Message on
Meaning Causes Action
LCD display
Door open ! Hall affect sensor ! Check set installation
alarm detects that the door ! Check the presence of the
is open door magnet
! Calibration of the door
sensor
! Replace the door sensor
Clamp ! The Optima does ! Check set installation
position not detect the ! Check the securing clamp
alarm presence of the set micro switch
securing clamp ! Replace it if necessary
Set ! Wrong position of ! Check set installation
insertion the set ! Check the presence of the
alarm door magnet
! Calibration of the door
sensor
! Replace the door sensor
OCS test ! Occlusivity test ! Check set installation
Alarm not successful. The ! Check the door and carry on
pumping mechanism a calibration if necessary
does not succeed to ! Verify the good condition of
increase the the membrane. Replace it if
pressure in the IV necessary; refer to the
set during the OCS maintenance protocol : PM 0094
test
Air bubble ! Air volume ! Prime the line to remove the
alarm detected is higher air
Than the limit ! Check set installation on the
programmed air detector Verify the air
! Set is not adjustment of the air size
properly inserted in programmed
the air detector ! Replace the air detector or
air bubble board
Troubleshooting
optima05.1_001a_en.fm 45
Message on
Meaning Causes Action
LCD display
Drop alarm ! The container is ! Verify the volume remaining
(If drop empty on the container
detector is ! The rolling clamp ! Verify if the roller clamp is
present) is closed open
! Over-flow , under ! Verify if the air inlet cap if
flow, free flow open when the pump is used
! The position of with a bottle
the drip sensor ! Verify the positioning of the
! Drop chamber is drop detector
full ! Verify the volume of the liquid
! Drop detector is in the drop chamber
disconnected while ! Verify the IV set position
the pump is ! Verify the drip sensor is well
configured to be connected
used with ! Verify the connector if it's
! The drip sensor damaged
is connected and ! Replace the drip sensor
disconnected
Free flow ! Over-flow ! Check set installation
alarm ! Set not properly ! Check drip sensor
(in case of positioned positioning
use with
drip
sensor)
46 optima05.1_001a_en.fm
Message on
Meaning Causes Action
LCD display
Air alarm ! The IV set is ! Remove the IV set and re-
present on the air install it
bubble detector but ! Carry on a calibration of the
not on the pumping door
mechanism
! The pump runs
over 24 hours, no air
bubble have
detected by the
sensor
Battery ! Low battery ! Connect the pump to the
alarm voltage mains power
! Check the charging voltage,
pump connected to the mains.
! Replace the battery
Preventive maintenance Preventive ! Preventive ! Proceed to the complete
24/10/2003 passed
maintenan maintenance date or check up of the pump and
ce time is passed update the maintenance date or
requested time ( test 1)
Mains LED's does ! Mains fuses out of ! Replace Check the mains fuses
not switch ON order ! Replace the power supply board
! Power supply out of ! If the problem remains, replace the
order CPU board
optima05.1_001a_en.fm 47
Problems Causes Action
Battery alarm while ! Battery is totally ! Check the charging voltage
it has been discharged ! Replace the battery
completely charged
Battery alarm ! Power supply or CPU ! Check the charging voltage
whereas the boards are faulty ! Check the power supply mains
Optima is fuses
connected to the
! Replace the power supply if
mains
necessary.
! If the problem remains, replace the
CPU board and carry on a complete
configuration
The module MVP ! Keyboard is defective ! Check the keyboard ( test 3)
turns ON or turns ! Replace the keyboard if necessary
OFF alone ! CPU board is ! Check that there was no liquid
defective infiltration on the CPU board
! Replace the CPU board and carry
on a complete configuration of the
device
Some keys from ! Keyboard is defective ! Check the keyboard (test 3)
keyboard out of ! Replace the keyboard if necessary
order
After a fall ! Mechanical elements ! Check that the mechanical
damaged assembly and the cases are intact
Pump noisy ! Power supply and ! Fix the wires as it shown on the
CPU wires are closed to technical manual
the motor (earth wire
also)
Troubleshooting
48 optima05.1_001a_en.fm
Troubleshooting
Er31 OCS DC motor failure ! Check the motor clamp for OCS and
rubber O ring
! Put some grease on the clamp finger
threaded screw. Replace the rubber
O ring if necessary (p/n: 174256)
Troubleshooting
optima05.2_002c_en.fm 49
Error codes Meaning Action
Er14 Motor period calculation ! Reconfigure the device with VP
anomaly Control software
! Replace the CPU board and carry
on a complete configuration of the
device
Er16 ! Date / time failure ! Check the date and the time clock
! Replace the CPU board and Carry
on a complete configuration of the
Troubleshooting
50 optima05.2_002c_en.fm
Error codes Meaning Action
Er17 ! Keyboard matrices default ! Replace the keyboard
! Check that there is no liquid
presence inside the device which could
damage keyboard and electronics
boards
Er47 ! OFF key out of order ! Replace the CPU board and carry
(no way to switch OFF the on a complete configuration of the
device) device via VP Control software
Troubleshooting
optima05.2_002c_en.fm 51
Troubleshooting
52 optima05.2_002c_en.fm