Beruflich Dokumente
Kultur Dokumente
GIOTTO
MAMMOGRAPHY
HF
GENERAL TECHNICAL
DESCRIPTION
t-
IME
1 GENERAL TECHMCAL DF^SCRIPTION
- 7'7-
tM5
cl: levelling capacitatorfor ! 24 V.
C 2 : levelLingcapacitatorfor t 90 V,
C 5 : startecapacitatofor rotaig anode.
PRI rcctifier bidge = 24 V.
Usedfor supplyingservicesin l Z V.
PRz
usedfo supplyingservicesi t 90 V.
PGI grid speedregulationptentiometer.
PG2 gid speedregulationpotentiometer.
TI genel electicalsupplycontactor.
T2 X-ay contacto.
The esistosmountedo[ top of the concactsare
usedin the minute of standby at the momentof
ignition il orderto bring theinvertecircuitsto the
propelevel.
R 1 relay cormandedby the micropocessorfor
enabling the Tl ON/OFFconctator.
R2 X-ray servicesrelay
R3 High speedrelay.ComeswhenmAs are< 40 mAs.
FllF2l senealmachinefuses25Al500 V 10x38
F3 4A 5x20
F4 10.A5x20
F5 8A 5x20
F6 6A 5x20
F"7 6A 5x20
F8 6A 5x20
F9 4A 5x2O
F l 0 4A 5x2O
F l 1 6A 5x20
Ft2 2A 5x2O
F l 4 250mA
F 1 5 10A
F 1 6 40A
-78-
IME
RACKMOWMENT
P.C.B. mocroprocessormovement(JPUmov)
This card managesall the machinemovementfunctions.
It alsoprogramsthe motorizedcompressioncorcuit with
the electuonicallyadjstableforce.
There are regulation tiruners on this card for the Up/
down - rctation - inclination motos.
Fo systemcalibration,seethesectionegadingegulation
procedwes.
P.CJ. Supply(JP4lmov)
Thissuppliestherackmovement
andits rclativeservices.
Thereaenoegulations.
-79 -
IME
X.RAYRACK
P.C.B.Supply
JPI/X-RAY: This cardenablestheX-ray micoprocessor
atthemomentcunentis suppliedby means
of a wall switch found abovethe machine.
Theeare no regutations.
P.C3. Starter
JPZX-RAY ContolslauDchitrgtime in preparation
(about 1 second)alongwith the normal
furiction of the X-there reno regulatios.
P.C.B. Filament Supply
JP3/X.RAY: Contolsthe electricalvoltageof
thetwo filaments,smallfocusand
largefocus.A vetical I Ktimmer
is mountedon the cardwhich
egulatesits voltage.We suggest
stabilizing it at 9,5 V. The two
leds Red led : largefocus
Geenled : small focus.
P.CS. Supply
JP4IX-RAY: This cardsuppliesall the circuits in theX-
ray rack uponignition of the mahine.
Thereareno rcgulations.
P.C.B. Filamert calibration
JPS/X-ray : This contolsfilamentscalibationbothfo
small andlarge focus,For any corrections
in calibation,consultthesectiorregardi[g
regulationprocedures.
P.CJ mAs cutting
YP6/X-RAY :This hasthe puposeof commencingand
terminatingexposuretime. It alsocaries
out conftols for:
-filament safety
-photo time ON / phototimer OFF
-invefer command
The greenled found on the cad,itrdicates
that all contols havebeencariedout. The
timmer mountedon onesideofthe cardis
used.foranyeventualcoections in the
tAsmeter with the setvalue.
For anyeventualclaifications,consultthe
sectionoregulationpoceduesin thepafi
regadingmAs cutting regulation.
JP7IX-RAY : P.C.B.interfacecadbetweenx-ray rack
ardinvete andH.T. trasfomercontol.
JP8IX-RAY : MicropocessorCPU P.C.B.to contol all
x-ray ack.
-80-
CONTROL INVERTER AND E.T. TRASFORMER
Seedetaileddiagramsandspecifications
onH.T.trasfomer
andinverter section,
. - 8 1-