2240751
Revision 0

Realtime Doppler Calculation


Software Option
Upgrade Instructions
LOGIQ 400

Copyright 1999 by General Electric Company

Operating Documentation
All packages should be closely examined at time of delivery. If damage is
apparent, have notation of bad order placed by the delivering driver on
all copies of the freight or express bill. If damage is of a concealed nature,
notify transportation agent as soon as possible to make an inspection report
of damage but in any event not later than 15 days after delivery. A trans-
portation company usually will not pay a claim for concealed damage if an
inspection is not requested within this 15 day period. Complete instructions
regarding claim procedure are found in section S of the Policy & Proce-
dure Bulletins.

If shipment was handled by moving van service - uncrated - call Traffic -


Milwaukee immediately when any damage is found. Do not attempt to call
any local agent. At this time be ready to describe type of damage, type of
equipment, serial numbers and if possible the order number.

The above paragraph is in regard to equipment requiring installation only,


and does not apply to supply items. The F.O.B. point for these items is as
shown in the Price Book.

GE Medical Systems: Telex 3797371


P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
(Asia, Pacific, Latin America, North America)

GE Medical Systems Europe: Telex 698626


283, rue de la Miniere, B.P. 34, 78533 Buc Cedex
France
GE YOKOGAWA MEDICAL SYSTEMS, LTD.
SERVICE ENGINEERING
TECHINICAL SUPPORT US
7127, Asahigaoka 4Chome
Hinoshi, Tokyo, 191 JAPAN
Stick along this line

(DO NOT TEAR) FOLD HERE AND SEAL (DO NOT TEAR)

REPORT ON TECHNICAL PUBLICATION RT


(OMISSIONS,ERRORS,SUGGESTIONS)
PRODUCT PUBLICATION #
PAGE # PAGE REV DATA REPORTED
SERVICE RECORD SUBMITTED YES NO
PLEASE BE SPECIFIC IN YOUR CORRECTIONS AND SUGGESTIONS

FOLD FOLD

IF ADDITIONAL PAGES ARE INCLUDED, FOLD AND STAPLE TO LOWER PART OF THIS
SHEET BEFORE MAILING.
PHONE
SERVICE REP FAX
COUNTRY
CITY

P9030CB REV2
US Advanced Document Group Of GEYMS
FAX Number: 81425855950

REV. 4
GE MEDICAL SYSTEMS

 THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


WARNING  IF A CUSTOMERS SERVICE PROVIDER REQUIRES A LANGUAGE OTHER
THAN ENGLISH, IT IS THE CUSTOMERS RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
 DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
 FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK,
MECHANICAL OR OTHER HAZARDS.

 CE MANUEL DE MAINTENANCE NEST DISPONIBLE QUEN ANGLAIS.


AVERTISSEMENT  SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE LANGLAIS, CEST AU CLIENT QUIL INCOMBE DE LE FAIRE
TRADUIRE.
 NE PAS TENTER DINTERVENTION SUR LES QUIPEMENTS TANT QUE LE
MANUEL SERVICE NA PAS T CONSULT ET COMPRIS.
 LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRANER CHEZ LE
TECHNICIEN, LOPRATEUR OU LE PATIENT DES BLESSURES DUES DES
DANGERS LECTRIQUES, MCANIQUES OU AUTRES.

 DIESES KUNDENDIENSTHANDBUCH EXISTIERT NUR IN


ENGLISCHER SPRACHE.
WARNUNG
 FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENTIGT,
IST ES AUFGABE DES KUNDEN FR EINE ENTSPRECHENDE BERSETZUNG
ZU SORGEN.
 VERSUCHEN SIE NICHT, DAS GERT ZU REPARIEREN, BEVOR DIESES
KUNDENDIENSTHANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN
WURDE.
 WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLGE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.

 ESTE MANUAL DE SERVICIO SLO EXISTE EN INGLS.


AVISO  SI ALGN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA
QUE NO SEA EL INGLS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
SERVICIO DE TRADUCCIN.
 NO SE DEBER DAR SERVICIO TCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
 LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELCTRICAS, MECNICAS O DE
OTRA NATURALEZA.

P9030CD REV. 0
GE MEDICAL SYSTEMS

 ESTE MANUAL DE ASSISTNCIA TCNICA S SE ENCONTRA


DISPONVEL EM INGLS.
ATENO
 SE QUALQUER OUTRO SERVIO DE ASSISTNCIA TCNICA, QUE NO A
GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIOS DE TRADUO.
 NO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTNCIA TCNICA.
 O NO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANA
DO TCNICO, OPERADOR OU PACIENTE DEVIDO A CHOQUES ELTRICOS,
MECNICOS OU OUTROS.

 IL PRESENTE MANUALE DI MANUTENZIONE DISPONIBILE


SOLTANTO IN INGLESE.
AVVERTENZA
 SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
 SI PROCEDA ALLA MANUTENZIONE DELLAPPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
 NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALLADDETTO ALLA
MANUTENZIONE, ALLUTILIZZATORE ED AL PAZIENTE PER
FOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751

LIST OF EFFECTIVE PAGES

REV DATE PRIMARY REASON FOR CHANGE


0 . . . . . April 23, 1999 . . . Initial release

PAGE REV PAGE REV PAGE REV PAGE REV PAGE REV
Title page . . . . . . . . . . 0 P9030CC . . . . . . . . . . 4 1 to 12 . . . . . . . . . . . . . 0
GE Logo page . . . . . . P9030CD . . . . . . . . . . 0
P9030CB . . . . . . . . . . . 2 A .................. 0

A
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751

Realtime Doppler Calculation Software Option


Upgrade Instructions

1. INTRODUCTION
These Instructions describe the installation of the optional Realtime Doppler Calculation Software for LOGIQ 400
system.

This installation enables the LOGIQ 400 system with the system software version 4.01 y or later (4.02 y or later
for CL console) to use the Realtime Doppler Calculation software.

2. TOOLS REQUIRED
Standard field Service Tool kit
LOGIQ 400 Service Manual, 2127661
LOGIQ 400 Proprietary Manual, 2127662

3. TIME REQUIRED
One (1) hour plus travel time

4. PARTS SUPPLIED
ITEM PART NAME QTY PART # REMARKS
LOGIQ 400 Realtime Doppler Calculation 1 2241560
Software Option Kit
1. Software Certification Sheet 1 2137225100
2. Option Label 1 P9524PZ
3. Upgrade Instructions 1 2240751

5. PREREQUISITE
1. The system software must be version 4.01y or later (4.02y or later for CL console).

Note
The Version 4.01 Upgrade kits (2239784 for Japan, 2239786 for Asia, and 2239785 for USA and Eu-
rope) upgrades the console to the system software version 4.01y from the version 3.40By or earlier.
The Version 4.02Upgrade kits (2239820 for Japan, 2239822 for Asia, and 2239821 for USA and Eu-
rope) upgrades the console to the system software version 4.02y from the version 3.41By or earlier.

1
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751

5. PREREQUISITE(continued)
2. The MR3 Upgrade Kit (2241566) must be installed.

Note
The optional MR3 Upgrade Kit, 2241566, upgrades two board assemblies to the appropriate ones
(DOPC2 Assy, and ASPR3 Assy) for using the Realtime Doppler Calculation software option on the
LOGIQ 400 system.
The software version must be upgraded to 4.01 y prior to the upgrade using the MR3 upgrade kit.

6. PROCEDURE
READ and UNDERSTAND these instructions thoroughly before proceeding with the installation. Perform each step in
sequence and check it off when completed. If a problem occurs after completion, go back through the procedure and
check for implementation errors before contacting your Ultrasound Region Engineer.

Check each step as it is completed.

61 Functional Checkout
1. Plug in the LOGIQ 400 system and power it ON.

2. Make sure that the part number of the OBC Assy is 2195747 or 21957472. The part number of the OBC Assy
currently installed is displayed on the CRT monitor screen during power-up sequence as follows:

PART NUMBER OF OBC ASSY


ILLUSTRATION 1

Note
The part numbers, 2195747 and 21957472 indicate that the MPUA Assy is installed in the con-
sole. The MPU Assy is upgraded to the MPUA Assy with the OBCA Assy and the ADPT2 Assy
supplied with the OBCA Assy Option (2219181).

2
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751

61 Functional Checkout (continued)


3. Make sure that the system software version 4.01 y or later (4.02y or later for the CL console) is indicated on the
CRT monitor screen during the power-up sequence as follows:

SOFTWARE VERSION DISPLAY


ILLUSTRATION 2

Note
Software version 4.02 y or later for the LOGIQ 400CL console.

4. Make sure that the MR3 is indicated on the CRT monitor screen during the power-up sequence as shown in
ILLUSTRATION 2.

Note
The MR3 is not displayed on the CRT monitor screen for the LOGIQ 400CL console.

5. Press the New Patient key to exit the patient entry mode when the PATIENT ENTRY MENU is displayed on the
CRT monitor screen.

6. Make sure that one or more probes are connected with the console.

7. Press the Top Menu Select key twice, located on the keyboard. Refer to ILLUSTRATION 3.

SOFTWARE MENU CONTROL KEYS


ILLUSTRATION 3

Note
The items displayed on the Top Menu are different depending on the component options

3
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751

61 Functional Checkout (continued)


Then top menu is displayed on CRT monitor screen as shown in ILLUSTRATION 4.

TOP MENU
ILLUSTRATION 4

8. Select the Set Up Menu using the or key of Sub Menu Select keys on the keyboard. The Set Up selec-
tion is displayed in reverse video as shown in ILLUSTRATION 4 on page 4.

9. Press the key to select the Set Up Menu. The Set Up Menu is displayed on the CRT monitor screen as
shown in ILLUSTRATION 5.

SET UP MENU
ILLUSTRATION 5

The Utility Menu will be displayed on the CRT monitor screen as shown in ILLUSTRATION 6.

UTILITY MENU
ILLUSTRATION 6

4
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751

61 Functional Checkout (continued)


10. Type 06 (or 6) and press the Return key to select the Option 6 Board Configuration Display when the Utility
Menu is displayed. The Board Configuration tables will be displayed as shown in ILLUSTRATION 7.

BOARD CONFIGURATION DISPLAY


ILLUSTRATION 7

Note
Press and hold the Ctrl key and then press F to go to the next page. Press and hold the Ctrl key and
then press B to return to the previous page.

11. Make sure that the versions of the following two board assemblies are as shown in table below:

TABLE 1
VERSION OF BOARD ASSEMBLIES

NO BOARD NAME VERSION


13 DOPC 8
15 ASPR d

12. Press and hold the Ctrl key and then press the R key twice to exit the Utility Menu.

13. Perform functional checks to verify that the system is functioning properly. Refer to Functional Checks in Chapter 4
of the LOGIQ 400 Service Manual, 2127661.

5
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751

62 Activating Software
1. Enter the service software. Refer to LOGIQ 400 Proprietary Manual, 2127662.
The Service Menu will be displayed on the CRT monitor screen as shown in ILLUSTRATION 5.

SERVICE MENU
ILLUSTRATION 8

2. Type 02 (or 2) and press the Return key to select the Option 2 Option Set when the Service Menu is displayed.
The Option Set menu will be displayed as shown in ILLUSTRATION 9.

OPTION SET MENU (a)


ILLUSTRATION 9

6
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751

62 Activating Software (continued)


3. Press and hold the Ctrl key and press F to go to next page.

OPTION SET MENU (b)


ILLUSTRATION 10

4. Type 34 and press Return to select the Realtime Doppler Calculation software option. The message 1:En-
able/0:Disable ? will be displayed.

5. Type 1 and press Return to activate the Realtime Doppler Calculation option. The status display of the Realtime
Doppler Calculation will change to Installed.

6. Press and hold the Ctrl key and press R to return to Service Menu.

7. Select Option 09 System Reset and press Return.

8. Type Y and press Return. The system will start the system reset process.

7
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751

63 System Check-out
1. Press the New Patient key to exit the patient entry mode when the PATIENT ENTRY MENU is displayed on the
CRT monitor screen.

2. Make sure that the one or more probes are connected with console.

3. Press the PD key to select the PWD mode.

4. Press the Top Menu Select key located on the keyboard. Refer to ILLUSTRATION 3 on page 3.

The PWD menu is displayed on the CRT monitor screen.

SOFTMENU ON VFD (b)


ILLUSTRATION 11

5. Press the or key of Sub Menu Select keys simultaneously to display page 4 of four measurement menu
pages. Refer to ILLUSTRATION 12.

SOFT MENU ON VFD (c)


ILLUSTRATION 12

6. Make sure that the CALC selection is indicated when changing the setting using the or key of Sub
Menu Select keys under the PWD section. Refer to ILLUSTRATION 12.

Note
The CALC selection is displayed only when the optional Realtime Doppler Calculation software is
set to ON.

8
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751

63 System Check-out (continued)


7. Make sure that the calculation items is displayed on the CRT monitor screen as shown in ILLUSTRATION 13
when the CALC is chosen.

CALCULATION ITEMS
ILLUSTRATION 13

Note
The Realtime Doppler Calculation is invalid when the Sweep Speed is set to the Very Fast.

9
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751

64 Final Procedures
1. Power the LOGIQ 400 system OFF.

2. Remove four screw caps (covering the screws 3 and 4 ) and unscrew four screws ( 1 4 ). Refer to
ILLUSTRATION 14.

3. Remove the left side cover. Refer to ILLUSTRATION 14.

SIDE COVER LEFT


LEFT SIDE COVER REMOVAL
ILLUSTRATION 14

4. Attach the Option Label, which indicates new version number of system software, onto the Nest Cover. Refer to
ILLUSTRATION 15.

NEST COVER
ILLUSTRATION 15

10
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751

64 Final Procedures (continued)


5. Reinstall the left side cover. Refer to ILLUSTRATION 14.

6. Fill out the proper customer information on the Product Locator Card. Refer to ILLUSTRATION 16. Mail the
installation card Product Locator to the address corresponding to your region.

PRODUCT LOCATOR CARD


ILLUSTRATION 16

Note
The Product Locator Card indicated in ILLUSTRATION 16 may be different from the one actually
provided.

7. Hand the Software Certification Sheet supplied with this kit to the customer and ask the customer to keep it.

8. Properly dispose of excess material.

9. This completes the upgrade of the optional LOGIQ 400 Realtime Doppler Calculation software.

11
LOGIQ 400 Realtime Doppler Calculation Software Option
GE MEDICAL SYSTEMS Upgrade Instructions
REV 0 2240751