Beruflich Dokumente
Kultur Dokumente
Essentials of
Anaesthetic Equipment
FOURTH EDITION
Senior Content Strategist: Jeremy Bowes
Content Development Specialist: Clive Hewat
Senior Project Manager: Beula Christopher
Designer: Miles Hitchens
Illustration Manager: Jennifer Rose
Illustrator: Richard Tibbits
iii
Essentials of
Anaesthetic
Equipment
FOURTH EDITION
Edinburgh London New York Oxford Philadelphia St Louis Sydney Toronto 2013
2013 Elsevier Ltd. All rights reserved.
No part of this publication may be reproduced or transmitted in any form or by any means,
electronic or mechanical, including photocopying, recording, or any information storage and
retrieval system, without permission in writing from the publisher. Details on how to seek
permission, further information about the Publishers permissions policies and our
arrangements with organizations such as the Copyright Clearance Center and the Copyright
Licensing Agency, can be found at our website: www.elsevier.com/permissions.
This book and the individual contributions contained in it are protected under copyright by
the Publisher (other than as may be noted herein).
ISBN: 978-0-7020-4954-5
Notices
Knowledge and best practice in this field are constantly changing. As new research and
experience broaden our understanding, changes in research methods, professional practices,
or medical treatment may become necessary.
Practitioners and researchers must always rely on their own experience and knowledge in
evaluating and using any information, methods, compounds, or experiments described
herein. In using such information or methods they should be mindful of their own safety
and the safety of others, including parties for whom they have a professional responsibility.
With respect to any drug or pharmaceutical products identified, readers are advised to check
the most current information provided (i) on procedures featured or (ii) by the manufacturer
of each product to be administered, to verify the recommended dose or formula, the method
and duration of administration, and contraindications. It is the responsibility of
practitioners, relying on their own experience and knowledge of their patients, to make
diagnoses, to determine dosages and the best treatment for each individual patient, and to
take all appropriate safety precautions.
To the fullest extent of the law, neither the Publisher nor the authors, contributors, or
editors, assume any liability for any injury and/or damage to persons or property as a matter
of products liability, negligence or otherwise, or from any use or operation of any methods,
products, instructions, or ideas contained in the material herein.
The
publishers
policy is to use
paper manufactured
from sustainable forests
Printed in China
Contents
Index 279
This page intentionally left blank
Preface
Over 20 years ago, we conspired to write our colour equipment textbook Essentials of Anaesthetic Equipment. It is
now in its fourth edition and hopefully as relevant to anaesthetic practice as ever.
We have tried to keep the book concise, however due to the sheer number of new anaesthetic equipment products
used in the clinical practice today, the size of the book has increased slightly. We have tried to freshen up the
photography/diagrams wherever possible. The text has been updated too and single best answer questions have
been included.
We hope this book will continue to be the equipment book of choice for both the trainees sitting FRCA exams and
their trainers and a useful reference tool for our Nursing and Operating Department Practitioner colleagues.
We are extremely grateful to the many manufacturers and others who have supplied the necessary information and
images for this edition. Without their help, this fourth edition could not have gone ahead in its current format.
Special mention goes to Andy Coughlan of Smiths Medical for his unflagging help with all things photographic.
Below is a list of the people and their companies who helped us by providing images during the preparation of
the book.
We are also grateful to the association of Anaesthetists of Great Britain and Ireland for granting permission to
reproduce their equipment checklist and monitoring recommendations.
Molly Bruton (Vygon) John van Kessel (B Braun) Ciska Proos (B Braun)
Tony Calvo (Olympus) Vanessa Light (Philips Health Care) Malcolm Pyke (Philips Heath Care)
Emma Christmas (Gambro) Sharon Maris (Teleflex) Siama Rafiq (BD Medical)
Andy Coughlan (Smiths Medical) Lucy Martin-Davis (Verathon Emma Richardson (Argon Medical)
Medical)
Inga Dolezar (Chart BioMedical) Rachel Stein (I-Flow)
Anne Pattinson (Draeger)
Bjrake Frank-Duab (Radiometer) Frank Toal (B Braun)
Mark Pedley (Blue Box Medical)
Andrew Garnham (Penlon) Jill Garratt (Zoll Medical)
Lee Pettitt (Rimer Alco)
John Jones (MAQUET)
This page intentionally left blank
Chapter 1
Medical gas
supply
2 1 Medical gas supply
Gas supply
Medical gas supply takes the form
of either cylinders or a piped gas
system, depending on the
requirements of the hospital.
Cylinders
Components
1. Cylinders are made of thin-
walled seamless molybdenum
steel in which gases and vapours Fig. 1.1 Nitrous oxide cylinder with its Fig. 1.2 Oxygen cylinder valve and pin
wrapping and labels. index.
are stored under pressure. They
are designed to withstand
considerable internal pressure.
2. The top end of the cylinder is
called the neck, and this ends
in a tapered screw thread into
which the valve is fitted. The
thread is sealed with a material
that melts if the cylinder is
exposed to intense heat. This
allows the gas to escape so
reducing the risk of an
explosion.
3. There is a plastic disc around
the neck of the cylinder. The
year when the cylinder was last
examined can be identified from
the shape and colour of the disc.
4. Cylinders are manufactured in Fig. 1.3 Nitrous oxide cylinder valve and Fig. 1.4 Carbon dioxide cylinder valve
different sizes (A to J). Sizes A pin index. and pin index.
and H are not used for medical
gases. Cylinders attached to the a fibreglass covering in epoxy resonance scanners. They have a
anaesthetic machine are usually resin matrix. These can be used flat base to help in storage and
size E (Figs 1.11.4), while size J to provide oxygen at home, handling.
cylinders are commonly used during transport or in magnetic
for cylinder manifolds. Size E
oxygen cylinders contain 680L,
whereas size E nitrous oxide Gas exits in the gaseous state at room temperature. Its liquefaction at
cylinders can release 1800L. room temperature is impossible, since the room temperature is above its
The smallest sized cylinder, size critical temperature.
C, can hold 1.2L of water, and Vapour is the gaseous state of a substance below its critical temperature. At
size E can hold 4.7L while the room temperature and atmospheric pressure, the substance is liquid.
larger size J can hold 47.2L of Critical temperature is the temperature above which a substance cannot be
water. liquefied no matter how much pressure is applied. The critical temperatures
5. Lightweight cylinders can be for nitrous oxide and oxygen are 36.5 and 118C respectively.
made from aluminium alloy with
Cylinders 3
Labelling
A full oxygen cylinder at The cylinder label includes the following details:
atmospheric pressure can deliver Name, chemical symbol, pharmaceutical form, specification of the product,
130 times its capacity of oxygen. A its licence number and the proportion of the constituent gases in a
typical size E full oxygen cylinder gas mixture.
delivering 4L per minute will last Substance identification number and batch number.
for 2 hours and 50 minutes but Hazard warnings and safety instructions.
will last only 45 minutes when Cylinder size code.
delivering 15L/min. Nominal cylinder contents (litres).
A full oxygen cylinder at Maximum cylinder pressure (bars).
atmospheric pressure can Filling date, shelf life and expiry date.
deliver 130 times its capacity of Directions for use.
oxygen. Storage and handling precautions.
At constant temperature, a
gas-containing cylinder shows a
linear and proportional reduction
in cylinder pressure as it empties.
Problems in practice and
For a cylinder that contains liquid a cylinder to the wrong yoke
safety features
and vapour, initially the pressure (Fig. 1.5).
remains constant as more vapour is 1. The gases and vapours should 3. Cylinders are colour-coded to
produced to replace that used. Once be free of water vapour when reduce accidental use of the
all the liquid has been evaporated, stored in cylinders. Water wrong gas or vapour. In the
the pressure in the cylinder vapour freezes and blocks the UK, the colour-coding is a
decreases. The temperature in such exit port when the temperature two-part colour, shoulder and
a cylinder can decrease because of of the cylinder decreases on body (Table 1.1). To improve
the loss of the latent heat of opening. safety, there are plans to
vaporization leading to the 2. The outlet valve uses the change the colours of the
formation of ice on the outside of pin-index system to make it bodies of cylinders using
the cylinder. almost impossible to connect medical gas to white while
4 1 Medical gas supply
Table 1.1 Colour coding of medical gas cylinders, their pressure when full and their physical state in the cylinder
Oxygen
Nitrous oxide
Entonox
(50% N2O/50%O2)
Air
Carbon dioxide
Helium/oxygen mixture
(79% He/21% O2)
Cylinder valves 5
Cylinders
Cylinders are made of thin-
walled molybdenum steel to
withstand high pressures, e.g.
13700kPa and 4400kPa for
Pin-index Side spindle Bullnose
oxygen and nitrous oxide valve pin-index valve
respectively. Lightweight valve
aluminium is also available.
They are made in different sizes:
size E cylinders are used on the
anaesthetic machine; size J
cylinders are used in cylinder
banks.
Oxygen cylinders contain gas
whereas nitrous oxide cylinders To Pin-index valve Star valve
contain a mixture of liquid and Valve types
vapour. In the UK, nitrous oxide Fig. 1.6 Chemical formula (N2O) Fig. 1.7 Cylinder valves.
cylinders are 75% filled with engraved on a nitrous oxide cylinder valve.
liquid nitrous oxide (filling ratio);
this is 67% in hotter climates.
At a constant temperature, the
pressure in a gas cylinder
decreases linearly and
proportionally as it empties.
This is not true in cylinders
containing liquid/vapour.
They are colour-coded (shoulder
and body).
Cylinder valves
They are mounted on the neck
of the cylinder.
Act as an on/off device for the
discharge of cylinder contents.
Pin-index system prevents
cylinder identification errors.
Bodok sealing washer must be
placed between the valve and
the yoke of the anaesthetic
Fig. 1.10 New cylinder valve which allows manual opening and closing.
machine.
A newly designed valve allows
2. A compressible yoke-sealing
keyless manual turning on and
washer (Bodok seal) must be
off.
placed between valve outlet and
the apparatus to make a
gas-tight joint (Fig. 1.11).
Components
1. Central supply points such as
cylinder banks or liquid oxygen
storage tank.
2. Pipework made of special
high-quality copper alloy, which Fig. 1.12 Inserting a remote probe into its matching wall-mounted
outlet socket.
both prevents degradation of
the gases it contains and has
bacteriostatic properties. The
fittings used are made from
brass and are brazed rather than
soldered.
3. The size of the pipes differs
according to the demand that
they carry. Pipes with a 42mm
diameter are usually used for
leaving the manifold. Smaller
diameter tubes, such as 15mm,
are used after repeated branching.
4. Outlets are identified by gas Fig. 1.13 Gas probes for oxygen (top),
colour coding, gas name and by nitrous oxide (middle) and air (bottom).
shape (Fig. 1.12). They accept Note the locking groove on the probe to
matching quick connect/ ensure connectivity.
disconnect probes, Schrader pipeline network. They are also
sockets (Fig. 1.13), with an known as area valve service
indexing collar specific for each units (AVSUs) (Fig. 1.16). They
gas (or gas mixture). can be accessed to isolate the Fig. 1.14 Outlet sockets mounted in a
5. Outlets can be installed as supply to an area in cases of fire retractable ceiling unit. (Courtesy of
flush-fitting units, surface-fitting or other emergency Penlon Ltd, Abingdon, UK (www.penlon.
units, on booms or pendants, or com).)
suspended on a hose and gang
Problems in practice and
mounted (Fig. 1.14). 5. Anaesthetists are responsible for
safety features
6. Flexible colour-coded hoses the gases supplied from the
connect the outlets to the 1. A reserve bank of cylinders is terminal outlet through to the
anaesthetic machine (Fig. 1.15). available should the primary anaesthetic machine. Pharmacy,
The anaesthetic machine end supply fail. Low-pressure alarms supplies and engineering
should be permanently fixed detect gas supply failure departments share the
using a nut and liner union where (Fig. 1.17). responsibility for the gas
the thread is gas specific and 2. Single hose test is performed to pipelines behind the wall.
non-interchangeable (non- detect cross-connection. 6. There is a risk of fire from worn
interchangeable screw thread, 3. Tug test is performed to detect or damaged hoses that are
NIST, is the British Standard). misconnection. designed to carry gases under
7. Isolating valves behind break 4. Regulations for PMGV pressure from a primary source
glass covers are positioned at installation, repair and such as a cylinder or wall-
strategic points throughout the modification are enforced. mounted terminal to medical
8 1 Medical gas supply
CYLINDER MANIFOLD
Components
1. Large cylinders (e.g. size J each
with 6800L capacity) are usually
divided into two equal groups,
primary and secondary. The two
groups alternate in supplying the
pipelines (Fig. 1.19). The number
Fig. 1.17 Medical gas alarm panel. (Courtesy of Penlon Ltd, of cylinders depends on the
Abingdon, UK (www.penlon.com).) expected demand.
Sources of gas supply 9
Liquid oxygen
LIQUID OXYGEN
A vacuum-insulated evaporator
(VIE) (Fig. 1.20) is the most
economical way to store and supply
oxygen.
Components
1. A thermally insulated double-
walled steel tank with a layer of
perlite in a vacuum is used as
the insulation (Fig. 1.21). It
can be described as a giant
thermos flask, employing the
same principles.
Fig. 1.19 An oxygen cylinder manifold. 2. A pressure regulator allows
gas to enter the pipelines
and maintains the pressure
2. All cylinders in each group are nearly empty. The changeover is through the pipelines at about
connected through non-return achieved through a pressure- 400kPa.
valves to a common pipe. This sensitive device that detects when 3. A safety valve opens at
in turn is connected to the the cylinders are nearly empty. 1700kPa allowing the gas to
pipeline through pressure 3. The changeover activates an escape when there is a build-up
regulators. electrical signalling system to of pressure within the vessel.
3. As nitrous oxide is only available alert staff to the need to change This can be caused by
in cylinders (in contrast to liquid the cylinders. underdemand for oxygen.
oxygen), its manifold is larger 4. A control valve opens when there
than that of oxygen. The latter Problems in practice and is an excessive demand on the
usually acts as a back up to liquid safety features system. This allows liquid oxygen
oxygen supply (see later). to evaporate by passing through
1. The manifold should be housed
in a well-ventilated room built of superheaters made of uninsulated
Mechanism of action coils of copper tubing.
fireproof material away from the
1. In either group, all the cylinders main buildings of the hospital.
valves are opened. This allows 2. The manifold room should not be
Mechanism of action
them to empty simultaneously. used as a general cylinder store.
2. The supply is automatically 3. All empty cylinders should be 1. Liquid oxygen is stored (up to
changed to the secondary group removed immediately from the 1500L) at a temperature of
when the primary group is manifold room. 150 to 170C (lower than
10 1 Medical gas supply
Entonox (BOC
Medical)
This is a compressed gas mixture
containing 50% oxygen and 50%
nitrous oxide by volume. It is
Fig. 1.24 The RA40/D/Mhospital oxygen concentrator. It produces 80L/min of oxygen. commonly used in the casualty and
(Courtesy of Rimer Alco Ltd.) labour ward settings to provide
12 1 Medical gas supply
Tank
Zeolite towers
Nitrogen vent
Centralized vacuum
or suction system
(Fig. 1.29)
Suction devices play a crucial
part in the care of patients in the
operating theatre, intensive care unit
and other parts of the hospital.
Components
1. A pump or a power source that
is capable of continuously
generating a negative pressure of
500mmHg.
2. A suction controller with a filter.
3. A receiver or a collection vessel.
4. A suction tubing and suction
Fig. 1.28 Compressed medical air plant. (Courtesy of Penlon Ltd, Abingdon, UK
nozzle (e.g. a Yankaeur sucker)
(www.penlon.com).)
or catheter.
to the supply of hypoxic mixtures, should be avoided because of the To determine the efficiency of
with less than 20% oxygen, as the effect of nitrous oxide on the central-piped vacuum systems
cylinder is nearly empty. bone marrow especially in the A negative pressure of at least
Rewarming and mixing of both critically ill patient. Adequate 53kPa (400mmHg) should
the cylinder and its contents reverses facilities for scavenging should be maintained at the outlet.
the separation and liquefaction. be provided to protect hospital Each central-piped vacuum
staff. outlet should be able to
Problems in practice and withstand a flow of free air of at
safety features least 40L/min.
A unit should take no longer
Liquefaction and separation of the
components can be prevented by:
Compressed air than 10 seconds to generate a
vacuum (500mmHg) with a
1. Cylinders being stored Medical air is supplied in a hospital displacement of air of 25L/min.
horizontally for about 24 hours for clinical uses or to drive power
at temperatures of or above 5C tools. The former is supplied at
before use. The horizontal a pressure of 400kPa and the
Mechanism of action
position increases the area for latter at 700kPa. The anaesthetic
diffusion. If the contents are machines and most intensive care 1. Negative pressure is generated
well mixed by repeated ventilator blenders accept a 400kPa by an electric motor and
inversion, cylinders can be used supply. The terminal outlets for the pneumatic-driven pumps using
earlier than 24 hours. two pressures are different to the Venturi principle.
2. Large cylinders are equipped prevent misconnection. 2. The amount of vacuum
with a dip tube with its tip Air may be supplied from cylinder generated can be manually
ending in the liquid phase. manifolds, or more economically adjusted by the suction
This results in the liquid being from a compressor plant with duty controller. This device has a
used first, preventing the and back-up compressors (Fig. 1.28). variable orifice with a float
delivery of an oxygen Oil-free medical air is cleaned by assembly, a back-up filter to
concentration of less than 20%. filters and separators and then dried prevent liquid entering the
Prolonged use of Entonox before use. system and ports to connect to a
14 1 Medical gas supply
Centralized vacuum
or suction system
Consists of a power source, a
suction controller, a receiver, a
suction tubing and suction
nozzle.
Efficiency of the system should
be tested before use.
The amount of negative pressure
should be adjusted according to
its use.
Trauma to tissues can be caused
by the suction.
FURTHER READING
Health Technical Memorandum 2022,
1997. Medical gas pipeline systems.
The Stationery Office, London
Fig. 1.29 Medical vacuum plant. (Courtesy of Penlon Ltd, Abingdon, UK (www.penlon. Health Technical Memorandum 02-01,
com).) 2006. Medical gas pipeline systems,
part A; design, installation, validation
and verification.
The Stationery Office, London.
Health Technical Memorandum 02-01,
2006. Medical gas pipeline systems,
collection vessel or reservoir 6. Bacterial filters are used to part B; operational management.
through flexible tubing. prevent spread of infectious The Stationery Office, London.
3. The reservoir must have bacteria, with a removal of Highly, D., 2009. Medical gases, their
sufficient capacity to receive the 99.999% of bacteria. Filters are storage and delivery. Anaesthesia
aspirated material. Too large a also used to prevent fluids, and Intensive Care Medicine 10 (11),
capacity will make the system condensate and smoke from 523527.
cumbersome and will take a contaminating the system. MHRA, 2011. Medical device alert:
long time to generate adequate 7. It is recommended that there anaesthetic machine: auxillary
negative pressure. are at least two vacuum outlets common gas outlet (ACGO)
4. The suction tubing should be per each operating theatre, manufactured by GE Healthcare
flexible and firm to prevent one per anaesthetic room (MDA/2011/118). Online. Available
collapse. Also it should be and one per recovery or at: http://www.mhra.gov.uk/
transparent so that the contents intensive care unit bed. Publications/Safetywarnings/
aspirated can be visualized, and MedicalDeviceAlerts/CON137664
of sufficient internal diameter National Health Service, 2009. Oxygen
Problems in practice and
and length for optimal suction. safety in hospitals. Online. Available
safety features
5. The negative pressure (or degree at: http://www.nrls.npsa.nhs.uk/resou
of suctioning) can be adjusted to To prevent trauma to the tissues rces/?entryid45=62811&p=7
suit its use; e.g. a lesser degree during suction, the nozzles should Poolacherla, R., Nickells, J., 2006.
of suctioning is required to clear taper, be smooth and have multiple Suction devices. Anaesthesia and
oral secretions in a child than in holes, so that if one is blocked the Intensive Care Medicine 7 (10),
an adult. others will continue suction. 354355.
MCQs 15
MCQs
In the following lists, which of the statements (a) to (e) are true?
Answers
Chapter 2
The
anaesthetic
machine
20 2 The anaesthetic machine
Pressure gauge
This measures the pressure in the
cylinder or pipeline. The pressure
gauges for oxygen, nitrous oxide
and medical air are mounted in a
front-facing panel on the anaesthetic
machine (Fig. 2.3).
Some modern anaesthetic
machine designs have a digital
display of the gas supply pressures
(Fig. 2.4).
Components
1. A robust, flexible and coiled
tube which is oval in cross-
section (Fig. 2.5). It should be
able to withstand the sudden
high pressure when the cylinder
Fig. 2.1 The Datex-Ohmeda Aestiva S/5 anaesthetic machine. is switched on.
Pressure gauge 21
2. The tube is sealed at its inner 2. The movement of the tube particular gas or vapour.
end and connected to a needle causes the needle pointer to The pressure measured indicates
pointer which moves over a dial. move on the calibrated dial the contents available in an
3. The other end of the tube is indicating the pressure. oxygen cylinder. Oxygen is
exposed to the gas supply.
Problems in practice and
Mechanism of action safety features
1. The high-pressure gas causes the 1. Each pressure gauge is colour-
tube to uncoil (Bourdon gauge). coded and calibrated for a
3
O2 supply
Reserve failure alarm Fig. 2.4 Digital display of pressure
cylinders gauges for oxygen (cylinder and pipeline),
Gas supply O2 nitrous oxide (pipeline) and air (pipeline).
1 5
at standard
Piped gases operating N2O O2 flush
Calibrated
2 pressure Coiled
dial
4 Pointer tube
Back-bar
0 200
vaporizer
Cylinder pressure regulator
Non-return pressure relief valve Common
activated at about 35 kPa gas outlet
Fig. 2.2 Diagrammatic representation of a continuous flow anaesthetic Mechanism
machine. Pressures throughout the system: 1. O2: 13700kPa, N2O: 4400kPa;
2. pipeline: about 400kPa; 3. O2 supply failure alarm activated at <250kPa; 4.
Gas under
regulated gas supply at about 400kPa; 5. O2: flush 45L/min at a pressure of
pressure
about 400kPa; 6. back-bar pressure 110kPa (depending on flow rate and
type of vaporizer). Fig. 2.5 The Bourdon pressure gauge.
Fig. 2.3 Pipeline pressure gauges for oxygen, nitrous oxide and air.
22 2 The anaesthetic machine
Pressure regulator
(reducing valve)
Pressure regulators are used
because:
Gas and vapour are stored under
high pressure in cylinders. A
regulator reduces the variable
cylinder pressure to a constant
safer operating pressure of about
400kPa (just below the pipeline
pressure) (Fig. 2.6).
The temperature and pressure of
the cylinder contents decrease
with use. In order to maintain
flow, constant adjustment is
required in the absence of
regulators.
Regulators protect the
components of the anaesthetic Fig. 2.7 Cylinder pressure regulators (black domes) positioned above the cylinder
machine against pressure surges. yokes in the Datex-Ohmeda Flexima anaesthetic machine.
Flow control (needle) valves 23
Second-stage
regulators and flow
restrictors
The control of pipeline pressure
surges can be achieved either by
using a second-stage pressure
regulator or a flow restrictor (Fig.
2.9) a constriction, between the
pipeline supply and the rest of
the anaesthetic machine. A lower
pressure (100200kPa) is achieved.
If there are only flow restrictors and
no regulators in the pipeline supply,
adjustment of the flowmeter
controls is usually necessary
whenever there is change in pipeline
pressure.
Flow restrictors may also be used
Fig. 2.8 Cylinder pressure regulator (the machines tray has been removed). downstream of vaporizers to prevent
back pressure effect (see later).
2. The diaphragm can rupture. opposing forces maintains a Fig. 2.9 A flow restrictor. The
3. Relief valves (usually set at constant operating pressure. constriction causes a significant pressure
700kPa) are fitted downstream drop when there is a high gas flow rate.
24 2 The anaesthetic machine
Components
1. The body, made of brass, screws
Bobbin
into the base of the flowmeter.
2. The stem screws into the body
and ends in a needle. It has
Tapered tube
screw threads allowing fine
adjustment.
3. The flow control knobs are
labelled and colour-coded.
4. A flow control knob guard is
fitted in some designs to protect
against accidental adjustment in
the flowmeters.
Bobbin stop
Control
Flowmeters knob Gas in
Mechanism of action
1. The calibration of each
vaporizer is agent-specific.
2. Fresh gas flow is split into
two streams on entering the
vaporizer. One stream flows
through the bypass channel
and the other, smaller stream,
flows through the vaporizing
chamber. The two gas streams
reunite as the gas leaves
Fig. 2.17 Tec Mk 5 vaporizers mounted on the back bar of an anaesthetic machine. the vaporizer.
3. The vaporization chamber is
Percentage designed so that the gas leaving
control valve Dial it is always fully saturated with
vapour before it rejoins the
Fresh bypass gas stream. This should
and be achieved despite changes in
Fresh vapour the FGF.
gas in out
4. Full saturation with vapour is
achieved by increasing the
Outlet port
Flow split surface area of contact between
the carrier gas and the
Vaporizing anaesthetic agent. This is
chamber achieved by having wicks
saturated by the inhalational
agent, a series of baffles or by
bubbling the gas through the
Bypass
liquid.
5. The desired concentration is
obtained by adjusting the
percentage control dial. This
alters the amount of gas flowing
through the bypass channel to
that flowing through the
Bypass Bypass vaporization chamber.
6. In the modern designs, the
Bimetallic Temperature
strip compensating
vapour concentration supplied
valve by the vaporizer is virtually
independent of the FGFs
Fig. 2.18 A schematic diagram of the Tec Mk 5, an example of a plenum
between 0.5 and 15L/min.
vaporizer.
7. During vaporization, cooling
contracts as the temperature of lever of the system is engaged (in occurs due to the loss of latent
the vaporizer decreases. Mk 4 and 5). The interlocking heat of vaporization. Lowering
5. The vaporizers are mounted extension rods prevent more the temperature of the agent
on the back bar (Fig. 2.20) using than one vaporizer being used makes it less volatile. In order to
the interlocking Selectatec at any one time, preventing compensate for temperature
system (Fig. 2.21). The contamination of the one changes:
percentage control dial cannot downstream (in Mk 4 and 5). a) the vaporizer is made of a
be moved unless the locking The FGF only enters the material with high density and
Vaporizers 29
Fig. 2.22 The Selectatec series mounted manifold bypass circuit. Only when a vaporizer
is locked in position and turned on can fresh gas enter. Vaporizer B is turned off and is Vaporizer filling
isolated from the fresh gas which only enters vaporizer A which is turned on. If no
vaporizer is fitted, the port valves are closed. (Reproduced with permission from
devices
Datex-Ohmeda.)
These are agent-specific being
Tec Mk 3. This ensures that pressure required to operate geometrically coded (keyed) to fit
the bypass channel is not commonly used ventilators the safety filling port of the correct
contaminated by retrograde c) both the bypass channel and vaporizer and anaesthetic agent
flow from the vaporizing the vaporizing chamber are of supply bottle (Fig. 2.23). They
chamber equal volumes so gas expansion prevent the risk of adding the
b) downstream flow restrictors: and compression are equal. wrong agent to the wrong vaporizer
used to maintain the vaporizer 4. Preservatives, such as thymol in and decrease the extent of spillage.
at a pressure greater than any halothane, accumulate on the The safety filling system, in
Emergency oxygen flush 31
Red Halothane
Orange Enflurane
Purple Isoflurane
Yellow Sevoflurane
Blue Desflurane
in 10% steps from 30% The UAM differs from standard less than the patients minute
to 100%. anaesthetic machines by the use ventilation) and a positive
d) This design prevents the of an electrically powered pressure relief valve prevents
delivery of hypoxic mixtures. oxygen concentrator (producing overpressure of the bag.
e) It is mainly used in dental 10L/min of 95% oxygen), A low-resistance vaporizer is
anaesthesia. drawover vaporizer, bellows and fitted downstream of the positive
Some newly designed balloon valve. The UAM can pressure valve. Vaporizers
anaesthetic machines have an function in both continuous flow calibrated for the use of
extra outlet with its own and drawover modes, entraining isoflurane or halothane are
flowmeter to deliver oxygen to air as necessary (e.g. if electricity available. Distally, the set of
conscious or lightly sedated supply to the concentrator fails), silicone bellows (up to 1600mL)
patients via a face mask. This with the vaporizer functioning allows manual ventilation
can be used in patients as normal. Alternatively, oxygen through a standard dual limb
undergoing surgery under can be provided via cylinder, breathing system. The expiratory
regional anaesthesia with pipeline or the side emergency valve is sited on the side of the
sedation. inlet. The UAM has two machine, and comprises a
5. Universal Anaesthesia Machine flowmeters, one for oxygen and long-life silicone balloon housed
(UAM) (Figs 1.23 and 2.35). the other for either nitrous oxide in a clear plastic tube.
This was developed to enable or air. A 2-L reservoir bag is A fuel cell oxygen
the provision of anaesthesia in positioned distal to the concentration monitor with a
poorly resourced countries flowmeters on the back bar. A touch-screen display, as well as a
where compressed gases and negative pressure valve allows safety anti-hypoxic feature, are
electricity supplies are unreliable. entrainment of air (if the FGF is included in the design.
Oxygen analyser
Oxygen and apnoea alarm DOV for
concentrator PPRV HAL/ISO
O2 inlet
REGS
NPRV
air inlet
Silicone
bellows
Y-piece outlet
To patient INSP
EXP FGF inlet
Bellows
block +
Scavenger 2 valves
MCQs
In the following lists, which of the following statements (a) to (e) are true?
9. The non-return valve on the 10. The oxygen emergency flush on SINGLE BEST ANSWER (SBA)
back bar of an anaesthetic an anaesthetic machine
machine between the vaporizer a) Operates at 20L/min. 11. Concerning a desflurane
and common gas outlet: b) Is always safe to use during vaporizer:
a) Decreases the pumping effect. anaesthesia. a) Is ready for use immediately.
b) Often is incorporated with a c) Operates at 40L/min. b) Can be used with other
pressure relief valve on d) Increases risk of awareness inhalational agents.
modern machines. during anaesthesia. c) Needs an electrical supply to
c) Is designed to protect the e) Can be safely used with a function.
patient. minute volume divider d) Fresh gas flow enters the
d) Is designed to protect the ventilator. vaporization chamber.
machine. e) Is colour-coded red.
e) Opens at a pressure of
70kPa.
Answers 41
Answers
8. The safety features found in an 9. The non-return valve on the 10. The oxygen emergency flush on
anaesthetic machine include: back bar of an anaesthetic an anaesthetic machine:
a) True. This is an essential machine between the vaporizer a) False. 3575L/min can be
safety feature in the and common gas outlet: delivered by activating the
anaesthetic machine. The ideal a) True. Minute volume divider oxygen emergency flush on
design should operate under ventilators exert back pressure the anaesthetic machine.
the pressure of oxygen itself, as they cycle. This causes b) False. The inappropriate use
give a characteristic audible reversal of the fresh gas flow of the oxygen flush during
signal, be capable of warning through the vaporizer. This anaesthesia increases risk of
of impending failure and give leads to an uncontrolled awareness (a 100% oxygen
a further alarm when failure increase in the concentration can be delivered) and
has occurred, be capable of of the inhalational agent. Also barotrauma to the patient
interrupting the flow of other the back pressure causes the (because of the high flows
gases and not require batteries fluctuation of the bobbins delivered).
or mains power to operate. in the flowmeters as the c) True. See a).
b) True. The flowmeters are ventilator cycles. The non- d) True. This can happen by
colour-coded and also the return valve on the back bar diluting the anaesthetic
shape and size of the oxygen prevents these events from mixture; see b).
flowmeter knob is different happening. e) False. Because of the high
from the nitrous oxide knob. b) True. The non-return valve on FGF (3570L/min), the
This allows the identification the back bar opens when the minute volume divider
of the oxygen knob even in a pressure in the back bar ventilator does not function
dark environment. exceeds 35kPa. Flowmeters appropriately.
c) False. The vaporizer level and vaporizer components
can be monitored by the can be damaged at higher 11. c)
anaesthetist. This is part of pressures.
the anaesthetic machine c) True. By preventing the effects
checklist. There is no alarm of back pressure on the
system. flowmeters and vaporizer as
d) True. A ventilator the minute volume divider
disconnection alarm is ventilator cycles, the non-
essential when a ventilator is return valve on the back bar
used. They are also used to provides some protection to
monitor leaks, obstruction the patient. The flows on the
and malfunction. They can flowmeters and the desired
be pressure and/or volume concentration of the
monitoring alarms. In inhalational agent can be
addition, clinical observation, accurately delivered to the
end-tidal carbon dioxide patient.
concentration and airway d) True. See b).
pressure are also e) False. The non-return valve
disconnection alarms. on the back bar of the
e) False. Only one vaporizer can anaesthetic machine opens at
be used at any one time. This a pressure of 35kPa.
is due to the interlocking
Selectatec system where
interlocking extension rods
prevent more than one
vaporizer being used at any
one time. These rods prevent
the percentage control dial
from moving, preventing
contamination of the
downstream vaporizer.
This page intentionally left blank
45
Chapter 3
Pollution in
theatre and
scavenging
46 3 Pollution in theatre and scavenging
maintenance (PPM) programme. At negative pressure variations in the demand in the scavenging
least once annually, the general breathing system. system. Without these valves,
ventilation system and the The performance of the excessive positive pressure
scavenging equipment should be scavenging system should be part of increases the risk of barotrauma
examined and tested by a the anaesthetic machine check. should there be an obstruction
responsible person. Scavenging systems can be divided beyond the receiving system.
into passive and active systems. Excessive negative pressure
could lead to the collapse of the
Pollution in the operating reservoir bag of the breathing
theatre PASSIVE SYSTEM
system and the risk of
In scavenged areas, there is
rebreathing.
no association between The passive system is simple to
3. The disposal system is a
occupational exposure to construct with zero running cost.
wide-bore copper pipe leading to
anaesthetic agents trace levels the atmosphere directly or via
and adverse health effects. Components
the theatre ventilation system.
There are no agreed
1. The collecting and transfer
international standards of system which consists of a
the maximum accepted Mechanism of action
shroud connected to the
concentrations of agents in the adjustable pressure limiting 1. The exhaled gases are driven by
theatre environment. (APL) valve (or expiratory valve either the patients respiratory
Routine monitoring and testing
of the ventilator). A 30-mm efforts or the ventilator.
(PPM) are mandatory. connector attached to transfer 2. The receiving system should be
tubing leads to a receiving mounted on the anaesthetic
system (Fig. 3.1). The 30-mm machine to minimize the length
wide-bore connector is designed of transfer tubing and resistance
Scavenging as a safety measure in order to to flow.
prevent accidental misconnection
In any location in which inhalation to other ports of the breathing
Problems in practice and
anaesthetics are administered, there system (Fig. 3.2).
safety features
should be an adequate and reliable 2. A receiving system (reservoir
system for scavenging waste bag) can be used. Two spring- 1. Connecting the scavenging
anaesthetic gases. A scavenging loaded valves guard against system to the exit grille of the
system is capable of collecting the excessive positive pressures theatre ventilation is possible.
waste anaesthetic gases from the (1000Pa) in case of a distal Recirculation or reversing of the
breathing system and discarding obstruction or negative pressures flow is a problem in this
them safely. Unscavenged operating (50Pa) in case of increased situation.
theatres can show N2O levels of
4003000ppm.
Ducted
A well-designed scavenging expiratory
system should consist of a collecting valve The safety block provides
device for gases from the breathing Reservoir bag positive and negative pressure relief.
system/ventilator at the site of (optional) If it is near to the collecting system,
overflow, a ventilation system to additional positive relief at the
carry waste anaesthetic gases from 30 mm conical collecting system is not required.
the operating theatre and a method connectors
for limiting both positive and Fig. 3.1 Diagram of a passive scavenging system.
48 3 Pollution in theatre and scavenging
Mechanism of action
1. The vacuum drives the gases
through the system. Active
scavenging systems are able
to deal with a wide range
of expiratory flow rates
(30130L/min).
2. A motorized fan, a pump or
a Venturi system is used to
generate the vacuum or negative
pressure that is transmitted
through the pipes.
3. The receiving system is capable
of coping with changes in gas
flow rates. Increased demands
(or excessive negative pressure)
allow ambient air to be
entrained so maintaining
Fig. 3.2 Attaching a 30-mm connecter to the APL valve of the breathing system. The the pressure. The opposite
30-mm wide bore is designed as safety measure. occurs during excessive
positive pressure. As a result,
a uniform gas flow is passed
2. Excess positive or negative between the receiving and to the disposal system.
pressures caused by the wind disposal systems can be used. A
at the outlet might affect the reservoir bag with two spring-
Problems in practice and
performance and even reverse loaded safety valves can also be
safety features
the flow. used as a receiving system.
3. The outlet should be fitted with 3. The active disposal system 1. The reservoir is designed to
a wire mesh to protect against consists of a fan or a pump used prevent excessive negative or
insects. to generate a vacuum (Fig. 3.5). positive pressures being applied
4. Compressing or occluding the
passive hose may lead to the
escape of gases/vapours into the
operating theatre so polluting it.
The disposal hose should be Flow Outlet plate
made of non-compressible controller/ assembly Copper pipe
materials and not placed on the indicator
floor.
To
ACTIVE SYSTEM Flexible atmosphere
connection
Components
1. The collecting and transfer
system which is similar to that
Ducted Reservoir assembly
of the passive system (Fig. 3.3). expiratory
2. The receiving system (Fig 3.4) is valve
usually a valveless, open-ended
Air vents Vacuum pump
reservoir positioned between
the receiving and disposal
components. A bacterial filter 30 mm conical
situated downstream and a connectors
visual flow indicator positioned Fig. 3.3 Diagram of an active scavenging system.
Charcoal canisters (Cardiff Aldasorber) 49
Waste
Air gas
Valve
Suction
Fig. 3.4 Anaesthetic gases receiving system. Its mechanism of action (right).
3. It is usually replaced after every FURTHER READING MHRA, 2010. Medical device alert:
12 hours of use. anaesthetic gas scavenging systems
American Institute of Architects, 1992. (AGSS) all manufacturers
Guidelines for construction and (MDA/2010/021). Online. Available
Problems in practice and equipment of hospitals and medical at: http://www.mhra.gov.uk/
safety features facilities. AIA, Washington DC. Publications/Safetywarnings/
1. It cannot absorb nitrous oxide. Department of Health, 1996. Advice on MedicalDeviceAlerts/CON076104.
2. Heating the canister causes the the implementation of the Health United States Department of Labor,
release of the inhalational agents. and Safety Commissions 1999. Anaesthetic gases: guidelines
occupational exposure standards for for workplace exposures. Online.
anaesthetic agents. DoH, London. Available at: http://www.osha.gov/
Charcoal canisters Health Service Advisory Committee, dts/osta/anestheticgases/index.html.
A canister with charcoal granules 1996. Anaesthetic agents: controlling
used to absorb halogenated exposure under the Control of
agents. Substances Hazardous to Health
Does not absorb nitrous oxide. Regulations (COSHH). HSAC,
Its weight indicates the degree London.
of exhaustion. Usually replaced Henderson, K.A., Raj, A., Hall, J.E.,
after every 12 hours of use. 2002. The use of nitrous oxide in
When heated, the agents escape anaesthetic practice: a questionnaire
back into the atmosphere. survey. Anaesthesia 57 (12),
11551158.
Further reading 51
MCQs
In the following lists, which of the statements (a) to (e) are true?
Answers
Chapter 4
Breathing
systems
56 4 Breathing systems
Breathing systems must fulfil three and C systems are used more 4. Efficient, requiring low FGF
objectives: frequently during the recovery rates.
period and in emergency situations. 5. Protects the patient from
1. Delivery of oxygen.
barotrauma.
2. Removal of carbon dioxide from
The fresh gas flow (FGF) rate 6. Sturdy, compact and lightweight
the patient.
required to prevent rebreathing of in design.
3. Delivery of inhaled anaesthetic
alveolar gas is a measure of the 7. Permits the easy removal of
agents. These agents are
efficiency of a breathing system. waste exhaled gases.
predominantly eliminated by the
8. Easy to maintain with minimal
lungs also, so the breathing
running costs.
system must be able to expel
them as necessary.
Properties of the
There are several breathing
systems used in anaesthesia.
ideal breathing
Mapleson classified them into system Components of the
A, B, C, D and E. After further
revision of the classification, a 1. Simple and safe to use. breathing systems
Mapleson F breathing system was 2. Delivers the intended inspired
added (Fig. 4.1). Currently, only gas mixture. ADJUSTABLE PRESSURE
systems A, D, E and F and their 3. Permits spontaneous, manual LIMITING (APL) VALVE
modifications are commonly used and controlled ventilation in all
during anaesthesia. Mapleson B age groups. This is a valve which allows the
exhaled gases and excess FGF to
leave the breathing system (Fig.
4.2). It does not allow room air
to enter the breathing system.
Synonymous terms for the APL
A valve are expiratory valve, spill
valve and relief valve.
Components
1. Three ports: the inlet, the patient
B
and the exhaust ports. The latter
can be open to the atmosphere
or connected to the scavenging
system using a shroud.
C
2. A lightweight disc rests on a
knife-edge seating. The disc is
held onto its seating by a spring.
The tension in the spring, and
D
therefore the valves opening
pressure, are controlled by the
valve dial.
Green
valve Scavenged gas Scavenged gas
FGF
Lack system
A
(Mapleson A)
This is a coaxial modification of the
Magill Mapleson A system.
FGF
Components
B
1. 1.8m length coaxial tubing
(tube inside a tube). The FGF
is through the outside tube,
and the exhaled gases flow
FGF through the inside tube
(Fig. 4.6A).
2. The inside tube is wide in
C diameter (14mm) to reduce
resistance to expiration.
The outer tubes diameter
is 30mm.
FGF 3. The reservoir bag is mounted
at the machine end.
4. The APL valve is mounted at
D the machine end eliminating the
drag on the connections at the
patient end, which is a problem
with the Magill system.
Alveolar gas containing CO2 Deadspace gas
Fig. 4.5 Mechanism of action of the Magill breathing system during spontaneous
ventilation; see text for details (FGF: fresh gas flow). Mechanism of action
1. A similar mechanism to the
Magill system except the Lack
system is a coaxial version. The
6. The Magill system is not an patients end, especially if connected
fresh gas flows through the
efficient system for controlled to a scavenging system. This places
outside tube whereas the exhaled
ventilation. A FGF rate of three a lot of drag on the connections at
gases flow through the inside
times the alveolar minute the patient end.
tube.
volume is required to prevent
2. A FGF rate of about 70mL/kg/
rebreathing.
min is required in order to
prevent rebreathing. This makes
Magill (Mapleson A) breathing
Problems in practice and system
it an efficient breathing system
safety features Efficient for spontaneous
for spontaneous ventilation.
3. Since it is based on the Magill
It is not suitable for use with ventilation. FGF required is equal
system, it is not suitable for
children of less than 2530kg to alveolar minute volume
controlled ventilation.
body weight. This is because of (about 70mL/kg/min).
the increased dead space caused Inefficient for controlled Instead of the coaxial design, a
by the systems geometry at the ventilation. FGF three times parallel tubing version of the system
patient end. Dead space is further alveolar minute volume. exists (Fig. 4.6B). This has separate
increased by the angle piece and APL valve is at the patients end. inspiratory and expiratory tubing,
face mask. Not suitable for paediatric and retains the same flow
One of its disadvantages is the practice. characteristics as the coaxial
heaviness of the APL valve at the version.
60 4 Breathing systems
A APL
valve
Bain system
FGF (Mapleson D)
The Bain system is a coaxial version
of the Mapleson D system (Fig.
4.7). It is lightweight and compact
at the patient end. It is useful where
B APL
access to the patient is limited, such
valve as during head and neck surgery.
A Manley ventilator which has
been switched to spontaneous
FGF
ventilation mode is an example of a
non-coaxial Mapleson D system.
Fresh gas flow (FGF)
Exhaled gas
Components
1. A length of coaxial tubing (tube
Fig. 4.6 (A) The coaxial Lack breathing system. (B) The parallel Lack breathing
system.
inside a tube). The usual length
is 180cm, but it can be supplied
at 270cm (for dental or
1.52 times the minute volume is
Lack breathing system ophthalmic surgery) and 540cm
required to prevent rebreathing.
Coaxial version of Mapleson A,
(for magnetic resonance imaging
During controlled ventilation, the
making it efficient for (MRI) scans where the
B system is more efficient due to
spontaneous ventilation. FGF anaesthetic machine needs to
the corrugated tubing acting as a
rate of about 70mL/kg/min is be kept outside the scanners
reservoir. A FGF of more than 50%
required. magnetic field). Increasing the
of the minute ventilation is still
FGF is delivered along the
length of the tubing does not
required to prevent rebreathing.
outside tube and the exhaled affect the physical properties of
gases flow along the inner tube. the breathing system.
Mapleson B and C breathing 2. The fresh gas flows through the
APL valve is at the machine end.
systems inner tube while the exhaled
Not suitable for controlled
B system has a tubing and bag gases flow through the outside
ventilation.
reservoir. tube (Fig. 4.8). The internal
Both B and C systems are not lumen has a swivel mount at the
efficient for spontaneous and patient end. This ensures that
controlled ventilation. the internal tube cannot kink, so
Mapleson B and C B system is more efficient than A ensuring delivery of fresh gas to
systems (see Fig. 4.1) system during controlled the patient.
ventilation. 3. The reservoir bag is mounted at
Components the machine end.
1. A reservoir bag. In the B system,
corrugated tubing is attached to
the bag and both act as a
reservoir.
2. An APL valve at the patients end.
3. FGF is added just proximal to
the APL.
Mechanism of action
Both systems are not efficient during
spontaneous ventilation. A FGF of Fig. 4.7 The Bain breathing system.
Bain system (Mapleson D) 61
A FGF C FGF
B FGF D FGF
Expired gas
Fig. 4.8 The proximal (machines) end of
coaxial Bains breathing system. The FGF Fig. 4.9 Mechanism of action of the Mapleson D breathing system during
flows through the narrow inner tube. spontaneous ventilation.
4. The APL valve is mounted at the removing the reservoir bag, 5. A parallel version of the D
machine end. a ventilator such as the system is available.
Penlon Nuffield 200 can
be connected to the bag
Mechanism of action Problems in practice and
mount using a 1-m length of
safety features
1. During spontaneous ventilation, corrugated tubing (the volume
the patients exhaled gases are of tubing must exceed 500mL 1. The internal tube can kink,
channelled back to the reservoir if the driving gas from the preventing fresh gas being
bag and become mixed with ventilator is not to enter delivered to the patient.
fresh gas (Fig. 4.9B). Pressure the breathing system). The 2. The internal tube can become
build-up within the system will APL valve must be fully disconnected at the machine end
open the APL valve allowing closed. causing a large increase in the
the venting of the mixture of
the exhaled gases and fresh gas
(Fig. 4.9C).
2. The FGF required to prevent
rebreathing (as seen in Fig.
4.9D) during spontaneous
ventilation is about 1.52 times FGF
the alveolar minute volume. A
flow rate of 150200mL/kg/min
is required. This makes it an
inefficient and uneconomical
system for use during
Corrugated tube
spontaneous ventilation.
3. It is a more efficient system for
controlled ventilation. A flow of
70100mL/kg/min will maintain
normocapnia. A flow of From control
100mL/kg/min will cause module of Exhaust
moderate hypocapnia during ventilator
controlled ventilation. Patient valve
4. Connection to a ventilator Fig. 4.10 The Bain breathing system connected to a ventilator (e.g. Penlon Nuffield
is possible (Fig. 4.10). By 200) via tubing connected to the bag mount.
62 4 Breathing systems
dead space, resulting in spontaneous and controlled 3. The bag can provide a degree of
hypoxaemia and hypercapnia. ventilation. continuous positive airway
Movement of the reservoir bag pressure (CPAP) during
during spontaneous ventilation spontaneous ventilation.
is not therefore an indication Components 4. Controlled ventilation is
that the fresh gas is being 1. A T-shaped tubing with three performed either by manual
delivered to the patient. open ports (Fig. 4.12). squeezing of the double-ended
2. Fresh gas from the anaesthetic bag (intermittent occlusion of
machine is delivered via a tube the reservoir tubing in the
Bain breathing system to one port. Mapleson E) or by removing the
Coaxial version of Mapleson D. 3. The second port leads to the bag and connecting the reservoir
A parallel version exits. patients mask or tracheal tube. tubing to a ventilator such as the
Fresh gas flows along the inner The connection should be as Penlon Nuffield 200.
tube and the exhaled gases flow short as possible to reduce dead 5. The volume of the reservoir
along the outer tube. space. tubing determines the degree of
Not efficient for spontaneous 4. The third port leads to reservoir rebreathing (too large a tube) or
ventilation. FGF rate required is tubing. Jackson-Rees added a entrainment of ambient air (too
150200mL/kg/min. double-ended bag to the end of small a tube). The volume of the
Efficient during controlled the reservoir tubing (making it reservoir tubing should
ventilation. FGF rate required is Mapleson F). approximate to the patients
70100mL/kg/min. 5. A recent modification tidal volume.
exists where an APL valve
is included before a closed-
ended 500mL reservoir bag.
A pressure relief safety
T-piece system mechanism in the APL valve
is actuated at a pressure of
(Mapleson E and F) 30cm H2O (Fig. 4.13). This
design allows effective
This is a valveless breathing system scavenging.
used in anaesthesia for children up
to 2530kg body weight (Fig.
4.11). It is suitable for both Mechanism of action
1. The system requires an FGF
of 2.53 times the minute
volume to prevent rebreathing
with a minimal flow of
4L/min.
2. The double-ended bag acts
as a visual monitor during
spontaneous ventilation. In Fig. 4.13 Intersurgical T-piece
addition, the bag can be used incorporating an APL valve and closed
for assisted or controlled reservoir bag to enable effective
Fig. 4.11 A T-piece breathing system. ventilation. scavenging.
A Expired B
gases
FGF FGF
Ventilator
Patient Patient
end end
Fig. 4.15 Mechanism of action of the parallel Humphrey ADE breathing system. With the lever up (A), the system functions in
its Mapleson A mode for spontaneous ventilation. For mechanical ventilation, the lever is down (B) and the system functions in
its Mapleson E mode. (Reproduced with permission from Dr D Humphrey.)
analyser are returned back to the reduce powder formation. widely used in the USA.
system. A dye or colour indicator is Another absorber is Amsorb
Minimal flow anaesthesia. The added to change the granules that consists of CaCl2 and
FGF is reduced to 0.5L/min. colour when the soda lime is Ca(OH)2.
Low-flow anaesthesia. The FGF exhausted. Colour changes can
used is less than the patients be from white to violet/purple
Mechanism of action
alveolar ventilation (usually below (ethyl violet dye), from pink
1.5L/min). Excess gases leave the to white (titan yellow dye) or 1. High FGF of several L/min is
system via the APL valve. from green to violet. Colour needed in the initial period to
changes occur when the pH is denitrogenate the circle system
less than 10. Newer types of and the functional residual
soda lime have a low capacity (FRC). This is
Components
concentration of a zeolite added. important to avoid the build
1. A vertically positioned canister This helps to maintain the pH at up of unacceptable levels of
containing soda lime. The a high level for longer and nitrogen in the system. In
canister has two ports, one to retains moisture so improving closed circle anaesthesia, a
deliver inspired gases to the carbon dioxide absorption and high FGF is needed for up
patient and the other to receive reducing the formation of to 15 minutes. In low-flow
exhaled gases from the patient. carbon monoxide and anaesthesia, a high FGF of
2. Inspiratory and expiratory compound A. up to 6 minutes is required. The
tubings connected to the 7. The size of soda lime granules is FGF can be later reduced to
canister. Each port incorporates 48 mesh. Strainers with 48 0.51L/min. If no N2O is used
a unidirectional valve. mesh have four and eight during anaesthesia (i.e. an
3. FGF from the anaesthetic openings per inch respectively. oxygen/air mix is used), it is not
machine is positioned distal to Therefore, the higher the mesh necessary to eliminate nitrogen
the soda lime canister, but number, the smaller the particles because air contains nitrogen. A
proximal to the inspiratory are. Recently produced soda short period of high flow is
valve. lime made to a uniform shape of needed to prime the system and
4. An APL valve is positioned 34-mm spheres allows more the patient with the inhalational
between the expiratory valve even flow of gases and a agent.
and canister and connected to a reduction in channelling. This 2. Exhaled gases are circled back
2-L reservoir bag. results in a longer life with to the canister, where carbon
5. A vaporizer mounted on the lower dust content and lower dioxide absorption takes place
anaesthetic machine back bar resistance to flow: 1kg can and water and heat (exothermic
(vaporizer outside circle VOC) absorb more than 120L of CO2. reaction) are produced. The
or a vaporizer positioned on the 8. Barylime, which consists of warmed and humidified gas joins
expiratory limb within the barium hydroxide (80%) and the FGF to be delivered to the
system (vaporizer inside circle calcium hydroxide (20%), is patient (Fig. 4.17).
VIC). 3. Chemical sequences for the then
6. Soda lime consists of 94% absorption of carbon dioxide by
calcium hydroxide and 5% soda lime: K2CO3 + Ca(OH)2
sodium hydroxide with a small CaCO3 + 2KOH
a) Note how both NaOH and
amount of potassium hydroxide KOH are regenerated at the or
(less than 0.1%). It has a pH of expense of Ca(OH)2. This
13.5 and a moisture content of explains soda limes mix CO2 + 2NaOH
1419%. Some modern types of only a little Na(OH) and Na2CO3 + H 2O + heat
soda lime have no potassium K(OH) and a lot of Ca(OH)2:
hydroxide. Soda lime granules then
are prone to powder formation, H 2O + CO2 H 2CO3 Na2CO3 + Ca(OH)2
especially during transport. 2NaOH + CaCO3
Disintegrated granules increase then
resistance to breathing. Because
of this, silica (0.2%) is added H 2CO3 + 2KOH
to harden the absorbents and K2CO3 + 2H 2O
66 4 Breathing systems
anaesthetic machine. They are dioxide and inhalational moiety (enflurane, isoflurane
high-efficiency vaporizers that can agent concentrations is and desflurane) are used with
deliver high-output concentrations essential. very dry grannules when the
at low flows. They have high 2. The unidirectional valves may water content is less than
internal resistance. stick and fail to close because of 1.5% in soda lime or less than
water vapour condensation. This 5% in barylime. This can occur
1. The vaporizer should be able leads to an enormous increase in when the system is left unused
to deliver accurate dead space. for a long length of time,
concentrations of inhalational 3. The resistance to breathing is e.g. overnight or at weekends,
agent with both high and low increased especially during or when a small basal flow
FGFs. This is easily achieved by spontaneous ventilation. The from the anaesthetic machine
most modern vaporizers, e.g. the main cause of resistance to occurs. Carbon monoxide
Tec series. breathing is due to the accumulation and subsequent
2. The volume of the circle system unidirectional valves. Dust carboxyhaemoglobin formation
is large in relation to the low formation can increase resistance is said to occur at less than
FGF used. Rapid changes in the to breathing further. It can also 0.1% per hour, so may become
concentration of the inspired lead to clogging and channelling, significant in smokers when
vapour can be achieved by so reducing efficiency. Newer ultra-low flows are used;
increasing the FGF to the circle soda lime designs claim less dust oxygen flushes of the system
system. Delivering the FGF formation. (e.g. once an hour) will
distally, using a coaxial 4. Compound A (a penta- prevent this.
inspiratory tubing design, allows fluoroisoproprenyl fluoro- More recent designs of soda
faster changes in inspired vapour methyl ether, which is lime claim less or no production
concentration compared to nephrotoxic in rats) is of carbon monoxide. The
conventional circle systems at produced when sevoflurane association of strong alkalis
low flows. is used in conjunction with such as KOH and NaOH to
VIC vaporizers (see Fig. 4.19) are soda lime. This is due to the the production of carbon
designed to offer minimal resistance degradation of sevoflurane monoxide has led to the
to gas flow and have no wicks on (dehydrohalogenation) as a subsequent removal of KOH
which water vapour might condense result of the alkali metal and reduction in amounts of
(e.g. Goldman vaporizer). The VIC hydroxide present in NaOH used. Some absorbers
is a low-efficiency vaporizer adding soda lime. (e.g. Amsorb) do not use strong
only small amounts of vapour to the Factors that increase alkalis at all.
gas recirculating through it. the production of compound 6. Other substances can accumulate
A are: such as methane, acetone,
1. FGF will be vapour free and a) increasing temperature ethanol and hydrogen.
thus dilutes the inspired vapour b) high sevoflurane However, they do not
concentration. concentrations generally become clinically
2. During spontaneous ventilation, c) use of barylime rather than significant.
respiration is depressed soda lime 7. Uneven filling of the
with deepening of anaesthesia. d) low fresh gas flow canister with soda lime
Uptake of the anaesthetic e) newer designs of soda leads to channelling
agent is therefore reduced. lime, being non-caustic of gases and decreased
This is an example of a (no KOH and only very efficiency.
feedback safety mechanism. low levels of NaOH), 8. The circle system is bulkier, less
The safety mechanism is claim less or no production portable and more difficult to
lost during controlled of compound A. For clean.
ventilation. substance A production, 9. Soda lime is corrosive. Protective
barylime is worse than soda clothing, gloves and eye/face
lime and Amsorb is the protection can be used.
Problems in practice and
safest. 10. Because of the many
safety features
5. Carbon monoxide production connections, there is an
1. Adequate monitoring of inspired can occur when volatile increased potential for leaks and
oxygen, end-tidal carbon agents containing the CHF2 disconnection.
68 4 Breathing systems
MCQs
In the following lists, which of the statements (a) to (e) are true?
8. The circle breathing system: 9. Regarding the circle system: SINGLE BEST ANSWER (SBA)
a) With low flow rates, a) A high FGF is needed in the
substance A can be produced first 15 minutes to wash out 10. The APL valve:
when sevoflurane is used. any CO2 remaining in the a) Is present in all breathing
b) The Goldman vaporizer is an breathing system. systems.
example of a VOC. b) The pH of soda lime is highly b) Is actuated at only very high
c) Failure of the unidirectional acidic. pressures when in the open
valves to close leads to an c) The lower the FGF, the slower position.
enormous increase in the dead the exhaustion of soda lime c) Can act as a scavenging
space. granules. system.
d) Patients should not be allowed d) The Waters to-and-fro system d) Scavenging systems can
to breathe spontaneously, is very efficient. usually be attached to it.
because of the high resistance e) Partially harmful substances e) Is coloured bright yellow for
caused by the soda lime. can be produced when using safety reasons.
e) Exhaustion of the soda lime soda lime.
can be detected by an
end-tidal CO2 rebreathing
waveform.
Answers 71
Answers
to avoid channelling and b) True. Jackson-Rees added a e) False. The circle system can
incomplete CO2 absorption. double-ended bag to the be used with low flows (e.g.
reservoir tubing of the 23L/min) as well as very
5. The Bain breathing system: Mapleson E thus converting it low flows (e.g. 0.51.5L/
a) False. The Bain breathing to a Mapleson F. The bag acts min). This can be achieved
system is an example of a as a visual monitor during safely with adequate
Mapleson D system. spontaneous ventilation and monitoring of the inspired
b) False. A FGF of about can be used for assisted or and exhaled concentration of
150200mL/kg/min is controlled ventilation. the gases and vapours used.
required to prevent c) False. The T-piece system is
rebreathing in the Bain system not an efficient system as it 8. The circle breathing system:
during spontaneous requires an FGF of 2.53 a) True. Substance A can be
ventilation. This makes it an times the minute volume to produced when sevoflurane is
inefficient breathing system. prevent rebreathing. used with soda lime under
c) False. The Bain system is d) False. If the reservoir is too low flow rates. Newer designs
usually made of a coaxial small, entrainment of ambient of soda lime claim lesser
tubing. A more recent design, air will occur resulting in production of substance A.
a pair of parallel corrugated dilution of the FGF. b) False. The Goldman vaporizer
tubings, is also available. e) True. Patients under the age is a VIC. It is positioned on
d) False. The Bain breathing of 6 years have a small FRC. the expiratory limb with
system is a Mapleson D General anaesthesia causes a minimal resistance to flow
whereas the T-piece system further decrease in the FRC. and no wicks.
is an E system. The Bain The reservoir bag in the c) True. The function of the
system has an APL valve Mapleson F provides a degree unidirectional valves in the
whereas the T-piece is a of CPAP during spontaneous circle system is crucial for its
valveless system. ventilation, helping to function. Failure of the valve
e) True. The Bain system can be improve the FRC. to close causes rebreathing
used for spontaneous resulting from the huge
ventilation requiring an FGF 7. Which of the following are true increase in the dead space of
of 150200mL/kg. It can or false? the system. This usually
also be used for controlled a) False. Mapleson classification happens because of water
ventilation requiring an FGF is used. The Mapleson A vapour condensing on the
of 70100mL/kg. During breathing system was valve.
controlled ventilation, the described by Magill. d) False. The circle system can
APL valve is fully closed, the b) False. As modern anaesthetic be used for both spontaneous
reservoir bag is removed and agents are not flammable, and controlled ventilation.
a ventilator like the Penlon modern breathing systems are Soda lime increases the
Nuffield 200 with a 1m not constructed using anti- resistance to flow but is
length of tubing is connected static materials. They are clinically insignificant.
instead. normally made of plastic. e) True. As the soda lime gets
c) True. Efficiency of a breathing exhausted, a rebreathing
6. T-piece breathing system: system differs during end-tidal CO2 waveform can
a) False. Although the T-piece spontaneous and controlled be detected. A dye is added
breathing system is mainly ventilation; e.g. the Mapleson that changes the granules
used in paediatrics, it can be A system is more efficient colour as they become
used in adults with a suitable during spontaneous than exhausted.
FGF and reservoir volume. An controlled ventilation whereas
FGF of 2.53 times the the opposite is true of the D 9. Regarding circle system:
minute volume and a reservoir system. a) False. High FGF (several
approximating the tidal d) False. The position of the L/min) is needed initially to
volume are needed. Such a valve in the breathing system denitrogenate the system and
system is usually used in is crucial in the function and the functional residual
recovery and for intensive efficiency of a breathing capacity (FRC) to avoid the
care patients. system. build-up of unacceptable
Answers 73
levels of nitrogen in the c) False. The lower the FGF, the d) False. It is not an efficient
system. In closed circle more rapidly soda lime system as the granules nearest
anaesthesia, a high FGF for granules are exhausted to the patient are exhausted
up to 15 minutes, and in because most of the exhaled first so increasing the dead
low-flow anaesthesia, a high gases pass through the space.
FGF of up to 6 minutes are absorber with very little being e) True. Substance A,
required and these can later discarded through the APL nephrotoxic in rats, can be
be reduced to 0.51L/min. If valve. For a 7080-kg patient produced when sevoflurane is
no N2O is used during with a tidal volume of used with soda lime although
anaesthesia, it is not necessary 500mL, respiratory rate of newer designs claim less or no
to eliminate nitrogen. A short 12/min and CO2 production substance A production.
period of high flow is needed of 250mL/min, using an FGF Carbon monoxide can occur
to prime the system and the of 1L/min, the soda lime will when dry soda lime is used.
patient with the inhalational be exhausted in 57h of use. Newer designs claim less or
agent. For the same patient but using no carbon monoxide
b) False. The pH of soda lime is an FGF of 3L/min, the soda production.
highly alkaline, 13.5, because lime will be exhausted in
of the presence of calcium 68h of use. 10. d)
hydroxide, sodium hydroxide
and small amounts, if any, of
potassium hydroxide. This
makes the soda lime a
corrosive substance. Colour
changes occur when the pH is
less than 10.
This page intentionally left blank
75
Chapter 5
Tracheal and
tracheostomy
tubes and
airways
76 5 Tracheal and tracheostomy tubes and airways
15 mm connector
Tracheal tubes
Tracheal tubes provide a means of Self-sealing valve
securing the patients airway. These
disposable plastic tubes are made of Pilot balloon
polyvinyl chloride (PVC) which
could be clear, ivory or siliconized.
As plastic is not radio-opaque,
tracheal tubes have a radio-opaque
line running along their length,
which enables their position to be
determined on chest X-rays. The
siliconized PVC aids the passage of
suction catheters through the tube. Radio-opaque line
In the past, tracheal tubes used to
be made of rubber allowing them to
be reused after cleaning and
autoclaving.
Cuff OD Outside diameter mm
Bevel
A B
2. The narrowest point in the
Fig. 5.2 (A) Correctly positioned tracheal tube. (B) The tracheal tube has been advanced adults airway is the glottis
too far, into the right main bronchus. (which is hexagonal). In order to
achieve an air-tight seal, cuffed
the tracheal tube tip within the tubes are used in adults.
trachea. The vocal cords should 3. The narrowest point in a childs
be at the black mark in tubes airway is the cricoid cartilage.
with one mark, or should be Since this is essentially circular,
between marks if there are two a correctly sized uncuffed tube
such marks. However, these are will fit well. Because of the
only rough estimates and correct narrow upper airway in children,
tracheal tube position depth post-extubation subglottic
should always be confirmed by oedema can be a problem. In
auscultation. order to minimize the risk, the
presence of a small leak around
Murphy the tube at an airway pressure of
The bevel 15cmH2O is desirable.
eye
1. The bevel is left-facing and oval 4. Cuffs can either be:
in shape in most tube designs. a) high pressure/low volume
A left-facing bevel improves the b) low pressure/high volume.
view of the vocal cords during
intubation.
High-pressure/low-volume cuffs
2. Some designs have a side hole
just above and opposite the 1. These can prevent the passing of
bevel, called a Murphy eye. vomitus, secretions or blood into
This enables ventilation to the lungs.
occur should the bevel become Fig. 5.3 Diagram showing a tracheal 2. At the same time, they exert a
occluded by secretions, blood or tube with an obstructed bevel against high pressure on the tracheal
the wall of the trachea (Fig. 5.3). the trachea wall but a patent Murpheys wall. If left in position for long
eye so allowing ventilation.
periods, they may cause necrosis
1. The cuff is connected to its of the tracheal mucosa (Fig. 5.5).
The cuff
pilot balloon which has a
Tracheal (oral or nasal) tubes can be self-sealing valve for injecting
Low-pressure/high-volume cuffs
either cuffed or uncuffed. The cuff, air. The pilot balloon also
when inflated, provides an air-tight indicates whether the cuff is 1. These exert minimal pressure on
seal between the tube and the inflated or not. After intubation, the tracheal wall as the pressure
tracheal wall (Fig. 5.4). This the cuff is inflated until no gas equilibrates over a wider area
air-tight seal protects the patients leak can be heard during (Fig. 5.6). This allows the cuff
airway from aspiration and allows intermittent positive pressure to remain inflated for longer
efficient ventilation during IPPV. ventilation (IPPV). periods.
78 5 Tracheal and tracheostomy tubes and airways
Small
contact area
Tracheal wall
High volume cuff
Large
contact area
(Fig. 5.13). It is used mainly during be either north- or south-facing, LASER RESISTANT
anaesthesia for maxillofacial surgery cuffed or uncuffed. TRACHEAL TUBES
as it does not impede surgical Because of its preformed shape,
access. Because of its design and there is a higher risk of bronchial These tubes are used in anaesthesia
shape, it lies over the nose and the intubation than with ordinary for laser surgery on the larynx or
forehead. It can be converted to an tracheal tubes. The cuffed RAE trachea (Fig. 5.14). They are
ordinary tracheal tube by cutting it tracheal tube has one Murphy eye designed to withstand the effect of
at the scissors mark just proximal to whereas the uncuffed version has carbon dioxide and potassium-
the pilot tube and reconnecting the two eyes. Since the uncuffed version titanyl-phosphate (KTP) laser
15-mm connector. An oral version is mainly used in paediatric practice, beams, avoiding the risk of fire or
of the polar tube exists. two Murphy eyes ensure adequate damage to the tracheal tube. One
The RAE (Ring, Adair and ventilation should the tube prove design has a flexible stainless steel
Elwyn) tube has a preformed shape too long. body. Reflected beams from the
to fit the mouth or nose without The tube can be temporarily tube are defocused to reduce the
kinking. It has a bend located straightened to insert a suction accidental laser strikes to healthy
just as the tube emerges, so the catheter. tissues (Fig. 5.15). Other designs
connections to the breathing system
are at the level of the chin or
forehead and not interfering with
the surgical access. RAE tubes can
Fig. 5.14 Laser resistant tracheal tubes. Note the stainless steel tube (left) with two
cuffs. The tube on the right is covered with laser protective wrapping.
Laser beam
Components
1. An introducer used for insertion.
2. Wings attached to the proximal
part of the tube to fix it in place
with a ribbon or suture. Some
designs have an adjustable flange
to fit the variable thickness of
the subcutaneous tissues
(Fig. 5.19).
3. They can be cuffed or uncuffed.
The former have a pilot balloon.
Fig. 5.16 Evoked potential tracheal tube. Note the electrodes (just above the cuff) with 4. The proximal end can have a
their cables. The other cable is earth. standard 15-mm connector.
82 5 Tracheal and tracheostomy tubes and airways
Problems in practice and Components this, double lumen tubes are used
safety features which allow the anaesthetist to
1. A scalpel and syringe selectively deflate one lung while
Percutaneous insertion of 2. A needle with a veress design maintaining standard ventilation
minitracheostomy has the risk of: and a dilator. The needle has a of the other.
red flag indicator. This helps in
1. pneumothorax
locating the tissues.
2. perforation of the oesophagus Components
3. 6-mm cuffed tube.
3. severe haemorrhage
4. ossification of the cricothyroid 1. The Mallinckrodt Bronchocath
membrane Mechanism of action double lumen tube has two
5. incorrect placement. separate colour-coded lumens,
1. After a 2-cm horizontal skin each with its own bevel (Fig.
incision has been made, the 5.29). One lumen ends in the
needle is inserted perpendicular trachea and the other lumen
to the skin. ends in either the left or right
Minitracheostomy 2. As the needle enters the trachea,
Tube inserted into the trachea
main bronchus.
the red indicator disappears. The 2. Each lumen has its own cuff
through the cricothyroid needle is advanced carefully until
membrane using the Seldinger
(tracheal and bronchial cuffs)
the red reappears, indicating and colour-coded pilot balloons.
technique. contact with the posterior wall
Used for clearing secretions and
Both lumens and pilot balloons
of the trachea. are labelled.
maintaining an airway in an 3. As the cricothyrotomy tube is
emergency.
3. There are two curves to the
advanced into the trachea, the tube: the standard anterior curve
needle and the dilator are to fit into the oropharyngeal
removed. laryngeal tracheal airway and
the second curve, either to the
Problems in practice and right or left, to fit into the right
Cricothyrotomy tube safety features or left bronchus respectively.
4. The proximal end of these tubes
(Fig. 5.28) The cricothyrotomy tube has is connected to a Y-shaped
complications similar to the catheter mount attached to the
This tube is used to maintain the minitracheostomy tube. breathing system.
airway in emergency situations such
as on the battlefield. It is inserted
into the trachea through the Cricothyrotomy Mechanism of action
cricothyroid cartilage. A cuffed 6-mm tube is inserted 1. Because of the differing anatomy
into the trachea through the of the main bronchi and their
cricothyroid cartilage. branches, both right and left
A veress needle is designed to versions of any particular double
locate the trachea. lumen tube must exist.
Used in emergencies to establish 2. Once the tubes are correctly
an airway. positioned, the anaesthetist can
selectively ventilate one lung. So,
for operations requiring that the
right lung is deflated, a left-sided
double lumen tube would be
Double lumen used that enabled selective
endobronchial tubes ventilation of the left lung alone
and vice versa.
During thoracic surgery, there is a 3. It is desirable, when possible, to
need for one lung to be deflated. insert a left double lumen tube
This offers the surgeon easier and instead of a right one. This
Fig. 5.28 Smiths Portex cricothyrotomy better access within the designated reduces the risk of upper lobe
set. hemithorax. In order to achieve bronchus obstruction by the
Double lumen endobronchial tubes 87
Oropharyngeal
airway
This anatomically shaped airway is
inserted through the mouth into the
oropharynx above the tongue to
maintain the patency of the upper
Fig. 5.31 Bronchocath double lumen tube with the bronchial lumen connected to a CPAP
valve assembly. The disconnected limb of the Y-shaped catheter mount has been clamped.
airway (Fig. 5.34) in cases of upper
airway obstruction caused by a
decreased level of consciousness in a
patient. Decreased consciousness
Double lumen endobronchial can lead to loss of pharyngeal tone
tubes that can result in airway obstruction
Two separate lumens each with by the tongue, epiglottis, soft palate
its own cuff and pilot tube. or pharyngeal tissues. There are
There are two curves, anterior various regularly used types of
and lateral. oropharyngeal airway. The most
The right-sided version has an common type is the Guedel airway,
eye in the bronchial cuff to named after its developer Arthur
facilitate ventilation of the right Guedel, an American anaesthetist
upper lobe. who served in France during the
Commonly used ones are First World War. It is available in
Robertshaw, Bronchocath and up to nine sizes, which have a
Fig. 5.32 The Arndt endobronchial
Carlens (and White). The latter blocker set. standardized number coding (the
has a carinal hook. smallest 000 to the largest 6).
Endobronchial
blocker Fig. 5.33 Detail of the blocker emerging
from a standard endotracheal tube. The
The endobronchial blocker is an guide loop is used to advance the blocker
over a paediatric fibreoptic bronchoscope
alternative means to the double
into the desired main bronchus.
lumen tube for providing one-lung
ventilation (Fig. 5.32).
Mechanism of action
Components
The patient is intubated with a
1. The blocker catheter. This is a standard endotracheal tube. A
9-FG, 78-cm catheter that has a specially designed multiport adapter
distal cuff inflated via a pilot is connected to the tubes standard
balloon. A guide loop emerges 15-mm connector. The blocker can
from its tip (Fig. 5.33). be advanced over a paediatric Fig. 5.34 An oropharyngeal (Guedel)
2. Multiport adapter. fibreoptic bronchoscope into the airway. (Courtesy of Smiths Medical.)
Nasopharyngeal airway 89
9. Extraglottic airways would tube and the cuff to prevent design improves the seal
normally elicit airway reflexes the epiglottis from obstructing pressure. A single-use version,
such as the gag reflex, and the lumen of the laryngeal mask. LMA Supreme, is available
therefore require depression of Newer designs, such as Portex which combines the best features
pharyngeal reflexes by general SoftSeal and Intersurgical of previous LMA versions, and
or topical anaesthesia. Solus, omit the bars with no contains an elliptical and
10. These devices are increasingly adverse clinical effects. anatomically shaped curve,
used in a variety of settings, 4. A modified design (LMA- which facilitates insertion
including routine anaesthesia, ProSeal) has an additional success and provides a double
emergency airway management lumen (drain tube) lateral to the seal. A first seal is important for
and as an aid to intubation. airway tube and traverses the adequacy of gas exchange, better
floor of the mask to open in the known as the oropharyngeal seal
mask tip opposite the upper (Fig. 5.38). It also incorporates
oesophageal sphincter allowing a second seal, designed to
blind passage of an orogastric reduce the risk of stomach
Laryngeal mask tube and helps in the drainage insufflation during ventilation,
of gastric air or secretions. Both to provide a passive conduit for
This very useful device is frequently
tubes are contained within an (unexpected) regurgitation or
used as an alternative to either the
integrated bite block. The cuff active suctioning of gastric
face mask or tracheal tube during
inflates in a three-dimensional content and enhances the
anaesthesia (Fig. 5.37).
manner with the elliptical cuff effectiveness of the first seal.
augmented by a second cuff 5. Low-cost disposable laryngeal
Components behind the bowl, known as the masks have been introduced and
rear boot or dorsal cuff. This are widely used (Fig. 5.39).
1. A transparent tube of wide
internal diameter. The proximal
end is a standard 15-mm
connection.
2. An elliptical cuff at the distal
end. The cuff resembles a small
face mask to form an air-tight
seal around the posterior
perimeter of the larynx and is
inflated via a pilot balloon with
a self-sealing valve. A non-
metallic self-sealing valve is
available for use during magnetic
resonance imaging (MRI) scans.
3. The original design (Intavent
Classic LMA) had two slits or
bars at the junction between the
Fig. 5.38 LMA-Supreme. Note the drainage lumen.
The intubating
laryngeal mask
airway (ILMA)
This is a modification of the Fig. 5.43 Single-use ILMA. (Courtesy of Intavent Direct.)
laryngeal mask designed to facilitate
tracheal intubation with a tracheal
tube either blindly or in conjunction 2. Despite the presence of the 4. Unlike the tracheal tube,
with a fibrescope while minimizing slits or bars, about 10% of rotation of the laryngeal mask
the requirements for head and patients develop airway may result in complete airway
neck manipulation. The specially obstruction because of obstruction. In order to assess
designed laryngeal mask is inserted down-folding of the epiglottis. the laryngeal masks orientation
first (Fig. 5.41). A specially designed Although clinically often when inserted, a black line is
tracheal tube is then passed through insignificant, a higher present on the tube. This should
the laryngeal mask through the proportion of obstructions by face the upper lip of the patient
vocal cords into the trachea (Fig. the epiglottis can be observed when the laryngeal mask is in
5.42). Single-use ILMA is endoscopically. position.
available (Fig. 5.43). 3. The manufacturers recommend 5. Cricoid pressure may prevent
using the laryngeal masks for correct placement of the
a maximum of 40 times. The laryngeal mask.
Problems in practice and
cuff is likely to perish after 6. A common cause of airway
safety features
autoclaving. A record card that obstruction during laryngeal
1. The laryngeal mask does not accompanies the laryngeal mask anaesthesia is down-
protect against the aspiration of mask registers the number of folding of the epiglottis, which
gastric contents. autoclaving episodes. occurs in 2056% of patients.
94 5 Tracheal and tracheostomy tubes and airways
COBRA perilaryngeal
airway (PLA)
The COBRA-PLA consists of a
large ventilation tube with a distal
circumferential inflatable cuff,
designed to reside in the
hypopharynx at the base of the
tongue, sealing off the upper Fig. 5.45 COBRA PLA Plus device.
Fig. 5.44 The i-gel airway. (Courtesy of oropharynx. It differs from other Note the temperature probe used for
Intersurgical.) extraglottic airway devices as the core temperature monitoring.
Further reading 95
MCQs
In the following lists, which of the statements (a) to (e) are true?
Answers
makes the laryngeal mask e) True. It is sometimes difficult
1. Concerning tracheal tubes:
more suitable for use in to ensure correct positioning
a) False. An RAE tube is a
spontaneously breathing of the double lumen
normal size preformed
patients for long periods endobronchial tube. By using
tracheal tube. It does not
of time. a fibreoptic bronchoscope, the
allow good visibility of the
d) False. The laryngeal mask can position of the tube can be
larynx because of its large
be autoclaved up to 40 times. adjusted to ensure correct
diameter. A microlaryngeal
The cuff is likely to perish positioning.
tracheal tube of 56mm ID
after repeated autoclaving. A
is more suitable for
record should be kept of the 4. Concerning the tracheal tube
microlaryngeal surgery,
number of autoclaves. cuff during anaesthesia:
allowing good visibility and
e) False. The standard laryngeal a) False. The design of the
access to the larynx.
mask has a metal component low-pressure/high-volume cuff
b) True. Because the shape of
in the one-way inflating valve. allows wrinkles to be formed
these tubes is fixed, they
This makes it unsuitable for around the tracheal wall. The
might not fit all patients of
use in MRI. A specially presence of the wrinkles
different sizes and shapes; e.g.
designed laryngeal mask with allows aspiration of gastric
a small, short-necked patient
no metal parts is available for contents to occur.
having an RAE tube inserted
MRI use. b) False. The rise in the intracuff
is at risk of an endobronchial
pressure is mainly due to the
tube position.
3. Double lumen endobronchial diffusion of N2O. Minimal
c) True. Some anaesthetists use
tubes: changes are due to diffusion of
the laryngeal mask in nasal
a) False. The Robertshaw double oxygen (from 21% to say
surgery with a throat pack.
lumen tube does not have a 33%) and because of increase
This technique has a higher
carinal hook. The Carlens in the temperature of the air
risk of aspiration.
double lumen tube has a in the cuff (from 21 to 37C).
d) False. RAE stands for the
carinal hook. The diffusion of inhalational
initials of the designers (Ring,
b) False. Left-sided tubes do not agents causes minimal changes
Adair and Elwyn).
have an eye in the bronchial in pressure due to the low
e) False. The Oxford tube is one
cuff to facilitate ventilation of concentrations used (12%).
of the few tracheal tubes with
the left upper lobe. This is New design material cuffs
a front-facing bevel. This
because the distance between prevent the diffusion of gases
might make intubation more
the carina and the upper lobe thus preventing significant
difficult as it obscures the
bronchus is about 5cm, which changes in pressure.
larynx.
is enough for the bronchial c) True. The high pressures
cuff. Right-sided tubes have achieved by the high-pressure/
2. Laryngeal masks:
an eye to facilitate ventilation low-volume cuffs, especially
a) False. Laryngeal masks do not
of the right upper lobe during nitrous oxide
protect the airway from the
because the distance between anaesthesia, can cause necrosis
risks of aspiration.
the carina and the upper lobe to the mucosa of the trachea
b) False. The bars in the cuff are
bronchus is only 2.5cm. if left in position for a long
designed to prevent the
c) True. Carlens double lumen period.
epiglottis from blocking the
tubes have relatively small d) True. Because of the design of
lumen of the tube.
lumens in comparison to the low-volume cuffs, a seal
c) True. The laryngeal mask has
the Robertshaw double can be maintained against a
a large internal diameter, in
lumen tube. relatively small area of the
comparison with a tracheal
d) True. CPAP can be applied to tracheal wall. In the case of
tube. This reduces the
the deflated lung to improve the high-volume/low-pressure
resistance to breathing which
oxygenation during one lung cuffs, a large contact area on
is of more importance during
anaesthesia. the tracheal wall is achieved.
spontaneous breathing. This
98 5 Tracheal and tracheostomy tubes and airways
e) False. The pressure in the cuff b) False. situation where the bevel of
may decrease because of a c) False. An armoured tube the tube is occluded.
leak in the cuff or pilot should not be cut, as that will e) False. The bevel of the tube is
balloons valve. cut the spiral present in its usually left-facing to allow
wall. This increases the risk of easier visualization of the
5. Concerning tracheal tubes: tube kinking. vocal cords. The tracheostomy
a) False. The ID is measured in d) False. A Murphy eye allows tube has a square-cut tip.
millimetres. pulmonary ventilation in the
6. e)
99
Chapter 6
Masks and
oxygen
delivery
devices
100 6 Masks and oxygen delivery devices
Nasal masks
(inhalers)
1. These masks are used during
dental chair anaesthesia.
2. An example is the Goldman
inhaler (Fig. 6.3) which has an
inflatable cuff to fit the face and
an adjustable pressure limiting
(APL) valve at the proximal end.
The mask is connected to tubing
which delivers the fresh gas flow.
Variable performance
masks (medium
concentration; MC)
These masks are used to deliver
oxygen-enriched air to the patient
Fig. 6.4 Catheter mount. (Fig. 6.5). They are also called
102 6 Masks and oxygen delivery devices
Components
1. The plastic body of the mask
with side holes on both sides.
2. A port connected to an oxygen
supply.
3. Elastic band(s) to fix the mask
to the patients face.
Mechanism of action
1. Ambient air is entrained through
the holes on both sides of the
mask. The holes also allow
exhaled gases to be vented out.
2. During the expiratory pause,
the fresh oxygen supplied helps
in venting the exhaled gases
through the side holes. The body
of the mask (acting as a reservoir)
is filled with fresh oxygen supply
and is available for the start of
the next inspiration.
3. The final concentration of Fig. 6.5 (A) Adult variable performance face mask. (B) Paediatric variable performance
inspired oxygen depends on: face mask.
a) the oxygen supply flow rate
b) the pattern of ventilation. Table 6.2 Factors that affect the delivered FiO2 in the variable
If there is a pause between performance masks
expiration and inspiration, the
mask fills with oxygen and a
High FiO2 delivered Low FiO2 delivered
high concentration is available
at the start of inspiration Low peak inspiratory flow rate High peak inspiratory flow rate
c) the patients inspiratory flow Slow respiratory rate Fast respiratory rate
rate. During inspiration, High fresh oxygen flow rate Low fresh oxygen flow rate
oxygen is diluted by the air Tightly fitting face mask Less tightly fitting face mask
drawn in through the holes
Nasal cannulae 103
the side holes is high (when the continuous fresh oxygen supply, The inspired oxygen
mask is a good fit). The patients oxygen present in the reservoir concentration depends on the
may experience a sense of (a mixture of the fresh oxygen oxygen flow rate, pattern and
warmth and humidity, indicating and exhaled oxygen) and rate of ventilation, maximum
significant rebreathing. ambient air. Higher variable inspiratory flow rate and how
6. A typical example of 4L/min of FiO2 can be achieved with such well the mask fits the patients
oxygen flow delivers an FiO2 of masks. A one-way valve is fitted face.
about 0.350.4 providing there between mask and reservoir to
is a normal respiratory pattern. prevent rebreathing.
7. Adding a 600800mL bag to 8. Some designs have an extra port
the mask will act as an extra attached to the body of the Nasal cannulae
reservoir (Fig. 6.6). Such mask allowing it to be connected
masks are known as partial to a side-stream CO2 monitor Nasal cannulae are ideal for patients
rebreathing masks. The inspired (Fig. 6.7). This allows it to on long-term oxygen therapy (Fig.
oxygen is derived from the sample the exhaled CO2 so 6.9). A flow rate of 24L/min
104 6 Masks and oxygen delivery devices
Open oxygen
delivery systems
These are designed to offer the
Fig. 6.10 Oxygen nasal catheter.
Fig. 6.9 Oxygen nasal cannula. maximum comfort to patients
while delivering variable FiO2
concentrations. They fit around the
delivers an FiO2 of 0.280.36 patients head like head phones, so
respectively. Higher flow rates are Nasal cannulae making minimal physical contact
uncomfortable. Entrainment of ambient air (Fig. 6.11). Such systems are
through the nostrils and during suitable for both nasal and mouth
mouth breathing. breathing patients. They may be
Components The FiO2 depends on the oxygen more suitable for patients on
1. Two prongs which protrude flow rate, tidal volume, long-term oxygen therapy.
about 1cm into the nose. inspiratory flow rate, respiratory A wide range of fresh oxygen
2. These are held in place by an rate and the volume of the flows can be used, so delivering a
adjustable head strap. nasopharynx. variable performance. As with the
Better compliance compared other devices of this kind, similar
with facial masks. factors will affect its performance.
Mechanism of action
1. There is entrainment of ambient
air through the nostrils. The
nasopharynx acts as a reservoir. Nasal catheters Fixed performance
2. The FiO2 achieved is
proportional to: Nasal catheters comprise a single devices
a) the flow rate of oxygen lumen catheter, which is lodged into
b) the patients tidal volume, the anterior naris (nostril) by a VENTURI MASK
inspiratory flow and foam collar (Fig. 6.10). Oxygen
respiratory rate flows of 23L/min can be used. These masks are fixed performance
c) the volume of the nasopharynx. The catheter can be secured to the devices (sometimes called high-air-
3. Mouth breathing causes patients face by using tape. It flow oxygen enrichment, or
inspiratory air flow. This should not be used when a nasal HAFOE).
produces a Venturi effect in the
posterior pharynx entraining
oxygen from the nose.
4. There is increased patient
compliance with nasal cannulae
compared to facial oxygen
masks. The patient is able to
speak, eat and drink.
Air
Air
O2
Mechanism of action
1. The Venturi mask uses the
Bernoulli principle, described
in 1778, in delivering a
Orifice predetermined and fixed
concentration of oxygen to
the patient. The size of the
constriction determines the final
concentration of oxygen for a
given gas flow. This is achieved
in spite of the patients
respiratory pattern by providing
a higher gas flow than the peak
inspiratory flow rate.
2. As the flow of oxygen passes
through the constriction, a
negative pressure is created. This
causes the ambient air to be
entrained and mixed with the
oxygen flow (Fig. 6.14). The
Fig. 6.13 Detail of the Venturi device. Design for administering 60% oxygen (left) and FiO2 is dependent on the degree
24% (right). Note the difference in the recommended oxygen flow rates and the size of of air entrainment. Less
orifice and air entrainment apertures. entrainment ensures a higher
FiO2 is delivered. This can be
recommended oxygen flow rate achieved by using smaller
Components
to provide the desired oxygen entrainment apertures or bigger
1. The plastic body of the mask concentration (Figs 6.12 and windows to entrain ambient
with holes on both sides. 6.13). air. The smaller the orifice is, the
2. The proximal end of the mask 3. Alternatively, a calibrated greater the negative pressure
consists of a Venturi device. variable Venturi device can generated, so the more ambient
The Venturi devices are colour- be used to deliver the desired air entrained, the lower the
coded and marked with the FiO2. FiO2. The oxygen concentration
106 6 Masks and oxygen delivery devices
can be 0.24, 0.28, 0.31, 0.35, 50L/min, well above the peak obstructive pulmonary disease.
0.4 or 0.6. inspiratory flow rate. However, caution should be
3. The Bernoulli effect can be 7. The masks side holes are used exercised as it has been shown
written as: to vent the exhaled gases only that the average FiO2 delivered
(as above) in comparison to in such masks is up to 5%
P + 1 2 2 =
the side holes in the variable above the expected value.
where is the density, v is the performance mask where the 2. The Venturi mask with its
velocity, P is the pressure. side holes are used to entrain Venturi device and the oxygen
4. The total energy during a fluid inspired air in addition to expel delivery tubing is often not well
(gas or liquid) flow consists of exhaled gases. tolerated by patients because it
the sum of kinetic and potential 8. The Venturi face masks are is noisy and bulky.
energy. The kinetic energy is designed for both adult and
Anaesthetic breathing systems
related to the velocity of the paediatric use (Fig. 6.14).
are other examples of the fixed
flow whereas the potential 9. The Venturi attachments, with
performance devices. The reservoir
energy is related to the pressure. a reservoir tubing, can be
bag acts to deliver a fresh gas flow
As the flow of fresh oxygen attached to a tracheal tube or
that is greater than the patients
passes through the constricted a supraglottic airway device as
peak inspiratory flow rate.
orifice into the larger chamber, part of a T-piece breathing
the velocity of the gas increases system (Fig. 6.15). This
distal to the orifice causing the arrangement is usually used in
kinetic energy to increase. As the recovery wards to deliver Venturi mask
total energy is constant, there is oxygen-enriched air to patents. Fixed performance device
a decrease in the potential (HAFOE).
energy so a negative pressure is Uses the Venturi principle to
Problems in practice and
created. This causes the ambient entrain ambient air.
safety features
air to be entrained and mixed No rebreathing or increase in
with the oxygen flow. The FiO2 1. These masks are recommended dead space.
is dependent on the degree of air when a fixed oxygen Changes in kinetic and potential
entrainment. Less entrainment concentration is desired in energy during gas flow lead to
ensures higher FiO2 is delivered patients whose ventilation is negative pressure and air
and smaller entrainment dependent on their hypoxic entrainment.
apertures are one method of drive, such as those with chronic
achieving this (Fig. 6.8). The
devices must be driven by the
correct oxygen flow rate,
calibrated for the aperture size
if a predictable FiO2 is to be
achieved.
4. Because of the high fresh gas
flow rate, the exhaled gases are
rapidly flushed from the mask,
via its holes. Therefore there is
no rebreathing and no increase
in dead space.
5. These masks are recommended
when a fixed oxygen
concentration is desired in
patients whose ventilation is
dependent on the hypoxic drive. A
6. For example, a 24% oxygen
Venturi mask has an air:oxygen
entrainment ratio of 25:1. This B
means an oxygen flow of 2L/ Fig. 6.15 (A) Adult and paediatric Venturi masks. (B) Venturi device as part of a
min delivers a total flow of breathing system.
Further reading 107
FURTHER READING the spontaneously breathing patient. of three devices for oxygen
CPD Anaesthesia 4 (1), 2730. administration in the late
Agusti, A.G., Carrera, M., Barbe, F., MHRA, 2011. Medical device alert: postoperative period. British Journal
et al., 1999. Oxygen therapy during oxygen masks manufactured by of Anaesthesia 74 (5), 607609.
exacerbations of chronic obstructive Lifecare Hospital Supplies Ltd Waldau, T., Larsen, V.H., Bonde, J.,
pulmonary disease. European (MDA/2011/015). Online. Available 1998. Evaluation of five oxygen
Respiratory Journal 14, 934939. at: http://www.mhra.gov.uk/ delivery devices in spontaneously
British Thoracic Society. Online. Available Publications/Safetywarnings/ breathing subjects by oxygraphy.
at: http://www.brit-thoracic.org.uk MedicalDeviceAlerts/CON108738 Anaesthesia 53, 256263.
Khakhar, M., Heah, T., Al-Shaikh, B., Stausholm, K., Rosenberg-Adamsen, S.,
2002. Oxygen delivery systems for Skriver, M., et al., 1995. Comparison
108 6 Masks and oxygen delivery devices
MCQs
In the following lists, which of the statements (a) to (e) are true?
1. Concerning the Venturi mask: 4. Concerning the oxygen nasal 7. Concerning fixed performance
a) Gas flow produced should be cannula: devices:
more than 20L/min. a) Is a fixed performance device. a) Anaesthetic breathing systems
b) Reducing the flow of oxygen b) Is a variable performance with reservoirs are fixed
from 12 to 8L/min results in device. performance devices.
a reduction in oxygen c) There is a Venturi effect in the b) Distal to the constriction of a
concentration. posterior pharynx. Venturi, there an increase in
c) With a constant oxygen supply d) An oxygen flow of 8L/min is potential energy.
flow, widening the orifice in usually used in an adult. c) In a Venturi mask, the higher
the Venturi device increases e) Has increased patient the entrainment ratio, the
the oxygen concentration compliance. higher the FiO2.
delivered to the patient. d) A nasal oxygen catheter is a
d) There is rebreathing in the fixed performance device.
5. Variable performance masks:
mask. e) Venturi masks are very well
a) During slow and deep
e) The mask is a fixed tolerated by patients.
breathing, a higher FiO2 can
performance device.
be achieved.
b) Ambient air is not entrained
SINGLE BEST ANSWER (SBA)
2. High-air-flow oxygen enrichment into the mask.
face masks: c) Alveolar gas rebreathing is
8. Catheter mounts:
a) Use the Venturi principle to not possible.
a) Should have a 22-mm
deliver a fixed O2 d) Normal inspiratory peak flow
connector at the distal
concentration to the patient. rate is 2030L/min for an
(patient) end.
b) The size of the constriction of adult.
b) Should have a 15-mm
the Venturi has no effect on e) Can be used safely on all
connector at the proximal
the final O2 concentration patients.
(machine) end.
delivered to the patient.
c) Gas sampling ports should
c) The holes on the side of the
6. Regarding variable performance always be built into the
mask are used to entrain
devices: structure.
ambient air.
a) They can offer greater patient d) Should never incorporate an
d) The gas flow delivered to the
compliance. angle piece.
patient is more than the peak
b) They can deliver an FiO2 that e) Should have a 15-mm
inspiratory flow rate.
can vary from breath to connector at the distal
e) There is significant rebreathing.
breath in the same patient. (patient) end.
c) The size of the medium
3. Face masks used during concentration oxygen face
anaesthesia: mask has no effect on
a) The rubber mask is covered rebreathing and CO2
by carbon particles which act elimination.
as an anti-static measure. d) Capnography can be used to
b) Masks have no effect on the measure the FiO2 delivered to
apparatus dead space. the patient.
c) The masks cuff has to e) With a variable performance
be checked and inflated mask, the FiO2 is higher when
before use. the face mask is a tight fit.
d) The dental nasal masks are
also known as nasal inhalers.
e) Masks have a 15-mm end to
fit the catheter mount.
Answers 109
Answers
in pressure at the constriction. apparatus dead space if the
1. Concerning the Venturi mask:
This leads to entrainment of wrong size is chosen. In an
a) True. The Venturi mask is a
ambient air leading to mixture adult, the dead space can
fixed performance device. In
of fixed oxygen concentration. increase by about 200mL. It
order to achieve this, the flow
b) False. It is the size of the is of more importance in
delivered to the patient should
orifice that determines the paediatric practice.
be more than the peak
degree of decrease in pressure c) True. The cuff of the face
inspiratory flow rate. A flow
at the constriction. This mask is designed to ensure a
of more than 20L/min is
determines the amount of snug fit over the patients face
adequate.
ambient air being entrained, and also to minimize the
b) True. It is the flow rate of
hence the final concentration masks pressure on the face.
oxygen through the orifice
of oxygen. Ensuring that the cuff is
that determines the final FiO2
c) False. In such a mask, the inflated before use is therefore
the patient receives. With a
holes are used to expel the important.
constant orifice, the amount
exhaled gases. There is no d) True. Nasal inhalers are nasal
of air entrained remains
entrainment of ambient air in masks used during dental
constant. So by reducing the
such a mask because of the anaesthesia allowing good
oxygen flow rate from 12 to
high gas flows. The holes in a surgical access to the mouth.
8L/min, there will be less
variable performance mask The Goldman nasal inhaler is
oxygen in the final mixture.
are used to entrain ambient an example.
c) True. The wider the orifice,
air. e) False. The face masks have a
the less the drop in pressure
d) True. The gas flow generated 22-mm end to fit the angle
across the orifice and the less
is higher than the peak piece or catheter mount.
the entrainment of the
inspiratory flow rate. This
ambient air, hence the less the
allows the delivery of a fixed 4. Concerning the oxygen nasal
dilution of the O2, resulting in
oxygen concentration to the cannula:
an increase in oxygen
patient regardless of the a) False. It is not a fixed
concentration delivered to the
inspiratory flow rate. It also performance device. The final
patient. The opposite is also
prevents rebreathing. FiO2 depends on the flow rate
correct.
e) False. There is no rebreathing of oxygen, tidal volume,
d) False. There is no rebreathing
because of the high flows inspiratory flow, respiratory
in the mask because of the
delivered to the patient. The rate and the volume of the
high fresh gas flow rates
exhaled gases are expelled nasopharynx.
causing the exhaled gases to
through the holes in the mask. b) True. See above.
be flushed from the mask
c) True. During mouth breathing,
through the side holes.
3. Face masks used during the inspiratory air flow
e) True. The Venturi mask is a
anaesthesia: produces a Venturi effect in the
fixed performance device that
a) True. Carbon particles prevent posterior pharynx entraining
delivers a constant
the build up of static oxygen from the nose.
concentration of oxygen in
electricity. The rubber face d) False. It is uncomfortable for
spite of the patients
masks and the rubber tubings the patient to have higher
respiratory pattern, by
used in anaesthesia are flows than 14L/min.
providing a higher gas flow
covered with carbon. With e) True. Patients tolerate the
than the peak inspiratory
modern anaesthetic practice nasal cannula for much longer
flow rate.
where no flammable drugs are periods than a face mask.
used, its significance has all Patients are capable of eating,
2. High-air-flow oxygen enrichment
but disappeared. drinking and speaking despite
face masks:
b) False. Face masks can have a the cannula.
a) True. Laminar flow through a
significant effect on the
constriction causes a decrease
110 6 Masks and oxygen delivery devices
Chapter 7
Laryngoscopes
and tracheal
intubation
equipment
112 7 Laryngoscopes and tracheal intubation equipment
Mechanism of action
1. Usually the straight blade is used
for intubating neonates and
infants. The blade is advanced
over the posterior border of the
relatively large, floppy V-shaped
epiglottis which is then lifted
directly in order to view the
larynx (Fig. 7.3B). There are
larger size straight blades that
can be used in adults.
2. The curved blade (Macintosh
blade) is designed to fit into the Fig. 7.2 A wide range of laryngoscope blades. (A) Miller blades (large, adult, infant,
premature); (B) Macintosh blades (large, adult, child, baby); (C) Macintosh polio blade;
(D) Soper blades (adult, child, baby); (E) left-handed Macintosh blade;
(F) Wisconsin blades (large, adult, child, baby, neonate); (G) Robertshaws blades (infant,
neonatal); (H) Seward blades (child, baby); (I) Oxford infant blade.
Epiglottis Epiglottis
Trachea Trachea
Fig. 7.13 Intubating bougie. Above is a single use bougie. Below is a reusable bougie
with its curved tip. (Courtesy of Smiths Medical.)
Fig. 7.14 The airway exchange catheter. Fig. 7.15 The Aintree intubation Fig. 7.16 The Cook retrograde
catheter. intubation set.
a tracheal tube to help direct it supraglottic airway device. It The proximal end has a 15-mm
through the vocal cords. allows any appropriate size of connector.
6. A gum elastic bougie is tracheal tube to be inserted
used when it is difficult to into the trachea which would
Mechanism of action
visualize the vocal cords. otherwise be limited by the
First, the bougie is inserted size of tube that could be 1. The introducer needle is inserted
through the vocal cords, passed through the supraglottic through the cricothyroid
then the tracheal tube is airway. membrane.
railroaded over it. Single use 2. The guidewire is advanced in a
intubating bougies are retrograde (cephalic) direction to
available (Fig. 7.13). exit orally or nasally.
7. The airway exchange catheter 3. The hollow guiding catheter is
(AEC) (Fig. 7.14) allows the Retrograde then introduced in an antegrade
exchange of tracheal tubes. It is direction into the trachea. The
a long hollow tube that can be
intubation set proximal end of the catheter can
inserted through a tracheal tube. (Fig. 7.16) be connected to an oxygen
This can then be withdrawn and supply.
another tracheal tube is inserted This set is used to assist in 4. A tracheal tube (5mm or
over it. Specially designed placement of a tracheal tube when a larger) can be introduced over
detachable 15-mm male taper fit difficult intubation is encountered. the guiding catheter into the
and Luer-Lok connectors can be trachea.
used to provide temporary
Components
oxygenation.
Problems in practice and
8. The Aintree intubation catheter 1. An introducer needle (18G and
safety features
(Fig. 7.15). This catheter is 5cm in length).
designed to be used with a 2. A guidewire with a J-shaped end. 1. Pneumothorax.
fibrescope being passed through 3. A 14-G 70-cm hollow guiding 2. Haemorrhage.
a laryngeal mask or other catheter with distal sideports. 3. Failure.
118 7 Laryngoscopes and tracheal intubation equipment
Anaesthesia 51, R1R5. Online. Thong, S.Y., Lim, Y., 2009. Video and
Retrograde intubation kit Available at: http://www.springerlink. optic laryngoscopy assisted tracheal
The introducer needle is inserted com/content/y8k03q8966rj8878/ intubation the new era.
through the cricothyroid. MHRA, 2011. Medical device alert: Anaesthesia and Intensive Care 37,
The guidewire is advanced in a reusable laryngoscope handles 219233. Online. Available at:
cephalic direction. all models and manufacturers http://xa.yimg.com/kq/groups/
Pneumothorax and haemorrhage (MDA/2011/0096). Online. Available 14982767/1890596062/name/Video
are potential complications. at: http://www.mhra.gov.uk/ +and+optic+laryngoscopy+assisted+
Publications/Safetywarnings/ tracheal+intubation+%C2%96+the+
MedicalDeviceAlerts/CON129213 new+era.pdf
FURTHER READING
Cooper, R.J., 2004. Laryngoscopy its
past and future. Canadian Journal of
MCQs 119
MCQs
In the following lists, which of the statements (a) to (e) are true?
Answers
Chapter 8
Ventilators
122 8 Ventilators
Ventilators are used to provide duration is reached, the a) volume: a preset volume is
controlled ventilation (intermittent ventilator changes to delivered
positive pressure ventilation; IPPV). exhalation. The cycling is not b) pressure: a preset pressure is
Some have the facilities to provide affected by the compliance of not exceeded.
other ventilatory modes. They can the patients lungs. Time 3. Source of power can be
be used in the operating theatre, cycling is the most commonly electric or pneumatic.
intensive care unit, during transport used method 4. Suitability for use in theatre and/
of critically ill patients and also at c) pressure cycling: when the or intensive care.
home (e.g. for patients requiring predetermined pressure is 5. Suitability for paediatric
nocturnal respiratory assistance). reached during inspiration, the practice.
ventilator changes over to 6. Method of operation (pattern of
exhalation. The duration gas flow during inspiration):
needed to achieve the critical a) pressure generator: the
pressure depends on the ventilator produces inspiration
Classification of compliance of the lungs. by generating a constant and
ventilators The stiffer the lungs are, the predetermined pressure.
quicker the pressure is Bellows or a moderate weight
There are many ways of classifying achieved and vice versa. The produce the pressure. The
ventilators (Table 8.1). ventilator delivers a different inspiratory flow changes with
1. The method of cycling is used to tidal volume if compliance or changes in lung compliance
change over from inspiration to resistance changes (Table 8.2)
exhalation and vice versa: d) flow cycling: when the b) flow generator: the ventilator
a) volume cycling: when the predetermined flow is reached produces inspiration by
predetermined tidal volume is during inspiration, the delivering a predetermined
reached during inspiration, the ventilator changes over to flow of gas. A piston, heavy
ventilator changes to exhalation. This method is weight or compressed gas
exhalation used in older design produce the flow. The flow
b) time cycling: when the ventilators. remains unchanged by changes
predetermined inspiratory 2. Inspiratory phase gas control: in lung compliance, although
pressures will change (see
Table 8.2). These ventilators
Table 8.1 Summary of the methods used in classifying ventilators have a high internal resistance
to protect the patient from
Method of cycling Volume cycling high working pressures.
Time cycling 7. Sophistication: new ventilators
Pressure cycling can function in many of the
Flow cycling above modes. They have other
Inspiratory phase gas control Volume modes, e.g. SIMV, PS and CPAP
Pressure (see pp 224225).
Source of power Electric 8. Function:
Pneumatic a) minute volume dividers: fresh
Suitability for use Operating theatre gas flow (FGF) powers the
Intensive care unit ventilator. The minute volume
Both
equals the FGF divided into
Paediatrics use Yes/no preset tidal volumes thus
Method of operation Pressure generator determining the frequency
Flow generator b) bag squeezers replace the hand
Sophistication SIMV, PS, CPAP ventilation of a Mapleson D
Function Minute volume divider or circle system. They need an
Bag squeezer external source of power
Lightweight portable c) lightweight portable: powered
by compressed gas and
CPAP, continuous positive airway pressure; PS, pressure support; SIMV, synchronized consists of the control unit
intermittent mandatory ventilation
and patient valve.
Manley MP3 ventilator 123
Components
Table 8.2 Differences between the pressure generator and flow
generator ventilators 1. Rubber tubing delivers the FGF
from the anaesthetic machine to
the ventilator.
Changes in lung
compliance Leak in the system 2. Two sets of bellows. A smaller
time-cycling bellows receives the
Pressure generator Can not compensate Can compensate (to a degree) FGF directly from the gas source
Flow generator Can compensate (to Can not compensate and then empties into the main
a degree) bellows.
3. Three unidirectional valves.
4. An adjustable pressure
limiting (APL) valve with
tubing and a reservoir bag
used during spontaneous
7. It should be easy to clean and
or manually controlled
sterilize.
Characteristics of the ventilation.
Some of the commonly used 5. The ventilator has a pressure
ideal ventilator ventilators are described below. gauge (up to 100cm H2O),
inspiratory time dial, tidal
1. The ventilator should be simple,
volume adjuster (up to
portable, robust and economical
1000mL), two knobs to change
to purchase and use. If
the mode of ventilation from
compressed gas is used to drive Manley MP3 and to controlled and
the ventilator, a significant
wastage of the compressed gas is
ventilator spontaneous (or manually
controlled) ventilation. The
expected. Some ventilators use a
This is a minute volume divider inflation pressure is adjusted
Venturi to drive the bellows, to
(time cycled, pressure generator). by sliding the weight to an
reduce the use of compressed
All the FGF (the minute volume) is appropriate position along its
oxygen.
delivered to the patient divided into rail. The expiratory block is
2. It should be versatile and supply
readily set tidal volumes (Fig. 8.1). easily removed for autoclaving.
tidal volumes up to 1500mL
with a respiratory rate of up to
60/min and variable I:E ratio.
It can be used with different
breathing systems. It can deliver
any gas or vapour mixture. The
addition of positive end
expiratory pressure (PEEP)
should be possible.
3. It should monitor the airway
pressure, inspired and exhaled
minute and tidal volume,
respiratory rate and inspired
oxygen concentration.
4. There should be facilities to
provide humidification. Drugs
can be nebulized through it.
5. Disconnection, high airway
pressure and power failure
alarms should be present.
6. There should be the facility to
provide other ventilatory modes,
e.g. SIMV, CPAP and pressure
support. Fig. 8.1 The Blease Manley MP3 with ventilator alarm (right).
124 8 Ventilators
2. A control unit with a variety of bellows causes a PEEP of Consists of driving and control
(640/min), I/E ratio, airway to an empty position and whereas the driving gas is within
pressure and power supply remains stationary in cases of the chamber.
(Figs 8.3 and 8.4). disconnection or leak.
126 8 Ventilators
themselves as their lungs at least two battery modules are tube (Fig. 8.8B), a cannula
compliance and inspiratory loaded for even the shortest positioned in the trachea via the
muscle strength improves. transport. cricothyroid membrane or a
This is shown by a gradual 4. The ventilator is heavier (20kg) modified tracheal tube with two
reduction in the peak airway than a dedicated transport additional small lumens opening
pressure measured by the ventilator. distally (Figs 8.8A and 8.9).
ventilator. Once the support is 2. Solenoid valves are used to
minimal, extubation can be deliver the jet gas.
considered Servo-i ventilator 3. Dials and display for driving
b) pressure support (PS): the Versatile intensive care ventilator pressure, frequency and
patients breath is supported suitable for both paediatric and inspiratory time.
with a set constant pressure adult use. 4. Built-in peristaltic pump for
above PEEP. This will give a Wide range of controls, displays nebulizing drugs or distilled
tidal volume that is dependent and alarms. water for humidifying the jet gas.
on the lung compliance and Portable with battery power. 5. High-flow air/oxygen or nitrous
patients inspiratory muscle oxide/oxygen blender determines
strength. The pressure support the mix of the jet gas.
setting needs reviewing
regularly to allow the patient High-frequency Mechanism of action
to wean from respiratory
support jet ventilator 1. Frequencies of 20500cycles/
c) continuous positive airway min can be selected, with minute
pressure (CPAP): a continuous This ventilator reduces the extent volumes ranging from 5 to
positive pressure is maintained of the side-effects of conventional 60L/min.
in the airways similar to that IPPV. There are lower peak airway 2. It is a time-cycled ventilator
developed with a conventional pressures with better maintenance of delivering gas in small jet
CPAP flow generator (see the cardiac output and less anti- pulsations. The inspiratory time
Chapter 13). This differs diuretic hormone production and is adjustable from 20% to 50%
from the conventional CPAP fluid retention. It is better tolerated of the cycle.
flow generator by allowing by alert patients than conventional 3. The fresh gas leaving the narrow
measurement of tidal volume, IPPV (Fig. 8.7). injector at a very high velocity
minute volume and respiratory causes entrainment of gas. The
rate, and trends can be amount of entrained gas is
Components
observed also. uncertain making measurement
1. A Venturi injector is used: a of tidal volume and FiO2
cannula positioned in a tracheal difficult.
A
Entrained gas
14 gauge cannula
B
Entrained gas
VentiPAC
Portable ventilator powered by
pressurized gas.
Controls include flow rate,
inspiratory time and expiratory
time.
An FiO2 of 0.45 or 1.0 can be
delivered.
It has a demand valve.
MRI compatible.
Pneupac VR1
Fig. 8.10 The VentiPAC ventilator. Emergency Ventilator
Mechanism of action Problems in practice and
(Fig. 8.11)
safety features
1. The source of power is dry, This is a lightweight hand-held,
oil-free pressurized gas (270 1. There is an adjustable time-cycled, gas-powered flow
600kPa) at 60L/min. Using inspiratory pressure relief generator ventilator. It is designed
air mix mode reduces gas mechanism with a range of for use in emergency and during
consumption by the ventilator 2080cm H2O to reduce the transport. It is MRI compatible up
by almost 70%. risk of overpressure and to 3Tesla.
2. The frequency is set by adjusting barotruma.
the inspiratory and expiratory 2. There are audible and visual
Components
times. low-pressure (disconnection)
3. The tidal volume is set by the and high-pressure (obstruction) 1. Tidal volume/frequency control.
adjustment of the flow and alarms. 2. Auto/manual control with a
inspiratory time. 3. A supply gas failure alarm. manual trigger and push button.
4. A choice of an FiO2 of 1.0 (no 4. The ventilator is magnetic 3. Air mix switch allowing the
air mix) or 0.45 (air mix). resonance imaging (MRI) delivery of oxygen at 100% or
5. The demand mode provides compatible. 50% concentrations.
100% oxygen to a
spontaneously breathing
patient. A visual indicator
flashes when a spontaneous
breath is detected.
6. CMV/demand mode provides
continuous mandatory
ventilation. If the patient makes
a spontaneous breath, this
causes the ventilator to operate
in a synchronized minimum
mandatory ventilation (SMMV)
mode. Any superimposed
mandatory ventilatory attempts
are synchronized with the
breathing pattern.
7. A PEEP valve can be added
generating a PEEP of up to
20cm H2O. Fig. 8.11 Pneupac VR1 Emergency Ventilator.
130 8 Ventilators
Self-inflating bag
Compact, portable, self-inflating
bag with a one-way valve.
Oxygen reservoir can be added
to increase FiO2.
Paediatric version exists.
PEEP valve
This valve is used during IPPV to
increase the functional residual
capacity (FRC) to improve the
patients oxygenation.
Fig. 8.13 A range of self-inflating resuscitation bags with oxygen reservoirs. It is a spring-loaded
unidirectional valve positioned on
Mechanism of action the expiratory side of the ventilator
2. The valve acts as a spillover
1. The non-rebreathing valve valve allowing excess inspiratory breathing system with a standard
(Ambu valve) incorporates a gas to be channelled directly to 22-mm connector. By adjusting the
silicone rubber membrane the expiratory outlet, bypassing valve knob, a PEEP of between zero
(Fig. 8.14). It has a small dead the patient port. and 20cm H2O can be achieved
space and low resistance to flow. 3. The valve is suitable for (Fig. 8.15).
At a flow of 25L/min, an both IPPV and spontaneous
inspiratory resistance of ventilation.
0.4cm H2O and an expiratory 4. The shape of the self-inflating
resistance of 0.6cm H2O are bag is automatically restored
achieved. The valve can easily be after compression. This allows
dismantled for cleaning and fresh gas to be drawn from the
sterilization. reservoir.
Valve housing
Inspiratory
connector
Expiratory Valve
connector membrane
Patient connector
Fig. 8.14 An Ambu valve disassembled. (Reproduced with permission from
AMBU International (UK) Ltd.) Fig. 8.15 The Ambu PEEP valve.
132 8 Ventilators
The valve provides almost manual, fifth ed. Butterworth- MHRA, 2010. Medical device alert: all
constant expiratory resistance over a Heinemann, Edinburgh. Oxylog 3000 emergency/transport
very wide range of flow rates. Merck, 2007. Overview of mechanical ventilators manufactured by Draeger
ventilation. Online. Available at: (MDA/2010/092). Online. Available
http://www.merckmanuals.com/ at: http://www.mhra.gov.uk/
FURTHER READING
professional/critical_care_medicine/ Publications/Safetywarnings/
Bersten, A.D., 2003. Mechanical respiratory_failure_and_mechanical_ MedicalDeviceAlerts/CON100176.
ventilation. In: Bersten, A., Soni, N., ventilation/overview_of_mechanical_
Oh, T.E. (Eds), Ohs intensive care ventilation.html.
MCQs 133
MCQs
In the following lists, which of the following statements (a) to (e) are true?
Answers
1. Bag in bottle ventilator: attached to the inspiratory c) False. The ventilator will cycle
a) False. The driving gas is limb and not the expiratory with time regardless of the
separate from the FGF. The limb. A Wright spirometer can compliance.
driving gas is usually either be attached to the expiratory d) False. In a lung with low
oxygen or, more economically, limb to measure the tidal compliance, the inspiration
air. There is no mixing volume. will be shorter because the
between the driving gas and e) True. During the spontaneous pressure will be reached more
the FGF. The volume of the (manual) breathing mode, the quickly leading the ventilator
driving gas reaching the Manley ventilator acts as a to cycle and vice versa.
chamber is equal to the tidal Mapleson D system. e) True. The ventilator will
volume. continue to deliver gases,
b) False. The tidal volume and 3. Bag in bottle ventilator: despite the leak, until a preset
respiratory rate can be a) True. The inspiratory and pressure has been reached.
adjusted separately in a bag in expiratory periods can be
bottle ventilator. determined by adjusting the 5. High-frequency jet ventilation:
c) True. Most of the bag in I:E ratio and the respiratory a) False. The ventilator uses the
bottle ventilators use rate. So, for example, with a Venturi principle to entrain
ascending bellows. This adds rate of 10breaths/min and an ambient air. The amount of
to the safety of the system as I:E ratio of 1:2, each breath entrainment is uncertain,
the bellows will collapse if lasts for 6 seconds with an making the measurement of
there is a leak. inspiration of 2 seconds and the FiO2 difficult.
d) False. See c). expiration of 4 seconds. b) False. Frequencies of up to
e) False. The ventilator can be b) False. There is no mixing 500/min (not Hz; i.e. per
used for both adults and between the fresh gas and the second) can be achieved with
children. A different size driving gas as they are a high-frequency jet ventilator.
bellows can be used for completely separate. c) True. Because of the lower
different age groups. c) False. In case of a leak in the intrathoracic pressures
system, the descending generated during high-
2. Manley ventilator: bellows will not collapse. The frequency jet ventilation,
a) True. The tidal volume can be opposite occurs with the causing a lesser effect on the
set in a Manley ventilator. The ascending bellows. venous return, the cardiac
whole FGF (minute volume) is d) True. A PEEP of 24cm H2O output is better maintained.
delivered to the patient is expected due to the d) True. It can be used both in
according to the set tidal compliance of the bellows. anaesthesia and intensive care,
volume, thus dividing the e) True. e.g. in the management of
minute volume. bronchopleural fistula.
b) False. There are two sets of 4. Regarding classification of e) False. Although the risk is
bellows in a Manley ventilators: reduced, there is still a risk of
ventilator, the time-cycling a) False. A pressure generator barotrauma.
bellows and the main bellows. ventilator cannot compensate
c) False. As it is a minute for changes in lung 6. a)
volume divider and the FGF is compliance. It cycles when the
the driving gas (see a) above), set pressure has been reached.
activating the emergency This can be a larger or smaller
oxygen flush will lead to tidal volume depending on the
considerable increase in the lung compliance.
minute volume. b) True. It will deliver the set
d) False. The pressure- flow whether there is a leak
monitoring alarm should be or not.
135
Chapter 9
Humidification
and filtration
136 9 Humidification and filtration
reservoir and in the breathing 4. The tubing has poor thermal Problems in practice and
system close to the patient. insulation properties causing a safety features
3. Tubing is used to deliver the decrease in the temperature of
humidified and warm gases to 1. The humidifier, which is
inspired gases. This is partly
the patient. It should be as short electrically powered, should be
compensated for by the release
as possible. A water trap is safe to use with no risk of
of the heat of condensation.
positioned between the patient scalding, overhydration and
5. By raising the temperature in
and the humidifier along the electric shock. A second backup
the humidifier above body
tubing. The trap is positioned thermostat cuts in should there
temperature, it is possible to
lower than the level of the be malfunction of the first
deliver gases at 37C and fully
patient. thermostat.
saturated. The temperature of
2. The humidifier and water trap(s)
gases at the patients end is
should be positioned below the
measured by a thermistor.
Mechanism of action level of the tracheal tube to
Via a feedback mechanism,
1. Powered by electricity, the water prevent flooding of the airway
the thermistor controls the
is heated to between 45C and by condensed water.
temperature of water in the
60C (Fig. 9.5). 3. Colonization of the water by
container.
2. Dry cold gas enters the bacteria can be prevented by
6. The temperature of gases at the
container where some passes increasing the temperature to
patients end depends on the
close to the water surface, 60C. This poses greater risk of
surface area available for
gaining maximum saturation. scalding.
vaporization, the flow rate and
Some gas passes far from the 4. The humidifier is large, expensive
the amount of cooling and
water surface, gaining minimal and can be awkward to use.
condensation taking place in the
saturation and heat. 5. There are more connections in a
inspiratory tubing.
3. The container has a large surface ventilator set up and so the risk
7. Some designs have heated
area for vaporization. This is to of disconnections or leaks
elements placed in the
ensure that the gas is fully increases.
inspiratory and expiratory limb
saturated at the temperature of of the breathing system to
the water bath. The amount of maintain the temperature and
gas effectively bypassing the prevent rain out (condensation)
Hot water bath humidifier
water surface should be minimal. within the tube.
Consists of a container with a
thermostatically controlled
Dry gas in heating element and tubing
with water traps.
The temperature of water in the
container, via a feedback
mechanism, is controlled by a
thermistor at the patients end.
Full saturation at 37C can be
achieved.
Colonization by bacteria is a
Humidified Overflow problem.
gas out
Anvil
Driving gas
Water column
Mechanism of action
Nebulizers 2. Minimal dead space,
Produce microdroplets of water particularly for paediatric There are five main mechanisms by
of different sizes, 120m. practice. which filtration can be achieved on
The quantity of water droplets is 3. Minimum resistance, especially a fibre:
not limited by the temperature when wet. 1. Direct interception: large
of the carrier gas. 4. Not affected by anaesthetic particles (=1m), such as dust
They can be gas driven, spinning agents and does not affect the and large bacteria, are physically
disc or ultrasonic. anaesthetic agents. prevented from passing through
5. Effective when either wet or the pores of the filter because of
dry. It should completely their large size.
prevent the passage of 2. Inertial impaction: smaller
contaminated body liquids particles (0.51m) collide with
Bacterial and (blood, saliva and other the filter medium because of
viral filters liquids) which may be present their inertia. They tend to
or generated in the breathing continue in straight lines, carried
These minimize the risk of cross- system. along by their own momentum
transmission of bacteria and/or 6. User friendly, lightweight, not rather than following the path
viruses between patients using the bulky and non-traumatic to of least resistance taken by the
same anaesthetic breathing systems. the patient. gas. The particles are held by
The British Standard defines them 7. Disposable. Van der Waals electrostatic
as devices intended to reduce 8. Provides some humidification if forces.
transmission of particulates, no other methods being used. 3. Diffusional interception: very
including micro-organisms, in Adequate humidification can small particles (<0.5m), such
breathing systems. It is thought that usually be achieved by the as viruses, are captured because
the incidence of bleeding after addition of a hygroscopic they undergo considerable
orotracheal intubation is 86%. The element to the device. Brownian motion (i.e. random
filter should be positioned as close 9. Transparent. movement) because of their very
to the patient as possible, e.g. on 10. Cost effective. small mass. This movement
the disposable catheter mount, to increases their apparent diameter
protect the rest of the breathing so that they are more likely
system, ventilator and anaesthetic to be captured by the filter
Size of micro-organisms
machine. It is recommended that a element.
Hepatitis virus 0.02m
new filter should be used for each 4. Electrostatic attraction: this can
Adenovirus 0.07m
patient. A humidification element be very important but it is
HIV 0.08m
can be added producing a heat and difficult to measure as it requires
Mycobacterium 0.3m
moisture exchanging filter (HMEF) knowing the charge on the
tuberculosis
(see Fig. 9.1). particles and on the fibres.
Staphylococcus aureus 1.0m
Increasing the charge on either
Cytomegalovirus 0.1m
the particles or the fibres
Characteristics of the ideal filter
increases the filtration efficiency.
Charged particles are attracted
1. Efficient: the filter should be
Components
to oppositely charged fibres by
effective against both air- and 1. Two ports designed to accept coulombic attraction.
liquid-borne micro-organisms. 15- and 22-mm size tubings and 5. Gravitational settling: this
A filtration action of 99.99 connections. affects large particles (>5m).
99.999% should be achieved. 2. A sampling port to measure the The rate of settling depends on
This allows between 100 and gases/agents concentrations the balance between the effect of
10 micro-organisms to pass positioned on the anaesthetic gravity on the particle and the
through the filter, respectively, breathing system side. buoyancy of the particle. In
after a 106 micro-organism 3. The filtration element can either filters used in anaesthesia, it has
challenge. The filter should be be a felt-like electrostatic minimal effect as most of the
effective bidirectionally. material or a pleated settling occurs before the
hydrophobic material. particles reach the filter.
Bacterial and viral filters 141
ELECTROSTATIC FILTERS
(FIG. 9.8)
PLEATED HYDROPHOBIC
FILTERS (FIG. 9.9)
high gas flows while retaining with no absorption. Such a filter
1. The very small pore size filter low resistance, a large surface can successfully prevent the
membrane provides adequate area is required. Pleated paper passage of water under pressures
filtration over longer periods of filters made of inorganic fibres as high as 60cm H2O.
time. These filters rely on the are used to achieve this. 4. Although hydrophobic filters
naturally occurring electrostatic 3. The forces between individual provide some humidification,
interactions to remove the liquid water molecules are a hygroscopic element can be
particles. A filtration efficiency stronger than those between added to improve
of 99.999% can be achieved. the water molecules and the humidification.
2. To achieve minimal pressure hydrophobic membrane. This 5. Currently there is no evidence
drop across the device with such leads to the collection of water showing any type of filter is
a small pore size, so allowing on the surface of the membrane clinically superior to another.
142 9 Humidification and filtration
MCQs
In the following lists, which of the statements (a) to (e) are true?
Answers
Chapter 10
Non-invasive
monitoring
146 10 Non-invasive monitoring
Amplitude mV
0.5 0.5 heart is a common error, leading
to false information.
2. Electrical interference can be a
0 0 50-Hz (in UK) mains line
interference because of
capacitance or inductive coupling
0.5 0.5
0 0.2 0.4 0.6 0.8 0 0.2 0.4 0.6 0.8 effect. Any electrical device
t.s t.s powered by AC can act as one
plate of a capacitor and the
A B
patient acts as the other plate.
Fig.10.6 ECG filters. (A) Unfiltered signal with noise. (B) Filtered clean signal. Interference can also be because
of high-frequency current
interference from diathermy.
b) the diagnostic mode has a Right arm lead Most modern monitors have the
wider frequency response of over manubrium facilities to avoid interference.
0.05150Hz. The high- sterni Shielding of cables and leads,
frequency limit allows the Indifferent lead differential amplifiers and
assessment of the ST segment, Left arm lead electronic filters all help to
QRS morphology and V5 position over produce an interference-free
tachyarrhythmias. The left ventricle monitoring system. Differential
low-frequency limit allows amplifiers measure the difference
representation of P- and between the potential from two
T-wave morphology and different sources. If there is
ST-segment analysis. interference common to the two
4. There are many ECG electrode input terminals (e.g. mains
configurations. Usually during frequency), it can be eliminated as
anaesthesia, three skin electrodes only the differences between the
are used (right arm, left arm and two terminals is amplified. This is
indifferent leads). The three limb called common mode rejection
leads used include two that are ratio (CMRR). Amplifiers used in
active and one that is inactive ECG monitoring should have a
(earth). Sometimes five electrodes Fig. 10.7 The CM5 ECG lead high CMRR of 100000:1
are used. Lead II is ideal for configuration. to 1000000:1, which is a
detecting arryhthmias. CM5 measurement of capability to
configuration is able to detect reject the noise. They should also
89% of ST-segment changes due left anterior axillary line) and have a high input impedance
to left ventricular ischaemia. In the indifferent lead is on the left (about 10M) to minimize the
CM5, the right arm electrode is shoulder or any convenient current taken from the electrodes.
positioned on the manubrium position (Fig. 10.7). Table 10.1 shows the various
(chest lead from manubrium), 5. The CB5 configuration is useful types and sources of interference
the left arm electrode is on V5 during thoracic anaesthesia. The and how to reduce the
position (fifth interspace in the right arm electrode is positioned interference.
Arterial blood pressure 149
Arterial
Absorbance
Venous
Skin
Tissue
Bone
Time
Fig. 10.12 Pulse oximeter probes. Finger probe (top) and ear probe (bottom).
Fig. 10.14 Schematic representation
of the contribution of various body
components to the absorbance of light.
Photodetector Display
Microprocessor
4. The microprocessor is
programmed to mathematically
analyse both the DC and AC
components at 660 and 940nm
calculating the ratio of
absorption at these two
660 940 frequencies (R/IR ratio). The
nm nm result is related to the arterial
LED sequence saturation. The absorption
of oxyhaemoglobin and
On Off
Sequence repeated many times per second
deoxyhaemoglobin at these two
Off On
wavelengths is very different.
Off Off
This allows these two
Fig. 10.13 Working principles of the pulse oximeter. The LEDs operate in sequence and wavelengths to provide good
when both are off the photodetector measures the background level of ambient light. sensitivity. 805nm is one of
the isobestic points of
oxyhaemoglobin and
constant (DC). The non-constant deoxyhaemoglobin. The OFF
Mechanism of action
absorption (AC) is the result part allows a baseline
1. The oxygen saturation is of arterial blood pulsations measurement for any changes in
estimated by measuring the (Fig. 10.14). The sensitive ambient light.
transmission of light, through a photodetector generates a 5. A more recent design uses
pulsatile vascular tissue bed (e.g. voltage proportional to the multiple wavelengths to
finger). This is based on Beers transmitted light. The AC eradicate false readings from
law (the relation between component of the wave is carboxy haemoglobin and
the light absorbed and the about 15% of the total methaemoglobinaemia.
concentration of solute in the signal. Advanced oximeters use more
solution) and Lamberts law 3. The high frequency of the LEDs than seven light wavelengths.
(relation between absorption of allows the absorption to be This has enabled the
light and the thickness of the sampled many times during each measurement of haemoglobin
absorbing layer). pulse beat. This is used to enable value, oxygen content,
2. The amount of light transmitted running averages of saturation carboxyhaemoglobin and
depends on many factors. The to be calculated many times per methaemoglobin concentrations.
light absorbed by non-pulsatile second. This decreases the 6. A variable pitch beep provides
tissues (e.g. skin, soft tissues, noise (e.g. movement) effect on an audible signal of changes in
bone and venous blood) is the signal. saturation.
End-tidal carbon dioxide analysers (capnographs) 153
E I E
CO2 (kPa)
D D
C
AB
Time (sec)
Fig. 10.15 Diagram of an end-tidal
carbon dioxide waveform. I = inspiration;
E = expiration; AB represents the
emptying of the upper dead space of the
airways. As this has not undergone gas
exchange, the CO2 concentration is zero.
BC represents the gas mixture from the
upper airways and the CO2-rich alveolar
gas. The CO2 concentration rises
continuously. CD represents the alveolar
gas and is described as the alveolar
plateau. The curve rises very slowly. D is
the end-tidal CO2 partial pressure where
the highest possible concentration of Fig. 10.16 A main-stream end-tidal carbon dioxide analyser.
exhaled CO2 is achieved at the end of
expiration. It represents the final portion concentrations during each which in turn produces heat.
of gas which was involved in the gas respiratory cycle. The heat is measured by a
exchange in the alveoli. Under certain temperature sensor and is
conditions (see text) it represents a
Components proportional to the partial
reliable index of the arterial CO2 partial pressure of carbon dioxide gas
pressure. DA represents inspiration 1. The sampling chamber can present in the mixture in the
where the fresh gas contains no CO2. either be positioned within the sample chamber. This produces
patients gas stream (main- an electrical output. This means
The end-tidal CO2 is less than stream version, Fig. 10.16) or that the amount of gas present is
alveolar CO2 because the end-tidal connected to the distal end of inversely proportional to the
CO2 is always diluted with alveolar the breathing system via a amount of infrared light present
dead space gas from unperfused sampling tube (side-stream at the detector in the sample
alveoli. These alveoli do not take version, Fig. 10.17). chamber (Fig 10.19).
part in gas exchange and so contain 2. A photodetector measures light 4. In the same way, a beam of light
no CO2. Alveolar CO2 is less than reaching it from a light source at passes through the reference
arterial CO2 as the blood from the correct infrared wavelength chamber which contains room
unventilated alveoli and lung (using optical filters) after air. The absorption detected from
parenchyma (both have higher CO2 passing through two chambers. the sample chamber is compared
contents) mixes with the blood from One acts as a reference whereas to that in the reference chamber.
ventilated alveoli. In healthy adults the other one is the sampling This allows the calculation of
with normal lungs, end-tidal CO2 is chamber (Fig. 10.18). carbon dioxide values.
0.30.6kPa less than arterial CO2. 5. The inspired and exhaled
This difference is reduced if the carbon dioxide forms a square
Mechanism of action
lungs are ventilated with large tidal wave, with a zero baseline
volumes. The Greek root kapnos, 1. Carbon dioxide absorbs the unless there is rebreathing
meaning smoke, give us the term infrared radiation particularly at (Fig. 10.20A).
capnography (CO2 can be thought as a wavelength of 4.3m. 6. A microprocessor-controlled
the smoke of cellular metabolism). 2. The amount of infrared infrared lamp is used. This
radiation absorbed is produces a stable infrared
End-tidal CO2 < alveolar CO2 source with a constant output.
proportional to the number of
< PaCO2
carbon dioxide molecules The current is measured
In reality, the devices used cannot (partial pressure of carbon with a current-sensing resistor,
determine the different phases of dioxide) present in the chamber. the voltage across which
respiration but simply report the 3. The remaining infrared radiation is proportional to the
minimum and maximum CO2 falls on the thermopile detector, current flowing through it. The
End-tidal carbon dioxide analysers (capnographs) 155
Fig. 10.17 The Penlon PM9000 Express which measures end-tidal CO2, oximetry and SIDE-STREAM ANALYSERS
inhalational agent concentration using a side-stream method. (Courtesy of Penlon Ltd,
Abingdon, UK (www.penlon.com).) 1. This consists of a 1.2-mm
internal diameter tube that
samples the gases (both inspired
Sample gas and exhaled) at a constant
rate (e.g. 150200mL/min).
The tube is connected to a
lightweight adapter near
Detector
the patients end of the
breathing system (with a
Light source Sample chamber Multigas filter pneumotachograph for
Fig. 10.18 Components of a gas analyser using an infrared light source suitable for spirometry) with a small increase
end-tidal carbon dioxide measurement. The reference chamber has been omitted for the in the dead space. It delivers the
sake of clarity. gases to the sample chamber.
It is made of Teflon so it is
impermeable to carbon dioxide
supply to the light source respiratory cycle, monitors are and does not react with
is controlled by the feedback designed to measure the anaesthetic agents.
from the sensing resistor respiratory rate. 2. As the gases are humid, there is
maintaining a constant 8. Alarm limits can be set for both a moisture trap with an exhaust
current of 150mA. high and low values. port, allowing gas to be vented
7. Using the rise and fall of the 9. To avoid drift, the monitor should to the atmosphere or returned to
carbon dioxide during the be calibrated regularly with the breathing system.
156 10 Non-invasive monitoring
Sample chamber
Filter
Fig. 10.19 Principles of infrared detector: due to the large amount of infrared
absorption in the sample chamber by the carbon dioxide, little infrared finally reaches
the detector.
sample.
intubation (no or very little
6. Portable hand-held side-stream
carbon dioxide is detected).
analysers are available (Fig.
Following manual ventilation or
10.21). They can be used during
the ingestion of carbonated
patient transport and out-of-
Time (sec) drinks, some carbon dioxide
hospital situations.
Fig. 10.20 (A) An end-tidal carbon
might be present in the
dioxide waveform which does not return stomach. Characteristically, this
to the baseline during inspiration may result in up to 56
MAIN-STREAM ANALYSER
indicating that rebreathing is occurring. waveforms with an abnormal
(B) An end-tidal carbon dioxide waveform shape and decreasing in
which illustrates the sloping plateau seen 1. The sample chamber is positioned
amplitude.
in patients with chronic obstructive within the patients gas stream,
2. As a disconnection alarm for
airways disease. The normal waveform is increasing the dead space. In
superimposed (dotted line). a ventilator or breathing
order to prevent water vapour
system. There is sudden
condensation on its windows,
absence of the end-tidal carbon
it is heated to about 41C.
3. In order to accurately measure dioxide.
2. Since there is no need for a
end-tidal carbon dioxide, the 3. To diagnose lung embolism as a
sampling tube, there is no
sampling tube should be sudden decrease in end-tidal
transport time delay in gas
positioned as close as possible to carbon dioxide assuming that
delivery to the sample chamber.
the patients trachea. the arterial blood pressure
3. Other gases and vapours are not
4. A variable time delay before remains stable.
measured simultaneously.
the sample is presented to the 4. To diagnose malignant
sample chamber is expected. The See Table 10.4 for a comparison hyperpyrexia as a gradual
transit time delay depends on of side-stream and main-stream increase in end-tidal carbon
the length (which should be as analysers. dioxide.
End-tidal carbon dioxide analysers (capnographs) 157
Sensing O2
Amplifier Out Thin electrolyte layer Electromagnet
membrane
Cathode Mixture
out
Ag-electrode Electrolyte
Pt-electrode Anode
Electrolyte Reference in
Switched
magnetic
Teflon membrane Circular Sample in field
Out contact plate
Fig. 10.23 Different types of oxygen analysers.
Oxygen concentration analysers 159
Table 10.6 summarizes the on and off and reset the pointer
Inhalational agent methods used in gas and vapour to the zero position.
concentration analysers analysis.
A sample of gas is used to Mechanism of action
measure the concentration of
inhalational agent using infrared 1. The Wright respirometer is a
light absorption. one-way system. It allows the
By selecting light of the correct
Wright respirometer measurement of the tidal volume
wavelengths, the inspired and if the flow of the gases is in one
expired concentrations of the This compact and light (weighs less direction only. The correct
agent(s) can be measured. than 150g) respirometer is used to direction for gas flow is
An infrared light of a
measure the tidal volume and indicated by an arrow.
wavelength of 4.6m is used minute volume (Fig. 10.29). 2. The slits surrounding the vane
for N2O. For other inhalational are to create a circular flow in
order to rotate the vane. The
agents, higher wavelengths are Components
used, between 8 and 9m. vane does 150 revolutions for
Ultraviolet absorption, mass
1. The respirometer consists of an each litre of gas passing through.
spectrometry and quartz crystal inlet and outlet. This causes the pointer to rotate
oscillation are other methods of 2. A rotating vane surrounded by round the respirometer display.
measuring the inhalational slits (Fig. 10.30). The vane is 3. The outer display is calibrated at
agents concentration. attached to a pointer. 100mL per division. The small
3. Buttons on the side of the inner display is calibrated at 1L
respirometer to turn the device per division.
MASS SPECTROMETER
Laminar
Gas flow
resistor
Heating coil
Fig. 10.31 A pneumotachograph.
Mica vane
See text for details.
Gas flow
Fig. 10.30 Mechanism of action of the Wright respirometer. Mechanism of action
1. The principle of its function is
4. It is usually positioned on the Problems in practice and sensing the change in pressure
expiratory side of the breathing safety features across a fixed resistance through
system, which is at a lower
1. The Wright respirometer tends which gas flow is laminar.
pressure than the inspiratory
to over-read at high flow rates 2. The pressure change is only a
side. This minimizes the loss of
and under-read at low flows. few millimetres of water and is
gas volume due to leaks and
2. Water condensation from the linearly proportional, over a
expansion of the tubing.
expired gases causes the pointer certain range, to the flow rate of
5. For clinical use, the respirometer
to stick, thus preventing it from gas passing through the
reads accurately the tidal volume
rotating freely. resistance.
and minute volume (510%)
3. The tidal volumes can be
within the range of 424L/min.
summated over a period of a
A minimum flow of 2L/min is
Wright respirometer minute to give the minute
required for the respirometer to
Rotating vane attached to a volume.
function accurately.
pointer. 4. It can measure flows in both
6. To improve accuracy, the
Fitted on the expiratory limb to inspiration and expiration (i.e.
respirometer should be
measure the tidal and minute bidirectional).
positioned as close to the
volume with an accuracy of
patients trachea as possible.
510%. Problems in practice and
7. The resistance to breathing is
The flow is unidirectional.
very low at about 2cm H2O at safety features
It over-reads at high flows and
100L/min.
under-reads at low flows. Water vapour condensation at the
8. A paediatric version exists with
resistance will encourage the
a capability of accurate tidal
formation of turbulent flow
volume measurements between
affecting the accuracy of the
15 and 200mL.
measurement. This can be avoided
9. A more accurate version of the Pneumotachograph by heating the parallel tubes.
Wright respirometer uses light
reflection to measure the tidal This measures gas flow. From this,
volume. The mechanical causes gas volume can be calculated.
of inaccuracies (friction and Combined
inertia) and the accumulation of
water vapour are avoided. Other
Components pneumotachograph
designs use a semiconductive 1. A tube with a fixed resistance. and Pitot tube
device that is sensitive to The resistance can be a
changes in magnetic field. Tidal bundle of parallel tubes (Fig. This combination (Fig. 10.32) is
volume and minute volume can 10.31). designed to improve accuracy
be measured by converting these 2. Two sensitive pressure and calculate and measure the
changes electronically. An alarm transducers on either side of the compliance, airway pressures, gas
system can also be added. resistance. flow, volume/pressure (Fig. 10.33)
164 10 Non-invasive monitoring
Sample gas
flow to be known with sensors
calibrated accordingly.
3. Gas temperature: A knowledge
gas temperatures is required.
Usually, the sensors software
provides default values for a
Gas flow typical patient.
4. Humidity: moisture can affect
measurement and generation of
pressure drop. Have the pressure
ports directed upwards to
To pressure prevent fluid from draining into
transducers them.
Fig. 10.32 Combined pneumotachograph and Pitot tube. (Courtesy of GE Datex 5. Apparatus dead space:
Ohmeda.) Sensors need to have a
minimum dead space; <10ml
for the adult flow sensors and
600 Vol Gas flow <1ml for the neonatal sensors.
mL
6. Operating range of flow
A B sensor: Sensors are designed to
function accurately with a very
wide range of tidal volumes, I:E
ratios, frequencies and flow
Paw
cmH20 ranges.
To pressure transducer 7. Inter-sensor variability:
0 20
Fig. 10.34 Cross-section of a Pitot tube Individual sensors can have
Fig. 10.33 Volume pressure loops in a flowmeter. The two ports are facing in different performances. There
patient (A) before and (B) during CO2 opposite directions within the gas flow. should be no need for individual
insufflation in a laparoscopic operation.
Note the decrease in compliance and device calibration of the flow/
increase in airway pressure (Paw). pressure characteristics
Mechanism of action
The effects of the density and
The pressure difference between the
viscosity of the gas(es) can alter the
and flow/volume loops. Modern ports is proportional to the square
accuracy. This can be compensated
devices can be used accurately even of the flow rate.
for by continuous gas composition
in neonates and infants. analysis via a sampling tube.
Problems in practice and
THE PITOT TUBE safety features Pneumotachograph
The effects of the density and A bidirectional device to measure
Components the flow rate, tidal and minute
viscosity of the gas(es) can alter the
1. Two pressure ports one facing accuracy. This can be compensated volume.
the direction of gas flow, the for by continuous gas composition A laminar flow across a fixed
other perpendicular to the gas analysis via a sampling tube. resistance causes changes in
flow. This is used to measure gas pressure which are measured by
flow in one direction only. transducers.
Factors affecting the readings in
2. In order to measure bidirectional Condensation at the resistance
pnuemotachograph
flows (inspiration and can cause turbulent flow and
expiration), two pressure ports 1. Location: should be placed inaccuracies.
face in opposite directions between the breathing system Improved accuracy is achieved
within the gas flow (Fig. 10.34). Y-piece and the tracheal tube. by adding a Pitot tube(s) and
3. These pressure ports are 2. Gas composition: nominal continuous gas composition
connected to pressure values of gas composition need analysis.
transducers.
Peripheral nerve stimulators 165
Peripheral nerve
stimulators
These devices are used to
monitor transmission across the
neuromuscular junction. The depth,
adequate reversal and type of
neuromuscular blockade can be
established (Fig. 10.36).
Components
1. Two surface electrodes (small
ECG electrodes) are positioned
over the nerve and connected via
Fig. 10.35 The Penlon pressure monitoring ventilator alarm. the leads to the nerve stimulator.
166 10 Non-invasive monitoring
A Normal
20 ms
750 ms
Fig. 10.38 The pattern of double-burst
B Total paralysis stimulation. Three impulses of 50Hz
tetanus, at 20-ms intervals, every 750ms
is shown.
MCQs
In the following lists, which of the statements (a) to (e) are true?
Answers
Chapter 11
Invasive
monitoring
178 11 Invasive monitoring
Invasive arterial
pressure monitoring
Invasive arterial pressure monitoring
provides beat-to-beat real-time
information with sustained accuracy.
Components
1. An indwelling Teflon arterial
cannula (20 or 22G) is used
(Fig. 11.1). The cannula has
parallel walls to minimize the
effect on blood flow to the distal
parts of the limb. Cannulation
can be achieved by directly
threading the cannula (either by
direct insertion method or a
transfixation technique) or by
using a modified Seldinger
technique with a guidewire to
assist in the insertion as in some
designs (Fig. 11.2).
2. A column of bubble-free
heparinized or plain 0.9%
normal saline at a pressure of
300mmHg, incorporating a
flushing device.
3. Via the fluid column, the
cannula is connected to a
transducer (Figs 11.311.5). Fig. 11.2 Argon Careflow arterial cannula with its guidewire. (Courtesy of Argon Medical.)
This in turn is connected to an
amplifier and oscilloscope. A Dome and diaphragm Flush
strain gauge variable resistor Connected to
transducer is used. arterial cannula Heparinized
4. The diaphragm (a very thin saline pressurized to
membrane) acts as an interface 300 mmHg
Three-way tap
Transducer
Electrical signal
Fig. 11.3 Components of a pressure measuring system.
Mechanism of action
between the transducer and the
fluid column. 1. The saline column moves back
5. The pressure transducer is a and forth with the arterial
device that changes either pulsation causing the diaphragm
electrical resistance or to move. This causes changes in
capacitance in response to the resistance and current flow
changes in pressure on a through the wires of the
Fig. 11.1 BD Flowswitch arterial cannula. solid-state device. The moving transducer.
Note the onoff switch valve. (Courtesy part of the transducer is very 2. The transducer is connected to a
of BD.) small and has little mass. Wheatstone bridge circuit
Invasive arterial pressure monitoring 179
10. Mean blood pressure is the in order to detect damping or pressure wave. Increased
average pressure throughout the resonance. The monitoring resonance elevates the systolic
cardiac cycle. As systole is system should be able to apply and lowers the diastolic
shorter than diastole, the mean an optimal damping value of pressures. The mean pressure
arterial pressure (MAP) is 0.64. should stay unchanged.
slightly less than the value half a) Damping is caused by Resonance can be due to
way between systolic and dissipation of stored energy. a stiff, non-compliant
diastolic pressures. An estimate Anything that takes energy diaphragm and tube. It is
of MAP can be obtained by out of the system results in a worse with tachycardia.
adding a third of the pulse progressive diminution of 2. To determine the optimum
pressure (systolic diastolic amplitude of oscillations. damping of the system, a square
pressure) to the diastolic Increased damping lowers wave test (fast flush test) is used
pressure. MAP can also be the systolic and elevates the (Fig. 11.8). The system is flushed
determined by integrating a diastolic pressures with loss by applying a pressure of
pressure signal over the duration of detail in the waveform. 300mmHg (compress and
of one cycle, divided by time. Damping can be caused by release the flush button or pull
an air bubble (air is more the lever located near the
The natural frequency compressible in comparison to transducer). This results in a
This is the frequency at which the the saline column), clot or square waveform, followed by
monitoring system itself resonates a highly compliant, soft oscillations:
and amplifies the signal by up to transducer diaphragm and a. in an optimally damped
2040%. This determines the tube. system, there will be two or
frequency response of the monitoring b) Resonance occurs when the three oscillations before
system. The natural frequency frequency of the driving force settling to zero
should be at least 10 times the coincides with the resonant b. an overdamped system settles
fundamental frequency. The natural frequency of the system. If the to zero without any
frequency of the measuring system natural frequency is less than oscillations.
is much higher than the primary 40Hz, it falls within the c. an underdamped system
frequency of the arterial waveform range of the blood pressure oscillates for more than three
which is 12Hz, corresponding to a and a sine wave will be to four cycles before settling
heart rate of 60120beats/min. superimposed on the blood to zero.
Stiffer (low compliance) tubing or a
shorter length of tubing (less mass)
produce higher natural frequencies. 150
This results in the system requiring a
much higher pulse rate before
(mmHg)
amplification. 100
The natural frequency of the
monitoring system is:
50 A B C
1. directly related to the catheter
diameter Fig. 11.7 Arterial pressure waveform. (A) Correct, optimally damped waveform. (B)
Underdamped waveform. (C) Overdamped waveform.
2. inversely related to the square
root of the system compliance
3. inversely related to the square 300 mmHg
root of the length of the tubing 300 mmHg
st
1 Oscillation
4. inversely related to the square
2nd Oscillation
root of the density of the fluid in
the system.
(mmHg)
midaxillary line. Raising or not be inserted in sites with 100
lowering the transducer above evidence of infection and
or below the level of the right trauma or through a
atrium gives error readings vascular prosthesis. 50
equivalent to 7.5mmHg for 11. Periodic checks, calibrations and
each 10cm. re-zeroing are carried out to 50 100 150
4. Ischaemia distal to the cannula prevent baseline drift of the True blood pressure (mmHg)
is rare but should be monitored transducer electrical circuits.
for. Multiple attempts at Zero calibration eliminates the Gradient Offset Ideal
insertion and haematoma effect of atmospheric pressure on drift drift responce
formation increase the risk of the measured pressure. This Fig. 11.9 Calibration of invasive pressure
ischaemia. ensures that the monitor monitor.
5. Arterial thrombosis occurs in indicates zero pressure in the
2025% of cases with very rare absence of applied pressure, so
adverse effects such as ischaemia eliminating the offset drift (zero
or necrosis of the hand. drift). To eliminate the gradient Central venous
Cannulae in place for less than drift, calibration at a higher
24h very rarely cause pressure is necessary. The
catheterization and
thrombosis. transducer is connected to an pressure (CVP)
6. The arterial pressure wave aneroid manometer using a
narrows and increases in sterile tubing, through a The CVP is the filling pressure of
amplitude in peripheral three-way stopcock and the the right atrium. It can be measured
vessels. This makes the systolic manometer pressure is raised directly using a central venous
pressure higher in the dorsalis to 100 and 200mmHg. The catheter. The catheter can also be
pedis than in the radial artery. monitor display should read the used to administer fluids, blood,
When compared to the aorta, same pressure as is applied to drugs, parenteral nutrition and
peripheral arteries contain the transducer (Fig. 11.9). sample blood. Specialized catheters
less elastic fibres so they are can be used for haemofiltration,
stiffer and less compliant. The haemodialysis (see Chapter 13,
arterial distensibility determines Haemofiltration) and transvenous
the amplitude and contour of Invasive arterial blood pressure pacemaker placement.
the pressure waveform. Consists of an arterial cannula, The tip of the catheter is usually
In addition, the narrowing a heparinized saline column, a positioned in the superior vena cava
and bifurcation of arteries flushing device, a transducer, an at the entrance to the right atrium.
leads to impedance of forward amplifier and an oscilloscope. The internal jugular, subclavian and
blood flow, which results in In addition to blood pressure, basilic veins are possible routes for
backward reflection of the other parameters can be central venous catheterization. The
pressure wave. measured and estimated such as subclavian route is associated with
7. There is risk of bleeding due to myocardial contractility, vascular the highest rate of complications but
disconnection. tone and stroke volume. is convenient for the patient and for
8. Inadvertent drug injection The waveform should be the nursing care.
causes distal vascular occlusion displayed to detect any The Seldinger technique is the
and gangrene. An arterial resonance or damping. common and standard method used
cannula should be clearly The measuring system should be for central venous catheterization
labelled. able to cover a frequency range (Fig. 11.10) regardless of catheter
9. Local infection is thought to of 0.540Hz. type. The procedure should be done
be less than 20%. Systemic The monitoring system should under sterile conditions:
infection is thought to be be able to apply an optimal
less than 5%. This is more damping value of 0.64. 1. Introduce the needle into the
common in patients with an vein using the appropriate
182 11 Invasive monitoring
PRESSURE TRANSDUCER
Vein
1. A similar measuring system to
that used for invasive arterial
pressure monitoring (catheter,
Catheter heparinized saline column,
Dilator in advanced
C D transducer, diaphragm, flushing
& out over J wire
device and oscilloscope system).
The transducer is positioned at
the level of the right atrium.
2. A measuring system of limited
frequency range is adequate
because of the shape of the
waveform and the values of the
central venous pressure.
J wire out
E
FLUID MANOMETER
(FIG. 11.12)
Fig. 11.15 An adult triple lumen catheter. (Courtesy of Vygon (UK) Ltd. Vygon (UK)
Dialysis catheters
Ltd.) These are large-calibre catheters
designed to allow high flow rates of
2. They are used when a central can have one, two or three
at least 300mL/min. They are made
catheter is required in situations lumens (Fig. 11.18).
of silicone or polyurethane. Most of
when it is undesirable to gain 2. The proximal end is tunnelled
them are dual lumen with staggered
access via the internal jugular under the skin for a distance of
end and side holes to prevent
or the subclavian veins, for about 10cm.
admixture of blood at the inflow
example during head and 3. A Dacron cuff is positioned
and outflow portions reducing
neck surgery or prolonged 34cm from the site of entry
recirculation.
antibiotic therapy. They are into the vein under the skin. It
made of soft flexible induces a fibroblastic reaction
Problems in practice and
polyurethane or silicone. to anchor the catheter in place
safety features
(Fig. 11.19). The cuff also
reduces the risk of infection as it 1. Inaccurate readings can be due
Hickman catheters
stops the spread of infection to catheter blockage, catheter
1. These central catheters are made from the site of entry to the inserted too far or using the
of polyurethane or silicone and skin. Some catheters also have wrong zero level.
are usually inserted into the a silver impregnated cuff that 2. Pneumohaemothorax (with an
subclavian vein. The catheter acts as an anti-microbial barrier. incidence of 210% with
Central venous catheterization and pressure (CVP) 185
catheterization), trauma to the and infection rates with the use of sulfadiazine or minocycline/
arteries (carotid, subclavian and central venous catheters rifampin-impregnated CVC in
brachial), air embolism, patients whose catheter is
Education and training of staff
haematoma and tracheal puncture expected to remain in place >5
who insert and maintain the
are complications of insertion. days.
catheters.
3. Sepsis and infection are common Use the maximum sterile barrier
complications with an incidence precautions during central (Centers for Disease Control. 2011.
of 2.820%. Staphylococcus venous catheter insertion.
Guidelines for the prevention of
intravascular catheter-related infections)
aureus and Enterococcus are the
most common organisms.
186 11 Invasive monitoring
Components
PA catheters are available in sizes
58G and are usually 110cm in
length (Fig. 11.21). They have up to
five lumens and are marked at
10-cm intervals:
1. The distal lumen ends in the Fig. 11.21 Argon balloon-tipped flow-guided pulmonary artery catheter with five
lumens. (Courtesy of Argon Medical Devices.)
pulmonary artery. It is used to
measure PA and pulmonary
capillary wedge (PCW) pressures
and to sample mixed venous
blood.
2. The proximal lumen should
ideally open in the right atrium, Up to 1.5mL of air is needed. distance from the internal
being positioned about 30cm When the balloon is inflated, the jugular or the subclavian vein
from the tip of the catheter. It catheter floats with the blood flow to the right atrium is about
can be used to continuously into a pulmonary artery branch 1520cm.
monitor the CVP, to administer (Fig. 11.22). 3. The balloon is partly inflated,
the injectate to measure enabling the blood flow to carry
the cardiac output (by the catheter tip through the
Mechanism of action
thermodilution) or to infuse tricuspid valve into the right
fluids. Depending on the design, 1. Before insertion, flush all the ventricle. Tall pressure waves
a second proximal lumen may lines and test the balloon with (1525mmHg systolic and
be present which is usually 11.5mL of air. 010mmHg diastolic) are
dedicated to infusions of drugs. 2. The distal lumen of the catheter displayed.
3. Another lumen contains two is connected to a transducer 4. As the balloon tip floats through
insulated wires leading to a pressure measuring system for the pulmonary valve into the PA,
thermistor that is about 3.7cm continuous monitoring as the the pressure waveform changes
from the catheter tip. Proximally catheter is advanced. As the with higher diastolic pressure
it is connected to a cardiac catheter passes via the superior (1020mmHg), but similar
output computer. vena cava to the right atrium, systolic pressures. The dicrotic
4. The balloon inflation lumen is low pressure waves (mean of notch, caused by the closure of
used to inflate the balloon which 38mmHg normally) are the pulmonary valve, can be
is situated at the catheter tip. displayed (Fig. 11.23). The noted. The distance from the
Balloon-tipped flow-guided pulmonary artery catheter 189
Components
1. Monitor housing containing:
Preload reduction
a) screen displaying real-time
cardiac output parameters
Positive inotropy Afterload reduction
b) software technology for
computing the cardiac output
Velocity and other parameters
Time
c) electrical connection feeding
the arterial waveform from
the patients bed-side invasive
pressure monitor.
Myocardial depression Afterload increase
Oesophageal Doppler
Minimally invasive and rapid
estimate of cardiac output using
the Doppler principle.
Insulated Doppler probe lies in the
distal oesophagus emitting a high
ultrasound frequency of 4MHz.
Continuous monitoring is
possible, although frequent
probe repositioning is a problem.
LiDCOrapid
This is a cardiac output monitor
that uses arterial pressure waveform Fig. 11.29 The LiDCOrapid cardiac output monitor.
Temperature probes 193
3. Dynamic preload parameters can The response is made linear measure accurately the core
also be generated by the software electronically. This property temperature within 3s.
to assess the stroke volume allows them to accurately 3. The rate of radiation by an
response to a fluid challenge. measure temperature to an order object is proportional to
of 0.1C. temperature to the fourth power.
2. It can be made in very small
Problems in practice and
sizes and is relatively cheap to
safety features Problems in practice and
manufacture.
safety features
The performance of the software 3. It is mounted in a plastic or
may be compromised in the stainless steel probe making it 1. Non-continuous intermittent
following patient groups: mechanically robust, and it can readings.
be chemically sterilized. 2. The probe has to be accurately
1. Patients with aortic valve
4. It is used in PA catheters to aimed at the tympanic
regurgitation.
measure cardiac output. membrane. False low readings
2. Patients being treated with
5. In the negative thermal from the sides of the ear canal
an intra-aortic balloon pump
conductivity thermistors, such can be a problem.
or cardiac arrythmias that
as cobalt oxide, copper oxide 3. Wax in the ear can affect the
will disrupt the usual
and manganese oxide, the accuracy.
arterial waveform pattern.
electrical resistance decreases
3. Patients with highly damped
as the temperature increases.
peripheral arterial lines or with
In the positive thermal THERMOCOUPLES
pronounced peripheral arterial
conductivity thermistors, such
vasoconstriction.
as barium titanate, the electrical These are devices that make use of
4. The Lidco Plus can be
resistance increases with the principle that two different
calibrated using a single-point
the temperature. metals in contact generate a voltage,
lithium indicator dilution
which is temperature dependent
process. This offers more
(Fig. 11.30).
accurate cardiac output Problems in practice and
measurement. safety features
Components
Thermistors need to be stabilized as
they age. 1. Two strips of dissimilar metals
(0.42-mm diameter) of
Temperature different specific heats and
probes INFRARED TYMPANIC in contact from both ends.
THERMOMETER Usually copper-constantan
Monitoring a patients temperature (copper with 40% nickel)
during surgery is a common and Components junctions are used.
routine procedure. Different types of 1. A small probe with a disposable 2. A galvanometer.
thermometers are available. and transparent cover is inserted
into the external auditory
meatus. Voltage measured
THERMISTOR 2. The detector (which consists of
a series of thermocouples called Hot junction
Components a thermopile).
1. A small bead of a temperature-
Metal A
dependent semiconductor.
Mechanism of action
2. Wheatstone bridge circuit. Metal B
1. The detector receives infrared
radiation from the tympanic
Mechanism of action membrane.
Cold junction
Mechanism of action
1. One junction is used as the
measuring junction whereas the
other one is the reference. The
latter is kept at a constant
temperature.
2. The metals expand and
contract to different degrees
with change in temperature
producing an electrical
potential that is compared to a
reference junction. The current
produced is directly
proportional to the temperature Fig. 11.31 Oesophageal/rectal Fig. 11.33 Tympanic membrane
difference between the two temperature probe. thermometer.
junctions, i.e. there is a linear
relationship between voltage
and temperature. accurately reflects core 5. Skin temperature, when
3. The voltage produced is temperature, compared with measured with the core
called the Seebeck effect or lower oesophageal temperature. temperature, can be useful
thermoelectric effect. Thermocouple and thermistor in determining the volaemic
4. The measuring junction probes as well as the infrared status of the patient
produces a potential of 40V probe can be used (Figs 11.32 (Fig. 11.35).
per C. This potential is and 11.33).
measured by an amplifier. 4. Bladder temperature correlates The axilla is the best location for
5. They are stable and accurate to well with the core temperature monitoring muscle temperature,
0.1C. when there is a normal urine making it most suitable for
6. If multiple thermocouples are output (Fig. 11.34). detecting malignant hyperthermia.
linked in series, they constitute a
thermopile. This is done to
improve their sensitivity
FURTHER READING
Temperature probes
They can be thermistors, Centers for Disease Control and
thermocouples or infrared Prevention, 2011. Guidelines for the
thermometers. prevention of intravascular catheter-
Core and skin temperatures can related infections. Online. Available
be measured. at: http://www.cdc.gov/hicpac/pdf/
Core temperature can be guidelines/bsi-guidelines-2011.pdf.
measured from the rectum, National Institute for Health and Clinical
oesophagus, tympanic Excellence, 2008. Inadvertent
membrane or the bladder. perioperative hypothermia. The
management of inadvertent
perioperative hypothermia in adults.
Online. Available at: http://www.nice.
org.uk/nicemedia/pdf/
CG65NICEGuidance.pdf.
NHS, 2008. Infusions and sampling
Fig. 11.35 Skin temperature probe. from arterial lines. Online. Available
at: http://www.nrls.npsa.nhs.uk/resou
rces/?entryid45=59891&p=10.
Shoemaker, W.C., Velmahos, G.C.,
Demetriades, D., 2002. Procedures
and monitoring for the critically ill.
WB Saunders, Philadelphia.
MCQs
In the following lists, which of the statements (a) to (e) are true?
Answers
Chapter 12
Pain
management
and regional
anaesthesia
202 12 Pain management and regional anaesthesia
Fig. 12.1 The Graseby Omnifuse PCA pump. Fig. 12.2 The CADD Legacy portable PCA.
Patient controlled analgesia (PCA) 203
TARGET CONTROLLED
INFUSION PUMPS
Elastomeric pumps
These recently designed light,
portable and disposable pumps
allow continous infusions of local
anaesthetic solutions. Continuous
incisional infiltration or nerve
blocks can be used so allowing the
delivery of continuous analgesia
(Fig. 12.5).
Components
1. A small balloon-like pump that
is filled with local anaesthetic.
Variable volumes of 100
600mL are available.
2. Specially designed catheters with
lengths of 730cm and of
different gauges.
3. A bacterial filter and a flow
Fig. 12.5 The On-Q elastomeric pump. Note the flow restricter, bacterial filter,
restrictor.
anti-syphon valve and the attached catheter. (Courtesy of I-Flow Corporation.)
Mechanism of action
3. An extra on-demand bolus
Problems in practice and
1. The balloon deflates slowly facility is available in some
safety features
and spontaneously delivering designs. This allows boluses of
a set amount of local 5mL solution with a lock-out 1. Some of the local anaesthetic
anaesthetic solution per hour. time of 60min. may get absorbed into the
Rates of 214mL/h can be 4. Some designs allow the balloon.
programmed. simultaneous infusion of two 2. The infusion rate profile can
2. Catheters are designed with surgical sites. vary throughout the infusion. It
multiple orifices allowing the 5. Silver-coated dressing for is thought that the initial rate is
infusion of local anaesthetic anti-microbial effect is higher than expected initially
solution over a large area. provided. especially if the pump is under
Epidural needles 205
filled. The infusion rates tend to 2. The needle wall is thin in order 6. A combined spinalepidural
decrease over the infusion to allow a catheter to be inserted technique is possible using a
period. through it. 26-G spinal needle of about
3. It is important to follow the 3. The needle is provided with a 12cm length with a standard
manufacturers instructions stylet introducer to prevent 16-G Tuohy needle. The Tuohy
regarding positioning of the occlusion of the lumen by a needle is first positioned in the
device in relation to the body core of tissue as the needle is epidural space then the spinal
and ambient temperature. inserted. needle is introduced through it
Changes in temperature can 4. The bevel (called a Huber point) into the subarachnoid space
affect the flow rate. A change is designed to be slightly oblique (Fig. 12.7). A relatively high
of 10C in the temperature of at 20 to the shaft, with a rather pressure is required to inject
water-based fluids results in blunt leading edge. through the spinal needle
altered viscosity, which causes a 5. Some designs allow the wings because of its small bore. This
2030% change in flow rate. at the hub to be added or might lead to accidental
removed. displacement of the tip of the
6. The commonly used gauges are needle from the subarachnoid
either 16G or 18G. space leading to a failed or
partial block. To prevent this
Epidural needles happening, in some designs, the
Mechanism of action
spinal needle is anchored to
Epidural needles are used to identify
1. The markings on the needle the epidural needle to prevent
and cannulate the epidural space.
enable the anaesthetist to displacement (Fig. 12.8).
The Tuohy needle is widely used in
determine the distance between
the UK (Fig. 12.6).
the skin and the epidural space.
Problems in practice and
Hence the length of the catheter
safety features
Components left inside the epidural space can
be estimated. 1. During insertion of the catheter
1. The needle is 10cm in length
2. The shape and design of the through the needle, if it is
with a shaft of 8cm (with 1-cm
bevel (Fig 12.7) enable the necessary to withdraw the
markings). A 15-cm version
anaesthetist to direct the catheter catheter, the needle must be
exists for obese patients.
within the epidural space (either withdrawn simultaneously. This
in a cephalic or caudal is because of the risk of the
direction). catheter being transected by the
3. The bluntness of the bevel also oblique bevel.
minimizes the risk of accidental 2. In accidental dural puncture,
dural puncture. there is a high incidence of
4. Some anaesthetists prefer winged postdural headache due to the
epidural needles for better epidural needles large bore (e.g.
control and handling of the 16G or 18G).
needle during insertion. 3. Wrong route errors: in order to
5. A paediatric 19-G, 5-cm long avoid administering drugs that
Tuohy needle (with 0.5-cm were intended for intravenous
markings), allowing the passage administration, all epidural
of a 21-G nylon catheter, is bolus doses are performed using
available. syringes, needles and other
Fig. 12.6 18-G Tuohy needle. Note the Fig. 12.7 Detail of a spinal needle introduced through a Tuohy needle (top); an epidural
1cm markings along its shaft. catheter passing through a Tuohy needle (bottom).
206 12 Pain management and regional anaesthesia
Mechanism of action
1. The catheters are designed
to pass easily through their
matched gauge epidural needles.
2. The markings enable the
anaesthetist to place the desired
length of catheter within the
epidural space (usually 35cm).
3. There are catheters with a single
port at the distal tip. These offer
a rather sharp point and increase
the incidence of catheter-induced
vascular or dural puncture.
4. An epidural fixing device can
be used to prevent the catheter
falling out. The device clips on
the catheter. It has an adhesive
Fig. 12.8 The Portex CSEcure combined spinal-epidural device. The spinal needle (top);
flange that secures it to the skin.
the epidural needle (middle); the spinal needle inserted and anchored to the epidural
needle (bottom). The device does not occlude the
catheter and does not increase
the resistance to injection
devices with safer connectors Teflon and biologically inert. (Fig. 12.11).
that cannot connect with The 16-G version has an
intravenous Luer connectors. external diameter of about
Problems in practice and
1mm and an internal diameter
safety features
of 0.55mm.
Epidural needle
2. The distal end has two or three 1. The patency of the catheter
10-cm Tuohy needle with the
side ports with a closed and should be tested prior to
oblique bevel (Huber point) is
rounded tip in order to reduce insertion.
most popular. Five- and 15-cm
the risk of vascular or dural 2. The catheter can puncture an
lengths exist.
puncture (see Fig. 12.7). epidural vessel or the dura at the
It has 1-cm markings to measure
Paediatric designs, 18G or time of insertion or even days
the depth of the epidural space.
19G, have closer distal side later.
A stylet introducer is provided
ports. 3. The catheter should not be
with the needle.
3. Some designs have an open end. withdrawn through the Tuohy
A combined epiduralspinal
4. The distal end of the catheter needle once it has been threaded
technique is becoming more
is marked clearly at 5-cm beyond the bevel as that can
popular.
intervals, with additional 1-cm transect the catheter. Both needle
markings between 5 and 15cm and catheter should be removed
(Fig. 12.10). in unison.
5. The proximal end of the catheter 4. It is almost impossible to predict
Epidural catheter, is connected to a Luer lock and in which direction the epidural
a filter (Fig. 12.10). catheter is heading when it is
filter and loss of 6. In order to prevent kinking, advanced.
resistance device some designs incorporate a 5. Once the catheter has been
coil-reinforced catheter. removed from the patient, it
(Fig. 12.9) 7. Some designs are radio-opaque. should be inspected for any signs
These catheters tend to be more of breakage. The side ports are
THE CATHETER
rigid than the normal design. points of catheter weakness
They can be used in patients where it is possible for the
Components
with chronic pain to ensure catheter to break. Usually, if a
1. 90-cm transparent, malleable correct placement of the portion of the catheter were to
tube made of either nylon or catheter. remain in the patient after
Epidural catheter, filter and loss of resistance device 207
Mechanism of action
1. The large 22-G needle is more
rigid and easier to direct. It gives
a better feedback feel as it passes
through the different tissue
layers.
2. The CSF is slower to emerge
from the smaller sized needles.
Aspirating gently with a syringe
can speed up the tracking back
of CSF.
3. Continuous spinal anaesthesia
can be achieved by inserting
34cm of the 28-G spinal
microcatheter into the
Fig. 12.12 An epidural catheter with a knot near its tip. (Courtesy of Dr MS Rao.) subarachnoid space.
Fig. 12.13 Different size spinal needles with their introducer. From left; 27G (grey), Fig. 12.15 Pencil-shaped Whitacre bevel
26G (brown), 25G (orange), 24G (purple) and 22G (black). (Courtesy of Smiths (left) and cutting Quincke bevel (right).
Medical.)
Spinal microcatheters
1. They are difficult to advance.
Fig. 12.17 The standard Luer connection spinal needle (left) and the Portex Correct
Inject spinal needle designed to avoid wrong route error injections.
2. There is a risk of trauma to
nerves.
3. Cauda equina syndrome is
thought to be due to the
potential neurotoxicity from the
anaesthetic solutions rather than
the microcatheter.
Spinal needles
They have a stylet and a
transparent hub.
Different gauges from 18G to
32G.
The bevel can be cutting (Yale
and Quincke) or pencil-like
(Whitacre and Sprotte).
Can cause dural headache.
Continuous spinal block using a
28-G microcatheter is possible.
Fig. 12.18 Portex Correct Inject range of spinal needles and syringe designed to avoid
wrong route error injections.
Components
1. They are made of steel with a
Luer-lock attachment.
2. They have short, rather blunt
bevels in order to cause minimal
trauma to the nervous tissue.
The bluntness makes skin
insertion more difficult. This can
be overcome by a small incision.
3. The needles have transparent
hubs which allow earlier
recognition of intravascular
Fig. 12.19 A range of Smiths Medical insulated peripheral nerve block needles of placement while performing
different lengths. blocks.
Nerve block needles 211
INSULATED NEEDLES
NHS, 2011. Safer spinal (intrathecal), Obstetric Anaesthetists Association, The Royal College of Anaesthetists,
epidural and regional devices. 2012. Spinal/epidural needle 2004. Good practice in the
Online. Available at: http://www.nrls. design. Online. Available at: management of continuous epidural
npsa.nhs.uk/resources/?entryid45=94 http://oaa-anaes.ac.uk/content. analgesia in the hospital setting.
529&p=2 asp?contentid=367 Online. Available at: www.rcoa.
ac.uk/docs/Epid.Analg.pdf.
216 12 Pain management and regional anaesthesia
MCQs
In the following lists, which of the statements (a) to (e) are true?
1. Epidural catheters and filters: 3. Nerve stimulators in regional 5. Which of the following is/are
a) A minimum of 10cm of the anaesthesia: true:
catheter should be inserted a) They enable the block to be a) Using ultrasound guidance in
into the epidural space. performed even without full regional anaesthesia, a
b) The catheter should not be knowledge of the anatomy. frequency range of 1014kHz
withdrawn through the Tuohy b) In the insulated nerve block is adequate.
needle once it has been needle, the current passes b) It is important to prevent free
threaded beyond the bevel. through the tip only. flow from the syringe pump.
c) Catheters with a single port at c) In the non-insulated nerve c) There is no need to use
the distal tip reduce the block needle, the current anti-reflux valves in other
incidence of vascular or dural passes through the tip and the infusion lines.
puncture. shaft. d) Sector transducers can achieve
d) The filter should be changed d) A catheter can be used for better images when in regional
every 8 hours. continuous nerve/plexus anaesthesia.
e) Catheters can be blockade. e) Syringe pumps should be
radio-opaque. e) The immobile needle positioned at the same level as
technique improves the the patient.
success rate of the block.
2. Regional anaesthesia using a
nerve stimulator:
SINGLE BEST ANSWER (SBA)
a) The needles used have sharp 4. Incidence of spinal headache:
tips to aid in localizing the a) Yale and Quincke needle
6. Spinal needles:
nerves/plexuses. design have lower incidence of
a) Can be used to perform
b) AC current is used to locate spinal headache.
epidural block.
the nerve. b) It is inversely proportional to
b) Have an opaque hub to allow
c) A current of 1A is usually the size of the needle used.
identification of CSF fluid.
used to locate a nerve. c) It is similar in young and
c) Dural puncture headache is
d) Paraesthesia is not required elderly patients.
eliminated with the cutting
for successful blocks. d) It is proportional to the
bevels.
e) 50-Hz frequency stimuli are number of dural punctures.
d) New guidelines make it
used. e) It is reduced using a pencil-
impossible to administer
shaped needle tip.
drugs.
e) Can be used as part of a
combined spinal/epidural
procedure.
Answers 217
Answers
needle will pass the different current, the tip of the needle
1. Epidural catheters and filters:
layers of tissues easily with has to be very close to the
a) False. 35cm of the catheter
minimal feedback. A blunt nerve before stimulation is
is left in the epidural space.
needle will provide much visible.
This reduces the incidence of
better feedback. c) True. In a non-insulated
vascular or dural puncture,
b) False. DC current from a needle, the current passes
segmental or unilateral block
battery is used to operate through both the tip and
(as the catheter can pass
nerve stimulators. By avoiding shaft. This might lead to
through an intervertebral
AC current, patient safety is nerve stimulation by current
foramina) and knotting.
improved. from the shaft even when the
b) True. The withdrawal of the
c) False. This is a very high tip is far away from the nerve.
catheter through the Tuohy
current. A current range of up This obviously leads to a
needle after it has been
to 5mA is needed in locating failed block.
threaded beyond the bevel can
the nerve. A current of d) True. After successful nerve
lead to the transection of the
0.250.5mA is used to stimulation, a catheter can be
catheter. This usually happens
stimulate the nerve fibres. inserted. This allows a
when the catheter punctures a
Using a very high current, the prolonged and continuous
vessel during insertion. The
tip of the needle might be far block using an infusion or
needle and the catheter should
away from the nerve but boluses.
be removed together and
might still lead to stimulation e) True. The immobile needle
another attempt should be
of the nerve fibres or the technique allows one operator
made to reinsert the needle
muscle fibres directly leading to maintain the needle in the
and catheter.
to the failure of the block. correct position while the
c) False. Catheters with a single
d) True. There is no need for second operator injects the
port at the distal tip increase
paraesthesia in order to local anaesthetic. This also
the incidence of vascular or
achieve a successful block reduces the risk of accidental
dural puncture. This is due to
using a nerve stimulator. intravascular injection.
the sharp point at the end of
Paraesthesia implies that the
the catheter. In contrast,
needle is touching the nerve. 4. Incidence of spinal headache:
catheters with side ports have
With a nerve stimulator, the a) False. Yale and Quincke have
a closed and rounded end
nerve can be stimulated a higher incidence of spinal
thus reducing the incidence of
electrically without being headache. This is due to the
vascular or dural puncture.
touched. traumatic bevel cutting the
d) False. The filter can be used
e) False. Stimuli with a dural fibres producing a
for up to 24h.
frequency of 12Hz are used. ragged tear which allows CSF
e) True. Some catheters are
Tetanic stimuli (e.g. 50-Hz leakage.
designed to be radio-opaque.
frequency) are not used b) False. The incidence is directly
They are more rigid than the
because of the discomfort proportional to the size of the
standard design. They are
caused. needle used. Using a 20-G
mainly used in patients with
spinal needle causes a 30%
chronic pain to ensure the
3. Nerve stimulators in regional incidence of spinal headache
correct placement of the
anaesthesia: whereas a 26-G needle has a
catheter.
a) False. Full knowledge of the 1% incidence of headache.
anatomical structure is c) False. The incidence of spinal
2. Regional anaesthesia using a
essential for a successful headache is much higher in
nerve stimulator:
block. the young than in elderly
a) False. There is a need for
b) True. As the tip is the only patients.
feedback from the needle as it
non-insulated part of the d) True. The incidence of spinal
goes through the different
needle, the current passes only headache is increased with
layers of tissue. A sharp
through it. Using a small multiple dural punctures.
218 12 Pain management and regional anaesthesia
e) True. The pencil-shaped the syringe should be securely The true echogenicity of a
needle tip separates rather clamped to the pump. nerve is only captured if the
than cuts the longitudinal Siphoning can also occur if sound beam is oriented
dural fibres. After removal of there is a crack in the syringe perpendicularly to the nerve
the needle, the dura has a allowing air entry. axis. This can best be
higher chance of sealing, c) False. An anti-reflux valve achieved with linear array
reducing the incidence of should be inserted in any transducers with parallel
spinal headache. other line that is connected to sound beam emission.
the infusion line. Anti-reflux e) True. Gravitational pressure
5. Which of the following is/are valves prevent backflow up can be generated to overcome
true: the secondary (usually with the friction between a
a) False. The frequencies needed lower pressure) should a distal non-secured plunger and
for nerve blocks are in the occlusion occur and avoid a barrel especially if the pump
range of 1014MHz. Most subsequent inadvertent bolus. is positioned more than
modern ultrasound devices d) False. Sector transducers emit 100cm above the patient.
can generate these frequencies. diverging sound waves, such
b) True. Anti-siphon valves are that the echotexture of the 6. e)
used to prevent free flow from nerves will only be displayed
the syringe pump. In addition, in the centre of the image.
219
Chapter 13
Additional
equipment
used in
anaesthesia
and intensive
care
220 13 Additional equipment used in anaesthesia and intensive care
Double lumen
venous line
Arterial blood gas
analyser (Fig. 13.5)
Roller
pump Heparin In order to measure arterial blood
gases, a sample of heparinized,
Replacement anaerobic and fresh arterial blood is
Blood flow in fluid needed.
1. Heparin should be added to the
sample to prevent clotting
during the analysis. The heparin
Filtrate should only fill the dead space
of the syringe and form a thin
Filtration
Haemofiltration
An effective method of renal
support in critically ill patients
using ultrafiltration of the blood.
Arteriovenous or venovenous
lines are connected to the
extracorporeal circuit (filter and
a pump).
Synthetic filters with a surface
area of 0.51.5m2 are used.
Fig. 13.5 The ABL800FLEX blood gas analyser. (Courtesy of Radiometer Medical ApS.)
224 13 Additional equipment used in anaesthesia and intensive care
oxygen electrode
Electrolyte (H2O + NaCl)
This measures the oxygen partial
pressure in a blood (or gas) sample Cathode(s) (platinum)
(Fig. 13.6).
- -
- -
PO2 Oxygen permeable membrane
Components
O2 O2 O2 O2 Blood sample (or calibration gas)
1. A platinum cathode sealed in a O2 O2
O2
glass body.
2. A silver/silver chloride anode.
Fig. 13.6 Mechanism of action of the oxygen electrode. (Reproduced with permission
3. A sodium chloride electrolyte from AVL Medical Instruments UK Ltd.)
solution.
4. An oxygen-permeable Teflon
Cathode reaction: 2. Protein particles can precipitate
membrane separating the
on the membrane affecting the
solution from the sample. O2 + 2H 2O + 4e = 4OH
performance.
5. Power source of 700mV.
Electrolyte reaction:
NaCl + OH = NaOH + Cl Polarographic oxygen electrode
Mechanism of action Consists of a platinum cathode,
Anode reaction: silver/silver chloride anode,
1. Oxygen molecules cross the
membrane into the electrolyte Ag + Cl = AgCl + e electrolyte solution, membrane
solution at a rate proportional and polarization potential of
3. The oxygen partial pressure in 700mV.
to their partial pressure in the
the sample can be measured since The flow of the electrical current
sample.
the amount of current is linearly is proportional to the oxygen
2. A very small electric current
proportional to the oxygen partial pressure in the sample.
flows when the polarization
partial pressure in the sample. Requires regular maintenance.
potential is applied across the
4. The electrode is kept at a
electrode in the presence of
constant temperature of 37C.
oxygen molecules in the
electrolyte solution. Electrons
are donated by the anode and
Problems in practice and pH electrode
safety features
accepted by the cathode,
This measures the activity of the
producing an electric current 1. The membrane can deteriorate
hydrogen ions in a sample.
within the solution. The circuit and perforate, affecting the
Described mathematically, it is:
is completed by the input performance of the electrode.
terminal of the amplifier. Regular maintenance is essential. pH = log[H + ]
Carbon dioxide electrode (Severinghaus electrode) 225
Amplifier pH display
Calomel reference
Saturated solution of Glass electrode
electrode
potassium chloride
H+ H+
H+ H+
Membrane Blood sample
(or calibration buffer solution)
is synchronized counterpulsation.
PCO2 CO2 permeable membrane
It is usually inserted using a
Blood sample (or calibration gas) percutaneous femoral approach,
CO2 CO2 CO2
over a guidewire, under fluoroscopic
or transoesophageal echo guidance.
The correct position of the pump is
Fig. 13.8 Mechanism of action of the carbon dioxide electrode.
in the descending aorta, just distal to
the left subclavian artery (Fig 13.9).
Diastolic augmentation
140
Coronary perfusion
Unassisted Assisted
120 systole systole
(mmHg)
100
Balloon
inflation
80
Unassisted aortic
end diastolic pressure
60 Assisted aortic
end diastolic pressure
MVO, Demand
Fig. 13.11 The changes that intra-aortic balloon pump therapy causes to the arterial
waveform and their consequences.
diastole, immediately after the balloon to aorta size, heart rate Aortic dissection.
closure of the aortic valve. This and rhythm, compliance of the Major coagulopathy.
leads to an increase in peak aorta and peripheral vessels and Severe bilateral peripheral arterial
Fig. 13.12 Intravenous giving sets: (A) Intrafix Safe Set adult fluid set; (B) SangofixB blood adult giving set with a 200-m filter;
(C) Sangopur adult blood set with a 40-m filter; (D) Dosifix paediatric fluid set with burrette. (Courtesy of B Braun Medical.)
Fig. 13.13 A range of intravenous cannulae. Fig. 13.14 Smiths Medical Protective
Acuvance cannula designed to reduce
the risk of needle stick injury.
and the patient and controls More recent designs are the internal diameter of 4mm is as
the drip rate. An injection site closed and integrated cannulae follows:
should be close to the patient (Fig. 13.15). A closed system may
20G: 4080mL/min.
to reduce the dead space. offer better protection against
18G: 75120mL/min.
c) Drop size is 60 drops per bacterial exposure than conventional
16G: 130220mL/min.
1mL of clear fluid. A burette open ports. As the blood does not
14G: 250360mL/min.
with a drop size similar to naturally escape from the catheter
the adults version (15 drops hub, these devices further minimize
per mL) is used for blood the risk of exposing the clinician
transfusion. to blood during the insertion
d) 0.2-micron filters can be procedure. Blood warmers
added in line to filter out air Using distilled water at a
and foreign bodies, e.g. glass temperature of 22C and under These are used to warm blood (and
or plastic particles. Infusion- a pressure of 10kPa, the flow other fluids) before administering
related thrombophlebitis can through a 110-cm tubing with an them to the patient. The aim is to
be reduced by the use of these
filters.
Intravenous
cannulae
Intravenous cannulae are made of
plastic. They are made by different
manufacturers with different
characteristics (Fig. 13.13).
Intravenous cannulae can be
either with or without a port. Some
designs offer protection against
the risk of needle stick injuries
(Fig. 13.14), covering the sharp
needle tip with a blunt end. Fig. 13.15 The BD Nexiva IV closed and integrated cannula. (Courtesy of BD Medical.)
230 13 Additional equipment used in anaesthesia and intensive care
20
defibrillator electrode pads are
currently used instead of 10
paddles, offering hands-free
defibrillation. 0
5. Most of the current is dissipated
10
through the resistance of the 0 1 2 3 4 5 6 7 8 9 10 11 12
skin and the rest of the tissues 20 Time (msec)
and only a small part of the A
total current (about 35A)
30
flows through the heart. The 25
impedance to the flow of current 20
Current (amps)
20
depending on transthoracic
impedance (e.g. larger 10
patients with high
0
transthoracic impedance
received considerably less 10
transmyocardial current than
20
smaller patients).
b) Biphasic defibrillators deliver C 0 1 2 3 4 5 6 7 8 9 10 11 12
a two-phased current flow Time (msec)
in which electrical current Fig. 13.22 Defibrillator waveforms. (A) Monophasic defibrillator waveform. (B) Biphasic
flows in one direction for truncated exponential defibrillator waveform. (C) Biphasic rectilinear defibrillator
a specified duration, then waveform.
Defibrillator 233
Defibrillators
AED and manual versions are
available.
A step-up transformer increases
mains voltage then a rectifier
converts it to direct current. DC
energy is discharged from a
capacitor.
Modern defibrillators use a
biphasic current flow.
Implanted automatic internal
defibrillators are becoming more
popular with pacemaker
capabilities.
Chest drain Fig. 13.25 Seldinger chest drainage kit. (Courtesy of Smiths Medical.)
is raised above the level of the the piezoelectric (PE) effect. The
patient. The collection chamber PE materials vibrate when a
should be kept below the level varying voltage is applied. The
of the patient at all times to frequency of the voltage applied
prevent fluid being siphoned into determines the frequency of the
the pleural space. sound waves produced. The
2. Absence of oscillations may thickness of the PE element
indicate obstruction of the determines the frequency at
drainage system by clots or which the element will vibrate
kinks, loss of subatmospheric most efficiently, i.e. its resonant
pressure or complete frequency (RF). RF occurs when
re-expansion of the lung. the thickness of element is half
3. Persistent bubbling indicates the wavelength of the sound
a continuing bronchopleural wave generated.
air leak. 3. An image is generated when the
4. Clamping a pleural drain in the pulse wave emitted from the
presence of a continuing air leak transducer is transmitted into
may result in a tension the body, reflected off the tissue
pneumothorax. interface and returned to the
transducer. Returning US waves
cause PE crystals (elements)
within the transducer to vibrate.
Fig. 13.27 The XD11XE Philips This causes the generation of a
Chest drain ultrasound machine. (Courtesy of Philips voltage. Therefore, the same
An air-tight system to drain the Health Care.) crystals can be used to send and
pleural cavity usually inserted at
receive sound waves.
the fourthfifth intercostal space
4. Two-dimensional images of
in the midaxillary line.
structures are displayed.
The underwater seal chamber
Procedures requiring precise
should be about 100cm below
connected to a control unit that needle placement such as venous
the level of the patient.
displays the ultrasound image. cannulation or nerve blocks can
Absence of oscillation is seen
be performed under direct
with complete lung expansion,
ultrasound control. This helps to
obstruction of the system or loss Components minimize the possible risks of
of negative pressure.
1. Beamformer: applies high- the procedure.
Persistent bubbling is seen with
amplitude voltage to energize the 5. The image can be displayed in a
a continuing bronchopleural
crystals. number of modes:
air leak.
2. Transducer: converts electrical a) A-mode (amplitude); not used
energy to mechanical (US) any more
energy and vice versa. b) B-mode (brightness); most
3. Receiver: detects and amplifies commonly used for regional
weak signals. anaesthesia
c) M-mode (motion); most
The ultrasound 4. Memory: stores video display.
commonly for cardiac and
machine (Fig. 13.27) foetal imaging
d) 2D-real time.
Mechanism of action
Ultrasound is a longitudinal 6. Structures can then be
high-frequency wave. It travels 1. The probe transmits and receives identified via their ultrasound
through a medium by causing local the ultrasound beam once placed characteristics and anatomical
displacement of particles. This in contact with the skin via relationships.
particle movement causes changes in acoustic coupling jelly. 7. Increasing the depth allows
pressures with no overall movement 2. Ultrasound is created by visualization of deeper
of the medium. An ultrasound converting electrical energy into structures. The depth of the
machine consists of a probe mechanical vibration utilizing image should be optimized so
236 13 Additional equipment used in anaesthesia and intensive care
MCQs
In the following lists, which of the statements (a) to (e) are true?
Answers
Chapter 14
Electrical
safety
242 14 Electrical safety
The electrical equipment used in the such that 1 watt (W) of power is
operating theatre and intensive care dissipated between two points, the
unit is designed to improve patient
Principles of potential difference between those
care and safety. At the same time, electricity (Fig. 14.1) points is 1 volt (V).
however, there is the potential of
exposing both the patient and staff Electric current (I)
Electrical resistance (R)
to an increased risk of electric
An electric current is the flow of
shock. It is essential for the Electrical resistance is the resistance
electrons through a conductor past
anaesthetist to have a thorough along a conductor to the flow
a given point per unit of time,
understanding of the basic principles of electrical current. It is not
propelled by the driving force, i.e.
of electricity, even though these dependent on the frequency of the
the voltage (potential difference).
devices include specific safety current. Electrical resistance is
The current is measured in amperes
features. measured in ohms ().
(A). One ampere represents a flow
In the UK, mains electricity is
of 6.24 1018 electrons (one
supplied as an alternating current
coulomb of charge) past a specific Impedance (Z)
with a frequency of 50Hz. The
point in 1 second.
current travels from the power Impedance is the sum of the forces
station to the substation where it is 1. Direct current (DC): the current that oppose the movement of
converted to mains voltage by a flows in one direction (e.g. flow electrons in an AC circuit. The unit
transformer. From the substation, from a battery). for impedance is the ohm (). The
the current travels in two 2. Alternating current (AC): the term impedance covers resistors,
conductors, the live and neutral flow of electrons reverses capacitors and inductors and is
wires. The live wire is at a potential direction at regular intervals dependent on the frequency of the
of 240V (or more accurately (e.g. mains supply); in the UK, current. Substances with high
240RMS (root mean square)). The the frequency of AC is 50 cycles impedance are known as insulators.
neutral is connected to the earth at per second (Hz). Substances with low impedance
the substation so keeping its are known as conductors. The
potential approximately the same as impedance through capacitors
Potential difference or
earth. The live wire carries the and inductors is related to the
voltage (V)
potential to the equipment whereas frequency (Hz) at which AC
the neutral wire returns the current It is the electrical force that drives reverses direction. Such an
back to the source, so completing the electric current. When a current impedance, i.e. frequency related,
the circuit. of 1A is carried along a conductor, is known as reactance (X).
Class I Equipment This type of equipment offers basic protection whereby the live, neutral and
earth wires do not come into contact with each other. There is a secondary
protection whereby parts that can be touched by the user, such as the metal
case, are insulated from the live electricity and connected to an earth wire via
the plug to the mains supply. There are fuses positioned on the live and
neutral supply in the equipment. In addition, in the UK, a third fuse is
positioned on the live wire in the mains plug. This fuse melts and disconnects
the electrical circuit in the event of a fault, protecting the user from electrical
shock. The fault can be due to deteriorating insulation, or a short circuit,
making the metal case live. Current will pass to earth causing the fuse to
blow (this current is called leakage current). Some tiny non-fault leakage
currents are always present as insulation is never 100% perfect. A faultless
earth connection is required for this protection to function.
Class II Equipment This type of equipment, also called double-insulated equipment, has double
or reinforced insulation protecting the accessible parts. There is no need to
earth this type of equipment. The power cable has only live and neutral
conductors with only one fuse.
244 14 Electrical safety
Class III Equipment This type of equipment does not need electrical supply exceeding 24V AC
or 50V DC. The voltage used is called safety extra low voltage (SELV).
Although this does not cause an electrical shock, there is still a risk of a
microshock. This equipment may contain an internal power source or be
connected to the mains supply by an adaptor or a SELV transformer. The
power input port is designed to prevent accidental connection with another
cable.
Type B Equipment This may be class I, II or III mains-powered equipment or equipment with an
internal power source. This equipment is designed to have low leakage
currents, even in fault conditions, such as 0.5mA for class I and 0.1mA for
class II. It may be connected to the patient externally or internally but is not
considered safe for direct connection to the heart.
Type BF Equipment This is similar to type B equipment, but the part applied to the patient is
isolated from all other parts of the equipment using an isolated (or floating)
circuit. This isolation means that allowable leakage current under single fault
conditions is not exceeded even when 1.1 times the rated mains voltage is
applied between the attachment to the patient and the earth. The maximum
AC leakage current is 0.1mA under normal conditions and under a single
fault condition is 0.5mA. It is safer than type B but still not safe enough for
direct connection to the heart.
Type CF Equipment This can be class I or II equipment powered by mains with an internal
electrical power source, but considered safe for direct connection to the
heart. Isolated circuits are used. There is no risk of electrocution by leakage
currents (allows 0.05mA per electrode for class I and 0.01mA per electrode
for class II). This provides a high degree of protection against electrical
shock. This is used in ECG leads, pressure transducers and thermodilution
computers.
Attention! The user must consult the accompanying documents for any equipment. A
black triangle on a yellow background with an exclamation mark means that
there is no standardized symbol for the hazard:
Protective Earth The equipment itself has its own earth connection via the green-and-yellow
lead in the conventional three-pin plug. The earth lead is connected to the
external case of the equipment so reducing it to zero potential. Although this
provides some protection, it does not guarantee it.
Functional Earth This is part of the main circuit. The current is returned via the neutral wire
back to the substation and so to earth. In effect, all conventional electrical
circuits have functional earth. It is necessary for proper functioning of
electrical equipment and is not a safety feature. On older equipment, the
same symbol may have been used to denote protective earth.
Additional Protective Earth This equipment carries an additional protective earth. This protects against
electric shock in cases of a single fault condition.
Equipotentiality This is used to ensure that all metalwork is normally at or near zero voltage.
Therefore, under fault conditions, all the metalwork will increase to the same
potential. Simultaneous contact between two such metal appliances would
not cause a flow of current because they are both at the same potential,
therefore no shock results. This provides some protection against electric
shock by joining together all the metal appliances and connecting to earth.
Drip Proof, Splash Proof, Depending on the nature and use of the equipment, some are drip proof,
Water Tight splash proof or water proof.
246 14 Electrical safety
Anaesthetic-proof equipment AP equipment standards are based on the ignition energy required to ignite
the most flammable mixture of ether and air. They can be used within
525cm of gas escaping from a breathing system. The temperature should
not exceed 200C. It is a less stringent requirement.
Anaesthetic-proof equipment APG standards are based on the ignition energy required to ignite the
category G most flammable mixture of ether and oxygen. Can be used within 5cm of
gas escaping from a breathing system. The temperature should not exceed
90C. This is a more stringent requirement as the energy level should be less
than 1mJ.
Isolated or Floating Circuit This is a safety feature whereby current is not allowed to flow between the
electrical source and earth. These circuits are used to isolate individual
equipment. An isolating transformer is used with two coils insulated from
each other. The mains circuit is earthed whereas the patients circuit is not
earthed, so floating. As current flows through the mains coil (producing an
electromagnetic field), a current is induced in the patients coil. To complete
the patients circuit, the wires A and B should be connected. Contact with
wire A or B alone will not complete a circuit, even if the subject is earthed.
Current-Operated Earth These safety features are also known as an earth trip or residual circuit
Leakage Circuit Breakers breakers. They consist of a transformer with a core that has an equal number
(COELCB) of windings of a live and neutral wire around it. These are connected via a
third winding to the coil of a relay that operates the circuit breaker. Under
normal conditions, the magnetic fluxes cancel themselves out, as the current in
the live and neutral wires is the same. In the case of a fault (e.g. excessive
leakage current), the current in the live and neutral wires will be different so
resulting in a magnetic field. This induces a current in the third winding
causing the relay to break circuit. The COELCB are designed to be very
sensitive. A very small current is needed to trip the COELCB (e.g. 30mA) for a
very short period of time reducing the risk of electrocution.
Class 1 Power not to exceed maximum permissible exposure for the eye, or safe because of engineering design
Class 2 Visible laser beam only (400700nm), powers up to 1mW, eye protected by blink-reflex time of 0.25s
Class 2m As class 2, but not safe when viewed with visual aids such as eye loupes
Class 3a Relaxation of class 2 to 5mW for radiation in the visible spectrum (400700nm) provided the beam is
expanded so that the eye is still protected by the blink reflex
Maximum irradiance must not exceed 25W/m for intrabeam viewing
For other wavelengths, hazard is no greater than class 1
Class 3b Powers up to 0.5W
Direct viewing hazardous
Can be of any wavelength from 180nm to 1mm
Class 4 Powers over 0.5W
Any wavelength from 180nm to 1mm
Capable of igniting inflammable materials. Extremely hazardous
MCQs
In the following lists, which of the statements (a) to (e) are true?
Answers
metal appliances as they are
1. Concerning electric current: 2. Electrical impedance:
both at the same potential
a) False. Inductance occurs when a) True. Impedance is the sum
and no shock results.
a magnetic field is induced as of the forces that oppose the
b) False. Functional earth is not
electrons flow in a wire. The movement of electrons in an
a safety feature. It is necessary
ability to store a charge is AC circuit. In capacitors, the
for the proper functioning of
known as capacitance. In an impedance is low to high-
the device. It is part of the
inductor, the impedance is frequency current and vice
main circuit where the
proportional to the frequency versa. The opposite is correct
current, via the neutral wire,
of the current. In a capacitor, in inductors.
is returned to the substation
impedance is inversely b) False. Inductors have low
and so to earth.
proportional to the current impedance to low-frequency
c) False. Ohms law states that
frequency. current and vice versa.
the potential difference (volts)
b) True. The frequency of the c) True. Capacitors have low
= current (ampere)
mains supply in the UK is impedance to high-frequency
resistance (ohms).
50Hz. At this relatively low current and high impedance
d) True. Type CF equipment can
frequency, the danger of to low-frequency current. The
be used safely in direct
electric shock is high. latter is of most importance in
contact with the heart. The
c) True. A central venous protecting the patient from
leakage current is less than
pressure monitoring device low-frequency current.
50A in class I and less than
can be in direct contact with High-frequency currents have
10A in class II, providing a
the heart. Ventricular low tissue penetration without
high degree of protection
fibrillation can occur with exciting the contractile cells,
against electrical shock.
very small current, between allowing the current to pass
e) False. Type B equipment can
50 and 100mA, as the directly across the heart
be provided with defibrillator
current is applied directly to without causing ventricular
protection. The same applies
the myocardium (microshock). fibrillation.
to type BF and type CF
Such devices should have a d) True. The skin forms the
equipment.
leakage current of less than main impedance against the
10mA to prevent microshock. conduction of the ECG signal.
4. Electrical shock:
d) True. The amount of current In order to reduce the skin
a) False. Electric shock can
flow per unit of area is known impedance, there should be
happen with direct current
as the current density. This is good contact between the skin
although the amount of
important in the function of and the electrodes.
current required to cause
diathermy. At the tip of the e) True. Ohms are used to
ventricular fibrillation is much
diathermy forceps, the current measure both impedance and
higher than that of alternating
density is high so heat is electrical resistance. Ohm =
current.
generated. At the patient volt/ampere.
b) False. The main impedance
plate, the current density is
is in the skin and not the
low and no heat is generated. 3. Which of the following
muscles. Skin impedance is
e) False. In alternating current, statements are correct?
variable and can be from
the flow of electrons reverses a) True. Equipotentiality is a
100000 to 1000000
direction at regular intervals. safety feature when, under
depending on the area of
In the UK, the AC is 50 fault conditions, all
contact and whether or not
cycles per second (Hz). In metalwork increases to the
the skin is wet.
direct current, the flow of same potential. Current will
c) True. The severity of the
electrons is in one direction not flow during simultaneous
electric shock depends on the
only. contact between two such
Answers 253
frequency of the current. current density at the plate patients plate are not
The lower the frequency, the should be low. The same connected to earth.
higher the risk. A current of current is passed through the d) True. Diathermy can cause
50Hz is almost the most tip of the diathermy forceps electrical interference with
lethal frequency. where the current density is ECG and other monitoring
d) True. A current of 100mA, high, thus producing heat. devices. The use of electrical
when applied to the surface The current density at the filters can solve this.
of the body, can cause plate is low because of its e) False. Very-high-frequency
ventricular fibrillation. Most large surface area. current (in the radiofrequency
of the current is lost as the b) True. The isolating capacitor range) of 500000 to
current travels through the protects the patient from 1000000Hz is used. This
body and only 50100A are low-frequency current (50Hz) high-frequency current
required to cause ventricular shock because of its high behaves differently from the
fibrillation. impedance to low-frequency standard 50-Hz current;
e) True. The electrical energy is currents. It has low because of its low tissue
dissipated throughout the impedance to high-frequency penetration, it passes directly
tissues of the body leading to (diathermy) currents. through the heart without
a rise in temperature and c) True. A floating patient circuit causing ventricular fibrillation.
resulting in burns. can be used to reduce the
risk of burns. The diathermy
5. Diathermy: circuit is earth free. The
a) False. In order to protect patient, the tip of the
the patient from burns, the diathermy forceps and the
This page intentionally left blank
255
Appendices
256 Appendices
This guideline is not a standard of medical care. The ultimate judgement with regard to a particular clinical procedure or treatment plan must be
made by the clinician in the light of the clinical data presented and the diagnostic treatment options available
The Association of Anaesthetists of Great Britain and Ireland 2012
Checking anaesthetic equipment 257
Breathing system Whole system patent and leak free using two-bag test
Vapourizers fitted correctly, filled, leak free, plugged in (if necessary)
Alternative systems (Bain, T-piece) checked
Correct gas outlet selected
Ventilator Working and configured correctly
Airway equipment Full range required, working, with spares
Suction Clean and working
The two-bag test
A two-bag test should be performed after the breathing system, vapourizers and ventilator have been checked
individually:
a) Attach the patient end of the breathing system (including angle piece and filter) to a test lung or bag.
b) Set the fresh gas flow to 5L/min and ventilate manually. Check the whole breathing system is patent and unidirectional
valves are moving. Check the function of the APL valve by squeezing both bags.
c) Turn on the ventilator to ventilate the test lung. Turn off the fresh gas flow, or reduce to a minimum. Open and close
each vapourizer in turn. There should be no loss of volume in the system.
This checklist is an abbreviated version of the publication by the Association of Anaesthesia of Great Britain and Ireland Checking Anaesthesia
Equipment 2012 (endorsed by the Chief Medical Officers)
258 Appendices
Infusion devices order to achieve this. Appropriate attach all monitoring before
clinical observations may include induction. In these circumstances,
When any component of anaesthesia
mucosal colour, pupil size, response monitoring must be attached as
(hypnotic, analgesic, muscle
to surgical stimuli and movements soon as possible and the reasons for
relaxant) is administered by
of the chest wall and/or the delay recorded in the patients notes.
infusion, the infusion device unit
reservoir bag. The anaesthetist
must be checked before use. Alarm
should undertake palpation of
settings and infusion limits must be B: Recovery from
the pulse, auscultation of breath
verified and set to appropriate levels anaesthesia
sounds and, where appropriate,
before commencing anaesthesia. It is
measurement of urine output and A high standard of monitoring
essential to verify that these drugs
blood loss. A stethoscope must should be maintained until the
are delivered to the patient. The
always be available. patient is fully recovered from
infusion site should be secure and
anaesthesia. Clinical observations
preferably visible.
must be supplemented by the
following monitoring devices:
MONITORING DEVICES
Alarms
1. Pulse oximeter.
Anaesthetists must ensure that The following monitoring devices 2. Non-invasive blood pressure
all alarms are set at appropriate are essential to the safe conduct of monitor.
values. The default alarm settings anaesthesia. If it is necessary to
The following must also be
incorporated by the manufacturer continue anaesthesia without a
immediately available:
are often inappropriate and during particular device, the anaesthetist
the checking procedure the must clearly record the reasons for 1. Electrocardiograph.
anaesthetist must review and reset this in the anaesthetic record. 2. Nerve stimulator.
the upper and lower limits as The AAGBI recommends that 3. Means of measuring
necessary. Audible alarms must be any monitor providing continuous temperature.
enabled when anaesthesia values, such as SpO2 and ECG, 4. Capnograph.
commences. should only display a static non-
If the recovery area is not
When intermittent positive invasive blood pressure value for a
immediately adjacent to the
pressure ventilation is used during maximum of 5min, after which the
operating theatre, or if the patients
anaesthesia, airway pressure alarms value should blink or disappear
general condition is poor, adequate
must also be used to detect altogether. The value should remain
mobile monitoring of the above
excessive pressure within the airway stored.
parameters will be needed during
and also to give warning of
transfer. The anaesthetist is
disconnection or leaks. The upper
A: Induction and responsible for ensuring that this
and lower alarm limits must be
maintenance of anaesthesia transfer is accomplished safely.
reviewed and set appropriately
Facilities and staff needed for the
before anaesthesia commences. 1. Pulse oximeter.
recovery area are detailed in the
Provision, maintenance, 2. Non-invasive blood pressure
Association booklets, The
calibration and renewal of monitor.
Anaesthesia Team and Immediate
equipment are institutional 3. Electrocardiograph.
Post Anaesthetic Recovery.
responsibilities. 4. Airway gases: oxygen, carbon
dioxide and vapour.
5. Airway pressure. C: Additional monitoring
The following must also be Some patients will require
available: additional, mainly invasive,
Monitoring the monitoring, e.g. vascular or
1. A nerve stimulator whenever a
patient muscle relaxant is used.
intracranial pressures, cardiac
output or biochemical variables.
2. A means of measuring the
During anaesthesia, the patients Specific devices designed to monitor
patients temperature.
physiological state and depth of loss of consciousness using
anaesthesia need continual During induction of anaesthesia adaptations of either surface EEG
assessment. Monitoring devices in children and in unco-operative monitoring or auditory evoked
supplement clinical observation in adults it may not be possible to potentials have become available.
Recommendations for standards of monitoring during anaesthesia and recovery 261
However, their routine use has yet controlled operating theatre document Recommendations
to be fully considered as part of our environment and that personnel for standards of monitoring
recommended minimum monitoring with adequate knowledge and during anaesthesia and recovery.
standards. The American Society of experience accompany the patient. This document noted that
Anaesthesiologists (ASA) recently The patient should be continuous capnography must
published a report from a task force physiologically as stable as possible be used during induction and
set up to assess the use of brain on departure. Prior to transfer, maintenance of general anaesthesia.
function monitoring to prevent appropriate monitoring must be It also recommended that
intraoperative awareness. This commenced. Oxygen saturation continuous expired carbon
report summarized the state of the and arterial pressure should be dioxide monitoring be used for
literature and reported the opinions monitored in all patients and patients whose tracheas are
derived from task force members, an ECG must be attached. intubated and who are undergoing
expert consultants, open forums and Intravascular or intracranial transfer from one clinical area
public commentary. It concluded pressure monitoring may be to another.
that brain function monitoring is necessary in special cases. A Because of the recognized
not routinely indicated for patients monitored oxygen supply of known safety advantages of using
undergoing general anaesthesia, content sufficient to last the capnography in these situations, it
either to reduce the frequency of maximum duration of the transfer is difficult to justify not using it
intraoperative awareness or to is essential for all patients. If the when caring for patients in similar
monitor depth of anaesthesia. It patients lungs are ventilated, circumstances throughout the
was the consensus of the task force expired carbon dioxide should be hospital. Following serious
that the decision to use a brain monitored continuously. Airway untoward incidents arising from
function monitor should be made pressure, tidal volume and patients being treated outside the
on a case-by-case basis by the respiratory rate must also be operating theatre in intensive care
individual practitioner for selected monitored when the lungs are units, high-dependency units and
patients. The task force reported mechanically ventilated. recovery rooms, the AAGBI has
that patients have experienced decided to update its guidance on
intraoperative awareness in spite of the use of capnography. The AAGBI
monitored values which would now makes the following
imply an adequate depth of Anaesthesia outside recommendation.
anaesthesia. The AAGBI endorses Continuous capnography should
the views of the ASA taskforce. hospital be used in the following patients,
regardless of location within the
The Associations view is that the hospital:
D: Regional techniques and standards of monitoring used during
sedation for operative general and regional analgesia and 1. Those whose tracheas are
procedures sedation should be exactly the same intubated.
in all locations. 2. Those whose airways are
Patients must have appropriate
being maintained with
monitoring, including a minimum of
supraglottic or other similar
the following devices:
airway devices.
1. Pulse oximeter.
2. Non-invasive blood pressure
AAGBI safety Capnographs should be available
for use wherever it is possible that a
monitor. statement patients trachea will be intubated,
3. Electrocardiograph.
such as anaesthetic rooms, operating
CAPNOGRAPHY OUTSIDE
theatres, recovery rooms, other
THE OPERATING THEATRE
treatment rooms in which general
MONITORING DURING anaesthesia is given, intensive care
A statement from the
TRANSFER WITHIN units, high-dependency units and
Association of Anaesthetists
THE HOSPITAL accident and emergency
of Great Britain and Ireland
departments. It is also recommended
(AAGBI), January 2009
It is essential that the standard of that a capnograph be immediately
care and monitoring during transfer In 2007, the AAGBI published available during the treatment of
is as high as that applied in the the fourth edition of its guidance cardiac arrests in hospital.
262 Appendices
Appendix C
Serial number
Batch code
Graphical symbols
for use in labelling Can be autoclaved
medical devices
Do not reuse (single use only)
Sterile
Cannot be autoclaved
Catalogue number
Method of sterilization using
ethylene oxide
Expiry date
264 Appendices
MR
Do not use if package is opened MR unsafe. This means an item is known
or damaged. Do not use if the to pose hazards in all MRI environments
product sterilization barrier or its
packaging is compromised
Cleaning is the removal of visible Disinfection describes a process that Sterilization is the
soil (e.g. organic and inorganic eliminates many or all pathogenic complete elimination or
material) from objects and micro-organisms on inanimate objects destruction of all forms
surfaces with the exception of bacterial spores of microbial life
It is normally accomplished by It is usually accomplished by the use of This is accomplished in
manual or mechanical means liquid chemical and heat (washer healthcare facilities by
using water with detergents or disinfector) either physical or
enzymatic products chemical processes
Cleaning does much to reduce risk
of vCJD
Items which enter normally sterile Items that touch the mucous Items that touch only intact skin
tissue or the vascular system or membranes or skin that is not intact require low-level disinfection
through which blood flows They require a high-level disinfection As skin is an effective barrier to
They have a high risk of infection process; i.e. mycobactericidal micro-organisms, such items
These items should be sterile As intact mucous membranes are pose a low risk of infection
Examples: surgical instruments and generally resistant to infection, such Examples: bedpans and blood
needles items pose an intermediate risk pressure cuffs.
Such devices should ideally be
sterilized but chemical disinfection is
usually reserved for those that are
intolerant of heat sterilization
Example: laryngoscopes
266 Appendices
This classification has been 2. Chemicals: detergents used are 4. Time: cleaning cycle requires a
successfully used by infection a complex formulation of suitable time period to achieve
control professionals and others chemicals designed to remove its desired effect.
when planning methods for soil (proteins, carbohydrates,
Cleaning can be achieved either
disinfection or sterilization. lipids, etc.) from instruments.
by:
The ease of inactivation differs Detergents have an optimal
according to the micro-organisms concentration and pH to work
Manual cleaning
involved (see Fig. AppD.1). effectively.
3. Energy: may take the form 1. Immersion in a diluted detergent
of manual washing, ultrasonic at 35C:
Cleaning
energy or water jets/sprays Non-immersion techniques
Involves physical removal of the in automated washer involve a cloth soaked in
infectious material or organic matter disinfectors. cleaning solution and used
on which micro-organisms thrive.
The critical parameters for cleaning
are the following:
1. Temperature: initial
wash temperatures must
be below 45C to prevent
coagulation of tissue/blood
residues.
Fig. AppD.1 Descending order of Fig. AppD.3 Bowie Dick test. This is a daily test to ensure that the steam sterilizer is
resistance of micro-organisms against functioning appropriately. Paper patches with heat-sensitive inks are used. The blue
inactivation. patch (left) is a PASS, the yellow patch (right) is FAILED.
Fig. AppD.2 Heat- or chemical-sensitive inks that change colour. Right: presterilization. Left: poststerilization with the sterilization
trace label. Note the date of sterilization that is valid for 1 year and details to track any object.
Decontamination of medical equipment 267
to wipe the items. This can glutaraldehyde 2% for 20min, to sterilization. Usually a
be used for electrical equipment. hydrogen peroxide 67.5% for temperature of 134C
2. Mechanical cleaning: uses 2030min, peracetic acid maintained for a period of 3min
thermal disinfection, chemical 0.20.35% for 5min and under a pressure of 2.25bars is
disinfection (see later) or orthophthalaldehyde (OPA) for used.
ultrasonic cleaners. 512min. 2. Ionizing radiation using rays
Ultrasonic cleaning is used in 3. Pasteurization (heat to produce sterility. It is ideal for
areas that are difficult to access. disinfection): heating to 60 prepacked heat-labile single-use
The ultrasonic waves create 100C for approximately 30min items such as IV cannulae and
small bubbles on the surfaces of to reduce the number of syringes. This technique of
the instruments. These bubbles pathogens by killing a significant sterilization is widely used in
expand until they cavitate and number of them. The higher the industry.
collapse, producing areas of temperature, the shorter the time 3. Dry heat sterilization (hot air
vacuum that dislodge the needed. oven): a constant supply of
contaminants. electricity is needed. Used
for reusable glass, metal
instruments, oil, ointments and
Sterilization
powders.
Disinfection
The complete destruction of all 4. Ethylene oxide can effectively
Involves reduction of micro- micro-organisms. Sterility is the sterilize most equipment that
organisms on devices. probability of complete sterilization. can withstand temperatures
This probability is known as the of 5060C. However, it is
1. Thermal washer disinfectors
sterility assurance level (SAL). A used under carefully controlled
combine cleaning and
sterile device has a SAL of 106, conditions because it is
disinfection. Powerful water
which means that the probability of extremely toxic, carcinogenic,
and detergent jets heated to
an organism surviving on that flammable and an explosion
about 80C are used. Most
device is one in a million using a risk. Although it is very
organisms are inactivated
validated process. versatile and can be used
except for bacterial spores,
The methods used to achieve for heat-sensitive equipment,
some heat-resistant viruses
sterility include the following: fluids and rubber, a long
and cryptosporidia.
period of aeration is necessary
2. Chemical disinfection is the 1. Steam sterilization is currently
to remove all traces of gas
destruction of micro-organisms the gold standard method (Table
before the equipment can be
by chemical or physiochemical AppD.3). It is reliable, easy to
distributed. The processing
means. This process is difficult monitor, non-toxic, inexpensive,
time ranges from 2 to 24h
to control and validate. It is sporicidal and has high lethality,
and is a very costly
frequently used for devices that rapid heating and good
process.
are heat sensitive in the penetration of fabrics. The
semicritical category such as temperature and pressure
endoscopes. Examples are reached determine the time
Sterilization monitoring
The sterilization process
should demonstrate a spore kill
Table AppD.3 Steam sterilization
to achieve a SAL of 1 106. To
ensure that sterilization has been
Advantages Disadvantages successful, indicators are used
(Table AppD.4). For steam
Highly effective Items must be heat and moisture resistant
sterilization, for example, this
Rapid heating and penetration Does not sterilize powders, ointments or oils
of instruments Needs good maintenance requires the direct contact of
Non-toxic saturated steam with the device in
Inexpensive question in the absence of air at
Can be used to sterilize liquids the required pressure/temperature
and time.
268 Appendices
These indicators are part of There are different chemical indicators These are rarely used in UK hospitals
the steam sterilizer itself There are tapes with heat- or Bacillus spores that are heat sensitive
They record and allow the chemical-sensitive inks that change can directly measure sterilization
observation of time, colour when the intended They are inherently unreliable but can
temperature and pressure temperature, time and pressure are be used as an additional method
readings during the reached (Fig. AppD.2) of validation for some forms of
sterilization cycle Such a tape does not assure sterility sterilization such as ethylene oxide
It merely states that the pack has
been through a heating process
Bowie Dick test uses heat-sensitive
inks to ensure that the steam
sterilizer is functioning appropriately
(Fig. AppD.3)
SINGLE-USE ITEMS rather than the quantity or source MHRA, 2010. Sterilization, disinfection
of the chlorine entering the and cleaning of medical equipment:
The use of single-use items should incinerator. It must be emphasized guidance on decontamination
be encouraged when possible. This that any equipment that is from the Microbiology Advisory
practice ensures the sterility of the designated single-use must be used Committee (the MAC manual).
equipment and prevents cross- for one patient only and for a single Online. Available at: http://www.
infection. The quality of such treatment episode, and not reused mhra.gov.uk/Publications/
devices must be the same as the even for the same patient during Safetyguidance/
reusable ones. As a large proportion subsequent visits. Otherdevicesafetyguidance/
of these devices are made from PVC CON007438
plastic materials, a balance should TSE Working Group, 2008. Transmissible
FURTHER READING
be struck between the reduction in spongiform encephalopathy agents:
infection risk and effect on the Association of Anaesthetists in Great safe working and the prevention
environment. Incinerating PVC has Britain and Ireland, 2008. Infection of infection. Guidance from the
no or very small effect on the levels control in anaesthesia. AAGBI, Advisory Committee on Dangerous
of dioxin produced. The operating London. Online. Available at: http:// Pathogens. Online. Available at:
conditions of an incineration plant www.aagbi.org/sites/default/files/ http://www.advisorybodies.doh.gov.
are the key factor in determining infection_control_08.pdf uk/acdp/tseguidance/Index.htm
dioxin production and emissions,
Latex allergy 269
Unit 3
COOK MEDICAL
Chiltern Court
Asheridge Rd Industrial Estate OHalloran Road
Asheridge Road National Technology Park
Chesham Limerick
Buckinghamshire HP5 2PX Ireland
UK Tel: 00353 61 334440
Tel: 01494 784422 www.cookmedical.com
www.spacelabshealthcare.com
272 Appendices
Glossary
276 Glossary
Absolute pressure This is the total pressure exerted on a system; i.e. the FRC The functional residual
gauge pressure plus atmospheric pressure. capacity. It is the sum of the
expiratory reserve volume and the
Absolute pressure = gauge pressure + atmospheric pressure.
residual volume. In an adult male, it
is normally about 2.53 litres.
Absolute temperature This is the temperature measured in relation to the
absolute zero using the Kelvin temperature scale with the absolute zero, or
0K, that corresponds to 273.15C. Gas laws
Daltons law of partial pressures:
Absolute zero The temperature at which molecular energy is minimum and in a mixture of gases, each gas
below which temperatures do not exist. exerts the same pressure which it
would if it alone occupied the
Cardiac index The cardiac output divided by the body surface area. container.
Normally, it is about 3.2L/min/m2. Boyles law: at a constant
temperature, the volume of a given
Dead space That volume of inspired air that does not take part in gas mass of gas varies inversely with the
exchange. It is divided into: absolute pressure.
1. Anatomical dead space: that part of the patients respiratory tract into 1
Volume = constant
which fresh gases enter without undergoing gas exchange. Gases are pressure
warmed and humidified in the anatomical dead space.
2. Alveolar dead space: that part of the lungs where perfusion is impaired Charles law: at a constant pressure,
resulting in ventilation/perfusion mismatch. the volume of a given mass of gas
varies directly with the absolute
End-tidal gas concentration This is an estimation of the alveolar gas temperature.
composition. In cases of severe ventilation/perfusion mismatch, it is an Volume = constant temperature
inaccurate estimation of the alveolar gas composition.
Gay-Lussacs law: at a constant
FGF Fresh gas flow from the anaesthetic machine or other source supplied to volume, the absolute pressure of a
the patient. given mass of gas varies directly
with the absolute temperature.
Flow The amount of fluid (gas or liquid) moving per unit of time.
Pressure = constant temperature
1. Laminar flow: flow (through a smooth tube with no sharp edges or
bends) is even with no eddies. Laminar flow can be described by the Humidity
HagenPoiseuille equation.
1. Absolute humidity: the mass of
Pr 4 water vapour present in a given
Q=
8 l volume of air. The unit is mg/L.
2. Relative humidity: the ratio of
where:
the mass of water vapour in a
Q = flow
given volume of air to that
P = pressure across tube
required to saturate the same
r = radius of tube
volume at the same temperature.
= viscosity
The unit is %.
l = length of tube
2. Turbulent flow: flow through a tube with a constriction (an orifice) is
I/E ratio The ratio of the length
uneven and eddies occur.
of inspiration to the length of
In this situation, flow (Q) is:
expiration, including the expiratory
proportional to the square of the radius of the tube r2
pause. The commonly used ratio is
proportional to the square root of the pressure gradient (P) P 1:2.
inversely proportional to the length of the tube (l) 1
1
inversely proportional to the density of the fluid () 1 Implantation testing (IT) In order
to ensure that tracheal tubes are
safe for use in the human body, tube
material is cut into strips and
Glossary 277
inserted usually into rabbit muscle. where compressed gases are driven under pressure over or through a liquid
After a period of time the effect anaesthetic.
of the implant on the tissue is
compared to controls. Pulmonary vascular resistance (PVR) The resistance against which the
right heart pumps. The unit is dyne/s/cm5. Normally, PVR = 80120dyne/s/
IPPV Intermittent positive pressure cm5.
ventilation includes both controlled
and assisted ventilation. The Pulse pressure The difference between the arterial systolic and diastolic
pressure within the lung increases pressures.
during inspiration (e.g. 15
20cm H2O) and decreases during SI units
expiration to atmospheric pressure.
Length Metre, m
This reverses the pressures found
Mass Kilogram, kg
during spontaneous ventilation. Time Second, s
Electric current Ampere, A
Latent heat of vaporization The Thermodynamic temperature Kelvin, K
energy needed to change a substance Luminous intensity Candela, cd
from liquid to gas, without Amount of substance Mole, mol
changing its temperature. Other units are derived, e.g.:
pressure (force + area) Pascal, P
Force Newton, N
Minute volume The sum of the Volume cubic metre, m3
tidal volumes in 1min. It is the
tidal volume respiratory rate.
SIMV Synchronized intermittent mandatory ventilation where the patient is
allowed to breathe spontaneously between the preset mechanical breaths
Oscilloscope A device capable
delivered by the ventilator. The ventilator synchronizes the patients
of displaying recorded electrical
spontaneous breaths if they coincide.
signals. It is particularly useful in
displaying high-frequency signals
Stroke volume The amount of blood expelled from a ventricle at each beat.
and allowing analysis of their
In an adult it is about 70125mL.
shapes.
cardiac output
Stroke volume =
Partial pressure The pressure heart rate
exerted by each gas in a gas
mixture. Systemic vascular resistance (SVR) The resistance against which the left
heart pumps. The unit is dyne/s/cm5. Normally, SVR = 10001500dyne/s/
PCWP Pulmonary capillary wedge cm5.
pressure is a reflection of the
pressure in the left atrium. It is Tidal volume This is the volume of a single breath. It is about 10mL/kg
obtained by using a balloon-tipped body weight.
flow-guided pulmonary artery
catheter. Transducer A device which changes one form of energy to another. An
example is the pressure transducer used to measure pressures in the body.
PEEP Positive end-expiratory Mechanical energy is converted to electrical energy.
pressure where the pressure during
expiration is prevented from Venturi principle A constriction in a tube causing an area of low pressure
reaching atmospheric pressure leading to the entrainment of a fluid (gas or liquid) via a side arm.
(zero). It prevents the closure of mean pulmonary artery pressure left atrial pressure
airways during expiration thus PVR = 80 (correction factor)
cardiac output
improving oxygenation.
mean arterial pressure right atrial pressure
Plenum A chamber where the SVR = 80 (correction factor)
cardiac output
pressure inside is greater than the
pressure outside. Most modern
vaporizers are plenum vaporizers
This page intentionally left blank
279
Index
280 Index
Page numbers followed by f indicate anaesthetic machine (Continued) balloon-tipped flow-guided pulmonary
figures, t indicate tables, and b indicate pressure gauges, 2022, 21f, 39 artery catheter, 187190, 188f,
boxes. pressure regulators, 22f23f, 23, 23b, 190b, 196
39 barium hydroxide, 65
A in remote areas, 37, 37f barylime, 65
safety, 20b, 39 B Braun Contiplex S Ultra, 197
Abbott cardiac output computer, 189f second-stage regulators, 23 BD Flowswitch arterial cannula, 178f
Abbott Oximetrix pulmonary artery vaporizer filling devices, 3031, 31f BD Nexiva IV closed and integrated
catheter, 190f vaporizers, 2730, 39 cannula, 229f
ABL800FLEX blood gas analyzer, 223f anaesthetic-proof equipment, 246 Beers law, 152
active scavenging systems, 4849, category G, 246 Bermann airway, 89, 89f
48f50f anaesthetists presence during Bernoulli principle, 105106
adenovirus, 140b anaesthesia, 259 bilevel positive airway pressure (BiPAP),
adjustable flange tracheostomy tube, 82f analgesia 220
adjustable pressure limiting (APL) valve, elastomeric pumps, 204205, 204f bimetallic strip, plenum vaporizer,
5657, 57b, 57f, 63, 65, patient controlled, 202, 202f 2728, 29f, 30
6970 anti-hypoxic safety features, 33b biphasic defibrillators, 232233
inhalers, 100 antipollution cap, 31 bipolar diathermy, 248
ventilators, 123 antistatic materials, 26, 247b bispectral index (BIS) analysis, 168169,
Aintree intubation catheter, 117, 117f aortic velocimetry, 187 168f169f, 169b, 172
air APL (adjustable pressure limiting) valve bite guards, 115117, 116f117f
compressed see compressed air breathing systems, 5657, 57b, 57f, bladder temperature, 194, 195f
cylinders, 4t 63, 65, 6970 Blease Manley MP3 ventilator, 123124,
cylinder valves, 5f inhalers, 100 123f, 124b, 133
air conditioning, 46 ventilators, 123 blood gases analyser, 223224, 223f,
airway exchange catheters, 117, 117f Aquinox hot water bath humidifier, 237
airway laser surgery, 249 137f blood giving sets, 228229, 228f
airways area valve service units (AVSUs), 7, 8f blood pressure monitoring
nasopharyngeal airway, 8990, 90b, argon accumulation, 11 invasive, 178181, 178f181f, 181b,
90f Argon cannula, 184f 196197
Nosworthy airway, 115117, 116f Argon Careflow arterial cannula, 178f non-invasive, 149151
oropharyngeal airway, 8889, armoured tracheal tube, 79, 79f blood sampling, catheters for, 184
88f89f, 89b Arndt endobronchial blocker, 88f blood warmers, 229230, 230f
alarms arterial blood gases analyser, 223224, Bodok seal, 6, 6f
monitoring, 260 223f, 237 body core temperature, 194
oxygen supply failure, 32, 32f33f arterial blood oxygen saturation bougie bite guards, 115117, 116f117f
ventilator, 165, 165b, 165f measurement, 151 Bourdon pressure gauge, 21f
allergy, latex, 269270 arterial blood pressure monitoring Bowie Dick test, 266f
alternating current, 242 invasive, 178181, 178f181f, 181b, Boyles gas law, 2122
Ambu laryngeal mask, 9293 196197 breathing systems, 5573, 106
Ambu PEEP valve, 131132, 131f MCQs on, 171, 196 adjustable pressure limiting (APL)
Ambu valve, 131, 131f non-invasive, 149151, 150f151f, valve, 5657, 57b, 57f
anaesthetic agents, inhaled, scavenging 150t, 151b Bain system, 56, 56f, 6062, 61f,
systems see Scavenging arterial pressure waveform, 179180, 62b, 69
anaesthetic equipment 180f checklist for, 256t
checking, 256, 256t arterial waveform analysis, 187 components of, 5658
monitoring, 259260 Association of Anaesthetists of Great fresh gas flow, 56
anaesthetic machine, 1943, 20f21f Britain and Ireland (AAGBI) Humphrey ADE system, 63, 63f64f,
backflow check valves, 23 checking anaesthetic equipment, 256, 64b
common gas outlet, 33, 33f 256t ideal, properties of, 56
compressed oxygen outlet(s), 32, 32f safety statement, 261262 Lack system, 59, 60b, 60f
emergency oxygen flush, 3132, 31f standards of monitoring, 258 Magill system, 5859, 59b, 59f
flow control (needle) valves, 2324, AVSUs (area valve service units), 7, 8f Mapleson A system, 56, 56f, 5859,
24f axillary block, 211 59b60b, 59f60f
flowmeters, 2426, 39 axillary temperature, 194 Mapleson B system, 56, 56f, 60, 60b
flow restrictors, 23, 23f Mapleson classification, 56, 56f
laminar flow, 3435, 39 B Mapleson C system, 56, 56f, 60, 60b
modifications and new designs, Mapleson D system, 56, 56f, 6062,
3336 backflow check valves, 23 61f, 62b
non-return pressure relief safety bacterial and viral filters, 140141, Mapleson E system, 56, 56f, 6263,
valve, 31, 31f, 40 140b, 142b, 143 62f, 63b
one-way valves, 23 bag in a bottle ventilator, 125, 125b, Mapleson F system, 6263, 62f, 63b
oxygen emergency flush, 3132, 31f, 125f, 133 MCQs on, 69
40 bag squeezers, 122 Monitoring, 259
oxygen supply failure alarm, 32, Bain breathing system, 56, 56f, 6062, objectives, 56
32f33f 61f, 62b, 69 reservoir bag, 5758, 58b, 58f
Index 281
breathing systems (Continued) cerebral ischaemia, BIS and, 169 cylinder valves (Continued)
soda lime and circle system, 6467, CE symbol, 246 problems in practice, 6
64f, 66f, 68b charcoal canisters, 4950, 50b, 51f safety, 56
T-piece system, 6263, 62f, 63b, 69 chemotherapy, catheters for, 184 types of, 5f6f
tubings, 58 chest drains, 234235, 234f, 235b, 237 cylinder yokes, 4f, 6f
Waters canister system, 68, 68f children see paediatric practice cytomegalovirus, 140b
Bronchocath double lumen tube, 88f circle breathing system, 46
bronchoscopy, ventilation during, 130 see also soda lime and circle D
breathing system
C Clark electrode oxygen analyzer, 159 damping, invasive arterial pressure
Classic LMA laryngeal mask, 9293 monitoring, 180
CADD Legacy portable PCA, 202f class I/II/III electrical equipment, Datex-Ohmeda Aestiva S/5 anaesthetic
calcium hydroxide, 65 243244 machine, 20f
capacitance, 243 cleaning of equipment, 265268, 265t Datex-Ohmeda Capnomac Ultima, 160f
capacitors, 242243 closed circle anaesthesia, 6465 Datex-Ohmeda compact monitor, 146f
capnograms, 153, 261262 coagulation diathermy, 248 Datex-Ohmeda Flexima anaesthetic
capnographs, 153157, 154f155f, 171 coaxial breathing systems, 5960, 60f, machine, 22f
capnometers, 153 63 Datex-Ohmeda 7900 ventilator, 125f
carbon dioxide coaxial fluid/blood warmer devices, dead space
absorption by soda lime, 6566 229230 breathing systems, 59, 6162, 6768
analyzers, 153157, 154f156f, 158b Cobb suction tracheal tube connector, MCQs on, 69
cylinders, 2f, 4t 79f decontamination of equipment, 265268,
infrared absorption spectrum, 157, COBRA-PLA, 9294, 94f 265t, 266f
157f common gas outlet, 33, 33f defibrillators, 230233, 231f233f,
monitoring, variable performance common mode rejection ratio (CMRR), 234b, 237
masks, 103, 103f 148 dental anaesthesia, 36, 100101
rebreathing in variable performance compound A, 67 dermatitis, latex allergy, 269270
masks, 103 compressed air, 13, 13f desflurane
carbon dioxide electrode, 225226, compressed oxygen outlet(s), 32, 32f concentration analysis, 160, 161f
226b, 226f, 237 contact dermatitis, latex allergy, description of, 3334
carbon monoxide production, encircle 269270 infrared absorption spectrum, 161f
breathing systems, 67 continuous low flow therapy, pollution by, 46
cardiac output monitoring, 187, haemofiltration, 221 vaporizer filling device colour code,
189192, 189f continuous positive airway pressure 31, 31f
cardiac pacemakers, diathermy and, 248 (CPAP), 127, 220, 220f, 221b, vaporizers, 3336, 34f, 40
Cardiff Aldasorber, 4950, 50b, 51f 237 device for indirect non-invasive
CardioQ oesophageal Doppler machine, Cook retrograde intubation set, 117f automatic mean arterial
191f CPAP (continuous positive airway pressure (DINAMAP), 149
Carlens tubes, 87 pressure), 127, 220, 220f, dialysis catheters, 184186
catheter mounts, 101, 101b, 101f, 108 221b, 237 diathermy, 248249, 249b, 251
catheters Crawford spinal needle, 208 dicrotic notch, arterial pressure
central venous catheterization and cricoid cartilage, 77 waveform, 179, 188189
pressure, 183186, 184f185f cricothyrotomy tube, 86, 86b, 86f diffusional interception, filtration, 140
pulmonary artery see pulmonary critical temperature, 2 DINAMAP (device for indirect non-
artery catheters current density, 202f, 247 invasive automatic mean
see also specific types current-operated earth leakage circuit arterial pressure), 149
catheter whip, 190 breakers (COELCB), 246 direct current (DC), 242
cauda equina syndrome, 210 cutting diathermy, 248 direct interception, filtration, 140
centralized vacuum suction system, 1314 cylinder manifold, 89, 9f disinfection, 265268, 265t
components, 13, 14f cylinders disposable circle breathing systems, 66
mechanism of action, 1314 colour-coding, 34, 4t disposable laryngeal masks, 9293
problems in practice, 14 components, 23, 2f disposable tracheal tubes, 76
safety, 14 engraved marks on, 3, 5f Doppler haemodynamic measurement,
central venous catheterization and filling ratio, 3 oesophageal, 190192,
pressure (CVP), 181186, labelling, 3b 191b192b, 191f192f, 197
182f183f, 186b lightweight, 2 double burst stimulation, neuromuscular
catheters, 183186, 184f185f MCQs on, 15 monitoring, 167
fluid manometer, 182183, 183f safety, 34, 4f, 4t double lumen endobronchial tubes,
long central catheters/peripherally sizes, 2, 8 8687, 87f88f, 88b
inserted central catheters, storage, 4 Drger D Vapor desflurane vaporizer, 34f
183184, 185f186f testing, 3 Drger Zeus IE anaesthetic workstation,
MCQs on, 196 cylinder valves, 56 3435, 35f
pressure transducer, 182 Bodok seal, 6, 6f draw-over vaporizers, 37
problems in practice and safety, components, 5, 6f drip proof, 245
184186 mechanism of action, 56, 6f dry heat sterilization, 267
sepsis/infection reduction, 185b pin-index system, 5, 5f dural headache, 208210, 216
282 Index
oxygen emergency flush, 3132, 31f PICC (peripherally inserted central post-tetanic facilitation/potentiation,
oxygen/nitrous oxide see Entonox catheters), 183184, 185f186f neuromuscular monitoring,
oxygen supply monitoring, 259260 piezoelectric quartz crystal oscillation, 167
161, 162t potassium hydroxide, 65
P pin-index system, cylinder valves, 5, 5f potential difference, 242
piped gas supply (piped medical gas and Poynting effect, 12
paediatric practice vacuum (PMGV)), 68 pressure control ventilation, 126, 133
bispectral index analysis, 168 colour-coding, 7, 8f pressure cycled ventilators, 122
breathing systems, 58, 6263, 62f, components, 7 pressure gauges, 2022, 21f
63b gas probes, 7f components, 2021
COBRA-PLA, 94 isolating valves, 7, 8f MCQs on, 39
epidural needles, 205 MCQs on, 15 mechanism of action, 21
face masks, 100f outlets, 7, 7f problems in practice, 2122
intravenous giving sets, 228229 problems in practice, 78 safety, 2122
multilumen catheters, 183, 185f safety, 78, 8f ventilators, 123
oropharyngeal airway in, 89 Pitot tube, pneumotachograph combined pressure generator ventilators, 122, 123t
tracheal tubes, 77 with, 163164, 164f pressure monitoring alarm, 165, 165f
variable performance face masks, pleated hydrophobic filters, 141, 141f pressure-regulated volume
102f plenum vaporizers, 2730 control(PRVC), 126
ventilators, 124125 components, 2728, 28f29f pressure regulators, 2223, 22f23f, 23b
Venturi masks, 106, 106f mechanism of action, 2829 MCQs on, 39
Wright respirometer, 163 problems in practice, 2930 second-stage, 23
pain relief see analgesia safety, 2930, 30f pressure support ventilation, 127
PA (pulmonary artery) catheters see PMVG see piped gas supply (piped pressure swing adsorption systems,
pulmonary artery catheters medical gas and vacuum 1011
paramagnetic oxygen concentration (PMGV)) pressure transducer, central venous
analyzers, 158159, 159f, pneumotachograph, 163, 163f, 164b catheterization and pressure,
162t, 172 combined with Pitot tube, 163164, 182
ParaPAC ventilator, 128 164f Prismaflex 1 haemofiltration system,
parenteral nutrition, catheters for, 184 MCQs on, 171 221f
partial rebreathing masks, 103 Pneupac VR1 emergency ventilator, protective earth, 245
passive scavenging systems, 4748, 129130, 129f, 130b additional, 245
47f48f, 52 Poiseuilles law, 24 psoas compartment block, 211
pasteurization, 267 polarographic (Clarke) oxygen electrode, pulmonary artery catheters, 187,
patient controlled analgesia (PCA), 202, 224, 224b, 224f 189f190f
202f, 203b polarographic oxygen concentration balloon-tipped flow-guided, 187190,
syringe pumps, 203, 203f analyzers, 158159, 162t, 171 188f, 190b, 196
target controlled infusion pumps, 204 polar tracheal tubes, 7980, 80f indications, 188b
volumetric pumps, 203, 203f polio laryngoscope blade, 112f, 113114 information gained from, 190
patient monitoring, 260261 pollution in theatre, 4647, 47b, 52 MCQs on, 196
see also invasive monitoring; causes of, 46t pulse oximetry, 151153, 151f152f,
non-invasive monitoring MCQs on, 52 153b, 153t
PCA see patient controlled analgesia Portex Correct Inject spinal needles, MCQs on, 171
(PCA) 210f
peak inspiratory flow, 5758 Portex cricothyrotomy set, 86f Q
PEEP valve, 131132, 131f Portex CSE cure combined spinal-
Penlon Nuffield 200 ventilator, 124125, epidural device, 206f Quantiflex anaesthetic machine, 3536,
124f, 125b Portex epidural catheter and filter, 207f 35f
with Bain breathing system, 61 Portex epidural set, 207f quartz crystal oscillation, piezoelectric,
with T-piece breathing system, 62 Portex Griggs percutaneous 161, 162t
Penlon PM9000 Express, 155f tracheostomy set, 84f Quincke needles, 208, 209f
Penlon pressure monitoring ventilator Portex LockIt Plus epidural catheter
alarm, 165f fixing device, 207f R
percutaneous localization of nerves, 212, Portex Silicone LM laryngeal mask,
212f 9293 radial artery, 179
percutaneous tracheostomy tubes, Portex single-use Soft-Seal laryngeal RAE tracheal tubes, 7980, 80f
8485, 84f85f mask, 91f Raman spectroscopy, 161, 162t
peripherally inserted central catheters Portex SoftSeal laryngeal mask, 9193, recovery from anaesthesia, 260
(PICC), 183184, 185f186f 91f rectal temperature, 194, 194f
peripheral nerve stimulators, 165168, Portex spinal microcatheter set, 209f reducing valves, 2223, 22f23f, 23b
166f, 168b Portex Tracer III peripheral nerve regional anaesthesia, ultrasound
neuromuscular monitoring, 166168, stimulator, 213f guidance in, 214, 214f
167f Portex tracheostomy tube, 82f relative humidity, 136, 143, 249
peripheral vascular resistance, 179 Portex Ultraperc single dilator remote areas, anaesthesia in, 37
pH electrode, 224225, 225b, 225f percutaneous tracheostomy set, Rendell-Baker face mask, 100, 100f
photo-acoustic spectroscopy, 155 85f reservoir bag, 5758, 58b, 58f
286 Index
residual circuit breakers, 246 safety (Continued) Severinghaus electrode, 225226, 226b,
resistive coupling, 248 Mapleson A breathing system, 59 226f
resistors, 242243 Mapleson D breathing system, sevoflurane
resonance, invasive arterial pressure 6162 concentration analysis, 160, 161f
monitoring, 180 Mapleson E breathing system, 63 pollution by, 46
respirometers, 162163, 162f163f, Mapleson F breathing system, 63 vaporizer filling device colour code,
163b, 171 minitracheostomy, 86 31, 31f
response entropy, 170 nasal cannulae, 104 Seward blades, 112f
retrograde intubation set, 117, 117f, nasopharyngeal airway, 90 Shannon entropy, 170
118b, 119 nerve stimulators, 213214 side-stream gas analyzer, 155156, 156f,
Reynolds number, 42 oesophageal Doppler haemodynamic 157t
Ring, Adair and Elwyn (RAE) tracheal measurement, 191192 Silicone LM laryngeal mask, 9293
tube, 7980, 80f oropharyngeal airways, 89 single-fault condition, 243
Robertshaw (rubber) tubes, 87 oscillometry, 150151 single hose test, 7
Robertshaws blades, 112f oxygen concentration analyzers, skin impedance, 146147, 247
RS7 peripheral nerve stimulator, 166f 159 skin temperature, 194, 195f
oxygen electrode, 224 Smiths gas-driven nebulizers, 139f
S patient controlled analgesia, 203 Smiths Medex pressure transducer, 179f
peripheral nerve stimulators, Smiths Medex silver impregnated triple
safety 167168 lumen central venous catheter,
adjustable pressure limiting valve, 57 pH electrode, 225 184f
anaesthetic machine, 20b, 37, 39 piped gas supply (piped medical gas Smiths Medical bladder catheter, 195f
anti-hypoxic safety features, 33b and vacuum (PMGV)), 78, 8f Smiths Medical FingerPrint oximeter,
bispectral index analysis, 169 plenum vaporizers, 2930, 30f 151f
breathing systems, 5759, 6163 pneumotachograph, 163164 Smiths Medical insulated peripheral
carbon dioxide electrode, 226 pressure gauges, 2122 nerve block needles, 210f
catheter mounts, 101 pressure regulators, 23 Smiths Medical Level 1 H 1200 fast
centralized vacuum suction system, pulmonary artery catheters, 190 fluid warmer, 230f
14 pulse oximetry, 153 Smiths Medical Protective Acuvance
central venous catheterization and reservoir bags, 58 cannula, 229f
pressure, 184186 retrograde intubation set, 117 Smiths Portex cricothyrotomy set, 86f
chest drains, 234235 scavenging systems, 4749 Smiths Portex Griggs percutaneous
continuous positive airway pressure, soda lime and circle breathing system, tracheostomy set, 84f
220 67 Smiths Portex LockIt Plus epidural
cricothyrotomy tube, 86 spinal needles, 208 catheter fixing device, 207f
cylinders, 34, 4f, 4t temperature probes, 193 Smiths Portex single-use Soft-Seal
cylinder valves, 56 tracheal tubes, 79 laryngeal mask, 91f
defibrillators, 233 tracheostomy tracheal tubes, 83 Smiths Portex tracheostomy tube, 82f
diathermy, 248249 ultrasound machine, 236 Smiths Portex Ultraperc single dilator
elastomeric pumps, 204205 variable performance masks, 103b percutaneous tracheostomy set,
electrical, 241253 ventilator alarms, 165 85f
electrocardiogram, 148149 ventilators, 124125, 127130 soda lime, 65, 67, 69
emergency oxygen flush, 32 Venturi masks, 106 humidifying properties, 136
end-tidal carbon dioxide analyzers, Wright respirometer, 163 soda lime and circle breathing system,
157 saturated vapour pressure (SVP), 29 6467, 64f, 68b, 70
Entonox, 13 scavenging, 4749, 49b components, 6566
epidural catheters, 206207 active systems, 4849, 48f50f diagrammatic representation of, 66f
epidural needles, 205206 checklist for, 256t mechanism of action, 6566, 66f
face masks, 100 ideal system, 47b safety and problems in practice, 67
fibreoptic intubating laryngoscopes, passive system, 4748, 47f48f, 52 use of vaporizers in, 6667
115 sciatic nerve block, 211 sodium hydroxide, 65
flowmeters, 2526, 27f sedation, monitoring of, 168170, 261 Soper blades, 112f
haemofiltration, 223 Seldinger chest drainage kit, 234f Spaulding classification, 265t
humidifiers, 137138 Seldinger technique spinal-epidural needle, 205, 206f
i-gel airway, 94 central venous catheterization and spinal microcatheters, 209f, 210
inhalational agent concentration pressure, 181182, 182f183f spinal needles, 208210, 209f210f,
analyzers, 161 tracheostomy tubes, 85 210b, 216
intra-aortic balloon pump, 227228 Selectatec vaporizer mount system, spinning disc nebulizers, 139
intubating laryngeal mask airway, 93 2829, 29f30f splash proof, 245
invasive arterial pressure monitoring, self-inflating bag and mask, 130131, Sprotte needles, 208209
180181 131b, 131f, 256t square wave test, 180, 180f
laryngeal masks, 93b sepsis reduction, central venous Staphylococcus aureus, 140b
laryngoscopes, 113114 catheterization and pressure, state entropy, 170
lasers, 249 185b static electricity, 26, 249
LiDCOrapid, 193 Servo-i ventilator, 126127, 126f, 127b steam sterilization, 267, 267t
Index 287
X Y Z
XD11XE Philips ultrasound machine, Yale needles, 208 zeolites, 10
235f Zoll Pro AED, 231f
Zoll R manual defibrillator, 231f