Designation: D 2149 97
Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1
D 2149 97
long. Precious metal alloys such as type B silver-magnesium- electrodes may be useful. Additional information on the
nickel have better overall properties but require special heat suitability of various electrode systems may be found in Test
treating. Methods D 150.
5.3.2 The insulators may be aluminum oxide, beryllium
oxide, or polytetrafluoroethylene. 7. Sampling
5.3.3 Use electrodes 50 mm in diameter and at least 5 mm 7.1 See ASTM standards for specific materials.
thick, with sharp corners. Maintain electrode parallellism to
within 0.01 mm. 8. Test Specimen
5.3.4 Select a length and cross-section for the lower tube so 8.1 Use a disk test specimen with a diameter of 40.006 0.01
that the temperature of each insulator does not exceed 100C mm and a thickness of 2 to 3 mm. Finish the surfaces to 1.8 m
when the oven is at 500C. Select a length and crosssection for or better and maintain parallel surfaces to within 0.01 mm. The
the upper tube so that the drive nut can be touched with the samples should be free of bubbles and other defects.
operators fingers (keep the drive nut less than 60C) when the
oven is at 500C. 9. Standard Test Frequencies
5.3.5 Use a micrometer or dial gage with a precision of 9.1 Unless otherwise specified, make measurements at one
0.005 mm to determine electrode separation and to monitor or more of the following frequencies:
specimen expansion. 60 Hz 100 000 Hz
100 Hz 1 MHz
6. Electrodes 400 Hz 10 MHz
1000 Hz
6.1 Prior to measurement, apply conducting film or foil
electrodes to both flat surfaces of the specimen. (The specimen Common test frequencies are 60 Hz, 1000 Hz, and 1 MHz.
thickness should be determined before applying electrodes.)
Silver paint, tin or tin-lead foil, or evaporated metal electrodes 10. Temperature Control
have ranges of usefulness. Evaporated metal electrodes are the 10.1 Take measurements at frequent temperature intervals
most suitable. When the specimen is porous sprayed-on metal (not to exceed 20C), until the required temperature range has
2
D 2149 97
been traversed. Reduce the temperature to the lowest required Cs 5 DC 1 C v (2)
test temperature and leave until equilibrium has been achieved. Ds 5 C t/Cs ~Di 2 Dv! (3)
Determine equilibrium by clamping a specimen between the
holder electrodes and balancing or peaking the measuring where:
circuit until no change takes place between balances made 2 Co = capacitance of the specimen holder with the specimen
min apart. After the required measurements have been made at out,
the lowest test temperature increase the temperature at the rate Ci = capacitance of the electrodes set at the average
of 2 6 0.5C/min to the next test temperature. Follow this measured thickness of the specimen (Note 2),
procedure for achieving the test temperature until the required Cv = equivalent geometric vacuum capacitance of the
temperature range has been traversed. Take measurements at specimen,
Ct = total capacitance at the unknown terminals of the
approximately the same test temperatures as the temperature is
measuring circuit,
increasing and as the temperature is decreasing. Measurements
Di = dissipation factor of the measuring circuit as indicated
as temperature is being increased and decreased are necessary
by the measuring circuit when the specimen is be-
to guard against possible hysteresis in electrical properties due
tween the electrodes, and
to such factors as moisture and chemical change. Do = dissipation factor of the measuring circuit as indicated
11. Conditioning by the measuring circuit when the circuit has been
rebalanced with the specimen removed from the
11.1 Prior to applying electrodes condition the specimens at
electrodes.
23 6 1C and 50 6 2 % relative humidity for a minimum of 40
h. Carry out room-temperature tests in the Standard Laboratory
Atmosphere of 23 6 1C and 50 6 2 % relative humidity. NOTE 2If the secondary electrodes are quite thin and the maximum
thickness of the specimen is close to the average thickness, this setting can
12. Procedure A (Using Micrometer Electrode) be considered the same as the micrometer reading when the specimen is
clamped between the electrodes.
12.1 Refer to Test Methods D 150. Center the specimen
between the electrodes and rotate the drive nut until the friction 14.2 Procedure BCalculate the capacitance of the speci-
is felt to suddenly decrease. Read this micrometer setting and men as follows:
check it against the setting at which the friction first increases Cs 5 DC 1 C RT 2 CT 1 Cv (4)
on increasing the electrode spacing. Balance or peak the
measuring circuit. Open the electrodes and remove the speci- where:
men. Then restore the balance of the measuring apparatus CRT = capacitance of the precision capacitor at room
without changing its capacitance setting by reducing the temperature when the measuring circuit is bal-
spacing between the electrodes and adjusting the measuring anced, and
circuit to balance the loss component. Note the new dissipation CT = capacitance of the precision capacitor at a tempera-
factor and micrometer setting, etc. ture test point when the measuring circuit is bal-
12.2 At each test temperature and each required frequency anced.
determine the capacitance and dissipation factor of each 14.3 Calculate the dissipation factor as in Procedure A (Eq
specimen. 3).
3
D 2149 97
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.