Sie sind auf Seite 1von 1

Accessed from 10.6.1.

1 by schering on Fri Jun 12 00:31:37 EDT 2015

2064 Adenosine / Official Monographs USP 38

CS = concentration of USP Adenosine RS in the Acceptance criteria: 19.5%23.5% of oxygen by


Standard solution (mg/mL) volume
CU = nominal concentration of adenosine in the
Sample solution (mg/mL) IMPURITIES
Acceptance criteria: 90.0%110.0% See Impurities Testing in Medical Gases Assay 413. The de-
tector tubes called for in the following tests are listed in
IMPURITIES Reagents, Indicators, and Solutions.
ORGANIC IMPURITIES If the label indicates that Medical Air is a synthetic mixture
Mobile phase, System suitability solution, Standard of oxygen and nitrogen, and where oxygen complies to
solution, Chromatographic system, and System suit- Oxygen USP and Nitrogen complies to Nitrogen NF, then
ability: Proceed as directed in the Assay. the Impurities tests are not required.
Sample solution: Nominally 0.3 mg/mL of adenosine LIMIT OF CARBON DIOXIDE
from a volume of Injection, in water Sample: Detector tube manufacturers recommended
Analysis volume 5% of Medical Air
Sample: Sample solution Analysis: Pass the Sample through a carbon dioxide de-
Calculate the percentage of each impurity in the vol- tector tube at the rate specified for the tube by the
ume of Injection taken: detector tube manufacturer.
Acceptance criteria: NMT 500 ppm
Result = (rU/rT) 100 LIMIT OF CARBON MONOXIDE
Sample: Detector tube manufacturers recommended
rU = peak response for each impurity volume 5% of Medical Air
rT = sum of the responses of all of the peaks Analysis: Pass the Sample through a carbon monoxide
Acceptance criteria detector tube at the rate specified for the tube by the
Any individual impurity: NMT 1.0% detector tube manufacturer.
Total impurities: NMT 1.5% Acceptance criteria: NMT 10 ppm
LIMIT OF SULFUR DIOXIDE
SPECIFIC TESTS Sample: Detector tube manufacturers recommended
PH 791: 4.57.5 volume 5% of Medical Air
PARTICULATE MATTER IN INJECTIONS 788: It meets the re- Analysis: Pass the Sample through a sulfur dioxide de-
quirements for small-volume injections. tector tube at the rate specified for the tube by the
BACTERIAL ENDOTOXINS TEST 85: When the product is detector tube manufacturer.
used for rapid intravenous injection, it contains NMT Acceptance criteria: NMT 5 ppm
11.62 USP Endotoxin Units/mg of adenosine. When the LIMIT OF NITRIC OXIDE AND NITROGEN DIOXIDE
product is used for continuous peripheral intravenous in- Sample: Detector tube manufacturers recommended
fusion, it contains NMT 5.95 USP Endotoxin Units/mg of volume 5% of Medical Air
adenosine. Analysis: Pass the Sample through a nitric oxideni-
OTHER REQUIREMENTS: It meets the requirements under trogen dioxide detector tube at the rate specified for
Injections 1. the tube by the detector tube manufacturer.
USP Monographs

ADDITIONAL REQUIREMENTS Acceptance criteria: NMT 2.5 ppm


PACKAGING AND STORAGE: Preserve in tight, single-dose LIMIT OF WATER AND OIL
containers, preferably of Type I glass, and store at con- Analysis: Support one container in an inverted position
trolled room temperature. (with the valve at the bottom) for 5 min. Cautiously
USP REFERENCE STANDARDS 11 open the valve slightly, maintaining the container in an
USP Adenosine RS inverted position. Vent the gas with a barely audible
USP Endotoxin RS flow against a stainless steel mirror for a few seconds.
Acceptance criteria: No liquid is discernible on the
mirror.
ADDITIONAL REQUIREMENTS
PACKAGING AND STORAGE: Preserve in pressurized contain-
.

Medical Air ers. Container connections shall be appropriate for air.


Adaptors shall not be used to connect containers to pa-
DEFINITION tient use supply system piping or equipment.
Medical Air is a natural or synthetic mixture of gases consist- LABELING: Label states if Medical Air is a synthetic mix-
ing largely of nitrogen and oxygen. It contains NLT ture of Oxygen USP and Nitrogen NF. Where it is piped
19.5% and NMT 23.5%, by volume, of oxygen (O2). directly from the collecting tank to the patient point of
use, label each outlet Medical Air.
IDENTIFICATION
A. The paramagnetic signal exhibited by the Sample gas
in the Assay confirms the presence of oxygen.
B. The Sample gas in the Assay meets the assay Accep-
tance criteria.
.

Alanine
ASSAY
PROCEDURE
The certified standards called for in the following test are
listed in Reagents, Indicators, and Solutions.
Zero gas: Nitrogen certified standard
Span gas: 21% Oxygen certified standard. [NOTESee
Reagents, Indicators, and Solutions.] C3H7NO2 89.09
Sample gas: Medical Air L-Alanine [56-41-7].
Mode: Paramagnetic oxygen measurement (see Medical
Gases Assay 415)
Analysis: Determine the concentration of oxygen in
percentage by volume of Medical Air using a suitable
paramagnetic analyzer.

Official from May 1, 2015


Copyright (c) 2015 The United States Pharmacopeial Convention. All rights reserved.

Das könnte Ihnen auch gefallen