1. OBJECTIVE
Objective of this SOP is to describe the procedure for operation of Universal Incubator
(Memmert).
2. SCOPE
This SOP describes the execution of operation of universal incubator in which
environmental conditions can be controlled for growing microbial cultures; this SOP is
applicable to microbiology laboratory.
3. GLOSSARY
Temp. = Temperature
SS = Stainless Steel
C = Degree Centigrade
SOP = Standard Operating Procedure
All or part of the information contained in this document should be treated as the confidential property of Ferozsons Laboratories
Limited. It cannot be published or divulged for whatever purpose to unauthorized persons, or 3 rd parties, in any form without the prior
written permission from DQO Ferozsons Laboratories Limited.
COPY NO.
FEROZSONS LABORATORIES LIMITED
Standard Operating Procedure
4. RESPONSIBILITIES
General Manager:
- To provide adequate resources for the implementation of this SOP as defined.
A. M. Microbiology:
- To follow the instructions laid down by this SOP.
- To keep the relevant SOPs valid and updated.
Microbiologist:
- To interface this SOP to the actual practices of flow patterns.
- To ensure that the SOP is updated & reflects the actual execution.
- To review the SOP in case of any change or whenever it is due for revision.
All or part of the information contained in this document should be treated as the confidential property of Ferozsons Laboratories
Limited. It cannot be published or divulged for whatever purpose to unauthorized persons, or 3 rd parties, in any form without the prior
written permission from DQO Ferozsons Laboratories Limited.
COPY NO.
FEROZSONS LABORATORIES LIMITED
Standard Operating Procedure
2 9
3
Universal Incubator
4 Components
All or part of the information contained in this document should be treated as the confidential property of Ferozsons Laboratories
Limited. It cannot be published or divulged for whatever purpose to unauthorized persons, or 3 rd parties, in any form without the prior
written permission from DQO Ferozsons Laboratories Limited.
COPY NO.
FEROZSONS LABORATORIES LIMITED
Standard Operating Procedure
5.2.2 Lay the power cable, so that, it doesnt come into contact with any hot part.
5.2.3 Plug the provided power cable into the rear of the appliance and connect it.
5.2.4 Turn the power switch on by pressing the On/Off switch on the front of the sterilizer, see
figure 2.
5.2.5 First start-up display appears on the screen, set the desired temperature by pressing the
activation key on the left or right of the respective display, figure 3.
All or part of the information contained in this document should be treated as the confidential property of Ferozsons Laboratories
Limited. It cannot be published or divulged for whatever purpose to unauthorized persons, or 3 rd parties, in any form without the prior
written permission from DQO Ferozsons Laboratories Limited.
COPY NO.
FEROZSONS LABORATORIES LIMITED
Standard Operating Procedure
All or part of the information contained in this document should be treated as the confidential property of Ferozsons Laboratories
Limited. It cannot be published or divulged for whatever purpose to unauthorized persons, or 3 rd parties, in any form without the prior
written permission from DQO Ferozsons Laboratories Limited.
COPY NO.
FEROZSONS LABORATORIES LIMITED
Standard Operating Procedure
All or part of the information contained in this document should be treated as the confidential property of Ferozsons Laboratories
Limited. It cannot be published or divulged for whatever purpose to unauthorized persons, or 3 rd parties, in any form without the prior
written permission from DQO Ferozsons Laboratories Limited.
COPY NO.
FEROZSONS LABORATORIES LIMITED
Standard Operating Procedure
5.2.9 Open the door completely by pulling the door handle from right side to the left, figure 7.
All or part of the information contained in this document should be treated as the confidential property of Ferozsons Laboratories
Limited. It cannot be published or divulged for whatever purpose to unauthorized persons, or 3 rd parties, in any form without the prior
written permission from DQO Ferozsons Laboratories Limited.
COPY NO.
FEROZSONS LABORATORIES LIMITED
Standard Operating Procedure
5.2.11 Do not load the chamber too tightly, so that the air circulates properly inside the working
chamber, also do not place any material on the chamber floor or touching the chamber
side walls, figure 8.
Note: If the door is open while the sterilizer is in operation, the incubator may overheat
and pose a fire hazard. Do not leave the door open during operation.
Note: Do not load such materials which produce poisonous explosive vapors or gases, if
there is any doubt then they must not be loaded into to the incubator.
All or part of the information contained in this document should be treated as the confidential property of Ferozsons Laboratories
Limited. It cannot be published or divulged for whatever purpose to unauthorized persons, or 3 rd parties, in any form without the prior
written permission from DQO Ferozsons Laboratories Limited.
COPY NO.
FEROZSONS LABORATORIES LIMITED
Standard Operating Procedure
All or part of the information contained in this document should be treated as the confidential property of Ferozsons Laboratories
Limited. It cannot be published or divulged for whatever purpose to unauthorized persons, or 3 rd parties, in any form without the prior
written permission from DQO Ferozsons Laboratories Limited.
COPY NO.
FEROZSONS LABORATORIES LIMITED
Standard Operating Procedure
5.3.2 A temperature sensor is installed near to ventilation flap at the top of the chamber, be
careful not to damage this sensor during cleaning the chamber, if sensor comes loose
from the fixing clip, reattach it.
5.3.3 Gently wipe stains from the body with a soft cloth, to remove stubborn stains, wring a
cloth moistened with neutral detergent diluted with water, and wipe off the stains with it.
5.3.4 Do not use benzene or thinner or any volatile substances to clean the body as these
substances may deteriorate the body or strip the paint, and also wipe off any moisture
with a dry cloth.
5.3.5 If rust spots appear on the surface of the working chamber due to impurities, the affected
area must be immediately cleaned and polished.
5.3.6 Grease the moving parts of the door (hinges and lock) with thin silicone grease and
check that the hinge screws are not loose, once a year.
5.4.2.2 Take care when operating chamber as electric shock may be hit when opened.
All or part of the information contained in this document should be treated as the confidential property of Ferozsons Laboratories
Limited. It cannot be published or divulged for whatever purpose to unauthorized persons, or 3 rd parties, in any form without the prior
written permission from DQO Ferozsons Laboratories Limited.
COPY NO.
FEROZSONS LABORATORIES LIMITED
Standard Operating Procedure
5.4.2.4 Do not pile specimen on top of one another, sufficient air should be allowed to circulate
round the material so that heat does not build-up and the materials combust.
5.4.2.5 Heat Labile plastic objects do not be placed inside the incubator.
5.5 DOCUMENTATION
Sterilization cycles (time) & cleaning entries are written in the following LOGs:
Retention Period:
The original reports will be retained in Microbiology Section for Three (3) years.
6. TRAINING
Training of the concerned SOP shall be given to the personnel who will be exercising the
SOP, and will be documented on the attached appendix-II.
7. APPENDICES
1) LOG Sheet No. FZ-202 (Temperature Monitoring Cycle = 30 35 C)
2) LOG Sheet No. FZ-203 (Temperature Monitoring Cycle = 20 25 C)
3) Form No. FZ-152 (Cleaning Record)
3) Training Record
8. REFERENCE
British Pharmacopoeia 2017.
Memmert Manufacturers Manual
United States Pharmacopoeia USP 40, NF 35 2017.
District Laboratory Manual Monica Cheesebrough Vol. 1.
All or part of the information contained in this document should be treated as the confidential property of Ferozsons Laboratories
Limited. It cannot be published or divulged for whatever purpose to unauthorized persons, or 3 rd parties, in any form without the prior
written permission from DQO Ferozsons Laboratories Limited.
COPY NO.
FEROZSONS LABORATORIES LIMITED
Standard Operating Procedure
All or part of the information contained in this document should be treated as the confidential property of Ferozsons Laboratories
Limited. It cannot be published or divulged for whatever purpose to unauthorized persons, or 3 rd parties, in any form without the prior
written permission from DQO Ferozsons Laboratories Limited.
COPY NO.