Sie sind auf Seite 1von 10

Blackwell Publishing IncMalden, USAAORArtificial Organs0160-564X 2006, Copyright the Authors; Journal compilation 2006, International Center for

r Artificial Organs and Transplantation? 2006307529538Original ArticleNUMERICAL MODEL OF FLOW IN A SAC-TYPE VADI.
AVRAHAMI ET AL.

Artificial Organs
30(7):529538, Blackwell Publishing, Inc.
2006, Copyright the Authors
Journal compilation 2006, International Center for Artificial Organs and Transplantation

Numerical Model of Flow in a Sac-Type Ventricular


Assist Device

*Idit Avrahami, Moshe Rosenfeld, Sagi Raz, and Shmuel Einav


*Option of Bioengineering, California Institute of Technology, Pasadena, California, U.S.A.; Department of Fluid Mechanics
and Heat Transfer and Department of Biomedical Engineering, Faculty of Engineering, Tel Aviv University, Tel Aviv, Israel

Abstract: This article addresses the growing need for com- coupled fluidstructure interaction numerical simulation
prehensive tools to investigate the hemodynamics of ven- and to experimental measurements using continuous digi-
tricular assist devices (VADs) in general and sac-type tal particle image velocimetry. Our results demonstrate
VADs in particular. Numerical simulations can be very that the flexible wall motion is sensitive to changes in pres-
helpful in these efforts. However, full simulation of flow sure distribution inside the chamber. However, small vari-
inside sac-type VADs poses several key problems, among ations in wall motion do not significantly affect the global
them simulation of the mechanical heart valves and calcu- features of flow inside the chamber. Therefore, the simpli-
lation of the motion of flexible walls. We present a sim- fied model can be used to predict the 3D time-dependent
plified three-dimensional (3D) numerical model of a flow field in the VAD. Key Words: Ventricular assist
sac-VAD chamber. The walls in the simplified model are devicePulsatile ventricular assist deviceSac-type ven-
defined to move according to experimental measurements, tricular assist deviceComputational fluid dynamics
and the valves are modeled in fully open or fully closed Fluidstructure interactionMoving boundariesContin-
positions. The model is validated by comparison to a fully uous digital particle image velocimetry.

Recently proven benefits of ventricular assist form pressure distribution in the blood chamber (5).
devices (VADs) for use as destination therapy (1) are However, numerical simulations of the flow inside
spurring the need to reduce complications by improv- sac-type VADs are highly complicated because the
ing their hemodynamics. Experimental investigation flow is unsteady at transitional Reynolds numbers,
of VAD hemodynamics relies primarily on flow visu- and it occurs within a complex 3D geometry. More-
alization techniques to identify flow patterns and over, such simulations pose two challenges: simu-
locate sites at risk of generating hemolysis or throm- lation of the flow across the passively moving
bus formation. Digital particle image velocimetry mechanical heart valves (MHVs), and simulation of
(DPIV) is a nonintrusive visualization technique that the flow inside a flexible chamber whose volume
can also provide quantitative data such as flow veloc- changes with transmural pressure drop.
ity or stress magnitudes (e.g., 2,3). Flow visualization Calculations of flow across MHVs require very
techniques, however, are limited when measuring demanding simulations. The geometry and orienta-
high velocities or three-dimensional (3D) flows inside tion of the MHVs greatly influence VAD hemody-
the VAD chamber (3,4). Numerical simulations may namics, as the flow across MHVs is characterized by
complement the experimental techniques by produc- high disturbances, high shear stresses, and bulk tur-
ing detailed 3D flow field data that can be compared bulence stresses (6,7). Proper simulation of the 3D
to experimental results for validation. flow across even a single valve requires a very fine
Pulsatile VADs with flexible chambers have some mesh model, the use of strongly coupled fluidstruc-
advantages over VADs with rigid chambers. These ture interaction (FSI) algorithms for the large motion
include smooth contacting walls and a relatively uni- of the leaflets (8), and implementation of low-
Reynolds turbulence models (7,9). To date, no simu-
Received August 2005; revised January 2006. lations were reported that incorporate all aspects of
Address correspondence and reprint requests to Dr. Idit Avra- flow and leaflet dynamics for a single MHV, not to
hami, Department of Bioengineering, California Institute of
Technology, 1200 East California Blvd., M/C 301-46, Pasadena, mention a detailed simulation of a flexible chamber
CA 91125, U.S.A. E-mail: idit@caltech.edu with two passively moving MHVs.

529
530 I. AVRAHAMI ET AL.

As for analysis of the flow in a flexible chamber, lected for this study. A more comprehensive study of
attempts were made to numerically simulate the flow the flow field inside the sac-type VAD will be given
inside the volume-changing ventricle of the natural elsewhere.
heart (e.g., 1012). In these simulations, the wall
motion is specified by myocardial activation. The METHODS
simulation of sac-type VADs, however, requires
coupling of the flow field with the motion of Berlin VAD and the experimental system
elastic chamber walls. The model of the VAD chamber was constructed
Combining these two challenges in the simulation based on the electrohydraulic Berlin sac-type VAD
of a sac-VAD chamber is beyond the routine capa- (4). It includes a discoid flexible chamber (8 cm in
bilities of existing computational resources. Conse- diameter and 2 cm in width) and two tilting disk
quently, only a few numerical analyses, using only MHVs (2 cm in diameter), placed in S-shape center-
simple models, were published for pulsating VADs. line ducts. The inlet and outlet conduits are parallel
The only numerical analysis of VAD that included to each other. The flexible walls of the sac are made
valve motion, by Kiris et al. (13), simulated the 3D of 0.09-cm transparent polyurethane. The mechanical
pulsatile flow in a rigid pusher-plate chamber during properties of the polyurethane walls were measured
one cardiac cycle. However, the motion of the rigid inside a water tank under strain rate as during pump-
pusher-plate and the two valves was specified accord- ing. The measured modulus of elasticity is 15 MPa,
ing to experimental data rather than coupled with the Poissons ratio is 0.3, and material density is 1.05 g/
flow. Chesler and Kamms simulation of a counter- cm3.
pulsating VAD (14) included a pusher plate and a The model of the VAD chamber was connected to
single valveless conduit. To allow the numerical sim- a mock-up flow loop that simulated cardiovascular
ulation, the Reynolds number was reduced from flow conditions (Fig. 1). The loop contained a com-
1600 to 36, and the Womersley number used was pliant chamber to model the arterial compliance,
reduced from 29 to 4.3. Moreover, the chamber was flow regulator representing afterload conditions,
divided into only 396 elements and only one pumping and reservoir representing preload conditions. The
cycle was analyzed to save computational resources. preload conditions (reservoir height) were set to a
Konig et al. (15,16) used a simple 3D model to sim- pressure of 35 mm Hg. The afterload conditions (reg-
ulate a sac-type VAD. The flexible walls were simu- ulator, compliance, and outlet pressure) were set to
lated using a rigid moving hemispherical plate and vary between 80 and 100 mm Hg. The VAD chamber
the valves were modeled indirectly as on/off valves. was enclosed inside a transparent rigid pressure con-
Shim et al. (17) were the only ones to use a fully tainer filled with transmitter fluid. The pressure in the
coupled method to compute the flow interacting with container varied by an attached piston pump and
a flexible sac-VAD chamber, but in order to simplify allowed the flexible chamber to expand and com-
their model, they also used on/off boundary condi- press at a rate of 30 bpm, circulating the fluid through
tions to simulate the valves. Their model ignored not
only the opening and closing phases of the valves,
but also the effect of the valves on profile and direc-
tion of the entering flow.
The objective of this study was to develop and
validate a numerical model for simulating the flow
inside a sac-type pulsating VAD. In the presented
model, the motion of the flexible wall is specified
according to experiments, and the valve leaflets are
modeled in their fully open or fully closed positions.
This article addresses these simplifications and their
effects on the models accuracy in predicting the flow
field in the chamber. In order to examine the conse-
quences of the model simplifications and assumptions,
the simulation results are compared with those
obtained by an FSI numerical analysis in which the
wall motion is fully coupled with the flow and with
an experimental analysis using continuous digital par-
ticle image velocimetry (CDPIV) measurements col- FIG. 1. Illustration of the experimental flow loop.

Artif Organs, Vol. 30, No. 7, 2006


NUMERICAL MODEL OF FLOW IN A SAC-TYPE VAD 531

the flow loop at flow rate of Q = 2 L/min. These chamber, inlet/outlet conduits, and models of the
under-physiological flow conditions were chosen to valves leaflets. In addition, the model of the FSI case
overcome the limitations of the experimental system included elastic walls.
in measuring high velocities in the chamber. Blood Even though the model and boundary condi-
was simulated by a waterglycerin mixture (60:40% tions are symmetric, nonsymmetric flow may occur
by weight) seeded with 10-m silver-coated glass as a result of flow instabilities in transient simula-
microspheres (dotted fill in Fig. 1). This mixture tions. Therefore, to avoid nonphysical effects of
is a transparent blood analog approximating both the symmetric boundary, complete 3D numerical
blood viscosity ( = 0.035 g/cm/s) and blood density models were used to incorporate the entire 3D
( = 1.1 g/cm3). The same solution without particles flow field.
was used as a transmitter fluid in the pressure con- A time-dependent, incompressible, and laminar
tainer (solid fill in Fig. 1). An electromagnetic flow- flow was simulated. The blood was assumed Newto-
meter (MAGFLO MAG 3000, EMCO Flow Systems, nian with a constant viscosity. Flow and pressure
Longmont, CO, U.S.A.), located downstream from fields were calculated numerically by solving the
the outflow valve, measured the instantaneous equations governing continuity and momentum in
and average flow rates, and a pressure transducer the fluid domain:
(PX143-05, Omega Engineering, INC., Stamford,
U=0 (1)
CT, U.S.A.) measured instantaneous pressure in the
pressure container. U
r + (U - U g ) U + p = u 2 U (2)
The experimental study used CDPIV. CDPIV, a t
continuous version of DPIV, is based on a continu- where U is the fluid velocity, Ug is the local coordi-
ous video recording, which is then divided into a nate velocity, p is the static pressure, t is the time,
sequence of images at the camera frame rate. is the blood density, and is the dynamic viscosity of
Image pairs are cross-correlated to create a blood.
sequence of instantaneous velocity maps. A In the FSI case, the maximal deformation of the
detailed description of the method and system can wall was relatively small (up to 6%); therefore the
be found in Grad and Einav (18). In the CDPIV stressstrain relationship was assumed to be linear:
system used for this study, a continuous laser light
sheet (5 W, water-cooled argon ion laser; Coherent = E( + 0) + 0 (3)
Innova 70, Coherent Inc., Santa Clara, CA, U.S.A.) where E is the wall Youngs elasticity modulus, and
was strobed using an electronic light shutter, trig- are the instantaneous strain and stress, respectively,
gered by a CCD video camera (29.97 fps, 8-Bit, and 0 and 0 are the initial strain and stress, respec-
grayscale, 768 480; Pulnix TM-9701, JAI PULNiX, tively. The dynamics of the flexible wall were calcu-
Inc., San Jose, CA, U.S.A.) to sample 15 image lated using the linear dynamics response of a system:
pairs per second. Time interval between images in
each image pair was set to t = 2.5 ms. A combina- + CS + KS = R (4)
MS
tion of mirrors and cylindrical lenses spread the
laser beam to a light sheet with a thickness of 1 mm where M, C, and K are the mass, damping, and stiff-
and an area that covered the entire mid-cross- ness matrices, respectively; R is the vector of exter-
section (XY plane), including the inflow and nally applied loads; and S, S, and S are the
outflow MHVs. The total image acquired by the displacement, velocity, and acceleration vectors of
camera covered an interrogation window of the structural domain, respectively. The set of equa-
76 47.5 mm. The subsamples were 16 16 pixels, tions was solved numerically using CFD-ACE (CFD
with a step size of 8 8 pixels. Research Corporation, Huntsville, AL, U.S.A.). The
fluid solver used the finite volume method to dis-
The numerical model cretize the governing equations and the SIMPLEC
Two numerical models of the Berlin VAD were pressure-based method to derive a stable pressure
simulated. The first model, FSI case, was a fully cou- velocity decoupling formulation. In both simula-
pled FSI model in which the flexible walls were tions, the grid deformation module of CFD-ACE
moving according to the time-dependent transmural was used to control the moving/deforming grid,
pressure drop between the calculated internal pres- bases on standard transfinite interpolation (TFI)
sure field and imposed external pressure. The motion scheme to determine the interior nodes movements
of walls was predefined from the experiments. Both based on the motion of boundary nodes. The arbi-
models included the internal cavity of the VAD trary Lagrangian-Eulerian approach was used to

Artif Organs, Vol. 30, No. 7, 2006


532 I. AVRAHAMI ET AL.

adjust the momentum equation to the grid motion in 140

FSI case
the fluid domain. 120
Base case

For the FSI case, the structural solver used the External pressure C
finite element method. Small deflections were 100

pressure (mm Hg)


assumed for the structure and first-order elements B
80
were used. The structural solver was combined with
the fluid solver using a strong implicit two-way cou- 60

pling mode, in which the structural solver is called


after each iteration. 40
A D

20
Boundary and initial conditions 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0
t n =t/T
For consistency with the experiments, similar pre-
load, afterload, and pump frequency were specified FIG. 3. Time-history of imposed external pressure (dark solid
for the simulations. The filling preload was modeled line) and of the pressure at the center of the chamber for the FSI
for both simulations by specifying a fixed pressure of (light solid line) and base case (dotted line) simulations. The
letters A, B, C, and D define four instants during the pumping
30 mm Hg in the inflow boundary. Ejection afterload cycle (peak filling, beginning of ejection, peak ejection, and
conditions were modeled by combining a linear flow beginning of filling, respectively).
resistor set at 10.5 mm Hgmin/L downstream from
the outflow valve and a fixed pressure of 75 mm Hg ter. Time-varying wall displacement was determined
at the outlet boundary (Fig. 2). To avoid the influence from the experiments. Side view of the VAD was
of boundary conditions on the flow inside the cham- recorded for 140 time samples during more than
ber, the inlet/outlet conduits were longer in the two pump cycles. Marked points on the central
numerical models than in the real VAD. sagittal cross-section were detected, and the distance
The entire VAD-sac chamber was elastic. How- between the central points of the walls was mea-
ever, the actual displacements of circumferential sured. This measured displacement was approxi-
regions were significantly smaller than in the central mated to a harmonic function. The imposed velocity
regions. Therefore, to simplify the numerical models, Wwall of each node on the front and back walls for the
it was assumed that only the central front and back base case was therefore:
regions are mobile, and the other parts of the wall
2p r2
are fixed (Fig. 2). Wwall = A sin (t - t 0 ) 1 - 2 (5)
T R
For the FSI case (the fully coupled simulation), the
time-dependent hydrostatic pressure acting on the where A = 2.39 cm/s, T is the pump cycle period, t0 is
outlet surface of the VAD walls was specified. Pres- the time shift between the beginning of the simula-
sure function was approximated from experimental tion and start of filling, r is the distance of the node
measurements to the curve shown by the solid line from the center of the membrane, and R = 3.25 cm
in Fig. 3. In the base case, the imposed front and back is the radius of the circular moving membranes
wall displacements were approximated to paraboloi- (as defined in Fig. 2). The resulting stroke volume
dal shape with the maximal displacement at the cen- was SV = 56 mL, with a mean flow rate of Q = 1.68 L/
min and a peak ejection rate of Qpe = 4.76 L/min.

Modeling of the valves


This study focuses on the hemodynamics of the
VAD chamber. To make the simulation feasible, the
effect of the MHVs was only approximated. The leaf-
lets were modeled in their fully open or fully closed
positions to create more realistic inflow/outflow con-
ditions on the VAD chamber. The dynamics of the
leaflets during opening and closing were ignored, as
they were shown to have only a secondary effect on
the flow in the chamber (19).
The mesh was built with both leaflets in their fully
open position. To model a closed-valve condition, the
FIG. 2. Boundary conditions during ejection phase (left) and fill- corresponding inlet/outlet boundary was declared as
ing phase (right). impenetrable, and the fluid in the corresponding con-

Artif Organs, Vol. 30, No. 7, 2006


NUMERICAL MODEL OF FLOW IN A SAC-TYPE VAD 533

duit distal to the closed-leaflet boundary line was 6.0


128 000
assigned a very high dynamic viscosity ( = 50 kg/m- 5.0
56 700
4.0
s). Consequently, the fluid in the conduit was blocked 32 100
3.0 Y component

Velocity (cm/s)
distal to the closed-leaflet boundary line (Fig. 2). The 16 500
2.0 10 000
flow was blocked downstream from the outflow valve
1.0
during the filling phase, and upstream from the 0.0
inflow valve during the ejection phase. Timing for 1.0
closing or opening of the valves was determined from 2.0 X component
the calculated moment on the valves. During the two 3.0
isovolumetric phases, both valves were designated as 4.0
closed, but minor flow (104 cm/s) was allowed to 1.0 0.8 0.6 0.4 0.2 0.0 0.2 0.4 0.6 0.8 1.0
Xn = x/R
avoid numerical instabilities.
The simulations started with an initial condition of FIG. 5. Velocity distribution along the X centerline for the different
zero velocity and a uniform pressure of 30 mm Hg mesh resolutions. The X centerline lies on the mid-plane of the
chamber, parallel to the x-axis.
inside the VAD chamber. The chambers walls were
in their neutral position and all simulations started at
peak filling. As the periodic flow characteristics of the evaluate the optimal spatial mesh resolution, five
VAD were investigated, at least two complete pump models with an increasing number of mesh nodes
cycles with 1000 time steps each were calculated for were built (10 000, 16 500, 32 100, 56 700, and 128 000
each model, and the results were taken from the nodes) for the base case and simulated over 20 time
second pump cycle. To assure a good convergence of steps (0.02 T), starting from the same initial flow
the solution, a minimum of three orders of mag- conditions of the beginning of the second ejection, at
nitude reduction in residuals was defined for all a nondimensional time of tn = t/T = 1.25 (marked as
the cases. A maximum of 300 iterations was set for B in Fig. 3).
each time step in the base case, and a maximum of Simulations for all the meshes shared similar flow
500 iterations per time step was set for the FSI patterns. An example of the Cartesian velocity com-
simulation. ponents (Ux and Uy) along X centerline (a line par-
The FSI case simulations required 0.166 central allel to the x-axis, passing through the center of the
processing unit (CPU) seconds per time step per chamber), are shown in Fig. 5 for the five meshes.
mesh node, while the equivalent CPU time required Negligible differences were found between the
for the base case is only 0.016 s. These data are for meshes. Average relative differences between veloc-
Pentium PC with 1000 MHz CPU and 512 MB ran- ities of the finest mesh (128 000 nodes) and of the
dom access memory. coarser meshes with 10 000, 16 500, 32 100, and
56 700 nodes were at 3.5, 3, 1.5, and 1%, respectively.
Mesh and time step independence tests Consequently, a mesh of 56 700 nodes was used for
The mesh models were built using structured mesh the base case simulation. To reduce the calculation
of hexahedral elements with clustering near the wall effort in the FSI analysis, a coarser mesh of 16 250
and near the valves (see a sample mesh in Fig. 4). To nodes was used.
Time step independence tests were also per-
formed. Five simulations using an identical mesh (of
56 700 nodes) were performed using 50, 100, 200,
400, and 1000 steps per cycle (of T = 2 s). All the
simulations started from the same initial conditions
at the peak of the second ejection (at tn = 1.5,
marked as C in Fig. 3), and were compared at the
end of the second ejection (tn = 1.75, marked as D
in Fig. 3).
A comparison of Ux and Uy at the X centerline is
shown in Fig. 6 for the five step intervals. In this case
as well, the simulations with different time steps
showed similar global features. However, the differ-
ences between the cases were more noticeable. The
FIG. 4. (a) Sample mesh of the VAD chamber and (b) magnified average relative difference between the simulation
view of the outlet valve. with the smallest time steps (1000 steps/cycle) and

Artif Organs, Vol. 30, No. 7, 2006


534 I. AVRAHAMI ET AL.

0.5
0.0
Y component
0.5
1.0
Velocity (cm/s)

1.5
2.0 1000
2.5 400
200
3.0
100
3.5 50
4.0 X component
4.5
FIG. 7. Wall displacement at three time instances of the FSI
1.0 0.8 0.6 0.4 0.2 0.0 0.2 0.4 0.6 0.8 1.0 simulation. Darker colors indicate larger wall displacement.
Xn = x/R

quite flexible, with E = 15 MPa and thickness of


FIG. 6. Velocity distribution along the X centerline for the different
time step intervals. 0.09 cm, relatively small variations in the pressure
inside the chamber caused noticeable deviations
from the theoretical paraboloidal shape, as shown in
the simulations with 50, 100, 200, and 400 time steps Fig. 7 for the three instances tn = 1.2, 1.7, and 1.8. The
per cycle were 18.5, 8, 2, and 0.5%, respectively. deviations are particularly significant at the end of
Therefore, all simulations were performed using 1000 ejection phase and at the beginning of filling phase,
time steps per cycle. as a result of the large pressure variations developed
For comparison, Shim et al. (17) used a mesh of inside the chamber in the vicinity of high-velocity
8677 nodes for their analysis of sac-type chamber. regions.
Konig et al. (15) used a mesh with 58 560 elements Side views with the shapes of the wall and velocity
for a similar case of a sac-type VAD with only one vector at the central sagittal cross-section are shown
moving wall, and reported only a small improvement in Fig. 8 for the ejection and filling phases in the
in results when they examined a simulation with 3200 experimental investigation and both simulation mod-
time steps per cycle over coarser time steps of 32, 64, els. In the left column, the experimental results are
and 100 per cycle. Kiris et al. (13) used a total of shown; the wall shape as obtained by connecting the
239 713 nodes in their Chimera grid model of the detected points on the sagittal cross-section is shown
Penn State pusher-plate VAD, and used only 40 time by dotted lines. Some asymmetry can be seen, espe-
steps per cycle for a relatively low Reynolds number cially in the filling phase. This asymmetry may be
(Re = 900).

RESULTS

Pressure distribution and wall motion


The pressure inside the sac chamber was found
to be almost spatially uniform for both the FSI and
base case simulations with a variation of less than
3.5 mm Hg. History of the pressure at the center
of the VAD, together with the external pressure
imposed in the FSI case, is shown in Fig. 3 for both
the FSI and base case simulations. The letters A, B,
C, and D define four specific phases in the cycle
peak filling, beginning of ejection, peak ejection, and
beginning of filling, respectively. A time-periodic
pressure solution is obtained after the second cycle.
The internal pressure in the FSI case follows the
external pressure with a small pressure difference of
less than 3 mm Hg across the wall. Small numerical
oscillations of 5 mm Hg are noticeable mainly dur-
ing the filling phase.
The shape of the wall, as found in the FSI simula-
FIG. 8. Side view of the experimental model (left column), and X
tion, indeed resembles a paraboloidal shape during cross-sections of the FSI case (middle column) and base case
most of the cycle. However, because the walls are (right column) models at two phases of the pump cycle.

Artif Organs, Vol. 30, No. 7, 2006


NUMERICAL MODEL OF FLOW IN A SAC-TYPE VAD 535

attributed to the asymmetric orientation of the valves


in the experimental model.
The Z-velocity vectors fields at the central YZ
plane are shown for the FSI case and the base case
in the central and right columns, respectively. In both
cases, the velocity field and shape of the wall are
symmetric in the mid-XY-plane, due to symmetric
geometry and boundary conditions. In the FSI simu-
lations, some asymmetry relative to the mid-XZ and
YZ planes could be observed due to the pressure FIG. 10. Comparison of vector velocity field at filling and ejection
variations developed in the vicinity of the valves phases.
(Fig. 8), while in the base case, a fully symmetric
shape of the wall was imposed.
In the base case, the wall velocities are distributed errors and the assumption of a symmetric motion.
according to the imposed parabolic profile. In the FSI Larger discrepancies were obtained with the FSI cal-
case, the wall velocity is calculated from the transmu- culations, where the wall displacement was calcu-
ral pressure gradient and therefore, although the wall lated coupled to the simulation of the flow field. The
shape is approximately parabolic, the wall velocity largest deviation between the experimental and FSI
does not follow a parabolic profile. In both the filling calculations was obtained at the beginning of filling
and ejection phases, larger velocities were calculated (5% relative to the neutral position). These devia-
at the center of the membrane, while at the periph- tions are mainly attributable to differences in the
eral regions of the membrane, small-to-negative pressure distributions in the entrance and exit
velocities were found. In addition, velocity and regions of the main chamber because the valves were
pressure gradients were found to be smaller than in only crudely modeled. The mean deviation, however,
the base case. During the isovolumetric phases, the was less than 3%.
imposed wall velocity in the base case was zero, while
the internal velocity field was nonzero. In the FSI Flow patterns at the mid-XY plane
case, however, although there was no change of Velocity vectors at the mid-XY plane of the VAD
volume, some wall motion was observed. are shown in Fig. 10 for the three cases at ejection
The displacement of the central points of the mem- and filling. The velocity vector fields for the CDPIV
branes for the experimental measurements and both measurement and two simulations are shown for the
simulation models is plotted in Fig. 9. The wall dis- interrogation window used in the CDPIV measure-
placement, as measured from the experiments, was ments. In all three cases, similar global flow features
calculated assuming symmetry along the XY plane. were observed. During filling, two high-velocity jets
The wall displacement for the base case was obtained flowed into the chamber from the inflow valve. A
from a best-fit of a harmonic function to the experi- large rotating vortex formed and grew until it filled
mental results. Deviation between the measured and the entire VAD chamber during the isovolumetric
approximate displacement was relatively small (less phase and at most of the ejection. In each case, at the
than 3%), and was mainly a result of measurement end of the ejection phase, the vortex was washed out
through the outflow valve. No significant stagnant
1.5 regions could be identified throughout the entire car-
diac cycle, an indication of good washout properties
W all displacement (cm)

1.0
B in the present design.
0.5
Distribution of the Ux and Uy velocity components
0.0
at filling and ejection along the X centerline (as
Experiment
Base case A C defined in Fig. 5) is plotted in Fig. 11. A global agree-
-0.5 FSI case
D ment was obtained for all three cases, but differences
-1.0 were noticeable in local features. The largest devia-
0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0
t n = t/T
tions were found in the FSI simulations at peak ejec-
tion (phase C). The average difference among the
FIG. 9. Wall displacements of the chambers center point as three models, relative to the mean velocity, was less
measured experimentally (solid line), as calculated in the FSI than 26%. However, correlation among velocity pro-
simulation (gray solid line) and as approximated to a harmonic
curve for the base case (dotted line). The letters A, B, C and D files of the three cases was high. The average velocity
correspond to the four instants defined in Figure 3. correlation (the covariance of the data sets divided

Artif Organs, Vol. 30, No. 7, 2006


536 I. AVRAHAMI ET AL.

a by the product of their SD values) between the


Filling CDPIV and base case results was 0.92, and the veloc-
(Phase A)
2.0 ity correlation between the two numerical models
was 0.93.
0.0

2.0
Ux (cm/s)

DISCUSSION
4.0

6.0
CDPIV The objective of this study was to develop a
FSI case numerical model for simulating the flow inside a sac-
8.0 Base case
type pulsatile VAD. The main advantage of the base
10.0
1.0 0.8 0.6 0.4 0.2 0.0 0.2 0.4 0.6 0.8 1.0
case is that it requires less computational effort and
xn = x/R therefore, allows the use of finer mesh. This could
be useful for a comprehensive investigation of the
b
Filling
hemodynamic properties of the VAD chamber and
(Phase A) for optimization of VAD design. The basic model
10.0
includes four times as many nodes as the FSI model,
5.0 and yet the simulations for the FSI case required 10
0.0 times as much CPU time as the base case. However,
Uy (cm/s)

5.0
several simplifications were made in the base model.
CDPIV
Two questionable assumptions are the laminar flow
10.0
FSI case assumption and modeling of the inflow and outflow
15.0 Base case valves.
20.0 The assumption of laminar flow may be relevant
1.0 0.8 0.6 0.4 0.2 0.0 0.2 0.4 0.6 0.8 1.0
mainly because of the under-physiological pumping
xn = x/R
conditions used in this verification study. In the cases
c analyzed, the mean Reynolds number was 560, and
Ejection
(Phase C) the peak filling Reynolds number was less than 1600.
3.0 In such flow conditions, the effect of transitional flow
2.0 CDPIV
FSI case
in the inlet cannula during deceleration phase of fill-
1.0
Base case ing is considered to be negligible (9,20). However,
Ux (cm/s)

0.0
according to Konig et al. (16), sudden expansion of
1.0
the flow as it enters the main chamber might tend to
2.0
3.0
cause transient turbulence even at Reynolds num-
4.0
bers as low as 750. Therefore, assumption of laminar
1.0 0.8 0.6 0.4 0.2 0.0 0.2 0.4 0.6 0.8 1.0 flow may introduce some inaccuracies, especially
xn = x/R
downstream from the valves. When higher flow rates
d are of interest, low-Reynolds turbulence models
Ejection should be employed and a considerably finer mesh
(Phase C)
12.0
should be used (7). In this case, the simplifications of
10.0 the basic model that do not require FSI simulation
8.0 may be useful to make the problem tractable.
6.0
Another simplification in the numerical simula-
Uy (cm/s)

4.0
2.0 CDPIV tions concerns the flow across the valves. A detailed
0.0 FSI case
description of the flow across MHVs is beyond the
2.0 Base case
4.0 scope of this study. Incorporating a 3D detailed FSI
6.0 simulation of two passively moving valves in a model
8.0
1.0 0.8 0.6 0.4 0.2 0.0 0.2 0.4 0.6 0.8 1.0
of a sac-type VAD is beyond the capabilities of rou-
xn = x/R tine simulations. Therefore, the valves were modeled
mainly to provide a more realistic entrance flow into
FIG. 11. X and Y velocity components along the X centerline at the VADs chamber in order to assess their contribu-
peak filling (phase A; a, b) and peak ejection (phase C; c, d).
tion to the flow inside the VAD chamber. Two prom-
inent simplifications were made regarding the valves:
(i) the passively moving leaflets were modeled as
either fully open or fully closed; and (ii) a relatively

Artif Organs, Vol. 30, No. 7, 2006


NUMERICAL MODEL OF FLOW IN A SAC-TYPE VAD 537

coarse mesh was used in the vicinity of the valves velocity was determined indirectly by the specified
(Fig. 4b). The same order of mesh resolution in the external pressure and internal flow field, while in the
vicinity of an MHV (15 000 nodes) was used by Kiris base case, internal flow was affected by the specified
et al. (13), Konig et al. (16), and Shim et al. (17). In wall velocity, resulting in higher internal velocity gra-
order to simulate properly the detailed flow through dients, especially in the Z-direction (Fig. 8). Similar
MHV, a grid mesh of 5 106107 nodes would be observations were reported by Chandran et al. (5),
required (7,9,19). who found smaller gradients of pressure and velocity
In both numerical simulations and in the experi- inside a flexible ventricle than in a rigid ventricle with
mental results, similar wall motion and flow patterns comparable flow rate conditions.
were observed inside the chamber, including the Bearing in mind the complexity of the flow field
formation of a large vortex that filled up the entire inside the VAD chamber and the assumptions made,
chamber during the end of filling and the beginning fair agreement was found between the experimental
of ejection. Similar large vortexes were identified results and those of the two numerical models. Based
in other VAD chambers with discoid chamber and on the calculated results, it may be concluded that
parallel conduits (13,2123). It was shown that the differences in wall motion and Z-velocity field among
characteristics of this vortex strongly affected the the various models do not significantly affect the glo-
hemodynamics of the VAD (6,24,25). bal features of the flow inside the chamber. The main
Some disagreement was found in the fine details flow features were dominated by the inlet and outlet
of the flow field between the numerical simulations flows and the large vortex inside the chamber. Both
and experiments as a result of assumptions and sim- numerical models succeeded well in representing the
plifications made in the numerical models and due to detailed 3D time-dependent flow behavior inside the
limitations of the experimental system. Main differ- flexible VAD chamber. Notwithstanding its limita-
ences were found in the details of the inflow and tions, the simplified base case model was found to be
outflow jets across the valves. These differences are as accurate as the more complex FSI case for predict-
mainly attributable to the limitations of the CDPIV ing the flow inside the VAD chamber. Because the
system to measure velocity larger than 20 cm/s. required computational resources are an order of
According to the numerical simulations, the maximal magnitude less than for the fully coupled FSI simu-
velocity across the valves reached 38 cm/s at peak lation, it may be reasonable to use the approximation
ejection. Therefore, the CDPIV vector field in the of prespecified wall motion when the flow patterns
inflow and outflow regions (Fig. 10) should be inter- and velocities in the sac chamber are of interest.
preted with care. However, as long as the flow inside Based on the conclusions of this article, a larger study
the chamber is considered (Fig. 11), the CDPIV of the hemodynamics and design of a sac-type VAD
results compare reasonably well with the numerical is being conducted using the simplified model. In
results. future work, the simplified model will include turbu-
The differences between specified boundary con- lence model as well. The results of a comprehensive
ditions in the two numerical models led to some dif- investigation of chamber hemodynamics in different
ferences in wall motion. The wall motion in the FSI VAD designs, comparing MHV type, conduit orien-
case was highly sensitive to internal pressure varia- tation, and compression mechanism will be given
tions. Even small variations in pressure inside the elsewhere in future publications.
chamber (of the order of 1.75 mm Hg) resulted in
nonparabolic wall shapes. This sensitivity may indi- ACKNOWLEDGMENTS
cate that the details of the wall motion might change
with even minor changes of the flow conditions. Fur- This research was conducted in collaboration with
thermore, as shown by the experimental model, the Prof. Klaus Affeld, Humboldt-University, Berlin. It
wall motion was found to be very sensitive to in- was partially supported by a grant from G.I.F., the
herent asymmetry in the model of the valve ori- German-Israeli Foundation for Scientific Research
entation as well. A similar sensitivity to pressure and Development and by the Fritz Brann Fellowship,
distribution and flow conditions in a sac-type cham- the Israel Science Foundation, the Joseph Drown
ber was also reported by Clark et al. (21), who Foundation, and the Berman Trust.
suggested that even device orientation with respect
to gravity or movement of the patient might have REFERENCES
noticeable influence on the motion of the wall.
1. Rose EA, Gelijns AC, Moskowitz AJ, et al. Randomized eval-
The specified boundary condition also affected uation of mechanical assistance for the treatment of conges-
internal velocity distribution. In the FSI case, the wall tive heart failure (REMATCH) Study Group. Long-term

Artif Organs, Vol. 30, No. 7, 2006


538 I. AVRAHAMI ET AL.

mechanical left ventricular assistance for end-stage heart 13. Kiris C, Kwak D, Rogers S, Chang ID. Computational
failure. N Engl J Med 2001;345:143543. approach for probing the flow through artificial heart devices.
2. Mussivand T, Day KD, Naber BC. Fluid dynamic optimiza- J Biomech Eng 1997;19:45260.
tion of a ventricular assist device using particle image veloci- 14. Chesler NC, Kamm RD. Performance analysis of a cardiac
metry. ASAIO J 1999;45:2531. assist device in counterpulsation. J Biomech Eng 1998;120:
3. Hochareon P, Manning KB, Fontaine AA, Tarbell JM, 43745.
Deutsch S. Wall shear-rate estimation within the 50cc Penn 15. Konig CS, Clark C. Flow mixing and fluid residence times in
State artificial heart using particle image velocimetry. J Bio- a model of a ventricular assist device. Med Eng Phys 2001;23:
mech Eng-T ASME 2004;126:4307. 99110.
4. Affeld K, Ziemann A, Goubergrits L. Technical obstacles on 16. Konig CS, Clark C, Mokhtarzadeh-Dehghan MR. Investiga-
the road towards a permanent left ventricular assist device. tion of unsteady flow in a model of a ventricular assist device
Heart and Vessels 1997:2830. by numerical modeling and comparison with experiment. Med
5. Chandran KB, Lee CS, Shipkowitz T, Chen LD, Yu LS, Eng Phys 1999;21:5364.
Wurzel D. In vitro hemodynamic analysis of flexible artificial 17. Shim EB, Yeo JY, Ko HJ, et al. Numerical analysis of the
ventricles. Artif Organs 1991;15:4206. three-dimensional blood flow in the Korean artificial heart.
6. Rose ML, Mackay TG, Wheatley DJ. Evaluation of four Artif Organs 2003;27:4960.
blood pump geometries: fluorescent particle flow visualisation 18. Grad Y, Einav S. Spectral and instantaneous flow character-
technique. Med Eng Phys 2000;22:20114. istics of vascular junctions using continuous DPIV. In: 9th
7. Bluestein D, Li Y, Krukenkamp I. Free emboli formation in International Symposium on Applications of Laser Techniques
the wake of bi-leaflet mechanical heart valves and the to Fluid Mechanics, Lisbon Portugal, 1998.
effects of implantation techniques. J Biomech 2002;35:1533 19. Rosenfeld M, Avrahami I, Einav S. Unsteady effects on the
40. flow across tilting disk valves. J Biomech Eng 2002;124:219.
8. Van de Vosse FN, de Hart J, van Oijen C, et al. Finite-element- 20. Bluestein D, Einav S. Techniques in the analysis of stability of
based computational methods for cardiovascular fluid- pulsatile flow through heart valves. In: Leondes CT, ed. Bio-
structure interaction. J Eng Math 2003;47:33568. mechanics Systems Techniques and Applications: Cardiovascu-
9. Ge L, Jones S, Sotiropoulos F, Healy T, Yoganathan AP. lar Techniques, Boca Raton, FL: CRC Press LLC, 2001;3:439.
Numerical simulation of flow in mechanical heart valves: grid 21. Clark C, Jin W, Glaser A. The fluid mechanics of a sac-type
resolution and the assumption of flow symmetry. J Biomech ventricular assist device. Int J Artif Organs 1990;13:81422.
Eng-T ASME 2003;125:70918. 22. Ryu GH, Kim J, Chang JK, et al. The fluid dynamic effect on
10. Lemmon JD, Yoganathan AP. Computational modeling of protein adsorption in left ventricular assist devices. ASAIO J
left heart diastolic function: examination of ventricular dys- 1993;39:M3326.
function. J Biomech Eng 2000;122:297303. 23. Baldwin JT, Deutsch S, Geselowitz DB, Tarbell JM. LDA
11. Saber NR, Wood NB, Gosman AD, et al. Progress towards measurements of mean velocity and Reynolds stress fields
patient-specific computational flow modeling of the left within an artificial heart ventricle. J Biomech Eng 1994;116:
heart via combination of magnetic resonance imaging with 190200.
computational fluid dynamics. Ann Biomed Eng 2003;31:42 24. Woodard JC, Shaffer FD, Schaub RD, Lund LW, Borovetz
52. HS. Optimal management of a ventricular assist system. Con-
12. McQueen DM, Peskin CS. Heart simulation by an immersed tribution of flow visualization studies. ASAIO J 1992;38:
boundary method with formal second-order accuracy and M2169.
reduced numerical viscosity. Proceedings of the 20th 25. Slater JP, Rose EA, Levin HR, et al. Low thromboembolic
International Conference on Theoretical and Applied risk without anticoagulation using advanced-design left ven-
Mechanics (ICTAM 2000), August 27September 2, Chicago, tricular assist devices. Ann Thorac Surg 1996;62:13217; dis-
IL, U.S.A. cussion 1328.

Artif Organs, Vol. 30, No. 7, 2006

Das könnte Ihnen auch gefallen