Beruflich Dokumente
Kultur Dokumente
Chris Magruder
72 d
egre
es
Advances in PA Scan Plan Designs Beam Resolution
The beam resolution will determine the number of A-scans in the compound
S-scan.
This parameter significantly affects flaw resolution, file size, and scanner speed.
72 d
egre
es
Advances in PA Scan Plan Designs Beam Resolution cont.
In the example below, all compound S-scans use the same beam aperture and
the same first and last A-scan angle and beam exit position in the wedge.
They only differ in beam resolution, which greatly affects acoustic results.
In addition to the S-scan beam resolution, results are also affected by the
proximity of the probe to the weld (beam spread) and probe frequency.
Advances in PA Scan Plan Designs Beam Resolution cont.
The beam resolution will determine the # of A-scans in the compound S-scan and
the beam-to-beam distance on the weld bevel that affects flaw height sizing.
Flaw resolution on the weld bevel is a function of many factors, including the
sound path distance from the probe to the weld, probe frequency, aperture, focus,
and beam orientation.
Advances in PA Scan Plan Designs ASME Compliance
The compound S-scan scan plan is defined for the PA calculator or work
procedure by populating the parameters of both the linear E-scan and S-scan
in ASME V Art 4 Appendix V Table V-421.
PWZ1 5L60
32mm
60 \ 10 \ .1
55 shear
40-72
1-29
1
72 d
egre 40-72
es .5 deg
32 (1-29)
Advances in PA Scan Plan Designs Calibration and Work Procedure
The compound S-scan is compatible with traditional phased array sensitivity
and time-of-flight (TOF) calibrations.
With regard to UT settings, gate position, C-scan configuration, and other
functions, the compound S-scan does not require any special accommodation
and is compatible with normal work procedure.
Advances in PA Scan Plan Designs Compound S-scan vs S-scan
Although a two group S-scan strategy from the front and back of the probe is
possible with any 32 or 64 element probe, use of a single compound S-scan
results in less setup and calibration time, more inspection coverage, and faster
analysis.
10 mm
15 mm
20 mm
Advances in PA Scan Plan Designs Data Sample 9% Ni LNG Shell
Ability to use one setup configuration for a range of thicknesses.
Thick section data sample.
35 mm 25 mm
Advances in PA Scan Plan Designs Data Sample ASME B31.3
25 mm double V weld bevel coverage comparison of compound and standard
S-scans with a 1 mm pitch, 60 element probe.
Compound S-scan Standard S-scan 2X Standard S-scan
Advances in PA Scan Plan Designs Data Sample ASME B31.3 cont.
Flaw detection, sizing, and characterization meet and exceed the needs of
ASME B31.3, I, V, VIII, and similar referencing codes.