Radharaman Institutes of Research & Technology Bhopal
Department of Electronics and Communication
Experiment No. 5
Objective: - To study the function of Multihole Directional Coupler by
measuring the following parameters.
Equipment :- Microwave Source(Klystron or Gunn Diode), Isolator,
Frequency Meter, Variable Attenuator, Slotted Line, Tunable Probe, Detector Mount Matched Termination, MHD Coupler, Wave guide stand , Cables and Accessories, VSWR meter.
Theory:- A Directional Coupler is a device with which it is possible to
measure the incident and reflected wave separately. It consist of two transmission lines the main arm and auxiliary arm, electromagnetically coupled to each other (as shown in fig.) The power entering, in the main arm gets divide between port 2 and 3, almost no power comes out in port (4) Power entering at Port(2) is Divided between port 1 and 4. The coupling factor is defined as Coupling C (db) = 10log10 [P1 / P2] Where port 2 is terminated Isolation I (db) = 10 log10 [P2 / P3] Where P1 is matched With built-in termination and power entering at port 1, the directivity of the directional coupler is a measure of separation between incident wave and reflected wave. Directivity is measured indirectly as follows: Directivity (db) = I C = 10 log10 [P2 / P1] Main line VSWR is SWR measured, looking into the main line input terminal when the matched loads are placed at all other ports. Auxiliary line VSWR is SWR measured, looking into the Auxiliary line output terminal when the matched loads are placed on other terminal. Main line insertion loss is the attenuation introduced in the transmission line by the insertion of coupler, It is defined as;
Insertion loss (db) = 10 log10 [P1 / P2]
Main Line SWR Measurement
1. Set up the equipment as shown in fig 2 2. Energize the Microwave source for particular frequency source for particular frequency of operation as described in the procedures given in the operation of klystron tube/Gunn Oscillator. 3. Follow the procedure as described for VSWR measurement (low and medium SWR measurement). 4. Repeat the same for other frequency.
Auxiliary line SWR Measurement
1. Set up the equipment as shown in fig 2 2. Energize the Microwave source for particular frequency source for particular frequency of operation as described in the procedures given in the operation of klystron tube/Gunn Oscillator 3. Measure SWR as described in exp. 4. Repeat the same for other frequency.
Measurement of coupling Factor, Insertion loss, Isolation & Directivity
1. Set up the equipment as shown in fig 2 2. Energize the Microwave source for particular frequency 3. Remove the multi hole Directional Coupler and connect the detector mount to the frequency meter. Tune the detector for maximum output. 4. Set any reference level of power on VSWR meter with the help of variable attenuator. Gain control knob of VSWR meter, and note down the reading. 5. Insert the Directional Coupler as shown in fig. 3 with Detector to the auxiliary port 3 and matched termination to port 2, without changing the position of variable attenuator and gain control knob of VSWR meter. 6. Note down reading on VSWR meter on scale with the help of range db switch if required >let it be Y. 7. Calculate coupling factor which will be X-Y=C(db) 8. Now carefully disconnect the detector from the auxiliary port 3 and matched termination from port 2 without disturbing the setup. 9. Connect matched termination to the auxiliary port 3 and detector to port 2 and measure the reading on VSWR meter. 10. Compute insertion loss. 11. Repeat steps 1-4. 12. Compute directivity. 13. Repeat same for other frequencies.