8810 - ORP
SULPHIDE
0 - 1 mg/l S 2-
Content
1. APPLICATION
1.1 Basic configuration
1.2 Principle of the measurement
1.3 Calibration principle
2. PROGRAMMING
2.1 Programming menu
2.2 PM XXXX
2.2.1 Application
2.2.2 Options
2.2.3 Adjust I out
2.2.4 Unit.
2.3 Titration parameters
2.4 Cycle parameters
2.5 Sample conditioning
2.6 Chemical cleaning
2.7 General test
2.8 Sequence
2.9 Analog outputs
2.10 Example of programming
3. STARTING UP
3.1 Initial calibration
3.2 Start up
3.3 Process calibration
3.4 Reading calibration parameters
4. REAGENT PREPARATION
4.1 R1 = Titration: Silver nitrate
4.2 R2 = Conditioning: NaOH
4.3 R3 = Cleaning: NaOH + EDTA
4.4 Calibration solution
Page 2 / 9
2-
8810 - ORP - Sulphide 0 - 1 mg/l S
1. APPLICATION :
1.2 Principle of the measurement : Low level sulphide analysis impose the so-called
two points calibration type to determine slope and
The cell of the analyser is automatically filled with a asymmetry. On a regular basis (once a week
constant and reproducible volume of sample. After maximum, once a month minimum), perform any of
alkaline conditioning, sulphide are titrated directly the following calibration mode to your convenience.
by a solution of silver. The titration is followed by an
Ion Specific Electrode Sulphide.
The titration curve is increasing from a potential of
-500 mV and up to +200 mV with a maximum
deviation around +130 mV. The exact determination a. MANUAL = Manual introduction of two
of the END POINT of the curve, is calculated by the STANDARD injections, directly in the reactor vessel
analyser (Auto-End point). of the analyser.
Page 3 / 9
2-
8810 - ORP - Sulphide 0 - 1 mg/l S
Connect the analyser to the power supply. SAMPLE COND. : YES (press ENTER)
O Switch on the red button on the front panel. CAL 2 PTS : YES (press ENTER)
O Press SELECT to enter the main menu. GRAB SAMPLE : (YES or NO) depends if option
External sample measurement was purchased
O Press the UP/DOWN keys to select
SAMPLE TIME : XXX s usually 030 depends
(press ENTER)
"PROGRAMMING".
on volumetric pump flow (press ENTER).
O Press ENTER and hold on 5 seconds to enter the
PROGRAMMING menu. ZINC : NO (press ENTER)
Page 4 / 9
2-
8810 - ORP - Sulphide 0 - 1 mg/l S
O With the scroll down key, select UNIT and O With the scroll down key select "SAMPLE
press ENTER. COND.", and press ENTER.
Enter and check the following information :
O Enter or check that the unit is in accordance with
usual practise (mg/l, USER = PPB S) and press REAGENT 1 : YES (press ENTER)
ENTER.
INJ. REAG. 1 : 060 s (press ENTER)
O Exit from PM XXXX by pressing SELECT.
TIME REAG.1 : 020 s (press ENTER)
O With the scroll down key select "TIMING" and RESIDENC t : XXX s usually 030 adjust to
press ENTER. Enter or check the following keep a clean reactor vessel (press ENTER)
information:
clean reactor vessel (press ENTER) O With the scroll down key select "SERVICE" and
rinsing water flow (press ENTER) O With the scroll down key, switch on/off each
analyser subassembly for testing purposes.
INJECT TIME : 01 s (press ENTER)
O Pass to next subassembly by pressing ENTER.
DELAY t. : 07 s (press ENTER)
Page 5 / 9
2-
8810 - ORP - Sulphide 0 - 1 mg/l S
2.7.1 Sample-taken volume: The unit has been previously selected in the
Using a volumetric cylinder, adjust the level of the UNIT submenu ( see chapter 2.2.4: Unit)
siphon tube (if necessary) to capture exactly 100
ml in the reactor vessel. O With the scroll down key select "ZERO" and
! ! ! Exact volume of 100 ml is required to adjust your chart recorder or DCS on the scale
achieve accuracy during calibration step ! ! ! origin.
Following instructions are the Standard Operation O With the scroll down key select "FULL" and
Procedure to measure this volume. Drain the cell adjust your chart recorder or DCS on full scale.
(DRAIN), fill it with rinse water (SPRINKLER), O With the scroll down key select "CONC" and
weight empty cylinder and place under the drain press ENTER.
valve connector, drain the cell (DRAIN) and
weight full cylinder.
3. STARTING UP :
2.7.2 Volumetric pump:
The exact flow is not critical. Verify the time 3.1 INITIAL CALIBRATION:
necessary for the pump to fill the cell and activate
the siphon tube (Usual value: 20 to 30 seconds). It is the first operation to do after the programming.
Input this value as "SAMPLE TIME : XXX" of
menu PM XXXX / OPTION / GRAB SAMPLE (see O With the scroll down key select CALIBRATION
chapter 2.2.3 options) and press ENTER.
O With the scroll down key select "ANALOG O Just after temperature and pH conditioning,
OUT" and press ENTER. quickly inject in the reactor vessel 2 ml of
STANDARD solution (with a micro-pipette). For
Enter your desired range(for example: 0 -1 mg/l): preparation see chapter 4.4 : calibration solutions.
OUT : Conc
Page 6 / 9
2-
8810 - ORP - Sulphide 0 - 1 mg/l S
3.3 "PROCESS" CALIBRATION (If you have to 3.4 READING CALIBRATION PARAMETERS
adjust the calibration slope).
O At any moment, reading calibration parameters is
O Press SELECT to access the main menu. Press possible.
ENTER and the display will show If sure : ENTER.
Press ENTER to stop the analyser. O Press SELECT to access the main-menu and the
scroll down key until display shows READING.
O Press SELECT to access the main menu and the
scroll down key until the display shows O Press ENTER and hold on 5 seconds to enter the
PROGRAMMING. READING menu.
O Press ENTER and hold on 5 seconds to enter the O Press the scroll down key until display shows
PROGRAMMING menu. CAL PARAMETER and press ENTER.
O With the scroll up key, select CALIBRATION O Press ENTER to read LAST CAL menu with
and press ENTER. following information:
O With the scroll down key, select SYSTEM CAL P / O / M / A / N letter indicates the calibration
and press ENTER. mode. Respectively Process, On line, Manual,
Automatic, Primary.
O With the scroll down key, select PROCESS SYS S= XXX.X with value from 50% to 200%.
CAL and press ENTER. Co= X.XX with values from 0 to 999.
O The system starts titrating using the rinsing water O Report the calibration parameters on a logsheet
(sulphide free) for the 1st point and the process and press ENTER.
sample for the 2nd point.
O Press SELECT twice to come back to the
O During each sampling phase, grab the solution at operating mode.
the outlet of the siphon tube and send them for a
laboratory comparison measurement.
Page 7 / 9
2-
8810 - ORP - Sulphide 0 - 1 mg/l S
Page 8 / 9
2-
8810 - ORP - Sulphide 0 - 1 mg/l S
4.4 CALIBRATION SOLUTIONS (for 500 ml): transfer quickly into the volumetric flask. Close
securely the flask and stir until crystals totally
dissolve. Adjust to 500 ml with demineralised water.
Shake to mix and transfer into a dark or amber glass
storage flask. Label and date. Store at +5 C.
All sulphide solution preparation must be
performed under the hood to avoid ESL of a storage flask: Maximum 6 months or S
breathing noxious fumes. value is below 0.2 g/l. Neutralise with appropriate
method.
Page 9 / 9