Beruflich Dokumente
Kultur Dokumente
M. Sundararajan
Abstract
Introduction
The presences of two oscillators in the human body namely the heart and the lungs are
well known and widely reported in scientific literature. While heart is a closed loop
oscillator whose frequency is mostly system controlled, the lungs form an open loop
oscillator whose frequency can be voluntarily and consciously varied by the subject.
Their presence and the coupling between these two oscillators is evident by studying
288 M. Sundararajan
Photoplethysmography
Photoplethysmograph is a non-invasive technique that measures relative blood
volume changes in the blood vessels close to the skin. The first paper on PPG dates
back to 1936 when Molitor and Kniazak [8] recorded peripheral circulatory changes
in animals. In recent years, it has developed into a popular non-invasive method for
assessing mean arterial blood pressure and oxygen saturation (Pulse Oximeter). The
measurement of blood volumetric changes in the skin perfusion by means of PPG
depends on the fact that blood absorbs infrared light many times more strongly than
the remaining skin tissues.
Optical Sensor Based Instrument for Correlative 289
Figure 2.1: Optical characteristics of the biological tissues in the Visible and Infrared
region.
As the blood pressure in the skin vessels decreases, the surface area of the vessels
will reduce. This increases the average reflection in the measuring window, so it will
be recorded as an increase in the PPG signal. The optoelectronic measuring principle
of the PPG thus depends on detecting the changes in reflection of the sub-epidermal
layers of skin during and after a defined movement or occlusion routine, which causes
variations in the volume of the vessel plexuses in the skin [9]. As the optical radiation
is introduced into tissue, part of the photons will be reflected directly by the skin
surface, another fraction will be distributed in the tissue by absorption or scattering,
while the remaining photons will travel into the tissue either straight through or with a
number of collisions.
The simplest PPG sensor consists of an infrared LED and a photo detector placed
in a small plastic housing. It is shown in Figure 2.2. The sensor is applied to the skin
290 M. Sundararajan
Figure 2.2: Photoplethysmograph Sensor and Measuring window under the Sensor.
Advantages of PPG
PPG offers several advantages over other in-vivo optical methods like Laser Doppler
Flowmetry (LDF). PPG uses inexpensive optical sensors, which are rugged and needs
little maintenance. Since it consumes very less power and can be powered by a battery
pack, it is an ideal ambulatory device. The PPG signal contains a rich source of
information related to cardio-pulmonary system. A range of clinically relevant
parameters like heart rate, respiratory rate, and respiratory induced intensity
variations-RIIV can be obtained from the PPG signal. PPG can be used in the study of
neurologically induced skin perfusion changes [10], asymmetry of the brain [11], etc.
Breathing
Breathing is the most important function of the human body which is fundamental for
its very survival. This maintains the oxygen level in the blood stream to the required
amounts, which when transported through the arteries will metabolize various
functions that the human body performs [12]. In maintaining good health, the quality
of breathing process plays an important role that is the manner in which oxygen is
inspired and carbon dioxide is expired. The external nose serves to gather air and
accelerate its flow, forming a rapid jet that enters the cavity within the face - the
internal nose. The internal nose is strategically connected to the brain. In addition, the
olfactory nerve responsible for the sense of smell is present in the upper compartment
of the nasal cavity and has its nerve endings in most parts of that compartment [13].
Normal Breathing
Normally, the air enters and leaves the lungs at a rate of 14 to 16 times per minute
with out one being aware of it. The depth and the rate of normal breathing is regulated
peripherally and automatically to meet the oxygen needed by the cells and to
discharge the carbon dioxide accumulated in it. It is interesting to note that in a
Optical Sensor Based Instrument for Correlative 291
Experimental Setup
Experimental Setup and Data Acquisition
The PPG sensors were properly placed on the left and right temples of the subject
with the Tx and the Rx in line with blood vessel, after thoroughly cleaning the surface
of the skin with clinical spirit to avoid dust related artifacts. The subject is also
requested not make any movements when the measurement is in progress in order to
record a superior quality signal free from motion artifacts. Then the thermistors with
dimensions in the order of few microns are introduced into the left and right nostrils
of the subject without causing any irritation to him, this will give a recording of the
breathing pattern for the left and right nostrils that are independent of each other.
Finally the ECG sensors are placed appropriately. The sensors placements,
amplification settings, filter settings, ambient light in the room and all the other
necessary connections are thoroughly crosschecked by two people before the
experiment began.
Figure 4.1 shows a typical PPG signal from the modified optical sensor with some
artifacts. Any variation in the optical coupling, between the sensor head and the
subject, or physiological changes which dynamically alter the transmitted light give
rise to what is commonly termed as motion artifact [14]. In fact even a simple
movement may produce a complex motion artifact. Ambient light can also cause
artifact by coupling to the probe receiver, either directly or by transmission through
tissue. Whilst it is theoretically straightforward to remove ambient artifact, practical
limitations mean that sufficiently bright or high frequency artificial light sources can
still cause artifact. The normal PPG signal, which is free from artifacts, is shown in
292 M. Sundararajan
the figure 4.1. Figures 4.2 and Figure 4.3 show the breathing signal and ECG signal
recorded under normal breathing conditions respectively.
Figure 4.1: Atypical PPG signal free of any Figure 4.2: Atypical breathing signal free of any
artifact. artifact.
measurement system has a built in notch filter for removing the 50 Hz noise. The
system also has an inbuilt variable filter confined to each channel, this was used as a
25Hz low-pass filter for removing other noises while making the measurements. The
resulting graph gives a good idea on how the signal develops with respect to time
focusing on the signal shape and amplitude.
Signal processing
Next step in evaluating the raw signal is the application of Fast Fourier Transform
(FFT). For filtration of the raw data, application of Fast Fourier Transform (FFT) /
Power Spectral Density (PSD) and plotting graphs, the Diadem 7.0 program is used.
FFT gives a graph that clearly portrays the power distribution of the signal over the
entire frequency spectrum. As the FFT of the PPG signal contains numerous peaks,
Power spectral density which nothing but the square of FFT has been used to evade
confusion.
Usually the cardiac peak is the most dominant peak in the power spectrum of any
normal PPG recording. A relatively weaker peak will be detectable at the breathing
frequency. Other than these two peaks, a peak may be detected varying over a wide
range 0.14 and less, depending on the mental and physical condition of the subject
under study. Though several explanations are given for the appearance of this low
frequency component, the origin is still debatable.
Figure 5.1: Spectral analysis of ECG, Breathing and PPG data before performing
autogenic relaxation
Figure 5.2 shows a similar spectral analysis done on a similar data that were
recorded on the same subject after performing the autogenic relaxation described in
section 5.1. By carefully comparing the several spectral components that are
discussed above in figures 5.1 and 5.2, several interesting features may be observed.
First of all a much stronger and sharper (narrow width) breathing peak may be clearly
noticed in the spectral analysis of the PPG signal recorded after autogenic relaxation
in comparison with the measurements done before the autogenic relaxation
Figure 5.2: Spectral analysis of ECG, Breathing and PPG data after performing
autogenic relaxation.
Apart from this an important low frequency peak at 0.12 Hz may be observed in
PSD of the PPG signal in figure 5.2 which was not present in the PPG spectrum of the
signal that was recorded before performing autogenic relaxation. This periodic low
frequency component around 0.12-0.14 Hz may disappear and reoccur after some
time number of times from time to time [5]. This low frequency rhythm is also
defined as relaxation rhythm in scientific literature.
In the second phase of the experiment the PPG breathing and ECG data were
recorded under two different breathing conditions viz. Normal breathing and Deep
breathing, the protocol for deep breathing is described in section 5.1.
Figure 5.1 the spectral analysis (PSD) of PPG signal picked up by the optical
sensors positioned appropriately on the left and right temples of a subject under
normal breathing conditions. The same figure shows the FFT of the breathing signal
picked from the left and right nostrils using the miniature thermistors sensors.
The PPG spectrum shows dominant peak at 1 Hz corresponding to the heart rate
and a relatively weaker peak may be noticed at 0.3 Hz corresponding to the breathing
rate. It is an established fact that one does not breath with the same force thorough
both the nostrils, one nostril is always more prominent than the other. This
prominence shifts from one nostril to the other with a periodicity of two to eight hours
[17]. This one-nostril prominence can be clearly noticed from the FFT spectrum of the
breathing signal recorded from the left and right nostrils shown in the Figure 5.3, the
strength of the breathing activity through the right nostril is roughly three times that of
the left nostril.
Figure 5.4 shows a similar spectral analysis done on a similar data that were
recorded on the same subject while he was performing deep, the protocol for deep
breathing is elaborated in section 5.1. By cautiously studying the PPG spectra shown
in figures 5.3 and 5.4, it may be noticed that the breathing peak in the spectrum is
much sharper and has a slightly lower frequency pertaining to the protocol followed
for deep breathing. More importantly a prominent peak can be noticed at a frequency
marginally less than 0.12 Hz which has no harmonic relation with the respiratory
296 M. Sundararajan
activity or the cardiac activity, this is similar to the low frequency peak obtained in
the PPG spectrum of the data recorded after autogenic relaxation. It may also be noted
that respiratory peak and the cardiac peak are at multiple of a frequency or in other
words the ratio of heart rate to breathing rate is an integer in the case of deep
breathing. It may also be noted that the harmonics of the cardiac and the respiratory
peaks are strengthened during deep breathing.
Figure 5.3: Spectral analysis Breathing and PPG data recorded during normal
breathing
Figure 5.4: Spectral analysis Breathing and PPG data recorded during deep breathing.
Optical Sensor Based Instrument for Correlative 297
References
[1] Rao.M.M: Influence of controlled breathing (Pranayama) on dermal perfusion
as monitored by optical sensors. Proc. CNVD98, Berlin, VDI Verlag
Dusseldorf (1999),p. 49-55.
[2] M. Mukunda Rao, Adalberto Sapia: On appearance of low frequency rhythms
in the human body, presented at the International conference on Medical
Diagnostic Techniques and Procedures and published in its proceedings
(2000), p. 207-215
[3] Schmitt. H.J : Note on the generation of breathing related sub harmonics in
arterial PPG rhythms., Proc. CNVD98, Berlin, VDI Verlag Dusseldorf
(1999),p. 41-48.
[4] Hoyer. D: Nonlinear dynamics of cardio respiratory system, IEEE Eng. Med.
And BIOl. 12 (1998), P 16-61.
[5] Schmitt. H.J, Blazek . V: low frequency blood volume rhythms : possible origins
and new measurements, Aachen university of technology, Achen , Germany.
[6] Swami mukthibodhananda Saraswati: Swara yoga, Published by Bihar school
of yoga, Munger, Bihar, India, 1983.
[7] Swami Sivananda Radha: kundalini Yoga for the West, Published by Timeles
books, P.O. Box 50905, Palo Alto, CA 94303-0673/USA, 1978.
[8] Molitor.H. and Knaizuk.M, A new bloodless method for continuous recording of
peripheral change, Jour. Phar. Expr. Ther., Vol.27,pp-5-16,1936.
298 M. Sundararajan