Beruflich Dokumente
Kultur Dokumente
MP-2000
INDEX
INTRODUCTION 3
INSTALLATION 7
OPERATING INSTRUCTIONS 9
FAULT CONDITION 16
IMPURE GAS 16
2 2
3
Introduction .
The MP-2000 Rare Gas Purifier is designed to purify rare gases. Rare gases include helium,
argon, xenon and krypton. Most users will be using the purifier to purify argon and the purifier
is shipped after being run on argon, so is ready for immediate use with this gas. Users
wishing to run on another rare gas must expect the outlet gas to be contaminated with argon
until all the argon has been purged out.
The purifier is capable of purifying impure argon at a typical impurity level of 10-50 vpm
(volumes per million) down to less than 1 vpm total impurity. The maximum flow rate
recommended is 10 litres/minute. Higher purification efficiency will be achieved at lower flow
rates. The MP-2000 purifier will remove the common impurities found in argon, that is:
oxygen, nitrogen, hydrogen, moisture, carbon monoxide, carbon dioxide and hydrocarbons.
The purifier works by passing the rare gas through a heated bed of titanium granules at
700C which removes oxygen and nitrogen by chemical reaction. The gas then passes
through a heated bed of copper oxide which removes hydrogen and hydrocarbons by
oxidation to carbon dioxide and water. Residual moisture and carbon dioxide is then
removed by passing the gas through a bed of molecular sieve.
The titanium, copper oxide and molecular sieve are contained in stainless steel tubes,
referred to as the titanium tube the copper oxide tube and the molecular sieve tube.
Eventually the titanium and copper oxide material will be consumed and the purification
process will become inefficient. At this stage the user will need to replace these tubes.
Details on how to do this are given later in this manual. The molecular sieve can be
regenerated by heating to around 250C. The molecular sieve tube is equipped with band
heaters to enable the process of regeneration to be carried out. During the regeneration
process impurities held on the molecular sieve will be released. Therefore, during
regeneration, rare gas must be passed through the system to remove these impurities and
the gas outlet from the purifier should be diverted away from the downstream equipment
The molecular sieve can be regenerated many times, but eventually the regeneration
becomes less effective and the user may have to change this tube as well. Spare tubes are
available on an ex-stock basis from Sircal.
Fig 1 overleaf shows the main parts of the purifier. Other features to note in the diagram
include the heat exchanger block and the ready relay contacts. The gas passing out of the
copper oxide tube is hot; if it passes directly to the
molecular sieve tube without cooling, the molecular sieve will heat up causing partial
regeneration to take place and impurities to be given off. For this reason the gas is passed
through a heat exchanger block which cools the gas to near ambient temperature before it
enters the molecular sieve tube.
The gas connections at the rear of the purifier are self-sealing, preventing ingress of air while
the purifier is disconnected from a gas supply.
3 3
MODEL MP-2000 RARE GAS PURIFIER 4
COPPER OXIDE
E F GETTER TUBE
TUBE FURNACE G
(Cover removed to show
internal details) MOLECULAR SIEVE
DRIER TUBE
HEAT EXCHANGER
BLOCK
C
TITANIUM B
GETTER TUBE A D
"READY" RELAY
CONTACTS
Drg P1567-00
4 4
5
The ready relay contacts are available at the rear of the purifier and can be used to provide
a remote confirmation that the purifier is switched on and is operating at the correct
temperature for purification.
The hot titanium will be rapidly consumed if exposed to high levels of oxygen or nitrogen. It is
very important that the user never turns the furnace on without the unit being connected to a
supply of rare gas. Never pass air, oxygen or nitrogen through the purifier when the titanium
is hot.
The rate at which the titanium and copper oxide are used up will depend on impurity level and
flow rate. Further information about this is given later in the manual. As an aid to the user a
digital display on the front panel indicates the number of hours the furnace has run. This time,
in conjunction with a knowledge of the impurity level and flow rate can be used to estimate
the number of hours remaining before the tubes should be changed.
The furnace heater is rated for continuous operation at 700C. The heater element is
contained within an aluminium oxide powder contained within an outer metal tube, If the
purifier has been left in storage for many months, or if a replacement furnace heater is
installed at a later date, the aluminium oxide powder may have absorbed some moisture.
This trapped moisture is released suddenly when the heater is turned on and can cause
heater failure. To prevent this the switches on the front panel can be set to gently bake out
the heater before full power is applied, (heater bake-out function).
5 5
6
FURNACE
HOURS
Fig. 2
Note that the main power switch is located at the rear of the purifier. When the main power
switch is turned on the power led will glow green. If neither the PURE switch nor the REGEN
switch is turned on the purifier is in a stand-by state. To bake out the heater prior to applying
full power turn on both the REGEN and PURE switches (only necessary after long-term
storage or after replacing a heater element).
6 6
7
Installation
The following accessories are provided:
2 plug-in gas connectors for or 6mm tube (1/4 is standard, 6mm is an alternative
available at time of ordering)
1 mains lead
2 spare fuses
In addition, the following parts may be supplied with the purifier if ordered when
purchased:
Spare titanium tube and/or copper oxide tube and/or molecular sieve drier
tube
Gas Side Entry Kit. See APPENDIX at rear of this manual for more
information on this option.
The tubes do not need to be fitted now as the purifier is supplied with these tubes
already fitted. The spare tubes should be kept safely for future use.
Wire up the mains lead to a suitable mains plug (Brown Live, Blue Neutral,
Green/Yellow Earth)
If local electrical requirements stipulate that only the live line should be fused, remove the
existing fuse draw and fuse and replace with the alternative single-pole fuse draw. The fuse
draw is located below the mains switch at the rear of the purifier. It may be removed by
inserting a thin-bladed electrical screwdriver into the horizontal slot on the fuse draw and
levering slightly to release it, as indicated in the diagram below. Due to the restricted access,
a blade bent as shown below will most effective.
7 7
8
Verify that the electrical supply matches the voltage requirement of the purifier. Refer to the
serial number label or the rear panel label to confirm the operating voltage of the purifier.
Connect gas inlet and outlet lines to the purifier using metallic tubing (e.g. copper or stainless
steel). Do not use plastic tubing as air can diffuse through the plastic wall and contaminate the
gas. The tubing should be inserted fully into the gas fittings and held securely while the nut is
tightened up to swage the ferrules on to the wall of the tube. The nut should be turned 1
turns from hand tight to ensure the correct degree of ferrule compression. Ensure that the
tube wall is free from surface scratches to ensure no leak path can exist between the ferrule
and tube.
If the Ready contacts are required for remote indication of purifier status, wire up the plug
with 2-core screened lead and ensure that the screen is connected to the plug shroud. Note
that the Ready relay contacts are rated at 50V dc 100mA max. The relay contacts close
when the purifier is at the correct temperature for purification.
Before connecting the rare gas supply to the purifier allow some rare gas to flow to sweep out
any air in the tubing.
Ensure that the main power switch is turned off. Plug the mains lead and gas lines into the
purifier. The power switch can now be turned on, but ensure that both the PURE and
REGEN switches are in the off position. Regulate the gas supply pressure to a maximum of
17 Bar (250 psig) and regulate the flow to the required level, either with a down-stream valve
or by adjusting the inlet pressure.
If the heater has not been used for a long time, (say more than 3-4 months) we recommend
to bake out the heater element prior to running at normal temperature (see heater bake out
below), otherwise proceed to the operating instructions overleaf.
2. Set rare gas flow to a low flow rate and divert gas flow away from downstream equipment
3. Set REGEN and PURE switches to on position and leave for 1 hours
4. Switch off REGEN and PURE switches and leave gas flowing for further hour.
5. Place purifier back into service and proceed to operating instructions overleaf
8 8
9
Operating Instructions
1. Purge gas lines before connection to purifier.
2. Set inlet pressure to a maximum of 17 Bar (250 psig) and the flow rate to no more than 10
litres/minute.
4. Allow rare gas to purge the purifier for 5 minutes and then press the PURE switch to the
on position.
5. While the furnace is heating up the amber WAIT led will illuminate.
6. The HOT led will light green to indicate the purifier is at the correct temperature.
8. To turn the furnace off, press the PURE switch to the off position.
9. The amber WAIT led will illuminate and will extinguish when the furnace has cooled
down. At this stage the COOL led will illuminate green. At this stage the gas supply can
be shut down.
10. To ensure that the purifier remains filled with argon it is recommended that the gas flow is
stopped by shutting the outlet line while keeping the inlet gas line pressurised.
Alternatively the inlet and outlet gas lines may be disconnected from the purifier by
unplugging the gas connection.
11. To turn the purifier completely off turn the mains switch at the rear of the purifier
to the off position.
9 9
10
1. Disconnect the outlet gas stream from the downstream equipment if you do not want the
equipment to be exposed to moisture and other impurities released from the molecular
sieve while it is being regenerated.
2. Turn off the furnace heater by pressing the PURE switch to the off position. It is not
necessary to wait for the furnace to cool down before proceeding to the next step.
4. Press the REGEN switch to the on position. The COOL or WAIT led will extinguish and
the REGEN led will start to flash amber
6. Press the REGEN switch to the off position. The REGEN led will extinguish and the
COOL led will light green. Wait a further 30 minutes while the drier tube cools down.
7. Press the PURE switch to the on position and reconnect the outlet gas stream to the
downstream equipment.
10 10
11
4. Remove the front cover of the purifier and disconnect the earth lead attached.
6. Release the gas couplings A and B as indicated in Fig. 3. Use two spanners, one to hold
the lower wrench pad and to prevent any rotational torque being applied to the bodies of
the couplings.
7. Pull the titanium and copper oxide tubes upwards a little until the tubes separate from the
couplings. Place small pieces of adhesive tape over the mouth of the couplings before
pulling the tubes up any further. This is to prevent any debris falling into the couplings
from the furnace. Alternatively operate with the purifier laying on its back so debris cannot
fall into the couplings.
8. Pull the tubes completely out of the purifier and separate them by undoing the top
couplings. Replace the tubes as required and reassemble using the steps as described
in reverse sequence. Do not remove the protective tapes until the tubes are close to the
couplings.
10. Apply rare gas to the inlet to pressurise the purifier and check for any gas leak using leak
detector fluid. If no proprietary fluid is available, diluted detergent will suffice. Use a
strong light and a magnifying glass to inspect for small bubbles which will indicate a leak.
11. Replace the top and front covers, not forgetting to re-install the earth lead on the front
cover.
12. Reconnect the purifier to the rare gas supply and electrical supply. Allow rare gas to flow
through the purifier for 5 minutes before turning on the power and pressing the PURE
switch.
11 11
12
Step 3: LIFT TUBES OUT OF PURIFIER
Step 1: REMOVE TOP COVER
E F
C
B
Step 2: RELEASE A D
COUPLINGS A & B
12 12
13
3. Remove the two split rings retaining the drier tube within the clips by grasping the clip with
pliers and rotating the ring so that the clip goes between the split part of the ring. By
rotating 360 the ring will come free. These rings need not be replaced unless the purifier
is to be transported somewhere. The rings prevent damage that may otherwise occur
while in transit.
5. Undo the couplings C and D using two spanners, one to hold the lower wrench pad and
to prevent any rotational torque being applied to the bodies of the couplings.
6. Slide the drier tube up to disengage the tubes from the couplings.
7. Pull the drier tube away from the clips and remove the drier tube assembly from the
purifier.
8. Remove the drier tube from the assembly and replace with a new drier tube. Note the
orientation of the band heaters on the tube and replace in the same fashion.
9. Do not tighten up the coupling G at the other end of the tube until the assembly has been
installed back in the purifier.
12. Apply rare gas to the inlet to pressurise the purifier and check for any gas leak using leak
detector fluid. If no proprietary fluid is available, diluted detergent will suffice. Use a
strong light and a magnifying glass to inspect for small bubbles that will indicate a leak.
13. Replace the front cover, not forgetting to re-install the earth lead on the front cover.
14. Reconnect the rare gas and electrical supply. Allow rare gas to flow for 5 minutes before
turning the purifier back on. To accelerate purging of the molecular sieve tube it is
recommended that the REGEN be used for around 30 minutes to accelerate desorption
of air from the molecular sieve.
13 13
14
E F
REMOVE SPLIT
G RINGS
C RELEASE
COUPLINGS
B
D C AND D
A
14 14
15
The lifetime of all three purifier tubes can be estimated by the following formula:
T = 100,000
Fx V
Example: F = 2 litres/minute
V = 15 vpm
T = 3333 hours
Assumptions:
Titanium tube:
The formula assumes a 50% consumption of the titanium and a 1:1 ratio for the chemical
reaction (worst case for nitrogen). This is a worst case scenario. In practice the tube life is
likely to be longer by up to 50% of this calculation.
The formula assumes a 50% consumption of the copper oxide and an average 2:1 ratio
(CuO:impurity) for the chemical reaction. The lifetime of the tube will probably be longer than
the calculation indicates, especially if the hydrocarbon impurity level is low.
The formula assumes the molecular sieve has undergone a large number of regenerations
and as a result its capacity has reduced by 65%. A 25% safety margin is added also.
15 15
16
Fault Condition
The front panel FAULT led will light red and the audible alarm will sound to indicate any of the
following problems:-
The fault detection circuit will shut down power to the furnace and turn off the COOL, WAIT
and HOT leds as these indicators may not be reliable under a fault condition.
The FAULT indicator can be reset by turning the PURE switch off which will put the purifier
into stand-by mode.
First action should be to check the integrity of the thermocouples. If both thermocouples are
OK the fault will be with the power control circuit and the control panel should be replaced or
returned for servicing.
Impure gas
Impure gas is not necessarily a fault condition. It may indicate that one of the purifier tubes
may need changing. Impure gas may arise, however, under one of the conditions listed
below.
The REGEN switch has been pressed in error, causing the furnace to cool down (equivalent
to heater bake-out mode)
The REGEN switch has been pressed instead of the PURE switch. In this case the drier will
be regenerating and giving off impurities and the purifier furnace will be cold.
The furnace heater has gone open circuit. In this case the WAIT led will still be on and the
HOT led will have failed to light. Check the furnace heater resistance to confirm this. Note
that the 230V model heater should measure around 65 ohms and the 110V model heater
should be around 14 ohms.
If none of these conditions apply, try using the REGEN facility to regenerate the molecular
sieve drier tube. If this fails to correct the situation, consider replacement of the titanium and
copper oxide tubes. If the molecular sieve drier tube has undergone more than 500
regenerations, consider replacing this tube also.
16 16
17
1. Turn off the purifier and wait for it to cool down if the furnace is still hot (it probably will be
cold as the heater will probably have failed unless the heater is being replaced as part of
a preventative maintenance program)
4. Remove the control panel from the purifier by unscrewing the supporting screws at the
base of the purifier and disconnecting the connecting wiring loom and earth wires.
5. Undo the 6 M6 nuts securing the purifier assembly to the enclosure. (Labelled X in Fig
.1)
7. Remove the titanium and copper oxide tubes as described in the procedure for replacing
these tubes (steps 6 &7)
8. Unscrew the front face of the furnace (screws A in Fig. 5) and remove the insulating panel
and metal plate below. Remove several layers of the ceramic fibre to expose the furnace
heater.
9. Note: Use a dust mask and gloves when handling the ceramic fibre.
10. Unscrew the side face of the furnace that supports the heater and pull the panel out from
the body of the furnace.
11. Mount a new furnace heater on the side panel. Adjust the positions of the thermocouples
so they are in contact with the furnace heater, but do not come into contact with the
titanium tube.
13. Re-insert the titanium and copper oxide tubes and tighten up the gas connections
14. Recheck that the thermocouples are in contact with the heater and then replace the
ceramic fibre insulation.
15. Replace the furnace top covers and reinstall the assembly within the enclosure
17 17
18
17. Apply rare gas to the inlet to pressurise the purifier and check for any gas leak using leak
detector fluid. If no proprietary fluid is available, diluted detergent will suffice. Use a
strong light and a magnifying glass to inspect for small bubbles which will indicate a leak.
18. Replace the enclosure front panel, not forgetting to re-install the earth lead on the front
cover
20. Set rare gas flowing at a slow rate (say 50-100 cc/min), with gas diverted away from
downstream equipment.
21. Press the PURE and the REGEN switches to the on position. This will bake out the
heater and release any moisture that may be trapped. Leave for 1 hours.
22. Switch REGEN and PURE switches to off position and allow gas to flow for a further
hour.
18 18
REMOVE TITANIUM AND 19
COPPER OXIDE TUBES
F
A A
E
A A
FIG. 5
19 19
20
Connections
All mains supplies and heater connections are made via PCB mounted connector PL1.
Although an earth connection is present on PL1, an additional cable and earth tag soldered to
GP1 is provided to ensure that the PCB ground is connected to the bracket.
Thermocouple connections are made via two screw terminal blocks ST1 and ST2, between
which is situated a thermistor cold junction sensor.
The elapsed timer connections are available on PL2.
The contact closure for remote indication of HOT is available on PL3.
Power Supplies
The power supply has a conventional format with a mains transformer, rectification,
smoothing and regulation. A mains selector switch SW3 controls the transformer primary
connections for either 230 Vac or 115 Vac supplies.
An additional selector switch SW4 controls the wiring for the REGEN heaters. When set to
115 they are wired in parallel, when set to 230 they are wired in series..
FS1 protects the low power control circuit. A 100 mA A/S fuse is fitted for 230 Vac, and 160
mA A/S fuse for 115 Vac operation.
U13 provides a regulated +12 Vdc rail. U14 provides a regulated -12 Vdc rail.
Reference
U15 is a precision reference voltage generator. Its wide initial tolerance (for this application)
requires VR2 and R57 for final adjustment.
U12 is used to generate the equivalent negative reference voltage. It is configured as a unity
gain invertor with R48 and R60 controlling the gain. VR1 is set at initial test to give -5.000 V
from this output. Using 0.1% resistors and a precision amplifier keeps the positive reference
voltage at +5.000 V 0.010 V.
Preamplifiers
There are two identical amplifiers, one for each thermocouple. Only one is described. Circuit
references that apply to the second channel are given in [] brackets.
The thermocouple input is filtered by R2 and C1 [R3 and C2] before being applied to the non-
inverting input of U1 [U2]. The amplifier gain is nominally set by R18 and R16 [R19 and R17]
although it is also slightly affected by R8, R9, R12 and R13 [R10, R11, R14 and R15]. The
gain is set for maximum accuracy around the 700 C area giving a +5 mV/C output. A type
K thermocouple has a non-linear output with temperature. In this circuit it will indicate about 6
C low at 300 C
20 20
21
21 21
22
22 22
23
G1 is a precision bead thermistor arranged as a single arm bridge across the two reference
supplies. The output is nominally 0.00 V at 25 C and varies at -0.1 V/C. It also is non-linear
and produces an error of about 1 C over the ambient range -10 to +50 C. This output is
subtracted from the amplified thermocouple signal by R8 and R9 [R10 and R11]. Because
the output is zero at 25 C an additional subtraction is made by R12 and R13 [R14 and R15]
to give a fully corrected output T1 [T2].
R5, R6 and R7 [R4] provide a slight positive bias to ensure that an open circuit thermocouple
input will produce an output equivalent to >1000 C which will trip the alarm circuits.
Proportional controller
Error Amplifier
U3 is an inverting error amplifier designed to amplify the difference between the average of
the two thermocouples and a set point. The set point is chosen by the position of the REGEN
switch to be either 685 C (REGEN off) or 150 C (REGEN on - sets a burn-in temperature
when both REGEN and PURE switches are on at the same time).
The positive inputs via R20 and R21 are combined with the negative reference from R22 (or
R23) and the difference is amplified with a gain of about 66. The output varies by -0.33 V/C
error and limits at the supply rails.
The accuracy of the set point is governed by the value and accuracy of R22 (or R23).
Precision 0.1% resistors are used which allows the set point to be within 7 C of any
required value. The accuracy required for burn-in is not critical and a 1% resistor is used.
R = 50000000
The required resistor value for any set point (C) can be calculated from Setpoint
.
An adjustment facility is provided by VR1, R24 and R25. The approximate change in set
point is 5 C/V allowing a 25 C adjustment. The adjustment voltage is available on SKT1 to
allow determination of the amount of adjustment added. The range of adjustment has been
set to cover the proportional band.
Ramp Generator
U6 is configured as a triangular wave oscillator. R34 and R35 control the amplitude which is
set for a nominal 8.7 Vp level. R34, R35, R36 and C9 all control the oscillator frequency
which is set to a nominal 0.37 Hz (2.7 s Period). The square wave output from U6/7 is gated
with the FAULT and REGEN outputs to create a modulated warning.
23 23
24
Proportional Band
U4 is a comparator with inputs from the ramp generator and the error amplifier. The output is
high (power off) when the output from the error amplifier is lower than the ramp (temperature
too hot), and low (power on) when the output from the error amplifier is higher than the ramp
(temperature too cold). When the output from the error amplifier is within the 8.7 Vp ramp
level the output will cycle at 0.37 Hz changing its on:off ratio proportionally with the error.
An assumption has been made that the correct power to maintain temperature will occur
when the error amplifier output is zero. As mains, gas flow rate, gas input temperature, and
ambient temperature vary, the actual temperature required to balance heat input against heat
loss will vary. The gain has been set so that the variation from zero to full power requires a
variation of 53 C in the set point (i.e. 21.5 C from nominal). Mains drive U5 and Q1
provide the conversion from the proportional band output (U4) to an isolated mains drive.
Mains drive
U5 is an opto-isolated triac designed to interface with the power triac U5. Drive to the internal
LED cannot occur until the PURE switch is on when EN1 and EN2 are connected together.
R39 and C10 are suppression components.
Safety
Two relays are used to isolate the mains from the heaters. RLY2 allows mains to go to the
regen heaters when the REGEN switch is set. RLY1 removes mains from both the purifier
heater and the regen heaters if the fault condition exists. The relays are driven between the -
12 V to +12 V rails to balance and minimise the current demand
Threshold1
U7 is a quad comparator used to detect the various threshold levels required for control of the
status indicators and fault conditions. U7a and U7b set /LIMIT if either thermocouple pre-
amplifier registers a temperature over 751 C. The threshold is set by R44 and R45.
U7c uses the mean of the two thermocouple pre-amplifiers to detect the high threshold (/HT)
of 651 C above which the purifier will be working. U7d uses the mean of the two
thermocouple pre-amplifiers to detect the low threshold (LT) of 279C below which the gas
can be disconnected. These two thresholds are set by R47, R48 and R49 and are
interactive.
The forward slash symbol (/) is used to denote the logical NOT function. E.g. /LIMIT means LIMIT is true when its voltage level is low.
24 24
25
Logic
U8, U9 and U10 are CMOS logic gates running between the -12 V and 0 V rails. The high
voltage increases the noise immunity over 5 V logic and does not require additional supplies.
They provide the combinatorial logic necessary for controlling the status LEDs, relays, buzzer,
and timer.
RELAY
This output controls RLY2, the primary mains switch. It controls the mains supply to both
heaters.
RELAY will be false while /LIMIT is true. It will stay latched false if the pure switch is on
(/PURE true) and /LIMIT goes false. It will change to true if /LIMIT is false and the pure switch
is off (/PURE false).
This ensures that no power can be applied to the heaters while the purifier temperature
exceeds its limit no matter what state the pure switch is in. It also ensures that power will not
be re-applied until the pure switch has been set off after the limit condition has been detected
or if the mains supply has been removed and re-instated.
FAULT
This output controls the FAULT LED and the buzzer. Both are always on or off at the same
time.
FAULT will be false during normal operation. It will go true when /LIMIT is true and the pure
switch is on (/PURE true). It will stay true after /LIMIT goes false provided that /PURE is still
true. It will always return false if /PURE is false.
This ensures that the fault condition can always be turned off by setting the pure switch off.
Note that power to the heaters will not be applied until the /LIMIT condition is false even if the
pure switch is off (see RELAY).
TIMER
HOT
25 25
26
WAIT
COOL
Drivers
All LED, relay and buzzer drivers (Q2-9) are N channel MOSFETs. Q8 is driven directly from
the regen switch. All others are driven from logic gates. Q6 and Q9 are gated with the ramp
square wave output.
They are all referenced to the -12 V rail and the gates are driven to 0 V to turn them on. The
LED, buzzer and elapsed timer returns are 0 V; the relay returns are +12 V.
Test Points
Edge of pcb
+REF 1 2 -REF
CJC 3 4 T1
T2 5 6 AVERAGE
ERROR 7 8 ADJUST
EN2 9 10 TGND
ERROR: Amplified voltage difference between thermocouple and set points (set
point is 685 with REGEN off and 150C with REGEN on)
26 26
27
EN2: Output of proportional band comparator I.C. U4. Near set point EN2
should ramp up & down about every 2 secs.
OPTIONAL FEATURES
Pressure Switch
The purifier may we supplied with a pressure switch which turns off the furnace if the inlet
gas pressure falls below a set value.
The pressure switch supplied has an electronic switch with an NPN output interface. It is
connected to PL5 as follows:
Pin 1: Power
Pin 2: Output 1 Set so output is on (i.e. conducting) when pressure is lower than the
required limit
Pin 3: Ground
Note that neither Pin 1 nor Pin2 are at 0V. A short from either if these pins to ground may
result in power supply failure
When pressure falls below the set limit the purifier is placed in FAULT mode and power to
the furnace heater will be cut. The alarm will sound. The operator must check the pressure
displayed by the pressure switch to see if the fault is a genuine fault condition or due to the
inlet pressure being too low.
Note also that heat sinks are fitted to U13 and U14 if the pressure switch option is present.
27 27
28
The titanium, copper oxide and molecular sieve tube are consumable items. The
warranty on these tubes is restricted to physical integrity, i.e. leak tightness of
welded joints. Tube lifetime is excluded from the terms of the warranty as SIRCAL
has no control over their conditions of use
SIRCAL undertakes to repair and return purifiers free of charge when the failed parts
are within the specified warranty periods listed above.
Service
Sircal Instruments may be contacted at the address below if you have any problems with the
use of this purifier:
UNITED KINGDOM
SM2 6RJ
28 28
29
Thermocouple MP2000/TC/01
Note 1 Factory fitted design not for use in field on standard MP2000 purifier
Always quote the serial number of the purifier when ordering spares.
29 29
30
APPENDIX
Accessories
A gas side entry kit is available as an optional extra for users who have restricted
access to the rear of the purifier, or where on old version Sircal Rare Gas Purifier is
being replaced with the new MP2000 model and the user wants to use the gas inlet
and outlet pipes set up without any additional re-piping work.
Two versions of the kit are available. The Type C Side Entry Kit is designed for use
on a standard MP2000 purifier and can be fitted by the user at any time. The Type F
Side Entry Kit is only provided in combination with a special version of the MP2000
purifier (designated MP2000-SE). In the MP2000-SE the rear Quick-Connect gas
connectors are replaced by simple inlet bulkhead connectors. In both types two
Quick-Connect gas connectors are mounted on a bracket on the left hand side of the
purifier.
The Type F kit has a lower profile than the Type C kit. The MP2000 with a Type F kit
can be placed almost flush against a rear wall, whereas the MP2000 fitted with a
type C kit will require an additional 50mm of space to accommodate the bends of the
tubes.
The Type C is also the more expensive option as it entails the use of 4 Quick-
Connect gas bulkhead connectors rather than two in the type F kit. The Type C kit
has the advantage, however, of being adaptable to the standard MP2000 purifier,
whereas the Type F kit can only be used on a specially ordered MP2000-SE purifier.
30 30
31
31 31