Sie sind auf Seite 1von 9

EVA-HF325/525

CALIBRATION

Technical Publication
Ver.3.2

CALIBRATION
EVA-HF325/525 generator

TABLE OF CONTENS

1. INTRODUCTION ------------------------------- 4

2. CALIBRATION PROCEDURES-------------------- 5

3. CALIBRATION----------------------------------- 6

COMED Medical Systems Co., Ltd.


EVA-HF325/525
CALIBRATION

4. DIP SWITCH------------------------------------

1. INTRODUCTION

The generator contains many self-diagnostic routines which greatly facilitates


troubleshooting. The self-diagnostic functions require that all microprocessors
function properly. Each microprocessor contains LEDs, LCD that indicates their proper
operation.

THE MAIN STORAGE CAPACITORS RETAIN A LARGE PORTION


OF THEIR FOR APPROXIMATELY 5MIN. AFTER THE UNIT IS

COMED Medical Systems Co., Ltd.


EVA-HF325/525
CALIBRATION

TURNED OFF.

ALWAYS HAVE IPM DRIVER PCB CONNECTED WHEN MAIN


POWER IS ACTIVATED IN THE GENERATOR. WHITOUT IPM
DRIVER PCB CONNECTED, PERMANENT DAMAGE WILL
OCCUR TO IPMS.

DO NOT USE THE GENERATOR POWER SUPPLIES TO


POWER EQUIPMENT EXTERNAL TO THE GENERATOR. ALL
GENERATOR INPUTS AND OUTPUTS MUST BE TTL LOGIC
LEVELS (TYPICALLY ACTIVE LOW) AND ISOLATED FROM
OTHER SUBSYSTEMS OPTICALLY OR WITH DRY CONTACT
RELAYS.

IF CHANGE EEPROM ON CONSOLE DISPLAY BOARD, WILL


LOSE CALIBRATION DATA, APR DATA, AND SOME DATA.
SO MUST CHECK EEPROM DATA.
AND IF CHANGE UV-PROM ON HT CONTROL CPU BOARD
AND CONSOLE DISPLAY BOARD, WILL CHECK THE VIRSION
OF PROGRAM.

2. CALIBRATION PROCEDURES

1. To enter the Calibration mode change position of 8Pin of DIP S/W1 to OFF
position.
Pic. Console display board

COMED Medical Systems Co., Ltd.


EVA-HF325/525
CALIBRATION

DIP SW1

2. Push power ON switch. And, right LCD window will display TUBE1 RAD CAL MODE
after initializing.
Pic. Display window and membrane switch.

Right LCD display window

3. Return to the DIP S/W original position after calibration complete.


(Anatomical mode)

3. CALIBRATION
3.1 DESCRIPTION
Pic. Right LCD display

COMED Medical Systems Co., Ltd.


EVA-HF325/525
CALIBRATION


: KV DATA OFFSET VALUE(not changeable)
: KV DATA ADJUST VALUE(changeable)
: MA DATA OFFSET VALUE(not changeable)
: MA DATA ADJUST VALUE(changeable)
: MA COMP ADJUST VALUE(changeable)
: *(star mark) MEANNING OF SELECTED

Pic. Membrane pannel

increasing kV data decreasing kV data


increasing mA data decreasing mA data
increasing mA comp data decreasing mA comp data
KV UNLOCK key : enable to increasing or decreasing kV data.

The star-mark (* ) display on window. If * is not


displaying, will not be able to increase or decrease
data

mA UNLOCK key : enable to increasing or decreasing mA data.


mA-COMP UNLOCK key : enable to increasing or decreasing mA-COMP
data.
LOCK key : disable to increasing or decreasing all data.
(The star-mark not visible)
SET key : save kV, mA, mA-COMP data to EEPROM

COMED Medical Systems Co., Ltd.


EVA-HF325/525
CALIBRATION

3.2 PROCEDURES
Tools & Equipment required : OSC or DYNALIZER
First, condition setting is 40kV, 200mA, 0.025sec. And exposure it.
Second, check it with two method, one is use with Oscilloscope, the other
is use with DYNALIZER. This Calibration method is example for using with
Oscilloscope
Pic. Testing point kV, mA

+mA

+kV

GND
When measure +kV wih OSC, real kV is 10kV per 1V.
Also, real mA is 20mA per 1V.(refer chapter 8 of troubleshooting)

Pic. Right LCD display


EEPOR
M
(all kV,mA

data)

Table 1. KV data restore procedure


40kV 45kV 50kV ....
150mA (A1x1) w1 (A1x1) w2 (A1x1) w3 ....

COMED Medical Systems Co., Ltd.


EVA-HF325/525
CALIBRATION

200mA A1x1 A2x2 A3x3 ....


250mA (A1x1)y1 (A1x1) y2 (A1x1) y3 ....
300mA (A1x1) z1 (A1x1) z2 (A1x1) z3 ....

A1,A2,A3 is value restored in EEP-ROM(program IC) : not changeable


W1,w2,w3 and x1,x2,x3 and y1,y2,y3 etc is saved in EEP-ROM :
Changeable

Table 1. mA data restore procedure


40kV 45kV 50kV ....
150mA B1w1 B2w2 B3w3 ....
200mA C1x1 C2x2 C3x3 ....
250mA D1y1 D2y2 D3y3 ....
300mA E1z1 E2z2 E3z3 ....

B1,B2,B3 and C1,C2,C3 and D1,D2,D3 etc is restored in


EEP-ROM : not changeable.
W1,w2,w3 and x1,x2,x3 and y1,y2,y3 etc is saved in EE-PROM.

Procedures
1. Condition setting 40kV,200mA,0.025sec
2. Push ORG1,ORG2,ORG3(see membrane pannel) will display star-mark(*).
3. Exposure and check kV,mA,comp data and then data up/down with membrane key
4. Save correct data with set key.
5. Program : 40kV,200mA,0.025sec 45kV,200mA,0.025sec 50kV,200mA,0.025sec
55kV,200mA,0.025sec . . . .
6. Comp data equal to all 200mA step(comp data is about +03 at 200mA step)
7. If all 200mA step calibration end, push ORG6. and then re-start 250mA step.
8. Condition setting 40kV,250mA,0.025sec.
9. Push ORG1,ORG2,ORG3
10. Exposure and job 3 ~ 6mumber excute.
11. 40kV,250mA,0.025sec 45kV,200mA,0.025sec 50kV,200mA,0.025sec . . . .
12. 40kV,300mA,0.025sec 45kV,200mA,0.025sec 50kV,200mA,0.025sec . . . .

4. DIP SWITCH

COMED Medical Systems Co., Ltd.


EVA-HF325/525
CALIBRATION

4.1 CONSOLE DISPLAY BOARD

DIP S/W 1

DIP S/W
DIP S/W 1
ONLY 8 PIN OFF : CALIBRATION MODE
ALL PIN ON : ANATOMICAL MODE

DIP S/W
ONLY 1 PIN OFF : EVA-HF325 MODE
ALL PIN ON : EVA-HF525 MODE

4.2 HT CONTROL CPU BOARD

DIP S/W 1

DIP S/W 2
DIP S/W 1

COMED Medical Systems Co., Ltd.


EVA-HF325/525
CALIBRATION

1PIN : BUCKY 1 method (OFF : spring ON : motor)


2PIN : BUCKY 2 method (OFF : spring ON : motor)
3PIN : BUCKY 3 method (OFF : spring ON : motor)
4PIN - 8PIN : NOT USED
DIP S/W 2
1PIN : ROTOR CHECK(OFF: CHECK , ON : NO CHECK)
2PIN : FILAMENT CHECK
3PIN : KV DATA CHECK
4PIN : MA DATA CHECK
5PIN : TEMP CHECK
6PIN : CHARGE CHECK
7PIN : NOT USED
8PIN : IPM CHECK

COMED Medical Systems Co., Ltd.