Abstract
Basing on strict requirements of wearability, portability, cheapness and modularity, an assistive and
rehabilitative device for hand opening disabilities, characterized by an innovative kinematics, has been
developed and tested. This robotic orthosis is designed to be a low-cost and portable hand exoskeleton to
assist people with hand opening disabilities in their everyday lives. The exoskeleton can also be used such as
a rehabilitative device in order to restore the gestures of the hand after an injury or caused by a functional
impairment.
The device is mainly realized in Acrylonitrile Butadiene Styrene (ABS) structural components using a rapid
prototyping technique. The cable-driven actuation is provided by means of four independent servomotors
placed on the back of the hand.
Concerning the hand opening disabilities, a methodology, which properly defines the novel kinematic
mechanism that better fits the finger trajectories starting from the geometrical characteristics of the patients
hand, has been developed.
The testing phase of the real prototype with some patients is currently on going.
1. Project Objectives
The objectives of this project can be summarized, according to the Work Packages (WPs) described
hereinafter in the proposal, in the following way:
Improvement of reliability and usability of the device through a closed loop control using angular
feedback. Moreover, the usability of the exoskeleton will be improved by the development of an
intention sensing method based on the Electromyography (EMG).
Implementation of an automatic scaling algorithm in order to extend the mechanical solution of the
device to many people with different hand size.
More specifically, we aim at acquiring a deeper understanding of the models and theories meaningful to
achieve these goals.
With these objectives in mind, we will focus on the development of some prototype parts in order to improve
the following critical aspects of a biomedical device: portability, wearability, usability and modularity.
The portability requirement regards the development of lightweight and limited encumbrance mechanisms
and actuation systems.
The wearability requires an ergonomic structure and the realization of a comfortable device for the patient.
The usability ensures the patients to use it in a simple way.
The modularity requires a device adaptable to different users, who have of course different hand
characteristics and different disabilities. This aspect is very critical due to the variability of the human hand
(e.g. bone positions and tissue deformations) and complicates the design of the exoskeleton.
2. Background Information
Today, the first cause of adult disability in Europe is the Cerebral Vascular Accident (CVA). In particular, at
least 80% of post-stroke patients suffer hemiparesis of the upper arm. Moreover, the number of elderly with
a hand impairment as a result of the age is increasing. Post stroke survivors, genetic disease patients and
elderly with hand disease need, on one hand, timely and persistent rehabilitative training to regain previously
dexterity and, on the other hand, an aid in Activities of Daily Living (ADLs).
From the viewpoint of rehabilitation, it is crucial for the patient performing intensive and continuous
therapeutic tasks for a successful rehabilitation. Robotics systems allow to provide prolonged and higher-
intensity rehabilitation treatments, with a reduction of costs and burden for the therapists.
These devices are also able to evaluate the patients progresses by measuring physical parameters and to
replicate a given protocol always in the same conditions.
Unfortunately, sometimes, hand functions may not be totally replaced even after an intense rehabilitation
process. In these cases, the hand exoskeletons can be used in order to assist the user in ADLs. In particular,
this device can be used to assist the hand performance by amplifying the hand gripping force or automating
the motion.
According to the state of the art [1], the hand exoskeleton can be classified using various criteria (e.g. actuator
type, Degrees Of Freedom (DOFs), intention sensing and control method).
As regards the linking system between the hand and the exoskeleton, different approaches are described in
literature: multi-phalanx devices [2] [3], which are able to control directly each phalanx of the hand and
single-phalanx exoskeletons [4], which are linked to only one phalanx and are able to actuate only that part
of the hand. A multi-phalanx approach requires a more complex mechanism and control strategy [5] [6] [7]
and, in most cases, these devices are not very portable, so that they are used for rehabilitative purposes [8]
[9] and in haptics [10], where the portability requirement is not a constraint. Instead, a single-phalanx device
allows the use of simpler actuation systems and control algorithms.
On the topic of the mechanisms, many examples of multi-DOFs kinematic chains can be found [11] [12] [13],
while single-DOF mechanism are not many [14] [15]. Furthermore, current single-DOF devices, which are
currently studied, have a very simplified kinematics [16] [17] [18] [19], which is far from the physiological
hand kinematics and so, even though they are useful for assistance, their use is not suitable for rehabilitative
purposes. Soft-robotic applications, which are designed and developed in recent years, present a very
different type of mechanism based on elastomeric materials or fluid structures [20] [21] [22] [23].
Concerning the type of actuator, hand exoskeletons may be driven by electric actuators [24] [25] and
pneumatic actuators [26]. The latter ones lead to more weight and encumbrance for the actuation system.
For this reason, these devices are not so suitable for assistance due to their low portability.
Last but not least, for assistive hand exoskeletons, accurate sensing of the users intended motion is a primary
concern. For the purposes of controlling a device or ergonomic evaluation, there have been various methods
for detection of motion intention (e.g. motion sensing, breath switch, surface EMG, mechanomyography).
3. Detailed Technical Description
Technical Description
The aim of the project is to improve the usability and portability of a hand exoskeleton.
In order to achieve these results, the research activity will be organized in the following themes.
Closed-loop angular control
Using an encoder placed on the joint A of the Figure 4.1, the measure of the angle 2, which identifies the
single DOF of the mechanism, is possible.
Figure 4.1: Actuation system and 1-DOF mechanism of the hand exoskeleton
Through this angular information, a closed-loop control of the servomotors is realized by means of an Arduino
single-board microcontroller such as shown in Figure 4.2.
The angular information is useful to know the pose of the exoskeleton mechanism (important in
rehabilitation use) and the closed-loop angular control can be used to stop the actuator when an object has
been grasped (this skill is so important for the use of the device during the ADLs).
EMG based sensing of the users intended motion
An electromyographyc signal, after a suitable processing phase, is used to trigger the servomotors actuation
both in opening and closing of the device. So, through the EMG signal, the user can put to work the
exoskeleton without using the other hand.
Automatic scaling algorithm
Using the trajectories of some points of interest of the hand (acquired by the patient by means of a MoCap
System developed by BTS Bioengineering or artificially produced by a parametric hand model implemented
in Matlab), an optimization algorithm is implemented to adjust the characteristic of the exoskeleton to
reproduce the desired trajectories.
The entire procedure is shown in Figure 4.3.
Figure 4.3: Flow chart of the automatic scaling procedure
Thanks to this automatic scaling procedure, the mechanism results easily adaptable to different hand sizes
and suitable to track the movement of different hands by modifying automatically only a few geometrical
parameters.
This represents an important feature for a large-scale production device which has to be used in the patients
everyday lives.
Resources Needed
Our project is based on the availability of data (trajectories of the hand) which have been acquired before
the eNTERFACE16 workshop, and the availability of some components, which are listed below:
Encoder (e.g. SuperModified V3.0 for RC-servos with driver)
Arduino Nano Boards
Servomotor (Hitec HS-5495BH)
Oscilloscope
Solderer and soldering wires
Breadboard
Prototyping board
Electrical wires
DC Power supply
LiPo batteries
Electrodes for EMG
PC104
Load cells
Tactile sensors
These components are already placed on the exoskeleton or linked to it, but there could be, during the
workshop, the need of substitute some of them.
For the same reason, we may need to use a 3D printer to print some some damaged exoskeletons
components (we use a Stratasys Dimension Elite).
Complementary equipment (e.g. cables for the transmission, jumper cables) will be brought by the project
investigators, if it is not available onsite.
We will also take some pre-recorded data to be used as an initial benchmark.
During the eNTERFACE16 workshop, we are taking responsibility for installing all needed software.
Project Management
The project should be carried out up to 12 people team, plus the principal investigators, which will provide
insights by participating either to a part or to the entirety of the eNTERFACE workshop.
The principal investigators will coordinate the activities listed at the beginning of the section.
7 team members have been already identified, they are already acquiring the background necessary to carry
out the work and they will attend all 4 weeks.
Furthermore, they will work on the research lines as described in the following.
Stefano Capitani and Matteo Venturi: closed-loop angular control development.
Arianna Cremoni, Lukas Lindenroth and Ali Shafti: EMG acquisition and sensing of the users intended
motion development.
Nicola Secciani and Agostino Stilli: Scaling algorithm implementation and development.
4. Work Plan and Implementation Schedule
We suggest this set of work packages in which to divide the work for the project:
WP0: Project coordination
WP1: Closed-loop control development
WP2: Electromyography (EMG) analysis
WP3: Automatic scaling algorithm development
WP4: Evaluation, testing and project finalization
Two reports are scheduled during the activity:
Midterm report: this report will be written down after the first 2 weeks in order to describe the
performed steps and the achieved results.
Final report: this report will be written down at the end of the activity describing the final project
results.
During the 4-week-activity, the team will develop and test a low cost wearable and portable hand
exoskeleton to assist people with physical disabilities in their everyday lives.
In particular, the work aim at improving the reliability of the device and its adaptability and modularity.
These important goals will be achieve through the following steps.
Development of a closed-loop control architecture implemented using an Arduino single-board
microcontroller.
Development of an extraction system for an EMG signal acquired from the user of the device.
Implementation of an automatic scaling algorithm implemented in Matlab environment and
development of a suitable procedure for adapting the characteristic of the exoskeleton 3D
SolidWorks model to the hand of the user.
6. Academic Supervisors
Agostino Stilli is a Ph.D. student at Centre for Robotics Research (CoRe) at Kings
College London. He received his BSc degree in Mechanical Engineering and his
MSc in Electrical and Automation Engineering from the University of Florence
(Italy). He spent six months at Southern Denmark University in Odense (Denmark)
during his master degree course as visiting student, joining the robotic research
centre. He joined the Centre for Robotics Research at Kings College London in
September 2013 focusing his research on continuum manipulators, soft robotics,
inflatable robotics and variable stiffness systems.
Since our proposed team is made up for the most part by mechanical engineering students, we welcome new
members in our team with a background in electronics, automation and control engineering.
In particular, we are seeking participants with skills in the following areas of interest:
wearable sensors
signal processing
control systems (closed-loop architectures)
EMG analysis
8. References
Papers about the hand exoskeleton designed by the MDM Lab of the University of Florence:
Conti, R., Allotta, B., Meli, E., Ridolfi, A., Development, design and validation of an assistive device
for hand disabilities based on an innovative mechanism, ROBOTICA, Cambridge University Press,
2015, DOI:10.1017/S0263574715000879.
B. Allotta, R. Conti, L. Governi, E. Meli, A. Ridolfi, Development and Testing of a low cost Wearable
and Portable Hand Exoskeleton based on a parallel mechanism, in Proceedings of the ASME 2015
International Design Engineering Technical Conferences, IDETC 2015, August 2-5, 2015, Boston,
Massachusetts, USA
[1] P. Heo, G. M. Gu, and S. Lee, Current hand exoskeleton technologies for rehabilitation and assistive engineering,
Int. J. Precis. Eng. Manuf., vol. 13, no. 5, pp. 807824, May 2012.
[2] L. Martinez, O. Olaloye, M. Talarico, S. Shah, R. Arends, and B. BuSha, A power-assisted exoskeleton optimized for
pinching and grasping motions, in Bioengineering Conference, Proceedings of the 2010 IEEE 36th Annual
Northeast, Mar 2010, pp. 12.
[3] M. S., K. Y., A. S., O. A. T. T., F. M. Q., and I. T., Muscle stiffness sensor to control an assistance device for the
disabled, Artificial Life and Robotics, vol. 8, no. 1, pp. 4245, Sep 2004.
[4] J. Iqbal, N. Tsagarakis, A. Fiorilla, and D. Caldwell, A portable rehabilitation device for the hand, in Engineering in
Medicine and Biology Society (EMBC), 2010 Annual International Conference of the IEEE, Aug 2010, pp. 3694
3697.
[5] C. Jones, F. Wang, C. Osswald, X. Kang, N. Sarkar, and D. Kamper, Control and kinematic performance analysis of
an actuated finger exoskeleton for hand rehabilitation following stroke, in Biomedical Robotics and
Biomechatronics (BioRob), 2010 3rd IEEE RAS and EMBS International Conference on, Sep 2010, pp. 282287.
[6] J. Li, R. Zheng, Y. Zhang, and J. Yao, ihandrehab: An interactive hand exoskeleton for active and passive
rehabilitation, in Rehabilitation Robotics (ICORR), 2011 IEEE International Conference on, Jun 2010, pp. 16.
[7] Y. Hasegawa, Y. Mikami, K. Watanabe, and Y. Sankai, Five-fingered assistive hand with mechanical compliance of
human finger, in Robotics and Automation, 2008. ICRA 2008. IEEE International Conference on, May 2008, pp.
718724.
[8] C. Schabowsky, S. Godfrey, R. Holley, and P. Lum, Development and pilot testing of hexorr: Hand exoskeleton
rehabilitation robot, Journal of NeuroEngineering and Rehabilitation, Jul 2010.
[9] C. Jones, F. Wang, C. Osswald, X. Kang, N. Sarkar, and D. Kamper, Control and kinematic performance analysis of
an actuated finger exoskeleton for hand rehabilitation following stroke, in Biomedical Robotics and
Biomechatronics (BioRob), 2010 3rd IEEE RAS and EMBS International Conference on, Sep 2010, pp. 282287.
[10] A. Wege, K. Kondak, and G. Hommel, Mechanical design and motion control of a hand exoskeleton for
rehabilitation, in Mechatronics and Automation, 2005 IEEE International Conference, Aug 2005, pp. 155 519.
[11] A. Chiri, N. Vitiello, F. Giovacchini, S. Roccella, F. Vecchi, and M. Carrozza, Mechatronic design and characterization
of the index finger module of a hand exoskeleton for post-stroke rehabilitation, Mechatronics, IEEE/ASME
Transactions on, vol. 17, no. 5, pp. 884 894, Aug 2012.
[12] A. Wege and G. Hommel, Development and control of a hand exoskeleton for rehabilitation of hand injuries, in
Intelligent Robots and Systems, 2005. (IROS 2005). 2005 IEEE/RSJ International Conference on, Aug 2005, pp. 3046
3051.
[13] C. Takahashi, L. Der-Yeghiaian, V. Le, R. Motiwala, and S. Cramer, Robot-based hand motor therapy after stroke,
Brain, vol. 131, no. 2, pp. 425437, Feb 2008.
[14] M. Mulas, M. Folgheraiter, and G. Gini, An emg-controlled exoskeleton for hand rehabilitation, in Rehabilitation
Robotics, 2005. ICORR 2005. 9th International Conference on, Jun 2013, pp. 371 374.
[15] K. Rotella, M.F.and Reuther, C. Hofmann, E. Hage, and B. BuSha, An orthotic hand-assistive exoskeleton for
actuated pinch and grasp, in Bioengineering Conference, 2009 IEEE 35th Annual Northeast, Apr 2009, pp. 12.
[16] E. Brokaw, I. Black, R. Holley, and P. Lum, Hand spring operated movement enhancer (handsome): A portable,
passive hand exoskeleton for stroke rehabilitation, Neural Systems and Rehabilitation Engineering, IEEE
Transactions on, vol. 19, no. 4, pp. 391399, Aug 2011.
[18] P. Medical. Kinetec maestra portable hand cpm. [Online]. Available: www.pattersonmedical.com
[19]K. M. Inc. Hand physical therapy with the hand mentor. [Online].
Available:http://www.kineticmuscles.com/handphysicaltherapy-handmentor.html
[20]H. Yap, F. Nasrallah, J. Lim, F. Low, J. Goh, and R. Yeow, Mrc-glove: A fMRI compatible soft robotic glove for hand
rehabilitation application, in Rehabilitation Robotics (ICORR), 2015 IEEE International Conference on, Aug 2015, pp.
735740.
[21]P. Polygerinos, Z. Wang, K. Galloway, R. Wood, and W. C.J., Soft robotic glove for combined assistance and at-home
rehabilitation, Robotics and Autonomous Systems, vol. 73, pp. 135143, Nov 2015.
[22] R. Deimel and O. Brock, A novel type of compliant, underactuated robotic hand for dexterous grasping, in
Proceedings of Robotics: Science and Systems, Berkeley, USA, July 2014.
[23] S. Delph, M.A.and Fischer, P. Gauthier, C. Luna, E. Clancy, and G. Fischer, A soft robotic exomusculature glove with
integrated semg sensing for hand rehabilitation. in Rehabilitation Robotics (ICORR), 2013 IEEE International
Conference on, Jun 2013.
[24] T. K.Y., S. Ho, P. Pang, X. Hu, W. Tam, K. Fung, X. Wei, P. Chen, and M. Chen, An intention driven hand functions
task training robotic system. in Engineering in Medicine and Biology Society (EMBC), 2010 Annual International
Conference of the IEEE, 2010.
[25] J. Iqbal and B. Khelifa, Stroke rehabilitation using exoskeleton-based robotic exercisers: Minireview,Biomedical
Research, vol. 26, no. 1, pp. 197201, 2015.
[26] L. Lucas, M. DiCicco, and Y. Matsuoka, An EMG-controlled hand exoskeleton for natural pinching, Journal of
Robotics and Mechatronics, vol. 16, no. 5, pp. 17, 2004.