DATA SHEET

1720E Low Range

Turbidity
Turbidimeter

Features and Benefits


USEPA Reporting
The 1720E Low Range Turbidimeter applies the instrument
design and meets performance criteria established by the
U.S. Environmental Protection Agency (USEPA) in Method
180.1, making it suitable for regulatory reporting.

Accuracy
Continuously flowing sample flows through the patented*
bubble removal system, which vents entrained air from
the sample stream and eliminates the most significant
interference in low level turbidity measurement. The
1720E Turbidimeter is not affected by variations in flow
and pressure.

Nephelometric Measurement
Incandescent light directed from the sensor head assembly
down into the turbidimeter body is scattered by suspended DW
particles in the sample. The sensor's submerged photocell
detects light scattered at 90 from the incident beam.
Sample enters the center column of the turbidimeter, rises
into the measuring chamber and spills over the weir into the
drain port. This configuration results in an optical flat surface
free of turbulence. PW
Simplicity
A simplified two-module design includes the sensor and IW
the controller interface. The controller accepts two turbidity
sensorsadding a second 1720E sensor makes a system
with two complete turbidimeters. Connections are simple The Model 1720E Low Range Turbidimeter is the newest is
plug & play. a long line of successful Hach turbidimetersfrom the
unsurpassed world leader in turbidity measurement.
Data Collection and Display
The 1720E Turbidimeter uses the sc100 Controller to
receive data from up to two sensors. A built-in data Fast Calibration and Verification
logger collects turbidity measurement at user selectable
intervals (1-15 minutes), along with calibration and Calibration and verification can be performed without loss of
verification points, alarm history, and instrument setup sample flow using the ICE-PIC Calibration/Verification
changes for 6 months. Communications using Module. One-point calibration with prepared StablCal
MODBUS/RS485, MODBUS/RS232, LonWorks Stabilized Formazin Solution eliminates the errors of
protocols or the wireless IR port are available. The sc100 user-prepared formazin suspension dilution. Features of
Controller is also compatible with AquaTrend Networks. the ICE-PIC Module include:
Local display, recall, graphing, and trending in CSV format Calibrate or verify the performance of each sensor
make chart recorders redundant. in less than one minute

Experience Factory calibrated and provided with a certificate


of accuracy.
The 1720E Turbidimeter reflects 45 years of Hach leadership
in turbidity measurement science. Hach has the largest Cost effective, one-time investment. No consumables
turbidimeter installation base in the world. And, Hach offers are needed.
a two-year warranty on the 1720E.
Small, lightweight design can be used for spot verification
in the facility.

*U.S. patent 5,831,727 Available in 20 and 1.0 NTU.

DW = drinking water WW = wastewater municipal PW = pure water / power


IW = industrial water E = environmental C = collections FB = food and beverage
2

Specifications*
Range Operating Humidity Compliance
0.001-100 Nephelometric Turbidity 5 to 95% non-condensing Standard Methods 2130B, USEPA 180.1,
Units (NTU) Hach Method 8195
Storage Temperature
Accuracy -20 to 60C (-4 to 140F) Certifications
(Defined according to ISO 15839.) Safety:
Power Requirements
2% of reading or 0.015 NTU Listed by ETL to UL 61010A-1: Certified
100-230 Vac, 50/60 Hz,
(whichever is greater) from 0 to 40 NTU; by ETL to CSA C22.2 No. 1010.1: CE
auto selecting; 40 VA
5% of reading from 40 to 100 NTU certified by Hach Company to EN
Sample Inlet Fitting 61010-1
Displayed Resolution
1/4" NPT female, 1/4" compression Immunity:
0.0001 NTU up to 9.9999 NTU;
fitting (provided) CE certified by Hach Company to
0.001 NTU from 10.000 to 99.999 NTU
Drain Fitting EN61326 (industrial levels)
Repeatability
1/2" NPT female, 1/2" hose barb (provided) Emissions:
(Defined according to ISO 15839.)
Better than 1.0% of reading or Recorder Outputs Class A: EN 61326, CISPR 11, FCC
0.002 NTU, whichever is greater Two selectable for 0-20 mA or 4-20 mA; Part 15, Canadian Interference-Causing
output span programmable over any Equipment Regulation ICES-003
Response Time
portion of the 0-100 NTU range; Mounting
Initial response in 1 minute, 15 seconds
built into the sc100 Controller Turbidimeter body and head assembly:
for a full-scale step change
Alarms wall and floor stand
Signal Average Time
Three set-point alarms, each equipped sc100 Controller:
User selectable from 6, 30, 60, 90
with an SPDT relay with unpowered wall, pole, panel, and floor stand
seconds; default 30 seconds
contacts rated 5A resistive load at
Dimensions
Sample Temperature 230 Vac; built into the sc100 Controller
Turbidimeter body and cap:
0 to 50C (32 to 122F)
Enclosure 25.4 x 30.5 x 40.6 cm (10 x 12 x 16 in.)
Sample Flow Required NEMA-4X (indoor)/IP66 Controller sc100 Controller:
200 to 750 mL/minute 14.4 X 14.4 X 15.0 cm
Digital Communication
(3.1 to 11.9 gal/hour) (5.67 x 5.67 x 5.91 in.)
Network card compatible;
Operating Temperature MODBUS/RS485, MODBUS/RS232, Shipping Weight
Single sensor system: LonWorks protocol (optional) 1720E Turbidimeter and
0 to 50C (32 to 122F) sc100 Controller:
Wireless Communication
Two sensor system: 6.12 kg (13.5 lbs.)
IR Port on the sc100 Controller to
0 to 40C (32 to 104F)
download into a handheld Personal 1720E Turbidimeter:
Digital Assistant (PDA), or laptop 4.54 kg (10 lbs.)
computer via MODBUS
*Specifications subject to change without notice.

Engineering Specifications
1. The turbidimeter shall be a 9. Optical components shall be mounted capability through the IR Port
microprocessor-based, continuous- in a sealed head assembly that can located on the interface unit to
reading, on-line nephelometric be removed for calibration/ service download and print realtime turbidity
instrument without disturbing sample flow. data, calibration history, and current
2. The turbidity monitoring system shall 10. The turbidimeter body shall be set points in a CSV format.
include one or two turbidimeter(s) constructed of corrosion-resistant 16. The Interface unit shall allow
and one interface unit. polystyrene. operators to control sensor and
3. The turbidimeter shall measure turbidity 11. An internal bubble removal system interface functions with menu-driven
in the range of 0.001-100 NTU shall be included to vent entrained software and shall provide data
4. Accuracy shall be 2% of reading or air from the sample stream. logging of measurement data from
0.015 NTU (whichever is greater) 12. Calibration of the turbidimeter up to two turbidimeters for 15
from 0 to 40 NTU; 5% of reading shall be either formazin-based minutes, 1 hour, 24 hours, 30 days,
from 40 to 100 NTU (20 or 1 NTU) or instrument or 180 days.
5. Displayed resolution shall be 0.0001 comparison-based calibration method. 17. The interface unit shall be able to
NTU from 0 to 9.999 NTU and 0.001 13. User selectable signal averaging, transfer data to a computer or
NTU from 10.000 to 9.999 NTU. bubble removal, alarm and recorder printer via direct MODBUS
output hold, and self-test communications or directly into a
6. Repeatability shall be better than
diagnostics shall be provided. Personal Digital Assistant (PDA) via
1.0% of reading or 0.002 NTU
a wireless IR Port.
(whichever is greater). 14. Connections between the
turbidimeter(s) and the controller 18. The interface unit shall have a built-
7. The turbidimeter shall meet all
shall be plug and play. in data logger with the capacity to
design and performance criteria
store data on 15-minute intervals for
specified by USEPA method 180.1. 15. All turbidimeters installed on a
up to 6 months with two sensors per
8. Light shall be directed through the network shall have the option for
controller.
surface of the sample and the MODBUS/RS232, MODBUS/RS485,
LonWorks serial input/output 19. The interface unit shall include two
detector shall be immersed in the
capability for two-way analog outputs and 3 unpowered
sample, eliminating glass windows
communication to a computer or a SPDT alarm contacts.
and flow cells.
have a wireless downloading
3

Engineering Specifications continued


20. The interface unit shall be housed in 23. All system components shall be ETL 25. All system components shall meet
a NEMA-4X (indoor) industrial listed to UL 61010A-1, certified to FCC Part 15 for North America and
metal/plastic enclosure. CSA C22.2 No. 1010.1, and CE Canadian Interference-Causing
21. The DC power supply shall be certified by manufacturer to EN Equipment Regulation ICES-003,
housed in the interface unit 61010-1. and CISPR 11 Class A levels for rest
22. The DC power supply shall 24. All system components shall be CE of the world.
automatically accept input in the certified by the manufacturer to EN 26. The turbidimeter shall be Hach
range of 100 to 230 Vac, 50/60 Hz. 61326 (industrial levels) for immunity Company Model 1720E Low Range
and emissions, Class A. Turbidimeter with sc100 Controller.

Dimensions
The sc100 controller unit can be installed on a pole, wall, panel or a floor stand. The 1720E turbidimeter can be installed on a
wall or a floor stand. No tools are needed to connect the controller unit to the turbidimeter. The distance between the two units
can be a maximum of 9.62 m (31.6 ft) with the use of an extension cable.
sc100 Controller
144.02 mm (5.67 inches)
144.0 mm (5.67 inches) 150.0 mm (5.91 inches) 72.01 mm
(2.84 inches)
40.14 mm
(1.58 inches)

(2.84 inches)
72.01 mm
sc100
144.0 mm (5.67 inches)

144.02 mm (5.67 inches)


80.27 mm
(3.16 inches)

40.14 mm (1.58 inches)


80.27 mm (3.16 inches) 4X
1720E Turbidimeter M6 x 1.0

312.48 mm (12.30 inches)


Allow at least
273.05 mm (10.75 inches) 254 mm (10.00 inches)
above the instrument 238.35 mm (9.38 inches)
19.72 mm 157.00 mm for removal of
the head assembly. 127.8 mm
(0.78 inches) (6.18 inches) (5.0 inches)

POWER STATUS

2 Slots for
0.25" fastener

384.30 mm
Sample Inlet (15.13 inches)

236.22 mm
(9.30 inches)

Sample Drain

Service Drain
1.00 I.D.
172.4 mm (6.79 inches) 164.18 mm (6.46 inches)
4

At Hach, its about learning


Ordering Information from our customers and
providing the right answers.
1720E Turbidimeter Calibration Supplies Its more than ensuring the
quality of waterits about
60101-00 1720E Turbidimeter with ICE-PIC Calibration/Verification ensuring the quality of life.
sc100 Controller Module / 1720E:
60101-01 1720E Turbidimeter, sensor only When it comes to the things
52250-00 20 NTU Module that touch our lives...
52215-00 1 NTU Module
1720E with DigitalDirect Keep it pure.
Communications StablCal Comparative Calibration Standards
26601-53 20.0 NTU, 1 L each Make it simple.
60101-02 1720E/sc100 with (Calibration Cylinder,
MODBUS/RS485 output P/N 44153-00, must be Be right.
60101-03 1720E/sc100 with ordered separately.)
MODBUS/RS232 output
60101-04 1720E/sc100 with StablCal Verification Standards
LonWorks output 26979-53 0.3 NTU, 1 L each For current price information,
26980-53 0.5 NTU, 1 L each
Cables 27233-53 0.1 NTU, 1 L each technical support, and ordering
26598-53 1.0 NTU, 1 L each assistance, contact the Hach
57960-00 Extension Cable, 7.7 m (25 ft.) 27463-53 40.0 NTU, 1 L each office or distributor serving
46306-00 Power Cord w/ strain relief, your area.
125 Vac Formazin Calibration Standards
46308-00 Power Cord w/ strain relief, 44156-00 Formazin Calibration Kit for
230 Vac, European-style plug user-prepared calibration In the United States, contact:
Note: Power cables must be ordered (includes 500 mL of 4000
separately. NTU Formazin, TenSette HACH COMPANY World Headquarters
Pipet, and calibration P.O. Box 389
Accessories cylinder) Loveland, Colorado 80539-0389
2461-49 Formazin Primary Standard, U.S.A.
57432-00 Floor Stand 4000 NTU, 500 mL Telephone: 800-227-4224
(replacement for P/N 44156-00)
44153-00 Calibration Cylinder, 1 L Fax: 970-669-2932
E-mail: orders@hach.com
www.hach.com
To complete your turbidity
measurement system, choose U.S. exporters and customers in Canada,
Latin America, sub-Saharan Africa, Asia,
the sc100 or the sc1000 controller and Australia/New Zealand, contact:
Model sc100 Controller HACH COMPANY World Headquarters
(see Lit. #2463) P.O. Box 389
Loveland, Colorado 80539-0389
LXV401.52.00002 sc100 Controller Standard U.S.A.
LXV401.52.01002 sc100 Controller with RS-232 MODBUS Telephone: 970-669-3050
LXV401.52.02002 sc100 Controller with RS-485 MODBUS
Fax: 970-461-3939
E-mail: intl@hach.com
www.hach.com

Model sc1000 Controller In Europe, the Middle East, and


(see Lit. #2403) Mediterranean Africa, contact:

LXV402.99.00002 sc1000 Display Module HACH + LANGE Europe


LXV400.99.1R572 sc1000 Probe Module, 4 sensors, Dr. Bruno Lange GmbH & Co. KG
4 mA Out, 4 mA In, 4 Relays, 110-230V Willsttterstrae 11
LXV400.99.1B572 sc1000 Probe Module, 4 sensors, D-40549 Dsseldorf
4 mA Out, 4 mA In, 4 Relays, GERMANY
RS-485 (MODBUS), 110-230V Tel: +49 (0) 211 5288-0
LXV400.99.1F572 sc1000 Probe Module, 4 sensors, Fax: +49 (0) 211 5288-143
4 mA Out, 4 mA In, 4 Relays, E-mail: info@hach-lange.de
PROFIBUS DP, 110-230V
www.hach-lange.com
LXV400.99.1R582 sc1000 Probe Module, 6 sensors,
4 mA Out, 4 mA In, 4 Relays, 110-230V
Lit. No. 2457
B55 Printed in U.S.A.
Hach Company, 2005. All rights reserved.