Beruflich Dokumente
Kultur Dokumente
KARL STAPPERT
Global Business Development Manager
Emerson Process Management - Micro Motion, Inc.
7070 Winchester Circle
Boulder, Colorado 80301
PP&RULROLVPHWHULQVWDOOHGLQ
PPOLQH
Having the ability to bidirectionally measure almost any Overall, Coriolis meters offer the least amount of
gas phase fluid from -400 to +660 degrees Fahrenheit measurement and maintenance concern in their use
without concern of error caused by flow profile compared to any other gas flow technology. This paper
disturbances, compressibility and density change, will discuss the theory of operation and guide the user
pulsations, regulator noise or damage due to flow surges through selection, application, maintenance, and provide
and pulsations, Coriolis meters are becoming the fiscally application examples Coriolis meters used in gas
responsible meter of choice in many applications. measurement.
1
conduit consists of one or more tubes that are vibrated at base volume are documented in AGA Report Number
their resonant frequency by a Drive Coil. Sensing 11, Measurement of Natural Gas by Coriolis Meter and
pickoff coils located on the inlet and outlet sections of AGA Report Number 8, Compressibility Factors for
the tube(s), oscillate in proportion to the sinusoidal Natural Gas and Other Hydrocarbon Gases.
vibration.
Note that the vibration frequency of the flow tubes is High fluid pressures cause the flow tube to be more
proportional to the flowing density of the fluid. For gas resistant to the twisting force of the Coriolis Effect than
DSSOLFDWLRQV WKH IORZLQJ RU OLYH GHQVLW\ measured by they are at low pressures. As the internal flow tube(s)
the Coriolis meter is typically not used for gas pressure increases, the Coriolis Effect or twisting of the
measurement, as its potential error in relation to gas flow tube(s) observed for a given mass flow rate
densities is not acceptable for gas flow measurement decreases. Likewise, as the internal pressure decreases
purposes. Although this is the case, density measurement the Coriolis Effect observed for a given mass flow rate
in gas applications can be used as an indicator of change increases.
LQD&RULROLVPHWHUVIORZIDFWRUDQGRUFOHDQYVGLUW\ The amount of under or over registration that occurs is
relative to the pressure the meter was calibrated at or at
Coriolis is a direct inferential mass meter eliminating the which the calibration factor was determined. If Coriolis
requirement to quantify gas at flowing conditions; i.e. is applied at a pressure other than calibration pressure,
the need to measure flowing temperature, flowing measurement engineers should investigate pressure
pressure, and calculate a flowing compressibility. effect and insure appropriate compensation is
Equations and methods for the conversion of mass to implemented.
2
Every Coriolis meter design and size has a different flow advanced designs to measure up to sonic velocities
pressure effect specification. To correct for flow pressure (choke point) removes any rate-of-change concern in the
effect the flow pressure effect compensation factor use of Coriolis.
detailed below is applied to the indicated mass flow rate.
Meter Selection - Over Range
F 1
p 1 (( P / 100) * ( P P )) Over ranging often causes mechanical damage and/or
Effect Static Cal loss of measurement in the use of almost any flow meter.
Some Coriolis meter designs can measure gas flow up to
Where: sonic velocity or choke point (approximately 1400 ft/sec
with natural gas mixtures from atmospheric to 2220 psi)
Fp = Flow pressure effect compensation factor without loss of measurement or damage. ManufacturerV
PEffect = Pressure effect in percent psi should always be consulted on the maximum velocity
limit of their particular Coriolis meter design, as not all
PStatic = Measurement fluid static pressure in psi Coriolis meter designs can measure up to choke point. It
should be noted that the stability of flow rate indication
PCal = Calibration static pressure in psi
decreases when gas velocities exceed approximately 400
ft/sec through a Coriolis sensor, but high speed
Most Coriolis transmitters have provisions for applying integration of the flow rates over an extended period of
an average flow pressure effect correction or for time when this instability is present has proven the
monitoring a static pressure transmitter to facilitate live average flow rate indication to still be very accurate.
pressure compensation. A Coriolis PHWHUV SUHVVXUH
effect typically ranges from a negative 0.001 to 0.0 Piping erosion caused by high flow velocities is a
percent per a psi (Worst case, a 100 psi error in pressure common concern in gas applications. Although this
compensation would equate to a 0.1% error in mass flow concern is valid for carbon steel piping, it is not valid for
rate). For applications were static pressure varies stainless steel piping or Coriolis meters. For a gas to
significantly, typically more than 200 psi (i.e. +/- 100 psi HURGHDPHWDOWKHPHWDOVVXUIDFHPXVWILUVWR[LGL]HDQG
from average), the use of a fixed/constant pressure effect then high gas velocities can erode the soft oxide layer. In
correction could potentially induce unacceptable errors. most gas applications, moisture in the gas causes
In these applications a live static pressure measurement oxidation of carbon steel piping driving the velocity
should be used to correct for the flow pressure effect, but concerns of many piping engineers. $ &RULROLV PHWHUV
a high accuracy pressure transmitter is typically immunity to high velocity gas erosion is similar to that
unwarranted. of an orifice plate or sonic nozzle, in that they are made
of stainless steel or other nickel alloys which are highly
immune to corrosion/oxidation and thus high velocity
Meter Selection - Compressibility, Density, Viscosity, gas erosion.
and Reynolds Number
If abrasive contaminants (sand, welding rods, rocks, etc.)
Although change in compressibility, density, viscosity, are present in the gas flow stream, erosion or damage of
and Reynolds Number are a concern with almost all the wetted meter components caused by this debris
metering technologies, the inferred mass flow rate of a traveling at high flow velocities is of concern. These
curved tube Coriolis meter is insensitive to error caused concerns are application specific, but when present,
by these changes. filtration should be used to protect the meter.
High rates-of-change in flow or flow surges are the most The pulsation of flow is typically of high concern in the
common cause of damage in flow meters with rotating use of every flow metering technology except Coriolis.
flow elements; e.g. Turbine, Rotary, Positive Pulsating flow can cause measurement error (e.g. Fluidic
Displacement, etc. High rates of change in flow cause Oscillation, Orifice/Differential Head, Rotary, Turbine,
these meters to over-speed and are typically inherent to and Ultrasonic meters) and mechanical damage in
engine, boiler, and burner fuel gas applications. metering technologies with load bearings and gears (e.g.
Rotary, and Turbine meters). Flow pulsations are
During a flow surge, the inlet flow splitter of a Coriolis typically a concern on fuel gas lines to reciprocating
meter chokes flow to the diameter of the flow tubes and engines, the inlet and outlet compression lines of
the flowing densities of natural gas mixtures do not reciprocating compressors, and the inlet and outlet lines
provide enough inertial force to be imparted on the flow of regulators.
tubes to damage them. This coupled with the ability of
3
Advancements in Coriolis flow meter design have 1.5
IUHTXHQF\RIWKHPHWHUVIORZWXEHV,QJDVDSSOLFDWLRQV -0.5
Although Coriolis meters, for the most part, are immune Meter Selection - Bidirectional
to mechanical vibration, they are sensitive to vibrations
at the resonant frequency of the flow tube(s). The Coriolis meters are bidirectional meters, during flow the
resonant frequency of the flow tube(s) is meter design signal from the inlet pick-off coil lags the outlet pick-off
and fluid density dependent. Testing of a Coriolis meter coil signal, by determining which pickoff signal is
subjected to mechanical vibrations is shown in the lagging flow direction is determined.
following figure. Note that the area of sensitivity is only
DWWKHUHVRQDQWIUHTXHQF\RIWKHPHWHUVIORZWXEHV
4
Meter Selection - Measurement Accuracy minimum acceptable flow accuracy is 1%. Since flow
through the meter increases with square root of static
The measurement accuracy of a Coriolis meter is design pressure change and the minimum flow through the
and fluid specific. Most bending mode Coriolis meters meter is constant and relative to the minimum acceptable
can measure gas mixtures at accuracies better than 1%. accuracy, applying a meter at higher pressures, in effect,
Some advanced Coriolis designs can achieve accuracies increases operating range and turndown.
of +/- 0.35%.
7KH KLJK HQG RI D &RULROLV PHWHUV IORZ UDQJH LV Coriolis Turndown versus Operating Pressure
determined by flow velocity. Most Coriolis meters can
measure gas velocities up to 200 ft/sec and advanced In summary the operating range for a given pressure
designs can measure gas flow at velocities up to sonic drop can be increased by installing a Coriolis meter at
velocity or choke point without loss of measurement or high-pressure locations or upstream of regulation versus
damage. Although some Coriolis meter designs can downstream.
measure gas flows up to sonic velocity, a maximum
allowable pressure drop dictated by the application in
which they will be applied typically determines Meter Selection Low Flow
maximum flow. This is quite different from traditional
flow technologies where maximum flow is where The following equation is the most utilized method for
measurement is lost and/or flow damage occurs to the determining the minimum flow rate of a Coriolis meter.
flow element.
ZeroStabil ty
Therefore, the appropriate size of Coriolis meter for an MinFlow
application is determined by the following. Accuracy% / 100
x Allowable Pressure Drop @ Maximum Flow Since Zero Stability can be expressed in standard volume
x Minimum Acceptable Accuracy @ Minimum (scf) units for a given relative density, the minimum
Flow standard volume flow rate at a user specified acceptable
accuracy never changes regardless of pressure or
In gas applications, if the non-linearity of gas temperature for a given meter design and size. This is
compressibility is ignored and temperature is assumed to different from other gas measurement technologies were
be constant, it can be stated that the flow through a the minimum flow rate varies with pressure and
Coriolis meter for a given pressure drop changes directly temperature.
with the square root of static pressure change. With a
constant pressure drop, if static pressure doubles, flow
would increase by a multiplier of 1.4; i.e. 2 x (flow Meter Selection - Maximum flow
prior to pressure change). Likewise, if static pressure
drops in half, with pressure drop constant, flow would Advanced Coriolis meter designs can measure gas flows
decrease by a multiplier of 0.707; i.e. 0.5 x (flow prior up to choke point or a flow velocity equivalent to sonic
velocity (Mach 1) of the gas mixture (Approximately
to pressure change).
1400 ft/sec for natural gas mixtures from atmospheric to
2220 psi). Although this is the case, in most applications
An example of this relationship is shown in the
Coriolis meters are sized to operate at Mach or less
following figure, where the turndown of multiple meters
due to flow noise. Although still accurate at flow
is trended across pressure range while pressure drop
velocities above Mach, flow noise or the trembling of
across the meters is held constant at 15 psi (the typical
the flow tubes at velocities higher than Mach will
pressure drop of a gas turbine at maximum flow) and the
cause instability in the instantaneous flow indication. In
5
applications were flow rates can exceed Mach the but some manufacturers have hardware configurations
flow averaging time, also referred to as flow dampening, that can be installed below ground if the electrical
can be extended to rectify this instability in indicated enclosures and conduits are properly sealed or weather
flow rate. proofed.
Where:
'PAppGas = Maximum allowable pressure drop
across the Coriolis meter with an
application gas density ( U f AppGas ) in psi
Uf Re fGas
= Density of reference gas at flowing
conditions in lb/cf
'PRe fGas = Reference differential pressure across
Coriolis meter with reference gas density
( U f Re fGas ) in psi
Uf AppGas
= Density of application gas at flowing Installation Effects (Complex)
conditions in lb/cf
Qvf Re fGas = Volume flow rate of reference gas at
Most Coriolis meters are designed to meet Class 1, Installation Effects (Common)
Division 1, and Class 1, Division 2 hazardous area
classifications. The test data shows that flow perturbations placed in
close proximity upstream of a bent tube Coriolis meter is
not of concern and flow conditioning or straight pipe up
Installation - Above and Below Ground and downstream of a Coriolis meter is not required.
6
Installation - Meter Mounting RIWKH&RULROLVVHQVRUVPHDVXUHGWHPSHUDWXUHVKRXOGEH
located upstream of the sensor due to the Joule Thomson
Consideration should be given to the support of the effect at high differential pressures across the sensor.
sensor and the alignment of the inlet and outlet piping
flanges with the sensor. For field fabrication of piping, a
spool piece should be used in place of the meter to align
pipe-work prior to welding the Coriolis sensor mating
flanges; i.e. slip fit is ideal.
7
intended for gas measurement. Although this is the case
Ub Pb x Mr
the user should review industry recommended practices,
standards, and regulatory requirements when
Zb x R x Tb
establishing calibration policy for Coriolis. Where:
Coriolis meters are an attractive technology when the
availability, capability, or economic viability of gas SCF( gas) = Gas volume at Tb and Pb
calibrations is limited. Highly accurate water calibrations
and construction of water calibration facilities are Mass = Weight of gas (Coriolis output)
achieved at a fraction the cost of their gas counterparts. Pb
Ub = Density at Tb and
8
bring the meterV performance back to its original Sensor diagnostics also provide insight into the process
specification. At any given level of coating, if the flow conditions and potential measurement problems
coating is stable, the meter can be re-zeroed without with them.
cleaning and meter performance can be restored. But if
coating of the sensor continues, the zero may continue to Advanced Coriolis sensor designs also provide on-line
drift. 6WUXFWXUDO,QWHJULW\9HULILFDWLRQRIWKHIORZWXEHVThe
flow tube structure of a Coriolis sensor dictates its flow
Although rare in gas applications, flow tube erosion or calibration factor. The structural integrity method of
corrosion will affect a meterV zero performance and verification measures flow tube(s) stiffness to infer
likewise its calibration. A repeated out of specification density and mass calibration factor change. Utilizing
zero found during the zero verification process is a resonant modal analysis techniques the transmitter
potential indicator of erosion and corrosion. actively tests the flow tube(s) stiffness during flowing
Confirmation of the condition can be made by conditions and determines if an out of tolerance stiffness
performing a density verification; i.e. a negative out of change has occurred.
specification bias in density is caused by reduction in
flow tube wall thickness. Structural Integrity Verification is a significant
advancement in Coriolis technology. This capability
allows for verification a Coriolis PHWHUV DFFXUDF\
Inspection and Re-zeroing ZLWKRXW LQWHUUXSWLRQ LQ IORZ LQVSHFWLRQ RI WKH PHWHUV
components for cleanliness or damage.
Improper zeroing will result in measurement error. Prior
to inspection of the meter zero and re-zero the meter, if
required, thermal equilibrium RI WKH PHWHUV IORZ WXEHV CAPEX
must be achieved. To insure meter thermal equilibrium, a
flow rate above the transitional flow must be established Capital Expenditures &$3(;WRLPSOHPHQW&RULROLV
DQGWKHPHWHUVLQGLFDWHGWHPSHUDWXUHPXVWEHYHULILHGDV measurement will vary dependent upon meter design,
stable. Once thermal equilibrium is achieved the meter is size, pressure ratings, materials of construction, and
WR EH EORFNHG LQ DQG WKH PHWHUV ]HUR YHULILHG. Even accuracy. Typical capital expenditures required in the
though the stream is not flowing, the flow meter may implementation of a Coriolis metering system include
indicate a small amount of flow bias, either positive or the following.
negative. Causes for a bias in the zero are usually
related to the differences between previous and current x Coriolis Sensor
zero flow conditions, which include. x Flow Computer or Transmitter
x Power System
x Differences between the calibration media x Installation and Startup
density and the gas density
x Differences in temperature Coriolis technology reduces or eliminates several of the
x Differing mounting conditions capital expenditures required in the application of other
gas flow technologies and typically associated with
The meter should read a mass flow rate that is less than natural gas metering. Typical capital expenditures that
WKH PDQXIDFWXUHUV ]HUR VWDELOLW\ VSHFLILFDWLRQ XQGHU WKH are reduced or eliminated are as follows.
no-flow condition.
x Pressure measurement - Typically not required
If the zero is within specification re-zeroing the meter is for flow measurement and if required, high
unwarranted. If outside of specification the current zero accuracy is unwarranted
value and meter temperature should be recorded for x Temperature measurement - Integral to Coriolis
future reference and the zeroing procedure specified by sensor design
the meter manufacturer should be followed. x Specialty upstream and downstream piping
and/or flow conditioning
x Gas flow calibration of sensor - Factory water
Field Maintenance Diagnostics flow calibration transfers to natural gas
measurement
Diagnostic LED(s) and display are typically provided to x Installation and startup
indicate the operating status of the sensor and
transmitter. The diagnostics of the Coriolis transmitter
verify the integrity of the CPU and insure operational
parameters are within tolerance. Coriolis sensor
diagnostics verify the sensing components of the sensor
are not damaged and operating within normal limits.
9
OPEX x Sudden changes in gas flow velocity (fuel gas
applications)
2SHUDWLQJ ([SHQGLWXUHV 23(; WR PDLQWDLQ &RULROLV x Pulsating gas flows (fuel gas and compression
measurement will vary dependent upon meter design, gas in the use of reciprocating compressors)
power system, cleanliness of process, and accuracy x Applications were abnormally high flow rates
verification procedures required by meter design, can occur.
DGRSWHG E\ WKH XVHUV RUJDQL]DWLRQ RU dictated by
regulatory requirements. Typical operating expenditures Coriolis meters can be sized for very low-pressure drop
required in the use a Coriolis metering system include +2O), but can also be installed upstream of the
the following. pressure regulator with high pressure drops for increased
turndown without concern of damage or malfunction due
x Routine inspection of sensor zero and to flow noise. For instance, in one application for
diagnostics custody transfer of nitrogen, a 50-SVLGGURS+2O)
x The replacement of battery backup power cells, was allowed across the Coriolis meter and the pressure
if used, when their efficiency has declined. regulator adjusted accordingly. This allowed the use of a
&RULROLV PHWHU LQVWHDG RI D PHWHU GRZQVWUHDP RI
Coriolis technology reduces or eliminates several of the the regulator and a 40:1 useable turndown (Better than
operating expenditures associated with gas flow 1% accuracy at minimum flow and an average 0.35%
technologies. Typical operating expenditures that are base volume accuracy over 95% of the upper flow
reduced or eliminated are as follows. range).
Application Examples
Coriolis meters are applied in a wide variety of Coriolis Liquid and Gas Separator Measurement
DSSOLFDWLRQV IURP WKH ZHOOKHDG WR WKH EXUQHU WLS
Coriolis meters are primarily a smaller to medium line
size meter, ideally suited to the following gas metering Fuel Control: A major US vendor of gas turbines designs
VZHHWVSRWV a high-efficiency, low emissions offering. This design
utilizes a trio of Coriolis meters to measure the natural
x Line sizes 250mm (12DQGVPDOOHU gas burned in each of three combustion zones. The
x 300 ANSI through 900 ANSI combination of no damage due to flow rate-of-change at
x High turndown requirements start-up, high turndown, high accuracy, immunity to
x Dirty, wet, or sour gas where maintenance can vibration in a very high vibration environment, along
be an issue with other technologies with ease of installation due to no straight pipe run
x There is no room for long straight-runs requirement, makes Coriolis the technology of choice.
x Changing gas composition and density
10
$IWHULQVWDOODWLRQVLQFHZLWh two operating and
Coriolis Fuel Gas Measurement on a Gas Turbine RQHKRWVSDUHPHWHUIRUWXUQGRZQ
Natural Gas Fiscal Transfer: One specific example of gas Natural Gas Storage: A storage field in Hungary utilizes
measurement capability is at a natural gas utility in 27 two-inch Coriolis meters for the injection and
:HVWHUQ $XVWUDOLD 7ZR PHWHUV DUH XVHG LQ SDUDOOHO withdrawal measurement of natural gas. The storage
ZLWK D WKLUG XVHG DV D KRW VSDUH IRU PRQWKO\ reservoir consists of a multilayer sandstone formation
verifications of the transfer meters. with an aquifer flowing through it. Due to the
complexity of managing the water level in a sandstone
The justification for using the Coriolis meters was based formation on the injection and withdrawal of natural gas,
on installation and calibration/maintenance cost multiple small wells are required. The withdrawal gas is
improvements over the more traditional gas metering also fully saturated, contains H2S, and during high flows
systems. Since Coriolis meters require no straight runs the wells produce sand. In this difficult application only
or flow conditioning the installed costs were reduced by Coriolis meters can provide bidirectional measurement,
five times, even with the parallel meters required to long-term accuracy, and achieve the wide turndowns
handle the highest flows. required for reservoir management.
11
Combustion control application to boilers
Summary
12
present, energy metering is required, and/or the gas is of
dirty, sour, or changing composition.
Parallel 3&RULROLVDQG16XOWUDVRQLFLQDIXHOJDV
metering installation.
13