Sie sind auf Seite 1von 2

QA-ST

Electrical Safety Tester


PRODUCT HIGHLIGHTS OVERVIEW
Manual and maximum simplicity The QA-ST is designed to quickly check electrical
Operating voltage from 90 -240 V (50-60Hz) safety on medical and industrial devices. It is palm
sized and very portable. The tests can be performed
Power configuration test
manually or in step by step mode. The results are
Supply Voltage measurements displayed in a bright 4-digit LCD display.
Current Consumption
Ground wire resistance The QA-ST performs all electrical safety tests neces-
sary to satisfy AAMI ESI, NFPA-99, VDE 701 and
Ground, chassis, and patient leakage current
VDE 751. It also features differential current meas-
Lead to Lead leakage current urements obtained by an internal current clamp. This
Replacement equipment current *) presents a new leakage current measurement
Replacement patient current *) technique for convenient testing of medical equip-
Differential current measurements *) ment.
Test loads: AAMI/ANSI ES1-1985 or IEC 601.1 The QA-ST is available in two versions:
1) US version which includes AAMI/ANSI ES1-
*) Included in the European version only 1985 test load.
2) European version which includes IEC 601.1 test
load.
QA-ST Specifications

LINE VOLTAGE MEASUREMENT GENERAL INFORMATION QA-ST


Temperature Requirements: Ordering Information
Range: 0 - 300 V +15/59 to +35/95 C/F while operating
Accuracy: 1% of reading, 0 to +50/122 C/F for storage Order no:
1 digit
US version:
Resolution: 1V Display: 11040: QA-ST Electrical Safety Tester
L-N Type: LCD
Numeric format: 4 characters European version:
Display control: Keypad 11041: QA-ST Electrical Safety Tester
CURRENT CAPACITY (Specify power supply socket)
Power: 100 V/110 V or
15 Amps @ 100 Volts 230 V/240 V 50/60 Hz
15 Amps @ 120 Volts
Accessories:
10 Amps @ 230 Volts Housing: Plastic case Various power supply socket including:
European Schuco, French Schuco, UK,
Dimensions: D x W x H Swiss, Australian, US.
RESISTANCE MEASUREMENT 191mm x 135mm x 45mm 17260: Hard Case
17458: Test lead (Kelvin cable)
DC-Source: Weight: 930 g
17024: Universal Banana Adapter
Range: 0 - 19.99 Ohms 11045: User/Service Manual QA-ST
Standard Accessories:
Accuracy: 2% of reading, User/Service Manual
1 digit Test lead
Resolution: 0.01 ohms
Current Source: 100 mA DC

AC-Source:
Range: 0 - 1999 mOhms Note:
Accuracy: 2% of reading, Specifications may be subject
1 digit to change without notice.
Resolution: 0.01 ohms
Current Source: 200 mA AC

LEAKAGE CURRENT

Range: 0 - 1999 A RMS


Accuracy: DC and 25 to 1.0 kHz,
2% of reading,
1 digit
1.0 kHz to 100 kHz,
2.5% of reading,
1 digit
100 kHz to 1 MHz,
5.0% of reading,
1 digit
Resolution: 1 A

INPUT IMPEDANCE

1000 ohm, AAMI ES1-1993 load or


IEC 601.1 load

POWER CONSUMPTION

110 V (50/60) Hz = 200 mA


115 V (50/60) Hz = 200 mA
240 V (50/60) Hz = 100 mA

MTD 0102.192.05 www.metron-biomed.com