Sie sind auf Seite 1von 50



Sensation
CT

Software
Configuration of SOMARIS/5 VA40
for factory delivered systems

This document is valid for: Sie mens AG 2001


The reproduction, transmission or
SOMATOM Sensation 4 use of this document or its contents
is not permitted without express
SOMATOM Volume Zoom written authority. Offenders will be
liable for damages. All rights,
SOMATOM Volume Access including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.

English
Print No.: CT02-021.816.02.02.02 Doc. Gen. Date: 11.01
Replaces: CT02-021.816.02.01.02
0-2 Revision
Chapter Page Revision
all all 02

Document revision level


The document corresponds to the version/revision level effective at the time of system delivery. Revi-
sions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho-
rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.

Sensation CT02-021.816.02 Page 2 of 4 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
Contents 0-1

Page
1 _______General _______________________________________________________ 1 - 1

Definitions and Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1


Overview of ICS / IRS / IES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Configuration overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 4

2 _______ Configuration _________________________________________________ 2 - 1

General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2- 1
Initial Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2- 1
Local Host . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Site Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 4
TCP/IP LAN "hospital network node"
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2- 6
Modem RAS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2- 7
Application (ICS only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2- 8
Technical (ICS only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2- 9
Serial (ICS only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 10
Version (ICS only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 10
Users . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 10
Monitor Type. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 10
Database . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 11
Satellite console (IES only shared database only) . . . . . . . . . . . . . . . . 2 - 11
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 13
Mail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 13
FTP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 13
Autotransfers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 13
Eventlog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 13
Backup/Restore . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 13
Licensing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 13
Escalation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 14
DICOM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 15
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 15
Character Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 17
Offline Devices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 18
Network Nodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 19
Print Devices (camera configuration)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 21
HC Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 24
LUT Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 25
HIS/RIS Nodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 26
Import/Export . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 28
Directories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 28
EPRI (Electronic Patient Record Interface) . . . . . . . . . . . . . . . . . . . . . . 2 - 29
Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 29
External Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 30
Paper Printer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 30
Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 31

Siemens AG Page 1 of 2  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
0-2 Contents

Pat. Registration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 31
Viewer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 32
CorRea (Correct & Rearrange) . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 33
MPPS (Modality Performed Procedure Steps) . . . . . . . . . . . . . . . . . . .2 - 34
Worklist Results. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 35
Completing the configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 35

3 ______ Miscellaneous _________________________________________________3 - 1

Customer UI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Camera test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Importing Reference Images . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Regional settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Time, timezone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Sinlge-Session or Multi-Session Mode for CD-Writer . . . . . . . . . . . . . . . . . 3-4

4 ______ Changes to previous version _____________________________________ 4 - 1

Sensation CT02-021.816.02 Page 2 of 2 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
*HQHUDO 1 

'HILQLWLRQVDQG$EEUHYLDWLRQV 1

$EEUHYLDWLRQ 0HDQLQJ
ACR-NEMA American College of Radiology - National Electrical
Manufacturers Association
BIOS Basic Input Output System, lowest PC configuration settings
AET Application Entity Title - Name of DICOM service participating in the
communication either as user (SCU) or provider (SCP)
CD Compact Disc
CDR Compact Disc Recordable
CT Computed Tomography
DICOM Digital Image Communication in Medicine (ACR-NEMA standard)
DHCP Dynamic Host Configuration Protocol - procedure of assigning
numeric IP addresses to nodes, in which blocks of addresses are
leased to clients on a first-come, first-serve basis. Typically, DHCP is
combined with WINS, so an IP address is both dynamically assigned
and automatically associated with a domain name. This can be com-
bined with DNS, resulting in a dynamically assigned numeric IP
address that is automatically associated with both NetBIOS and Inter-
net names
DNS Domain Name Services - address resolution procedure which associ-
ates numeric IP addresses with Internet-style names
FTP File Transfer Protocol - a client/server application that transfers files
from one computer to another
Gateway A display that controls traffic between two networks using factors
above the network layer
ICS = Navigator Image Control System - control computer for the entire CT scanner,
with interface to the acquisition system
IES = Wizard Imaging Evaluation System
IMS Imaging System, consisting of ICS, IRS and console
IRS Image Reconstruction System - reconstruction computer
LAN Local Area Network (here, always Ethernet)
LMHOSTS Address resolution procedure which statically associates numeric IP
addresses with NetBIOS names. The LMHOSTS file is checked to
see if the name is being addressed in this file
MOD Magneto-Optical Disk
NetBIOS Network Basic Input Output System - a collection of network services
and programming interfaces. NetBIOS services run on network proto-
cols such as TCP/IP
NIC Network Interface Card - an expansion card that provides a physical
connection to the network

Siemens AG Page 1 of 4 CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 *HQHUDO

$EEUHYLDWLRQ 0HDQLQJ
NT "New Technology" Operating System - trademark of Microsoft, Inc.
OI Operating Instructions
PB Patient Browser
PPP Point-to-Point Protocol - a data link and network protocol designed for
use over serial connections (such as dial-up modems)
RAS Remote Access Services - a Windows NT service that supports dial-
in and dial-out connections using modems and ISDN
Router A multiport computer or piece of dedicated hardware that examines
network layer addresses of packets on a network and retransmits
them on a different network if necessary
SCU Service Class User
SCP Service Class Provider
SRS Siemens Remote Services - a siemens service tool that supports dial-
in and dial-out connections using anlog or ISDN rooter for remote
TCP/IP Transport Control Protocol/Internet Protocol - the most widely used
standard for networking
UI User Interface
USC Uptime Service Center
WINS Windows Internet Name Service - address resolution procedure
which dynamically associates numeric IP addresses with NetBIOS
names. It is possible to combine DNS and WINS
WIP Work in Progress

Sensation  CT02-021.816.02 Page 2 of 4 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
*HQHUDO 

2YHUYLHZRI,&6,56,(6 1

,&6 1DYLJDWRU
In a standard configuration, the ICS is the main console for all scanning and imaging func-
tions.
,56
The IRS is not equipped with a keyboard, mouse or monitor.
,(6 :L]DUG
The IES console can be installed parallel to the main operating console ICS using a
shared database link. The IES can be used as a second console for post-processing
applications.

127,&( 7KH1DYLJDWRU ,&6 QHHGVWREHVZLWFKHGRQWRDFFHVVWKHVKDUHG


,&6SDWLHQWGDWDEDVHIURPWKH:L]DUG ,(6 

1DYLJDWRU
:L]DUG

,(6

:DUQLQJ 8QDXWKRUL]HGRULPSURSHURSHUDWLRQRUFKDQJHVWRWKHVRIWZDUH
FDQFDXVHV\VWHPPDOIXQFWLRQV7KLVFDQUHVXOWLQLQMXU\RUGDP
DJHWRWKHV\VWHP7KHUHIRUHWKHXVHRULQVWDOODWLRQRIVRIWZDUH
QRWDSSURYHGE\6LHPHQVLVVWULFWO\SURKLELWHG

Siemens AG Page 3 of 4  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 *HQHUDO

&RQILJXUDWLRQRYHUYLHZ 1

All configuration steps for service software functions are described in chapter Configura-
tion. Other routines for installing a newly delivered system are described in chapter Mis-
cellaneous.
.

127,&( To prevent misinterpretation as an octal number, enter all IP


addresses without leading zeros in the respective fields of the net-
work masks.

All configuration steps are identical for ICS and IES except for those in the following table.

Configuration differences between Navigator (ICS) and Wizard (IES)

Configuration Local host Application ICS only


Initial setup Technical ICS only
Version ,&6UHDGRQO\SDJH
Serial ICS only
Database Satellite console ,(6RQO\

127,&( For DGGLWLRQDODVVLVWDQFHIRUWKHFXVWRPHUVSHFLILFVRIWZDUHFRQ


ILJXUDWLRQ, see the online help (Help button in the upper left corner in
the service platform) and the "Software Release Note" included with
the software (on the Syngo platform open pull down menu "System -->
Run") or ask the application specialist or read the white colored cus-
tomer user manuals

Sensation  CT02-021.816.02 Page 4 of 4 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
&RQILJXUDWLRQ 2 

*HQHUDO 2

Within the Service software, the IXQFWLRQFRQILJXUDWLRQ allows easy setup and customiz-
ing of your system in terms of networking, DICOM, customer user interface, image data-
base, remote diagnostics, camera, and other service functions.
Configuration is performed both at the factory and during system installation.

127,&( ,IWKHVHUYLFH8,UHTXHVWVDUHVWDUWRIWKHDSSOLFDWLRQFRPSOHWHWKH
FRQILJXUDWLRQDQGVHOHFW+20(This will automatically perform a
restart. Never perform a manual restart on your own!

Caution After the restart, do not close or manipulate any of the active win-
dows (e.g. IRS installation window).

Notice To prevent misinterpretation as an octal number, enter the IP


addresses without leading zeros in the respective fields of the net-
work masks.

Initial Setup 2

The contents of the configuration frames will depend on the checkbox selections made in
the list of system options; i.e., only selected options will be displayed.
From the user interface, select 2SWLRQV/RFDO6HUYLFH

Fig. 1 Authentication page

In the Authentication page, type in the Service Password. This corresponds to the
servicekey which can be found on the "Prfprotokoll 1" (white owner manual, chapter 4).
Select the 6HWDVGHIDXOW checkbox to save the password.
Click 2.

Siemens AG Page 1 of 36  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &RQILJXUDWLRQ

Fig. 2 Home menu

In the Home menu, click the &RQILJXUDWLRQbutton.


The Initial Setup will start with the "list of system options".

List of system options

,(6RQO\

Fig. 3 List of system Options

Select/deselect the options according to your requirements.


e.g. - if using a router for RDiag, deselect 0RGHP> Modem RAS setting will not appear
e.g. - only for IES, the "0DLQ&RQVROHZLWK..." checkbox is available for shared database
e.g. - select +,65,6 only if applicable
e.g. - select 3DSHUSULQWHU if applicable

Notice If you deselect a checkbox and continue with Next, the corresponding
configuration pages and navigations will not appear.

Sensation  CT02-021.816.02 Page 2 of 36 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
&RQILJXUDWLRQ 
How to proceed
- Continue through the initial setup (using "Next" and ">") until the next configuration
mask is displayed.
- If you need more detailed information during configuration:
- Click the "enable help pages" button in the " list of system options" mask.
- When you modify settings in the corresponding mask, click the <Save> button and
confirm the information dialogs with "OK" to save the settings. Click <Finish> to cancel
without saving.
On the following pages, navigating and saving will no longer be explicitly mentioned!

Notice After saving, the service UI often requests a restart ("restart system
pending" appears in the bottom line). You do not need to restart imme-
diately. Complete the configuration steps first and then select HOME.
The restart will be performed automatically. Depending on the modifi-
cations, a restart of the application or a reboot of the system will be
performed. It may take several minutes until the system is up again -
do not interrupt the restart procedure.

Siemens AG Page 3 of 36  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &RQILJXUDWLRQ

/RFDO+RVW 2

6LWH,QIR 2

Select Configuration/Local Host -> Site Info.


Some of these data will appear in the authentication page and in the customer UI.

Fig. 4 Site Info page1

Under System, enter Department (site). (Serial No and Maintenance Unit are
preconfigured - do not change this information!)
Under Customer and Address, enter the location data .........
To allow different hospital (Institution) names, see configuration of Customer UI.
Under Service, enter Site Identification and Headquarter Phone Number.
This entry will appear in the customer UI (check it under Options/Maintenance).
Some of these data will appear in the authentication page and in the customer UI.

Notice The Maintenance Unit has to be entered as "CT".


This entry is (by mistake) used to identify the BU. Together with the
serial no., it serves as an identifier for transfers to the RDiag server.
Observe the restrictions regarding length of alphanumerics for DICOM
relevant entries (65 character limitation).

Click ">" to continue.


-------------------------------------------------------------------------------------------------------------------
If the system data entered e.g. for Site Info does not appear on the authentication page,
delete the temporary Internet pages in the browser as follows:
- Press <CTRL> and <ESC>.
- Start Internet Explorer.
- Select View/Internet options and click the Delete Files button.

Sensation  CT02-021.816.02 Page 4 of 36 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
&RQILJXUDWLRQ 
Page 2 for Site Info includes Date & Time settings.

Fig. 5 Site Info page 2

Set Timezone, Date and Time and AutoDaylightTime according to your local
requirements.

Siemens AG Page 5 of 36  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &RQILJXUDWLRQ
TCP/IP LAN "hospital network node"
2

Notice To prevent misinterpretation as an octal number, enter the IP


addresses without leading zeros in the respective fields of the net-
work masks.

Select Configuration/Local Host -> TCP/IP LAN.

Notice Ensure that the correct adapter card for the hospital LAN is selected:
Intel(R) PRO/100+ Server Adapter.
Due to Windows NT Service Pack 6 in VA40, the order of network
cards in the scroll list has changed (in VA40, the ICS LAN card is now
at the top of the selection for CT systems).

Enter the ,3DGGUHVVIRUORFDOKRVW


Delete the existing default IP address (192.168.1.1 for ICS and 192.168.1.2 for IES).
EntertheVXEQHWPDVN (if applicable).
Under gateway, enter the ,3DGGUHVVRIWKHJDWHZD\URXWHU

Notice If a router is used for RDiag, it has to be entered as gateway in the


TCP/IP LAN configuration mask.
The configuration of the router itself is described in the document
"Installation of Siemens Remote Services", CT02-021.816.03.xx.

Sensation  CT02-021.816.02 Page 6 of 36 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
&RQILJXUDWLRQ 
Modem RAS 2

Notice CT systems are usually connected to RDiag by a router.


The configuration of the modem is described in the document "Installa-
tion of Siemens Remote Services", CT02-021.816.03.xx.

Example
Select Configuration/Local Host -> Modem RAS.
Click the > button until the following page is displayed.

CT system RDIAG server

Fig. 6 TCP/IP RAS

Select the WKLVFRPSXWHURQO\ checkbox.


Select the DSUHGHWHUPLQHG,3DGGUHVV checkbox.
Under TCP/IP address pool, enter the,3DGGUHVVHVfor &7V\VWHP and 5'LDJVHUYHU

The IP addresses have to be created by the USC / TSC / HSC.


As a recommended standard, use 110 and 111 as the last three digits for a single system
in a network.
If you install a second system in your hospital, use 210 and 211, etc.

Siemens AG Page 7 of 36  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &RQILJXUDWLRQ
$SSOLFDWLRQ ,&6RQO\ 2

Select Configuration -> Local Host -> Application.


Click the > button until the following page is displayed.

Fig. 7 Application

Select the Standard View Direction for Body and Head.


In the US, deselect the 'LVSOD\&7',:.
Adjust the settings according to customer and site requirements.

Sensation  CT02-021.816.02 Page 8 of 36 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
&RQILJXUDWLRQ 
7HFKQLFDO ,&6RQO\ 2

Select Configuration -> Local Host -> Technical.


Click the > button until the following page is displayed.

Fig. 8 Technical

Enter the Osteo phantom VHULDOQXPEHU, RIIVHWandVFDOHIDFWRU(refer to document


,QVWDOODWLRQRI2SWLRQV&7[[).
Adjust the settings (ORA, Non Linear Lookup Table) according to customer
requirements. By default, ORA is selected.

Siemens AG Page 9 of 36  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &RQILJXUDWLRQ
6HULDO ,&6RQO\ 2

Select Configuration -> Local Host -> Serial.


Click the > button until the following page is displayed.
Serial is a read only page and preconfigured at the factory.
You only have to enter the UPS serial number if this option is installed in the PDC.

Fig. 9 Serial page (example)

9HUVLRQ ,&6RQO\ 2

Version is a read only page; no input is necessary.

8VHUV 2

Keep the default settings for "Users"; the default password VLHPHQV is preconfigured at
the factory.

0RQLWRU7\SH 2

Keep the default settings for Monitor Type; it is preconfigured at the factory. The standard
monitor type for CT system is Siemens (True Color).

Sensation  CT02-021.816.02 Page 10 of 36 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
&RQILJXUDWLRQ 

'DWDEDVH 2

6DWHOOLWHFRQVROH ,(6RQO\VKDUHGGDWDEDVHRQO\ 2

For configuration of the IES (Wizard) as a satellite console, the network between ICS
(Navigator) and IES (Wizard) must be connected and the ICS must be running.
In the ICS configuration, the IES (hostname, TCP/IP etc.) must be configured under
DICOM "network nodes" (see page 2-20).

Under Select Host, select "define new".


Configure the ICS (Navigator) with Host Name and TCP/IP address.

Fig. 10 Satellite console configuration of ICS

Click the VDYH button and continue by clicking !

Clicking the "Test" button checks the network connection by sending a network ping to the
ICS.

Siemens AG Page 11 of 36  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &RQILJXUDWLRQ
Select the satellite console checkbox.
Select the ICS from the "remote main console Host".
Set the Database Offset to 500.

Fig. 11 Satellite console configuration and database offset

Click the VDYH button.


The shared database (satellite console connection) will be active after rebooting the IES
(the ICS must be running during rebooting of the IES).

Sensation  CT02-021.816.02 Page 12 of 36 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
&RQILJXUDWLRQ 

6HUYLFH 2

0DLO

Notice The configuration of PDLO is described in the document "Installation of


Siemens Remote Services", CT02-021.816.03.xx.

)73 2

Notice The configuration of )73 is described in the document "Installation of


Siemens Remote Services", CT02-021.816.03.xx.

Autotransfers 2

Notice The configuration of $XWRWUDQVIHUV is described in the document


"Installation of Siemens Remote Services", CT02-021.816.03.xx.

(YHQWORJ 2

Keep the GHIDXOW settings for Eventlog.


The checkbox is selected for: 'LVSOD\6HUYLFHPHVVDJHV and 'LVSOD\VHTXHQFH
QXPEHU

%DFNXS5HVWRUH 2

Keep the GHIDXOW settings for backup/restore.

/LFHQVLQJ 2

Keep the GHIDXOWsettings for Licensing; it is preconfigured at the factory.

Siemens AG Page 13 of 36  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &RQILJXUDWLRQ
(VFDODWLRQ 2

Keep the GHIDXOW settings for Escalation.

Sensation  CT02-021.816.02 Page 14 of 36 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
&RQILJXUDWLRQ 

',&20 2

*HQHUDO 2

Fig. 12 Local AEs

Set the AETs of the CT for the HIS/RIS, Study Transfer and Print services.

EXAMPLE: Service <service-prefix>_hostname> e.g. 40001


HIS/RIS +5,B3./
Study Transfer $1B3./
Print 35B3./

For all services, the own DICOM port is by default set to 104.

1RWLFH The entries for the AE Titles are case-sensitive!

Siemens AG Page 15 of 36  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &RQILJXUDWLRQ

Fig. 13 Genral DICOM settings

Keep the GHIDXOWsettings for General.


Do not select the "implicit raw data transfer" button, otherwise a lot of data (rawdata)
will be transferred over the network which could result in network overload.

Sensation  CT02-021.816.02 Page 16 of 36 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
&RQILJXUDWLRQ 
Character Set 2

Fig. 14 Character Set

Keep the default character set,62B5


(This enables use of special characters such as the German "Umlaut".)
ISO_IR100: The 8-bit character sets are defined in ISO 8859, which is a series of more
than 10 standardized multilingual graphical character sets.

Siemens AG Page 17 of 36  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &RQILJXUDWLRQ
2IIOLQH'HYLFHV 2

Fig. 15 Offline Devices (example: CD-RW drive)

1RWLFH CD device is not supported for DICOM archiving. The CD device and
media should only be used for exporting.

Fig. 16 shows the standard &'5:GULYH configuration.


- Drive letter: 5
- Bus: 
- Target: 
- Logical Name: e.g. CD_WRITER
- Access: UHDGZULWH
- Recording speed:(refer to the type of CD-RW)
The configuration of optional MOD devices is described in the document "Installation of
Options", CT02-021.814.02.xx.

Sensation  CT02-021.816.02 Page 18 of 36 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
&RQILJXUDWLRQ 
1HWZRUN1RGHV 2

Enter all network nodes (except camera and HIS/RIS; enter these network components
under "Print Devices" for camera and "HIS/RIS nodes" for HIS/RIS).

Fig. 16 Network Nodes ( define Hostproperties)

Select Host -> define new.


For Host Name, enter the KRVWQDPH
For TCP/IP address, enter the 7&3,3DGGUHVV
Save and continue by clicking !.

Siemens AG Page 19 of 36  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &RQILJXUDWLRQ

Fig. 17 Network Nodes (define network node-specific data)

Enter the ORJLFDOQDPHin the "HGLW1DPH" box.


(This name appears in the "transfer" pull-down menu in the browser.)
Select the 6\QJREDVHG checkbox if a syngo-based system is connected.
Select the KRVWto be configured.
Enter the $(WLWOH in the $(7LWOH box.
Enter the SRUWQXPEHU in the 32571XPEHU box.
Ask the network administrator responsible which DICOM services will be supported by
the configured network node and select the required ',&20VHUYLFHV
(Storage, Storage Commitment, Query, Retrieve).
Ask for the settings (syntax, compression, etc.) for the connected devices.
For additional assistance, see the online help (Help button in the upper left corner, see
Fig. 18) and the "Software Release Note" included with the software.
Click the DGG button.
Click the 9HULILFDWLRQbutton to test the DICOM connection.
Click the VDYH button.

For additional nodes / DICOM services, repeat the complete "Network Nodes" configura-
tion.

Sensation  CT02-021.816.02 Page 20 of 36 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
&RQILJXUDWLRQ 
Print Devices (camera configuration)
2

Notice If autofilming is active only:


Everytime the hostname of the WIZARD (IES) has been changed, the
following comand "perl c:\Somaris\utils\determinewizard.pl" must be
performed on the NAVIGATOR (ICS) in the comand prompt window
(select Run... in the Windows Start Menu and enter the comand).

For details on camera installation and autofilming for all released cameras, refer to the
"Software Release Note" included with the software or to the Siemens TD Intranet pages
(Product Information -> CAMERA). (General CT02-021.814.04.xx.)

Fig. 18 Print Devices (define host properties)

Select Host -> define new.


For Host Name, enter the KRVWQDPH
For TCP/IP address, enter the 7&3,3DGGUHVV
Save and continue by clicking !.

Siemens AG Page 21 of 36  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &RQILJXUDWLRQ

Fig. 19 Print Devices (define logical name, DICOM services and port)

Enter the ORJLFDOQDPH in the "/RJLFDOQDPH" box.


Select the KRVW to be configured.
Enter the $(WLWOH in the $(7LWOH box.
Enter the SRUWQXPEHU in the 32571XPEHU box.
For ',&20VHUYLFHV select SULQWonly.
Click the DGG button.
Click the 9HULILFDWLRQbutton to test the DICOM connection.
Click the VDYH button and continue by clicking the ! button.

Sensation  CT02-021.816.02 Page 22 of 36 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
&RQILJXUDWLRQ 

Fig. 20 camera specific

Select HC device -> define new.


Enter the +&GHYLFHQDPH in the "+&'HYLFH" box.
(This name will appear in the Filming card after a reboot.)
Select "'HIDXOW&DPHUD" if you want this camera to be listed first.
For ',&207\SH, select ',&203ULQWHU
For &ODVV, select one of the predefined cameras for the available camera.
Enter ORFDWLRQ and FRPPHQW (optional).
+ROGSULQWHGILOPMRE (default = 3; max = 10) defines the number of films that can be
reprinted.
Delete the )LOPVKHHW)RUPDWV3RUWUDLW not supported by the camera.
Delete the )LOPVKHHW)RUPDWV/DQGVFDSHnot supported by the camera.
0HGLXPW\SH (default = blue film)!see the information about the respective camera in
the appendix.
For )LOPGHVWLQDWLRQ, select SURFHVVRU (default).
Set the "Filming Properties" values forFRORUDSSHDUDQFH (monochrome),EDFNJURXQG
(black), and 7UDQVIRUPDWLRQ (no magnification).
Click the VDYH button and wait for the message GHYLFHSURSHUWLHVVXFFHVVIXOO\
VDYHG.
Repeat all steps for "Print Devices" to configure additional cameras.

Siemens AG Page 23 of 36  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &RQILJXUDWLRQ
HC Overview 2

Fig. 21 hc overview

Under +&2YHUYLHZ, check if the camera configuration was successful.

Sensation  CT02-021.816.02 Page 24 of 36 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
&RQILJXUDWLRQ 
/87)LOHV 2

For importing external LUT files (e.g., from floppy disk).

For LUT definitions, refer to the specific hardcopy camera information on the Siemens TD
Intranet pages or the Release Note VA40.
Default for Wet cameras: set the CT Image and SC Image to CTLUT and 2.80 for maxi-
mum density.
Default for Dry cameras: set the CT Image and SC Image to CTLUTdry and 2.80 for max-
imum density.

Siemens AG Page 25 of 36  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &RQILJXUDWLRQ
+,65,61RGHV 2

Select Host -> define new.


For Host Name, enter the KRVWQDPH
For TCP/IP address, enter the 7&3,3DGGUHVV
Save and continue by clicking !.

Sensation  CT02-021.816.02 Page 26 of 36 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
&RQILJXUDWLRQ 

Enter the ORJLFDOQDPH in the "/RJLFDOQDPH" box.


Select the KRVW to be configured.
Enter the $(WLWOH in the $(7LWOH box.
Enter the SRUWQXPEHU in the 32571XPEHU box.
Select Basic Worklist for "DICOM services" and adjust the settings to the
requirements on site (HIS/RIS) and by the customer.
Please ask the HIS/RIS technician responsible if the HIS/RIS system support MPPS
(Modality Performed Procedure Steps). If yes, select "Performed Procedure Step".

Attention If "Performed Procedure Step" is selected without the proper support,


database problems can occur (patients can no longer be deleted from
the database).

Click the DGG button.


Click the 9HULILFDWLRQbutton to test the DICOM connection.
Click the VDYH button.

Siemens AG Page 27 of 36  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &RQILJXUDWLRQ

,PSRUW([SRUW 2

'LUHFWRULHV 2

Adjust the settings according to customer and site requirements. Default directories
have been preconfigured.

Sensation  CT02-021.816.02 Page 28 of 36 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
&RQILJXUDWLRQ 

(35, (OHFWURQLF3DWLHQW5HFRUG,QWHUIDFH 2

6HUYHU 2

Fig. 22 TCP/IP LAN ICS-IRS

Adjust the settings according to customer and site requirements.

Siemens AG Page 29 of 36  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &RQILJXUDWLRQ

External Devices 2

Paper Printer 2

Fig. 23 Paper Printer

Adjust the settings according to customer and site requirements.

Sensation  CT02-021.816.02 Page 30 of 36 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
&RQILJXUDWLRQ 

Applications 2

3DW5HJLVWUDWLRQ 2

Fig. 24 Pat. Registration

Adjust the settings according to customer requirements.

Siemens AG Page 31 of 36  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &RQILJXUDWLRQ
Viewer 2

Allows you to configure the viewer for loading a new study or series.

Fig. 25 Viewer

Adjust the settings according to customer requirements.

Sensation  CT02-021.816.02 Page 32 of 36 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
&RQILJXUDWLRQ 
CorRea (Correct & Rearrange) 2

Allows you to set the attributes that can be modified in the Correct & Rearrange platform.

Fig. 26 Correct & Rearrange

Adjust the settings according to customer and site requirements.

Siemens AG Page 33 of 36  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &RQILJXUDWLRQ
MPPS (Modality Performed Procedure Steps) 2

Allows you to set the attributes to be displayed in the customer UI when a MPPS message
is sent to the RIS.
.

Fig. 27 Modality Performed Procedure Steps

Adjust the settings according to customer and site requirements.

Sensation  CT02-021.816.02 Page 34 of 36 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
&RQILJXUDWLRQ 
Worklist Results 2

Allows you to set the attributes displayed after a DICOM worklist request.
.

Fig. 28 Worklist Results

Adjust the settings according to customer and site requirements.

&RPSOHWLQJWKHFRQILJXUDWLRQ 2

After going through all configuration masks, click +RPH and confirm the reboot.

 1RWHWKDWWKHUHERRWWDNHVVRPHWLPH

Siemens AG Page 35 of 36  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &RQILJXUDWLRQ
This page intentionally left blank.

Sensation  CT02-021.816.02 Page 36 of 36 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
0LVFHOODQHRXV 3 

&XVWRPHU8, 3

The defaults for the Customer UI are preset during software installation at the factory.
Perform changes regarding the Customer UI on request of the customer.

You can make the Institution Name available in the Patient Registration by selecting the
Institution name checkbox. This allows entering different hospital names, e.g. if several
doctors are sharing the CT system.
Select Options/Configuration/Patient Registration to enter defaults for the institution
name.

Siemens AG Page 1 of 4  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 0LVFHOODQHRXV
Page 2 for configuring the institution name

Enter the ,QVWLWXWLRQ1DPH according to the customer requirements.

Sensation  CT02-021.816.02 Page 2 of 4 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
0LVFHOODQHRXV 

&DPHUDWHVW 3

Instruct the local camera service to set the camera to linear LUT for our camera PRINT
AET (PR_<hostname>).
Set the CT Image to OLQHDUin the LUT settings platform.
Print the reference image with the linear grayscale bar (located under
c:\Somaris\service\Reference_images) to the camera using the filming task card.
Use the densometer to check the printed linear grayscale bar on the film sheet for linear
density. If it passes, the camera responds with a linear look-up table for our PRINT AET.
Switch the LUT for CT Image back to CTLUT in the configuration dialog under the LUT
settings platform.

,PSRUWDQW 3OHDVHPDNHVXUHWRSHUIRUPWKHDERYHWHVW
An incorrect AE Title in the camera does not necessarily result in a
completely incorrect camera printout. Usually, the camera sets a
default behavior for an unknown AE title. This means that images are
printed, but they are not of the desired quality.

*HQHUDO,QIRUPDWLRQIRU6XSSRUWHG&DPHUDV
See the product information on our Intranet page at http://www-td.med.siemens.de.

1RWH ,I\RXLQVWDOODFDPHUDZKLFKLVQRWRQWKHOLVWRIVXSSRUWHGFDP
HUDV\RXPXVWXVHDWHVWLPDJHGLVSOD\LQJD'REMHFWRINQRZQ
VL]HWRFRQILUPWKDWWKHUHDOVL]HLVFRUUHFWZLWKHDFKVXSSRUWHG
ILOPVL]H

,PSRUWLQJ5HIHUHQFH,PDJHV 3

Perform the following steps:


From the syngo user interface, select 7UDQVIHU!,PSRUWIURP2II/LQH
From the drop-down list, select the path "c:\Somaris\service\ReferenceImages\*".
The images are now available in the patient browser.

Siemens AG Page 3 of 4  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 0LVFHOODQHRXV

5HJLRQDOVHWWLQJV 3

From the syngo user interface, select 2SWLRQV&RQILJXUDWLRQ


Select Regional Settings.

7LPHWLPH]RQH 3

Select the 'DWH 7LPH tab and set the correct local time.
Select the 7LPH=RQH tab and set the correct time zone.

.H\ERDUG 3

Select the .H\ERDUG tab and set the available keyboard type.

/DQJXDJH 3

Select the /DQJXDJHtab and set the required language.


The following native languages are supported: English, French, German and Spanish.

1RWH The regional settings, date & time, keyboard and language are not
stored in the backup set. Therefore, the regional settings have to be
reconfigured after each SW installation.

6LQOJH6HVVLRQRU0XOWL6HVVLRQ0RGHIRU&':ULWHU 3

Thhe standard mode for the CD-Writer is defined as Muli-Session Mode.


If the Single-Session Mode is required by customer, doubleclick the file
c:\Somaris\utils?VLQJOHBVHVVLRQUHJwithin the Windows NT Explorer

1RWH This setting is not contained in the backup and has therefore to be
repeted after each SW installation.

To re-configuret the standard Multi-Session Mode, doubleclick the file


c:\Somaris\utils\PXOWLBVHVVLRQUHJwithin the Windows NT Explorer

Sensation  CT02-021.816.02 Page 4 of 4 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions
Changes to previous version 4 

&KDSWHU 3DJH &KDQJH 5HDVRQ


Some editorial changes:
-No. 107826,
all all - Changed confiuration paramater for VA40B
VA40B for P20
- CD_RW information single/multi-session

Siemens AG Page 1 of 2  CT02-021.816.02 Sensation


Medical Solutions CTL PM Rev. 02 11.01
 &KDQJHVWRSUHYLRXVYHUVLRQ
This page intentionally left blank.

Sensation  CT02-021.816.02 Page 2 of 2 Siemens AG


Rev. 02 11.01 CTL PM Medical Solutions

Das könnte Ihnen auch gefallen