Beruflich Dokumente
Kultur Dokumente
The tracings on the EKG paper are a reflection of electrical activity of the heart.
A series of body surface electrodes are placed at specific points on the arms, legs
and thorax that sense and record the heart's electrical activity. The electrodes
are assigned a specific polarity - i.e. - either negative or positive.
For any lead, the EKG machine looks at specific combinations of leads in order
to configure the tracing we expect to see for Lead I, II, III, aVF, aVL, aVR or for
V1-V6. The electrode that we care about the most is called the sensing
electrode which is always given a positive polarity (+). Hence, we call it
the positive sensing electrode.
In order to understand what the EKG tracing is saying to us, there are a few
general rules to remember for depolarization and repolarization.
The EKG paper is a ruled paper that is usually heat sensitive. The smallest
division on the paper is a one millimeter box. The largest division on the paper is
a five millimeter box. By international convention, the speed of the paper moves
at a rate of 25 millimeters per second past the stylus. This is the speed that makes
all of the tracings look appropriate for any lead that is being viewed. Because the
speed of the paper is 25 mm/second, a small 1 mm box is traversed in .04 seconds
and a large 5 mm box is traversed in .2 seconds.
Because of the predictable EKG paper speed, we can count the number of boxes
traversed over a period of time between heart beats to determine HR.
Place the beginning point of a cardiac ruler over an R wave. Look at the
number on which the next R wave falls and that becomes the heart rate for
that patient. Use the following numbers to indicate what the heart rate is
between two successive R waves : 300, 150, 100, 75, 60, 50, 43, 37, 33, 30
Method # 2 - The Six Second Tracing Method
Obtain a six second tracing (30 five mm boxes) and count the number of R
waves that appear within that 6 second period and multiply by 10 to obtain
the HR/min.
Example : If there were 12.5 small boxes between two successive R waves,
then the heart rate would be : 1500/12.5 small boxes = 120 bpm.
The six limb leads (Leads I, II, III, aVL, aVF, and aVR) look at the electrical
output of the heart in the frontal plane. By international convention, a circle is
drawn from the chin to the symphysis pubis to describe the area in which the
limb leads view the heart. The circle is divided along the horizontal X axis and
the verticle Y axis. The top half of this circle is designated to be wholely negative
with the circle enumerated as 0 to -180 degrees. The bottom half of this circle is
designated as wholely positive with the circle enumerated from 0 to +180 degrees.
The circle is further divided into 90 degree halves such that the frontal plane is
now cut into four equal 90 degree quadrants : the normal axis quadrant from 0
to + 90 degrees; the left axis deviation quadrant from 0 to -90 degrees; the right
axis deviation quadrant from +90 to +180 degrees; and finally the extreme right
axis deviation quadrant from -180 to -90 degrees.
1) Lead I is created by making the left arm positive (+) and the right arm
negative (-). Its angle of orientation is + 0°. Lead I looks across the heart
from the right side of the thorax to the left side of the thorax along the +0°
axis in the frontal plane.
2) Lead II is created by making the left leg positive (+) and the right arm
negative (-). Its angle of orientation is +60°. Lead II looks across the heart
from the right shoulder down through the thoracic cage to the left hip
along the + 60° axis in the frontal plane.
3) Lead III is created by making the left leg positive (+) and the left arm
negative (-). Its angle of orientation is a +120°. Lead III looks across the
heart from the left shoulder down through the thoracic cage to the right
hip along the +120° axis in the frontal plane.
4) aVL is created by making the left arm positive (+) and the other
extremities negative (-). Its angle of orientation is -30°. aVL looks across
the heart from the right leg up through the thoracic cage to the left
shoulder along the -30° axis in the frontal plane.
5) aVR is created by making the right arm positive (+) and all other
extremities negative (-). Its angle of orientation is -150°. aVR looks across
the heart from the left hip up through the thoracic cage to the right
shoulder along the -150° axis in the frontal plane.
6) aVF is created by making the legs positive (+) and all other extremities
negative (-). Its angle of orientation is +90°. aVF looks through the heart
from the chin down through the thoracic cavity to the feet along the +90°
axis in the frontal plane.
The leads looking at the inferior portion of the heart are : II, III, aVF.
The leads looking at the left lateral portion of the heart (left ventricle) are : I,
aVL.
The augmented lead aVR looks toward the right side of the heart and is
considered to be a lead in "no man’s land".
Last revised : August, 2001