Instrumentation of YSZ Oxygen Sensor Calibration in Liquid Lead-Bismuth Eutectic

Xiaolong Wu1, Jian Ma2, Yingtao Jiang1, Bingmei Fu2, Wei Hang3, Jinsuo Zhang3, and Ning Li3
1. 2. 3.
Dept. of Electrical & Computer Engineering Dept. of Mechanical Engineering Los Alamos National Laboratory
University of Nevada, Las Vegas University of Nevada, Las Vegas Department of Energy
Las Vegas, NV 89154 Las Vegas, NV89154 Los Alamos, NM 87544
USA USA USA

Abstract Liquid lead-bismuth eutectic (LBE), a good is the corrosion of the structural materials employed in the
candidate for the coolant in the subcritical transmutation circuits. One way to reduce corrosion is to protect the
blanket, is known to be very corrosive to stainless steel that structural materials with a stable oxide layer [10]. Solid-
transports it. Such a corrosion problem can be prevented by state YSZ oxygen sensors have been utilized for monitoring
producing and maintaining a protective oxide layer on the dissolved oxygen in molten LBE. By controlling the
exposed surface of stainless steel. Proper formation of the oxide temperature and the concentration of oxygen dissolved in
layer critically depends on the accurate measurement and LBE, it's possible to maintain a sufficient oxide layer on the
control of the oxygen concentration in liquid LBE. An oxygen
surface of the steel containers. Oxygen required to protect
sensor calibration/measurement apparatus thus has been
designed and built to deliberately calibrate Yttria Stabilized steel surface is established by the control of the ratio
Zirconia (YSZ) oxygen sensor. A detailed description of this between H2 and O2 introduced into LBE [6] [9]. It is known
system with main components and their functions is presented. that oxygen partial pressure in the gas phase must be higher
Set of calibration curves of voltage vs. temperature ranging than that required for oxidation of iron, the main steel
from 350 oC to 550 oC under various oxygen concentrations in component, and lower than that for PbO formation. In terms
liquid LBE for the YSZ oxygen sensors have been obtained, of free energy of formation of Fe3O4 and PbO, this condition
and are presented and analyzed here. Analysis on the is given by [11]
characteristics of this YSZ oxygen sensor is also discussed.
1 / 2 F Fe 3 O 4 RT ln PO 2 2 FPbO 2 RT ln a Pb (1)
NOMENCLATURE
Furthermore, PH2O/PH2 should be controlled in the
E : voltage output of the oxygen sensor
following range [11]:
Emax : maximum voltage output of the oxygen sensor P
1
4
FFe3O4 FH2O RT ln( PH2O ) FPbO FH O RT ln aPb (2)
H2 2
Emin : minimum voltage output of the oxygen sensor
In a typical operating temperature range of 350~ 550 oC,
F : Faraday constants, F = 964 85.30929 C/mol the maximum oxygen concentration in LBE is 5.510-5
PH2 : partial pressure of the hydrogen gas wt%, corresponding to PH2O/PH2=104.42; the minimum
oxygen concentration is 1.0510-8 wt%, corresponding to
PH2O : partial pressure of the water steam
PH2O/PH2=10-0.75. Taking into account all of these, reasonable
PO2 : oxygen partial pressure of the working electrode to PH2O/PH2 ratio in cover gas of LBE loop will be in range
PO2 : oxygen partial pressure of the reference electrode of -0.75~4.42.

R : ideal gas constants, R=8.3144 J mol-1K-1 II. OXYGEN SENSOR


T : the absolute temperature of the melt LBE A. Solid Electrolyte Materials and Operation Principles
aPb : activity of lead Measurement of relative oxygen concentration in liquid
AxBy: free energy of formation of oxide AxBy metals using solid electrolyte membranes is well established
[2][3]. Solid electrolytes are materials permeable to specific
I. INTRODUCTION oxygen ions. In the case of a sintered ceramic zirconia,
ZrO2, O2- ions may pass through solid from high
Liquid LBE has been studied worldwide as a spallation concentration side to low concentration side if they have
target in Accelerator Driven Systems (ADS) recently. LBE sufficient thermal energy. Oxygen ions move by hopping
has also been a primary candidate material for nuclear between oxygen vacancy sites. Energy needed to hop
coolant of high-power spallation target due to its specific between these sites is reduced by stabilizing (or partially
thermal-physical and chemical properties [1]. However, one stabilizing) the crystal structure of ZrO2 by addition of 8-
of the problems using this lead alloy as a coolant in the ADS 18% yttria [2].

0-7803-8834-8/05/$20.00 2005 IEEE. 1746


gasket. A second flange with a high temperature BNC feed
through also seals with a copper gasket and through a spring
and the innermost tube presses the sensor onto its gasket.
Electrical connections are made via the BNC feed through to
a liquid reference through a Ta wire down to the center of the
ceramic tube [7].

Fig. 1. Schematic of the Electrochemical Cell in OS. Connector


to Voltmeter

Because the hot solid electrolyte membrane is porous to


oxygen, an imbalance in oxygen concentration will cause
oxygen to migrate so as to equalize concentrations on each
side. Oxygen on higher concentration side will pick up
two electrons to become ions, travel through the YSZ and
re-formed into a neutral molecule at lower concentration Water

surface [3]. This process is shown in Figure 1. Jacket

O2 + 4e- 2O2- (3) Stainless


Stell Tube

A metallic connection is used for sourcing and sinking


electrons, and chemically catalyzing the reaction shown
above. In liquid metals, a certain amount of the dissolved
oxygen exists as ions. The liquid metal is a good connection
to YSZ, both for making electrical connection to voltmeters
and providing the ions to pass through YSZ [3]. Sensor
Element

Bi/Bi2O3
In electrochemical terms, oxygen sensor can be Reference

represented by the following equation:

PO' 2 orao' // solid electolyte // O, PbO(liquid metal ) (4)

where superscript prime denotes the reference, and the solid Fig. 2. Schematic Drawing of an Oxygen Sensor.
electrolyte is typically YSZ. Since we only want to measure
C. Oxyen Sensor Signal
the relative oxygen concentrations, we must have a stable,
In the reference electrode, the reaction is:
known oxygen concentration reference to measure against.
The outside of the conical sensor will be immersed in LBE 2 Bi + 3O = Bi2O3 (5)
with an unknown oxygen concentration. If Bi is in Potential difference across the electrode is calculated
equilibrium with chunks of its solid oxide, the dissolved from the Nernst equation, which is
oxygen in the liquid (Bi and Bi2O3) will have a known PO'
E= RT
ln( P 2 )
saturation value, depending on the temperature. This 4F O2

saturation value will serve as a good reference. = 1


( 13 FBi 2 O3 FPbO RT ln aPbO + RT ln aPb ) 6)
Due to the oxygen concentration difference in oxygen- 2F

saturated reference and liquid LBE, there exists a chemical These activities are assumed to be simply related with
potential difference, resulting in the flow of oxygen ions, and corresponding oxygen concentrations by Henrys law.
accumulating charges. When it finally reaches equilibrium, Therefore, oxygen concentration dissolved in LBE melt can
there is an EMF (electromagnetic force) across. EMF is the be determined from the EMF reading of the YSZ Oxygen
Sensor. Two limits of sensor voltage output can be described
measure of the oxygen concentration difference [7].
as [11]:
B. Oxygen Sensor Design
( Fe3 O4 )
Emax = 8.123 10 5 T + 0.4304[V ] (7)
Oxygen sensors used in experiments are manufactured by 5
PbO
E min = 7 .264 10 T + 0 .142 [V ] (8)
Delphi [5]. The conical material with a rounded end, as
shown in Fig.1, is the sensing element of the sensor made of The maximum voltage output of YSZ Oxygen Sensor
YSZ. corresponds to minimal oxygen concentration in LBE that
Oxygen sensor shown in Fig.2 has a graphite seal at the forms minimal iron oxide Fe3O4 protective layer on the LBE
bottom of the tube so that it protrudes into the LBE flow. carrying tube, while the minimum voltage of YSZ Oxygen
The interior of the tube has several concentric tubes of Sensor corresponds to the maximum saturated concentration
stainless steel and alumina ceramic. A standard vacuum in LBE that lead oxide starts to precipitate. Figure 3 shows
flange is welded to the top end. A mating flange seal via a those maximum and minimum outputs, as well as shows
spring and the first inner tube, compresses the support ring

1747
output changes with temperature under different constant Controlled quantities of 6% H2 and O2 in He are
oxygen concentrations in LBE. introduced into system via a solid state mass flow controller
0.6
(MKS Multi Gas Controller 647C) in all SS system. The
PbO Precipitation
exhaust line includes a HEPA filter to trap particles, and a
0.5
vacuum oil trap for residual vapors. Sensor output is fed
Sensor Voltage Output [V]

Maximum Voltage Output


through a BNC cable to a high impedance electrometer
0.4
(Keithley 6514 Electrometer). Computer based data
-5
acquisition software/LabVIEW was utilized to continuously
0.3 Co=5.5x10
-4
record the temperature inside the LBE loop and sensor signal
Co=1.4x10
output. A Residual Gas Analyzer is also connected to the
0.2 -4

Minimum Voltage Output


Co=6.3x10
system to continuously monitor the partial pressures of all
Co=2.0x10
-3
gases present in the system, further to monitor the system
0.1
tightness.
Lliquid Metal Corrosion

0.0 B. Calibration Results


300 350 400 450 500 550 600 650 700 750
0 Experiment is conducted by initially setting up
Temperature [ C]
temperature of the system to 500 oC, and oxygen
Fig. 3. Maximum and Minimum Sensor Voltage Output to
concentration to a certain under-saturation condition. The
Maintain a Proper Oxide Layer.
apparatus is then closed. After a waiting time until the
oxygen activity reaches a fixed point, we start to slowly
III. EXPERIMENTAL SETUP AND RESULTS decrease temperature by 10 oC for every 20 minutes.
Temperature and sensor voltage signal were recorded by
A. Apparatus Description LabVIEW for each time interval. This procedure was
Calibration of Oxygen Sensor has been performed using repeated for various oxygen concentrations.
the experimental set up shown in Fig.4. This Oxygen Sensor Co=1.86e-5 Co=2.95e-5 Co=4.47e-5
calibration system consists of a temperature controlled U- 200

shape container, gas supplies and exhaust, a residual gas 190


OSB OSA
analyzer (RGA), a high-impedance electrometer and a PC for 180 Co=9.77e-5

data acquisition. This SS U-shape container is tightly sealed 170


OSA
from outside atmosphere using conflat flange except for gas
Voltage Output [mV]

160
Co=1.95e-4
inlet and outlet and other several openings for insertion of 150
OSB
OSA

thermocouple or RGA signal wire. Flexible heating tapes 140


OSB
Co=3.63e-4
around the SS tube were used to heat up the liquid metal up 130 OSA
to the required temperature controlled by a temperature 120
controller. The container is thermally insulated and placed on 110 E -OxHB
min
rocker to provide fluid motion that promotes mixing and
100
homogenization of oxygen concentration in LBE [7] [11]. 300 320 340 360 380 400 420 440 460 480 500 520

o
Temperature [ C]

Fig.5. Calibration Results from Sensors A & B.

Figure 5 shows the calibration results from Sensors A &


B. It finds after some initial transients, voltage-temperature
curves first follow constant oxygen concentration lines
according to Nernst equation, then turn to concentration
saturation line. These curves display linear trend with slopes
close to the theoretical values. A singular point appears when
turn to the oxygen concentration saturation line, this
indicates the moment when solution becomes oxygen
saturated. Due to insufficient waiting time to let oxygen
activity to reach a fixed point before the temperature starts
to decrease, the linear tread of the curves from Sensor B are
not as good as those of Oxygen Sensor A. It's clear that their
curve slopes are almost the same, both in the range from 0.33
to 0.5. This indicates the YSZ sensors we used are of high
sensing quality under our experimental conditions. The
turning points on the curves clearly depict the region at
which the solution becomes saturated in oxygen.
Fig. 4. Experimental Setup. (OS: YSZ Oxygen Sensor).

1748
Overlapping of the calibration curves indicates consistency of oxygen sensor to temperature change is really fast and has
of the sensors of the same design. linear trend. And figure 7 shows that how oxygen sensor
In the oxygen-saturated domain, for sensor A the response when oxygen is introduced following by hydrogen
variation between the experimental result and the theoretical dilution.
predication is limited to about +5mV. However, for oxygen
Sensor B the variation between the experiment result and IV. CONCLUSION
the theoretical prediction is about +10 mV or even higher.
This shows that there still exit some characteristic An YSZ oxygen sensor calibration/measurement
differences between them even they both have the same apparatus has been designed and built to deliberately
design mechanism. This is partly due to the variations of calibrate the sensor to be used in LBE. Current experiments
assembly issues have been operating between 350 oC and 550 oC. Experiment
data showed that oxygen sensor responses well under this
C. Sensor Characteristics temperature range. YSZ oxygen sensors will be tested at
Following sensor characteristics are usually expected for higher temperature up to 800 oC in the coming experiment
liquid metal oxygen sensors: accuracy, reproducibility, time schedule. Current available technology proves that it's easy
to response to changes about operating conditions, large to heat the liquid LBE up to 800 oC. However, experience
operating temperature and oxygen concentrations range, strongly suggests that new signal wires that can stand high
limited or no time drift, long service life, mechanical temperature with good electricity conductivity and with no
resistance, limited sensor to sensor differences. Figure 5 excess oxidation need to be exploited and tested first prior to
demonstrates linear trend close to the theoretical predictions the filed deployment.
even there exists 5~10 mV difference between the turning
point and the saturation line. Both Sensors A and B indicate REFERENCES
good accuracy of the oxygen sensor under such a low
[1] F. Barbier, G. Benamati, C. Fazio, and A. Rusanov,
oxygen concentration level (1.0510-8 ~ 5.510-5 wt%). Compatibility tests of Steels in Flowing Liquid Lead-
Bismuth, J. Nuc. Mat.,vol. 295, pp. 149-156. 2001.
[2] N. Li, Active Control of Oxygen in Molten Lead-
Bismuth Eutectic Systems to Prevent Steel Corrosion
and Coolant Contamination, J. Nuc. Mat. vol. 300 pp.
73-81. 2002.
[3] E. M. Logothetis and R. E. Htrick, High-Temperature
Oxygen Sensors Based on Electrochemical Oxygen
Pumping, Scientific Research Laboratories, Ford
Motor Company, Dearborn, MI 48121-2053.
[4] G.V. Samson(Ed.), The Oxide Handbook, 2nd Ed.,
IFI/Plenum: New York. 1982.
[5] N. Li and T. W. Darling, Oxygen Control
Methodology and Sensor Calibration Strategy, LANL,
Fig. 6. Voltage Response of Sensor vs. Temperature Change. May 30, 2002.
[6] V. A. Gulevsky, et al. Application of Hydrogen/Water
Figure 6 shows the reproducibility of sensor with respect Vapor Mixtures in Heavy Coolant Technology, Proc.
of Heavy Liquid Metal Coolants in Nuclear Technology
to temperature under this set of experiments. Although (HLMC-98), Obnisk: SSC RF IPPE, vol. 2. pp. 668-
hysteresis exists in the response time during the reversing 67. 1992.
temperature change process, it is within 25% for the [7] T. W. Darling and N. Li, Oxygen Concentration
temperature range of interest. This might be due to two Measurement in Liquid Pb-Bi Eutectic, LANL.
major factors: sensor's response time and response time of [8] G. Muller, A. Heinzel, G. Schumacher and A.
the oxygen in liquid LBE. Figures 5 shows voltage response Weisenburger, Control of Oxygen Concentration in
Liquid Lead and Lead-bismuth, Journal of NuC. Mat.
Data from 08-08-02(10:32 AM) - 08-16-02(9:46AM)
vol. 321. 2003.
460 440
[9] J. Zhang, N. Li and J. S. Elson, Review of Studies on
440
400 Fundamental Issues in LBE Corrosion, Progress in
420
360
Material Science, in press. 2004.
400 Temperature [10] X. L. Wu, et al., Calibration of YSZ Senxors for the
Measurement of Oxygen Concentration in Liquid Pb-
Voltage Output [mV]

320
380
Hydrogen with 94% He
Temperature [ C]

Bi Eutectic, Intl High-Level Radioactive Waste


o

360 was introducted ! 280

340
Voltage Management Conference, Las Vegas, NV. 2003.
240

320
[11] X. L. Wu, Calibration of YSZ (Yttira Stabilized
200
Zirconia) Oxygen Sensor Used in Liquid Bismuth,
300
Oxygen with 94% He 160 Master Thesis, UNLV, Nov. 2004.
280
was introduced! Oxygen with 94% He
120
260 was introduced!
240 80
0 1000 2000 3000 4000 5000 6000 7000 8000 9000 10000 11000 12000

Time [minute]
Fig.7. Time to Response to Oxygen Concentration Chantge

1749