Sie sind auf Seite 1von 11

Eur Spine J (2009) 18:16951705

DOI 10.1007/s00586-009-1146-y

ORIGINAL ARTICLE

The effect of different design concepts in lumbar total disc


arthroplasty on the range of motion, facet joint forces and
instantaneous center of rotation of a L4-5 segment
Hendrik Schmidt Stefan Midderhoff
Kyle Adkins Hans-Joachim Wilke

Received: 9 March 2009 / Revised: 29 June 2009 / Accepted: 16 August 2009 / Published online: 4 September 2009
Springer-Verlag 2009

Abstract Although both unconstrained and constrained implant with a constrained core. All implants caused a
core lumbar artificial disc designs are in clinical use, the moving center of rotation. Except for axial rotation, the
effect of their design on the range of motion, center of unconstrained and constrained configurations mimicked
rotations, and facet joint forces is not well understood. It the intact situation. In axial rotation, only the Slide-
is assumed that the constrained configuration causes a Disc with mobile core reproduced the intact behavior.
fixed center of rotation with high facet forces, while the Results partially support our hypothesis and imply that
unconstrained configuration leads to a moving center of different implant designs do not lead to strong differ-
rotation with lower loaded facets. The authors disagree ences in the range of motion and the location of center
with both assumptions and hypothesized that the two of rotations. In contrast, facet forces appeared to be
different designs do not lead to substantial differences in strongly dependent on the implant design. However, due
the results. For the different implant designs, a three- to the great variability in facet forces reported in the
dimensional finite element model was created and sub- literature, together with our results, we could speculate
sequently inserted into a validated model of a L4-5 that these forces may be more dependent on the indi-
lumbar spinal segment. The unconstrained design was vidual spine geometry rather than a specific implant
represented by two implants, the Charite disc and a design.
newly developed disc prosthesis: Slide-Disc. The con-
strained design was obtained by a modification of the Keywords Mobile artificial discs  Finite
Slide-Disc whereby the inner core was rigidly con- element analysis  Arthroplasty devices  Back pain 
nected to the lower metallic endplate. The models were Implants
exposed to an axial compression preload of 1,000 N.
Pure unconstrained moments of 7.5 Nm were subse-
quently applied to the three anatomical main planes. Introduction
Except for extension, the models predicted only small
and moderate inter-implant differences. The calculated Lumbar total disc arthroplasty devices have been intro-
values were close to those of the intact segment. For duced to clinics as an alternative to fusion with the aim of
extension, a large difference of about 45% was calcu- preserving spinal motion. This should also alleviate back
lated between both Slide-Disc designs and the Charite pain, decrease the incidence of adjacent segment degen-
disc. The models predicted higher facet forces for the eration, avoid complications related to fusion, and allow
implants with an unconstrained core compared to an early return to function [3, 14, 15, 26, 33]. Due to these
advantages, a large number of different arthroplasty devi-
ces have been developed and are currently available and in
clinical use. The first successful arthroplasty device was the
H. Schmidt (&)  S. Midderhoff  K. Adkins  H.-J. Wilke
SB Charite disc (Depuy Spine; Raynham, MA, USA).
Institute of Orthopaedic Research and Biomechanics,
University of Ulm, Helmholtzstrasse 14, 89081 Ulm, Germany Other concepts followed such as Prodisc (Synthes; Paoli,
e-mail: h.schmidt@uni-ulm.de USA), MaverickTM (Medtronic; Minneapolis, MN, USA),

123
1696 Eur Spine J (2009) 18:16951705

Flexicore (Stryker; Kalamazoo, MA, USA), Mobidisc (RoM) and to similar facet joint forces. We investigated
(LDR medical; Troyes, France), and Activ L (Aesculap our hypotheses in a finite element (FE) analysis.
AG; Tuttlingen, Germany).
Most artificial discs consist of two metallic endplates
and a polyethylene core. The core is either separated Materials and methods
between both endplates (=unconstrained design), partially
or intrinsically connected with the lower endplate (=semi- Implants
constrained or constrained design). The Charite disc is
representative of an unconstrained and the Prodisc In the current study, we used three different implant
prothesis of a constrained design. designs: The first design was the SB Charite III disca
The unconstrained design allows the adjacent vertebrae three piece construct comprised of a biconvex core
to translate relatively to each other without any accompa- sandwiched between two concave endplatesrepresenting
nying rotation [11]. This behavior leads to a moving an unconstrained design. The second implant, also rep-
instantaneous center of rotation (ICR) and therefore, resenting an unconstrained design, was a newly developed
mimics the physiological behavior of a functional spinal disc prosthesis: Slide-Disc (Weber Instrumente GmbH;
unit, as supposed by Cunningham et al. [3]. In contrast, it Emmingen-Liptingen, Germany) (Fig. 1). Similar to the
is stated that the constrained configuration causes a fixed Charite disc, the Slide-Disc consists of two endplates
ICR. This assumption is explained in the way that the and a mobile sliding core. This core is articulated with the
segmental translation occurs concurrently with rotation upper endplate by a spherical fully congruent surface. In
given by the radius of the inner core. We disagree with this contrast to the Charite disc, the articulation between the
assumption, because the kinematical behavior of an core and the lower endplate is realized by a slightly
implanted spinal segment is not only dependent on the curved contact surface. This allows the core to move
implant itself. The articulating facet joints, which also play almost freely within the transversal plane. The developers
a major role for spinal load transmission, the remaining assume that the Slide-Disc is able to better reproduce
soft disc tissue, and the elastic behavior of the adjacent the ICRs of a healthy intact spinal segment, in particular
bony structures also influence the spinal motion behavior for axial rotation than the Charite, as a result of the
and therefore, control the location of the ICRs. freely sliding core. For axial rotation it has been found
We therefore proposed the hypothesis that different from in vitro and FE studies that the ICR migrates outside
design concepts do not lead to substantial differences in the the disc close to the facet joints under higher load con-
location of the ICRs. We furthermore hypothesized that the ditions [9, 27, 34]. The third implant design was a
different implant designs lead to a similar range of motion modification of the Slide-Disc. Here, the inner core was

Fig. 1 Left Finite element mesh


of the L4-5 lumbar spinal
motion segment with the
implanted Slide-Disc. Right
Detail view of both investigated
dynamic disc implants: Slide-
Disc (above) and SB Charite
III (below). The location of both
implants was varied by 4 mm in
both an anterior and posterior
direction. Label A indicates the
contact areas which were varied
between standard unilateral
contact (normal pressure equals
zero if separation occurs) and
bonded (contact surface is
always attached to the target
surface along the normal and
tangent directions)

123
Eur Spine J (2009) 18:16951705 1697

rigidly connected to the lower endplate representing a modulus of 2 GPa and a Poissons ratio of 0.3 [7]. A
constrained design. titanium alloy (Ti6Al4 V) with a Youngs modulus of
113.8 GPa and a Poisson ratio of 0.3 was assigned to the
Finite element modeling Slide-Disc endplates. For the inlay a Youngs modulus of
2 GPa and a Poissons ratio of 0.3 were similarly chosen.
A three-dimensional, non-linear FE-model of an intact L4- All the values for the SlideDisc were provided by Weber
5 ligamentous human lumbar motion segment was used in Instrumente.
this study (Fig. 1). This FE-model has been used previ- Two cases were simulated for the Slide-Disc. In the
ously to investigate a number of clinically relevant issues first case, the inner core was assumed to be movable (Slide-
[2832]. The model validation has been extensively docu- Disc mobile core: SD-MC), as it is intended by the com-
mented in these studies. In the following is given a brief pany. In the second case, the inner core was rigidly fixed
description of this FE-model. (bonded) at the lower metallic endplate of the implant,
The commercial software ANSYS 11.0 (ANSYS INC., simulating a constrained design (Slide-Disc immobile core:
Canonsburg, PA, USA) was utilized to perform the FE SD-IMC), as it is realized by the ProDisc.
analysis. The model consists of two vertebrae, the inter-
vertebral disc, and the seven main ligaments. The inter- Implantation
vertebral disc considers the nucleus pulposus and the
surrounding annulus fibrosus. The annulus was modeled as Disc placement procedure for both implants requires an
a composite of solid matrix with embedded fibers, which anterior surgery for implantation. The protocol involves the
are organized in seven concentric rings around the nucleus. stepwise removal of the anterior longitudinal ligament, the
Fibers and ligaments were represented by unidirectional anterior portion of the annulus, and the entire nucleus
spring elements with a non-linear force-deflection curve pulposus. Only the posterior and lateral portion of the
and no compression capabilities. The articulating facet annulus remains in place. To mimic this surgical proce-
surfaces were modeled using surface-to-surface contact dure, the elements representing these structures were
elements in combination with the penalty algorithm with a removed in the FE-model (Fig. 1).
normal contact stiffness of 200 N/mm and a friction
coefficient of zero. The facet cartilage layer was assumed Investigation of different implantation situations
to yield a thickness of 0.2 mm. The initial gap between the
cartilage layers was assumed to be 0.4 mm. The capsular First, both implants were integrated in the intervertebral
ligament was simulated by using spring elements with a space in a geometrically central position. Subsequently, the
non-linear force-deflection curve forming a ring around the locations of both implants were varied by 4 mm in both an
articular contact; they linked the borders of the inferior anterior and posterior directions (Fig. 1).
vertebra superior articular process to the borders of the The contact condition between the metallic endplates
superior vertebra inferior articular process. and the adjacent bony structures was verified between
perfect bond and standard unilateral contact (indicated by
Implant modeling the label A in Fig. 1) with a friction coefficient of 1. While
a bonded contact represents a perfect bone osteo-integra-
Both implants were meshed using eight-node isoparametric tion between the metallic endplates and the adjacent ver-
solid elements. The spiked endplate surfaces of both tebrae, the standard contact investigates the treated
implants were simplified to a flat surface. The Slide-Disc segment right after the surgical procedures. The latter
and the Charite disc are available in different sizes. For additionally represents the poorest osteo-integration, which
our simulations, we used the Slide-Disc type III with a implies that a gap between implant and adjacent bony
height of 11 mm and a lordotic angle of 4 and the Charite endplates may exist under certain loads.
III with an approximate height of 13 mm and a lordotic
angle of 5. Loading and boundary conditions
A standard unilateral contact was assumed at the arti-
cular surface between the core and the concave metallic The inferior endplate of the lower vertebral body was
endplates. A friction coefficient of 0.02 was chosen for rigidly fixed. An axial compression preload of 1,000 N was
both implants [7]. For the Charite endplates, a chrome- applied simulating upper body weight plus muscle forces.
cobalt alloy was assumed with a Youngs modulus of This load was applied using the follower load technique
300 GPa and a Poissons ratio of 0.27 [7]. The inlays were [19], which is thought to have a similar stabilization effect
represented by a polyethylene core (ultra high molecular as that of the local muscles [23]. The load follows the
weight polyethylene (UHMWPE)) with a Youngs curvature of the spine through the proximities of the ICRs

123
1698 Eur Spine J (2009) 18:16951705

and therefore, avoids the generation of additional larger intact situation, we defined a standard evaluation scale for
moments. This was realized by using connector elements all result parameters. This scale was defined as follows: A
which were spanned between the centers of both vertebral difference between 0 and 15% indicated a slight difference,
bodies. Subsequently, the spinal segment was loaded with between 16 and 30% indicated a moderate difference, and
unconstrained moments of 7.5 Nm in the sagittal, lateral, from 31% to higher values indicated a large difference. All
and axial directions simulating flexion, extension, lateral results were adapted according to this scale.
bending, and axial rotation.
Range of motion
Data analysis
Except for extension, the models predicted only small and
1. Range of motion moderate inter-implant differences, particularly for both
2. Facet joint force and pressure distribution in the facet Slide-Disc designs, where the maximal percentage differ-
surfaces ence was 15% (Fig. 2). For extension, a large difference of
about 45% was calculated between both Slide-Disc
The resulting forces for each contact element in one designs and the Charite.
facet joint were added together to give a total facet force In flexion, all implants led to an averaged decreased
(FF). In each facet surface the pressure distribution was RoM of 22% compared to the intact state when the artificial
displayed independent of the calculated maximum value discs are centrally placed. In extension, only the Charite
and divided into a pressure area greater than 70%, greater caused a strong increase of 52%, while both Slide-Disc
than 40%, and less than 40% of the maximum value. designs showed a good representation of the intact state. In
3. Center of rotation lateral bending, the different Slide-Disc designs led to a
slight increase of the RoM. The RoM predicted by the
The ICR was calculated according to the Reuleaux Charite is 36% higher than for the intact segment. In axial
method [21]. This was performed by intersecting the per- rotation, all three implant designs restored almost to the
pendicular vectors from the midpoint of the translation intact state.
vectors for two nodes in the upper vertebral body. The In flexion, placing the implants more posteriorly led to
applied moments were incrementally increased from zero an increased RoM and a better representation of the intact
to the predetermined maximum value of 7.5 Nm, which
were attained in ten equally sized incremental steps. The
ICRs were evaluated between two consecutive steps. In the
corresponding figures, the ICRs were only shown for three
different load magnitudes: 1.5, 3.75, and 7.5 Nm in relation
to their previous step, i.e., load magnitudes of: 0.75, 3.00,
and 6.75 Nm, respectively.

Model verification

Before undertaking the present study we performed mesh


convergency tests with both disc prostheses. In previous
investigations we also conducted convergency tests with
the facet joints. As critical result parameters for both cases
we used the RoM and the ICRs, and for the facet joints we
additionally used the contact pressure, the pressure distri-
bution, and the contact forces. The element edge length
was reduced until the percentage difference of these critical
results between two consecutive mesh densities was less
than 2%.

Results Fig. 2 Influence of implant position in an antero-posterior direction


on range of motion in flexion, extension, lateral bending, and axial
rotation. The dashed line indicates the range of motion of the intact
In order to better interpret the differences between the model. MC Slide disc with mobile core and IMC Slide disc with
individual implant designs and between the implanted and immobile core

123
Eur Spine J (2009) 18:16951705 1699

state. In extension only the Charite was influenced by the inter-implant differences up to 15% were calculated. The
implant position. Similar to flexion, a more posteriorly artificial disc models caused an increase of 46% for the SD-
placed implant led to a better representation of the RoM MC, 32% for the Charite, and 27% for the SD-IMC
calculated for the intact state. A similar behavior was seen compared to the intact state.
for lateral bending. While both Slide-Disc designs Changing the position of the artificial discs affected the
showed only a slight influence on the implant position, the facet joint forces for all investigated loading cases differ-
Charite predicted a strong influence. In axial rotation, the ently. The SD-MC showed a slight effect for flexion,
implant position did not greatly influence the RoM. extension, and axial rotation. In lateral bending, strong
differences up to 36% of the resulting forces were calcu-
Facet joint forces lated. Here, the forces decreased when shifting the implant
posteriorly.
In flexion, the facet joints remained unloaded for the intact The influence of different implant positions was still
model (Fig. 3). In contrast, the presence of any of the disc stronger emphasized by the Charite and the SD-IMC,
prostheses led to high facet forces, especially for the especially in extension. For the Charite we calculated
Charite. Here, the calculated force was 138 N when the completely unloaded facets when shifting the implant
implant is centrally placed. Up to 32% inter-implant dif- posteriorly, whereas, a more anterior position increased the
ferences were calculated. In extension, much larger inter- facet forces by 240%. The SD-IMC showed the largest
implant differences were found than was indicated under effect in flexion. Here the implant led to a force increase of
flexion. The models predicted strongly increased forces for 110% when shifting the disc from the most anterior to the
the SD-MC (60%) and a strong decrease of 70% for the most posterior position.
Charite compared to the intact state. Totally unloaded
facet joints were predicted for the SD-IMC. In lateral Pressure distribution in the facet surfaces
bending, strong inter-implant differences were also calcu-
lated. Compared to the intact situation the force decreased In flexion and in extension, both the left and the right facet
by 4% for the Charite and 83% for the SD-IMC. In joints were equally loaded: in flexion more in the central
contrast, the facet joint forces strongly increased by 214% region (Fig. 4) and in extension more in the inferior tip of
for the SD-MC prosthesis. In axial rotation, only small the facet (Fig. 5). In lateral bending, the facet joint was
loaded on the ipsilateral side of bending for the intact
model. The contralateral side remained almost unloaded
(Fig. 6). The different artificial discs also caused loaded
facets on the contralateral side. However, the calculated
forces were always smaller compared to the ipsilateral side.
For the intact situation and for all disc implants, left axial
rotation led to an increase of contact forces on the right
facet joint, while the left facet remained nearly unloaded
(Fig. 7). Only for small torsional moments, forces for the
left facet were calculated. These forces were mainly pro-
duced by the applied preload.

Center of rotation

In flexion, the intact model predicted ICRs in the mid-


height of the disc (Fig. 4). The ICR was located in the
center of the disc for small moments (up to 1.5 Nm). With
increasing moment (up to 7.5 Nm) the ICR moved slightly
toward the anterior direction. The location of the ICR is
only slightly affected by the presence of any of the
investigated disc prostheses. For small moments they are
Fig. 3 Influence of implant position in an antero-posterior direction shifted slightly caudally close to the superior endplate of
on acting forces in the facet joints in flexion, extension, lateral the lower vertebral body and slightly altered their positions
bending, and axial rotation. The dashed line indicates the facet joint
with increasing flexion, except for the SD-MC model
forces of the intact model. In flexion, the facets remained unloaded.
MC Slide disc with mobile core and IMC Slide disc with immobile where it moves somewhat cranially close to the inferior
core endplate of the upper vertebral body.

123
1700 Eur Spine J (2009) 18:16951705

Fig. 4 Predicted location of the


center of rotations (ICR)
(above) and the resulting
pressure distribution in the
inferior facet of L4 (below) with
increasing flexion. The ICRs are
shown for three different
moments (M): 1.5, 3.75, and
7.5 Nm in relation to the
moment resulting from the
previous sub-step: 0.75, 3.00,
and 6.75 Nm, respectively. The
dark points show the current
ICR for each of the three
moment intervals and the light
points show the ICR always for
the first moment interval (0.75
1.5) to illustrate how the ICRs
migrate with increasing load.
The maximum forces for each
facet joint and the pressure
distribution are shown for
moments of 1.5, 3.75, and
7.5 Nm. MC Slide disc with
mobile core and IMC Slide disc
with immobile core

In extension, the ICRs of the intact model were also moments. With ongoing extension moment, the ICRs
located in the center of the disc for small moments of migrated towards the lower vertebral body and moved
1.5 Nm (Fig. 5). With increasing moment, the ICR slightly in the posterior direction.
migrated slightly posteriorly. Under the maximum moment In lateral bending, the ICRs of the intact spinal model
of 7.5 Nm the ICR was calculated as being in the posterior were calculated as being almost in the center of the disc for
nucleus. Similar to flexion, the differences in ICR locations small moments and migrated towards the side of the
between the investigated prostheses are small. The ICRs bending with increasing moment (Fig. 6). Under the
were located almost in the center of the disc for small maximum moment of 7.5 Nm, the ICR was located on

Fig. 5 Predicted locations of


the center of rotations (ICR) and
the resulting pressure
distribution in the inferior facet
of L4 with increasing extension.
Explanations for the ICRs see
Fig. 4. The maximum forces for
each facet joint and the pressure
distribution are shown for
moments of 1.5, 3.75, and
7.5 Nm. MC Slide disc with
mobile core and IMC Slide disc
with immobile core, FF Facet
joint forces

123
Eur Spine J (2009) 18:16951705 1701

Fig. 6 Predicted locations of


the center of rotations (ICR) and
the resulting pressure
distribution in the inferior facet
of L4 with increasing right
lateral bending. Explanations
for the ICRs see Fig. 4. The
maximum forces for each facet
joint and the pressure
distribution are shown for
moments of 1.5, 3.75, and
7.5 Nm. MC Slide disc with
mobile core and IMC Slide disc
with immobile core, FF Facet
joint forces

the right side of the disc, near the inferior endplate of L4. In axial rotation, the ICRs were found close to the center
For small moments, the artificial disc models predicted of the disc for small moments (Fig. 7). With increasing
ICRs also almost in the center of the disc. With increasing moment, the ICR migrated to the posterior direction. With
moment, the SD-MC showed a larger upward movement a moment of 7.5 Nm the model predicted an ICR location
towards the center of the L4 vertebral body. The Charite outside of the disc, close to the compressed facet joint. Up
and the SD-IMC altered slightly their locations in the to a moment of approximately 5 Nm all three artificial
direction of the lower vertebral body. discs showed similar tendencies: with increasing moment,

Fig. 7 Predicted locations of


the center of rotations (ICR) and
the resulting pressure
distribution in the inferior facet
of L4 with increasing left axial
rotation. Explanations for the
ICRs see Fig. 4. The maximum
forces for each facet joint and
the pressure distribution are
shown for moments of 1.5, 3.75,
and 7.5 Nm. MC Slide disc with
mobile core and IMC Slide disc
with immobile core, FF Facet
joint forces

123
1702 Eur Spine J (2009) 18:16951705

the location of ICRs migrated to the postero-lateral direc- The results of our FE study supported our hypothesis in
tion of the disc. However, for the maximum moment of that the different mobile implants do not lead to substantial
7.5 Nm, only the SD-MC prosthesis led to an ICR location differences in the location of the ICRs. Except in axial
outside the disc, close to the compressed facet joint. rotation, the ICRs of all three implant designs mimicked
In flexion and extension, placing the implant ventrally almost the non-treated intact situation. The implant with
showed a shift of the ICR in the anterior location, while a the constrained core did not cause a fixed ICR. This result
more posterior implant position caused the ICR to migrate can be explained as follows: The artificial disc should not
slightly posteriorly. For both load directions, the SD-IMC be considered separately but with the surrounding elastic
led to ICRs which were located in the region of the adja- structure. Our FE results showed that the adjacent bony
cent lower vertebral body. In lateral bending, the ICRs endplates may strongly deform under certain loads. At the
were calculated to be more in the adjacent lower bony same time high forces in the facet joints occurred, espe-
structures when shifting the implant anteriorly as well as cially in axial rotation (Fig. 3). These high facet forces in
posteriorly. The ICR pattern caused by the SD-IMC was combination with the endplate deformation allow the
spread over a much larger area compared to the intact adjacent vertebrae to translate relatively to each other and
situation. Axial rotation led to a shift of the ICR slightly in therefore, lead to a moving ICR.
the anterior location. In axial rotation, the different implant designs showed a
slightly different behavior. While the Charite and SD-
Perfect bond and standard unilateral contact IMC prostheses predicted ICRs inside the disc, the SD-MC
prosthesis caused an ICR location outside the disc, close to
Changing the contact condition between the metallic end- the compressed facet joint and therefore, better mimicked
plates and the adjacent bony endplates did not strongly the intact situation. This is caused by the sliding core.
affect the results, neither for the central nor for the different Under small moments no differences in ICRs were calcu-
implant positions. A maximum difference of 2% for the lated between SD-MC and SD-IMC (Fig. 7). Under these
facet joint forces was calculated. small moments, the inner core of the SD-MC prosthesis
moved only marginally within the implant (less than
0.1 mm). From approximately 5 Nm upward to 7.5 Nm,
Discussion the SD-MC led to a sudden change of the resulting ICR
locations. This moment increase caused the inner core to
The biomechanical behavior of dynamic non-fusion migrate 0.8 mm to the ipsi-lateral direction. The movement
implant systems were often tested in experimental in vitro of the core is the crucial factor, which caused an ICR
studies [10, 12, 13, 17, 18] and evaluated by analyzing migration to the compressed facet joint.
clinical radiographs [14]. These studies were principally In flexion, extension, and lateral bending we calculated a
focused on the estimation of remarkable changes in RoM at cranial-caudal migration of the ICRs for the SD-MC. In the
the treated and adjacent segments. The authors of these first instance such a result would seem unexpected, because
studies showed that the various disc prostheses generally an implant which is restricted to a transversal movement of
preserve the mobility at near physiological level, for both the inner core should only allow the migration of the ICRs
constrained and unconstrained designs, which was con- in same direction. The cranial-caudal migration can,
firmed by our FE simulations. However, the RoM alone is however, be induced by two different parameters. One
not sufficient to evaluate the spinal stability. Therefore, point is that the lower metallic endplate is slightly con-
authors used the ICR as an additional parameter to analyze cavely curved allowing the core to move slightly vertically
the motion behavior of the treated segment [5, 6, 20]. In a during transverse motion. A second point is that the adja-
FE analysis, we estimated the measurement error, which is cent bony structures are deformed during loading so that
always given when ICRs are experimentally determined the relative distance of the core in relation to the bony
[27]. In this study we showed that a small domain of input structures is altered.
data led to severe changes on the position of the ICR. This Both results, a posterior migration of the ICRs towards
implies that simply alignment of two X-rays generated in the facet joints as well as the constrained core did not lead
different postures, as done in clinical practice, is not suf- to a fixed ICR, seem to be caused by a lift-off of the
ficient to determine the ICR. Indeed, a high accuracy is metallic endplate from the core. However, this phenome-
needed because of small movements which in particular non did not occur in all load situations. The pressure dis-
occur when the spine is exposed to axial rotation. Data tribution of the core and the metallic endplates was
evaluations using FE models with a high validity and analyzed by the authors and both parts were always found
predictability are a suitable tool for evaluating ICR loca- to be under stress, which is an indication that this lift-off
tions, since there are no further measurement errors. would not occur. Such a lift-off was found in a previous

123
Eur Spine J (2009) 18:16951705 1703

study by us (unpublished data), however, for only one strong facet forces. The same inconsistencies can be seen
specific load direction: pure moment in extension without for lateral bending and axial rotation.
an axial compression preload. In this case, the facet joints One reason for such a large variation in the results may
led to an opening of the implant. We made additional have arisen through the use of FE models based on dif-
investigations in which we defined a no separation con- ferent geometries. Consequently, slight changes in the gap
tact condition between the metallic endplates and the inner distance, the degree of curvature or the facet orientation
core. The results were the same as that found in our study. can lead to completely different results. In our investiga-
In addition, no tensile stresses between the metallic end- tions the gap size was set at 0.4 mm. An increase in the gap
plates and the inner core were calculated, which would of only 0.1 mm led to an average decrease of 45% for all
indicate a separation between both contact pairs. the three implants. A probabilistic FE analysis by Rohl-
In contrast to the RoM and ICRs, the facet joint force mann et al. [22] confirmed our supposition. The authors
strongly depends on the implant design. For all load investigated how, for example, the facet forces are influ-
cases, our model predicted higher facet joint forces for the enced by parameters such as the implant position, the
implants with an unconstrained core (Charite and SD- presence of scar tissue or the gap size in the facet joints
MC) compared to an implant with a constrained core when implanting the ProDisc. In extension, Rohlmann
(SD-IMC). This result can be explained by the resulting et al. reported a zero force for the instrumented segment in
forces passing through the implant itself. An uncon- 70% of all investigated cases (1,000 randomly generated
strained design transfers only normal forces from the FE models), which is in agreement with our predictions.
upper to the lower vertebral body. In contrast, a con- Rohlmann et al. suggested that the resulting forces are
strained design additionally transfers shear forces through strongly dependent on the gap size. Furthermore, the
the implant and therefore, is stronger loaded than the authors found that in flexion the ProDisc caused facet
unconstrained design. Our FE model predicted implant forces in 37% in the instrumented segment, which is also
forces in flexion, extension, lateral bending, and axial confirmed by our predictions.
rotation of 623, 1,013, 785, and 852 N for the SD-MC, The described methodology has some assumptions and
and forces of 985, 1,130, 945, and 963 N for the SD-IMC limitations. In the current study, we decided on a follower
model, respectively. On average, the constrained implant load magnitude of 1,000 N, which contrasts with our pre-
is 19% higher loaded than an unconstrained implant. This vious studies in which we only simulated a follower load of
force difference is compensated by the articulating facet 500 N. The reason for this is that Rohlmann et al. [24]
joints meaning that higher implant forces resulted in suggested in a FE study that a load of 500 N was not
lower facet forces. sufficient for simulating flexion, since it did not take into
The resulting forces in the facet joints have been dis- account the global muscle forces. However, when the
cussed at length in the literature. While Zander et al. [35] authors increased the load up to 1,175 N, they obtained
calculated a strong increase of the acting forces in the results that were comparable with in vivo data. In com-
facets in extension for the Charite, Grauer et al. [8], Goel parison to 500 N, this increase had only a minor influence
et al. [7], and Moumene and Geisler [16] found a slight to on the RoM and a neglectable effect on the resulting facet
moderate decrease in these forces for the same implant. forces. For extension when they simulate a follower load of
These opposing findings were intensified by our results. For 500 N they were already able to calculate an acceptable
the Charite we predicted even less forces, down to intradiscal pressure. However, the calculated facet joint
unloaded conditions. Whereas, with the Slide-Disc-MC forces under this 500 N were much lower in comparison to
we determined a moderate increase in the resulting forces. the simulation when the global muscle forces were con-
In the case of the ProDisc (which corresponds to the sidered, which Rohlmann et al. indicated as being a more
Slide-Disc-IMC in our study) these inconsistencies were physiological load application. A more realistic facet
similarly found. For this implant, both Moumene and joint force was much more important in our study than the
Geisler [16], and Zander et al. [35] predicted a slight intradiscal pressure, because the facet joint forces have a
decrease in the acting forces, whereas, the study results of stronger influence on our results than the disc pressure,
Rundell et al. [25] as well as our results indicated either which are moreover absent in our different implant models.
almost or completely unloaded facets. In flexion, Zander The follower load was applied to the spinal segment by
et al. [35] found no forces in the facets for intact as well as using connector elements which passed almost the ICRs
for all the investigated implants. In contrast, Grauer et al. (center of the disc) found for small moment applications
(2006) indicated that the Charite led to a slight force in and therefore, avoid the generation of additional larger
flexion, although more details were not provided. The moments. However, the ICRs are not fixed in space and
study results of Rundell et al. [25] for the ProDisc toge- changed with motion. This may produce additional
ther with our results for all three implant designs predicted moments when further bending or torsional moments are

123
1704 Eur Spine J (2009) 18:16951705

applied. The current study showed that the ICRs are mostly together with our results, we could speculate that these forces
located in the center of the disc, except for axial rotation. may be more dependent on the individual spine geometry
Axial rotation produced the largest changes in the ICRs rather than a specific implant design. In order to be able to
during loading, meaning that the ICRs migrate from the make a suggestion, it is necessary to know which load
disc center to the corresponding compressed facet joint. directions of the spine and which parameter is most signifi-
This larger distance generates additional moments for the cant. If it is, for example, flexion and the RoM, it does not
motion segment, which cannot be neglected. make any difference which implant design is used. If it is in
We used a fixed combination of material properties for contrast axial rotation and the ICR, our results suggest the
the different artificial disc implants as used in a prior FE need of a mobile core design.
study of Goel et al. [7] and defined by Weber Instrumente.
The exact values of the Youngs modulus and of the Acknowledgments This study was financially supported by the
German Research Foundation (Wi 1352/14-1). The authors would
Poissons ratio of the titanium alloy and the UHMWPE like to thank Depuy Spine (Raynham, MA, USA) and Weber
were neither published by the company Depuy Spine nor Instrumente GmbH (Emmingen-Liptingen, Germany) for the supply
known by Weber Instrumente and therefore, may slightly of their implants. We also would like to thank Dr. Robert Blakytny for
differ from the values used in the current study. Different his help in editing the manuscript.
values might have a small influence on our findings.
The form, such as the radius and the size of the inlay, of
References
an actual fixed core design was not considered in the cur-
rent study and therefore a specific functional goal of this 1. Auerbach JD, Wills BP, McIntosh TC, Balderston RA (2007)
design is lacking. The advantage of this approach lies in the Evaluation of spinal kinematics following lumbar total disc
fact that the actual functional difference between a fixed replacement and circumferential fusion using in vivo fluoroscopy.
and a mobile core can be determined independently of Spine 32(5):527536
2. Bertagnoli R, Kumar S (2002) Indications for full prosthetic disc
other inter-implant differences. In our study the only dif- arthroplasty: a correlation of clinical outcome against a variety of
ference between mobile and immobile core design was a indications. Eur Spine J 11 Suppl 2:S131S136
single condition in the contact definition (standard vs. 3. Cunningham BW, Gordon JD, Dmitriev AE, Hu N, McAfee PC
bonded). If we had used the ProDisc prosthesis as an (2003) Biomechanical evaluation of total disc replacement
arthroplasty: an in vitro human cadaveric model. Spine
example of a fixed-core design, other parameters would 28(20):S110S117
have had an influence on our results and therefore would 4. Cunningham BW, McAfee PC, Geisler FH, Holsapple G, Adams
have obscured the differences that we found. K, Blumenthal SL, Guyer RD, Cappuccino A, Regan JJ, Fedder
Our controversial findings in facet forces raise the IL, Tortolani PJ (2008) Distribution of in vivo and in vitro range
of motion following 1-level arthroplasty with the CHARITE
question of the reliability of deterministic FE studies, artificial disc compared with fusion. J Neurosurg Spine 8(1):712
which are characterized by a fixed configuration of material 5. Gertzbein SD, Seligman J, Holtby R, Chan KH, Kapasouri A,
and geometrical properties. The results generated by this Tile M, Cruickshank B (1985) Centrode patterns and segmental
type of model should always be interpreted with caution, instability in degenerative disc disease. Spine 10(3):257261
6. Gertzbein SD, Seligman J, Holtby R, Chan KW, Ogston N,
and certain results, such as specific pressure values at the Kapasouri A, Tile M (1986) Centrode characteristics of the
facet surfaces should not be interpreted in a clinical con- lumbar spine as a function of segmental instability. Clin Orthop
text; they simply provide a trend. Deterministic models are, Relat Res 208:4851
however, legitimate when directly comparing the bio- 7. Goel VK, Grauer JN, Patel T, Biyani A, Sairyo K, Vishnubhotla
S, Matyas A, Cowgill I, Shaw M, Long R, Dick D, Panjabi MM,
mechanical outcome of two different disc prosthesis, with Serhan H (2005) Effects of charite artificial disc on the implanted
the qualification that a small modification of the geometry and adjacent spinal segments mechanics using a hybrid testing
could produce the opposite results. protocol. Spine 30(24):27552764
8. Grauer JN, Biyani A, Faizan A, Kiapour A, Sairyo K, Ivanov A,
Ebraheim NA, Patel T, Goel VK (2006) Biomechanics of two-level
Charite artificial disc placement in comparison to fusion plus sin-
Conclusions gle-level disc placement combination. Spine J 6(6):659666
9. Haher TR, OBrien M, Felmly WT, Welin D, Perrier G, Choueka
Total disc arthroplasty inherently alters the kinematics at J, Devlin V, Vassiliou A, Chow G (1992) Instantaneous axis of
rotation as a function of the three columns of the spine. Spine
implant level for all clinical used prostheses. Compared to 17(6 Suppl):S149S154
these alterations, our finite element model predicted that 10. Hitchon PW, Eichholz K, Barry C, Rubenbauer P, Ingalhalikar A,
inter-implant differences are small regarding the RoM and Nakamura S, Follett K, Lim TH, Torner J (2005) Biomechanical
ICR. In contrast, facet joint forces are strongly dependent on studies of an artificial disc implant in the human cadaveric spine.
J Neurosurg Spine 2(3):339343
the implant design. In contrast, facet forces appeared to be 11. Huang RC, Girardi FP, Cammisa FP Jr, Wright TM (2003) The
strongly dependent on the implant design. However, due to implications of constraint in lumbar total disc replacement.
the great variability in facet forces reported in the literature, J Spinal Disord Tech 16(4):412417

123
Eur Spine J (2009) 18:16951705 1705

12. Kafer W, Cakir B, Mattes T, Reichel H (2008) Orthopaedic spine 25. Rundell SA, Auerbach JD, Balderston RA, Kurtz SM (2008)
surgery: an instruction course textbook. Steinkopff, Springer, Total disc replacement positioning affects facet contact forces
New York, pp 191199. ISBN:978-3-7985-1828-5 and vertebral body strains. Spine 33(23):25102517
13. Kotani Y, Cunningham BW, Abumi K, Dmitriev AE, Hu N, Ito 26. Schlegel JD, Smith JA, Schleusener RL (1996) Lumbar motion
M, Shikinami Y, McAfee PC, Minami A (2006) Multidirectional segment pathology adjacent to thoracolumbar, lumbar, and lum-
flexibility analysis of anterior and posterior lumbar artificial disc bosacral fusions. Spine 21(8):970981
reconstruction: in vitro human cadaveric spine model. Eur Spine J 27. Schmidt H, Heuer F, Claes L, Wilke HJ (2008) The relation
15(10):15111520 between the instantaneous center of rotation and facet joint for-
14. Lee CK (1988) Accelerated degeneration of the segment adjacent cesa finite element analysis. Clin Biomech (Bristol, Avon)
to a lumbar fusion. Spine 13(3):375377 23(3):270278
15. Mayer HM, Korge A (2002) Non-fusion technology in degener- 28. Schmidt H, Heuer F, Drumm J, Klezl Z, Claes L, Wilke HJ
ative lumbar spinal disorders: facts, questions, challenges. Eur (2007) Application of a calibration method provides more real-
Spine J 11(Suppl 2):S85S91 istic results for a finite element model of a lumbar spinal segment.
16. Moumene M, Geisler FH (2007) Comparison of biomechanical Clin Biomech (Bristol, Avon) 22(4):377384
function at ideal and varied surgical placement for two lumbar 29. Schmidt H, Heuer F, Simon U, Kettler A, Rohlmann A, Claes L,
artificial disc implant designs: mobile-core versus fixed-core. Wilke HJ (2006) Application of a new calibration method for a
Spine 32(17):18401851 three-dimensional finite element model of a human lumbar
17. Panjabi M, Henderson G, Abjornson C, Yue J (2007) Multidi- annulus fibrosus. Clin Biomech (Bristol, Avon) 21(4):337344
rectional testing of one- and two-level ProDisc-L versus simu- 30. Schmidt H, Heuer F, Wilke HJ (2008) Interaction between finite
lated fusions. Spine 32(12):13111319 helical axes and facet joint forces under combined loading. Spine
18. Panjabi M, Malcolmson G, Teng E, Tominaga Y, Henderson G, 33(25):27412748
Serhan H (2007) Hybrid testing of lumbar CHARITE discs versus 31. Schmidt H, Heuer F, Wilke HJ (2009) Which axial and bending
fusions. Spine 32(9):959966 discussion 967 stiffnesses of posterior implants are required to design a flexible
19. Patwardhan AG, Havey RM, Meade KP, Lee B, Dunlap B (1999) lumbar stabilization system? J Biomech 42(1):4854
A follower load increases the load-carrying capacity of the 32. Schmidt H, Kettler A, Heuer F, Simon U, Claes L, Wilke HJ
lumbar spine in compression. Spine 24(10):10031009 (2007) Intradiscal pressure, shear strain, and fiber strain in the
20. Pearcy MJ, Bogduk N (1988) Instantaneous axes of rotation of intervertebral disc under combined loading. Spine 32(7):748755
the lumbar intervertebral joints. Spine 13(9):10331041 33. Shah RR, Mohammed S, Saifuddin A, Taylor BA (2003)
21. Reuleaux F (1876) The kinematics of machinery: outline of a Radiologic evaluation of adjacent superior segment facet joint
theory of machines. Macmillan, London violation following transpedicular instrumentation of the lumbar
22. Rohlmann A, Mann A, Zander T, Bergmann G (2008) Effect of spine. Spine 28(3):272275
an artificial disc on lumbar spine biomechanics: a probabilistic 34. Shirazi-Adl A, Ahmed AM, Shrivastava SC (1986) Mechanical
finite element study. Eur Spine J response of a lumbar motion segment in axial torque alone and
23. Rohlmann A, Neller S, Claes L, Bergmann G, Wilke HJ (2001) combined with compression. Spine 11(9):914927
Influence of a follower load on intradiscal pressure and inter- 35. Zander T, Rohlmann A, Bergmann G (2009) Influence of dif-
segmental rotation of the lumbar spine. Spine 26(24):E557E561 ferent artificial disc kinematics on spine biomechanics. Clin
24. Rohlmann A, Zander T, Rao M, Bergmann G (2009) Realistic Biomech (Bristol, Avon) 24(2):135142
loading conditions for upper body bending. J Biomech
42(7):884890

123

Das könnte Ihnen auch gefallen