2240751
Revision 0
Operating Documentation
All packages should be closely examined at time of delivery. If damage is
apparent, have notation of bad order placed by the delivering driver on
all copies of the freight or express bill. If damage is of a concealed nature,
notify transportation agent as soon as possible to make an inspection report
of damage but in any event not later than 15 days after delivery. A trans-
portation company usually will not pay a claim for concealed damage if an
inspection is not requested within this 15 day period. Complete instructions
regarding claim procedure are found in section S of the Policy & Proce-
dure Bulletins.
(DO NOT TEAR) FOLD HERE AND SEAL (DO NOT TEAR)
FOLD FOLD
IF ADDITIONAL PAGES ARE INCLUDED, FOLD AND STAPLE TO LOWER PART OF THIS
SHEET BEFORE MAILING.
PHONE
SERVICE REP FAX
COUNTRY
CITY
P9030CB REV2
US Advanced Document Group Of GEYMS
FAX Number: 81425855950
REV. 4
GE MEDICAL SYSTEMS
P9030CD REV. 0
GE MEDICAL SYSTEMS
PAGE REV PAGE REV PAGE REV PAGE REV PAGE REV
Title page . . . . . . . . . . 0 P9030CC . . . . . . . . . . 4 1 to 12 . . . . . . . . . . . . . 0
GE Logo page . . . . . . P9030CD . . . . . . . . . . 0
P9030CB . . . . . . . . . . . 2 A .................. 0
A
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751
1. INTRODUCTION
These Instructions describe the installation of the optional Realtime Doppler Calculation Software for LOGIQ 400
system.
This installation enables the LOGIQ 400 system with the system software version 4.01 y or later (4.02 y or later
for CL console) to use the Realtime Doppler Calculation software.
2. TOOLS REQUIRED
Standard field Service Tool kit
LOGIQ 400 Service Manual, 2127661
LOGIQ 400 Proprietary Manual, 2127662
3. TIME REQUIRED
One (1) hour plus travel time
4. PARTS SUPPLIED
ITEM PART NAME QTY PART # REMARKS
LOGIQ 400 Realtime Doppler Calculation 1 2241560
Software Option Kit
1. Software Certification Sheet 1 2137225100
2. Option Label 1 P9524PZ
3. Upgrade Instructions 1 2240751
5. PREREQUISITE
1. The system software must be version 4.01y or later (4.02y or later for CL console).
Note
The Version 4.01 Upgrade kits (2239784 for Japan, 2239786 for Asia, and 2239785 for USA and Eu-
rope) upgrades the console to the system software version 4.01y from the version 3.40By or earlier.
The Version 4.02Upgrade kits (2239820 for Japan, 2239822 for Asia, and 2239821 for USA and Eu-
rope) upgrades the console to the system software version 4.02y from the version 3.41By or earlier.
1
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751
5. PREREQUISITE(continued)
2. The MR3 Upgrade Kit (2241566) must be installed.
Note
The optional MR3 Upgrade Kit, 2241566, upgrades two board assemblies to the appropriate ones
(DOPC2 Assy, and ASPR3 Assy) for using the Realtime Doppler Calculation software option on the
LOGIQ 400 system.
The software version must be upgraded to 4.01 y prior to the upgrade using the MR3 upgrade kit.
6. PROCEDURE
READ and UNDERSTAND these instructions thoroughly before proceeding with the installation. Perform each step in
sequence and check it off when completed. If a problem occurs after completion, go back through the procedure and
check for implementation errors before contacting your Ultrasound Region Engineer.
61 Functional Checkout
1. Plug in the LOGIQ 400 system and power it ON.
2. Make sure that the part number of the OBC Assy is 2195747 or 21957472. The part number of the OBC Assy
currently installed is displayed on the CRT monitor screen during power-up sequence as follows:
Note
The part numbers, 2195747 and 21957472 indicate that the MPUA Assy is installed in the con-
sole. The MPU Assy is upgraded to the MPUA Assy with the OBCA Assy and the ADPT2 Assy
supplied with the OBCA Assy Option (2219181).
2
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751
Note
Software version 4.02 y or later for the LOGIQ 400CL console.
4. Make sure that the MR3 is indicated on the CRT monitor screen during the power-up sequence as shown in
ILLUSTRATION 2.
Note
The MR3 is not displayed on the CRT monitor screen for the LOGIQ 400CL console.
5. Press the New Patient key to exit the patient entry mode when the PATIENT ENTRY MENU is displayed on the
CRT monitor screen.
6. Make sure that one or more probes are connected with the console.
7. Press the Top Menu Select key twice, located on the keyboard. Refer to ILLUSTRATION 3.
Note
The items displayed on the Top Menu are different depending on the component options
3
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751
TOP MENU
ILLUSTRATION 4
8. Select the Set Up Menu using the or key of Sub Menu Select keys on the keyboard. The Set Up selec-
tion is displayed in reverse video as shown in ILLUSTRATION 4 on page 4.
9. Press the key to select the Set Up Menu. The Set Up Menu is displayed on the CRT monitor screen as
shown in ILLUSTRATION 5.
SET UP MENU
ILLUSTRATION 5
The Utility Menu will be displayed on the CRT monitor screen as shown in ILLUSTRATION 6.
UTILITY MENU
ILLUSTRATION 6
4
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751
Note
Press and hold the Ctrl key and then press F to go to the next page. Press and hold the Ctrl key and
then press B to return to the previous page.
11. Make sure that the versions of the following two board assemblies are as shown in table below:
TABLE 1
VERSION OF BOARD ASSEMBLIES
12. Press and hold the Ctrl key and then press the R key twice to exit the Utility Menu.
13. Perform functional checks to verify that the system is functioning properly. Refer to Functional Checks in Chapter 4
of the LOGIQ 400 Service Manual, 2127661.
5
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751
62 Activating Software
1. Enter the service software. Refer to LOGIQ 400 Proprietary Manual, 2127662.
The Service Menu will be displayed on the CRT monitor screen as shown in ILLUSTRATION 5.
SERVICE MENU
ILLUSTRATION 8
2. Type 02 (or 2) and press the Return key to select the Option 2 Option Set when the Service Menu is displayed.
The Option Set menu will be displayed as shown in ILLUSTRATION 9.
6
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751
4. Type 34 and press Return to select the Realtime Doppler Calculation software option. The message 1:En-
able/0:Disable ? will be displayed.
5. Type 1 and press Return to activate the Realtime Doppler Calculation option. The status display of the Realtime
Doppler Calculation will change to Installed.
6. Press and hold the Ctrl key and press R to return to Service Menu.
8. Type Y and press Return. The system will start the system reset process.
7
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751
63 System Check-out
1. Press the New Patient key to exit the patient entry mode when the PATIENT ENTRY MENU is displayed on the
CRT monitor screen.
2. Make sure that the one or more probes are connected with console.
4. Press the Top Menu Select key located on the keyboard. Refer to ILLUSTRATION 3 on page 3.
5. Press the or key of Sub Menu Select keys simultaneously to display page 4 of four measurement menu
pages. Refer to ILLUSTRATION 12.
6. Make sure that the CALC selection is indicated when changing the setting using the or key of Sub
Menu Select keys under the PWD section. Refer to ILLUSTRATION 12.
Note
The CALC selection is displayed only when the optional Realtime Doppler Calculation software is
set to ON.
8
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751
CALCULATION ITEMS
ILLUSTRATION 13
Note
The Realtime Doppler Calculation is invalid when the Sweep Speed is set to the Very Fast.
9
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751
64 Final Procedures
1. Power the LOGIQ 400 system OFF.
2. Remove four screw caps (covering the screws 3 and 4 ) and unscrew four screws ( 1 4 ). Refer to
ILLUSTRATION 14.
4. Attach the Option Label, which indicates new version number of system software, onto the Nest Cover. Refer to
ILLUSTRATION 15.
NEST COVER
ILLUSTRATION 15
10
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751
6. Fill out the proper customer information on the Product Locator Card. Refer to ILLUSTRATION 16. Mail the
installation card Product Locator to the address corresponding to your region.
Note
The Product Locator Card indicated in ILLUSTRATION 16 may be different from the one actually
provided.
7. Hand the Software Certification Sheet supplied with this kit to the customer and ask the customer to keep it.
9. This completes the upgrade of the optional LOGIQ 400 Realtime Doppler Calculation software.
11
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751