Beruflich Dokumente
Kultur Dokumente
2009-E-0206
SUBJECT: Releasing the Revised Reinstallation Procedure for CR-IR 391CL and CR-IR
391V CL
DESCRIPTION OF CHANGE:
When you reinstall the OS and the application for CR-IR 391CL and CR-IR 391V CL, the reinstallation fails if
you perform the procedure according to the service manual. This ECN is to announce the correct
reinstallation procedure.
Affected document:
CR-IR 391CL Service Manual / 020-221-01E
DETAILS:
The contents described or shown in our ECN documents and/or any information provided from FUJIFILM in verbal or written communication is intended for the
addressee and/or its organization. Transfer and/or copy distribution of the information without written consent of FUJIFILM is not permitted.
PAGE 1 of 2 K12N0028-6-J
ECN No. 2009-E-0206
- NOTE
1) Important information about SUPPLY OF THE ARTICLES:
Charge:
FREE OF CHARGE UNTIL (mmddyyyy):
WITH YOUR CHARGE
NOT APPLICABLE
The contents described or shown in our ECN documents and/or any information provided from FUJIFILM in verbal or written communication is intended for the
addressee and/or its organization. Transfer and/or copy distribution of the information without written consent of FUJIFILM is not permitted.
PAGE 2 of 2 K12N0028-6-J
ECN 2009-E-0206 1/4
P448 (MC-36)
1.4.1 Installing .NET Framework
The installation fails if you perform "install-ASPNET". Do not perform the procedure 14 and 15.
Incorrect correct
13. Click [OK]. 13. Click [OK].
14. Open folders in the following sequence: D drive,
dotNet, and DNET1.1. The installation fails if you perform "install-ASPNET".
15. Duble-click Install-ASPNET. Do not perform the procedure 14 and 15.
The following window opens.
enter system as the password. Click . PC, enter system as the password. Click .
The desktop window open. The desktop window open.
P451 (MC-39)
1.4.3 Adding the Windows Components
Incorrect correct
4. Put checkmarks for the following items. 4. Put checkmarks for the following items.
ActiveX Installer Service ActiveX Installer Service
FTP Publishing Service FTP Publishing Service
Web Management Tools Web Management Tools
.NET Extensibility .NET Extensibility
ASP .NET ASP .NET
ISAPI Extensions ISAPI Extensions
ISAPI Filters ISAPI Filters
Common Http Features Common Http Features
HTTP Logging HTTP Logging
Security Security
XPS Viewer XPS Viewer
Remote Differential Compression Remote Differential Compression
Tablet PC Optional Components
Telnet Client Telnet Client
Windows DFS Replication Service Windows DFS Replication Service
Windows Process Activation Service Windows Process Activation Service
ECN 2009-E-0206 2/4
ECN 2009-E-0206 3/4
P557 (MU-17)
5.5 Setting the Image Reader
Unfailingly remove the checkmark from "Check the box to use Patient information display". If it is
checkmarked, CR-IR 391RU cannot be connected because it cannot display patient information.
Incorrect correct
1. Enter an Image Reader name (default: Image Reader) in 1. Enter an Image Reader name (default: Image Reader)
the Equipment name field. in the Equipment name field.
2. Enter an Image Reader host name (default: RU0) in the 2. Enter an Image Reader host name (default: RU0) in
Host name field. the Host name field.
3. Enter an Image Reader IP address (default: 172.16.1.10) 3. Enter an Image Reader IP address (default:
in the IP Address field. 172.16.1.10) in the IP Address field.
4. Checkmark Use in the Check the box to use Patient 4. Remove the checkmark from Check the box to use
information display field ( ). Patient information display.
NOTE NOTE
To have the patient information be displayed, the relevant If "Check the box to use Patient information display" is
setting must be turned ON also on the RU side. checkmarked, CR-IR 391RU cannot be connected.
For details of the procedure to use, see Appendix 4: Setting Unfailingly remove the checkmark from it.
the RU-Config in MC: Checks, Replacement and
Adjustment of Parts.
P739 (IN-27)
4.4.6 Setting the Image Reader
Unfailingly remove the checkmark from "Check the box to use Patient information display". If it is
checkmarked, CR-IR 391RU cannot be connected because it cannot display patient information.
Incorrect correct
1. Enter an Image Reader name (default: Image Reader) in 1. Enter an Image Reader name (default: Image Reader)
the Equipment name field. in the Equipment name field.
2. Enter an Image Reader host name (default: RU0) in the 2. Enter an Image Reader host name (default: RU0) in
Host name field. the Host name field.
3. Enter an Image Reader IP address (default: 172.16.1.10) 3. Enter an Image Reader IP address (default:
in the IP Address field. 172.16.1.10) in the IP Address field.
4. Remove the checkmark from Check the box to use
Patient information display .
NOTE
If "Check the box to use Patient information display" is
checkmarked, CR-IR 391RU cannot be connected.
Unfailingly remove the checkmark from it.
4. Click [Next]. 5. Click [Next].
The Simplified settings completed window opens. The Simplified settings completed window opens.