Beruflich Dokumente
Kultur Dokumente
2.0 Responsibility
All Indoor Air Quality (IAQ) technicians and persons handling this equipment are
responsible for following this procedure.
2.0 Equipment
MultiRae Pro PID Multigas meter
Unplug MultiRAE from charger, hold MODE Key to turn on; Alarm will beep once,
watch display screen for messages such as:
Check the status of the meter. Battery life at three segments (max), date and time,
appropriate sensors calibrated (check last calibration date). No alarms and sensors
reading zero indicate monitor ready for use. If alarms (pump or lamp) appear,
troubleshoot (see operational manual). Verify external filters fitness for use (see
QA/QC).
1. Install and calibrate the relevant sensors as specified by the above instructions
and job-specific Measurement Survey
2. While in survey mode, enter the menu, and then select Monitor, which will
display settings for date and time. Make the necessary adjustments to these
fields as needed (Refer to manual).
3. Generate calibration certificate, print, sign and file in the appropriate
instrument folder.
4. It is the responsibility of whomever is removing the MultiRae from the office to
update the MultiRae folder with the relevant log out information prior to
leaving the office.
5.1 Calibration
2
2. Calibrate monitor? is the first item on the menu. Press Y/+ to perform
calibration.
3. Select Fresh Air Calibration from the Calibration sub menu
4. Attach the regulator to the zero air gas cylinder (ensure that the cylinder is
dated in use and not expired), turn the regulator to allow gas flow, and
connect the tubing to the monitor's intake valve.
5. Press the Y/+ key to start fresh air calibration of the monitor. The display
shows zero . In progress followed by the name of each sensor and the
message zeroed. The display should show a reading of 20.9 for the oxygen
sensor and 0.0, (or a very small value), for all other sensors.
6. After a five second pause the display will show the message Zero Cal Done!
Turn off gas and disconnect the regulator from the zero gas cylinder.
4
5.4 Downloading of Data
1. Connect the data cable to the MultiRae and the computer. Press Mode on the
MultiRae until "Enter Communication Mode" is found. Press Y/+ when
prompted until the MultiRae enters communication mode.
2. Open the ProRAE Studio II Program, and select Administrator.
3. Enter the password rae and press Enter
4. Click Operation followed by Auto detect, and select the device from the pop
up list. Subsequently, click Receive Data
5. A new window is opened with the newly received logged data. To save the
data, select File Save As and save to a specified job folder. Always save in
appropriate folder and label "Save As" file with the Serial Number of the
meter being downloaded.
6.0 QA/QC
5
zeroed in an area containing VOCs. Perform Fresh Air Calibration in an area
clear of all VOCs.
o Pump display along with 3 beep audible alarm indicates that pump has
stopped due to line clog. This is a latching alarm; Clear line/filter and press
Y/+ key to clear alarm and restart pump.
o A flashing Bat display along with a 1 beep alarm every 10 seconds
indicates that the battery voltage has dropped below 4.35 V and it will shut
down in 20-30 minutes. Quickly complete testing and charge MultiRAE or
install alkaline battery.
o A flashing MEM display along with a 1 beat/second audible/tactile/visual
alarm indicates that the datalog memory is full. Clear datalog in Change
Datalog? If you get this alarm you might want to consider changing. Select
memory full type? to Wrap around.
User safety
The MultiRAE will provide audible and visual alarms to alert users to unsafe
conditions (according to preset limits).
It is extremely important to note that during the following conditions the alarm
signals are disabled:
When entering the Communicate with PC? standby mode.
When entering the Calibrate Monitor menu
Alarms -
o High display along with 3 beat/second audible/visible alarm and flashing
display backlight indicates that VOC sensor (PID) has exceeded the high
alarm set point.
o Low display along with 2 beat/second audible/visual alarm and flashing
display backlight indicates that Oxygen sensor has gone into low alarm (less
than 19.5% Oxygen default). Press Y/+ key to clear if latching alarm.