Beruflich Dokumente
Kultur Dokumente
Drug Quality
R&D
Discovery Control
Logistics Retail
Laboratory Solutions
Manufacturing Packaging Industrial Solutions
Retail Solutions
6
Guaranteed Accuracy Reduced Costs Secure Audits
Problem source:
How to prevent bad
Inaccurate weighing
batches and OOS*?
-QUALITY-
Testing is costly
Do we test too little and a lot of effort
or too often?
-COST-
Good Manufacturing Practice
9
Regulations Affecting Scales & Balances: cGMP
Capacity: 3'000kg
Readability: 1kg
min. 50%
Uncertainty
History
Revision of USP General Chapter 41
Revision of USP General Chapter 1251
Reference to General Chapter 841 and 1251
Scope
"States the requirements for balances used for materials that must be
accurately weighed (see General Notices, 8.20)."
Weighing of analytes for quantitative measures
Scope unchanged but better defined/explained
Requirements
"Weighing shall be performed using a balance that is calibrated []"
"[] and meets the requirements defined for repeatability and
accuracy."
"For balances used for other applications the balance repeatability and
accuracy should be commensurate with the requirements of its use."
Requirement
"Repeatability is assessed by weighing one test weight NLT 10 times."
"Repeatability is satisfactory if two times the standard deviation of the
weighed value, divided by the desired smallest net weight, does not
exceed 0.10%."
"If the repeatability obtained is smaller than 0.41 d, where d is the scale
interval, replace this standard deviation with 0.41 d."
Additional Statement
"Because repeatability is virtually independent of sample mass within the
balances capacity, use of a small test weight, which may be difficult to
handle, is not required."
Indication
Load
5% 5%
"To facilitate handling, the test weight that is used for the repeatability
test does not need to be at the minimum weight value but can be larger
because the standard deviation of repeatability is only a weak function
of the test weight value." Internal usage only
Requirement for Repeatability
Correct sensitivity
Indication
Not allowed
Stipulated
Load
< 5% 5% Capacity Internal usage only
USP <41>: Accuracy Requirements
Scope
"States the requirements for balances used for materials that must be
accurately weighed (see General Notices, 8.20)."
Weighing of analytes for quantitative measures
Scope unchanged but better defined/explained
Requirements
"Weighing shall be performed using a balance that is calibrated []"
"[] and meets the requirements defined for repeatability and
accuracy."
"For balances used for other applications the balance repeatability and
accuracy should be commensurate with the requirements of its use."
History
Revision of USP General Chapter 41
Revision of USP General Chapter 1251
Reference to General Chapter 841
All USP Generel Chapters with a number equal or larger than 1000 are
for information only and not legally binding.
A proposal to change General Chapter 1251 "Weighing on an Analytical
Balance" was made by the USP Expert Panel in parallel to the proposal
to change General Chapter 41 "Weights and Balances"
Provides information on
- Installation and Operational Qualification
- Performance Qualification and Balance Checks
- Minimum Weight
- Balance Operation
"Table 1 provides a list of the most important balance properties that should be
assessed during performance qualification. Depending on the risk of the
application and the required weighing process tolerance, some of these tests
may be omitted. Tests also can be omitted if there is evidence that the property
in question has only minimal effect on the weighing performance."
Property Examples
Sensitivity The test load at or sufficiently close
to the capacity of the balance.
Eccentricity Performed in the center of gravity and the four quadrants [].
Test load usually should be 30% of the capacity of the balance
or higher.
Performance Qualification
"Performance qualification should be performed periodically as described
in standard operating procedures, and the frequency of each of the
individual tests can vary depending on the criticality of the property."
Balance Check
"The balance check is performed at appropriate intervals based on
applicable standard operating procedures. The frequency of the balance
check depends on the risk of the application and the required weighing
tolerance."
"Checks with external weights can be replaced partially using (automatic
or manually triggered) adjustment by means of built-in weights."
Indication
Load
5% 5%
"In order to satisfy the required weighing tolerance, when samples are
weighed the amount of sample mass (i.e., the net weight) must be equal
to or larger than the minimum weight. The minimum weight applies to
the sample weight, not to the tare or gross weight."
Safety Factor = 1
Weighing process is currently OK,
but reproducibility may fail over time.
Safety Factor 2
(depending on risk)
Assured accuracy over time
US Minimum Weight beyond USP <41>
What we have evolved to....
Minimum Weight?
Testing Parameters?
Training
Certified weights
Tolerances
Uncertainty
Service
Calibration Departments
4 Simple Questions to Answer
Am I doing too much or
Are my balances/scales accurate?
too little testing?
40
Copyright 2009 Mettler-Toledo, Inc.
Electromagnetic compensating analytical balance
1 weighing pan
2 hanger
3 guide
4 flexible joint
5 coupling
6 lever
7 lever bearing
8 compensation coil
9 permanent magnet
10 magnetic flux
11 optical position sensor
12 position vane
13 temperature sensor
41
Copyright 2009 Mettler-Toledo, Inc.
Measurement Uncertainty what is it?
Measurement uncertainty
[] Parameter that describes the dispersion of the quantity values
that is assigned to a scale [].
International Vocabulary Of Basic And General Terms In Metrology (VIM) JCGM
200:2008, entry 2.26
A scale is ACCURATE
if it satisfies the requirements of the user.
In other words:
OK Fair Not OK
m1 kg m2 kg m3 kg Display [kg]
Nonlinearity NL
Eccentricity EC
Repeatability RP
Max
For small sample weights,Weight
the [g]
relative uncertainty can become
so high that the weighing results cannot be trusted anymore!
Accuracy limit = Minimum weight
Measurement uncertainty %]
Accuracy limit:
Minimum weight
Measurement uncertainty %]
1%
0.5%
0.1%
Metrology
Quality
management
Process
requirements
Content of GWP Verification: Per Instrument
Recommended performance
verification activities and frequencies