Beruflich Dokumente
Kultur Dokumente
Quality Assurance
2.Fluoroscopic Systems
A. Exposure Reproducibility -Fluorscopic spot film devices use automatic
exposure controls (AEC)
-they should be evaluated according to the AEC
tests that follow
B. Field Size Accuracy & Beam -fluoroscopic units should not be capable of
Alignment irradiating tissue outside the IR area
-should display everything within 1cm of the
edges of the image intensifier tube
-the primary beam should be aligned to the
center of the image intensifier
-correction of the tube collimation, image
intensification system, or video display system
may be required
3.Tomographic Systems
A. Uniformity & Completeness -tomography relies on the motion of the tube
of Motion during exposure to produce a sectional image of
the subject
-test tool is a lead mask with a pinhole that is
positioned several cm above the tomographic
fulcrum
-if the pinhole is centered to the IR and then
imaged with a full tomographic motion this is
seen:
B. Ion Chamber Sensitivity -most AECs utilize 3 ion chambers and permit
activation of various combinations during
exposures
-need to make sure individual ion chambers are
equally sensitive
-do this by using AEC without a subject at
extremely low kVp
-then the test is done by using a lead brick
to block all chambers but one to be tested
-the AEC reproducibility test is then
performed for the unblocked chamber
-if all chambers dont respond within +/-10%
then a service engineer should be consulted to
recalibrate the generator AEC circuitry
E. Backup Timer Verification -when you forget to set AEC or activate the
proper tube, patient protection requires that
AEC backup timers be functioning to terminate
exposures and protect both patient and tube
-when a lead plate is placed over the AEC
Notebook #11
Technologist Responsibilities: