Sie sind auf Seite 1von 2

HAXO-8 Humidity

Temperature Recorder

Using the LogTag Interface and LogTags freely


available companion software LogTag Analyzer, the
LogTag is easily set-up for recording conditions including
delayed start, sampling interval, number of readings,
continuous or fixed number of readings and configuration
of conditions to activate the ALERT indicator.

Readings are downloaded using LogTag Analyzer


The LogTag HAXO-8 Humidity & Temperature Recorder which provides facilities for charting, zooming, listing
measures and stores up to 8000 sets of high resolution data statistics and allows exporting the data to other
humidity and temperature readings over a measurement applications such as Excel.
range of 0 to 100%RH & -40C to +85C (-40F to +185F).
The LogTag HAXO-8 complies with the relevant
Enclosed in a robust and durable polycarbonate case - the international standards for temperature monitoring
HAXO-8 is equipped with a unique external temperature devices:- FCC, CE, C-TICK and RoHS. This
sensor arrangement providing fast reaction time to demonstrates the quality and suitability of the LogTag
temperature change and a real time clock which provides HAXO-8 for temperature monitoring applications where
date/time stamps for each temperature reading. accuracy and consistency is required.

Product Highlights Recommended Applications

A real time clock provides date/time stamps for each


Cold Chain Pharmaceuticals
temperature reading.
Push-to-start button with optional delay or a specific time Produce/Meat/Dairy Laboratories
& date.
Comprehensive customisation options including alert Agriculture & Floral Clinical Trials
settings, sample interval and trip duration.
Robust and durable polycarbonate case with lug for Animals Chemicals
secure mounting.
Blood/Organs/Tissue Museums
Up to 8,000 recordings - enough for the longest trip.
In-transit inspections can be recorded at the push of a
button. Server Room Office & Warehouse

Industry best download time - less than 5 seconds for


HVAC Manufacturing
fully memory.

Sports/Leisure

RELEASE SPECIFICATION : DOC REV C 110715 - Copyright 2015, LogTag Recorders Ltd
LogTag Recorders Ltd reserves the right to change this specification without notice. Page 1 of 2
Accuracy/Resolution Charts Accessories
Rated Relative Humidity Reading Accuracy @ 25C Rated Temperature Reading Accuracy
6%RH 3.0C

5%RH 2.5C

Rated Absolute Accuracy


Rated Absolute Accuracy

4%RH 2.0C

3%RH 1.5C

2%RH 1.0C

1%RH 0.5C

0 10 20 30 40 50 60 70 80 90 100 -40 -30 -20 -10 0 10 20 30 40 50 60 70 80 Wall Mount Bracket


%RH (Relative Humidity) Temperature (C)

Specifications
Product Model HAXO-8 Our FREE LogTag Analyzer
Humidity Measurement Range 0 ~ 100%RH but with limitations as detailed in Humidity Measurement software provides an easy
Operating and Storage conditions below. to use, powerful platform
Temperaturte Measurement -40C to +85C (-40F to +185F) for configuring any LogTag
Range recorder product before
Humidity Resolution Better than 0.1%RH deployment and for data
Temperature Resolution Better than 0.1C or 0.1F download & analysis when the
recorder is retrieved.
Capacity 8000 pairs of humidity & temperature readings
(32Kb memory)
Sampling Interval Configurable from 30 seconds to several hours

Environmental IP61 (when hung or mounted vertically)


Power Source 3V Lithium
Battery Life 2~3 years typical use
Size 86mm(H) x 54.5mm(W) x 8.6mm(T) LogTags unique interface
cradle design provides rapid &
Weight 35g
reliable LogTag data transfer.
Case Material Polycarbonate

%RH Note: Users do not need


Additional RH uncertainty for temperatures different than 25C to purchase more than one
100% Interface Cradle per LogTag
This graph shows the possible additional
90%
2-4% uncertainty in %RH compared to the accuracy
product.
80%
specifications for temperatures different than 25C
70%
for the standard factory calibration. The shown
2-3%
60%
1-2% 0-1% 1-2% uncertainties may be positive or negative.
50%

40% The performance can be improved by re-


calibrating the product for %RH at the specific
30%
temperature of interest after which case this chart
20%
would no longer apply.
10%
C
-20 -10 0 10 20 30 40 50 60 70 80

Humidity Measurement Operating and Storage Conditions


100%
80% This chart shows the normal recommended
operating range of the humidity sensor. Conditions
outside the recommended range may temporarily
60%
offset the RH signal up to 3 %RH.
40% After return to normal conditions it will slowly
return towards calibration state by itself. See Exposure to Chemicals
20% Reconditioning Procedure to accelerate
this process. Prolonged exposure to extreme Chemical vapors may interfere with materials
0% conditions may accelerate ageing. used for the humidity sensor. The diffusion
-40C 0C 40C 80C of chemicals into the sensors polymer may
cause a shift in both offset and sensitivity. In a
clean environment the contaminants will slowly
Re-Conditioning Procedure outgas.

Exposure of the internal sensor to chemical vapors may interfere with the internal sensor and cause inaccurate readings to be logged. The reconditioning procedure described to the
In a clean environment, this will slowly rectify itself. However, exposure to extreme conditions or chemical vapors will require the left will accelerate this process. High levels of
following reconditioning procedure to bring the internal sensorback to calibration state. pollutants may cause permanent damage to the
humidity sensors polymer.
80C (176F) at <5%RH for 36 hours (baking) followed by
20-30C (70-90F) at >74%RH for 48hours (re-hydration)

High levels of pollutants may cause permanent damage to the internal sensor.

Page 2 of 2