In this chapter, electrocardiograph problems are listed along with possible causes and recommended corrective
actions. Refer to the tables to check the electrocardiograph, identify and eliminate the troubles.
The troubles we list here are frequently arisen difficulties and the actions we recommend can correct most
problems, but not all of them. For more information on troubleshooting, contact our Customer Service
Department.
4.2Part Replacement
Printed circuit boards (PCBs), major parts and components in the electrocardiograph are replaceable.Once you
isolate a PCB you suspect defective, follow the instructions in 7 Disassembly and repair to replace the PCB
with a known good one and check that the trouble disappears or the patient monitor passes all performance
tests. Defective PCB can be sent to us for repair. If the trouble remains, exchange the replacement PCB with
the original suspicious PCB and continue troubleshooting as directed in this chapter.
Some troubleshooting may involve software compatibility. Thus it requires you to your electrocardiograph
configuration and software version. For detailed information on version compatibility, please contact our
Customer Service Department.
Before troubleshooting the electrocardiograph, check for technical alarm messages. If an alarm message is
presented, eliminate the technical alarm first.
For detailed information on technical alarm message, possible cause and corrective action refer to the
electrocardiographs Operation Manual.
4.5 Troubleshooting Guide
AC power supply cable is not Check whether the power supply cable is
connected well. reliably connected
The two fuses defective Remove power supply cable and open the
fuse-boxes then ensure each of them is good
by multi-meter. If not, change new fuses..
The electrocardiograph
AC power supply unit defective Switch the multi-meter to AC voltage gear
fails to start with AC
and test the transformer output. The
power.
reasonable output should be 18V-20V
LCD cable didnt connect well. Check whether LCD cable is reliably
The electrocardiograph connected
fails to start with Battery defective. Check the battery voltage is around 14.8V
battery.
Main board defective Change the new main board
Keypad cable is not connect well. Check the keypad cable or connector cable
is reliably connected
White or Black screen
System failure Change another main board
LCD cable is not connect well. Check whether LCD cable is reliably
connected
Abnormal screen
display LCD cable defective Change new LCD cable
Identical electrode falls off or has no re-install the electrode or paste some water
reliable linked with patient body or conductive ointment at the contact
position
Cable has been broken due to long The cable and electrode of the same model
time use or unit aging/clamp shall be used to replace the old one.
electrodes or chest bulbs lose
No ECG output from
flexibility or weak connection
certain lead.
ECG board and main board cable Check the internal wire between ECG board
didnt connect well. and main board connect well.
Identical electrode had no reliable re-install the electrode or paste some water
linked with patient body; or conductive ointment at the contact
position
ECG machine, electrode, metal bed or Check ECG machine, electrode, metal bed
ECG disturbance lead cable reliably connected with or lead cable reliably connected with
grounding grounding
the patient wearing jewelry such as switch on the ECG filtering function of the
glasses or gems, AC disturbance can machine
not be cleared.
Communication disturbance
Switch on the shift filtering function of the
Electromyograhic The doctor talk to the patient machine to 25Hz or 35Hz, and then the
disturbance during ECG recording process recorded ECG waveforms will reduce a
little bit.
The Electrode clippers on the
arms and legs too tightened
electrode and patient skin clean Switch on the shifting filtering function of
The electrocardiograph
electrode and skin pasted with the machine to 0.1Hz, 0,2Hz or 0.5Hz, the
fails to start with
sufficient conductive ointment recorded ECG waveforms will reduce a
battery.
little bit.
it is due to body movement or
respiration of the involved patient or
mixed an old electrodes with new
electrodes for together use
Keypad cable between the keypad Re-connect keypad cable between the
Partial buttons or all
and main board out of proper keypad and main board out of proper
buttons failure
connection. connection.
Keypad cable between the keypad Replace the Keypad cable between the
and main board broken due to keypad and main board
long time of use.
Inverse putting of paper Check whether the print paper in the right
position.