Beruflich Dokumente
Kultur Dokumente
Alecia Eliason
February 17, 2017
Lucite Block Tray Attenuation
Purpose: Any device placed in the path of a radiation beam changes the dose delivered to the
patient via attenuation.1 This beam interference must be accounted for with a transmission factor
(Cattn) when performing monitor unit calculations in order to accurately calculate radiation dose.
A transmission factor is a ratio of dose with a device in the path of a radiation beam to the dose
without the device in the beam, as demonstrated here1:
(1)
Specifically, the lucite block trays upon which cerrobend field-shaping blocks are
mounted are inserted into the head of a linear accelerator directly in the radiation beams path,
consequently attenuating the beam. Therefore, using Equation 1, the block tray transmission
factor can be calculated by using the following formula:
(2)
Methods and Materials: An Elekta Infinity linear accelerator was used as the radiation source
for data collection. Both 6 MV photons and 18 MV photons were employed at an SSD of 100
cm, with outputs recorded at a depth of dmax specific to beam energy (1.5 cm and 3 cm,
respectively) as well as at 10 cm. 3 symmetric field sizes (10x10 cm, 20x20 cm and 30x30 cm)
were also tested for each energy and depth mentioned previously so to report an accurate
transmission factor while accounting for different field sizes. A PTW ion chamber was placed at
Eliason 2
measurement depth in solid water, and outputs were read on a Standard Imaging Supermax
electrometer. 100 monitor units (MU) were delivered per measurement. Measurements were
recorded with and without a 3/8 inch lucite block tray inserted into the head of the linear
accelerator.
Figure 1. Ion chamber measuring dose at 10 cm depth in solid water with 100 SSD.
Results: Measurements (nC) with and without the lucite block tray in place were recorded in
Tables 1 and 2, for 6 MV photons and 18 MV photons respectively. Each measurement displayed
in Tables 1 and 2 is an average of 3 measurements recorded per setting.
The block tray factor for each setting was calculated using Equation 2 and is displayed in
Table 3.
6 MV 18 MV 6 MV 18 MV 6 MV 18 MV
10x10 cm 10x10 cm 20x20 cm 20x20 cm 30x30 cm 30x30 cm
1.5 cm 0.95 0.96 0.97
depth
Eliason 4
Discussion: The results displayed in Table 3 show that the block tray factor increases slightly as
field size increases. Scatter increases as field size increases2; therefore, more dose (scatter +
primary) reaches the ion chamber with larger field sizes, resulting in a larger block tray factor.
Results also reveal a slightly larger block tray factor with higher energy. Higher energy photon
beams have greater penetrating power, while lower energy beams have less. Higher energy
beams will then be less attenuated by a lucite block tray, consequently leading to a higher tray
factor. Overall, the results verify that a 3/8 inch lucite block tray attenuates approximately 3% of
the incident radiation beam.
(3)
Dose per field(cGy)
MU=
cGy
Dose rate ( )
MU
x Scp x TMR x Cattn x ISF
Since the linear accelerator used for this project was calibrated at 100 SAD to a depth of
dmax (1 cGy/MU) and the blocked equivalent square for the example field is just less than 20
cm, the tray factor used (per measurements previously documented) for the following
calculations will be 0.97.
Eliason 5
The following monitor unit calculation demonstrates the use of the block tray factor. Also
demonstrated is the same calculation without the block tray factor.
Figure 3. Transverse slice at isocenter showing field with blocked left kidney.
Treatment parameters for another clinical example of a plan requiring custom cerrobend
blocks is demonstrated in Figure 6. Of note is the different tray factors used for different energies
and equivalent field sizes; a 6 MV photon beam with 13.02x13.02 cm equivalent square uses a
tray factor of 0.956 and a 18 MV photon beam with 13.69x13.69 cm equivalent square uses 0.97,
both of which are consistent with my reported results.
Eliason 8
Figure 6. Treatment parameters for mantle fields with custom cerrobend blocks.
Eliason 9
Conclusion. Although not used as frequently as in the past, effective photon-beam shielding of
vital organs using cerrobend blocks instead of MLC leaves is sometimes necessary, especially for
structures requiring blocking which are in the middle of an open field. When a cerrobend block is
deemed appropriate for treatment, the dose attenuated by the lucite tray on which the block is
mounted must be accounted for via a block tray factor in dose calculations. Omitting this tray
factor in calculations when a lucite block tray is in place may lead to underdose of patient
treatment volumes. To the other extent, incorrectly adding a tray factor to a dose calculation
when a block is not used for treatment may cause patient overdose. Using appropriate factors in
dose calculations are crucial to accurate delivery of radiation per the prescription.
References:
1. Stanton R, Stinson D. Applied Physics for Radiation Oncology. Madison, WI: Medical Physics
Publishing; 1996.
2. Bentel GC. Radiation Therapy Planning. 2nd ed. New York, NY: McGraw-Hill; 1996.
3. Khan FM, Gibbons JP. Khans The Physics of Radiation Therapy. 5th ed. Philadelphia, PA:
Lippincott Williams & Wilkins; 2014.