Sie sind auf Seite 1von 1

Application

Analytical Quality Assurance,


Standard for Nonionic Surfactants
Preparation of a standard solution for nonionic surfactants

Reagents:
Cat.No. 112298 Triton X-100
Cat.No. 116754 Water for analysis

Preparation of a standard solution:


Dissolve 1.00 g of Triton X-100 with distilled water in a calibrated or conformity-checked 1000-ml volumetric flask
and make up to the mark with distilled water. The stock solution prepared according to this procedure has a
concentration of 1000 mg/l non-ionic surfactants.

Further investigational concentrations may be prepared from this stock solution by diluting accordingly.

Stability:
When stored in a cool place (refrigerator), the stock solution remains stable for one month. The diluted
investigational solutions must be used immediately.

For use with Measuring range


test kits: in mg/l
101787 Spectroquant Surfactants (nonion.) Cell Test 0.10 - 7.50

Ordering Information
Product Catalog No.
Triton X-100 for gas chromatography 112298
Water for analysis EMSURE 116754
Surfactants (nonion.) Cell Test Method: photometric 0.10 - 7.50 mg/l Triton X-100 101787
Spectroquant

Merck KGaA, 64271 Darmstadt, Germany, Tel. 2012-08-30 Page 1 of 1