prior
the three units (intensive care unit of trauma, postoperative
to cleaning, instruments should be cleaned as soon
intensive care unit and general intensive care unit) had gained
as possible after use to prevent organic material
a diploma in nursing), Therefore, these factors may affect the
drying on instruments, place the instrument in a
unsatisfactory level of knowledge of nurses [19] . Stated that
pre-soak of water or an enzymatic solution, this
many medical personnel responsible for monitoring and
will prevent drying of organic material, organic
instructing patients in optimal inhaler use. Nurses seldom
material must be removed before disinfection or
receive formal training in the use of inhaling devices, and
sterilization procedures are initiated and the
newer inhaling devices designed to obviate problems of
process for cleaning should include written
technique are at present less likely to be used well by medical
protocols for disassembly, sorting and soaking,
personnel soon after their introduction [20] . In another
physical removal of organic material, rinsing,
study also discovered that knowledge is acquired primarily
drying physical inspection and wrapping. In this
through experience, usually observation of and
regard Labrague, et al., [24] found that age, gender,
working withmentors and continuing education.
length of clinical experience, and number of
Tacit knowledge may also be described as practical
trainings attended are not determinants of the
that is, derived from experience or practice [21].
Recommended that lack of knowledgeable nurses knowledge on the principles of sterile technique.
leads to increased risks to patients and may influ- This implies that nurses regardless of their age,
ence patient's outcomes. Nursing shortage also gender, length of clinical experience, and numbers
leads to increased workload as the nurse: Patient of trainings attended do not differ in knowledge on
ratio decreases and one nurse has to provide care sterile technique.
for more than one critically ill patient. Shift leaders The present study of revealed that nurses'
in both public and private sectors are working knowledge about checking all connections of neb-
under great pressure and a large number of them ulizer and the majority of nurses in postoperative
are also not trained, but simply have the
I. C.0 were at unsatisfactory level. McNeal, [25]
experience of working in an ICU for a number of
presented that There are lots of models of
years.
nebulizer out there, and each one is a little
The current study revealed that the majority of different from the others. This general guide to
the study sample (81.6%) werehaving satisfactory nebulizer parts will give a basic idea of what
in nurses' knowledge about pulse oximetery. These makes a nebulizer system function.
finding is in accordance with the results of the
The current study revealed that of nurses'
study conducted by [22] whomentioned that a
knowledge about pulmonary function test were at
pulse oximetry (Sp02) knowledge survey was
satisfactory level (85%). In this respect,
conducted with 551 experienced critical care
(Encyclopedia of Nursing) [26], reported that when
nurses at the 2002 American Association of
the results of pulmonary function testing are
Critical Care Two hundred and seven
accurate, the most frequent reason is adequate
questionnaires were completed (a response rate of
patient education and/or technician training. It is
74.5%). Mean pulse oximetry knowledge, with
recommended that personnel conducting
ICU nurses having significantly higher scores than
pulmonary function testing have one of the
anesthesia department nurses and those with more
following credentials: Certified respiratory therapy
than 10 years of experience having significantly
technician (CRTT); registered respiratory therapist
higher scores. Correct responses did not exceed
(RRT); certified pulmonary function technologist
50% for six questionnaire items, five of which
(CPFT); or registered pulmonary function
covered principles of pulse oximetry function. ICU
technologist (RPFT).
nurses had significantly more correct responses in
five items compared to nurses, and in two of them The results of the present study showed that the
compared to anesthesia department nurses. majority of studied sample were having unsat-
isfactory level of knowledge about complications
Fares A.A. Khodish, et al. 13

of the nebulizer. according to (Hoyle, 2012) [27] The present study showed that There was a
stated that nurses should have knowledge about significant positive correlation between score of
any complications that may occur during nebulizer knowledge and score of performances (p-value
therapy, the patient may become dizzy,complainse =0.013). This result agreed with Whyte, [32] who
of headache or may become disoriented during the have reported that selection of a group of nurses
course of treatment. If this happens, have patient participants was classified as either novice or
take off his or her mask or remove his mouthpiece experienced nurses on the basis of their years of
and take a break for a few minutes. Once his head experience. There were 12 experienced nurses and
is clear, have him continue the treatment and focus 10 novice nurses. Using an experimental research
on breathing more slowly than before. In some design based on the expert performance approach,
cases, slower breathing may not resolve dizziness a simulated task environment was developed for
or disorientation. Kulas, [28] who also found that the study using a patient simulator and a fully
the act of inhaling the medication through a nebu- equipped true-to-life intensive care unit suite.
lizer can cause dry mouth, sore throat, a bad taste Nurses were required to control the physiologic
in the mouth or hoarseness? Sometimes, nebulizers deterioration of the patient with respiratory com-
can cause thrush, which is a yeast infection in the promise in 4 scenarios and were also tested on
mouth. Symptoms of thrush include white spots in their knowledge of the constructs present in the
the mouth, bleeding and pain. scenarios.
The current study results denoted that the ma- The present study revealed significant differ-
jority of the studied nurse's were having a satisfac- ences in nursing knowledge regarding to
tory level of knowledge about the definition of the department (p-value=0.027) and education (p-
nebulizer. The present study also showed that there value=0.001) & there was no significant
was significant difference in nurses' knowledge differences regarding to experiences. Hassan &
about explaining procedures and the majority of Aboulazm, [32] stated that the highest mean scores
nurses in nursing Bachelor were at satisfactory of knowledge was found among nurses' who have
level of knowledge (68.7%). According to Ethan, bachelor of Science in nursing, and also found
[29] the patient should be in a comfortable there were improvement of nurses' knowledge
position throughout the procedure. Once the related to nebulizer.
patient puts the face mask on, he or she can start In the present study we found significant dif-
taking slow, deep breaths. He or she should ference in nurses' performance of arterial blood
continue to breathe slowly and deeply until there is gases (p-value=0.007). Nurses having Bachelor of
no medication left in the Nebulizer cup. This Sciences in nursing had significantly higher scores
generally takes anywhere from ten to thirty than those having technical institute of nursing and
minutes, depending on the prescribed quantity of those having nursing diploma. Coombs, [33] found
medication. that the ABG analysis provides useful monitoring,
The current study revealed the majority of especially for carbon dioxide. In most wards,
nurses in nursing diploma were unsatisfactory taking arterial samples has traditionally been a
knowledge regarding hearing the lung sound medical role, but some specialist nurses are now
(0.00%). This result agrees with Layman, et al., taking samples and so need to be able to interpret
[30] who emphasized that the assessment of the measurements. Nurses who are not taking samples
breath sound, pulse rate, respiratory status, oxygen may be able to initiate earlier intervention if they
saturation and assess the heart rate during the are able to interpret results. Understanding results
treatment. In pregnant patient, the fetal heart rate can help nurses to understand treatments and
should also be assessed, instruct the patient to take interventions, so making nursing more interesting.
slow deep breaths for 10 seconds through the The present study revealed a significant differ-
mouth and hold at end inspiration and wash face if ence in nurses' performance about reassessing
using a mask to prevent rash. breath sounds, pulse rate, oxygen saturation, res-
For performance regarding nebulizer, the find- piratory rate, and peak (p-value=0.036) and the
ings showed that total practice was at an unsatis- majority of nurses in nursing diploma were at
factory level. The reasons may be due to lack of unsatisfactory level (89.1%).
equipments, ignorance and the activities not in the A report by Creed & Spiers, [34] entailed that
duties of nurses Plaza, et al., [31] found that a professional nurse must perform by assessing
shortage of equipment was identified as obstacle to patient status from breath sounds, respiratory
the facilitation of the nurses' performance in status, pulse rate and other significant respiratory
(*) Statistical significant difference (p<0.05).
clinical areas.
functions needed. Compare, record significant Bachelor Were at satisfactory level (68.7%). in this regard,
changes and improvement. Refer if necessary and Quino, [41] reported that vital signs are necessary reassessing
perform other significant respiratory functions. patients status; breath sounds, respiratory status, pulse rate
The current study revealed a great lack of and other significant respiratory functions needed. Compare
performance as regarding of hands washing as all and record significant changes and improvement. Refer if
nurses were having an unsatisfactory performance necessary after inhalation therapy. Gamal, [42] found that
score level and had statistical significant (p- highest mean scores of knowledge and performance among
value=0.019). Hand hygiene is the most important younger nurses' those who have least experience and who
single infection control measure used in nursing have Bachelor of Science in nursing.
[35] .
Pratt, et al., [36] presented that hand contami- The current study revealed a great highof performance as
nation is one of the main contributing factors in regards toattach mechanical ventilator to endotracheal tube or
the current infection threat; contaminated hands tracheostomy tube (p-value =0.10).
are responsible for transmitting infection. British Thoracic Society, [43] found that nebulizers
Effective hand cleaning can significantly reduce should not be connected directly to a cuffed endotracheal or
infection rates in high risk area. Unfortunately, the tracheostomy tube. This could result in a pneumothorax,
result of the present study revealed that the which occurs when there is no expiratory route for exhaled
majority of nurses didn't carry out certain gases When nebulizing with endotracheal or tracheostomy
procedures in relation to infection control tubes check compatibility with the nebulizer machine.
precautions such as hands washing, wearing In thepresent there was significant difference in nurses'
gloves and cleaning equipments. Lam, et al., [37] performance about flow rate of 02 determined by the patient's
presented that hand hygiene has been singled out as condition (p-value=0.021) this result agreed with
the most important measures in preventing hospital acquired Nationallnstitute for Health & Clinical Excellence, [44] who
infection Ahmed, [38] has found lack of knowledge related reported that driving gas, either an air compressor or
universal precautions and infection control among nurses. piped/cylinder oxygen at 6-8Lt/min flow rate for most
efficient nebulization. And the patients with chronic obstruc-
The current study revealed a great lack of performance as tive pulmonary disease (COPD) should not routinely use
regards to washing all parts of the nebulizer. Scores of nurses' oxygen driven nebulizers due to the risk of carbon dioxide
performance (p-value =0.001) and the majority of nurses in retention.
nursing diploma were at unsatisfactory level (82.6%).
According to Ahmed, [39] who found that the majority of The current study showed that there was significant
nurses practices related to cleaning instruments mechanical differences nurses' knowledge regarding to education (p-
value=0.07) and experiences (p-value =0.32). The findings of
ventilation inadequately performed by nurses having diploma
the present study are in accordance with Abolwafa, [45] who
of Science. Robert, et al., [40] mentioned that ineffectual dis-
found that the majority of score of nurses' knowledge who
infection of inhalation equipment between therapy sessions of
have bachelor of Science in nursing had significant higher
different patients has been reported to produce serious
score than those who have technical institute and nursing
outbreaks of infection. The role of properly disinfected
diploma, scores of nurses' who have work experience ranged
equipment that becomes colonized during use was assessed.
from 5 to 10 years were higher than those who have work
In the Vancouver General Hospital gram-negative bacilli in
experience less than 5 years.
the nebulizer water were demonstrated in 10.5% of nebulizers
being used for periods of 24 to 72 hours. In at least 15% of Conclusion:
patients exposed to contaminate nebulizers the organism was Nurses' knowledge and performance about nebulizer
recovered from the respiratory tract 48 hours after removal of therapy were unsatisfactory with slightly higher level of
the equipment. Only one out of 85 patients exposed to knowledge and this could explain that the nurses acquired
contaminated nebulizers developed pneumonia, and in this experience during their nebulizer therapy.
instance a clear history of aspiration of vomitus was present.
The low incidence of pneumonia may be related to the high Recommendations:
percentage of nebulizers contaminated by organisms of low The nursingstaff in critical careunits are responsible for
virulence and might be much greater in hospitals where more giving care to the patients connected to nebulizer machine
highly pathogenic organisms are commonly found as should be trained enough to give effective and safe care to
nebulizer contaminants. such patients.

References
The present study revealed that there was significant 1- HOWARD L.: Current trends in the management of
difference in nurses' performance about reassessing vital critically 1ll patients: Drug delivery via nebulization,
signs (p-value=0.035) and the majority of nurses in Nursing mediscape, available at http://www.medscape.org/
viewar- ticle/413053_4, 2011.
Fares A.A. Khodish, et al. 15

2- PRISCILL, TASHKIN D. and ZIEVE D.: The science of 20- EVENS J.R. and DONNELLY G.W.: Amodel to describe
nebulized drug delivery available from http://kidshealth. the relationship between knowledge, skill, and judgment
org/parent/asthma center/words know/nebulizer.html, in nursing practice, Journal Nursing Forum, 1744-6198,
2011. 2006.
3- GEORGOPOULOS D., MAULOUDI E. and KONDILI 21- DANIEL R., GRENDELL R.N. and WILKINS F.R.:
0.: Bronchodilator delivery with metered-dose inhaler Nursing fundamentals, caring & clinical decision making,
during mechanical ventilation. Journal of Respiratory medication administration, (2nd edition), Australia. Brasil.
Critical Care, 4 (4): 227-234, 2002. Japan. Spain, Delmercengage, p. 903-04, 2010.
4- GUERIN C., et al.: Inhaled bronchodilator administration 22- GIULIANO and LIU L.M.: Knowledge of Pulse oxime-
during mechanical ventilation: Journal of Aerosol. Med- tryamong critical care nurses; Journal of Dimensions
icine Pulmonary Drug, pp. 85-95, 2008. Critical Care Nursing, Jan.-Feb., 25 (1): 44-9, 2006.
5- ALTMAN G.B., KERESTZES P. and WCISEL M.: Fun- 23- Canadian Committee on Antibiotic Resistance: Infection
damental & advanced. Nursing skills, administration prevention and control best practices for Long Term Care,
nebulizer medications. Skills 5-7, Austrlia. Brasil. Japan, home and community care including health care offices
Delmarcengage, p. 584-85, 2010. and ambulatory clinics. Available at on line at: http://
www.cpsa.ab.ca/collepe
6- MAN L.S. and GRACE: What are the advantages and programs/attachments_ipac/IPAC-
disadvantages of each aerosol delivery device? Is there
Best_Practices_general.pdf, 2007.
any benefit of one over another? Journal of Chest, 127:
335-371, 2008. 24- LABRAGUE L.J., DOLORES L. and LEDRORO J.:
Operating room nurses' knowledge and practice of sterile
7- PINNOCK C. and JONES R: Fundamentals of
technique. Journal of Nurse Care, 1: 113.Doi.:10. 4172/
anaesthesia, anaesthesiaequipments, (4th edition),
2167-1168.1000113, 2012.
London, Gmm., p. 904-05, 2011.
25- DAUTZENBERG B. and CARONE M.: European respi-
8- BRAND C.: Nebulizer therapy complications available at
ratory society guidelines on the use of nebulizers, Journal
https: //www.healthtap.com/#topics/nebulizer_therapy
of Anaesth., 109: 655P-668P, 2012.
-complications, 2012.
9- DIGGORY P., BAILEY R. and VALLONE A.: Efective- 26- Encyclopedia of Nursing: Pulmonary function test, avail-
able from http://www.enotes.com/pulmonary-function-
ness of inhaled bronchodilator delivery systems for pa-
test--172.341..3. 2006.
tients; Journal of Chest, 26: 412-586, Aging, 2007.
27- HOYLE G.M.: How should you instruct a patient to
10- KELLY C. and LYNES D.: Nebulizer therapy, best
breathe for a nebulizer treatment? available from http://
practice in the provision of nebulizer therapy. Journal of www.ehow.Com/way 5678698 should-patient-breathe-
Nursing Standard. Apr., p. 25 (31): 50-56, 2011. nebulizer-treatment .html, 2012.
11- BOE J., DDNNIS J.H., DRISCOLL B.R., BAUER T.T. 28- KULAS M.: Side effects of nebulizer use available from
and HESLOP K.: European respiratory society guidelines http : //www. live strong. com/article/23178-side effects-
on the use of nebulizers. European Respiratory Journal, 28 nebulizer-use, 2011.
(3): 278-263, 2011.
29- ETHAN P.: Nebulization Procedure, available at http://
12- DAY T., FARNELL S., HAYNES S., WIANWRINGHT www.reference.com/motif/health/nebulzer. 2012.
S. and WILSON-BARNETTS: Tracheal suctioning: An
exploration of nurses knowledge and competence in acute 30- LAYMAN M.E. and PROEHL J.A.: Nebulizer therapy,
and high dependency ward areas. S. Adv. Nurs., 39 (1): procedures steps, (4th edition), p. 171-72, 2009.
35- 45, 2002. 31- PLAZA V., GINER J. and GOMEZ J.: Knowledge and
13- SACHDEVA, 2010. skills regarding the use of the turbuhaler inhaler, Journal
14- LAYMAN M.E. and PROEHL J.A.: Nebulizer therapy, of Archivos De Bronconeumologia, 33 (3): 113-117, 2009.
procedures steps, (4th edition ), p. 171-72, 2009. 32- LEWIS S., HEITKEMPER M.M., DIRKSEN S.R. and
15- HOYLE G.M.: How should you instruct a patient to BUCHER L.: Medical surgical nursing, obstructive pul-
breathe for a nebulizer treatment? available from http:// monary diseases, (7 th edition), Mosby El Sliver, p. 64344,
www.ehow. Com/ way_5678698_should-patient-breathe- 2009.
nebulizer-treatment_.html, 2012. 33- WHYTE J.: The relationship between knowledge and
16- WOODROW P.: Intensive care nursing, a framework for clinical performance in Novice and experienced critical
practice, airway management, (3rd edition), London & care nurses,available at http://www.medscape.com
New York. Routledge, p. 55, 2011. /viewar- ticle/714090, 2009.
17- SMELTZER S. and BARE B.: Textbook of medical- 34- HASSAN H.E. and ABOULAZM S.F.: Infection control
surgical nursing, gas exchange and respiratory function, education. The New Egyptian Journal of Medicine, 36 (1):
Philadelphia. New York. London. Lippincott, p. 730-31, pp. 67-73, 2007.
2008. 35- COOMBS M.: Making sense of arterial blood gases,
18- MARQUIS L.B. and HAUSTON J.C.: Leader ship role & Journal of Nursing Times, 97 (27): 36-38, 2001.
practice foundations founctions in nursing, (6th ed), 36- CREED F. and SPIERS C.: Care of the acutely ill adult
Lippincott, Hong Kong, p. 371, 2009. -an essential guide for nurses, humhdifition; (6th edition),
19- HANANIA, et al.: Medical personnel's knowledge of and Oxford University Press, p: 459, 2011.
ability to use inhaling devices. Metered-dose inhalers, 37- COBEN B., SAIMAN L. and CIMIOTTI J.: Factors
spacing chambers, and breath-actuated dry powder associated with hand hygienepractices in two neonatal
inhalers Journal of Chest, 105 (1): 111-6, 2009. intensive care units. Pediatric Infection Control, 2008.
(*) Statistical significant difference (p<0.05).
38- PRATT RJ., PELLOWE C.M. and WILSON J.A.: nursinp-crib.com/demo-checklist/nebuliza.tion-therapy.
National evidence-based NHS hospital in England. 2011.
Journal of Hospital Infection 65.wwwe epicivu-Acuk, 43- GAMAL L.M.: Establishing standards for prevention of
2007. Nosocomial infection in the recovery rooms and surgical
39- LAM B.C., LEE J. and LAU Y.L.: Hand hygiene ward at El meniauniversity hospital. Doctorate thesis,
practices in neonatal intensive care unit: A multimodal Faculty of Nursing, Assiut University, 2006.
intervention and impact on nasocomial infection.
Departement of pediatric and adolescent medicine, queen 44- British Thoracic Society: British guideline on the man-
Mary Hospital, Hong kong, china, lamccb @ha. agement of asthma. Available at www.brit-thoracic.org.uk,
org.hk.114 (5): pp. 565571. 2; 6, 2007. 2008.
40- AHMED N.F.: Assessment of nurses' knowledge and 45- National Institute for Clinical Excellence: Management
performance about universal precautions during cesarean of chronic obstructive pulmonary disease in adults and
section, master degree, Faculty of Nursing, Assiut Uni- primary and secondary care. NICE. July, Clinical Guide-
versity, 2010. line, 101, 2010.
41- ROBERT: Infant nebulizer treatment, available at http:// 46- ABOLWAFA N.F.: In assessment of nurse's knowledge
www.livestrong.c om/article/269008_infant-nebulizer- and performances units at El-Menia City Hospitals.
treatments, 2011. Master degree, Faculty of Nursing, Assiut of Hospital,
2009.
42- QUINO F.: Nebulization therapy, available at http://

Das könnte Ihnen auch gefallen