Sie sind auf Seite 1von 2

Green Fume HoodTechnologies

Authorized access: User R

Model: M2 - M3 - M4 - M5
(2)
RFID(1)reader LCD screen
User Multifunction
Keyboard
ok
Yellow LED
+ Enter -
Up Next Down Level 1 alarm
Visual: Flashing LED
Audible: Beeping
WSE134 10/2008
FPH.JML.RZ.CH

Level 2 alarm
Cancel Visual: Fixed LED
04/29/2009 05:01PM Audible: Beeping
Return
Access granted
(Simple user) Level 3 alarm
Visual: Flashing LED
Y o u c a n t u r n Mode
o nSecuret h e
Hood Audible: Continuous
Level 4 alarm
Visual: Fixed LED
Audible: Continuous
ok (3)
Red LED
A Global Solution For Green Buildings.
WSE130 09/2008

Authorizes access
ON
to the hood ON/OFF Lighting Ventilation ON/OFF Green LED
by RFID card
OFF

User A
Open Mode Function
User Operator User Card Fans OFF
This authorizes the holder to:
1 - Start and stop the fans (Ventilation ON/OFF key)
2 - Switch the lighting on and off (Lighting ON/OFF key)
3 - Deactivate the disengageable audible alarms if the right has bee granted by the
04/29/2009 05:01PM
administrator (Cancel key).
Access granted
(Simple user)

You can turn on the


Hood

Secure Mode function


Fans OFF
User X
User
B A
04/29/2009 05:01PM

04/29/2009 05:01PM Work in progress


04/29/2009 05:01PM
Saved energy
Insert a valid user id 3.25 kWh
card to use this
Access granted
Machine
(Simple user)

You can turn


the light on

User X
B
04/29/2009 05:01PM 04/29/2009 05:01PM

Secured mode Shut down in progress


function disabled 00:10

User A

B
04/29/2009 05:01PM

(1)
RFID - Acronym (Radio Frequency Identification) referring to an identification Shut down in progress
Technology using radio frequency data communication. 00:00
(2)
LCD - Liquid Cristal Display Technologie where the image is produced by
liquid crystals excited by an electric current.
(3)
LED- Light-Emitting Diode.
Alarms
Causes and solutions

R

For Green Fume Hood Technologies


Model: M2 - M3 - M4 and M5
Number of
the alarm

the alarm

Clearable
Level of
Solutions

Audible
Visual

alarm

alarm
alarm
Causes Alarm
Message *Authorized by the administrator

Keyboard used by an unauthorized user Fixed To use the hood,insert Place an authorized USER card
11A (or no authorized access card) 4 Red
Continuous
an authorizeduser card No in the RFID reader.

This alarm has priority even though it Flashing Reinsert card. Reinsert an authorized USER card
8D is level 3. Access card removed 3 Red
Continuous emergency stop in
00:XX
No in the RFID reader.

Fan adjustment problem. Fan X regulation


Fixed Contact the
8G The rotation setpoint in rpm cannot be achieved by one
or more fans.
2 Yellow
Beeping problem,setpoint
cannot be reached
No Administrator

Fan out of service. Rotation speed less than 700rpm. Stop working
Faulty fan management electronic Fixed Fan X: out of order
8H Card or poor contact of the fan 4 Red
Continuous
stop working now No and
contact the Administrator
Connexion cable.

Solvent alarm threshold temporarily exceeded *Press <ENTER> to


Fixed Filters end of life
8J for longer than the "solvent alarm delay
Time configured in minutes.
2 Yellow
Beeping
(Sensor solvent) YES mute current alarm sound
and call Administrator

Solvent alarm threshold permanently Fixed Replace filters Contact the


8K Exceeded.saturated filter modules 4 Red
Continuous
(Sensor solvent) No Administrator

Flashing Sensor solvent Contact the


8L Faulty solvent detector electronic card. 1 Yellow
Beeping
Out of order No Administrator
Acid alarm threshold temporarily exceeded for a period *Press <ENTER> to
Fixed Filters end of life
8M greater than the "Solvent alarm delay"
Time configured in minutes.
2 Yellow
Beeping
(Sensor acid) YES mute current alarm sound
and call Administrator

Acid alarm threshold permanently exceeded. Fixed Replace filters Contact the
8N Saturated filter modules. 4 Red
Continuous
(Sensor acid) No Administrator

Flashing Sensor acid Contact the


8P Faulty acid detection electronic card. 1 Yellow
Beeping
out of order No Administrator

Door open too wide: Opening greater than the Fixed Sensor acid Lower the door until the opening
8S configured alarm threshold. 2 Yellow
Beeping
out of order No complies with the threshold

*Press <ENTER> to
Fixed Check filters
8R Maintenance alarm 2 Yellow
Beeping
saturation YES mute current alarm sound
and call Administrator

The filter expiry date has passed. This date is Fixed Filters end of Partial or total replacement
8T saved when the filters are installed. 4 Red
Continuous
life(timer) No of the filter modules.

The lifetime saved when the filters were installed


Fixed Filters expiration Contact the
8U has been exceeded. Value configured between 5 4 Red
Continuous
date is over No Administrator
and 700 days.
Fixed Change Contact the
8Q End of acid sensor lifetime" alarm 4 Red
Continuous
sensor acid No Administrator
the established

the established
the administrator

1 - Stop handling
According to

According to
Determined by

2 - Contact safety officer


Alarms triggered remotely by 3 - Lower the sash Contact the
level

leve

Xxxx the administrator via a surveillance


PC and the gguard software
4 - Warning solvent
detection in laboratory
No Administrator
5 - Warning solvent or acid
Detected in your handling

Flashing Histo file is full, Contact the


8Y Log file has reached a critical size. 3 Red
Continuous
call maintenance No Administrator

Configuration does not comply with Fixed Hardware configuration Contact the
9A the factory configuration 4 Red
Continuous
Mismatch No Administrator

Anormal temperature in
40C temperature alarm threshold Flashing Check the heat sources.
8V exceeded in the working volume. 3 Red
Continuous working area.
call maintenance
No Contact the Administrator

Top priority alarm 60C Critical temperature. Stop your handling operation immediately do not
Fixed
8W temperature alarm threshold exceeded 3 Red
Continuous Fans stopped,
stop working now
No forget the heat sources. Completely close the
in the working volume. door. Call the Administrator
Faulty heat detection Flashing Sensor temperature Contact the
8V electronic card 1 Yellow
Beeping out of order No Administrator