Sie sind auf Seite 1von 37

practical Strain Gage

measurements

Introduction

W ith todays emphasis on


product liability and energy
efficiency, designs must not only be
lighter and stronger, but also more
measurements using bonded
resistance strain gages. We will
introduce considerations that affect
the accuracy of this measurement
and suggest procedures for
Appendix B contains schematics
of many of the ways strain gages
are used in bridge circuits and the
equations which apply to them.
Readers wishing a more thorough
thoroughly tested than ever before. improving it. discussion of bridge circuit
This places new importance on theory are invited to read Item
the subject of experimental stress We will also emphasize the 7 referenced in the bibliography.
analysis and the techniques for practical considerations of strain
measuring gage measurement, with an
strain. The main theme of this emphasis on computer controlled
application note is aimed at strain instrumentation.

STRAIN GAGES
E

E-94
practical Strain Gage measurements

Symbols
s normal stress Rg gage resistance in ohms Rl lead wire resistance
t shear stress K t transverse sensitivity T temperature in C
ratio
e strain (normal) VIN bridge excitation voltage
L length
me micro-strain (e x 10 ) 6
VOUT bridge output voltage
g shear strain DL change in length
E modulus of elasticity
DRg change in gage resistance
(due to strain)
or Youngs modulus
%DGF % change in gage factor
n Poisson Ratio
(due to temperature)
GF gage factor

Vr [(VOUT/VIN)[strained] (VOUT/VIN)[unstrained]]

stress &strain

T
Force Force
(F) (F)
he relationship between stress
and strain is one of the most
fundamental concepts from the
study of the mechanics of materials L
and is of paramount importance to
the stress analyst. In experimental
stress analysis, we apply a given
load and then measure the strain on
Figure 1: Uniaxial Force Applied
individual members of a structure or
machine. Then we use the stress-
strain relationships to compute the See Figure 2. When this is written
stresses in those members to verify
Tensile (+) Strain
This Surface
e
in equation form, = DL/L, we
that these stresses remain within can see that strain is a ratio and,
the allowable limits for the particular Force (F) therefore, dimensionless.
materials used. To maintain the physical
significance of strain, it is often
written in units of inches/inch. For
Strain Compressice (-) Strain
This Surface most metals, the strains measured
When a force is applied to a body, in experimental work are typically
Figure 2: Cantilever in Bending less than 0.005000 inch/inch. Since
the body deforms. In the general
case, this deformation is called practical strain values are so small.
See Figure 1. This is the strain that they are often expressed as micro-
strain. In this application note, we
will be more specific and define the
we typically measure with a bonded
resistance strain gage. Strain
e
strain, which is x 106 (note this
term STRAIN to mean deformation is equivalent to parts per million
per unit length or fractional change
may be either tensile (positive) or
compressive (negative).
or ppm) with the symbol me . Still
in length and give it the symbol, .e another way to express strain is as
E-95
practical Strain Gage measurements

e
percent strain, which is x 100. For
example: 0.005 inch/inch = 5000
D
me =0.5%. F F
As described to this point, strain is
fractional change in length and is
directly measurable. Strain of this D D L
type is also often referred to as L + L
normal strain. = L/L t = D/D

Figure 4: Poisson Strain


Shearing Strain
Another type of strain, called that of Poisson strain. The dashed and structures as small and light as
SHEARING STRAIN, is a measure lines indicate that the bar not only possible, component parts should
of angular distortion. Shearing strain elongates but that its girth contracts. be stressed, in service, to the
is also directly measurable, but not This contraction is a strain in highest permissible level. STRESS
as easily as normal strain. If we had the transverse direction due to a refers to force per unit area on a
a thick book sitting on a table top property of the material known as given plane within a body.
and we applied a force parallel to Poissons Ratio. Poissons ratio, ,n The bar in Figure 5 has a uniaxial
the covers, we could see the shear is defined as the negative ratio of
tensile force, F, applied along the
strain by observing the edges of the the strain in the transverse direction
x-axis. If we assume the force to be
pages. to the strain in the longitudinal
uniformly distributed over the cross-
direction. It is interesting to note
sectional area, A, the average
that no stress is associated with the
stress on the plane of the section
Poisson strain. Referring to Figure
Shear Strain = is F/A. This stress is perpendicular
Force 4, the equation for Poissons ratio
n e e
is = t / 1. Note that is n to the plane and is called NORMAL
s
STRESS, . Expressed in equation
dimensionless.
form, s = F/A, and is denoted
Book in units of force per unit area.
Since the normal stress is in the x
Normal Stress

STRAIN GAGES
direction and there is no component
of force in the y direction, there is
While forces and strains are
Figure 3: Visualizing Shearing Strain no normal stress in that direction.
measurable quantities used by the
The normal stress is in the positive
designer and stress analyst, stress
See Figure 3. Shearing strain, , g is the term used to compare the
x direction and is tensile.
is defined as the angular change loading applied to a material with its
in radians between two line ability to carry the load. Since it is
segments that were orthogonal in
the undeformed state. Since this
usually desirable to keep machines E
angle is very small for most metals,
shearing strain is approximated by Y
the tangent of the angle.

Poisson Strain
F X F
In Figure 4 is a bar with a uniaxial
tensile force applied, like the bar in
Figure 1. The dashed lines show the Area (A)
shape of the bar after deformation,
pointing out another phenomenon, Figure 5: Normal Stress

E-96
practical Strain Gage measurements

Shear Stress
Just as there are two types of components, not the stresses, and stress in the y direction and in this
strain, there is also a second type that the resulting stresses are a case zero is the minimum normal
of stress called SHEAR STRESS. function of the orientation of the stress. So the requirements for the
Where normal stress is normal to section. This means that stresses principal axes are met by the x-y
the designated plane, shear stress (and strains), while having both axes. In Figure 6, the x-y axes are
is parallel to the plane and has the magnitude and direction, are not the principal axes, since that bar is
t
symbol . In the example shown in vectors and do not follow the laws also loaded uniaxially. The n-t axes
Figure 5, there is no y component of vector addition, except in certain in Figure 6 do not meet the zero
of force, therefore no force parallel special cases, and they should shear stress requirement of the
to the plane of the section, so there not be treated as such. We should principal axes. The corresponding
is no shear stress on that plane. also note that stresses are derived STRAINS on the principal axes is
Since the orientation of the plane is quantities computed from other also maximum and minimum and
arbitrary, what happens if the plane measurable quantities, and are not the shear strain is zero.
is oriented other than normal to the directly measurable. [3]
The principal axes are very
line of action of the applied force?
important in stress analysis because
the magnitudes of the maximum
and minimum normal stresses are
usually the quantities of interest.
Y
t Once the principal stresses are
known, then the normal and shear
stresses in any orientation can be
n computed. If the orientation of the
Fn
principal axes is known, through
F X F knowledge of the loading conditions
F
Ft or experimental techniques, the
Area (A) task of measuring the strains and
computing the stresses is greatly
simplified.
Figure 6: Shear Stress
In some cases, we are interested
in the average value of stress or
Figure 6 demonstrates this concept
with a section taken on the n-t Principal Axes load on a member, but often we
want to determine the magnitude
coordinate system at some arbitrary
In the preceding examples, the of the stresses at a specific point.
angle, f, to the direction of action
x-y axes are also the PRINCIPAL The material will fail at the point
of the force.
AXES for the uniaxially loaded where the stress exceeds the load-
We see that the force vector, F, can bar. By definition, the principal carrying capacity of the material.
be broken into two components, axes are the axes of maximum This failure may occur because of
Fn and Ft , that are normal and and minimum normal stress. They excessive tensile or compressive
parallel to the plane of the section. have the additional characteristic normal stress or excessive shearing
This plane has a cross-sectional of zero shear stress on the planes stress. In actual structures, the area
area of A and has both normal that lie along these axes. In Figure of this excessive stress level may
and shear stresses applied. The 5, the stress in the x direction is be quite small. The usual method of
s
average normal stress, , is in the the maximum normal stress, and diagramming the stress at a point
n direction and the average shear we noted that there was no force is to use an infinitesimal element
t
stress, , is in the t direction. Their component in the y direction and that surrounds the point of interest.
equations are: s = Fn /A and therefore zero shear stress on the The stresses are then a function of
t = Ft /A. Note that it was the plane. Since there is no force in the orientation of this element, and,
force vector that was broken into the y direction, there is zero normal in one particular orientation, the
E-97
practical Strain Gage measurements

materials. Written in equation form,


element will have its sides parallel this stress-strain relationship is Y

to the principal axes. This is the s e


= E . Some materials do not
orientation that gives the maximum have a linear portion (for example,
X
and minimum normal stresses on cast iron and concrete) to their
the point of interest. stress-strain diagrams. To do
q
accurate stress analysis studies
for these materials, it is necessary p

Stress-Strain to determine the stress-strain


properties, including Poissons
Relationships ratio, for the particular material
Now that we have defined stress on a testing machine. Also, the Figure 8: Shaft in Torsion and Tension
and strain, we need to explore the modulus of elasticity may vary
stress-strain relationship, for it is with temperature. This variation
this relationship that allows us to may need to be experimentally
calculate stresses from measured determined and considered when the biaxial stress state is the
strains. If we have a bar made of performing stress analysis at most common. Figure 8 shows
mild steel and incrementally load temperature extremes. There are an example of a shaft with both
it in uniaxial tension and plot the two other points of interest on the tension and torsion applied. The
strain versus the normal stress in stress-strain diagram in Figure 7: point of interest is surrounded by an
the direction of the applied load, the the yield point and the ultimate infinitesimal element with its sides
plot will look like the stress-strain strength value of stress. oriented parallel to the x-y axes.
diagram in Figure 7. The point has a biaxial stress state
The yield point is the stress level at and a triaxial strain state (remember
which strain will begin to increase Poissons ratio). The element,
rapidly with little or no increase in rotated to be aligned with the
Ultimate Strength
stress. If the material is stressed principal (p-q) axes, is also shown

beyond the yield point, and then the in Figure 8. Figure 9 shows the
Rupture stress is removed, the material will element removed with arrows added
not return to its original dimensions,

STRAIN GAGES
Yield Point to depict the stresses at the point for
Proportional
but will retain a residual offset or both orientations of the element.
Limit strain. The ultimate strength is the
maximum stress developed in the We see that the element oriented
material before rupture. along the x-y axes has a normal
stress in the x direction, zero
Figure 7: Stress-Strain Diagram for
Mild Steel The examples we have examined normal stress in the y direction
to this point have been examples and shear stresses on its surfaces.
of uniaxial forces and stresses.
In experimental stress analysis,
The element rotated to the p-q E
From Figure 7, we can see that, up axes orientation has normal stress
to a point called the proportional
limit, there is a linear relationship
between stress and strain. Hookes q

Law describes this relationship. The Y p


XY
slope of this straight-line portion q
of the stress-strain diagram is x
x
the MODULUS OF ELASTICITY p q
p

or YOUNGS MODULUS for


the material. The modulus of
elasticity, E, has the same units X

as stress (force per unit area) and


is determined experimentally for Figure 9: Element on X-Y Axes and Principal Axes
E-98
practical Strain Gage measurements

in both directions but zero shear analyst uses measured strains in part would be locally higher than
stress as it should, by definition, if conjunction with other properties the measured value because of
the p-q axes are the principal axes. of the material to calculate the the reduced cross-sectional area
s
The normal stresses, p and s q stresses for a given loading carrying the load. The stresses
, are the maximum and minimum condition. There are methods of will also be highest in the narrow
normal stresses for the point. The measuring strain or deformation region; the part will rupture there
strains in the p-q direction are also based on various mechanical, before the measured average strain
the maximum and minimum, and optical, acoustical, pneumatic, and value indicates a magnitude of
there is zero shear strain along electrical phenomena. This section stress greater than the yield point
these axes. Appendix C gives the briefly describes several of the more of the material as a whole.
equations relating stress to strain for common methods and their
Ideally, we want the strain
the biaxial stress state. relative merits.
measuring device to have an
If we know the orientation of the infinitesimal gage length so we can
principal axes, we can then measure
the strain in those directions
Gage Length measure strain at a point. If we had
this ideal strain gage, we would
and compute the maximum and The measurement of strain is the place it in the narrow portion of the
minimum normal stresses and measurement of the displacement specimen in Figure 10 to measure
the maximum shear stress for a between two points some distance the high local strain in that region.
given loading condition. We dont apart. This distance is the GAGE Other desirable characteristics for
always know the orientation of the LENGTH and is an important this ideal strain measuring device
principal axes, but if we measure the comparison between various strain would be small size and mass,
strain in three separate directions, measurement techniques. Gage easy attachment, high sensitivity to
we can compute the strain in any length could also be described as strain, low cost and low sensitivity
direction including the principal the distance over which the strain is to temperature and other ambient
axes directions. Three- and four- averaged. For example, we could, conditions. [2,6]
element rosette strain gages are on some simple structure such as
used to measure the strain when the part in Figure 10, measure the
the principal axes orientation
is unknown. The equations for
part length with a micrometer both
before and during loading. Then we
Mechanical Devices
computing the orientation and would subtract the two readings to The earliest strain measurement
magnitude of the principal strains get the total deformation of the part. devices were mechanical in nature.
from 3-element rosette strain data Dividing this total deformation by We have already considered an
are found in Appendix C. the original length would yield an example (using a micrometer to
average value of strain for the entire measure strain) and observed
For further study of the mechanics
part. The gage length would be the a problem with that approach.
of materials, refer to Items 1, 4, and
original length of the part. Extensometers are a class of
6 referenced in the Bibliography.
mechanical devices used for
Properties of several common If we used this technique on the
measuring strain that employ a
engineering materials are listed in part in Figure 10, the strain in
system of levers to amplify minute
Appendix A. the reduced width region of the


MEASURING Area of High Local
Strain and Stress
STRAIN

S tress in a material cant be


measured directly. It must be
computed from other measurable
parameters. Therefore, the stress
F F

Figure 10

E-99
practical Strain Gage measurements

Figure 11: Large Area Strain Gages

strains to a level that can be read. gages have been constructed, but
A minimum gage length of 12 inch Optical Methods sensitivity to vibration, mounting
and a resolution of about 10 me is
Several optical methods are used
difficulties, and complex circuit
the best that can be achieved with requirements keep them from being
for strain measurement. One very practical for stress analysis
purely mechanical devices. The
of these techniques uses the work. These devices are, however,
addition of a light beam and mirror

STRAIN GAGES
interference fringes produced by often employed in transducers.
arrangements to extensometers
optical flats to measure strain. This The piezoelectric effect of certain
improves resolution and shortens
device is sensitive and accurate, crystals has also been used to
gage length, allowing
2 me resolution and gage lengths
but the technique is so delicate that
laboratory conditions are required
measure strain. When a crystal
down to 14 inch. strain gage is deformed or strained,
for its use. Item 5 referenced in a voltage difference is developed
Still another type of device, the Bibliography gives excellent across the face of the crystal. This
the photoelectric gage, uses a
combination of mechanical, optical,
introductions to the optical methods
of photoelasticity, holography, and
voltage difference is proportional to E
the strain and is of a relatively high
and electrical amplifications to the moir method of strain analysis. magnitude. Crystal strain gages are,
measure strain. This is done [2,5] however, fairly bulky, very fragile,
by using a light beam, two fine and not suitable for measuring
gratings and a photocell detector to static strains.
generate an electrical current that
is proportional to strain. This device
Electrical Devices Probably the most important
comes in gage lengths as short as Another class of strain measuring electrical characteristic which
1
16 inch, but it is costly and delicate. devices depends on electrical varies in proportion to strain is
All of these mechanical devices characteristics which vary in electrical resistance. Devices
tend to be bulky and cumbersome proportion to the strain in the body whose output depends on this
to use, and most are suitable only to which the device is attached. characteristic are the piezoresistive
for static strain measurements. Capacitance and inductance strain or semiconductor gage, the carbon-

E-100
practical Strain Gage measurements

resistor gage, and the bonded The bonded resistance strain gage the same basic principle, and both
metallic wire and foil resistance is by far the most widely used strain can be treated in the same fashion
gage. The carbon-resistor gage measurement tool for todays from the measurement standpoint.
is the forerunner of the bonded experimental stress analyst. It The semiconductor gage, having
resistance wire strain gage. It is consists of a grid of very fine wire a much higher sensitivity to strain
low in cost, can have a short (or, more recently, of thin metallic than metallic gages, can have other
gage length, and is very sensitive foil) bonded to a thin insulating considerations introduced into its
to strain. A high sensitivity to backing called a carrier matrix. measurement. We will use the
temperature and humidity are The electrical resistance of this grid term STRAIN GAGE or GAGE to
the disadvantages of the carbon- material varies linearly with refer to the BONDED METALLIC
resistor strain gage. strain. In use, the carrier matrix is FOIL GRID RESISTANCE STRAIN
attached to the test specimen with GAGE throughout the rest of this
The semiconductor strain gage is
an adhesive. application note. These foil gages
based on the piezoresistive effect
are sometimes referred to as metal-
in certain semiconductor materials When the specimen is loaded,
film gages.
such as silicon and germanium. the strain on its surface is
transmitted to the grid material by Strain gages are made with a
Semiconductor gages have elastic printed circuit process using
the adhesive and carrier system.
behavior and can be produced to conductive alloys rolled to a thin foil.
The strain in the specimen is found
have either positive or negative The alloys are processed, including
by measuring the change in the
resistance changes when strained. controlled-atmosphere heat treating,
electrical resistance of the grid
They can be made physically small to optimize their mechanical
material. Figure 12 is a picture of a
while still maintaining a high properties and temperature
bonded resistance strain gage with
nominal resistance. The strain coefficients of resistance. A grid
a Constantan foil grid and polyimide
limit for these gages is in the configuration for the strain sensitive
1000 to 10000 me range, with
carrier material. The bonded
element is used to allow higher
most tested to 3000 me in tension.
resistance strain gage is low in cost,
can be made with a short gage values of gage resistance while
Semiconductor gages exhibit a high maintaining short gage lengths.
length, is only moderately affected
sensitivity to strain, but the change Gage resistance values range
by temperature changes, has small
in resistance with strain is nonlinear. from 30 to 3000 , with 120 and
physical size and low mass, and
Their resistance and output are 350 being the most commonly
has fairly high sensitivity to strain. It
temperature sensitive, and the high used values for stress analysis.
is suitable for measuring both static
output, resulting from changes in Gage lengths from 0.008 inch to 4
and dynamic strains. The remainder
resistance as large as 10-20%, inches are commercially available.
of this application note deals with
can cause measurement The conductor in a foil grid gage
the instrumentation considerations
problems when using the has a large surface area for a
for making accurate, practical strain
devices in a bridge circuit. given cross-sectional area. This
measurements using the bonded
However, mathematical corrections keeps the shear stress low in the
resistance strain gage. [2, 5, 6]
for temperature sensitivity, the adhesive and carrier matrix as the
nonlinearity of output, and the
nonlinear characteristics of the  strain is transmitted by them. This
larger surface area also allows
bridge circuit (if used) can be made
automatically when using computer- the bonded good heat transfer between grid

resistance
and specimen. Strain gages are
controlled instrumentation to small and light, operate over a
measure strain with semiconductor
strain gage wide temperature range, and can

T
gages. They can be used to respond to both static and dynamic
measure both static and dynamic strains. They have wide application
strains. When measuring dynamic and acceptance in transducers as
strains, temperature effects are he term bonded resistance
strain gage can apply to the well as in stress analysis.
usually less important than for static
strain measurements and the high nonmetallic (semiconductor) gage In a strain gage application, the
output of the semiconductor gage is or to the metallic (wire or foil) gage. carrier matrix and the adhesive must
an asset. Wire and foil gages operate on work together to faithfully transmit

E-101
practical Strain Gage measurements

STRAIN GAGES
Figure 12: Foil Bonded
Resistance Strain Gages

E
strain from the specimen to the gage factor
grid. They also act as an electrical When a metallic conductor is
insulator between the grid and the of carrier material, grid alloy, strained, it undergoes a change in
specimen and must transfer heat adhesive, and protective coating electrical resistance, and it is this
away from the grid. Three primary for the given application cannot change that makes the strain gage
factors influencing gage selection be over-emphasized. Strain gage a useful device. The measure of
are 1) operating temperature; 2) manufacturers are the best source this resistance change with strain is
state of strain (including gradients, of information on this topic and GAGE FACTOR, GF. Gage factor is
magnitude and time dependence); have many excellent publications to defined as the ratio of the fractional
and 3) stability requirements for the assist the customer in selecting the change in resistance to the
gage installation. The importance proper strain gages, adhesives and fractional change in length (strain)
of selecting the proper combination protective coatings. along the axis of the gage. Gage
E-102
practical Strain Gage measurements

factor is a dimensionless quantity,


and the larger the value, the more
sensitive the strain gage. Gage Overall Pattern Length
factor is expressed in equation

Transition
form as: Gage Length Solder Tab Length
End Loops

GF =
DR/R = DR/R
DL/L e Tab Spacing Grid Width

Equation No. 10
Solder Tab Width
Outer Grid Lines Grid Center
It should be noted that the change Alignment Marks Inner Grid Lines
in resistance with strain is not due
solely to the dimensional changes
in the conductor, but that the
resistivity of the conductor material Figure 13: Strain Gage Nomenclature
also changes with strain: The term
gage factor applies to the strain
gage as a whole, complete with transverse direction.
carrier matrix, not just to the strain- TRANSVERSE SENSITIVITY Temperature
sensitive conductor. The gage factor FACTOR, Kt, is defined as: Effects
for Constantan and nickel-chromium
alloy strain gages is nominally 2, and Ideally, we would prefer the strain
GF (transverse)
various gage and instrumentation Kt = gage to change resistance only in
specifications are usually based on GF (longitudinal) response to stress-induced strain in
this nominal value. the test specimen, but the resistivity
and is usually expressed in percent. and strain sensitivity of all known
Values of K range from 0 to 10%. strain-sensitive materials vary with
Transverse temperature. Of course this means
Sensitivity To minimize this effect, extra that the gage resistance and the
gage factor will change when the
material is added to the conductor
if the strain gage were a single in the end loops, and the grid temperature changes. This change
straight length of conductor of small lines are kept close together. This in resistance with temperature for
diameter with respect to its length, serves to minimize resistance in a mounted strain gage is a function
it would respond to strain along its the transverse direction. Correction of the difference in the thermal
longitudinal axis and be essentially for transverse sensitivity may expansion coefficients between the
insensitive to strain applied be necessary for short, wide- gage and the specimen and of the
perpendicularly or transversely to grid gages, or where there is thermal coefficient of resistance of
this axis. For any reasonable value considerable misalignment between the gage alloy. Self-temperature-
of gage resistance, it would also the gage axis and the principal axis, compensating gages can be
have a very long gage length. When or in rosette analysis where high produced for specific materials by
the conductor is in the form of a grid transverse strain fields may exist. processing the strain-sensitive alloy
to reduce the effective gage length, Data supplied by the manufacturer in such a way that it has thermal
there are small amounts of strain- with the gage can be entered into resistance characteristics that
sensitive material in the end loops the computer that controls the compensate for the effects of the
or turn-arounds that lie transverse to instrumentation, and corrections mismatch in thermal expansion
the gage axis. See Figure 13. This for transverse sensitivity can thus coefficients between the gage and
end loop material gives the gage a be made to the strain data as it is the specific material. A temperature
non-zero sensitivity to strain in the collected. compensated gage produced in this
E-103
practical Strain Gage measurements

Figure 14: Typical Temperature-Induced Apparent Strain

manner is accurately compensated because of the nonlinear character or 240 micro-ohms. That means we
only when mounted on a material of both the thermal coefficient

STRAIN GAGES
need to have micro-ohm sensitivity
that has a specific coefficient of of expansion and the thermal in the measuring instrumentation.
thermal expansion. Table 2 is a list coefficient of resistance. Since it is the fractional change in
of common materials for which self-
resistance that is of interest, and
temperature-compensated gages
since this change will likely be only in
are available.
The Measurement the tens of milliohms, some reference
point is needed from which to begin
Approximate Thermal
E
Expansion Coefficient From the gage factor equation, the measurement. The nominal value
Material PPM/C we see that it is the FRACTIONAL of gage resistance has a tolerance
Quartz 0.5 CHANGE in resistance that is the equivalent to several hundred
Titanium 9 important quantity, rather than the microstrain, and will usually change
Mild Steel 11 absolute resistance value of the when the gage is bonded to the
Stainless Steel 16 gage. Lets see just how large this specimen, so this nominal value cant
Aluminum 23 resistance change will be for a strain be used as a reference.
Magnesium 26 of 1me . If we use a 120 strain
An initial, unstrained gage
Table 2: Thermal Expansion Coefficients gage with a gage factor of
of Some Common Materials for Which +2, the gage factor equation tells us resistance is used as the reference
Temperature Compensated Strain Gages
Are Available
that 1me applied to a 120 gage against which strain is measured.
Typically, the gage is mounted on
produces a change in resistance of
the test specimen and wired to the
The compensation is only effective DR = 120 x 0.000001 x 2 = 0.000240 instrumentation while the specimen
over a limited temperature range is maintained in an unstrained state.
E-104
practical Strain Gage measurements

A reading taken under these application to strain gage specimen to which it is attached.
conditions is the unstrained measurement. By using a VOUT is a function of VIN, R1, R2, R3
reference value, and applying a computer in conjunction with the and Rg. This relationship is:
strain to the specimen will result measurement instrumentation,
in a resistance change from this we can simplify use of the bridge
R3 R2
value. If we had an ohmmeter
that was accurate and sensitive
circuit, increase measurement
accuracy, and compile large
VOUT = VIN [ R3 + Rg

R1 + R2 ]
enough to make the measurement, quantities of data from multichannel Equation No. 11
we would measure the unstrained systems. The computer also
gage resistance and then subtract removes the necessity of balancing
When (R1/R2) = (Rg/R3), VOUT
this unstrained value from the the bridge, compensates for
becomes zero and the bridge is
subsequent strained values. nonlinearities in output, and
balanced. If we could adjust one of
Dividing the result by the unstrained handles switching and data storage
the resistor values (R2, for
value would give us the fractional in multichannel applications.
resistance change caused by
strain in the specimen. In some
cases, it is practical to use this very
method, and these cases will be
discussed in a later section of this
application note. A more sensitive
way of measuring small changes R1 Rg
in resistance is with the use of a
Wheatstone bridge circuit, and,
in fact, most instrumentation for
measuring static strain uses this
circuit. [2,5,6,7,8] R2 V OUT

R3

measurement
methods
Figure 16: Bridge Circuit with Provision for Balancing the Bridge

wheatstone Bridge Balanced Bridge Strain


example), then we could balance
circuit Gage Measurement
the bridge for varying values of the
other resistors. Figure 16 shows a
Because of its outstanding schematic of this concept.
In Figure 15, VIN is the input
sensitivity, the Wheatstone bridge
voltage to the bridge, Rg is the Referring to the gage factor
circuit (depicted in Figure 15) is
resistance of the strain gage, R1, equation,
the most frequently used circuit for
R2 and R3 are the resistances of
static strain measurement. This
the bridge completion resistors, and
section examines this circuit and its DRg/Rg
VOUT is the bridge output voltage. A GF =
Rg
1
4 bridge configuration exists when e
R1 one arm of the bridge is an active Equation No. 10
gage and the other arms are fixed
VIN VOUT
value resistors or unstrained gages,
we see that the quantity we need to
as is the case in this circuit. Ideally,
measure is the fractional change in
the strain gage, Rg, is the only
R2 R3 gage resistance from the unstrained
resistor in the circuit that varies,
value to the strained value. If, when
and then only due to a change in
Figure 15:Wheatstone Bridge Circuit the gage is unstrained, we adjust R2
strain on the surface of the
E-105
practical Strain Gage measurements

until the bridge is balanced and becomes cumbersome to the This new equation is:
then apply strain to the gage, the user. Multichannel systems, under
change in Rg due to the strain computer control, eliminate
will unbalance the bridge and these adjustments by using an DRg -4Vr
=
VOUT will become nonzero. If we unbalanced bridge technique. Rg 1 + 2Vr
adjust the value of R2 to once Equation No. 14
again balance the bridge, the
amount of the change required in Unbalanced Note that it was assumed in this
resistance R2 will equal the change Bridge Strain Gage derivation that DRg was the only
in Rg due to the strain. Some strain Measurement change in resistance from the
indicators work on this principle unstrained to the strained condition.
by incorporating provisions for The equation for V OUT can be
Recalling the equation for gage
inputting the gage factor of the rewritten in the form of the ratio of
factor:
gage being used and indicating the VOUT to VIN:
change in the variable resistance,
R2, directly in micro-strain.
R3 R2 DRg/Rg
In the previous example, the bridge
becomes unbalanced when strain
VOUT
VIN
= [ R3 + Rg

R1 + R2 ] GF =
e
Equation No. 10
is applied. VOUT is a measure of this Equation No. 12
imbalance and is directly related and combining these two equations,
to the change in Rg, the quantity of This equation holds for both the
we get an equation for strain in
interest. Instead of rebalancing the unstrained and strained conditions.
terms of Vr and GF:
bridge, we could install an indicator, Defining the unstrained value of
calibrated in micro-strain, that gage resistance as Rg and the
responds to VOUT. Refer to Figure change due to strain as DRg, the -4Vr
16. If the resistance of this indicator strained value of gage resistance e =
GF(1 + 2Vr )
is much greater than that of the is Rg + DRg. The actual effective
strain gage, its loading effect on the value of resistance in each bridge Equation No. 15
bridge circuit will be negligible, i.e., arm is the sum of all the resistances
negligible current will flow through in that arm, and may include The schematic in Figure 17
such things as lead wires, printed shows how we can instrument the

STRAIN GAGES
the indicator. This method often
assumes: 1) a linear relationship circuit board traces, switch contact unbalanced bridge.
between VOUT and strain; 2) a resistance, interconnections, etc.
A constant voltage power supply
bridge that was balanced in the As long as these resistances
furnishes VIN, and a digital voltmeter
initial, unstrained, state; and 3) remain unchanged between the
(DVM) is used to measure VOUT.
a known value of VIN. In a bridge strained and unstrained readings,
The DVM for this application
circuit, the relationship between the measurement will be valid.
should have a high (greater than
E
VOUT and strain is nonlinear, but for Lets define a new term, Vr, as the
109 ) input resistance, and 1
strains up to a few thousand micro- difference of the ratios of VOUT
microvolt or better resolution. The
strain, the error is usually small to VIN from the unstrained to the
gage factor is supplied by the
enough to be ignored. At large strained state:
gage manufacturer. In practice,
values of strain, corrections must we would use a computer to have
be applied to the indicated reading VOUT VOUT the DVM read and store VOUT
to compensate for this nonlinearity. Vr = under unstrained conditions, then
VIN strainedVIN unstrained
The majority of commercial strain take another reading of VOUT
indicators use some form of Equation No. 13 after the specimen is strained.
balanced bridge for measuring Since the values for gage factor
resistance strain gages. In By substituting the resistor values and excitation voltage, VIN, are
multichannel systems, the number that correspond to the two (VOUT/ known, the computer can calculate
of manual adjustments required VIN) terms into this equation, we the strain value indicated by the
for balanced bridge methods can derive an equation for DRg/Rg. change in bridge output voltage.

E-106
practical Strain Gage measurements

the strained reading, except for


the change in the gage resistance
due to strain. If any of the bridge
R1 Rg arm resistances changed during
Constant
Voltage (VIN ) that time span, there would be a
Power DVM
corresponding change in bridge
Supply output voltage which would be
V OUT interpreted as strain-induced, so
we would see an error. The same
would be true of any other variation
R2 R3 that changed the bridge output
voltage. Any switching done in the
bridge arms can cause a change in
resistance due to variations in the
switch or relay contact resistance
Figure 17:Instrumentation for Unbalanced Bridge Strain Gage Measurement and can affect the bridge output
voltage. For that reason, it is not
If the value of VIN is unknown or VIN and the gage factor were desirable to do switching inside
subject to variation over time, both known quantities. the bridge arms for multichannel
we can have the DVM measure systems, but, rather, to allow
it at the time VOUT is measured Appendix B shows the schematics those interconnections to be
to get a more precise value for of several configurations of bridge permanently wired and switch the
Vr . This timely measurement of circuits using strain gages, and DVM from bridge to bridge. Since
VIN greatly reduces the stability gives the equation for strain as a a DVM has extremely high input
requirements of the power supply, function of Vr for each. impedance compared to the bridge
allowing a lower-cost unit to be arms, it doesnt load the bridge,
used. Note that, in the preceding and switching the DVM has no
1
4 bridge example, the bridge was Multichannel effect on the bridge output voltage
not assumed to be balanced nor Wheatstone Bridge level. Figures 18 and 19 show the
its output approximated as truly Measurements schematics of these two methods
linear. Instead, we just derived of switching. We can see that
the equation for strain in terms of In the preceding example, switching inside the bridge arms
quantities that are known or can measurement accuracy was allows the same bridge completion
be measured, and let the computer dependent upon all four bridge resistors to be used for multiple
solve the equation to obtain the arms resistances remaining gages, but that the power to the
exact strain value. constant from the time of the gage is removed when it is not
unstrained reading to the time of
In the preceding example, we
made some assumptions that
affect the accuracy of the strain
measurement:
0 1 Rg1
resistance in the three inactive Rg2
bridge arms remained constant 2

from unstrained to strained R1

readings, 0 1
V IN V OUT
DVM accuracy, resolution, and 2

stability were adequate for the


R2 R3
required measurement,
resistance change in the active
bridge arm was due only to
change in strain, and Figure 18: Switching Inside Bridge Arms
E-107
practical Strain Gage measurements

Rg1 Rg2
R1
Constant DVM 0 1
Voltage (VIN) D
C E
Power (VOUT) 2
Supply 3
R2
Rg2
R3

Figure 19: Schematic of Bridge Circuit with Shared Internal Half-Bridge and Power Supply

being read. Also, any variations in excitation voltage, VIN, must be become inoperative. However, the
switch contact resistance will appear measured across points A-B, and measurement of the excitation
in series with the gage resistance it may be desirable to measure
voltage permits the power
and will be indistinguishable from this voltage each time a new set of
supply to drift, be adjusted, or
resistance changes due to strain. readings is taken from this group
even be replaced with no loss in
of channels. The DVM is switched
Figure 19 shows a multiple-channel measurement accuracy.
between points C-D, C-E, etc., to
arrangement that switches the DVM
read the output voltages of the
and also shares the power supply
various channels in the group. This
and internal half-bridge. This circuit
method keeps all of the gages Four-Wire OhM Strain
is known as a Chevron Bridge and
is often used for strain measurement
energized at all times, which Gage Measurement
minimizes dynamic heating and
on rotating machine elements to As we mentioned before, we can
cooling effects in the gages and

STRAIN GAGES
minimize the number of slip rings. measure the change in absolute
eliminates the need for switching
One channel is shown as a 14 bridge value of gage resistance to compute
inside the bridge arms. If the DVM
and the other as a 12 bridge (two strain. This can be done quite
has good low-level measurement
active gages). The midpoint of the accurately using a four-wire
capability, the power supply voltage
internal half-bridge for either of measurement technique with a high
can be maintained at a low level,
these configurations serves as a resolution (e.g., 1 milliohm per least
thereby keeping the gages self-
voltage reference point for the DVM significant digit) digital multimeter
heating effects to a minimum. For
and isnt affected by strain. Since
the bridge completion resistors
example, using a 2 volt power
(DMM). Figure 20 depicts the four-
wire method of resistance
E
supply for the bridge yields a power
must have excellent stability measurement. The current source
dissipation, in a 350 gage, of
specifications, they are relatively is connected internally in the DMM
only 3 milliwatts. Yet even with this
expensive, and there is a cost
low power, 1 microstrain sensitivity
advantage to sharing the internal Rl
is still maintained with a 14 bridge
half-bridge in multichannel systems.
configuration (assuming GF=2), Rl
For this method to function properly, when using a DVM with 1 microvolt
V Rg
the circuit must be designed and resolution. Since several channels Rl

constructed such that a change are dependent upon one power Current Voltmeter
Source Rl
in current due to strain in one arm supply and one resistor pair, a
does not change the current in failure of one of these components
any of the other arms. Also, the will cause several channels to Figure 20:Schematic of Four-Wire
Ohm Circuit

E-108
practical Strain Gage measurements

to the source terminals, while the To use four-wire for measuring constant voltage supply which is
voltmeter is connected to the strain, we first make a resistance shared by multiple bridges cannot
sense terminals of the DMM. When measurement of the gage in the be directly replaced by a constant
a measurement is taken, the current unstrained condition and store this current source, since we wouldnt
source supplies a known fixed reading. Then we apply strain to know how the current was divided
value of direct current through the the specimen and make another among the various bridge circuits.
circuit from the source terminals, measurement of gage resistance. In some cases, the bridge output
while the voltmeter measures the The difference between these two is more nearly linear when using
dc voltage drop across the gage readings divided by the unstrained constant current rather than
resistance. The absolute resistance reading is the fractional change in constant voltage excitation, but
value is computed from the values resistance that we use in the gage that is of little consequence if we
of current and voltage by the factor equation to compute strain. solve an equation for strain versus
DMM and displayed or output to a Of course the DMM can input these output voltage with a computer.
computer. The lead resistances, R l readings directly to a computer, The use of a constant current
, from the source terminals to the which calculates strain using the source for a full-bridge configuration
gage, are in series with the gage gage factor for the particular gage. does eliminate the need to sense
resistance, but do not affect the This technique also lends itself the voltage at the bridge, which
accuracy of the measurement, since to multichannel systems, since eliminates the need to run two wires
the voltage is read directly across variations in switch resistance in the to the bridge. In general, there is
the gage. The input impedance to circuit have the same effect as lead no real measurement advantage
the sense terminals is extremely resistances and do not affect the to using constant current rather
high, so the current flow in that loop accuracy of the measurement. than constant voltage excitation for
is negligible. The source current bridge circuits as applied to strain
value is typically very low, which gage measurements.
means the power dissipated in the Constant Current The four-wire measurement
strain gage is also very low, and Techniques discussed in the preceding section
self-heating effects are virtually used a constant current source
eliminated. For example, In the discussion of bridge circuits,
for excitation, and we noted that
1 milliamp is a typical value we assumed that the bridge
the lead wires had no effect on
for the source current, and this excitation was furnished by a
the measurement. That method
corresponds to a power dissipation constant voltage source. We could
required four wires to be connected
of 120 microwatts in a 120 gage have assumed constant current
to the gage. Constant current
or 350 microwatts in a 350 gage. excitation for those discussions
excitation is sometimes used with
and derived the corresponding
A technique for voltage offset a two-wire gage connection for
equations for strain as a function of
compensation can be used with dynamic strain measurements
voltage out and current supplied.
four-wire measurements to In the example of Figure 19, the
where temperature drift effects are
correct for these effects. This is
accomplished by first measuring
the voltage across the gage without Rl
current flow from the source
terminals, and then subtracting
this value from the voltage read
with source current flow. The
resulting net voltage is then used
to compute the gage resistance. V Rg
Offset compensated four-wire
measurements can be made Current
automatically by the DMM if it Source Voltmeter
has that capability, or the offset
compensation can be accomplished Rl
by the computer controlling the
instrumentation. Figure 21:Constant Current Circuit Dynamic Strain Measurement
E-109
practical Strain Gage measurements

negligible or can be filtered out also be used in a circuit similar to of the possible sources of difficulty.
from the strain data. In the circuit that for four-wire (see Figure 22). The results of this type of electrical
of Figure 21, changes in gage In this case, the current source interference can range from a
resistance result in proportional and the DVM should be separate negligible reduction in accuracy
changes in VOUT. Note that VOUT instruments, to allow the current to deviances that render the data
is also affected by changes in the level to be adjusted to obtain the invalid.
lead resistances, Rl . By measuring best output voltage for the expected
only the time-varying component maximum strain level.
of VOUT, the dynamic strain can be
observed, while slowly- changing
The lead wires do not affect the The Noise Model
measurement, since the voltage,
effects, such as temperature, are In Figure 23, the shaded portion
as in four-wire , is measured
rejected. includes a Wheatstone bridge
directly across the gage. This
strain gage measuring circuit seen
The use of very sensitive DMMs arrangement also allows the use of
previously in Figures 15 and 17.
to measure the bridge imbalance a less sensitive, higher speed DVM
The single active gage, Rg, is shown
voltage or the gage resistance while maintaining reasonable strain
mounted on a test specimen
directly with four-wire limits the resolution. For example, a DVM
e.g., an airplane tail section. The
speed at which the measurement with 100 microvolt sensitivity gives a
bridge excitation source, VIN , bridge
can be taken, and only low strain resolution of 6 me with a
completion resistors, R1 , R2 and
frequency dynamic strains can be 0.44 milliamp current source
R3, and the DVM represent the
measured with these methods. (350 semiconductor gage with
me
measurement equipment located a
Higher speed analog-to-digital GF = 100).
significant distance (say, 100 feet)
converters typically have lower
from the test specimen. The strain
sensitivities, so higher signal levels
gage is connected to the measuring
are needed when measuring higher
Shielding and Guarding equipment via three wires having
frequency dynamic or transient
strains. One way to achieve this is
Interference Rejection resistance Rl in each wire. The
electrical interference which
to amplify the bridge output voltage The low output level of a strain
degrades the strain measurement
to an acceptable level. Another gage makes strain measurements
is coupled into the bridge through
method is to use a semiconductor susceptible to interference from
a number of parasitic resistance
strain gage and exploit its large other sources of electrical energy.
and capacitance elements. In this
gage factor. A semiconductor gage Capacitive and magnetic coupling to

STRAIN GAGES
context, the term parasitic implies
can be used in a bridge circuit long cable runs, electrical leakage
that the elements are unnecessary
(such as Figure 19) with a DVM from the specimen through the
to the measurement, are basically
having lower resolution and higher gage backing, and differences in
unwanted, and are to some extent
speed than that required with metal grounding potential are but a few
unavoidable. The parasitic elements
gages. A semiconductor gage can
result from the fact that lead wires
have capacitance to other cables,

Rl
gages have capacitance to the test
specimen, and gage adhesives
E
Constant and wire insulation are not perfect
Current Rl
insulators giving rise to leakage
Source resistance.
Rg Examining the parasitic elements
High Speed
(Semiconductor) in more detail, the active gage Rg
DVM
Gage is shown to be made up of two
Rl equal resistors with Ciso connected
at the center. Ciso represents
the capacitance between the
Rl airplane tail section and the gage
foil. Since the capacitance is
Figure 22:Circuit for Semiconductor Gage and High Speed Digital Voltmeter
E-110
practical Strain Gage measurements

Adjacent Power or Signal Cable

Cc Rc
Measurement Unit Rl

+ R1 Rg
+ Rl
Vs VIN +
DVM

R2 Ciso
R3 Rl

Riso
Cps Rps 100 FT

Vcm
Measurement Earth
Connection

Figure 23:Remote Quarter-Bridge Measurement Illustrating Parasitic Elements and Interference Sources

distributed uniformly along the gage mechanisms causing electrical


grid length, we approximate the interference in a practical, industrial
powered, floating output power
effect as a lumped capacitance environment. Typical values for
supply usually has no deliberate
connected to the gages midpoint. Cps and Rps for floating output,
electrical connection between
Riso and Ciso determine the degree laboratory grade power supplies
the negative output terminal and
of electrical isolation from the test are 0.01 mf and 100 megohms,
earth via the third wire of its power
specimen, which is often electrically respectively. It is important to
cord. However, relatively large
grounded or maintained at some realize that neither the measuring
amounts of capacitance usually
floating potential different than equipment nor the gages have been
exist between the negative output
the gage. Typical values of Riso grounded at any point. The entire
terminal circuits and the chassis and
and Ciso are 1000 megohms and system is floating to the extent
between the primary and secondary
100 pF, respectively. Elements allowed by the parasitic elements.
windings of the power transformer.
Cc and Rc represent the wire-to-
The resistive element Rps is caused To analyze the sources of
wire capacitance and insulation
by imperfect insulators, and can electrical interference, we must
resistance between adjacent power
be reduced several decades by first establish a reference potential.
or signal cables in a cable vault or
ionic contamination or moisture due Safety considerations require that
cable bundle. Typical values for Cc
to condensation or high ambient the power supply, DVM, bridge
and Rc are 30 pF and 1012 per
humidity. If the power supply completion, etc., cabinets all be
foot for dry insulated conductors in
does not feature floating output, connected to earth ground through
close proximity.
Rps may be a fraction of an . It the third wire of their power cords.
The power supply exciting the will be shown that use of a non- In Figure 23, this reference potential
bridge is characterized by parasitic floating or grounded output power is designated as the measurement
elements Cps and Rps. A line- supply drastically increases the earth connection. The test
E-111
practical Strain Gage measurements

Cc Rc
Rl

+ R1 Rg
+ Rl
Vs Ciso
VIN +
DVM

R2
R3 Rl
Riso

Cps Rps

Vcm

Figure 24:Current Leakage from Adjacent Cable Flows Through Gage Wires Causing Measurement Error

specimen is often grounded an electric motor). The disturbing


is flowing through the gage
(for safety reasons) to the power source, Vs, is shown connected to
measurement leads as a result of
system at a point some distance measurement earth and represents
the source Vs. Negligible current

STRAIN GAGES
away from the measurement the electrical potential of some cable
flows through the DVM because of
equipment. This physical separation in close proximity (but unrelated
its high impedance. The currents
often gives rise to different functionally) to the gage wires. In
through Rg and Rl develop error-
grounding potentials as represented many applications, these adjacent
producing IR drops inside the
by the voltage source Vcm. In some cables may not exist or may be so
measurement loops.
cases, functional requirements far removed as to not affect the
dictate that the test specimen be measurement. They will be included In Figure 25, a shield surrounds the
floated or maintained many volts
away from the power system ground
here to make the analysis general
and more complete.
three measurement leads, and the
current has been intercepted by
E
by electronic power supplies or the shield and routed to the point
signal sources. In either case, Vcm where the shield is connected to
may contain dc and time- varying
Shielding of the bridge. The DVM reading error
components most often at power-
line related frequencies.
Measurement Leads has been eliminated. Capacitive
coupling from the signal cable to
The need for using shielded unshielded measurement leads
Typical values of Vcm, the common
measurement leads can be seen by will produce similar voltage errors,
mode voltage, range from
examining the case shown in even if the coupling occurs equally
millivolts due to IR drops in clean
Figure 24. Here, an insulation to all three leads. In the case
power systems to 250 volts for
failure (perhaps due to moisture) where Vs is a high voltage sine
specimens floating at power-line
has reduced parasitic dc to a wave power cable, the DVM error
potentials (for example, parts of
few thousand , and dc current will be substantially reduced if the
E-112
practical Strain Gage measurements

Cc Rc
Measurement Unit Rl

+ R1 Rg
+ Rl Ciso
Vs VIN +
DVM

R2
R3 Rl
Riso

Cps Rps

Vcm

Figure 25:Addition of a Metal Shield Around the Gage Wires Keeps Current Due to Vs out of Measurement Leads

voltmeter integrates the input for reducing measurement degradation entering the measurement loop
a time equal to an integral number as a result of magnetic induction. via the gage parasitic elements,
of periods (e.g., 1, 10, or 100) of The flat, three-conductor side-by- Ciso and Riso. In the general case,
the power line wave form. The side, molded cable commonly used both ac and dc components must
exact amount of the error reduction for strain gage work approaches be considered. Again, current flow
depends upon the DVMs normal the effectiveness of a twisted through gage and lead resistances
mode rejection, which can be as pair by minimizing the loop area result in error voltages inside the
large as 60-140 dB or 103:1 - 107:1. between the wires. The use of bridge arms. Tracing either loop
If the DVM is of a type having a shielded, twisted leads and a DVM from the DVMs negative terminal
very short sampling period, i.e., less which integrates over one or more to the positive terminal will reveal
than 100 msec, it will measure the cycles of the power line wave form unwanted voltages of the same
instantaneous value of the dc signal should be considered whenever polarity in each loop. The symmetry
(due to strain) plus interference. leads are long, traverse a noisy of the bridge structure in no way
Averaging the proper number of electromagnetic environment, provides cancellation of the effects
readings can reduce the error due or when the highest accuracy is due to current entering at the gage.
to power line or other periodic required.
interference. Whereas shielding kept error-
producing currents out of the
Where the measurement leads measurement loop by intercepting
run through areas of high Guarding the the current, guarding controls
magnetic field, near high-current Measuring Equipment current flow by exploiting the fact
power cables, etc., using twisted that no current will flow through
measurement leads will minimize Figure 26 shows the error- an electrical component that has
the loop areas formed by the producing current paths due to both of its terminals at the same
bridge arms and the DVM, thereby the common mode source, Vcm, potential.
E-113
practical Strain Gage measurements

Cc Rc
Rl

+ R1 Rg
+ Rl
Vs VIN +
DVM

R2 Ciso
R3 Rl
Riso
Guard Lead
Cps Rps

Vcm

Figure 26:Error-Producing Common Mode Current Path

Cc Rc
Rl

STRAIN GAGES
+ R1 Rg
+ Rl
VIN +
DVM


R2 Ciso
R3 Rl
Riso E
Cps Rps

Vcm

Figure 27
E-114
practical Strain Gage measurements

In Figure 27, a guard lead has In situations where it is possible the DVM and the associated guard
been connected between the test to ground the test specimen at connection.
specimen (in close proximity to the measurement earth potential, the
Figure 29 illustrates the currents
gage) and the negative terminal of common mode source, Vcm, will be
flowing due to the specimen
the power supply. This connection essentially eliminated.
potentials Vcm1 and Vcm2 . Note
forces the floating power supply that, regardless of which channel
and all the measuring equipment is selected, the guard line (also
including the gage to the Extension to functioning as the shield for the
same electrical potential as the Multichannel wires to the gage) keeps the
test specimen. Since the gage Measurements common mode current out of
and the specimen are at the same the gage leads selected for the
potential, no error-producing current Figure 28 shows the extension of
measurement. Common mode
flows through Riso and Ciso into the the guarding technique to a multi-
current flows harmlessly through
measuring loops. Another way of channel strain gage measurement
the gage leads of the unselected
interpreting the result is to say that using a shared power supply and
channel. It should be noted that
the guard lead provides an alternate internal half-bridge completion
each lead wire shield is grounded
current path around the measuring resistors. For simplicity, only the
at only a single point. The
circuit. It should be observed that, if capacitive parasitic elements
common mode current through
the power supply and the rest of the are shown. In ordinary practice,
each combined guard and shield
measuring circuits could not float capacitive coupling is usually more
is limited by the relatively high
above earth or chassis potential, significant and more difficult to
impedance of the parasitic element
the guarding technique would avoid than resistive coupling. For
Cps, and should not be confused
reduce the interference by factors generality, weve used two test
with the heavy shield current
of only 2:1 or 4:1. Proper guarding specimens at different potentials
which might occur if a shield were
with a floating supply should yield with respect to measurement earth.
grounded at both ends, creating a
improvements on the order of The switching shown in the figure
ground loop.
105:1 or 100 dB. allows simultaneous selection of

Shield and Guard Combined

Rg1 Ciso
1

Measurement Unit
R31
+
VIN
Cps
R1
+ DVM
DVM
Switch
R2 Rg2
Ciso2

R32
Guard
Cps Switch

Shield and Guard Combined Vcm +


1

Vcm
2

Figure 28:Multichannel Strain Measurement Including Two Separate Test Specimens


E-115
practical Strain Gage measurements

Rg1 Ciso
1

R31
+
VIN
Cps
R1
+
DVM DVM Switch

R2
Rg2
Ciso2

R32
Guard
Cps Switch

+ Vcm +
1

Vcm
2

Figure 29:Multichannel Guard Switch Keeps Common Mode Current out of Selected Gage Leads

Rg1 Ciso
1

STRAIN GAGES
R31
+
VIN
Cps
R1
+
DVM DVM Switch

R2
Rg2 E
Ciso2

R32
Cps Guard
Switch

+ +
Vcm
1
Vcm
2

Figure 30:Unguarded Capacitance of Multiplexer and DVM Result in Measurement Error Due to Vcm2 of Selected Channel
E-116
practical Strain Gage measurements

A dc voltmeters NMRR is a function and if the rejection ratios are


CMR Limitations of input filtering and the analog- expressed in dB,
to-digital conversion technique
The schematics and discussion
employed.
of guarding presented thus far ECMRR(dB) = CMRR(dB) + NMRR(dB)
might convey the impression Additionally, the DVM and
Equation No. 19
that infinite rejection of common multiplexer system reject ac
mode interference is possible. It interference via guarding and
seems reasonable to ask what, design control of parasitics. The Thus, ECMRR describes how
if anything, limits common mode quantitative measure of a systems well the parasitics are controlled
rejection? Figure 30 includes a ability to reject common mode in the system and the sampling
new parasitic element, Cug, the ac voltage is the common mode characteristics of the DVM, i.e.,
unguarded capacitance to chassis rejection ratio, (CMRR), defined as: integrating or instantaneous
associated with the DVM and sampling.
multiplexer. In practice, the DVM Reference 10 provides additional
and multiplexer are usually realized Vcm (ac)
CMRR = 20 log information on the subjects of
as guarded instruments [Reference VDVM (ac)
floating, guarded measurements
13] featuring three-wire switching Equation No. 17 and rejection ratios. Appendix D
and measurements, but the guard contains measurement sensitivity
isolation is not perfect. Capacitance data which can be used to compute
where Vcm and VDVM are both
ranging from 15 pF to 20 mf can measurement error (in me ) as a
sinusoids at the power-line
be found between the instrument function of DVM, power supply,
frequency of interest - 50, 60,
low connection and chassis. In and bridge completion resistor
or 400 Hz. Note that VDVM is an
Figure 30, this capacitance causes specifications.
ac wave form presented to the
a portion of the common mode
terminals of a dc voltmeter. Thus,
current in the selected channel
CMRR is an ac voltage transfer
to flow through the internal half-
bridge resistors R1 and R2, giving
ratio from the common mode Bridge Excitation Level
source to the DVM terminals.
rise to a measurement error. In The bridge excitation voltage level
Caution must be exercised in
a multichannel system, all of the affects both the output sensitivity
comparing CMRR specifications
unselected channels (gages) and the gage self-heating. From a
to insure that identical procedures
sharing the same power supply measurement standpoint, a high
are employed in arriving at the
also contribute current, but this excitation level is desirable, but a
numerical result.
current exits the bridge via the
power supply and returns through
the guard wire, causing no EMCRR = CMRR x NMRR
additional error. ac interference = ac interference x ac @DVM
In Figure 30, the ac interference dc response error ac @DVM dc response error
}
}
voltage presented to the terminals
of the DVM causes an error transmission dc voltmeter
because the dc measuring via parasitics response to
and guarding ac input
voltmeter does not totally reject
the ac. A DVMs ability to measure Equation No. 18
dc voltage in the presence of ac
interference is called the normal The overall figure of merit for lower level reduces gage self-
mode rejection ratio (NMRR) and a measurement system is the heating. The electrical power in the
is usually stated for 50 and 60 Hz effective common mode rejection gage is dissipated as heat which
interference. ratio (ECMRR), which reflects the must be transferred from the gage
systems ability to measure dc to the surroundings. In order for
VNM (ac) voltage (strain) in the presence of this heat transfer to occur, the gage
NMRR = 20 log
VDVM (dc) ac common mode interference. If temperature must rise above that of
all measurements same frequency, the specimen and the air. The gage
Equation No. 16
E-117
practical Strain Gage measurements

temperature is therefore a function current through the gage, and V is POWER DENSITY is a parameter
of the ambient temperature and the bridge excitation voltage. From used to evaluate a particular gage
the temperature rise due to power Equation 20, we see that lowering size and excitation voltage level
dissipation. the excitation voltage (or gage for a particular application. Power
current) or increasing the gage density is the power dissipated by
An excessive gage temperature
resistance will decrease power the gage divided by the gage grid
can cause various problems. The
dissipation. Where self-heating may area, and is given in units of watts/
carrier and adhesive materials will
be a problem, higher values of gage in2. Recommended values of power
no longer be able to transmit strain
resistance should be used. Table 3 density vary, depending upon
faithfully from the specimen to the
illustrates the relationship between accuracy requirements, from 2-10
grid if the temperature becomes
voltage, gage resistance and power for good heat sinks (such as heavy
too high. This adversely affects
dissipation. aluminum or copper sections),
hysteresis and creep and may
show up as instability under load. The temperature rise of the grid is to 0.01-0.05 for poor heat sinks
Zero or unstrained stability is also difficult to calculate because many (such as unfilled plastics). Stacked
affected by high gage temperature. factors influence heat balance. rosettes create a special problem,
Temperature-compensated gages Unless the gage is submerged in in that the temperature rise of the
suffer a loss of compensation when a liquid, most of the heat transfer bottom gage adds to that produced
the temperature difference between will occur by conduction to the by the two gages above it, and that
the gage grid and the specimen specimen. Generally, cooling of the of the center gage adds to the top
becomes too large. When the gage gage by convection is undesirable gages. It may require a very low
is mounted on plastics, excessive because of the possibility of creating voltage or different voltages for
power dissipation can elevate time-variant thermal gradients each of the three gages to maintain
the temperature of the specimen on the gage. These gradients the same temperature at each gage.
under the gage to the point that the can generate voltages due to the [6 ]
properties of the specimen change. thermocouple effect at the lead One way we can determine the
wire junctions, causing errors in maximum excitation voltage that
The power that must be dissipated
the bridge output voltage. Heat can be tolerated is by increasing
as heat by the gage in a bridge
transfer from the gage grid to the voltage until a noticeable
circuit with equal resistance arms is
the specimen is via conduction. zero instability occurs. We then
given by the following equation:
Therefore, the grid surface area and reduce the voltage until the zero
the materials and thicknesses of is once more stable and without

STRAIN GAGES
P = V2/4Rg = (I2)Rg the carrier and adhesive influence a significant offset relative to the
Equation No. 20
gage temperature. The heat sink zero point at a low voltage. Bridge
characteristics of the specimen are completion resistors also dissipate
also important. power and in practice may be more
where P is the power in watts,
Rg is the gage resistance, I is the susceptible to drift from self-heating
effects than the strain gage. The
stability of the bridge completion
bridge
strain gage power dissipation resistors is related to load-life, and E
gage power in milliwatts maintaining only a fraction of rated
excitation
voltage 1000 500 350 120 power in them will give better long
0.1 0.0025 0.005 0.007 0.021 term stability. If the above method of
0.2 0.010 0.020 0.029 0.083 finding the maximum voltage level
0.5 0.0625 0.125 0.179 0.521 is used, care should be exercised to
1.0 0.250 0.500 0.714 2.083 insure that the power rating of the
2.0 1.000 2.000 2.857 8.333
completion resistors is not exceeded
as the voltage is increased.
3.0 2.250 4.500 6.429 18.750
4.0 4.000 8.000 11.429 33.333 Reducing the bridge excitation
5.0 6.250 12.500 17.857 52.083 voltage dramatically reduces gage
10.0 25.000 50.000 71.400 208.300 power, since power is proportional
Table 3 to the square of voltage. However,

E-118
practical Strain Gage measurements

bridge output voltage is proportional changes in the gage. The other actually Rg + Rl . If we substitute this
to excitation voltage, so reducing it error is known as LEAD WIRE into Equation 14, it becomes:
lowers sensitivity. If the DVM used DESENSITIZATION and becomes
to read the output voltage has significant when the magnitude
1 microvolt resolution, 1 micro-strain of the lead wire resistance
DRg -4Vr Rg + Rl
=
resolution can be maintained with a exceeds 0.1% of the nominal gage Rg 1 + 2Vr Rg
1
4 bridge configuration, using a 2 volt resistance. The significance of this Equation No. 21
bridge excitation level. If the DVM source of error is shown in Table 4.
has 0.1 microvolt resolution, the
Rewriting the equation to solve for
excitation voltage can be lowered
strain, we see that the previous
to 0.2 volts while maintaining the
lead wire desensitization strain equation is in error by a
same strain resolution. From Table (Refer to figure 32) factor of the ratio of the lead wire
3 we see that, at these excitation 1
4 and 12 bridge, 3-wire connections resistance to the nominal gage
levels. the power dissipated by a awg rg = 120 rg = 350 resistance.
350 ohm gage goes from 2.857
18 .54% .19%
to 0.029 milliwatts. Thus, using a
20 .87 .30
sensitive DVM for measuring the -4Vr Rl
e
( )
bridge output permits the use of low 22 1.38 .47 =
excitation voltages and low gage 24 2.18 .75
GF(1 + 2Vr ) 1+
Rg

}
self-heating while maintaining good 26 3.47 1.19
measurement resolution. 28 5.52 1.89 Error Term
30 8.77 3.01 Equation No. 22
The four-wire technique is also a
Magnitudes of computed strain values will be
good way to keep the power in the low by the above percent per 100 feet of hard
gage extremely low. This is due to drawn solid copper lead wire at 25C (77F)
This factor is lead wire
the low value of constant current Table 4 desensitization, and we see from
supplied to the gage by the DMM,
Equation 22 and from Table 4 that
typically 1 milliamp. This current If the resistance of the lead wires the effect is reduced if the lead
(1 milliamp) corresponds to a power is known, the computed values of wire resistance is small and/or the
dissipation of 0.12 milliwatts in a strain can be corrected for LEAD nominal gage resistance is large. If
120 gage and 0.35 milliwatts WIRE DESENSITIZATION. In ignoring this term (1 + Rl /Rg) will
in a 350 gage. With four-wire a prior section, we developed cause an unacceptable error, then
, a gage is energized only equations for strain as a function it should be added to the computer
when it is selected and is actually of the measured voltages for a 14 program such that the strains
being measured by the DMM. As bridge configuration:
mentioned previously, resolution will
be lower using four-wire than with
a bridge, but will be adequate for DRg -4Vr -4Vr
many applications. Rg
=
(1 + 2Vr)
e =
GF(1 + 2Vr )
Equation No. 14 Equation No. 15

Lead Wire Effects These equations are based on the


computed in Equation 15 are
multiplied by this factor. Appendix
In the preceding chapter, reference assumptions that Vr is due solely B gives the equations for various
was made to the effects of lead wire to the change in gage resistance, bridge configurations and the lead
resistance on strain measurement DRg, and that the total resistance of wire resistance compensation terms
for various configurations. In the arm of the bridge that contained that apply to them. Appendix A has
a bridge circuit, the lead wire the gage was Rg. Referring to a table containing the resistance,
resistance can cause two types Figure 32, we see that one of the at room temperature, of some
of error. One is due to resistance lead wire resistances, Rl , is in commonly used sizes of copper
changes in the lead wires that are series with the gage, so the total wire.
indistinguishable from resistance resistance of that bridge arm is

E-119
practical Strain Gage measurements

Rl

R1
Rg

Rl
V IN V OUT

R2 R3

Figure 31:Two-Wire 14 Bridge Connection

Rl

R1
Rg

A Rl
V IN V OUT C

B Rl
R2 R3

STRAIN GAGES
Figure 32:Three-Wire 14 Bridge Connection

The most common cause of with the gage in the bridge arm, effected by using the same size
changes in lead wire resistance further contributing to the lead wire and length wires and keeping them
is temperature change. The desensitization error. physically close together.
copper used for lead wires has a
nominal temperature coefficient
The THREE-WIRE method of Temperature compensation is E
connecting the gage, shown in possible because resistance
of resistance, at 25C/77C, of
Figure 32, is the preferred method changes occur equally in adjacent
0.00385 / C. For the 2-wire
of wiring strain gages to a bridge arms of the bridge and, therefore,
circuit in Figure 31, this effect will
circuit. This method compensates the net effect on the output voltage
cause an error if the temperature
for the effect of temperature of the bridge is negligible. This
during the unstrained reading is
on the lead wires. For effective technique works equally well for
different than the temperature
compensation, the lead wires must 1
4 and 12 bridge configurations.
during the strained reading. Error
have approximately the same The lead wire desensitization
occurs because any change in
nominal resistance, the same effect is reduced over the two-wire
resistance in the gage arm of the
temperature coefficient of resistance connection because only one lead
bridge during this time is assumed
and be maintained at the same wire resistance is in series with the
to be due to strain. Also, both
temperature. In practice, this is gage. The resistance of the signal
lead wire resistances are in series
wire to the DVM doesnt affect the
E-120
practical Strain Gage measurements

measurement, because the current performed on the gage installation resistance of less than 500 M
flow in this lead is negligible due to and instrumentation. In a stress usually indicates the presence of
the high input impedance of analysis application, the entire some type of surface contamination.
the DVM. gage installation cant be calibrated Contamination often shows up as a
as can be done with certain time-varying high resistance shunt
Mathematical correction for lead
transducers. Therefore, potential across the gage, which causes an
wire desensitization requires that
error sources should be examined error in the strain measurement. For
the resistances of the lead wires
prior to taking data. this reason, an isolation resistance
be known. The values given in wire
value of at least 150 M should be
tables can be used, but, for extreme
maintained.
temperatures, measurement of the
wires after installation is required for Mounted Gage A properly mounted gage with
utmost accuracy. Two methods for Resistance fully cured adhesive will usually
arriving at the resistance of the lead have 1000 M or higher isolation
The unstrained resistance of the
wires from the instrumentation side resistance, so any gages with low
gage should be measured after
of the circuit in Figure 32 follow: values should be suspect.[2]
the gage is mounted but before
(1) If the three wires are the the wiring is connected to the
same size and length, the instrumentation. This test will help
resistance measured between identify gages that may have been Diagnostic Bridge
points A and B, before the damaged during installation. Under Measurements
wires are connected to the laboratory conditions with room-
Additional errors occur when
instrumentation, is 2Rl . temperature curing adhesives, the
voltages are induced in the
mounted resistance value of metal
(2) Measure the voltage from A-B measurement circuit by sources
foil gages will usually fall within the
(which is equivalent to B-C) other than strain. These voltages
package tolerance range for the
and the voltage from B-D. may be in the form of static
gage. Under field conditions, the
Since R3 is typically a precision offsets (such as a thermally
shift in gage resistance will usually
resistor whose value is well induced voltage) or time-
be less than 2%. Greater shifts may
known, the current in the C-D varying disturbances (such as a
indicate damage to the gage. The
leg can be computed using magnetically induced voltage).
farther the gage resistance value
Ohms Law. This is the current Other sources of interference are:
deviates from the nominal value, the
that flows through the lead capacitive coupling of signals to the
larger the unstrained bridge output
resistance, so the value of Rl gage or wiring; resistive leakage
voltage. This limits the strain range
can be computed, since the paths to the gage or from the
at maximum resolution when using
voltage from B-C is known. The wiring to adjacent signal carriers;
the unbalanced bridge technique.
equation for computing Rl is: a leakage path in the excitation
The easiest, most accurate way to
supply; a poor connection to a
measure this resistance is with the
guard; or a damaged shield. Since
VAB four-wire function of a DMM.
Rl =
VBD
R3 error-producing interference can
arise from so many unexpected
Equation No. 23 sources, what can be done to detect
Gage Isolation the presence of unwanted voltages?
These measured values for lead The isolation resistance from the The first step is to disconnect the
resistance should be retained for gage grid to the specimen, if the excitation supply from the bridge
later calculations. specimen is conductive, should also and power up all equipment that is
be measured before connecting the to be operating during the test. This
lead wires to the instrumentation. insures that all possible interference
Diagnostics This check should not be made
with a high-voltage insulation
sources are activated. Next, take
several consecutive bridge output
To insure strain data that is as tester, because of possible damage voltage readings for each strain
error free as possible, various to the gage, but rather with an gage channel. The voltages should
diagnostic checks can be ohmmeter. A value of isolation be very nearly zero. If there is an

E-121
practical Strain Gage measurements

Figure 33:Gage Wiring to Minimize Magnetic Induction

offset voltage, it could be thermally wires and forming minimum but readings could be corrected by
induced or due to a resistive equal loop areas in each side of this offset voltage amount. To
leakage path. A time-varying the bridge. These loops should be monitor the thermally induced
cyclic voltage could be caused by arranged as shown in Figure 33 voltages, the bridge power can be
resistive, magnetic or capacitive to have minimum effect on bridge connected with switches so that
coupling to the interfering source. output. In severe magnetic fields, the voltage readings can be taken
Erratic voltage readings could be magnetic shielding for the wiring with both power supply polarities.
due to an open input to the DVM. may be required. If the measured thermally induced
An integrating voltmeter which voltages are more than a few micro-
The next step is to connect the
samples over a whole number of volts, the source should be found
excitation supply to the bridge. A
power line cycles greatly increases and eliminated rather than trying

STRAIN GAGES
series of readings taken by the
rejection of magnetic induction to correct the voltage readings. If,
DVM of the excitation voltage is a
and other interference sources at after a reasonable time for the gage
good verification that the excitation
power- line frequency. When we and bridge resistors to reach steady
supply is set to the correct voltage
use non-integrating voltmeters, state temperatures, the voltage is
level and is stable enough to allow
several readings can be averaged still drifting, the excitation level may
the accuracy expected. Some
to minimize the effect on static be too high.
thermally induced voltages may be
strain readings.
Thermally induced voltages are
due to heating effects from power
dissipated in the bridge circuit, so
Another test on the gage,
particularly on the gage bond, can
E
caused by thermocouple effects at a check should be made with the be performed at this time. While
the junctions of dissimilar metals power applied. This is done by monitoring the bridge output with
within the measurement circuits taking a sequence of readings of the DVM, press lightly on the
in the presence of temperature the bridge output, then reversing strain gage with a pencil eraser.
gradients. These can occur at the polarity of the excitation supply The output voltage should change
connectors, where the lead wire and repeating the sequence. One- slightly but then return to the
meets the gage metal, in switches half the difference in the absolute original value when the pressure
or in the DVM. Magnetically induced values of the bridge output voltages is removed. If the output voltage
voltages occur when the wiring is is the thermally induced voltage. If doesnt return to the original value
located in a time-varying magnetic the temperature and power levels or becomes erratic, the gage is
field. Magnetic induction can be will remain at this level during the probably imperfectly bonded or is
controlled by using twisted lead test, then subsequent voltage damaged and should be replaced.

E-122
practical Strain Gage measurements

The unstrained bridge output


voltage level also has diagnostic in resistance from the parallel
value. A shorted or open gage will combination of resistors, we can
give an output of approximately compute the equivalent strain value
one-half the excitation voltage. for a given gage factor, i.e.,
In many cases, the unstrained
bridge output should be 2 millivolts
e = (1/GF)(DR/R). By using the
same program subroutines and
or less per excitation volt. For instrumentation which will be used
example, if each of the four bridge in the actual test, we verify most of
arms had a tolerance of 1%, the the system and gain confidence in
unstrained output would at most the test setup.
be 10 millivolts per excitation volt.
So, if the unstrained output is more The value of the shunt resistor is
than a few millivolts per volt of often in the 10-500 k range, so
excitation, the installation should be the current through it is low, less
Figure 34: Typical Installation
inspected. If the test entails some than 1 milliamp. This resistor is
type of temperature cycle and a also outside the bridge arms, so
temperature compensated gage is the effects of switching and lead
utilized, recording the unstrained wires are not as important as for
output over the temperature cycle is the gage. Any of the bridge arms Gage Factor versus
a method of verifying the adequacy for any bridge configuration can be Temperature
of the compensation. shunted and a corresponding value
of equivalent strain computed. The gage manufacturer supplies a
nominal gage factor and tolerance
with each gage. If this gage factor
Shunt Calibration is per NAS 942, Reference [9],
(Verification) Temperature Effects it is the nominal gage factor and
We have examined ways to tolerance as measured at room
When using the unbalanced bridge
compensate for the effects of temperature, in a uniaxial stress
method of strain measurement with
temperature on the lead wires to field, on a material with a Poissons
instrumentation under computer
the gage. Now lets look at some ratio of 0.285, for that particular
control, there are no adjustments for
methods to compensate for the lot of gages. The tolerance on the
bridge balance or span. Since shunt
temperature effects on the gage gage factor directly affects the
calibration was originally used to
resistance and the gage factor. accuracy of the strain computation.
adjust the span of balanced bridge
Some of these methods require the In other words, the computed
instruments, what is the role of
temperature to be measured at the strain value will have a tolerance
shunt calibration with an unbalanced
gage. This can be accomplished at least as great as the gage factor
bridge? Shunt calibration with this
by several different temperature tolerance. A plot showing how
technique might more correctly
sensors such as thermocouples, gage factor varies with temperature
be termed shunt verification,
thermistors and resistance is also furnished with the gage.
since the instrumentation wont
temperature detectors (RTDs). This plot is in the form of % gage
actually be calibrated by shunt
Since we want to sense the factor variation (%DGF) versus
calibration. Shunt verification is the
temperature of the strain gage itself, temperature (T). The temperature
placing in parallel with one of the
problems can arise when large at which these variations become
bridge arm resistors, or gages, of
thermal gradients exist or when the significant depends upon the gage
a resistor of known value. This will
temperature is rapidly changing. We alloy and the accuracy required.
change the bridge output voltage
by a predictable amount and, if we need a sensor that has adequate In practice, the temperature must
measure this output change just as thermal response, and we need be measured at the gage during the
if it were caused by strain, we can to locate it such that it senses the strained measurement and the gage
compute the equivalent strain value. same temperature that exists at factor variation computed or looked
Since we already know the change the gage. up. The actual gage factor is then
E-123
practical Strain Gage measurements

For gage alloys with nonlinear


computed using this variation and characteristics, we need to use
the nominal gage factor. a point-by-point correction or using this plot, since the actual
some type of curve-fitting routine gage value may be different and
to approximate the temperature temperature-dependent. A fourth- or
(1 + % DGF) dependence. In general, gage fifth-order polynomial can be used
GFA = GF factor temperature compensation is to describe the apparent strain
100
required only for large temperature curve and can be obtained from the
Equation No. 24 extremes or for tests requiring the manufacturer or derived from the
utmost accuracy. plot. Thermally induced apparent
strain occurs because perfect
This actual gage factor, GFA, is then temperature compensation over a

STRAIN GAGES
used in the equation for computing
Temperature-Induced broad range cant be achieved. It
strain, e.g., Equation 15, instead
of GF. The value of strain thus
Apparent Strain results from the interaction of the
thermal coefficient of resistance of
computed is compensated for the For temperature compensated the gage and the differential thermal
effect of temperature on the gage strain gages, the manufacturer expansion between the gage and
factor. supplies a plot of temperature- the specimen. Also, the specimen
induced APPARENT STRAIN will seldom be the exact alloy
For most metallic gage alloys
commonly used for static strain
versus temperature. This plot is
obtained by installing a sample
used by the gage manufacturer
in determining the apparent
E
measurement, the gage factor
of gages from the lot on a piece strain curve. Apparent strain is, of
variation with temperature is nearly
of unstrained material having a course, zero for the temperature
linear over a broad temperature
thermal coefficient of expansion at which the gage is mounted. If
range and is less than 1%
matching that for which the that temperature were maintained
for temperature excursions of
compensated gage was intended, for the duration of the test, no
100C/180F. For example, the
and then varying the temperature. correction would be required, but if
equation for gage factor variation
The apparent strain value can thus the temperature varies during the
versus temperature in C for a
be computed and plotted versus course of the test, compensation for
typical temperature-compensated
temperature. The apparent strain the apparent strain may be required
Constantan alloy gage, as taken
curve may have been plotted depending upon the temperature
from the plot enclosed by the
by using a gage factor of +2. changes, the gage alloy and the
manufacturer, was found to be:
This should be considered when accuracy required.
%DGF=0.007T0.1.
E-124
practical Strain Gage measurements

of the apparent strain.


If the temperature changes between the desired accuracy. Any time the
For multichannel systems where
the time of the unstrained and temperature varies during the test,
many gages are mounted in an
strained readings, errors may be the accuracy of the apparent strain
area of uniform temperature, it is
incurred, as can be seen from the compensation can be improved by
more efficient to read the dummy
apparent strain plot. These errors using the actual characteristics of
gage directly. The value of strain
are in the form of a strain offset. the mounted gage. To accomplish
read from the dummy gage will be
If the gage temperature and the this, the mounted but mechanically
the value of the apparent strain.
apparent strain characteristics are unstrained gage must be subjected
The strain readings from the
known, this offset can be calculated to temperature variation, and
active gages that are mounted
and the strain value compensated the apparent strain computed at
on the same material at the same
accordingly. Another way of appropriate values of measured
temperature can then be corrected
achieving compensation is to use an temperature. With computer
by subtracting this amount
unstrained dummy gage mounted controlled instrumentation, the data
from them.
on the same material and subjected can be taken automatically while
to the same temperature as the There are some cases where it is the temperature is varied. If the
active gage. This dummy gage desirable to generate a thermally temperatures of the actual test are
and the active gages that are to be induced apparent strain curve for known, the apparent strain values
compensated should all be from the particular gage mounted on can be recorded at only those
the same manufacturers lot so they the test specimen. Such would be temperatures and used as a look
all have the same apparent strain the case if a compensated gage up table for correction of the test
characteristics. The dummy can be werent available to match the data. The temperature compensated
used in a bridge arm adjacent to thermal coefficient of expansion gage factor of the mounted gage
the active gage, thereby effecting of the specimen material, or if the should be used for computing these
electrical cancellation compensation werent adequate for apparent strain values. If the test
E-125
practical Strain Gage measurements

temperatures at which data will be type of backup for this resistor pair, in as little as 10 microstrain loss in
taken are not known, then it will be since several data channels will be measurement accuracy.
necessary to generate the equation unstrained raw data strain readings
for the apparent strain curve over chan vout Vin Ratio channel microstrain
the temperature range of interest. 0 -0.000589 1.980 -0.000298 0 286
Curve-fitting computer programs are 1 -0.000528 1.980 -0.000267 1 410
2 -0.000065 1.980 -0.000033 2 1165
available to generate an equation 3 -0.000101 1.980 -0.000051 3 417
that approximates the measured 4 -0.000128 1.980 -0.000065 4 291
characteristics. [5,6] 5 -0.000418 1.980 -0.000211 5 776
6 -0.000275 1.980 -0.000139 6 257
7 -0.001345 1.980 -0.000679 7 142
8 -0.000276 1.980 -0.000139 8 351
Data: Input, 9 -0.000244 1.980 -0.000123 9 117

Output, Storage Unstrained Data Should Be Stored Strain Readings from 10 Channels
in Nonvolatile Media
When using unbalanced bridge
techniques with computer control, lost should a resistor fail. This can Use of a computer to control
data storage becomes an important be accomplished by reading the instrumentation, data manipulation,
consideration. Storage of the voltage across each of the two and storage gives us almost
unstrained bridge imbalance voltage resistors and the power supply unlimited data output capability.
ratio is especially critical, since for voltage, and storing these voltages With the wide variety of printers,
some tests it may be impossible to in a nonvolatile medium. Should displays and plotters available,
return to the unstrained condition. the resistor pair fail, they can be the test data can be reduced and
This unstrained data should be replaced with a new pair and a new output by the computer in almost
stored in nonvolatile media such set of voltage readings taken. These any conceivable format, often while
as magnetic tape or disc, with a two sets of readings would then be the test is still in progress. With
redundant copy if the test is critical used to compute an offset voltage computational power and smart
or of long duration. Storage of the to compensate for the difference in instrumentation, we can greatly
subsequent strained readings can the ratio of the two pairs of resistors. increase the speed and accuracy of
be done during or after the test This offset voltage would be added the measurement while eliminating
as required for data reduction or to all strained imbalance voltage the tedious manual-adjustment
archival purposes. Large amounts readings taken with the new pair of process. Now we have more time to

STRAIN GAGES
of data can be stored quickly resistors. This technique can result concentrate on the test results.
and inexpensively with the media
available today, and frequent
storage of data is good insurance
against power interruptions and date and time:
equipment failures.
Previously we discussed using
one power supply for several E
different channels of strain gages
and measuring it with an accurate
DVM. This enables us to use an
inexpensive supply and also allows
its replacement should it fail, with no
loss in measurement accuracy. We
also discussed a circuit that used
a common internal half-bridge for
several channels of strain gages.
For very expensive and/or long term
tests when using this technique,
it may be desirable to have some
Plot of Stress vs. Time Computed from Strain Gage Mounted on a Cantilever Beam
E-126
practical Strain Gage measurements

appendices and average properties of selected engineering materials


Exact values may vary widely
bibliography modulus of elastic
poissons elasticity, E strength (*)
material ratio, n psi x 106 tension (psi)
ABS (unfilled) 0.2-0.4 4500-7500
appendix a: tables Aluminum (2024-T4) 0.32 10.6 48000
Aluminum (7075-T6) 0.32 10.4 72000
wire resistance Red Brass, soft 0.33 15 15000
solid copper wire Iron-Gray Cast 13-14
awg /foot (25C) diameter (in)
Polycarbonate 0. 285 0.3-0.38 8000-9500
18 0.0065 0.040
Steel-1018 0.285 30 32000
20 0.0104 0.032
Steel-4130/4340 0.28-0.29 30 45000
22 0.0165 0.0253
Steel-304 SS 0.25 28 35000
24 0.0262 0.0201
Steel-410 SS 0.27-0.29 29 40000
26 0.0416 0.0159
Titanium alloy 0.34 14 135000
28 0.0662 0.0126
(*) Elastic strength can be represented by proportional limit, yield point,
30 0.105 0.010 or yield strength at 0.2 percent offset.
32 0.167 0.008

appendix B: BRidge Circuits Equations compute strain from unbalanced bridge voltages:
sign is correct for VIN and VOUT as shown Vr = [(VOUT /VIN) strained (VOUT /VIN) unstrained]:
GF = Gage Factor n = Poissons ratio: e = Strain:Multiply by 10 for microstrain:
6

tensile is (+) and compressive is ()


Quarter-Bridge Configurations
Rl Rl

R1 R1
Rg () Rg ()
+ Rl + Rl
+ +
V
IN VOUT OR V
IN
VOUT

Rl Rg (Dummy)
R2 R3 R2
Rl

e =
4Vr
GF(1 + 2Vr ) (
1+
Rl
Rg )

Half-Bridge Configurations (AXIAL) (BENDING)


Rl Rl

R1 R1
Rg ( + ) Rg ( + )
+ Rl + Rl
V + +
V OUT V VOUT
IN IN

Rg ( ) Rg ( )
R2 Rl R2 Rl

e =
4Vr
GF[(1 + n ) -2Vr (n 1)] 1+
Rg
Rl e =
2Vr
GF (1+
Rl
Rg )
Agilent Technologies, Inc. 1999. Reproduced with
Permission, Courtesy of Agilent Technologies, Inc.

E-127
practical Strain Gage measurements

Full-Bridge Configurations (BENDING) (AXIAL)

 +  +  +

+ + +
V + V + V V +
IN VOUT IN VOUT IN OUT

+  +   + 

e =
Vr
e =
2Vr
e =
2Vr
GF GF(n + 1) GF[(n + 1) Vr (n 1)]

appendix C: Equations
BIAXIAL STRESS STATE EQUATIONS

sX sy sX sy E
ex = n ez = n n sy = (ex + n ex)
1-n
2
E E E E

sy sx sx =
E
(ex + n ey )
ey = n sz = 0
1-n
2
E E

ROSETTE EQUATIONS
Rectangular Rosette:

3
2
ep,q = 1
2 [
e e =e e
1 +e e e ] 3 ( 1 3)
2
+ (2 2 1 3)
2

[ e e e e ]
45
sp,q = E e + 1
n =
1
e e 3
( 3)
2
+ (2 3)
2

n 1 2 1

STRAIN GAGES
2 1 1+

Up,q = 1 TAN 2e2 e1 e3


45

1 2 -1
e1 e3
Delta Rosette:

3 2 ep,q = 1
3 [ e
1 e e = [e e
+ e2 + 3 2( 1 2)
2
+( 2 e3 )2 + (e3 e1) 2] ] E
60
sp,q = E
3 [ e
1 e e
n
+
n
= [
e e
1
2 + 3

1+
1
2( 1 2)
2
+ (e2 e3 ) 2 + (e3 e1) 2] ]
Up,q = 1 TAN =3 (e2 e3 )
60
1 2 -1
2e1 e2 e3
Agilent
Technologies, Inc. 1999. Reproduced with
WHERE: Permission, Courtesy of Agilent Technologies, Inc.
e p,q = Principal strains
sp,q = Principal stresses
Up,q = the acute angle from the axis of gage 1 to the nearest principal axis. When positive,
the direction is the same as that of the gage numbering and, when negative, opposite.
NOTE: Corrections may be necessary for transverse sensitivity. Refer to gage manufacturers literature.
E-128
practical Strain Gage measurements

appendices and appendix D: instrumentation accuracy


bibliography Measurement error (in me) due to instrumentation is often difficult to
(continued) determine from published specifications. However, accuracy can be
computed using the following simplified error expressions. For the 14 bridge,
add equations 1-6 (N = 1). For the 12 bridge with two active arms, add
R1 Rg equations 2-6 (N=2). For the full bridge with four active arms, add equations
3-6 (N = 4).
V VOUT
The total error for a measurement must also include gage, lead wire, and,
IN
if applicable, bridge nonlinearity errors. These are discussed in the body of
this application note. Additionally, other equipment imperfections which vary
R2 R3 from instrument to instrument must occasionally be considered (e.g., offsets
caused by leakage currents due to humidity or ionic contamination on PC
boards and connectors).
4 Bridge Circuit
1

(1) R3 change from unstrained to


strained reading (due to
eerror DR3 /R3
temperature, load life, etc.)
GF

(2) R1 change from unstrained to


R2 strained reading (due to
eerror D
R1 R1
temperature, load life, etc.) R2 R2
GFN
Digital voltmeters and A/D converters are specified in terms of a gain error (% of reading) and a offset
error (number of counts, in volts). Since strain calculations require two measurements, a repeatable offset
error, e.g., due to relay thermal EMF, etc., will cancel, but offset due to noise and drift will not. Assuming
that noise and drift dominate, the offset on two readings will be the root sum of squares of the two offsets.
This is incorporated into the formulas.

(3) DVM offset error on bridge measurement

eerror , 4

VINGFN = (Offset error
strained)2
+ (Offset error
unstrained)2

Error terms 4-6 can usually be ignored when using high accuracy DVMs (e.g., 512 digit). These error terms
are essentially the product of small bridge imbalance voltages with small gain or offset terms. For equations
4-6, VOUT, the bridge imbalance voltage, is a measured quantity which varies from channel to channel. To
calculate worst case performance, the equations use resistor tolerances and measured strain, eliminating
the need for an exact knowledge of VOUT.

(4) DVM gain error on bridge measurement

eerror 4
[(VOUT)(Gain Error) strained reading (VOUT) (Gain Error) unstrained reading]
GFVINN
,
e
measured (Gain Error)
strained reading
S tolerances on R1/R2,R3,Rg Gain Error change
( )
GF N
strained-unstrained

The bridge excitation supply can be monitored with a DVM or preset using a DVM and allowed to drift. In
the first case, supply related errors are due only to DVM gain and offset terms, assuming a quiet supply. In
the second case, since power supply accuracy is usually specified in terms of a gain and a offset from
the initial setting, identical equations can be used. Also for the second case, note that the strained reading
gain error is the sum of the DVM and excitation supply gain errors, while the strained reading offset error is
the root sum of squares of the DVM and excitation supply offset errors.
E-129
practical Strain Gage measurements

(5) Offset error on supply measurement (or on supply drift)

eerror 4 [(VOUT)(Offset Error) strained reading (VOUT) (Offset Error) unstrained reading]
GF V IN
2
N
,
e
measured (Offset Error)strained reading + S tolerances on R1/R2,R3,Rg = (Offset error
strained)2
+ (Offset error
unstrained)2
V IN GF N
V IN
(6) Gain error on supply measurement (or on supply drift)

eerror 4
[(VOUT)(Gain Error) strained reading (VOUT) (Gain Error) unstrained reading]
GFVINN
,e
measured (Gain Error) strained reading + S tolerances on R1/R2,R3,Rg (Gain Error change )
GF N
strained-unstrained

Equation 4 bridge 12 bridge full bridge


1 even when using state-of-the-art
bridge completion resistors and
(1) R3 15.5
EXAMPLE (2) R1/R2 31.0 15.5
measuring equipment.
Although typical measurements will
Evaluate the error for a 24-hour (3) VOUT offset 2.3 1.1 0.6 have a smaller error, the numbers
strain measurement with a 5C/9F (4) VOUT gain 0.4 0.4 0.3
computed reflect the guaranteed
instrumentation temperature variation. instrumentation performance.
This includes the DVM and the bridge
(5) VIN offset 2 0.2 0.2 Measuring the excitation supply for
completion resistors, but not the gages. (6) VIN gain 1.3 1.1 1.0 both the unstrained and strained
The hermetically sealed resistors have Sum 50.08 me 18.3 me 2.1 me readings not only results in smaller
errors, but allows the use of an
a maximum TCR of 3.1 ppm/C, and inexpensive supply.
have a 0.1% tolerance. The DVM/
Bridge completion resistor drift
Scanner combination, over this time CONCLUSIONS limits quarter- and half-bridge
and temperature span, has a 0.004% performance. Changes due to
Based upon this example, several
gain error and a 4 mV offset error on important conclusions can be drawn: temperature, moisture absorption
the 0.1 volt range where the bridge and load life require the use of ultra-
output voltage, VOUT, will be measured. Surprisingly large errors can result stable hermetically sealed resistors.
The excitation supply is to be set at 5 V

bibliography and credits


using the DVM. The DVM has a 0.002%
gain error and a 100 mV offset error on

STRAIN GAGES
the 10 volt range. Over the given time
and temperature span, the supply has a Application Note 290-1Practical Strain Gage Measurements
0.015% gain error and a 150 mV offset
1. Higdon, Ohlsen, Stiles, Weese and Riley: MECHANICS OF MATERIALS, 3rd Edition,
error and will not be remeasured. The John Wiley & Sons, Inc., New York, 1976.
mounted gage resistance tolerance is
2. C.C. Perry and H.R. Lissner: THE STRAIN GAGE PRIMER. McGraw-Hill, Inc.,
assumed to be 0.5% or better. The strain New York, 1962.
to be measured is 3000 me and the gage
3. I.S. Sokolnikoff: TENSOR ANALYSIS (Theory and Applications to Geometry and
factor is assumed to be +2.
E
Mechanics of Continua), Second Edition, John Wiley & Sons, New York, 1964.
Notice that the temperature, as given, 4. Hornsey, McFarland, Muhlbauer, and Smith: MECHANICS OF MATERIALS
can change by as much as 10C/18F (An Individualized Approach), Houghton Mifflin Company, Boston, 1977.
between the unstrained and strained 5. MANUAL ON EXPERIMENTAL STRESS ANALYSIS, Third Edition, Society for
measurements. This is the temperature Experimental Stress Analysis, Westport, Ct, 1978.
change that must be used to evaluate the 6. James W. Daily and William F. Riley: EXPERIMENTAL STRESS ANALYSIS,
resistor changes due to TCR. The R1/R2 Second Edition, McGraw-Hill, Inc., New York, 1978.
ratio has the tolerance and TCR of two 7. William M. Murray and Peter K. Stein: STRAIN GAGE TECHNIQUES, 1958.
resistors included in its specification, so the 8. Peter K. Stein: ADVANCED STRAIN GAGE TECHNIQUES, Stein Engineering Services,
ratio tolerance is 0.2% and the ratio TCR Inc., Phoenix, AZ, 1962.
is 6.2 ppm/C. The gain error change 9. National Aerospace Standard 942, Revision 2. 1 July 1964: STRAIN GAGES, BONDED
on the bridge output measurement and RESISTANCE, Aerospace Industries Association of America, Inc., Washington DC
on the excitation measurement can be as 10. Hewlett-Packard Co., Application Note 123: FLOATING MEASUREMENTS
much as twice the gain error specification. AND GUARDING, Hewlett-Packard Co., Palo Alto, CA, 1970.
The following table shows the total error
and the contribution of individual error Agilent Technologies 1999.
equations 1-6. Reproduced with Permission, Courtesy of Agilent Technologies Inc.
E-130

Das könnte Ihnen auch gefallen