Beruflich Dokumente
Kultur Dokumente
Organs at risk (OR) in the treatment area (list organs and desired objectives
in the table below):
Contour all critical structures on the dataset. Place the isocenter in the
center of the PTV (make sure it isnt in air). Create a single AP field using the
Bijoy Anand Page | 2
lowest photon energy in your clinic. Create a block on the AP beam with a
1.5 cm margin around the PTV. From there, apply the following changes (one
at a time) to see how the changes affect the plan (copy and paste plans or
create separate trials for each change so you can look at all of them).
Bijoy Anand Page | 3
Plan 1: Create a beam directly opposed to the original beam (PA) (assign
50/50 weighting to each beam)
a. What does the dose distribution look like?
has to provide more MUs to achieve 50-50 split between each beam.
back. The maximum dose is 66.48 Gy, or 110.8% of the prescribed dose.
Plan 2: Increase the beam energy for each field to the highest photon
cooler. The coverage to the PTV is more homogenous, but the 95% isodose
however, it has moved deeper (anteriorly) by about 1.1 cm. Now, the
The 15 MV beam has a Dmax ~ 3 cm compared to 1.5 cm for 6MV. This build-
Plan 3: Adjust the weighting of the beams to try and decrease your hot
spot.
a. What ratio of beam weighting decreases the hot spot the most?
I increased the weighting for AP gradually and watched the maximum dose
b. How is the PTV coverage affected when you adjust the beam
weights?
As I increase the AP weight, the isodose lines get pushed posteriorly and
the hot spot gets pushed the same. Now, the plan is more uniform, but
Maximum dose to PTV = 60.82 Gy; Average dose to PTV = 58.34 Gy;
Plan 4: Using the highest photon energy available, add in a 3rd beam to the
plan (maybe a lateral or oblique) and assign it a weight of 20%
a. When you add the third beam, try to avoid the cord (if it is being
treated with the other 2 beams). How can you do that?
i. Adjust the gantry angle?
ii. Tighter blocked margin along the cord
iii. Decrease the jaw along side of the cord
Doing any of the above would allow us to avoid the cord. However, tighter
blocked margins may possibly reduce the coverage to the PTV. Then, if we
increase the MUs to increase the dose to PTV, we get larger hot spots. I
chose to adjust the gantry angle and decrease the jaw along side of the
cord.
Bijoy Anand P a g e | 10
Bijoy Anand P a g e | 11
b. Alter the weights of the fields and see how the isodose lines
change in response to the weighting.
This was done and the best result is shown below.
Bijoy Anand P a g e | 12
c. Would wedges help even out the dose distribution? If you think so,
try inserting one for at least one beam and watch how the isodose
lines change.
Yes, inserting a wedge in the correct orientation definitely helped even
out the dose. I played around with various wedge angles and found that
only kept the AP beam and the LPO beam. The AP beam had a 15 wedge.
Bijoy Anand P a g e | 13
Which treatment plan covers the target the best? What is the hot spot
for that plan?
The last treatment plan covered the PTV the best. The hot spot = 61.80 Gy
Did you achieve the OR constraints as listed above? List them in the table
above.
The plan met the constraints for Heart, Esophagus, Skin and Lung. But, it did
not meet them for Brachial Plexus, Spinal cord.
Bijoy Anand P a g e | 14
I was able to see the effect of that one variable on the coverage and isodose
lines. We started by increasing the energy, then changing the beam weights,
then studying the impact of introducing the wedge, changing the gantry angle to
though contours for organs such as Heart, Brachial Plexus etc. normally would
doing them. This was very helpful. For example, I learned how to use the
Hopefully, I will able to spot some of the things quicker and thus avoid
repeating some of these learning steps; but, I fear I have ways to go before I