Beruflich Dokumente
Kultur Dokumente
Aspirators
An aspirator is often known as a suction pump or a vacuum pump. It uses suction to remove gas, fluid, tissue,
or other materials from a body cavity. An aspirator typically consists of a suction pump, a collection con-
tainer, tubing, a pressure gauge, a safety valve (to prevent suctioned material from overflowing the container)
and a means for adjusting the vacuum pressure. The motorized suction pump creates a vacuum in the suction
tubing. When the tubing is inserted into a body cavity, material is sucked through the tubing and deposited
into the collection container.
Recommended functional test frequency: Annual
Equipment information
Control number:_________________________________ Hospital:_________________________________________
Manufacturer:___________________________________ Model:___________________________________________
Serial number:__________________________________ Location:_________________________________________
Test information
Technician:_____________________________________ Date:____________________________________________
Test type: Incoming_________ Post repair_________
Test equipment needed: ESA615 Electrical Safety Analyzer (or equivalent)
DPM4 Pressure Meter or equivalent
Stopwatch or watch with a second hand
Tubing and connectors to connect to DPM4
Test Result
Pass Fail N/A
Physical condition
Device is clean and decontaminated
No physical damage to case, display, mounts, cart, or components
Switches and controls operable and correctly aligned
Display intensity adequate for daytime use
Control numbers, labeling, and warnings present and legible
Inlets and hoses
Power cord, accessory cables, charger
Filters and vents clean
Electrical safety Criteria IEC 62353
Ground wire resistance < 0.3
< 100 A NC
Chassis leakage
< 500 A SFC
< 100 A B and BF
Patient leakage current
< 10 A CF
< 100 A BF
Patient lead leakage current: isolation test (mains on applied part)
< 10 A CF
Insulation test (optional) 500 V < 2 M
Physical condition
Check the physical condition of the device as described in the General Equipment Procedure.
Electrical safety
Perform electrical safety checks as described in Chapter 6: Electrical Safety. Check ground wire resistance
and chassis leakage.
Preventive maintenance
Replace battery
The battery should be replaced every 24 months. Replace if necessary.
Replace filters
Inspect filters and replace as necessary. Refer to the device service manual for filter replacement.
Lubricate motor
Follow the manufacturers instructions in the service manual for lubricating the pump motor. Not all motors
will need to be lubricated.
Complete model-specific preventive maintenance
Refer to the monitors service manual for preventive maintenance tasks specific to
the device. Complete the preventive maintenance per manufacturers procedure.
Performance inspection
Verify unit operates on battery
Confirm the ac power indicator is lit when the power cord is plugged into an
outlet. Unplug the ac power cord and perform the remainder of the functional test
on battery power. The ac power indicator should go out when the power cord is
unplugged and the battery indicator should light. Be sure to
Figure 7-1. Fluke Biomedical
plug the power cord in at the conclusion of the test. DPM4 Pressure Meter
Figure 7-2.
Maximum vacuum
Disconnect the suction tubing from the DPM4 and occlude the suction tubing. The tubing can be occluded
with a stopcock in the off position or by simply covering the tubing with your thumb. Adjust the suction to
its maximum setting. The accuracy of the vacuum gauge must be verified prior to this test. The values on the
test form are guidelines based on common practice. Refer to the aspirators operator manual for its actual
performance capability. If the measured vacuum is low, look for air leaks, particularly in collection bottle
caps and hoses. Release the occlusion on the tubing.
Vacuum rise time
With the aspirator set to its maximum setting, occlude the suction tubing. Use a stopwatch or a watch with a
second hand to measure the time it takes to reach the vacuum level indicated on the test form. Thoracic aspira-
tors should reach 30 mmHg in less than four seconds. Emergency, surgical, or tracheal aspirators should reach
300 mmHg in less than four seconds. Uterine aspirators should reach 300 mmHg in less than three seconds.
Breast pumps should reach 150 mmHg in less than two seconds. Again, these are general guidelines based on
common practice. Refer to the aspirators operator manual for its actual performance capability.
Complete model-specific performance testing
Refer to the service manual for performance inspection tasks specific to the device. Complete the perfor-
mance inspection per manufacturers procedure.
Return to service
Before returning to use, return the suction setting to its original setting. Plug in the power cord to ensure the
battery remains charged.