Beruflich Dokumente
Kultur Dokumente
Calorimeter
Instruction Manual
Version 1.2x
Part Number 200-999-004
January 2012
1 Introduction
Illustrations .................................................................................. 16
Warranty...................................................................................... 18
LECO-Supplied Dell Computers ....................................................... 19
WEEE......................................................................................... 111
Interference to Other Devices ....................................................... 116
Declaration of Conformity ............................................................. 117
About this Manual ....................................................................... 118
Using the Software Help System ................................................... 119
Software License ......................................................................... 121
Equipment Packages .................................................................... 124
Optional Equipment ..................................................................... 126
Components and Accessories List .................................................. 127
Specifications ............................................................................. 129
Certificate of Traceability .............................................................. 130
Oxygen Combustion Vessel Safety Notice ....................................... 133
Hydrostatic and Proof Testing ....................................................... 135
Combustion Vessel Hydrostatic and Proof Testing Return Form ......... 136
2 Installation
Illustrations .................................................................................. 22
Installing Calorimeter .................................................................... 23
International Installation .............................................................. 214
Computer Installation .................................................................. 215
Data Transmit Installation ............................................................ 216
3 Option Installation
Illustrations .................................................................................. 32
Balance Installation ....................................................................... 33
Printer Installation......................................................................... 37
Smartline Installation .................................................................... 38
Introduction AC500 13
4 System Setup
Illustrations .................................................................................. 43
Controls and Indicators .................................................................. 44
Software Main Window ................................................................... 47
Menu Bar ..................................................................................... 48
Using Toolbars ............................................................................ 414
Customizing Sample Spreadsheet .................................................. 415
Importing Sample Data ................................................................ 419
Creating Sample Templates .......................................................... 421
Creating Transmission Formats ..................................................... 422
Creating Export Format ................................................................ 428
Determining User Permissions....................................................... 432
Selecting a Language ................................................................... 436
Configuring the System ................................................................ 437
Defining Calibration Standards ...................................................... 442
Sample Templates ....................................................................... 444
Using Sample Filters .................................................................... 447
Creating and Modifying User Defined Fields .................................... 451
Managing Databases .................................................................... 453
5 Operation
Illustrations .................................................................................. 53
Logging in a Sample ...................................................................... 55
Preparing Crucibles........................................................................ 57
Preparing Samples ........................................................................ 58
Combustion Vessel Preparation for Wire Fuse ................................... 59
Combustion Vessel Preparation for Thread Fuse .............................. 512
Combustion Bucket Preparation .................................................... 515
Creating or Modifying Methods ...................................................... 517
Analyzing a Sample ..................................................................... 521
Abort an Analysis ........................................................................ 523
Calibration ................................................................................. 524
Editing Samples in the Spreadsheet ............................................... 525
Printing Analysis Results .............................................................. 530
Exporting Data ............................................................................ 534
Importing Data ........................................................................... 542
Transmitting Data ....................................................................... 543
Transmitting Selected Samples ..................................................... 550
Data Backup ............................................................................... 551
Error Messages ........................................................................... 554
14 AC500 Introduction
6 Maintenance
Illustrations .................................................................................. 62
Periodic Maintenance Schedule ....................................................... 63
Combustion Vessel Maintenance ...................................................... 69
Fan Filter Removal and Cleaning ................................................... 610
Particle Filter Replacement ........................................................... 611
Water Return Filter Replacement ................................................... 612
Water System Draining ................................................................ 614
Water System Filling .................................................................... 615
Adjusting the Pressure Safety Release Valve ................................... 616
7 Theory of Operation
Illustrations .................................................................................. 73
Theory of Operation....................................................................... 75
Calculations and Corrections for Nitrogen, Sulfur, and Net Calorific Value76
Correction For The Energy Of Formation Of Nitric Acid ....................... 77
Correction For Energy Of Formation Of Sulfuric Acid .......................... 79
Calculation for Net Calorific Value .................................................. 712
Calculation to Other Bases ........................................................... 716
Fuse (Ignition) Wire Correction ....................................................... 79
Example Calculations ................................................................... 718
8 Diagnostics
Monitoring Ambient Chart............................................................... 83
Solenoids and Switches .................................................................. 85
Backup (Hardware) Calibration ....................................................... 86
Hardware Calibration ..................................................................... 89
Communications ......................................................................... 814
9 Service
Illustrations .................................................................................. 92
Installing LECO Software................................................................ 93
Configuring the Ethernet Adapter .................................................. 911
Configuring the Firewall Settings................................................... 915
Registering the Software.............................................................. 916
Troubleshooting Copy Protection Key............................................. 922
Combustion Vessel Charger O-ring Replacement ............................. 929
Troubleshooting .......................................................................... 930
10 Illustrations
11 Schematics
12 Index
Introduction AC500 15
Illustrations
16 AC500 Introduction
Figure 11-1 Cabinet Schematic ........................................................111
Figure 11-2 Igniter Box Schematic ...................................................112
Introduction AC500 17
Warranty
Notice
The warranty on LECO equipment remains valid only when genuine LECO
replacement parts are employed. Since LECO has no control over the quality or
purity of consumable products not manufactured by LECO, the specifications for
accuracy of results using LECO instruments are not guaranteed unless genuine LECO
consumables are employed in conjunction with LECO instruments. If purchaser defaults
in making payment for any parts or equipment, this warranty shall be void and shall not
apply to such parts and equipment. No late payment or cure of default in payment shall
extend the warranty period provided herein.
LECO Corporation is not responsible for damage to any associated instruments,
equipment, or apparatus nor will LECO be held liable for loss of profit or other special
damages resulting from abuse, neglect, or use in disregard of instructions. The Buyer,
their employees, agents, and successors in interest assume all risks and liabilities for the
operation, use, and/or misuse of the product(s) described herein and agree to
indemnify, hold harmless, and defend the seller from any and all claims and actions
arising from any cause whatsoever, including seller's negligence for personal injury
incurred in connection with the use of said product(s) and any and all damages
proximately resulting there from.
18 AC500 Introduction
LECO-Supplied Dell Computers
Dell Support
All LECO supplied Dell PC's include Dell's ProSupport Service, which is designed
to provide a base level of customer security. This service option can place a
Dell-certified technician at your location the following day, if necessary (after
phone-based troubleshooting). In many cases, remote troubleshooting can
minimize overall system downtime by getting the issue quickly fixed remotely
or helping Dell to dispatch the right part (if necessary).
The inclusion of Dell ProSupport with LECO supplied PCs ensures that any
computer related service issues are handled directly by the experts at Dell.
This eliminates the unnecessary step of working through the LECO service
professionals for a resolution from Dell, and provides PC support 24 hours a
day, seven days a week.
Dell ProSupport provides expert assistance when you need rapid problem
resolution to help maximize your PC system's uptime and productivity.
ProSupport provides the following benefits:
24 x 7 Access to advanced-level North American-based phone technicians
Live Chat tech support
DellConnect remote diagnosis and resolution
Next Business Day On-Site Service after Remote Diagnostics
For more information, call 1-800-WWW-DELL (1-800-999-3355) or access
Dell's website by copying and pasting the following link into your internet
browser: http://www.dell.com/content/topics/global.aspx/services
/prosupport/end_users?c=us&cs=555&l=en&s=biz
Please have your Service Tag and Express Service Code numbers available
when contacting Dell to expedite service. These can be found on the back or
side of the tower.
Introduction AC500 19
For Domestic Customers:
1. Access Dell's website by copying and pasting or entering the following
website URL into your internet browser:
http://support.dell.com/support/topics/global.aspx/support/change_order/
tag_transfer?c=us&l=en&s=gen&redirect=1
It is recommended to complete the transfer of ownership from
the computer from which the ownership will be transferred. If
this is not possible, follow the instructions provided on Dell's
website.
2. Complete the transfer by completing and submitting the form provided on
Dell's website.
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to Part 15 of the FCC Rules and Regulations. These
limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses and
can radiate radio frequency energy and, if not installed and used in accordance
with the instructions, may cause interference to radio communications.
However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more of
the following methods:
Reorient or relocate the radio or television receiving antenna.
Increase the separation between the equipment and the radio or television
receiver.
Connect the equipment to an outlet on a circuit different from that which
the radio or television receiver is currently connected.
Consult the radio or television dealer or an experienced radio/TV technician
for help.
The screen shots used throughout the manual are examples only and
may or may not be applicable to specific procedures.
Bookmarks
For versions prior to 8, after opening the pdf document, select the Bookmarks
tab located on the left side of the screen to access a tree that displays main
section headings. Select a heading to go immediately to the corresponding
section. If a + appears, select it to display more section headings.
Find
Select Edit on the Menu bar and then Find to highlight a search box where you
can enter a word or phrase. Each time you select Next, the software will
highlight the next occurrence of the word or phrase in the manual. Find is most
useful when the word or phrase is used infrequently in the manual.
Search
For versions prior to 8, select Edit and then select Search on the Menu bar to
display a side bar with a search box. Enter the desired word or phrase and each
occurrence of the word or phrase in context will display in the Results area.
For version 8, the Search box appears in a separate window rather than a
sidebar.
Search makes it easier to find words or phrases that occur several times in the
manual. Search also provides the ability to search words or phrases in several
pdf documents, if applicable. To search the instruction manual only, select the
Search this Pdf Document Only check box.
Links
In the manual, text or a page number highlighted in blue provides a direct link
to the associated topic. In the Index, select the page number to go directly to
the associated topic. For illustrations, a label and a page number are provided.
The label links to the caption of the illustration while the page number links to
the body of illustration.
In version 8, there are two types of arrows, page arrows and view
arrows . Page arrows move from one page to the next page and follow
the page numbering of the manual while view arrows move from one view to
the next and depend on the pages you have viewed. To display both types of
arrows, click Tools on the Menu bar and then click Customize Toolbars. From
the list, locate Page Navigation Toolbar and then if all of the arrows do not
appear, select the Previous Page check box, Next Page check box, Previous
View check box, and Next View check box to display both sets of arrows.
I. Definitions
"Licensed Software" shall mean the software supplied with this Software
License, which is software developed and distributed by LECO.
"Instruction Manual" shall mean a guide describing the use and operation of the
Licensed Software, together with any related supporting documentation.
"Licensee" shall mean and include any of the Licensee's divisions and
departments within Licensee's organization, but shall not include wholly or
partially owned subsidiaries, affiliates, successors or assigns, or independent
third parties.
a. the right to use a copy of the Licensed Software in executable form by one
user at a time;
b. the right to use and copy the Instruction Manual for Licensee's internal
operations; and
c. the right to make one backup copy of the Licensed Software.
In accepting the license granted by LECO, Licensee agrees that it shall not:
VIII. Miscellaneous
a) This Software License shall be binding upon and inure to the benefit of the
parties hereto and their respective successors and assigns, provided that
Licensee shall not have any right to assign its rights or obligations under this
Software License (whether by operation of law or otherwise) without the prior
written consent of LECO.
b) All notices hereunder shall be in writing and shall be sent by registered or
certified mail postage prepaid to LECO at the addresses first written above.
The date of postmark shall be deemed the date on which such notice is given.
c) No delay on the part of LECO in exercising any right, power or remedy
hereunder shall operate as a waiver thereof, nor shall any single or partial
exercise of any such right, power or remedy preclude any other or further
exercise thereof or the exercise of any other right, power or remedy.
d) No amendment, modification or waiver of, or any consent with respect to,
any provision of this Software License shall be effective unless the same shall
be in writing and signed and delivered by each of the parties hereto, and then
such amendment, modification, waiver or consent shall be effective only in the
specific instance and for the purpose for which given.
e) This Software License shall be construed in accordance with and governed
by the internal laws of the State of Michigan. Wherever possible, each
provision of this Software License shall be interpreted in such manner as to be
effective and valid under applicable law, but if any provision of this Software
License shall be prohibited by or invalid under such law, such provision shall be
ineffective to the extent of such prohibition or invalidity, without invalidating
the remainder of such provision or the remaining provisions of this Software
License.
f) Licensee agrees that it is responsible for obtaining proper authorization and
complying with any export, re-export, or import laws or regulations and that
may apply if Licensee exports, re-exports, or imports the Licensed Software,
technology or technical data licensed hereunder.
Items listed below are repeated throughout this manual and are subject to
revision. Please consult the packing slip received with the instrument. For a
complete listing of kits and packages, see Components and Accessories List,
page 127.
603-350-0op
1 051-063 SPEC SHEET AC500
1 259-052UPGR SOFTWARE KIT AC500 UPGRADE
1 502-266 WIRE FUSE 34AWG AC-350/500
1 608-948 ASSY BASIN WASH
1 614-459 CARTRIDGE BLK PRNTR HP700/950
1 617-237-110 KIT REPL RESERVOIR FIBERGLASS
1 619-995 READER BAR CODE CCD USB*
1 620-632 CARTRIDGE COLOR PRNTR SERIES C
1 621-192 KIT ADD ON MONITOR 19 FLT SCR AGP
1 621-245 KIT VESSEL THREAD IGNITER 500
1 621-246 KIT VESSEL WIRE IGNITER
1 621-322 ASSY PRESS PELLET SAMPLE AC
1 621-325 PURIFIER POWER 60 AMP 10KVA DOMESTIC*
1 621-434-110 KIT PRINTER DESKJET HP6940
1 621-435 CARTRIDGE COLOR PRNTR HP6940
1 621-436 CARTRIDGE BLACK PRNTR HP6940
1 751-300-140 KIT BALANCE L-250 4PL
1 259-052-CPK SOFTWARE KIT AC500 W/KEY
1 621-429 KIT ADD ON MONITOR 19 INCH PCI EXPRESS
1 502-815 SAMPLE MINERAL OIL 118/ML
1 774-208 ACID BENZOIC 1.00GR 10/BT
1 774-208-150 ACID BENZOIC 1.00GR 50/BT
1 621-445-110 KIT PRINTER INKJET WORKFORCE 30
1 621-446 CARTRIDGE COLOR PRINTER WORKFORCE 30
1 621-447 CARTRIDGE BLACK PRINTER WORKFORCE 30
1 502-680 STD RM COAL PROX PLUS LOW
1 502-681 STD RM COAL PROX PLUS MED
1 502-682 STD RM COAL PROX PLUS HIGH
1 621-376 PURIFIER POWER 50 AMP 7.5 KVA DOMESTIC
1 611-351-277 KIT REPL 606-429/614-840 CABLE AC500
1 611-351-283 KIT REPL 777-904C AND SWITCH
1 709-806-808 RECEPTACLE SNGL 15A 250V BLK
1 709-806-809 COVER RECEPTACLE SNGL SST
1 709-806-719 CORD POWER ASSY 10FT 15A/250V
Illustrations .................................................................................. 22
Installing Calorimeter .................................................................... 23
International Installation .............................................................. 214
Computer Installation .................................................................. 215
Data Transmit Installation ............................................................ 216
Installation AC500 21
Illustrations
22 AC500 Installation
Installing Calorimeter
The following procedure should be used for first time installation of the
calorimeter. This procedure assumes that the calorimeter has been
unpacked and positioned in its permanent location.
The Balance and Printer are not supplied with the calorimeter. They
must be purchased separately. Refer to Optional Equipment, page
125.
The connectors and hardware referred to in this procedure are
included in the component pack supplied with the calorimeter.
4. Remove the access door above the power switch and locate the
115/230 V~ voltage select switch. Refer to Figure 2-2, page 28.
5. Depending on line voltage available, position the voltage select
switch to either the 115 V~ or 230 V~ position.
A. Move the switch to the right for 115 V~ as viewed from the
rear.
B. Move the switch to the left for 230 V~ as viewed from the
rear.
6. Remove the fuse and check its current rating. Refer to Figure 2-2,
page 28.
A. For 115 V~ operation, install a 3.15 amp fuse.
B. For 230 V~ operation, install a 1.6 amp fuse.
Installation AC500 23
7. Remove the Left Shroud Assembly and locate the igniter fuse
switchbox. Refer to Figure 2-3, page 28.
8. Verify that the Fuse Selection Switch is in the correct position. For
the thread fuse position the switch towards Thread (cotton). For
the wire fuse position the switch towards Wire.
9. If a power cord with the proper plug is not available, use the
wiring diagram to install the proper plug. Refer to Figure 2-4,
page 29.
10. Connect the AC power cord, supplied with the calorimeter, to the
power connector on the rear of the calorimeter. DO NOT plug it
into facility power. For installation outside the U.S.A. Refer to
International Installation, page 214.
11. Install the Computer near the calorimeter. Refer to Computer
Installation, page 215.
12. Install the Balance near the computer and instrument. Refer to
Balance Installation, page 33.
13. Complete the oxygen connection as explained in the steps below.
A. Remove the cap plugs from the Oxygen In fitting on the rear
panel of the calorimeter. Refer to the Facility Requirements
diagram, page 27.
B. Using the required connector, connect one end to the
regulator on the oxygen supply and the other end to the
Oxygen In connector on the rear panel of the calorimeter.
C. The regulator is equipped with a pressure safety release
valve set to 500 psi (34.5 bar). Check this valve by adjusting
the regulated pressure to, or just above, 500 psi (34.47
bars). The safety valve should release. Reset the regulated
pressure to 450 psi (31.0 bar) for normal operation.
D. The pressure safety valve should be checked periodically for
proper operation. Turn the main tank valve off when the
calorimeter is not in use. Refer to Adjusting the Pressure
Safety Release Valve, page 616.
14. Complete the combustion vessel charger connection as explained
below:
A. Remove the cap plugs from the Oxygen Out fitting on the
rear panel of the calorimeter. Refer to the Facility
Requirements diagram, page 27.
B. Using the required connector, connect one end of the
connector to the combustion vessel charger assembly and
the other end to the Oxygen Out connector on the rear panel
of the calorimeter.
C. Connect the combustion vessel fill switch plug to the fill
switch jack on the rear of the calorimeter.
15. Complete the Pipet Installation as explained below:
A. Install the holder base to the block on the base of the
calorimeter. Refer to Figure 2-5, page 210.
B. Install the brace to the shroud directly above the holder
base.
24 AC500 Installation
C. Slide the rod, threaded end down, into the holder base.
Secure the rod by screwing it into the holder base.
D. Slide one of the pipet claws onto the rod until it is 10 inches
below the top of the rod.
E. Turn it until it is at a right angle with respect to the right
side of the calorimeter. The pipet claw should be positioned
away from the calorimeter. Tighten the setscrew to secure it.
F. Insert the pipet into the installed pipet claw positioning the
valve down. The valve knob should be positioned towards
the front of the calorimeter.
G. Hold the pipet and slide the other pipet claw onto the rod
positioning it on the top of the pipet. Tighten the set screw
to secure it.
H. Connect the interconnecting tubing and particle filter to the
pipet.
I. If necessary, adjust the claws up or down so that the tubing
just clears the combustion vessel bucket rim.
16. Complete the water return tank installation as explained below:
A. Thread the three standoff legs into the bottom of the water
return tank. Refer to Figure 2-5, page 210.
B. Place a water return filter over the filter holder (wetting the
filter first will aid the installation). Secure the water return
filter to the filter holder with the o-ring.
C. Place the filter assembly and filter screen into the bottom of
the water return tank. The o-ring is intended to make a seal
against the tank wall. Some interference will be felt when
sliding the filter down into the water return tank.
17. The AC Power Switch on the left side of the calorimeter should be
in the OFF position.
18. Refer to Computer Installation, page 215, and install the
computer.
19. Plug the calorimeter into the facility AC power source.
20. Turn the Oxygen supply ON with the regulator at the supply tank.
21. Turn AC Power ON to the computer and calorimeter. Permit the
computer to boot-up and display the login screen.
22. When the Windows login screen appears, login to the Windows
operating system. On computers supplied by LECO, the
administrator password will be defined as LECO. The password
may be changed after logging on.
23. Double-click the AC500 ICON on the desktop to start the AC500
application. The following screen will appear. If the computer was
not supplied by LECO, the AC500 software must be installed.
Refer to Installing Leco Software, page 95.
Installation AC500 25
24. Fill the water system as explained below.
A. Keep the combustion vessel bucket lid closed as the system
is filled.
B. Close the reservoir tank drain, jacket tank drain, and pump
purge valves.
C. Remove the water return tank from the calorimeter. Refer to
Figure 2-6, page 211.
D. Pour 9 liters of distilled water into the water reservoir.
E. Slowly open the pump purge valve and allow air to escape
from the system. As soon as water flows from the valve
nozzle, close the pump purge valve.
F. Close the pipet valve and turn the power On to the
calorimeter. Permit water to flow through the system for a
few minutes. Refer to Figure 2-7, page 212.
G. If the water level goes below the low limit switch, the pump
will turn off. It will be necessary to add water to raise the
water level above the low limit switch.
H. Add 5 milliliters of wetting agent to the water return tank.
The wetting agent will improve water flow through the
system.
I. Insert the water return tank into the water reservoir and
slowly add approximately 7 liters of distilled water into the
water return tank. When the water level stabilizes it should
be 31/4 inches (8.3 cm) below the rim of the water return
tank. If the water level is lower, add more water to reach
this point. If the water level goes beyond this point, an
alarm will sound and it will be necessary to drain some water
from the water reservoir tank.
26 AC500 Installation
J. Open the pipet valve and permit the pipet to fill. The water
level in the water return tank will drop. Do not add any more
water.
K. With 2000 ml of water in the bucket and 2000 ml of water in
the pipet, the level of water in the water return bucket
should be just above the filter holder.
MONITOR
COM 5
COM 6
COM 4
COM 2
COM 3
SWITCH
ON/OFF
POWER
FUSE
COM 1
KEYBOARD
MOUSE
AC POWER
PRINTER
POWER CORD
SWITCH ACCESS
BREAKER
VOLTAGE
CIRCUIT
DOOR
O2 IN
RETURN TANK
JACKET
DRAIN
TANK
CONNECTOR
WATER
DRAIN
O2 OUT
FAN FILTER
SWITCH
776-465
FILL
HOSE ASSEMBLY
VESSEL CHARGING
776-692
COMBUSTION
ASSEMBLY
O2 REGULATOR
608-658
774-235
OXYGEN
Figure 2-1
Facility Requirements
Installation AC500 27
VOLTAGE SELECT SWITCH
(ACCESS DOOR REMOVED)
115 V~
230 V~
POWER CORD
POWER
SWITCH
FUSE
Figure 2-2
Voltage Select Switch
Figure 2-3
Igniter Fuse Switch Position
28 AC500 Installation
POWER CORD
BLACK GREEN
POWER
115 V~
NEUTRAL WHITE
0 V~ BLACK WHITE
GROUND GREEN
CHASSIS
(CABINET)
GROUND
POWER CORD
POWER BROWN
220 V~
NEUTRAL LIGHT BLUE
0 V~
GROUND GREEN/YELLOW
CHASSIS
(CABINET)
GROUND
PLUG BLACK
BLACK T
POWER Y Y
115 V~ GREEN/YELLOW GREEN/ LIGHT *
GROUND G G YELLOW G B BLUE
230 V~ 115 V~ BLACK
POWER X X
LIGHT BLUE X
B B
BLACK
CHASSIS
(CABINET)
RECEPTACLE GROUND
PLUG BLACK
BLACK T
POWER Y Y
115 V~ 230 V~ GREEN/YELLOW GREEN/ LIGHT *
GROUND G G YELLOW G BLUE
B
115 V~ BLACK
POWER X X
0 V~ LIGHT BLUE
NEUTRAL B X
B
BLACK
CHASSIS
(CABINET)
RECEPTACLE GROUND
Installation AC500 29
608-751
FILTER
HOLDER
502-296
WATER
RETURN
FILTER
608-750
0-RING
608-749
FILTER
SCREEN
608-747
WATER RETURN
TANK
608-876
CLAW
608-665
190-381 (2 REQ'D)
2000 ML
608-748 SCREW PIPETTE TANK
WATER RETURN (2 REQ'D)
TANK
STAND-OFF
(3 REQ'D)
773-162
BUTTON
CAP
608-671
PARTICLE
FILTER
515-149
GROMMET
608-762-004
TUBING
190-269
SCREW
608-668 (2 REQ'D)
BRACE 193-185
WASHER
(2 REQ'D)
608-666 608-667
BLOCK HOLDER BASE
193-110 190-332
WASHER SCREW
(2 REQ'D) (2 REQ'D)
Figure 2-5
Pipet and Water Return Tank Filter
COMBUSTION
VESSEL
BUCKET LID
HEAT
EXCHANGER
PUMP
PURGE VALVE
Figure 2-6
Pump Purge Valve Location
Figure 2-7
Pipet Valve Positions
608-655
FILAMENT
FUSE HOLDER 608-879
608-658 SAMPLE HOLDER
COMBUSTION BRACKET
VESSEL
CHARGER
ASSEMBLY
608-780
COMBUSTION
VESSEL
HANDLE
608-656
COMBUSTION VESSEL
FILL NOZZLE
HOLDER
608-866
COMBUSTION
VESSEL
DISCHARGE
HOSE DRAIN
608-946
BASIN SHELF
Figure 2-8
Wash Basin Assembly
For international installation, a HAR approved line cord for use in the
end country must be obtained. This product requires a disconnect
device that may either be the plug of the power cord or a disconnect
device installed at the installation site if the equipment is permanently
wired.
When the plug on the power supply cord is intended to serve as the
disconnect device, the socket/outlet must be installed near the
equipment and shall be easily accessible. When the wiring is
permanently connected, a disconnect device (such as a circuit breaker
or switch) must be lockable, installed near the equipment, and be
easily accessible.
3. For Port, click the arrow and select the port location from the list
that matches the port selected on the PC tower. Default is COM1.
4. For Baud Rate, click the arrow and select the baud rate from the
list. Refer to the Sartorius Instruction Manual. Default is 1200
bps.
5. For Data Bits, click the arrow and select the number of data bits
from the list. Refer to the Sartorius Instruction Manual. Default is
7.
6. For Parity, click the arrow and select either odd or even. Refer to
the Sartorius Instruction Manual. Default is Odd.
7. For Stop Bits, click the arrow and select either 1 or 2 from the
list. Refer to the Sartorius Instruction Manual. Default is 1.
8. For Readability, the default is 4.
9. For Balance Mode, select Normal if a balance is connected to the
system. Select Disabled if a balance is not connected to the
system.
10. For Send Command, click the arrow and select the desired
command from the list. The default is <ESC>P<CR><LF>.
11. For Tare Command, click the arrow and select the desired
command from the list. The default is <ESC> T.
Modem Installation
1. Turn the instrument and computer off.
2. Unpack the Smartline package. Refer to Components and
Accessories List, page 127, for package contents.
3. Connect the modem to the rear of the computer using the modem
cable.
4. Connect the power cube, supplied with the modem, to the power
input jack on the modem.
5. Connect the modem to the same outlet as the computer.
6. Turn the modem, computer, and instrument on.
7. Insert the Modem Installation CD-ROM in the CD-ROM disk drive
(typically D:). The Installation program should start
automatically.
8. When the Welcome screen is displayed, click Next.
9. When the Installation screen is displayed, select Software then
select US Robotics control center and click Install.
10. When the Setup screen is displayed, click Next then Next again.
The program will automatically choose the best location (folder)
in which to install the program.
11. Click OK when finished.
12. Turn the computer off then back on again to reboot the system to
activate the modem software.
13. After Modem Installation, page 38, the NetOp software must be
installed. Refer to NetOp Software Installation and install the
software, page 39.
BUCKET 608-665
LID 2000 ML
PIPET TANK
608-671
PARTICLE
FILTER
608-658
COMBUSTION
VESSEL CHARGER
589-923
COMBUSTION
VESSEL FILL
SWITCH
608-662
PIPET OXYGEN
VALVE PRESSURE
GAUGE
608-948
COMBUSTION
VESSEL CLEANING/
WASH BASIN
STATION
Figure 4-1
Controls and Indicators
The following section explains the menus available on the menu bar.
Database Menu
Selections on the Database menu are available when logging has been
enabled for a database connection. Refer to Managing Databases,
page 454.
Presentation Time: Refer to Presentation Time,
page 468.
View History: Refer to View History, page 469.
Archive Log: Refer to Archive Log, page 470.
Delete by Date Refer to Delete by Date, page
471.
Exit: Select Exit to close the software program.
Edit Menu
Select the Edit menu to access commands for entering sample data
into the spreadsheet.
Insert: Refer to Inserting a Sample Out of Sequence,
page 525.
Fill: Refer to Repeating Data in Multiple Cells,
page 525.
Cut: Select Cut to remove data from an entire row in
the spreadsheet to the clipboard. Data can then be
pasted into another row.
Copy: Select Copy to duplicate data from a row in the
spreadsheet to the clipboard. Data can then be pasted
into another row of the spreadsheet, duplicating the
copied row.
Paste: Select Paste to insert data from the clipboard to
a selected row in the spreadsheet.
Diagnostics Menu
Select the Diagnostics menu to monitor the instrument and help
maintain proper instrument operation.
Chart: Refer to Monitoring Ambient Chart, page 83.
Solenoids/Switches: Refer to Solenoids and
Switches, page 85.
Hardware Calibration: Refer to Hardware
Calibration, page 89.
Communications: Refer to Communications, page
814.
Help Menu
Select the Help menu to access further information about the
instrument.
Information and Help: Select Information and Help
to access the instruction manual in a pdf.
About: Select About to access instrument
information such as the version number, serial
number, and copyright information and to register the
instrument using the Edit License dialog box. Refer to
Registering the Software, page 922.
Toolbar
Toolbar Button Action
This section explains how to customize the font and columns in the
spreadsheet and how to navigate the spreadsheet.
Customizing the samples that display can be accomplished using
sample filters. Refer to Using Sample Filters, page 447.
Modifying Font
Use the Font dialog box to change the font, style, and size of text.
The current selections are highlighted and appear in the boxes.
1. Right-click anywhere in the spreadsheet and select Properties or
click the Configuration menu and select Display.
2. Click the Font tab.
3. While making changes, use Sample Text to view the text as it will
display before accepting the changes.
4. Change the Font Name, Style, or Size by single-clicking the
desired selection in the list.
5. Select OK when the text displays as desired.
Text Import Data provides the means to import sample data into the
instrument software. For example, the sample weights could be
obtained from a balance that is not connected to the instrument.
Sample information including the mass weight is then entered into a
text editing or spreadsheet program, and Text Import Data can be
used to import these sample weights into the sample spreadsheet.
2. Select Add. A new row will display with Sample- in the Name
column.
3. Double-click in the Name column of the newly added row and
enter a new name for the sample, if desired.
4. Select a method for the sample from the menu. If the desired
method has not been created, refer to Creating or Modifying
Methods, page 517.
5. Enter information in the Comments and Description column, if
desired.
6. Click OK. The sample template can now be selected from the
Name column in the spreadsheet.
7. After a sample template has been created, samples can be easily
logged in to this template. Refer to Logging In Using Sample
Template, page 446.
2. Select Add to insert a blank row into the table. Transmit- will
appear in the new row, which will be positioned above the
currently selected row.
3. Double-click in the Name column of the newly added row and
enter a new name for the sample.
4. Refer to Properties Dialog Box, below, to configure the
transmission format.
Fields Tab
Select Fields to determine the Fields that will transmit. Fields are
defined in the Glossary, page 121.
To transmit a field, select the field from the Fields selection box and
then select Add. To remove a field from transmit, select the field from
the Transmit selection box and then select Remove.
To change the order in which the fields will transmit, select the field to
move. Select Promote to move the field to a higher position or select
Demote to move the field to a lower position.
Delimiters Tab
Select Delimiters to configure the display of the transmitted data.
1. Select the XML Format checkbox, and the data will export in XML
format. When the XML checkbox is cleared, data will transmit in
normal text.
2. Select the Column Headers check box to have the transmitted
data organized with column headers.
3. Select one delimiter as a means to separate the data. Comma
separates the data by commas; Tab separates the data by a
blank space equivalent to one tab; Other separates the data by a
user-determined punctuation mark or symbol.
4. Select how the Start of Transmission will appear. None displays
nothing before transmitted data; Other allows you to enter the
text or symbol that will appear, or you can click the arrow and
then select an item from the list.
Automation Tab
On the Automation tab, select the Automatically Transmit Each New
Analysis checkbox, and each new analysis will transmit.
To add a field to the Transmission Format, select the desired field from
the Fields selection box and then select Add Field to Transmit Format.
The field appears in the Export Format selection box.
In the Transmission Format box, information such as nested xml
formulas or formulas with complicated formatting can be typed or cut
and pasted into the box; however, the XML check box on the
Delimiters tab must be cleared. When additional information is entered
directly, it is not necessary to use []. The software uses [] to identify a
field. When transmit information is added or changed in the
Transmission Format box, the appearance of the transmission format
can be viewed using the Preview tab.
Preview Tab
Select the Preview tab to display the appearance of the data before it
is transmitted. Any changes can then be made before the data is
actually transmitted. The following example displays the data with the
following parameters selected on the Delimiters tab: Column Headers
checkbox checked, comma checked for Delimiter, and the Carriage
Return checkbox checked for End of Transmission. Refer to Delimiters
Tab, page 423, to change any delimiters.
The way the date and time display can be changed using the Date
and Time Properties dialog box on the Microsoft Windows Control
Panel.
2. Select Add. Export- will display in the new row that has been
added to the Export Formats spreadsheet.
3. Double-click in the new row to change the name of the export, if
desired.
4. Select Properties and determine the properties of the text export
format. Refer to Properties Dialog Box, page 428.
Filename Tab
Use the Filename tab to configure the file for the export.
1. Enter the name for the file or select Browse and then select the
file from the list.
2. Select one of the following:
Select Append to File to attach the sample analysis data
highlighted in the spreadsheet to the previously saved file.
Fields Tab
Select Fields to determine the Fields that will transmit.
To export a field, select the field from the Fields selection box and then
select Add. To remove a field from export, select the field from the
Export selection box and then select Remove.
To change the order in which the fields will export, select the field to
move in the Export list box and then select Promote to move the field
to a higher position or select Demote to move the field to a lower
position.
For Average, select the arrow and then select Off to export the
individual results for each sample. Select On to calculate and then
export an average value for each field selected. Select Group by
Sample Name to calculate an average value for each field and then
export the fields grouped by sample name.
Delimiters Tab
Select Delimiters to configure the display of the exported data.
1. Select the XML Format checkbox, and the data will export in XML
format. When the XML checkbox is cleared, data will transmit in
normal text.
2. Select the Column Headers check box to have the exported data
organized with column headers.
3. Select one delimiter as a means to separate the data. Comma
separates the data by commas; Tab separates the data by a
blank space equivalent to one tab; Other separates the data by a
user-determined punctuation mark or symbol.
Automation Tab
On the Automation tab, select the Automatically Transmit Each New
Analysis checkbox, and each new analysis will transmit.
Preview Tab
Select the Preview tab to display the appearance of the data before it
is exported. Any changes can then be made before the data is actually
exported. The following example displays the data with the following
parameters selected on the Delimiters tab: Column Headers checkbox
checked, comma checked for Delimiter, and the Carriage Return
checkbox checked for End of Transmission. Refer to Delimiters Tab,
page 429.
The way the date and time display can be changed using the Date
and Time Properties dialog box on the Microsoft Windows Control
Panel.
Test Commands
SendUsed to test the balance interface. Click Send to send and
display a weight sent from the balance to the computer. The weight
will appear in the sample grid.
TareUsed to test the balance interface. Click Tare to set the weight
displayed on the balance to zero.
3. Click the drop down arrow in the serial port selection box and
select the serial port used to communicate with the instrument.
Selections are COM 1 through COM 8. The other entries on this
screen cannot be changed from their default settings.
4. When finished, click OK.
1. From the Samples menu, click Login. The Sample Login screen
will appear.
2. Click the drop down arrow to the right of sample name and enter
the desired sample name from the drop down list. The sample
name, method comments, and description will automatically be
entered into the sample grid.
3. Continue with sample login and enter the other necessary
information.
2. Click the drop down arrow to the right of sample name and select
the desired sample name from the drop down list. The sample
name, method comments, and description will automatically be
entered into the sample grid.
3. Continue with sample login and enter the other necessary
information. Refer to Sample Login for additional information.
Adding Filters
Use the Add command to create a new filter.
1. To add a new filter, click Add. New Filter will display in the
Filters dialog box. Each time Add is clicked New Filter will
increment by one and display as New Filter 1, New Filter 2,
etc.
Filter Properties
Use the Properties command to determine the parameters for a filter.
1. To determine or modify the properties for a filter, click on the
name of the filter in the Sample Filter dialog box and then click
Properties. The Sample Filters dialog box will display.
2. Click New Filter and then click Properties. The Sample Filter dialog
box will display.
3. In the Filter Name entry box, type in the name for the filter.
4. For Export Format, click the arrow to select the export format for
this filter.
5. For Transmit Format, click the arrow to select the transmit format
for this filter.
6. Add a parameter, delete a parameter, or modify columns for the
sample filter.
A. To add a parameter,
1) In the Field entry box, click the arrow and select the
desired field.
2) In the Compare entry box, click the arrow and select the
desired comparator. The comparator selections change
depending on the selected field.
3) In the With entry box, enter the desired parameter
depending on the selected field. For example, if Analysis
Date is selected for Field and On is selected for
Compare, in the With column, enter the desired date for
the analysis results.
Deleting Filters
Use the Delete command to remove sample filters from the Filters
dialog box.
1. Click on the desired sample to remove.
2. Click Delete.
Predefined Filters
Filters that may be frequently used have been predefined in the
software. Default filters are All Samples and Today's Samples. The
remaining filters that display are based on created methods. To select
a predefined filter, click on the name of the filter, and then click OK.
In the Sample Filter dialog box, the Filter Name displays the name of
the predefined filter. The Description explains the filter and whether or
not the fields or automatic updates can be modified. For example, for
All Samples, only the Fields can be modified. To modify the fields, click
the Fields button. For further information about the fields, refer to
Modifying Columns, page 416.
When the Automatically Reapply Sample Filter check box is checked,
the software will update the sample filter according to the parameters
entered for Start Time and Update Frequency. To determine the
automatic update, refer to step 7, page 449.
Adds unique fields to the sample grid and sample login screen. The
field can include a formula for calculation or contain only text.
1. From the Configuration menu, click User Defined Fields. The User
Defined Fields screen opens.
2. Click the Add button to add a user defined field to the sample
grid. The field will also appear on the Sample Login screen.
3. Click in the cell under name and enter a name for the added field.
4. If the field will be used for a calculation enter a formula for
calculation within the field. Refer to Functions, page 452.
5. Check Read Only to restrict the operator from making changes to
the field.
6. Check Numeric to restrict text from the field and only display
numeric information.
7. Under Format enter the number formatting. Refer to Number
Formatting, page 453.
8. Click OK to add the new user defined field to the sample grid and
sample login screen.
The term exp means expression, which may be a single number
or a function that returns a number.
The term cond(x) means a comparison user (<, >, =, etc. . . )
or the @AND, @OR, and @XOR functions.
The @IF function evaluates its condition expression and then
performs the operations in either the true or false section.
Functions
Function Definition
+ Plus
- Minus
* Multiply
/ Divide
= Equal
!= Not equal
Number Formatting
Formatting creates a string representation of the floating point
numeric value from the supplied format string:
[A|, [Sn.m], [En.m], [Fn.m], [#~##.#~###] and [In]
where:
A = Specifies automatic or default format. Also used if psz Format is
NULL or points to a NULL String.
S = Specifies scientific notation.
E = Specifies engineering format (Scientific format where exponent
mod 3 == 0).
F = Specifies fixed point format.
# = Specifies a digit format to use. # is replaced with digits. When
used with F,S,E or I, # preceding the decimal defines the final number
of significant digits and # after the decimal defines precision, as
interpreted in the original fixed point format.
I = Specifies integer.
n = Specifies the number of significant digits to be represented. When
n is not specified the default value is DBL_DIG.
m = Specifies the maximum number of digits after the decimal place.
When m is not specified the default value is DBL_DIG.
Examples:
FormatDoubleString( 0.123456, "F4.5", ...) ==> "0.1235"
FormatDoubleString( 0.000789, "F3.5", ...) ==> "0.00079"
FormatDoubleString( 0.000789, "F3.2", ...) ==> "0.00"
FormatDoubleString( 321.8765, "F3.5", ...) ==> "322"
FormatDoubleString( 0.123456, "S4.5", ...) ==> "1.234E-1"
FormatDoubleString( 0.000789, "S3.5", ...) ==> "7.9E-4"
FormatDoubleString( 0.000789, "S3.2", ...) ==> "0.00E+0"
FormatDoubleString( 321.8765, "S3.5", ...) ==> "3.22E+2"
FormatDoubleString( 0.123456, "E4.5", ...) ==> "123.5E-3"
FormatDoubleString( 0.000789, "E3.5", ...) ==> "790E-6"
FormatDoubleString( 0.000789, "E3.2", ...) ==> "0.00E+0"
FormatDoubleString( 321.8765, "E3.5", ...) ==> "322E+0"
FormatDoubleString( 321.8765, "I2", ...) ==> "320"
Create Backup
1. In the Start dialog box, highlight the database connection of the
database to back up. The file location of the corresponding
database will appear as shown below.
Retrieve Backup
Data can be retrieved using either method described below. The
instrument software must be closed to perform either procedure.
If there is a problem with the original database and the backup is to
replace it, copy the backup database folder from where it was stored
Validating a Database
During database validation, the database files will be reviewed for
corruption. If orphan objects are found, refer to Reviewing Orphan
Objects, page 463.
To validate a database,
1. Exit the instrument software.
2. Click Start on the Windows desktop, select Programs, select
LECO, select the instrument, and then click Database
Management. The Start dialog box will appear.
Database Menu
Items on the Database menu in the instrument software will only
appear if logging was enabled when the database connection was
created. Refer to Enabling Logging for an Existing Database, page 4
64.
2. Select the time and date for the presentation time as follows:
Select the Live check box to view current information or clear the
Live checkbox in order to select a different date and time.
When the Live check box is cleared, Date and Time are active.
View History
When logging is enabled for a database connection, View History
displays a log of database changes that could be used to identify the
time/date when some particular change took place. It can only be used
to identify who made a change because there is no description of what
each change represents.
1. Click the Database menu in the instrument software and select
View History. The View Log dialog box will appear.
2. The log items that appear in the View Log dialog box are
explained below.
Date/Time displays the date and time that the database change
occurred.
Entry ID displays a unique identifying number for the database
object that was changed. For example, each sample would have a
unique number.
Action displays Created, Modified or Deleted to indicate the type
of action that occurred.
User displays the username of the person who performed the
action.
3. Select the time zone to use for displaying Date/Time. Local
displays the local time zone. UT displays the universal time that
corresponds with the UTC (Universal Coordinated Time) as
defined by the National Bureau of Standards.
Operation AC500 51
Advanced Tab ................................................................... 538
Preview ............................................................................ 539
Exporting Data ....................................................................... 539
Exporting Buffers ................................................................... 541
Importing Data ........................................................................... 542
Transmitting Data ....................................................................... 543
Transmit Format Configuration................................................. 543
Fields Tab ........................................................................ 544
Delimiters Tab .................................................................. 545
Automation Tab ................................................................ 546
Communication Settings .................................................... 547
Advanced Tab ................................................................... 548
Preview Tab ...................................................................... 549
Transmitting Selected Samples ..................................................... 550
Data Backup ............................................................................... 551
Transferring Export ................................................................ 551
Transferring Import ................................................................ 552
Error Messages ........................................................................... 554
52 AC500 Operation
Illustrations
Operation AC500 53
This page intentionally left blank.
54 AC500 Operation
Logging in a Sample
2. Enter the Sample Name in the sample name selection box or click
on the drop down arrow to select from a list of sample names.
3. For Mass, enter the mass of the sample. Click Balance to enter a
mass from an external balance.
4. For Method, enter the method used for analysis. Click the down
arrow to select from the list of methods.
5. For Vessel, enter the identification number of the combustion
vessel that the logged in sample is being analyzed in.
6. For Comments, enter additional information about the sample if
desired.
Operation AC500 55
7. For Description, enter a statement used to explain or identify a
sample if desired.
8. For Operator, enter the name of the operator if desired.
9. Click OK to login the sample.
10. Select Add to End of List to enter the logged in sample in the last
row of the sample grid.
11. Select Add as Next Sample to Run to enter the logged in sample
after the last sample that was analyzed.
12. For Keep Logging Samples until Cancel is Pressed, select the
check box to log consecutive samples.
56 AC500 Operation
Preparing Crucibles
Operation AC500 57
Preparing Samples
Do not touch the sample with your fingers or hands as this will
contaminate the sample.
58 AC500 Operation
Combustion Vessel Preparation for Wire Fuse
1. Prepare the sample and place it into the crucible. Refer to Sample
Preparation, page 58.
2. To ensure repeatable results, dry all moisture from the outside of
the combustion vessel with a clean, lint-free cloth.
3. Prepare Combustion Pressure Closure using the steps below.
A. Remove, rinse, and thoroughly dry combustion vessel cap.
Refer to Figure 5-2, page 511.
B. Remove the combustion vessel closure.
C. Rinse electrode arms and slide the sleeves up and down until
they move freely, dry arms and underside of combustion
vessel closure.
D. Turn the combustion vessel closure over and shake it to
remove the water from the cavities in the valve assembly and
electrode connectors.
E. Thoroughly dry the top of combustion vessel closure.
F. Dry the combustion vessel cleaning station sample holder
bracket. Refer to Figure 2-7, page 213.
G. Place the combustion vessel closure on the sample holder
bracket.
4. Place the crucible containing the sample into the crucible holder.
Center the sample in the crucible holder to prevent misfires.
5. Attach the fuse wire. The length of the fuse wire must be 10 cm
long.
A. Lift the sleeve on the electrode above the slot in the crucible
holder.
B. Feed one end of fuse wire into the notch on the electrode arm.
Allow approximately 2mm of the fuse to overhang.
C. Slide the sleeve firmly over the fuse wire.
D. Repeat steps A through C with the other end of the fuse wire.
Operation AC500 59
0.5 cm
Figure 5-1
Sample PreparationWire Fuse
1. Prepare the sample and place it into the crucible. Refer to Sample
Preparation, page 58.
2. To ensure repeatable results, dry all moisture from the outside of
the combustion vessel with a clean, lint-free cloth.
3. Prepare Combustion Pressure Closure using the steps below.
A. Remove, rinse, and thoroughly dry combustion vessel cap.
Refer to Figure 5-2, page 511.
B. Remove the combustion vessel closure.
C. Rinse and dry electrode arms and underside of combustion
vessel closure.
D. Turn the combustion vessel closure over and shake it to
remove the water from the cavities in the valve assembly
and electrode connectors.
E. Thoroughly dry the top of combustion vessel closure.
F. Dry the combustion vessel cleaning station sample holder
bracket. Refer to Figure 2-7, page 213.
G. Place the combustion vessel closure on the sample holder
bracket.
4. Attach the 502-461 wire between the electrode posts
A. Loosen the two Phillips screws at the bottom of the electrode
posts of the combustion vessel closure.
B. Wrap each end of the wire around the threads of the screws.
C. Tighten the screws keeping the wire tight.
5. Place the crucible containing the sample into the crucible holder.
Center the sample in the crucible holder to prevent misfires.
6. Attach the Thread Fuse. The length of the fuse wire must be 10
cm long.
A. Fasten the Thread onto the middle of the ignition wire. Refer
to Figure 5-3, page 513. One end should be long enough to
come in contact with the sample.
The 502-461 fuse wire should be replaced when necessary to prevent
misfires. Refer to Maintenance Counters, page 64.
SCREWS
TO ATTACH
STRING WIRE
(ONE ON
EACH SIDE
OF WIRE)
Figure 5-3
Sample PreparationThread Fuse
1. Thoroughly dry both the inside and the outside of the combustion
vessel bucket. Buff the bucket if required to remove any
fingerprints.
2. Position the combustion vessel bucket under the pipet Refer to
Figure 2-4, page 210.
3. Fill the combustion vessel bucket with 2000 ml of water from the
pipet by turning the pipet valve to the Fill position. Refer to Figure
2-6, page 212.
4. Allow the water stream from the pipet to hit the inside wall of the
combustion vessel bucket. This will make the water swirl and
reduce the amount of bubbles and splashing in the bucket.
5. Leave the combustion vessel bucket under the pipet valve for
several seconds after the pipet has been emptied to allow any
remaining drops of water to fall into the combustion vessel
bucket.
6. When the pipet is empty, turn the pipet valve back to the Water
Flows Through position. Refer to Figure 2-6, page 212.
7. Pull up on the combustion vessel bucket lid latch and lift the cover
to the complete vertical position.
8. Place the combustion vessel bucket inside the combustion vessel
bucket well being careful not to spill any water. Make sure the
indent in the bottom of the combustion vessel bucket, is
positioned towards the front of the calorimeter. This allows room
for the stirrer blades and prevents them from hitting the
combustion vessel during analysis. Refer to Figure 4-1 on
page 46.
9. Fold the handle down, flat on top of the combustion vessel
bucket, towards the back of the instrument.
10. Attach the carrying handle to the combustion vessel by inserting
the pins on the end of the handle into the holes on the
combustion vessel cap. Refer to Figure 5-2, page 511.
11. Lower the combustion vessel, using the carrying handle, into the
combustion vessel bucket taking care not to disturb the sample.
Center the combustion vessel over the indent in the bottom of the
bucket.
12. Remove the combustion vessel handle and tap it gently over the
bucket to return any water droplets into the bucket. Be careful
not to splash water on jacket insulation.
13. Place the combustion vessel handle on the wash station hanger.
14. Connect the combustion vessel fuse leads to the combustion
vessel electrodes. Grasp the insulated portion of the combustion
vessel fuse leads when inserting them into the combustion vessel
electrodes so the water in the bucket is not touched.
Create a Method
1. From the Configuration menu, click Methods. The Methods dialog
box shown below will appear. For further information about the
buttons on the Method Dialog Box, refer to Dialog Box Buttons,
page 412.
Standard Calibration
Before this procedure is performed, standard calibration samples must
be analyzed. Refer to Sample Analysis and perform analysis using
defined standard calibration samples. Refer to Analyzing a Sample,
page 521.
1. From the Configuration menu, click Calibration. The Standards
screen will appear.
1. Click and select a row in the sample grid after the last analyzed
sample. Insert will insert a row before the selected row.
2. From the Edit menu, click Insert. A row will be inserted into the
sample grid before the selected row.
2. From the Edit menu, click Fill. The Fill screen will appear.
Cut
Cut is used to remove a sample from the sample spreadsheet and copy
it to the Windows Clipboard. From there it can be Pasted into another
location in the sample grid.
1. Select a sample by clicking on a row in the sample grid. The row
should be highlighted.
2. Select the Edit menu and click Cut. The sample will be removed
and cut to the clipboard.
Copy
Copy is used to copy a sample from the sample grid and copy it to the
Windows Clipboard. From there it can be Pasted into another location
in the sample grid.
1. Select a sample by clicking on a row in the sample grid. The row
should be highlighted.
2. Select the Edit menu and click Copy. The sample will be copied to
the clipboard.
Paste
Paste is used to copy a sample that was cut or copied to the Windows
Clipboard and insert it into the sample grid.
1. Select a sample location in the sample grid by clicking on a row.
The row should be highlighted.
2. Select the Edit menu and click Paste. The sample will be copied to
the sample grid.
Deleting a Sample
Use Delete a Sample to remove a sample row from the sample grid. If
a database is created with log activity enabled, a deleted sample can
be viewed by setting the Presentation Time to a time prior to the
deletion.
1. Select a row in the sample grid to delete. If more than one row
should be deleted, click and drag the mouse pointer down the
desired number of rows to delete. More than one row will be
selected.
2. From the Samples menu, click Delete.
3. Click Yes to delete the row or rows of sample information. The
selected sample information will be deleted from the sample grid.
It is not possible to restore this data once its deleted.
2. Click the Samples menu and select View History. The View History
screen will appear. Refer to View History Definitions below.
3. Click the Local Time button, in the Time screen, to display the
time in local 12-hour time.
4. Click the UTC button, in the Time screen, to display the time in
Universal Coordinated Time.
5. Click OK to exit view history.
2. Click the Samples menu and select Sign. The Sign screen will
appear.
View Signatures
View Signatures permits the operator or laboratory manager to view a
sample signature and determine if it is valid. A sample must be signed
before the signature can be viewed. Refer to Applying Sample
Signature, page 528. If information associated with the sample was
changed after it was signed, the signature will be cleared and the view
signatures screen will not appear in step 2.
1. Click on a row to select a sample from the sample grid to view.
2. Click the Samples menu and select View Signatures. The View
Signatures screen will appear.
When multiple columns are selected to appear in the report, the font
and page orientation can be adjusted to improve the readability of
the report. The font size can be changed using the font tab as
described in Configuring Report before Printing, page 530.
6. Click OK. The Print dialog box will appear.
7. Click OK to print.
Export Properties
Filename Tab
1. Select the Configuration menu and click Text Export Formats. The
Filename Format Screen will appear.
Fields Tab
1. Click the Fields tab. The Fields Format Screen will appear. Select
the desired fields to export in the file.
2. Position the cursor in the Export Format and click the mouse to
place the cursor at the point in the format to edit a field.
3. Select a Field from the Fields box and click Add Field to Export
Format to add a field.
4. Highlight a Field in the Export Format and then press Delete on
the keyboard to delete it.
5. Click OK to save the changes and exit.
6. Click Cancel to exit this procedure without making changes.
Exporting Data
Export Text Data permits the operator to select sample data and
export that data to a file. Sample data can be inserted into an Excel
sample grid from the exported file. If no sample is selected only the
last sample in the sample grid will be exported and saved.
1. Click and drag the mouse pointer to select the rows with sample
data to export to a file.
3. Select the Export Format and click OK. The Export Filename
selection screen will appear.
2. Click the Sample menu and select Text Export Buffers. The Export
Buffers Filename selection screen will appear. File extensions of
(.txt) or (.csv) are good choices to insert the saved files into an
Excel sample grid.
2. Enter a filename and extension. Click Save to export and save the
file.
Text Import Data permits the operator to import sample data from
another application such as a Microsoft Excel sample grid.
1. Click the Sample menu and select Text Import Data. The Import
File selection screen will appear.
2. To add a new transmission format to the list, click Add and enter a
transmission format name in the row that appears at the bottom of
the list.
3. To move a selected transmission format up towards the top of the
list select it and click Promote.
4. To move the selected transmission format down towards the
bottom of the list select it and click Demote.
5. Select Ascending to list the transmission formats in ascending
alphabetical order.
6. Select Descending to list the transmission formats in descending
alphabetical order.
7. To delete a transmission format, select it and click Delete.
8. Click OK to save any changes and exit.
2. Click XML to export the text data in XML format. If XML is selected
no other parameters in the screen can be configured.
3. Click Column Header to transmit the column header.
4. Select or enter the Field, Start of Transmission, and End of Line
delimiters.
5. Click OK to save the changes and exit.
6. Click Cancel to exit this procedure without making changes.
Advanced Tab
1. Click the Advanced tab. The Advanced Transmission Format
screen will appear.
2. Position the cursor in the Export Format and click the mouse to
place the cursor at the point in the format to edit a field.
3. Select a Field from the Fields box and click Add Field to Export
Format to add a field.
4. Highlight a Field in the Export Format and then press Delete on
the keyboard to delete it.
5. Click OK to save the changes and exit.
6. Click Cancel to exit this procedure without making changes.
Transmit Data permits the operator to select sample data and transmit
the data to a peripheral device. Sample data can be transmitted from
a serial port or over a network.
1. Before data is transmitted a transmission data format must be
defined. Refer to Transmit Data Format Configuration, page 543.
2. Click and drag the mouse pointer to select the rows with sample
data to transmit. If no sample is selected only the last sample in
the sample grid will be transmitted.
Transferring Export
1. Click and drag the mouse pointer to select the rows with sample
data to export to a file.
2. Click the Sample menu and select Transfer Export. The export file
selection screen will appear.
4. Select the folder to store the sample data and enter a unique file
name without an extension. The software will automatically add the
proper file extension.
Transferring Import
Transfer Import restores sample data that was saved as a file by
Transfer Export. Refer to Transferring Export, page 551.
1. Click the Sample menu and select Transfer Import. The Import
File selection screen will appear.
Maintenance AC500 61
Illustrations
62 AC500 Maintenance
Periodic Maintenance Schedule
Maintenance AC500 63
Hydrostatic and Proof Test Maintenance
After every 500 firings the oxygen combustion vessel should be
disassembled for cleaning and the o-rings should be replaced.
Hydrostatic and proof tests shall be repeated if any of the following
have occurred:
Five thousand (5000) firings.
Firing with an excessive charge of either sample (>35000 Joules*)
or oxygen (>500 PSI).
Ignition of any internal part of the vessel, including crucible.
Excessive corrosion or surface defects.
A noticeable change in thread play between vessel enclosure and
capnut; thread tolerance should not be more than 0.010 inch axial
and 0.020 inch diametrical.
See the Oxygen Combustion Vessel Safety Notice, page 131.
64 AC500 Maintenance
Maintenance Counters
The Counters dialog box tracks the periodic maintenance performed on
selected instrument components and alerts the operator when periodic
maintenance is due. Until the required maintenance is logged in, the
Maintenance is Required Soon dialog box will continue to display each
time an analysis is initiated. Refer to Logging In Periodic Maintenance,
page 67.
Only users with permission can edit the Counters dialog box. Refer to
Adding or Changing User Permission for Instrument, page 434.
Maintenance AC500 65
D. In the Warning column, enter a value to indicate the number
of items from the Type column that can occur before periodic
maintenance is necessary. Analysis will not stop when this
counter limit is reached, but a warning prompt will display
until the maintenance is performed. To turn this function off,
enter zero.
E. In the Stop column, enter a value to indicate the number of
items from the Type column that can occur before the
instrument will stop and no longer analyze samples until
periodic maintenance is performed. To turn this function off,
enter zero.
F. The Resets column displays the total number of times that
periodic maintenance has been performed.
G. The Reset column displays the date and time periodic
maintenance was last performed.
H. In the Comments column, enter any notes regarding
maintenance.
4. Modify the Counters dialog box, if desired, as follows:
To delete a counter, select the desired counter and then select
Delete.
To return the counters in the list to the original settings and/or
return system counters that have been deleted, select Restore.
To sort the Counters, select Ascending, and the Counters will
automatically display in alphabetical and/or numerical order from
lowest to highest. Select Descending, and the Counters will
automatically display in alphabetical and/or numerical order from
highest to lowest.
To change the order of the Counters, select desired counter to
move. Select Promote to move a counter to a higher position in
the table. Select Demote to move a counter to a lower position in
the table.
66 AC500 Maintenance
Logging In Periodic Maintenance
Use the Maintenance dialog box to log in periodic maintenance after it
has been performed. The maintenance counter will be reset when
periodic maintenance is logged in.
1. Select Maintenance and then select Login. The Maintenance dialog
box will display.
2. Select the check box for the item that received periodic
maintenance. If an item is already checked, periodic maintenance
is necessary based on the parameters set for that item in the
Counters dialog box. Refer to Maintenance Counters, page 64.
3. In the Maintenance dialog box, select and enter information in the
Comments dialog box if desired.
4. Select OK. A dialog box, such as the following example, will
display for each item checked.
Maintenance AC500 67
Viewing Log
The Log File displays a history of past events. This information can
help to troubleshoot problems that may occur during operation. Every
time one of the following procedures is performed, it will be recorded
in the log file.
Change of Mass by Operator
Data Out of Range (Ambient Monitor parameter values)
Date of last calibration
Date scheduled maintenance is performed
Network Errors
Network Sign-on Errors
Sample Aborted
Sample Deleted
Standard check out-of-range
1. Select Maintenance and then select View Log File. The log will
display.
68 AC500 Maintenance
Combustion Vessel Maintenance
General Maintenance
Maintenance AC500 69
Fan Filter Removal and Cleaning
776-465
FAN FILTER
776-426
FILTER
HOLDER
190-269
SCREW
(4 RE'D)
Figure 6-1
Fan Filter Removal
602-646
CLAMP
608-671
PARTICLE
FILTER
PIPET
VALVE
Figure 6-2
Particle Filter Replacement
If the water return filter is not excessively dirty, steps two, three,
five, six, nine and ten may be skipped. In this case, tilt the water
return filter assembly sideways and pull it out of the water return
bucket. After replacement, turn the water return filter assembly
sideways and insert it into the bucket. As the filter assembly
approaches the bottom of the bucket, turn the filter assembly to the
horizontal position and position it on top of the water return filter
screen.
1. Open the lid of the calorimeter that covers the water return
bucket.
2. Slowly remove the water return bucket from the water return
tank by pulling it up. Stop the flow of water from the bottom of
the bucket by covering the small hole in the bottom.
3. Lift the water return bucket out of the water return tank and
drain the water into a sink or the combustion vessel fill station.
Be careful that the filter assembly and filter screen remains in the
bucket. Do not drain the water from the water return bucket back
into the water return tank.
4. Remove the water return bucket filter assembly from the water
return bucket and set it aside for use in step 7. Refer to Figure 6-
4, page 613.
5. Remove the water return bucket screen and wipe it dry with a
clean lint free towel or set it aside to air dry.
6. Dry the water return bucket with a clean lint-free towel or set it
aside to air dry.
7. Remove the o-ring from the filter holder and discard the wet
filter.
8. Install a new filter over one end of the filter holder and secure it
with the o-ring removed in step 7. Set it aside for use in step 11.
9. Insert the water return bucket into the water return tank and let
it settle to the bottom of the water return tank. Permit the water
level in the water return bucket to stabilize.
10. Insert the filter screen into the water return bucket until it is
positioned on the posts on the bottom of the water return bucket.
11. Insert the filter assembly into the water return bucket until it is
positioned on top of the filter screen. Tilt the filter assembly to
permit water to flow around it as it is being inserted. The filter
should be positioned down and set on the filter screen.
502-296
WATER
RETURN FILTER
608-750
O-RING
608-749
FILTER
SCREEN
608-747
WATER
RETURN
BUCKET
608-748
BUCKET STAND-OFF
(3 REQ'D)
Figure 6-3
Water Return Filter Replacement
When draining the system water, replace the reservoir bucket filter.
Refer to Water Return Filter Replacement, page 612.
1. Close the reservoir tank drain, jacket tank drain, and pump purge
valves.
2. Turn the calorimeter power Off.
3. With the water return bucket removed, pour 9 liters of distilled
water into the water return tank. Keep the lid on the combustion
vessel bucket closed as the system is being filled.
4. Slowly open the pump purge valve and allow air to escape from
the system. As soon water flows from the valve nozzle, close the
pump purge valve. Refer to Figure 2-5, page 211.
5. Close the pipet valve and then turn the calorimeter power On.
Allow water to flow through the system for a few minutes then
add 5 milliliters of wetting agent to the water return tank. Refer
to Figure 2-6, page 212.
6. Insert the water return filter bucket into the water reservoir.
7. Slowly pour approximately 7 liters of distilled water into the water
return tank keeping the water level below 31/4 inches (8.3 cm)
from the top rim of the water return bucket. When the water
stops equilibrating it should be approximately 31/4 inches (8.3 cm)
from the top rim of the water return bucket.
8. Open the pipet valve and allow the pipet to fill. The water level in
the reservoir tank will drop.
Figure 6-4
Pressure Safety Valve
4. Use the 1/8" Allen wrench to adjust the set screw inside the valve
locking nut until the pressure gauge reads approximately 500 psi
(34.47 bar).
5. When the target pressure of 500 psi (34.47 bar) is reached, with
the Allen wrench still in the hole of the valve locking nut, use the
open-ended wrench to tighten the valve locking nut.
6. Use the adjusting screw to set the pressure regulator to 450 psi
(31.03 bar).
ASTM Mode
The sample weight and %N can be entered during sample login or after the analysis has been completed
when editing result parameters.
When a normality other than 0.394 is used, the Volume is adjusted by the ratio of the normality specified
in System Options for "Normality N" to the normality specified by ASTM of (0.394 N). A typical volume V
can be entered during sample login or the actual volume can be entered after the analysis has been
completed when editing result parameters.
The Nitrogen Factor (NF) can be defined in the Method, during sample login or after the analysis has been
completed when editing result parameters. The standard Std is entered and defined on the calibration
screen.
The Energy Equivalent Factor (EE) can be defined in the Method, during sample login or after the analysis
has been completed when editing result parameters.
18.416 The energy associated with the formation of Nitric Acid (60kJ/mol / 14.0067 = 4.284kJ/g =
18.416 wt% Btu/lb)
The sample weight and %N can be entered during sample login or after the analysis has been completed
when editing result parameters.
18.416 The energy associated with the formation of Nitric Acid (60kJ/mol / 14.0067 = 4.284kJ/g =
18.416 wt% Btu/lb)
The concentration for the solution use is defined in System Options C1 = Normality_N. The volume V1 =
Nitrogen can be defined in the Method, entered during sample login or after the analysis has been
completed when editing result parameters.
When V1(Nitrogen) = 0.00 no correction is calculated and e1 is forced
to be zero.
Nitrogen Units (NF and EE)
e1 = same as ASTM
GB Mode
18.416 The energy associated with the formation of Nitric Acid (60kJ/mol / 14.0067 = 4.284kJ/g =
18.416 wt% Btu/lb)
Although no editable parameters are used in this calculation the nitrogen value should be set to some
value (e.g. 1.0) to enable this correction. The value for nitrogen can be defined in the Method, entered
during sample login or after the analysis has been completed when editing result parameters.
When Nitrogen = 0.00, no correction is calculated and e1 is forced to
be zero.
Fuse Length
e2 = F x FE[Btuglb ] cm-1
-1
e2 The correction for the heat of combustion of the ignition wire in Btuglb -1
The fuse length and fuse energy can be defined in System Options, page 416, and the final fuse length
can be entered/edited during sample login or after the analysis has been completed when editing result
parameters
0.0090
Btu
cm
X
453.59 g
lb
= 4.1 ( Btu
cm
) ( g
lb
)
ASTM Mode
39.6 The energy associated with the formation of sulfuric Acid (295kJ/mol/32.064=9.20kJ/g =
39.6 wt% Btu/lb).
The sample weight and %S can be defined in the Method entered during sample login or after the analysis
has been completed when editing result parameters.
The sample weight and %S can be defined in the Method, entered during sample login or after the
analysis has been completed when editing result parameters.
The concentration for solutions use are defined in System Options C1 = Normality_N and C2 =
Normality_S. The volumes V1 = Nitrogen and V2 = Sulfur can be defined in the Method entered during
sample login or after the analysis has been completed when editing result parameters.
When V2(Sulfur) = 0.00 no correction is calculated and e3 is forced to
be zero.
The sample weight and %S can be defined in the Method, entered during sample login or after the
analysis has been completed when editing result parameters.
GB Mode
The concentration for the solution use is defined in System Options C2 = Normality_S. The volume V2 =
Sulfur can be defined in the Method then entered during sample login or after the analysis has been
completed when editing result parameters.
The sample weight and %S can be defined in the Method then entered during sample login or after the
analysis has been completed when editing result parameters.
When V2(Sulfur) = 0.00 no correction is calculated and e3 is forced to
be zero.
E= [ mH+e1+e2
t
]
E Calorimeter energy equivalent (Btu/oC) (g/lb)
H Heat of combustion of benzoic acid (Btu/lb)
m Mass of benzoic acid (g)
e1 Nitrogen titration collection (Btug)/lb
e2 Fuse collection (Btug)/lb
t Temperature change (oC)
Qraw = (tE)
ASTM Mode
Hydrogen
Har The known weight percent of Hydrogen As-Received basis in the sample
9 The approximation for the ratio of the molecular weight of water H2O and diatomic
hydrogen H2
The default Hydrogen Factor HF = 10.3 x 9 = 92.7[Btulb-1]. The % Har and Hydrogen Factor (HF) can be
defined in the Method then entered during sample login or after the analysis has been completed when
editing result parameters.
0.1119 The ration of the molecular weight diatomic Hydrogen H2 and water H2O
DIN Mode
Hydrogen
Qnet = Qgr - 24.41(Had x 8.936)[J/g] = Qgr - (HF x Had)[Btulb-1]
Qnet The net calorific value in Btu/lb
Had The known weight percent of Hydrogen As-Determined basis in the sample
8.936 The ratio of the molecular weight of water H2O and diatomic Hydrogen H2
The default Hydrogen Factor HF = 24.41 x 8.936 = 218.13[J/g] or 93.78[Btulb -1]. The %Had and
Hydrogen Factor (HF) can be defined in the Method then entered during sample login or after the analysis
has been completed when editing result parameters.
Hydrogen
The %H, M and Mt can be defined in the Method then entered during sample login or after the analysis has
been completed when editing result parameters.
As-Received Basis
Qar = Qad X 100 - Mar
100 - Mad
Dry Basis
Qd = Qad X 100
100 - Mad
Ash-Free Basis
Qd = Qad X 100
100 - Aad
ar As-Received Basis
ad As-Determined Basis
d Dry Basis
af Ash-Free Basis
DIN Mode
DIN is when Q = Qnet only.
As-Received Basis
Qar = [Qad + k(Mad) X 100 - Mar - k(Mar)
100 - Mad
Dry Basis
Qd = [Qad + k(Mad)] X 100
100 - Mad
ar As-Received Basis
ad As-Determined Basis
d Dry Basis
af Ash-Free Basis
ASTM Mode
Given:
Qraw = 11,597[Btu]
Sample Mass = 0.9996[g]
Fuse Length = 6.0[cm]F
Fuse Btu = 4.1[Btucm-1]FE
ml of Na2CO3 = 2.03[ml]V
% Nitrogen = 1.12[%]N
% Sulfur=0.68[%] = S
% Hydrogen=4.24[%] = H
% Total Moisture=9.2[%] = Mar
% Moisture=2.16[%] = Mad
% Ash=14.8[%] = Aad
Hydrogen Factor = 92.7
As-Received Basis
100 Mar 100 Mar
Qar = Qad X = 11,091 x = 10,293[Btuib-1]
100 - Mad 100 - Mad
DIN Mode
Given:
Qraw =11,597[Btu]
Sample Mass =0.9996[g]
Fuse Length =6.0[cm] =F
Fuse Btu =4.1[Btucm-1] =FE
ml of Hydrochloric Acid =12.01[ml] =V1
ml of Barium Hydroxide =7.01[ml] =V2
% Nitrogen =1.12[%] =N
% Sulfur =0.68[%] =S
% Hydrogen =4.24[%] =H
% Total Moisture =9.2[%] =Mar
% Moisture =2.16[%] =Mad
% Ash =14.8[%] =Aad
Hydrogen Factor =92.7
Concentration of Hydrochloric Acid Solution =0.1N =C1
Concentration of Barium Hydroxide Solution =0.1N =C2
As-Received Basis
100 - Mar
Qar = [Qad + k(Mad) X - k(Mar) = 10,254[Btulb-1]
100 - Mad
GB Mode
Given:
Qraw =11,597[Btu]
Sample Mass =0.9996[g]
Fuse Length =6.0[cm] =F
Fuse Btu =4.1[Btucm-1] =FE
ml of Hydrochloric Acid =12.01[ml] =V1
ml of Barium Hydroxide =7.01[ml] =V2
% Nitrogen =1.12[%] =N
% Sulfur =0.68[%] =S
% Hydrogen =4.24[%] =H
% Total Moisture =9.2[%] =Mar
% Moisture =2.16[%] =Mad
% Ash =14.8[%] =Aad
Hydrogen Factor =92.7
Concentration of Hydrochloric Acid Solution =0.1N =C1
Concentration of Sodium Hydroxide Solution =0.1N =C2
As-Received Basis
Pump Purge
Valve
Jacket Drain Reservoir
Valve Drain Valve
Particle
Filter
Lid
Figure 7-1
Water Flow Diagram
Diagnostics AC500 81
This page intentionally left blank.
82 AC500 Diagnostics
Monitoring Ambient Chart
3. Click the right mouse button, with the mouse pointer inside the
graph, to reset, restore, zoom in, zoom out, and set the graph
range. The parameter value is displayed on the vertical axis and
the time is displayed on the horizontal axis. The box in the lower
left corner of the graph displays the hours since the instrument
was turned on or the graph reset. The hours reset to 0 after 24.
4. Click Reset Chart to reset the maximum and minimum parameter
values in the chart and time axis on the graph.
5. Click Print to print the parameter values on the system printer.
Diagnostics AC500 83
Ambient Nominal Values
84 AC500 Diagnostics
Solenoids and Switches
2. Click the Check Box next to the desired solenoid valve or switch
to activate or deactivate it. A check mark in the box indicates its
activated.
3. Click Print to print the solenoid and switch states on the system
printer.
Diagnostics AC500 85
Backup (Hardware) Calibration
86 AC500 Diagnostics
3. Enter the file name and folder location in the screen.
4. Click Save to save the backup file.
Diagnostics AC500 87
3. Select the folder and backup file from the screen.
4. Click Open to load the calibration values from the selected backup
file.
88 AC500 Diagnostics
Hardware Calibration
2. Edit the Bucket Gain and Offset or the Jacket Gain, Offset and
Calibration values by clicking the box to the right of the
parameter and entering a new value.
3. After the desired values are entered close the diagnostics screen.
4. Click Print to print the calibration values on the system printer.
Diagnostics AC500 89
Preamplifier Calibration
3. Open the combustion vessel bucket lid and remove the four
screws securing the cover. Hold the cover while the screws are
being removed to prevent it from falling. Refer to Figure 8-1,
page 815.
4. Remove the jacket inlet cap from the jacket inlet and set it aside.
5. Remove the bucket thermistor and insert it into the jacket inlet.
Tighten the thermistor finger tight.
8. Click Yes to turn the water flow OFF. In either case the calibration
will be saved.
Yes must be pressed to turn the water pump off and remove the
bucket thermistor. If No is pressed and the thermistor is removed
water will rush out of the calorimeter.
2. Click the Pause Check Box to stop the data scrolling in the
communications data window.
3. Click the Hide Analysis State check box to stop the analysis status
messages from being displayed.
4. Click the Hide Outgoing check box to stop outgoing messages
from the computer from being displayed.
5. Click the Hide Incoming check box to stop incoming messages
from the computer from being displayed.
If Hide Analysis State, Hide Outgoing and Hide Incoming are all
checked only network and error messages will be displayed.
6. To search for data or messages enter search data into the edit
box and click Find. Communications will pause during the search.
7. Click Copy to Clipboard to save the communications data scrolling
on the screen in the computer clipboard. The data can be pasted
from the clipboard to another file for analysis.
8. Click Close to exit diagnostics.
Figure 8-1
Thermistor Location
Service AC500 91
Illustrations
92 AC500 Service
Installing LECO Software
The following section explains the required settings, below, and the
installation procedure, page 95, for LECO software. Additional
information can be found in Configuring the Ethernet Adaptor,
page 912, Configuring Firewall Settings, page 921, and Registering
the Software, page 922.
Service AC500 93
Data File Permissions
To use LECO software, read-write access is required for all the data
files. The proper permissions are automatically set up during the
software installation; however, if these permissions are ever changed,
file access errors can occur. The following table shows the data folders
that require read-write permissions.
Antivirus Software
LECO recommends that antivirus checking be turned off
for the files listed below to avoid interference with data
collection or storage.
94 AC500 Service
Control Panel Settings
Do not install the Copy Protection Key until after the software
has been installed.
Service AC500 95
9. Select Next. The following message box will appear to indicate
Adobe Reader is preparing for installation.
10. Determine the folder location for Adobe Reader. Select Next for
the default location, or select Browse, select a different location,
and then select Next.
96 AC500 Service
11. When the "Ready to Install Program" dialog box appears, click
Install.
12. When the "Setup Completed" dialog box appears, click Finish.
Service AC500 97
13. If the instrument software Welcome dialog box does not appear
A. Select the Start button in the lower left-hand corner.
B. Select Run. The Run dialog box is displayed.
C. Enter D:\Setup.exe and then select OK. (The letter D
indicates the location of the CD Rom drive. Typically this
drive is D, but the letter that is entered should correspond
with the CD-ROM drive on the computer's hard drive.)
14. At the instrument software Welcome dialog box, click Next.
98 AC500 Service
15. Determine the folder location for the software. Select Next for the
default location, or select Browse, select a different location, and
then select Next.
Service AC500 99
17. Once the software has installed successfully, select Finish. The
icon for the instrument software will appear on the desktop.
Windows 7
9. Click OK.
Windows Vista
The following steps apply when using the Windows Vista operating
system.
Windows XP
6. Select the Internet Protocol (TCP/IP) check box. The other check
boxes should be cleared.
7. Click on Internet Protocol (TCP/IP) and select Properties. The
TCP/IP Properties dialog box will appear.
The following steps apply only when a firewall program other than
Microsoft Windows is used.
Do not insert the copy protection key into the USB port until
the LECO software is installed and the PC has been
rebooted.
4. Click Copy.
5. Open a blank email message and press Ctrl and V on the
keyboard to paste the license information into the email message.
6. Send the email or just the Key number to the LECO designee.
The Copy Protection Key Not Found or License Not Valid dialog box
may appear if the copy protection key is not installed, the software is
not registered, or the software is already running. Follow the steps
below.
Verify the copy protection key is installed. If not,
1. Click OK, and then insert the copy protection key.
2. On the Licenses dialog box, click Close. The software login will
appear.
If the copy protection key is installed, the software may not be
registered.
1. Click OK.
2. Insert the copy protection key. The Licenses dialog box will
appear.
3. Click on the old version and then click Change. Follow the
instructions in Registering the Software, beginning with step 5.
1. Use the o-ring removal tool and pry out the two o-rings. Discard
the removed o-rings. It may be necessary to heat the end of the
o-ring removal tool and bend the end at a right angle to access
both o-rings. Refer to Figure 9-1 below.
762-515
O-RING REMOVAL
TOOL
760-224
O-RING
(2 REQ'D)
Figure 9-1
Combustion Vessel Charger O-ring Replacement
Problem Solution
The bucket water temperature does not Position the Preamplifier Calibration
rise even though the sample has ignited. switch in run position. Refer to Hardware
Calibration, page 89.
Dirt in pipet. Particle filter should be replaced. Refer to
Particle Filter Replacement, page 611.
Pipet fills too slow (longer than 2 Particle filter should be replaced. Refer to
minutes). Particle Filter Replacement, page 611.
Air bubbles in the pipet. Low water Add more water to the water return tank
alarm. until it settles slightly over the top of the
filter holder. This depth is only correct
when there has been 4 liters of water
pumped from the water return tank
(2000 ml in the bucket and 2000 ml in
the pipet).
Water too low in water return tank. Pump Water return filter is plugged and not
is sucking air. letting the water drain through fast
enough. Replace the water return filter.
Refer to Water Return Filter
Replacement, page 612.
Water is too low in the water return tank Add more water to the water return tank
and water return tank filter is okay. until it settles slightly over the top of the
filter holder. This depth is only correct
when there has been 4 liters of water
pumped from the water return tank
(2000 ml in the buck and 2000 ml in the
pipet).
Pipet valve leaks water. Nut on rear of pipet valve is too loose,
tighten nut. Always turn pipet valve
clockwise.
Water overflows from the water return Water return filter is plugged. Replace
tank when the pipet valve is off. the water return filter. Refer to Water
Return Filter Replacement, page 612.
The calorimeter displays an open fuse The fuse wire is not making good contact
alarm. with the electrodes. Clean the electrodes
with an abrasive tool or cleaning
solution.
608-876
CLAW
(2 REQ'D)
608-665
PIPET
TANK
AC500
Figure 10-1
AC500 Exploded View #1
608-666 193-110
BLOCK WASHER
(2 REQ'D)
193-154 190-332
621-207 WASHER SCREW
KEYBOARD (2 REQ'D) (2 REQ'D)
COVER 621-240
CALORIMETER 608-667
190-323
SHROUD HOLDER BASE
SCREW
(2 REQ'D)
103
608-751
HOLDER
502-296
WATER
RETURN FILTER
608-750
O-RING
608-749
SCREEN 608-770
RESERVOIR
617-294 COVER
BUCKET 608-747
TANK 608-777
SPACER
(2 REQ'D)
190-366
SCREW 789-066
(2 REQ'D) HINGE
(2 REQ'D)
608-735 608-748
STRAIN STANDOFF 194-056
RELIEF PLATE (3 REQ'D) NUT
608-734
(4 REQ'D)
PROBE
J217 HARNESS
190-267
SCREW
(4 REQ'D)
617-236
COMBUSTION 617-237 190-269
VESSEL RESERVOIR TANK SCREW
TANK ASSEMBLY (5 REQ'D)
ASSEMBLY
193-185
WASHER
(5 REQ'D)
190-323
SCREW
(8 REQ'D)
193-108
WASHER
(8 REQ'D) 194-056
NUT
(2 REQ'D)
618-542
BACK PANEL
621-250 608-776
THREAD HINGE
IGNITER BRACKET
ASSEMBLY
Figure 10-2
AC500 Exploded View #2
190-267
SCREW
(4 REQ'D)
190-267
SCREW
608-738 (4 REQ'D)
STIRRER
608-177 193-185
LID MECHANISM WASHER
ASSEMBLY (4 REQ'D)
190-269 621-210
SCREW DISPLAY BOX 190-883
(8 REQ'D) ASSEMBLY SCREW
619-818 (2 REQ'D) 608-895
193-185 BRACKET BRACKET
WASHER 193-127
(8 REQ'D) 190-269 WASHER
SCREW 190-527
(2 REQ'D) SCREW
(2 REQ'D)
(4 REQ'D)
193-185 608-720
WASHER MANIFOLD
(2 REQ'D) ASSEMBLY
608-717
DRAIN
ASSEMBLY
608-754 194-173
190-269 BRACKET
NUT
SCREW (4 REQ'D)
(2 REQ'D)
193-185
WASHER
(2 REQ'D) 193-108 194-057
621-212 WASHER NUT
POWER SUPPLY (4 REQ'D) (4 REQ'D)
ASSEMBLY 193-185
WASHER
(2 REQ'D)
190-269
SCREW
621-250 (2 REQ'D)
THREAD 618-547
IGNITER HEAT EXCHANGER
ASSEMBLY ASSEMBLY
789-419
PUMP
ASSEMBLY
619-422
BASE
780-970
FOOT
(4 REQ'D)
Figure 10-3
AC500 Exploded View #3
RUN
CAL.2
CAL.1
CALIBRATION
SWITCH
ELECTRONICS
CHASSIS
621-212
POWER
J217 SUPPLY
FUSE LEAD
CONNECTOR
775-604 775-604
PUMP PURGE JACKET
VALVE TANK DRAIN
606-202 VALVE
FAN
618-547
HEAT
EXCHANGER
608-722
789-419 FILL
WATER PUMP SOLENOID
ASSEMBLY
608-720
775-604 MANIFOLD
RESERVOIR
TANK
DRAIN VALVE
778-246
RESTRICTOR
603-065
776-465 PRESSURE
FAN FILTER SWITCH
608-722
RELIEF
SOLENOOID
Figure 10-4
Rear View AC500
Figure 10-5
Combustion Vessel 621-605-878
Figure 10-6
Combustion Vessel 621-251
608-732
INSULATION
608-736
INNER JACKET
602-788
PLUG
608-728
TUBING
608-727
FITTING
620-816
OVAL TANK
ASSEMBLY
Figure 10-7
Combustion Vessel Tank 620-817
767-629
EXTENSION
(4 REQ'D)
777-767-110
EXTENDED
194-055 POWER CARD
NUT
(4 REQ'D)
608-786
FAN
194-055
602-581
NUT STANDOFF
(4 REQ'D) (2 REQ'D)
777-904
190-227 THERMISTOR
CARD 190-237
SCREW
SCREW
(4 REQ'D) 190-269 (4 REQ'D)
SCREW
190-227
SCREW
(2 REQ'D)
608-861
NUT
(2 REQ'D)
608-759
COVER 608-932
621-208 KNOB
194-055 BOX
666-001
NUT
8 POSITION
(2 REQ'D)
666-339-271 SOLENOID
608-790 PROM CARD CARD WITH
SPEAKER 666-046 WITH 602-581
190-229 FAN CONTROL 602-581 STANDOFF
SCREW CARD WITH STANDOFF (4 REQ'D)
(2 REQ'D) 602-581 (4 REQ'D)
STANDOFF
(4 REQ'D)
Figure 10-8
Electronic Assembly 621-210
608-719
FITTING
(2 REQ'D)
Figure 10-9
Drain Bracket 608-717
620-627
608-713
HEAT
FITTING
EXCHANGER
608-714
FITTING
608-715
FITTING
606-202
FAN
615-583
FAN GUARD
(NOT SHOWN)
193-158
WASHER
(4 REQ'D)
190-295 608-716
SCREW FITTING
(4 REQ'D)
Figure 10-10
Heat Exchanger 618-547
193-178
193-107 WASHER
190-463 WASHER (2 REQ'D)
SCREW (2 REQ'D)
193-178
(2 REQ'D)
WASHER
(2 REQ'D)
608-684
BLOCK 608-679
ROD
608-685
SHAFT 190-877
(2 REQ'D) SCREW
(2 REQ'D)
608-681
BEARING
(2 REQ'D)
193-177
WASHER
(2 REQ'D)
608-682
RETAINING RING
(2 REQ'D)
515-151
GROMMET
(2 REQ'D)
608-676
190-268
GUIDE
SCREW
(2 REQ'D)
608-683 (2 REQ'D)
193-110
BRACKET
WASHER
(6 REQ'D)
193-137
WASHER 190-713
(6 REQ'D) SCREW
(6 REQ'D)
608-677
BRACE
190-930
SCREW
(8 REQ'D)
603-215
PILLOW BLOCK
(2 REQ'D)
193-138
WASHER
(2 REQ'D)
190-930
SCREW
(2 REQ'D)
608-176
GUSSET
CASTING
Figure 10-11
Lid Assembly 608-177
190-312
SCREW
(12 REQ'D)
608-695
PLATE
190-433
SCREW
(4 REQ'D)
608-992
WEATHER
STRIPPING
608-694
O-RING
762-357
O-RING
589-551
608-693
O-RING
LID
Figure 10-12
Lid Plate 608-893
608-722
SOLENOID
WITH O-RINGS
(2 REQ'D)
603-065
SWITCH
775-333
FITTING
ADAPTOR
775-611
PLUG
Figure 10-13
Manifold Assembly 608-720
J201
604-708
FAN
J1
J215
J210 J2
J216
Figure 10-14
Power Supply Chassis 621-212
608-618
190-933
BRACKET
SCREW 194-056
(NOT VISIBLE) NUT
608-605 (4 REQ'D)
CAPACITOR
190-268
SCREW
(2 REQ'D)
194-056 193-127
NUT WASHER
(4 REQ'D) (2 REQ'D)
608-609
TRANSFORMER 602-581
STANDOFF
608-606 (4 REQ'D)
BRACKET
193-185 777-887
WASHER INTERFACE
190-930
(2 REQ'D) CARD
SCREW
(2 REQ'D)
608-608
BRACKET
193-127 190-268
WASHER 589-275 SCREW
(2 REQ'D) TERMINAL (2 REQ'D)
BLOCK
193-185
194-055 802-340
WASHER
NUT INSULATOR
(2 REQ'D) BOARD
(4 REQ'D)
(2 REQ'D)
604-708 767-571
FAN STANDOFF
806-547 (2 REQ'D)
CIRCUIT
BREAKER
(WITH NUT)
190-269
190-355 SCREW
SCREW (3 REQ'D)
(4 REQ'D) 193-158
WASHER
900-537
(4 REQ'D)
STANDOFF
(4 REQ'D)
618-545
BOX 608-604
POWER SUPPLY
190-293 194-055
SCREW NUT
193-158 (4 REQ'D)
(4 REQ'D) WASHER
(4 REQ'D)
190-925
SCREW
608-607 193-127 (4 REQ'D)
POWER ENTRY WASHER
SWITCH (4 REQ'D)
190-269
SCREW
(4 REQ'D)
Figure 10-15
Power Supply Chassis Exploded View
775-604
VALVE
782-396
FITTING 608-710
FITTING
Figure 10-16
Pump Assembly 789-419
776-021
FITTING
760-285
FITTING
608-878
FLOAT SWITCH
EXTENSION
617-235
RESERVOIR TANK
Figure 10-17
Reservoir Tank 617-237
F O
Field Delimiter: Determines how the Origin: The intersection of the X and
transmitted data will be separated. Y axes. On the calibration curve this is
zero concentration and zero signal.
Flow Control: Selects the means to
control the data flow from one device Orphan: The database keeps track of
to another to ensure the receiving data objects such as samples,
device can handle all of the incoming calibrations, standards, methods, etc.
data. that are associated with each other.
Orphan database objects are created
when the connection between data
I objects is broken due to a rare
Instance: A unique number assigned occasion such as when a sample is
to an electronic assembly that shows deleted and the software does not
how many variations of each electronic perform the save correctly.
assembly are present in the
instrument.
P
Parity: The quality of being odd or
L even. Parity is used to ensure the
LCL: Lower Control Limit validity of the data.
Pause: Select to stop or delay an
analysis and then resume the analysis
M manually or at a predetermined time.
MAC- ID: The network ID number pdf: portable document format
assigned to the electronic assembly
connected to the network. This Pneumatic Pressure: The pressure
number is assigned at power up and of the pneumatic gas (compressed
can vary. air). The value is in psi (pounds per
square inch).
Method: A set of analysis parameter
values specified for an analysis. Poll Devices: Determines which
electronic assemblies are signed on to
the network.
U
UCL: Upper Control Limit
Uncertainty: The range of values
within which the true value is
estimated to be. It is the best estimate
of the possible inaccuracy due to both
random and systematic error. (Taylor,
John Keenan. Statistical Techniques
for Data Analysis. Lewis Publishers,
Inc.: Chelsea, Michigan. 1990).
V
VDC: Volts Direct Current. Current
flows in a single direction.
Version: The version number of the
application software located on the
electronic assembly. This software can
be upgraded from the network
procedure.
B
Balance
Balance Configuration Definitions ................................................440
Configuration ............................................................................439
Control Panel ............................................................................34
Entering a Mass ........................................................................55
Entering a Sample Weight ..........................................................36
Installation ...............................................................................33
Resetting .................................................................................36
Setting the Data Characters Parameters.......................................34
Setting Up Communication with Balance ......................................35
Testing ....................................................................................36
C
Cabinet Schematic .........................................................................111
Calculations and Corrections ............................................................76
Calibration ....................................................................................524
Backup Calibration Values ..........................................................86
Backup (Hardware) ...................................................................86
Restore Calibration Values ..........................................................87
Standard Calibration ..................................................................524
Clone Button .................................................................................412
Combustion Vessel
Bucket Preparation ....................................................................515
D
Data
Backup ....................................................................................551
Exporting .................................................................................539
Importing .................................................................................542
Transmitting .............................................................................543
Database Menu
Archive Log .................................................................................470
Data Transmit Installation ...............................................................216
Declaration of Conformity ................................................................117
Defining Calibration Standards .........................................................442
Deleting a Sample ..........................................................................526
Deleting a Sample Template ............................................................445
Determining User Permissions..........................................................432
Diagnostics Menu ...........................................................................411
E
Edit Menu ......................................................................................48
F
Facility Requirements (Illustration) ...................................................27
Fan Filter Removal and Cleaning ......................................................610
Functions ......................................................................................452
H
Hardware Calibration ......................................................................89
Edit Preamplifier ..........................................................................89
Peamplifier Calibration ..................................................................810
Thermistor Calibration ..................................................................812
Heat Correction ..............................................................................518
Hydrostatic and Proof Testing ..........................................................135
I
Igniter Box Schematic.....................................................................112
Illustrations List .............................................................................16
Importing Data ..............................................................................542
Indicators ......................................................................................44
Inserting a Sample out of Sequence .................................................525
Instrument
Installation ...............................................................................23
Rear View (Illustration) ..............................................................106
Interference to Other Devices ..........................................................111
International Installation .................................................................214
J
Jacket
Assembly (Illustration)...............................................................109
Thermistor (Illustration) .............................................................812
L
Log Bar .........................................................................................49
Logging in a Sample .......................................................................55
Logging in Using a Calibration Standard ............................................443
Logging in Periodic Maintenance .......................................................67
M
Maintenance Menu .........................................................................412
Maintenance Schedule ....................................................................63
Maintenance Counters ....................................................................64
Modifying Periodic Maintenance Counter .......................................65
Managing Databases .......................................................................454
Manifold Assembly (Illustration) .......................................................1015
Methods ........................................................................................517
Creating ...................................................................................517
Importing .................................................................................542
Monitoring Ambient Chart...........................................................83
Menu Bar ......................................................................................48
N
Needle Valve (Troubleshooting) .......................................................69
Number Formatting ........................................................................453
O
Optional Equipment ........................................................................125
O-ring Pack ...................................................................................128
Oxygen Connection ........................................................................24
P
Particle Filter Replacement ..............................................................611
Periodic Maintenance Schedule ........................................................63
Pipet Installation ............................................................................24
Power Cord Wiring Diagram .............................................................29
Preamplifier Calibration ...................................................................810
Preparing Crucibles.........................................................................57
Preparing Samples .........................................................................58
Printer Installation..........................................................................37
S
Safety Notice .................................................................................131
Sample
Analyzing .................................................................................521
Applying Sample Signature .........................................................528
Creating Sample Templates ........................................................ 421
Filters ......................................................................................447
Logging....................................................................................55
Menu .......................................................................................410
Preparation ..............................................................................58
Selecting a Language ......................................................................436
Serial Port Configuration .................................................................440
Setup, Balance...............................................................................36
Smartline Installation .....................................................................38
Software
Installing ..................................................................................93
Registering ...............................................................................922
Solenoids and Switches ...................................................................85
Specifications ................................................................................129
Status Bar .....................................................................................49
Standard Calibration .......................................................................524
T
Text Commands Definition ..............................................................440
Theory of Operation........................................................................75
Thermistor Calibration ....................................................................812
Thermistor (Illustration) ..................................................................815
Thread Vessel Accessory Pack ..........................................................128
Toolbar .........................................................................................414
Transferring Export ........................................................................551
Transferring Import ........................................................................552
Transmit Format Configuration.........................................................543
Transmitting Data ..........................................................................543
Troubleshooting .............................................................................930
U
User Defined Fields.........................................................................451
Using Toolbars ...............................................................................414
W
Warranty.......................................................................................17
Water
Water System Filling ..................................................................615
Water System Draining ..............................................................614
Water Return Filter Replacement .................................................612
Pump Assembly (Illustration) ......................................................1018
Manifold Assembly (Illustration) ..................................................1015
Water Flow Diagram .......................................................................722
Wire Vessel Accessory Pack .............................................................128
Water Return Tank Installation ........................................................25