Sie sind auf Seite 1von 7

4.

8 Miscellaneous and Discontinued Sensors


T. J. CLAGGETT, R. W. WORRALL (1969, 1982) B. G. LIPTK (1995, 2003)

J. E. JAMISON (2003), REVIEWED BY K. C. SLONEKER

Pneumatic and Suction Pyrometers


Applications: Temperature profiling of boilers and furnaces during commissioning

Length of Probes: 6 to 30 ft (2 to 10 m); the permissible insertion length is a function of the maximum
operating temperature.

Inaccuracy: 2% of actual temperature

Response: 98% in 10 to 20 s

Probe: Stainless steel water-jacketed units with 2 to 3 in. (50 to 75 mm) diameters

Temperature Range: 0 to 2900F (0 to 1600C) with thermocouple and 0 to 4500F (0 to 2500C) with
pneumatic probes

Fan or ejector requirements: The fan must generate about 20 in. of water column (0.05 bar, or 37 mmHg) vacuum
and pull a flow rate of about 40 cfm (2000 lpm) of furnace gas.

Former Supplier: Land Combustion, a Division of Land Instruments International

Quartz Crystal Thermometry


Applications: Temperature and temperature difference measurements, usually in the laboratory

Temperature Range: 112 to 482F (80 to 250C)

Inaccuracy: 0.135F (0.075C) throughout the full range; 0.072F (0.04C) for the range of 58
to 302F (50 to 150C). For temperature difference measurement over an 18F
(10C) span, the error under laboratory conditions is within 0.018F (0.01C).

Former Supplier: Micro Crystal of Switzerland

In this section, several unrelated methods of temperature relationship of pressure to temperature is given by the rela-
measurement will be briefly discussed. Some of these are tionship: pv = nrt. This relationship can be exploited in indus-
still in the experimental stage, while others have not been try where fixed volumes of gas may exist and where the
exploited commercially because they are not yet competitive pressure is known. A constant volume gas thermometer is
with the more common sensors. This fast-changing field is precisely this type of device.
so broad that only select examples will be covered. Similarly, the resistance of an electric heating coil can
be measured and used to define its own temperature, because
the resistance changes as a function of temperature.
SELF-MEASURING DEVICES Actually, any property having a consistent rate of change
for a temperature variation will serve to indicate tempera-
Sometimes temperatures can be measured by allowing a ture. The frequency of the chirp of crickets on a summer
material to serve as its own thermometer. This can be done by night is an indication of their temperature environment. The
measuring the pressure of a constant volume of confined gas, rate at which a viscous substance, such as oil drips through
because that indicates its mean temperature. The proportional a small hole in the bottom of its container, is just as much an

623
2003 by Bla Liptk
624 Temperature Measurement

indication of its temperature (provided the time-temperature acoustic thermometers (see Section 4.15), these devices can
relationship is known) as would be obtained by inserting a be highly accurate and much consideration has been given
glass-stem thermometer into the fluid itself. All of these to using this apparatus to define the lower end of the tem-
relationships may be derived by equation, but reasonably perature scale. Accuracies in the sub-milli-Kelvin range are
accurate measurements can be made from empirical data. achievable.
These indirect measurement techniques are quite useful for
correlating known temperatures. For example, in a fixed
volume of gas having a measured temperature, the pressure
CARBON RESISTORS
gauge should read the appropriate pressure for that temper-
ature. Because of the known relationship the parameters can
Commercially available carbon resistors have been used as
confirm each other.
temperature sensors in the cryogenic temperature area near
In addition to such self-measuring devices, there are also
absolute zero (from about 424F, or 253C, downward to
many new configurations of older systems. For example, the
below 458F, or 272C). One major benefit of the carbon
bimetallic spring can now be used as an optical temperature
resistor at low temperature is its lower susceptibility to
switch. In this design, the bimetallic element (Section 4.2) is
adverse effects caused by a magnetic field. They are some-
upwardly convex until the temperature is reached. At that what less accurate than competing sensors and do require
point it becomes downwardly convex, which blocks an opti- individual calibration to keep the measurement error under
cal light path. The number of devices that exploit old concepts 1%. Carbon resistors may be incorporated into resistor net-
in new ways is very large, and no attempt will be made here works to improve linearity in some ranges.
to fully cover them. Resistor sizes of 0.1 to 1 W and ambient resistance values
up to 150 exhibit a large increase in resistance below 424F
(253C). Reproducibility on the order of 0.2% are obtainable
ACOUSTIC TIME DOMAIN REFLECTOMETRY when calibrated individually. Small size, low cost, and general
availability make their use attractive in cryogenic work.
This thermometer operates on the principle that in solids The influence of stray radio interference and a loss of
ultrasound pulses travel at speeds, which are a function of sensitivity are drawbacks. Variation in the resistors make
the temperature of the solids (Figure 4.8a). The measurement calibration difficult and generally impose an individual
is made by detecting the time needed for the acoustic pulses device calibration requirement.
to travel from the transducer to the impedance demarcation In addition to carbon resistors and conventional resis-
point (which may be the junction between the wire and the tance temperature detectors (RTDs) (Section 4.10) or ther-
wall of the tank) and back to the transducer. This device is mistors (Section 4.12), there is a variety of special resistors
in the development stage and shows good potential, although used as temperature sensors. One example consists of a thin
some drift in the measurement has been reported. nickel film deposited onto an electrically insulating substrate
Gas temperature can be measured by detecting the time in a reducing atmosphere. The sheet resistance is a function
needed for an ultrasonic pulse to travel through a fixed dis- of the heat treating temperature, cycle time, and thickness.
tance in the hot gas (Section 4.15). This type of measurement The different designs of special temperature-sensitive resis-
is useful where conventional techniques may be less suitable, tors are too numerous to mention. These systems are engi-
such as nuclear reactor environments. Often referred to as neered for specific applications and lack the widespread
applicability required of most temperature measurement sys-
Temperature, F tems. They are, however, a very effective method of solving
0 1,000 2,000 3,000 4,000 5,000 6,000 7,000
5 0.20 specific measurement problems.
Molten Aluminum
Reciprocal Velocity m/sec
Sound Velocity m/sec

4 Aluminum 0.25
Molten
Thin Wire Iron 0.5C Prenium
3 Silver Thin Wire 0.30 CAPACITANCE CABLE FOR DETECTING HOT SPOTS
Thin Wire
Molten
Sodium 0.40
2 0.50 A cable consisting of an electrical conductor, a dielectric,
and a conductive plastic can be used to detect hot spots.
1 10 The polymer in the over-temperature locator cable increases
20 its electrical resistance with temperature. Therefore when a
0 hot spot evolves anywhere along the length of the cable,
0 1,000 2,000 3,000 4,000
this will result in a drop in capacitance, as measured from
Temperature, C
one end of the cable. The location of the hot spot can be
FIG. 4.8a determined by comparing the measured capacitance from
The velocity of sound is higher in thin wires when they are cold, the control end of the cable with the capacitance of the full
2
and it drops as the wires are heated. length of the cable.

2003 by Bla Liptk


4.8 Miscellaneous and Discontinued Sensors 625

Input Two will provide a corresponding temperature that is determined


Gases at Chamber using the Johnson noise measurement technique. It is inter-
Unknown Resonator esting that much of the current research in this area is cen-
Temperature Oscillator
tered around cryogenic temperatures. Inaccuracies of 0.001%
may be achievable in the range from 84 to 430 K. This type
Steady
of instrument and sensor is collectively referred to as Johnson
Beat noise thermometers.
Frequency Output
Detector Frequency Pressure Is
Made of Converter Proportional
Fluid
Amplifiers to Beat LIQUID CRYSTALS
Frequency

Reference Used in nondestructive testing for surface temperature mea-


Signal surements, liquid crystals undergo a series of color changes
as temperature varies. They are organic compounds that are
FIG. 4.8b physically liquid, but exhibit optical properties that are sim-
Fluidic temperature-to-pressure converter. ilar to those of a crystalline solid.
A number of solutions are available, from the minimum
temperature of about 68F (20C) to a maximum of approx-
FLUIDIC SENSORS
imately 340F (170C). The solutions are packaged in kit
form for various ranges within these limits. Mixtures are made
The fluidic sensor is a device for converting gas temperatures
covering spans as narrow as 4F (2.2C) within the selected
into gas pressure. Actually, it is a beat-frequency detector
range.
system, which contains no moving parts other than the gas.
Temperature is read by comparing the color exhibited
One type of design is shown in Figure 4.8b. The oscil-
when a thin coating is subjected to the conditions under
lator is a two-chamber resonator in which the entering gases
question to a standard reference color. Response speed is less
are split by a knife-edge. The gases that are reflected from
than 1 s. The indication is continuous and reversible. Cost is
one chamber into the other, set up oscillations whose fre-
low. Disadvantages include manual preparation and limited
quency is proportional to the square root of the absolute
range, in addition to lack of automatic readout.
temperature.
Liquid crystals have also been used in fiber-optic ther-
A reference signal input from a temperature-sensitive
mometers. A pellet of liquid crystal is inserted at the tip of
resonant oscillator is compared with the unknown in a beat-
an optical fiber. As light at different wavelengths is sent
frequency detector made of beam-deflection fluid amplifiers.
through the fiber, the reflection peaks from the liquid crystal
The frequency of its output is the beat frequency of the
are related to temperature.
combined reference and oscillator signals. The components
in the frequency converter create a steady pressure propor-
tional to the beat frequency. Accuracies of about 2% and
PARAMAGNETIC SALTS
temperatures up to 2000F (1093C) are claimed for the
system.
Magnetic thermometry has been developed chiefly to measure
temperatures near absolute zero (below 458F, or 272C).
The temperatures are obtained by adiabatic demagnetization
JOHNSON NOISE THERMOMETER of a paramagnetic salt. An isothermal magnetization at the
lowest attainable liquid helium temperature (about 458F,
Investigation into noise thermometry has been evolving over or 272C) followed by an adiabatic demagnetization is used.
the last few decades. All resistive devices produce a random The entropy is decreased, with a simultaneous heat flow from
noise power level based upon their absolute temperature. the sample, when the magnetic ions are oriented parallel to
In most cases this noise is considered interference and elec- the field. During subsequent adiabatic demagnetization the
tronic circuit designers work to reduce it. Equation 4.8(1) entropy of the salt remains constant, if demagnetization is
gives a general relationship that relates temperature to the reversible, and temperature decreases.
VRMS value: To obtain the temperature, some temperature-dependent
quality of the salt under investigation is used, such as the
f1 magnetic susceptibility.
VRMS = 4 KT
R df
f2
4.8(1) If a sphere or rotational ellipsoid of an isotropic paramag-
netic salt is located in the homogeneous part of the magnetic
field of a coil of a mutual inductance or a self-inductance, the
For very high temperature measurements, tungsten rhe- inductance of the coil is a function of the temperature. Induc-
nium thermocouples viewed in terms of a resistive device tance can be measured with an AC bridge (as shown in

2003 by Bla Liptk


626 Temperature Measurement

Oscillator of this material is placed on the surface to be investigated


and is excited with ultraviolet (UV) radiation in a darkened
room. The brightness of the coating is compared to the bright-
Shielded Transformer ness of the same coating at a known temperature as an indi-
cation of the unknown temperature.
The sensitivity of the phosphors used gives a 10% bright-
Filter
Shielded ness change per F and this can be picked up with a relatively
Transformer crude system of photometry. Temperature range for this type
Tuned
of measurement is from 32 to about 750F (0 to about 400C).
Amplifier AC Bridge
Detector

FIBER BRAGG GRATING TEMPERATURE SENSORS


FIG. 4.8c
Circuitry for temperature measurement by paramagnetic salts. The Fiber Bragg Grating (FBG) offers a promising technol-
ogy for future temperature-sensing applications as well as
Figure 4.8c) whose balance is independent of frequency. A
possible strain sensor applications. Research has been done
galvanometer can be used for detection. Effective shielding is
indicating considerable promise for FBG to be a viable tem-
a requirement.
perature measurement technology.
In a paramagnetic salt with a coil surrounding it, self-
In this optical technology, Bragg wavelength shifts are
inductance is related to temperature. An Anderson AC bridge
measured to determine the temperature of a specifically
has been used to measure magnetic temperature in such a
treated section of the fiber core material. FBG technology
situation. The relationship between self-inductance and sus-
can be measured in a distributed manner by wavelength divi-
ceptibility of a salt has been found to be linear when the
sion multiplexing. This would make it suitable for petro-
ellipsoidal or spherical salt piece is placed in the homoge-
chemical plant applications in hazardous, classified locations.
neous part of the measuring field.
An FBG is formed by exposure to a fringe of UV light.
Accuracy of the magnetic method has been estimated on
Only the specified Bragg wavelength as related to its grating
the order of 0.001F (0.00056C). The method is the best
period is reflected from broadband-source input light. As a
available for measurements near absolute zero.
temperature change occurs, a Bragg wavelength shift occurs
that can be measured and can correlate back to the measured
SPECTROSCOPIC TEMPERATURE MEASUREMENT temperature. The measurement hardware can be completely
placed within a conventional thermowell just like a thermo-
Spectroscopic methods are often used to measure the tem- couple and its sheathing would.
perature of hot gases. They are, in fact, the only possible way
to measure the surface temperature of stars. DISCONTINUED TEMPERATURE SENSORS
The spectroscope in its simplest form is the familiar
triangular glass prism, which breaks up light from a hot object While the temperature sensors described in these paragraphs
into its constituent colors (its spectrum). The chemical com- are not marketed at this time, they are still in use. Whether
position of glowing gas is determined from the pattern of
they will or will not be manufactured in the future is
dark (Fraunhofer) lines that appear across the spectrum.
unknown. For sake of completeness, they are described here.
Many procedures for temperature determination from the
spectrum have been developed, such as measurement of bright- Pneumatic and Suction Pyrometers
ness and actual color, reversal temperatures, population tem-
perature estimates, measurements made of spectral line shifts The measurement of hot gas temperature of combustion pro-
in ionized gases, and many others. These are all laboratory cesses does not give accurate results with conventional ther-
techniques seldom employed industrially because of their com- mometers because of interference from the radiated heat of
plexity and relatively high costs. For further information on the surroundings. When a conventional thermocouple (TC)
this subject, refer to Chapter 56, Spectroscopic Methods of is used, the amount of interference is a function of the relative
Temperature Measurement, in Temperature: Its Measurement values of the radiant heat transfer coefficient of the surround-
and Control in Science and Industry, Vol. III, Pt. I. ings and the convective heat transfer coefficient of the gas.
When a standard TC is used in a cold furnace, the resulting
error can be several hundred degrees. Pneumatic and suction
THERMOGRAPHY pyrometers have been developed in order to reduce this error.

The strong temperature dependence of the brightness of cer- Suction Pyrometers The suction pyrometer consists of a
tain luminescent materials may be converted into a pattern sheathed TC (sheathed against chemical attack) located inside
of color that can be recorded photographically. A thin layer a radiation shield at the tip of a suction pipe (Figure 4.8d).

2003 by Bla Liptk


4.8 Miscellaneous and Discontinued Sensors 627

Radiation Shield

Sheathed Thermocouple Expansion Bellows Thermocouple Output

Stainless Steel Probe Protection Tube


Hot Gas
C

A
Operating Length
B
To Aspiration Equipment

Cooling Water Spare Thermocouple Wire

FIG. 4.8d
Sheathed thermocouple detects the hot gas temperature in an aspirating suction pyrometer. (Courtesy of Land Combustion, a Division of
Land Instruments International Inc.)

The combustion gases are sucked through the shield and over and cold venturi are inversely proportional to the densities
the TC at high velocity by aspirating equipment. The effi- of the hot and cold gas. Therefore they are directly propor-
ciency of this aspirating TC is a function of the quality of tional to the absolute temperatures at the hot and cold venturi.
the radiation shield and of the suction flow rate. If, for exam- As the temperature at the cold venturi (Tc) is measured, the
ple, a suction pyrometer has a 100C error without suction hot gas temperature (Th) can be calculated by multiplying Tc
and a 10C error when the suction flow is on, it is said to and a configuration constant K with the ratio of Ph /Pc. There-
have an efficiency of 90%. fore, the absolute temperature of the gas is:
The suction pyrometer probe is usually made of stainless
steel and is water-cooled. These probes are designed for high Th = KTc ( Ph / Pc ) 4.8(2)
gas velocities of 500 fps (152 mps). At such velocities the
radiation shields usually produce better than 90% pyrometer Relative to the suction pyrometer, the pneumatic pyrom-
efficiencies. The furnace gases can be pulled out by fans or eter requires more utilities and more design attention as
by air or steam ejectors. The main limitations of this design almost all units must be uniquely designed.
include plugging of the probe when the combustion products
are dusty (as in pulverized coal burners). In addition, they Quartz Crystal Thermometry
cannot be used in applications where the temperatures exceed
2912F (1600C). The effect of temperature on the frequency of quartz crystal
controlled oscillators has been known since work on this
Pneumatic Pyrometers The pneumatic probe was invented subject been published in 1946. Temperature measurement
in 1893 for use in blast furnaces. It is superior to suction using quartz thermometers is based on the very linear change
pyrometers in that dust in the combustion gases does not plug that occurs in resonant frequency in response to a change
it as easily and that it can measure very high temperatures, in temperature. Under ideal conditions, this temperature sen-
such as in turbojet engines or rockets. This pyrometer can sitivity is on the order of 0.0005F (0.0003C), and temper-
6
detect higher gas temperatures than can the aspirating TC ature deviations on the order of 10 F have been measured
because the thermometer does not need to reach the gas in the laboratory.
temperature (Figure 4.8e). The pneumatic probe can handle A reference and a sensor oscillator are used. The refer-
gases heavily laden with dust because it does not have a ence oscillator frequency is selected to provide zero reference
radiation shield, which is prone to plug. beat when the probe is at 32F (0C). Sensor oscillators
The pneumatic pyrometer measures the pressure drop of provide a sensitivity of 500 cps/F and dividers are arranged
the hot combustion gases through one restriction at the probe to obtain a sample period of .01 s. A resolution of 0.2F (0.1C)
tip and measures their pressure drop again after they have has been obtained with a digital readout (Figure 4.8f). When
been cooled through an identical restriction (see Figure 4.8e). arranged for differential measurement, gating circuits are
Because the mass flow rate of the gas is the same across these added to enable the oscillators to be heterodyned against each
restrictions, the only reason these two pressure drops are dif- other.
ferent is because of the change in the gas temperature, which In past applications, probes were provided with the quartz
affects its density. The pressure differentials across the hot crystal hermetically sealed inside a well, formed from a 3/8

2003 by Bla Liptk


628 Temperature Measurement

Th
Th

Ph
Water
Ph
Expansion Bellows Platinum Resistance Thermometer

Hot Venturi Cold Venturi


Hot Gas Protection Tube
Th Stainless Steel Probe
C

A
Operating Length
B
To Aspiration Equipment
Tc
Ph Ph
Cooling Water

FIG. 4.8e
Pneumatic pyrometer determines hot gas temperature by multiplying cold and temperature (Tc) with the ratio of pressure drops across the
hot and cold venturis. (Courtesy of Land Combustion, a Division of Land Instruments International Inc.)

Sensor in. (9.5 mm) outside diameter stainless steel cylinder (Figures
Sensor
Indicator Oscillator 4.8g and 4.8h). The probes can be used at pressures up to
+ or 3000 PSIG (21 MPa) and can stand shocks of 10,000 g
Recorder D/A Digital
without changing their calibration.
Input Output Counter
Mixer In a well-stirred water bath, the response time for a step
Digital change is 1 s. Long-term stability, for periods of a month or
to more, is on the order of 0.02F (0.01C). Oscillator drift can
Analog
Gate lower the stability, but, in an environmentally controlled area,
Converter Dividers
Control this is not much of a problem.
Reference
Oscillator Advantages and Disadvantages The main advantages of
quartz crystal thermometry are that:
FIG. 4.8f
Quartz thermometer system block diagram. 1. There are no lead resistance or noise problems because
the temperature is converted to frequency.
Sleeve 2. It has excellent short-term stability.
3. It has good accuracy.
4. It has a 1-s response time.
Insulated 5. It has accurate differential measurement.
Probe Coaxial
6. It is ruggedit can withstand shocks without chang-
Cable
ing calibration.
FIG. 4.8g
Quartz crystal sealed in a capsule. The disadvantages include that:

1. These units are expensive.


2. Their accuracy is somewhat lower than those of resis-
tance thermometers or thermistors.
3. Their probe size is larger than TC or resistance tem-
FIG. 4.8h perature detector.
Quartz crystal in stainless steel well. 4. They are better suited for the laboratory environment.

2003 by Bla Liptk


4.8 Miscellaneous and Discontinued Sensors 629

References Johnson, R., Measuring the Hot, Difficult, and Inaccessible, Control Engi-
neering, June 2001.
Kerin, T.W. and Katz, E.M., Temperature Measurement in the 1990s,
1. Plumb, H.H., Temperature: Its Measure and Control in Science and
InTech, August 1990.
Industry, Vol. 4, 5th Symposium on Temperature, National Bureau of
Krigman, A., Temperature Instrumentations: Recent Patent Activity, InTech,
Standards, American Institute of Physics, Instrumentation, Systems,
August 1983.
and Automation Society, Pittsburgh, PA, 1972.
Langan, P.E. and Ironside, D.S., Calibration Ensures Accurate Temperature
2. Lynnworth, L.C., Sound Ways to Measure Temperature, Instrumen-
Measurement, InTech, October 1989.
tation Technology, April 1969.
Li, B., Liu, B.Y., Zheng, X.R., Lai, P.T., Wu, Z.H., and Sin, J.K.O., Ultra-
Low Power Thermal Sensor with Silicon-on-Insulator Structure for
High-Temperature Applications, Paper #28.4, First IEEE International
Conference on Sensors, Orlando, FL, June 2002.
Bibliography Magison, E., Selecting the Right Temperature, InTech, February 1993.
Magison, E., Temperature Measurement, InTech, October 25, 2001.
Michalski, L. et al., Temperature Measurement, 2nd ed., London: John Wiley
Adler, C.B., Reliability Aspects of Temperature Measurement, Instrumen- & Sons, 2001.
tation, Systems, and Automation Society Conference, 2001. Moldover, M.R., Primary Acoustic Thermometry, Dubbeldam, J.F., Ed.,
Allen, M.G., Wang, T.P., and Bediones, D.P., Recent Advances in Temper- Gaithersburg, MD: National Institute of Standards and Technology,
ature Sensors, Paper #95080, ISA/95: Advances in Instrumentation TEMPMEKO, 1999.
& Control, Chicago, October 1995. Moray, W.W., Dunphy, J.R., and Meltz, G., Multiplexing Fiber Bragg
Benz, S.P. and Tew, W., A New Approach to Johnson Noise Thermometry, Grating Sensors, Applied Optics, September 1996.
Fellmuth, B., Ed., National Institute of Standards and Technology Owen, T., No Need to Cool Down, Control, June 1998.
Proceedings, TEMPMEKO, Boulder, CO, 2001. Peterson, W., Choosing the Right Temperature Sensor, InTech, April 1991.
Bidini, G., A Method for the Thermo-Fluid Dynamic Design of Gas Turbine Plumb, H. H., Temperature: Its Measure and Control in Science and Indus-
Total Temperature Probes, Paper #910511, Instrumentation, Sys- try, Vol. 4, 5th Symposium on Temperature, National Bureau of Stan-
tems, and Automation Society Conference, Toronto, 1991. dards, American Institute of Physics, Instrumentation, Systems, and
Bliss, P. and Morgan, R.K., Errors in Temperature Measurement, Instru- Automation Society, Pittsburgh, PA, 1972.
mentation Technology, March 1971. Prentice, G.R., Seven Reasons Why Temperature Transmitters Are Better
Bluestein, I., Understanding Contact Temperature Sensors, Sensors, Octo- than Direct Wiring, Process Heating, June 1999.
ber 2001. Sade, S., Eyal, O., and Katzir, A., Embedded Silver Halide Optical-Fiber
Desmarais, R. and Breuer, J., How to Select and Use the Right Temperature Temperature Sensor, Paper #31.2, The First IEEE International Con-
Sensor, Sensors, January 2001. ference on Sensors June 2002.
Fraden, J., Handbook of Modern Sensors, 2nd ed., Heidelberg: Springer- Schooley, J.F., State of the Art of the Instrumentation for High Temperature
Verlag, 1997. Thermometry, Publication Number ANL-787, Argonne National
Hage, J., Smart Temperature Transmitter Accents Accuracy, Control, Sep- Laboratories Symposium, Argonne, IL,1977.
tember 1999. Shepard, R.L. et al., Development of Long Life, High Reliability, Remotely
Hashemian, H.M. and Peterson, K.M., Assuring Accurate Temperature Operated, Johnson Noise Thermometer, Paper #910306, Instrumenta-
Measurement, InTech, October 1989. tion, Systems, and Automation Society Conference, Toronto, 1991.
Hirayama, N., Sano, Y., and Kuroiwa, S., Fiber Bragg Grating Temperature Stockham, R., Temperature Transmitters Take the Lead Over Direct Wir-
Sensor for Practical Use, ISA/99 Tech/Expo,: Advances in Instrumen- ing, Control and Instrumentation, Product Survey Directory, 1999.
tation & Control: Emerging TechnologiesPhiladelphia, October 1999. Volbrecht, A. and Gordon, W., Temperature Measurement: Making Sense
Hormuth, G.A., Ways to Measure Temperature, Control Engineering, of it All, Sensors, June 1998.
Reprint No. 948, 1971. Yoder, J., Taking the Mystery Out of IR Sensors, Control, June 2000.

2003 by Bla Liptk

Das könnte Ihnen auch gefallen